High-Field Surface-Coil MR Imaging of Localized Anatomy

Similar documents
1 Introduction. 2 The basic principles of NMR

HETERONUCLEAR IMAGING. Topics to be Discussed:

Cardiac MR. Dr John Ridgway. Leeds Teaching Hospitals NHS Trust, UK

MR Advance Techniques. Flow Phenomena. Class II

MR Basics: Module 8 Image Quality

The SENSE Ghost: Field-of-View Restrictions for SENSE Imaging

Simultaneous Multi-Slice (Slice Accelerated) Diffusion EPI

Magnetic Resonance Imaging Principles, Methods, and Techniques

H 2 O and fat imaging

Magnetic Resonance Imaging

(N)MR Imaging. Lab Course Script. FMP PhD Autumn School. Location: C81, MRI Lab B0.03 (basement) Instructor: Leif Schröder. Date: November 3rd, 2010

Pulse Sequence Design and Image Procedures

Experience in implementing continuous arterial spin labeling on a commercial MR scanner

Advanced MSK MRI Protocols at 3.0T. Garry E. Gold, M.D. Associate Professor Department of Radiology Stanford University

MRI SYSTEM COMPONENTS Module One

Image Quality/Artifacts Frequency (MHz)

M R I Physics Course. Jerry Allison Ph.D., Chris Wright B.S., Tom Lavin B.S., Nathan Yanasak Ph.D. Department of Radiology Medical College of Georgia

MAGNETIC RESONANCE IMAGING

MRI Metal Artifact Reduction

Downloaded from by on 02/07/18 from IP address Copyright ARRS. For personal use only; all rights reserved

MARP. MR Accreditation Program Quality Control Beyond Just the Scans and Measurements July 2005

High Field MRI: Technology, Applications, Safety, and Limitations

Works-in-Progress package Version 1.0. For the SIEMENS Magnetom. Installation and User s Guide NUMARIS/4VA21B. January 22, 2003

BOLD fmri: signal source, data acquisition, and interpretation

1.5T HIGH FIELD SMALL ANIMAL MRI

Evaluation of MWT Materials Accusorb MRI Shield. Dr. E. Kanal, Department of Radiology, UPMC. Wednesday, April 21, 2010

Magnetic Resonance Research Facility (MRRF) Resources

ISSN X CODEN (USA): PCHHAX. The role of dual spin echo in increasing resolution in diffusion weighted imaging of brain

COMMUNICATIONS Volume-Selective Multipulse Spin-Echo Spectroscopy

MRI Systems and Coil Technology

Analysis of magnetic and electromagnetic field emissions produced by a MRI device

Anatomy and Metabolism of the Normal Human Brain Studied by Magnetic Resonance at 1.5 Tesla

Pulse Sequence Design Made Easier

Noninvasive Blood Flow Mapping with Arterial Spin Labeling (ASL) Paul Kyu Han and Sung-Hong Park

Spatial Localization in NMR Spectroscopy in Vivo

Supplementary Figure 1. Scanning Electron Microscopy images of the pristine electrodes. (a) negative electrode and (b) positive electrode.

TITLE: Prostate Cancer Detection Using High-Spatial Resolution MRI at 7.0 Tesla: Correlation with Histopathologic Findings at Radical Prostatectomy

Module 2. Artefacts and Imaging Optimisation for single shot methods. Content: Introduction. Phase error. Phase bandwidth. Chemical shift review

2014 M.S. Cohen all rights reserved

2 Hardware for Magnetic Resonance Imaging

Nuove tecnologie per ecografia ad ultrasuoni: da 2D a 4D

Background (~EE369B)

The ACR magnetic resonance accreditation phantom (ACR MRAP) has been designed to examine a broad range of instrument parameters.

Weber State University Radiologic Technology 4603

TimTX TrueShape. The parallel transmit architecture of the future. Answers for life.

SIGNA Explorer Lift revives our MR

Magnetic Resonance Imaging and Radio Frequency. Part 1. Produced on behalf of Mid Sussex Amateur Radio Society by M5BTB

Chapter 11 Coherence Editing: Pulse-field Gradients and Phase Cycling

Introduction. Parametric Imaging. The Ultrasound Research Interface: A New Tool for Biomedical Investigations

Applications Guide. Spectral Editing with SVS. (Works-in-Progress) MAGNETOM TaTs and Verio Systems (3T)

Supplementary Material

Siemens AG, Healthcare Sector. syngo MR D Operator Manual - Breast 0.0.

Signal-to-Noise Ratio and Absorbed Power as Functions of Main Magnetic Field Strength, and Definition of 90 RF Pulse for the Head in the Birdcage Coil

Photomultiplier Tube

10. Phase Cycling and Pulsed Field Gradients Introduction to Phase Cycling - Quadrature images

Review of Progress in Quantitative Nondestructive Evaluation, Vol. 14 Edited by D.O. Thompson and D.E. Chimenti, Plenum Press, New Yorlc,l99S 1749

Gradient Spoiling. Average balanced SSFP magnetization Reduce sensitivity to off-resonance. FFE, FISP, GRASS, GRE, FAST, Field Echo

Medical Imaging. X-rays, CT/CAT scans, Ultrasound, Magnetic Resonance Imaging

Hardware. MRI System. MRI system Multicoil Microstrip. Part1

Fast Field-Cycling Magnetic Resonance Imaging (FFC-MRI)

2015 Spin echoes and projection imaging

ACRIN 6686 / RTOG 0825

EPISTAR MRI: Multislice Mapping of Cerebral Blood Flow

Improve Image Quality of Transversal Relaxation Time PROPELLER and FLAIR on Magnetic Resonance Imaging

PET/CT Instrumentation Basics

MRI imaging in neuroscience Dr. Thom Oostendorp Lab class: 2 hrs

Digital Imaging CT & MR

MR in RTP. MR Data for Treatment Planning: Spatial Accuracy Issues, Protocol Optimization, and Applications (Preview of TG117 Report) Acknowledgements

Spiral MRI on a 9.4T Vertical-bore Superconducting Magnet Using Unshielded and Self-shielded Gradient Coils

The development of the RF-pulse for the low level SAR used by the MRI.

CARING DESIGN. PRACTICAL TECHNOLOGY.

diagnostic examination

A Complete Digital Magnetic Resonance Imaging (MRI) System at Low Magnetic Field (0.1 Tesla)

Alae Tracker: Tracking of the Nasal Walls in MR-Imaging

Numerical Evaluation of an 8-element Phased Array Torso Coil for Magnetic Resonance Imaging

Passive Tracking Exploiting Local Signal Conservation: The White Marker Phenomenon

NEMA Standards Publication MS (R2014) Determination of Signal-to-Noise Ratio (SNR) in Diagnostic Magnetic Resonance Imaging

Analysis of spatial dependence of acoustic noise transfer function in magnetic resonance imaging

Effect of RF Pulse Sequence on Temperature Elevation for a Given Time-Average SAR

3T Unlimited. ipat on MAGNETOM Allegra The Importance of ipat at 3T. medical

Acoustic noise reduction of MRI systems by means of magnetic shielding

Beam-Restricting Devices

Fundamental and Clinical Studies for Effectiveness of Zero-filling Interpolation on k-space for Improvement of Sharpness in Magnetic Resonance Imaging

MR Basics: Module 6 Pulse Sequences

Enhancing Gray-to-White Matter Contrast in 3T T1 Spin-Echo Brain Scans by Optimizing Flip Angle

Development of a 1.0 T MR microscope using a Nd-Fe-B permanent magnet

MR in Tx Planning. Acknowledgements. Outline. Overview MR in RTP

PET Performance Evaluation of MADPET4: A Small Animal PET Insert for a 7-T MRI Scanner

Electromagnetic Field Exposure Feature of a High Resonant Wireless Power Transfer System in Each Mode

Model Based Tissue Differentiation in MR Brain Images

HISTORY. CT Physics with an Emphasis on Application in Thoracic and Cardiac Imaging SUNDAY. Shawn D. Teague, MD

IR/SR TrueFISP. Works-in-Progress package Version 1.2. For the SIEMENS Magnetom. Installation and User s Guide NUMARIS/4VA21B.

2. Sources of medical images and their general characteristics

Liver imaging beyond expectations with Ingenia

Field Simulation Software to Improve Magnetic Resonance Imaging

Automatic Selection of Mask and Arterial Phase Images for Temporally-Resolved MR Digital Subtraction Angiography

INNOVATION BY DESIGN. Toshiba A History of Leadership REMOTE CONTROL R/F SYSTEM

Echo-Planar Imaging for a 9.4 Tesla Vertical-Bore Superconducting Magnet Using an Unshielded Gradient Coil

NMR Basics. Lecture 2

Multi-Access Biplane Lab

Transcription:

181 High-Field Surface-Coil MR Imaging of Localized Anatomy John F. Schenck,' Thomas H. Foster,' John l. Henkes,' William J. Adams,' Cecil Hayes,2 Howard R. Hart, Jr.,' William A. Edelstein,' Paul A. Bottomley,' and Felix W. Wehrli2 For some time it has been recognized that surface-receivercoil designs can enhance certain aspects of magnetic resonance (MR) signal detection [1]. During the last several years, surface coils have been the predominant means of recording the MR spectra of metabolites from animals and humans in vivo [2-4]. Efforts to produce surface-coil images at low magnetic field strengths have been undertaken as well [5, 6]. Recently, we exploited the properties of these coils to achieve enhanced sensitivity in MR imaging at 1.5 T [7]. Initial results demonstrate that significantly improved resolution of small anatomic structures is possible when surface coils are used as receiver elements in high-field MR imaging systems. We have been exploring this new technique as part of an ongoing effort to improve the diagnostic usefulness of MR in specific regions of the anatomy. Presented here are three series of images of the neck, the inner ear, and the lumbar spine obtained using the surface-coil method. Materials and Methods Our MR imaging system is built around an Oxford Instruments 1- m-bore 1.5 T superconducting magnet. The 'H resonant frequency at this magnetic field strength is 63.9 MHz. All of the images in our study were recorded using separate radiofrequency (RF) transmitter and receiver coils. 'H excitation was accomplished using a wholebody cylindrical coil with a diameter of about 51 cm. The images of the neck and of the inner ear were acquired using the same 13.5-cmdiam surface coil as the receiver element. In both cases, this flexible coil was shaped to conform closely to the region of anatomy under study. The lumbar spine was imaged with a slightly larger diameter (",22 cm) surface-coil receiver to optimize sensitivity in the deeper pixels of interest. The RF fields of the transmitter and receiver coils were orthogonalized to minimize their mutual inductive coupling. Throughout this investigation, the imaging slice thickness was 5 mm. There was no "dead space" between images. Images comprised 128 x 256 pixels, corresponding to a resolution in the plane of the image (except where noted in the figure legends) of about 1 x 0.5 mm. Pulse repetition times (TRs) and total scan time for each section are specified in the figure legends. A partial-saturation pulse sequence was used throughout, and images were recorded with our spin-warp spin-echo (SE) method [8). Healthy volunteers served as the subjects for these studies. The offset slice capability of our imaging system was used so that the complete series of images in each anatomic region was acquired without repositioning the patient. Results A series of 12 transverse sections through the neck is shown in figure 1. The sequence proceeds from sections closest to the chest upward toward the head with a distance between centers of consecutive scans of 5 mm. Artifact caused by respiratory motion is apparent in figures 1 A-1 D. The detail of the arteries and veins of the neck is particularly encouraging. Good discrimination between thyroid and adjacent muscular (and possibly parathyroid) tissue is noted as well. Figure 2 shows nine consecutive sagittal sections through the ear, external meatus, and inner ear; fi,gure 3 is a coronal sequence in the same region. The sensitivity of the surface coil is well illustrated in figure 3A. Signal strength is excellent within the semicircular region that has a radius approximately equal to that of the surface coil and falls off rapidly for more remote pixels. Resolution of fine anatomic detail in the inner ear is afforded by the technique in the two imaging planes. The sensitive region of a larger surface coil receiver is demonstrated in figure 4A. Figures 4B-4E show a series of consecutive sagittal sections through the lumbar spine. Pulse TRs, total scantimes, and resolution in the plane of the image are provided in the figure legends. Discussion The development of the surface-coil technique is an important aspect of advancing the capabilities of MR imaging and establishing MR 's role in specific clinical applications. These experiments confirm our earlier findings that the use of modified surface-coil receivers in high-magnetic-field MR imaging. systems generates sufficiently improved signal-to-noise (SjN) ratio to permit a significant reduction in the imaging voxel size. This reduction, achieved through the application of stronger imaging gradient fields, is responsible for the im- Received July 18, 1984; accepted after revision September '" 1984. 1 General Electric Corporate Research and Development Center, P.O. Box 8, Schenectady, NY 12301. Address reprint requests to T. H. Foster. 2 General Electric Medical Systems, Milwaukee, WI 53201. AJNR 6:181-186, March/April 1985 0195-6108/85/0602-0181 $00.00 American Roentgen Ray Society

182 SCHENCK ET AL. AJNR:6, March/April 1985 A B c D E F Fig. 1.-Axial sections through neck using 13.5-cm-diam flexible surface-coil receiver. Slice thicknesses and distances between centers of consecutive slices are all 5 mm. Image size is 128 x 256 pixels; pixel size is 1 x 0.5 mm. A-C, TR = 500 msec; O-L, TR = 300 msec. Scan times were 256 sec for A-C and 154 sec for O-L. (A-F are above; G-L are on facing page.) proved resolution of anatomic detail observed in these images. The 1 x 0.5 x 5 mm picture-volume elements achieved here are smaller by a factor of two than those typically obtained in our imaging system when standard head-sized cylindrical RF coils are used as transmitter/receiver elements [10]. The ability to obtain high-resolution images of regions of anatomy not readily accessible to standard MR coils is an-

AJNR:6, March/April 1985 HIGH-FIELD SURFACE-COIL MR IMAGES 183 G H J K L other advantage offered by the surface-coil technique. When, for example, the neck is imaged with an RF coil large enough to accept the whole body, the arrangement yields relatively poor coupling between' H nuclei in the neck and the coil and results in a suboptimal SIN ratio. The surface-coil receiver, on the other hand, may be fashioned to conform closely to the surface of the neck, thereby enhancing the magnetic flu x linkage between the nuclear spins and the coil while at the

184 SCHENCK ET AL. AJNR:6, March/April 1985 A B c G H Fig. 2.-Parasagittal sections through ear and auditory canal. TR = 300 msec. Field of view in E is 250 x 250 mm, corresponding to resolution in plane of image of 2 x 1 mm; total scan time was 77 sec. Other images have 1 x 0.5 mm resolution and were recorded in 154 sec/slice. Distance between centers of consecutive sections is 5 mm in all but one instance: F and G are separated by only 2.5 mm. Slice thicknesses are 5 mm.

AJNR:6, March/April 1985 HIGH-FIELD SURFACE-COIL MR IMAGES 185 Fig. 3.-Coronal images through inner ear. Field of view in A is 250 x 250 mm, corresponding to resolution in plane of image of 2 x 1 mm; total scan time was 77 sec. Other images have 1 x 0.5 mm resolution and were recorded in 154 sec/slice. Distances between centers of consecutive slices are 2.5 mm in B-O and 5 mm between A and B. (C is reprinted from [9].) same time greatly reducing the received patient generated noise voltage. The method we describe involves no modification to the MR imaging system other than the introduction of separate transmitter and receiver cells. This fact, together with the improvement in image content obtained, immediately raises the possibility of coupling the surface-coil technique with other emerging MR methods. Recently, for example, a chemicalshift imaging scheme was described and demonstrated whereby water (H20) or lipid (-CH2-) protons may be imaged independently [11]. The ability to perform this type of imaging under the high-resolution conditions offered by surface coils may enhance MR 's capability to diagnose atherosclerotic lesions of the carotid and vertebral arteries. Work will continue in an effort to further optimize surface-coil design for specific regions of the anatomy. ACKNOWLEDGMENTS We thank Patrick Jaskolski, Linda Hutton, and Cheryl Tauber for technical assistance in performing these studies; W. M. Leue, R. D. Darrow, D. R. Eisner, G. D. Brower, L. S. Smith, D. Vatis, O. M. Mueller, J. E. Piel, Jr., and R. E. Argersinger for contributions to our MR program; and Arlene Agresta for help in manuscript preparation. REFERENCES 1. Zhernovoi AI, Latyshev GD. Nuclear magnetic resonance in a flowing liquid. New York: Plenum, 1965:70 2. Ackerman JJH, Grove TH, Wong GG, Gadian DG, Radda GK. Mapping of metabolites in whole animals by 31p using surface coils. Nature 1980;283 :167-170 3. Alger JR, Behar KL, Rothman DL, Shulman RG. Natural-abundance 13C NMR measurement of hepatic glycogen in the living rabbit. J Magnetic Resonance 1984;56:334-337 4. Bottomley PA, Hart HR, Edelstein WA, et al. NMR imaging/ spectroscopy system to study both anatomy and metabolism. Lancet 1983 ;2 :273-274 5. Alfidi RJ, Haaga JR, EI Yousef SJ, et al. Preliminary results in humans and animals with a superconducting whole-body, nuclear magnetic resonance scanner. Radiology 1982;143:1 75-181

186 SCHENCK ET AL. AJNR:6, March/April 1985 A B c Fig. 4.-Sagittal sections through lumbar spine. For each image, TR = 300 msec and scan time is 308 sec. A is midsagittal section with 250 x 250 mm field of view and resolution in imaging plane of 2 x 1 mm. B is the same slice with field of view narrowed to 125 x 125 mm. B-E have 1 x 0.5 mm pixel sizes. Imaging plane was moved laterad at 5 mm steps to acquire C-E: 6. Axel L. Surface coil magnetic resonance imaging. J Comput Assist Tomogr 1984;8:381-384 7. Edelstein WA, Schenck JF, Hart HR, Hardy CJ, Foster TH, Bottomley PA. High resolution NMR imaging at 1.5 Tesla using surface coils. JAMA (in press) 8. Edelstein WA, Bottomley PA, Hart HR, Smith LS. Signal, noise and contrast in nuclear magnetic resonance (NMR) imaging. J Comput Assist Tomogr 1983;7:391-401 9. Daniels DL, Pech P, Haughton VM. Magnetic resonance imaging of the temporal bone. Milwaukee: General Electric, 1984 10. Bottomley PA, Hart HR, Edelstein WA, et al. Anatomy and metabolism of the normal human brain studied by magnetic resonance at 1.5 Tesla. Radiology 1984;150:441-446 11. Bottomley PA, Foster TH, Leue WM. Chemical imaging of the brain by NMR. Lancet 1984;1 : 1120