Film Holder Accessories

Similar documents
Automated Double Aperture Accessory

Solid Sample Holder Accessory

Unique Scattering Measurements Using the Agilent Universal Measurement Accessory (UMA)

Agilent Cary 7000 Universal Measurement Spectrophotometer (UMS)

Vinyl Cutter Instruction Manual

Educational Spectrophotometer Accessory Kit and System OS-8537 and OS-8539

Measuring optical filters

Single-Slit Diffraction. = m, (Eq. 1)

General Guidelines:

Automated Spectrophotometric Spatial Profiling of Coated Optical Wafers

EDUCATIONAL SPECTROPHOTOMETER ACCESSORY KIT AND EDUCATIONAL SPECTROPHOTOMETER SYSTEM

Activity P35: Light Intensity in Double-Slit and Single-Slit Diffraction Patterns (Light Sensor, Rotary Motion Sensor)

LED Thin Frame Fixed Frame Screen User Guide

Stainless Steel Bench Stand

1. ASSEMBLING THE PCB 2. FLASH THE ZIP LEDs 3. BUILDING THE WHEELS

CAM SYSTEM. Eccentric self-locking clamping devices. Set. CAM SYSTEM t. CAM SYSTEM t. Set. CAM SYSTEM s pag CAM SYSTEM s pag.

Section 11 CABLE & ROD SYSTEMS, SIGN FIXING & SIGN STANDOFFS. Web. acrylicdesign.ie .

Assembly and installation help

ASSEMBLY INSTRUCTIONS RIGHT & LEFT HAND "J" TABLE

a.k.a. casegoods instructions

S6 User s Manual USER S MANUAL ver. 1.0

WARNING: Prior to installation, turn the power off to the vending machine and unplug it from its power source. Also, make sure to level the machine.

Ribcage Installation. Part 2 - Assembly. Back-Bone V1.06

1503 Follow Spot Yoke, Source Four LED

UV-Vis-NIR Spectrophotometer Quick Start Guide

FixLogix CMM Fixture System Instructions

Kai Installation Instructions

SETUP INSTRUCTIONS If you would like to tell us about your experience with your setup instructions please us at

ROLL-A-GLIDE INSULATED ROLLER DOOR

BEST PRACTICE GUIDE. Socket Bases. Working with Concrete Slabs

WHEN DISASSEMBLING ALUMINUM EXTRUSION, TIGHTEN ALL SETSCREWS AND LOCKS TO PREVENT LOSS DURING SHIPPING

The Useless Machine. Parts Only - Build Guide v0001

REC Series Rack Installation Guide

University of Minnesota Nano Fabrication Center Standard Operating Procedure Equipment Name:

Rotary Fixture M/V/X CLASS LASER SYSTEMS. Installation and Operation Instructions

MAIN PARTS

7878 K940. Checkpoint Antenna. Kit Instructions. Issue B

Fig. 2 DORMA-Glas Stand/Issue 02/03 Seite/Page 1/7

Pickup Box Utility Rack Package Installation (Instruction ID: )

A500 ASSEMBLY & INSTALLATION INSTRUCTIONS

Installers guide Deadbolt 02.

Installation Instructions

ASSEMBLY AND ADJUSTMENT

INSTALLATION INSTRUCTIONS

ASSEMBLY INSTRUCTION CONTENT

SQ2 User Instructions SQ2 Overview:

Ohbot. Eyes turn. servo. Eyelids open. servo. Head tilt. servo Eyes tilt. servo. Mouth open servo. Head turn servo

PACLEASE Hybrid Sacagawea VK-1222 Display Instructions / PAGE 1 - General Layout

Elara NanoEdge Fixed Frame Screen User Guide

SLIDING MECHANISM TROLLEY CATCH TROLLEY ASSEMBLY FLOOR GUIDE

7. Operating instructions: EFL 300

GlideRite Retractable Cover System For Hot Spot Spas (SE & SLX only)

Replacement of Drive Nuts in TTL/Adding Drive Nut Safety Bracket

18600 Angular Momentum

GAGING AND INSPECTION

THE ROGUE TM FUNSLIDE TM

SS1062, SS10621 & SS10621E Free Standing PWC & Fishing Boat Hoist SS1062 SS10621 SS10621E

43107 Rhino Jerry Can Holder Rhino Jerry Can Holder - Horizontal

3 wide x 3 deep Video Wall Display Installation Guide

Orbital Truss Kiosk 02

GlideRite Retractable Cover System For HotSpring & Tiger River Spas (except Classic & pre-2000 Landmark Spas)

EPPA2-KIT DUAL MONITOR ARM CONVERSION

EPS Duct Fabrication and Installation Sheet

Quill Stop V2 Installation Guide 11/16/2014

Isuzu D-Max Part Numbers Black Polished

Cover Stitch Wide. Cover Stitch Looper Thread. Cover Stitch Wide page 1

COACT MEDIA UNIT ASSEMBLY WITH X9 OPEN SHELF STORAGE OPTION PART # INSTALLATION INSTRUCTIONS

POP PLUS / SPIDER SET-UP INSTRUCTIONS

EDGE2 DUAL MONITOR ARM

Written By: Walter Galan

HOW TO INSTALL HORIZONTAL ROD RAILING TREX SIGNATURE STANDARD

Greenslade & Company, Inc. USA

CS Unitec, Inc. 22 Harbor Avenue, Norwalk, CT Phone: Toll-free:

OPTICAL BENCH - simple type

Codonics SLS 500i Shelf Installation

SCHACHT STANDARD FLOOR LOOMTM

SmartView Mounting Frame 3 Wide x 3 Deep Video Wall Display Installation Guide

Delrin Cone T-Handle Spring Stop w/pad...9 Screw Jack Large Toggle Clamp w/pad...9 Adjustable Ball Positioner...14

Installation Manual Using Existing Handrail Version 4.0 Oct 2017

FitWork Walkstation Series 7 AdjusTables

Read Below! Read Below! Read Below! Read Below! Read Below! Read Below! STOP READ TIPS BELOW TO MAKE ASSEMBLY MUCH EASIER

v1.0 ASSEMBLY GUIDE Mia Wide Bookcase

Interactive Monitor Arm

Installing Your Electronic Deadbolt

S48-L12-SC AND G48-L12-GC STEEL PORTA-DOCK S82 SC 6 X 12 PLATFORM AND G82 GC 6 X 12 PLATFORM

Measuring photometric accuracy using the double aperture method

C70 Window Roller Repair Taken from: Heres the problem:

Thank you for purchasing out product! *Please read these instructions and follow them step by step. *

US RACK, Inc Falcon Drive, Madera, CA

ATLANTIS RAIL Contact Information

Range height adjustable assembly

JEEP JK ( 5 DOOR ) SLIMLINE II - FULL TRAY EXTREME RACK KIT

Horizontal Cable Systems

OWNER S MANUAL. But that s just half the story. The fence INCRA Miter Gauge really does work

HOUSE PARTS PACKED IN HOUSE BOX PARTS IN PLASTIC BAG (HARDWARE) PARTS IN SMALL PLASTIC BAG (FLOOR CLIPS) PARTS PACKED IN BUNDLE

Deauville Installation Guide

GF Machining Solutions. System 3R. Mecatool WEDM. Productportfolio. Version 2, valid from July 1st. 2017

ASSEMBLY INSTRUCTIONS FOR DESK & CREDENZA PRODUCTS

Bunk Pod Front Entry Assembly Instructions

BombiniBot Parts and Assembly

Transcription:

Page 1 of 9 Film Holder Accessories For Cary 1/3/4/5/100/300/400/500/500i/4000/5000/6000i/Deep UV Part numbers: Cary 100/300-0010048500; Cary 400/500/500i/4000/5000/6000i - 0210125400 Contents This document is also available in PDF format for improved print quality. PDF files are stored in the 'Manuals' folder on the software disk. Introduction Getting started Unpacking Installing Accessories Applications Measuring multiple filters Measuring gels Spare parts Introduction The Film Holders are versatile accessories for use in conjunction with a Sample Transport Accessory. They provide vertical mounting capabilities for films, gels, sheets of material or multiple filters. By moving the platform across the sample beam you can plot photometric mode versus position in millimeters, with photometric mode being the ordinate and millimeters the abscissa. The Film Holder accessories are fitted with Mylar strips to prevent samples being scratched. All holes in the Cary 4000/5000/6000i version of the accessory are pre-threaded: you need only use thumbscrews. The Film Holder accessory requires the Extended Sample Compartment (part number for Cary 100/300: 0010041800; Cary 400/500/500i: 0010046600; Cary 4000/5000/6000i: 0010079800) and the Sample Transport accessory (part number for Cary 100/300: 0010040600; Cary 400/500/500i: 0010044700). These are available from Agilent. Getting started Unpacking Unpack the accessory and check that you have received all of the components listed in the packing list. For the Cary 100/300 you should have: 1 x Film Holder 4 x V-plates 1 x gel boat mount 2 x plastic clamps with mounting pins 14 x thumbscrews 1 x resting plate 2 x 5 mm magnetic masks 2 x 10 mm magnetic masks 4 x magnetic strips For the Cary 400/500/500i/4000/5000/6000i you should have:

Page 2 of 9 1 x Film Holder 16 x V-plates 2 x gel boat mounts 4 x plastic clamps with mounting pins 25 x thumbscrews 2 x resting plates 2 x 5 mm magnetic masks 2 x 10 mm magnetic masks 4 x magnetic strips Installing Cary 1/3/100/300 You will need: Inspect all parts for damage during transit. Any damage should be reported immediately. Refer to the accessory by part number. Extended Sample Compartment Sample Transport Accessory 1. Install the Cary 1/3/100/300 Sample Transport accessory according to its instructions. 2. Pull the upper section of the Sample Transport accessory towards you. 3. Hold the Film Holder so that the Mylar strips are on the side to the right. 4. Place it lengthwise on top of the Sample Transport accessory. 5. Line up the holes at each end of the Film Holder with the corresponding holes in the Sample Transport accessory. 6. Fasten with thumbscrews. 7. Slide the Sample Transport accessory back into the sample compartment. 8. Remove the standard base from the Extended Sample Compartment and replace it with the one supplied with the Film Holder. 9. Install the Cary 1/3/100/300 Extended Sample Compartment according to its instructions. Cary 4/5/400/500/500i/4000/5000/6000i You will need: Extended Sample Compartment Sample Transport Accessory 1. Install the Cary 4/5/400/500/500i/4000/5000/6000i Sample Transport accessory according to its instructions. 2. Pull the upper section of the Sample Transport accessory towards you. 3. Hold the Film Holder so that the mylar strips are on the side to the right.

Page 3 of 9 4. Place it lengthwise on top of the Sample Transport accessory, with the lower end of the vertical section towards the back of the Sample Compartment and the higher end towards you. 5. Line up the two smaller holes on the horizontal section of the Film Holder with the two locating pins on the Sample Transport accessory. 6. Push the thumbscrew with the brown standoff through the keyhole and tighten it. 7. Slide the Sample Transport accessory back into the Sample Compartment. 8. Install the Cary 4/5/400/500/500i Extended Sample Compartment or the Cary 4000/5000/6000i Extended Sample Compartment according to its instructions. Accessories V-plates Up to 9 V-plates can be attached underneath each window of the Film Holder. This allows you to, for example, measure a batch of small round filters in turn without having to measure each one individually. Gel boat mounts With the exception of the magnetic masks, all accessories must be attached to the right side of the Film Holder. These hold gel boats in which strips of plate gel can be placed and then scanned. For the Cary 1/3/100/300, one gel boat mount is provided. For the Cary 4/5/400/500/500i/4000/5000/6000i, one gel boat mount is provided for the reference beam and one for the sample beam. Gel boats (part number 6610004200, 8 mm pathlength; 6610004300, 5 mm pathlength) are available from Agilent. Plastic clamps with mounting pins These are used for attaching heavier samples to the Film Holder. The mounting pins can be attached to any hole in the Film Holder, then the clamps attached to them. The clamps may also be used in conjunction with the resting plates.

Page 4 of 9 1. Screw a thumbscrew through the hole in the Film Holder where you want to attach the mounting pin. 2. Screw the mounting pin onto the thumbscrew. 3. Push the plastic clamp onto the mounting pin. The rubber grommets hold the sample in position. Resting plates These are used for supporting heavier samples on the Film Holder. One is provided for the reference beam and one for the sample beam.

Page 5 of 9 1. Attach with thumbscrews beneath the appropriate window of the Film Holder. 2. Rest the bottom of the sample on the resting plate and clamp the top of the sample to the Film Holder. Magnetic masks These allow you to reduce the light beam height if you want to measure only part of a sample. Two widths are provided -- 5 and 10 mm. Place the magnetic masks over the Film Holder's windows, on the left side.

Page 6 of 9 Magnetic strips These are used to attach light, thin samples to the Film Holder. Place a magnetic strip at the top and bottom of the sample to hold it flat against the right side of the Film Holder.

Page 7 of 9 Applications Measuring multiple filters 1. Attach the V-plates underneath each window of the Film Holder, screwing them with thumbscrews onto the hole closest to the base of the Film Holder. The 'V' section should be pointing upwards. The V-plates for the Cary 4000/5000/6000i need one thumbscrew each; those for the Cary 100/300 need two each, placed side-by-side. 2. Rest each filter on the 'V' section of a V-plate. 3. Perform a wavelength scan on each filter in turn. With a filter in each 'V'-holder, the centers of the filters are 20 mm apart. 4. Switch on the computer, then the spectrophotometer and wait for the initialization procedure to be completed. 5. Click the Start button on the Windows taskbar, then Programs, Cary WinUV and Scan. 6. Click the Setup button and set these parameters: Cary page X mode: Mode X mode: Start/Stop Y mode: Mode Y mode: Y min 0 Y mode: Y max 2 Nanometers User set range Abs Ave time (s) 0.100 Data interval (nm) 1.000 Scan rate (nm/min) 600.00 Cycle: Cycle mode Cycle: Cycle count 1 On

Page 8 of 9 Options page SBW/Energy: SBW (nm) 2.000 SBW/Energy: Beam mode Double Source/Detector Vis Source/Detector: Source changeover 350.0 Baseline page Correction None Auto Accessories page Sample Transport On Sample Transport: Start (mm) 0.0 Sample Transport: Stop (mm) 100.0 Sample Transport: Interval 20.0 Auto Store page Storage on (prompt at start) On 7. Place the samples on the V-holders. 8. Click the Start button to commence reading. You will also be prompted to select a location in which to store the data files. Each filter will be scanned in turn. Measuring gels You may need to store a separate baseline for each filter position if masks are used. 1. Cut the gel to the correct size to fit in the gel boat. 2. Place the gel in the gel boat. 3. Line up the window on the gel boat mount with the appropriate window of the Film Holder. The lip on the window of the gel boat mount should be at the bottom. For the Cary 1/3/100/300, match up the slot on the top of the gel boat mount with the fifth hole from the front and the second hole from the top of the Film Holder. For the Cary 4/5/400/500/500i, match up the slot on the top of the gel boat mount with the fourth hole from the front of the Film Holder. Because of the positioning of the gel boat mount, you will need to set up the instrument to scan from 15 to 96 mm. 4. Fasten this hole and the two other holes on the gel boat mount with thumbscrews. 5. Place the gel boat on the lip of the window of the gel boat mount. The rubber grommets at each end of the window hold the boat in place. 6. Click the Start button on the Windows taskbar, then Programs, Cary WinUV and Scan. 7. Click the Setup button and set these parameters: Cary page X mode: Mode Nanometers

Page 9 of 9 X mode: Start/Stop User set range Y mode: Mode Abs Y mode: Y min 0 Y mode: Y max 2 Ave time (s) 0.100 Data interval (nm) 1.000 Scan rate (nm/min) 600.00 Cycle: Cycle mode On Cycle: Cycle count 1 Options page SBW/Energy: SBW (nm) 2.000 SBW/Energy: Beam mode Double Source/Detector Vis Source/Detector: Source changeover 350.0 Baseline page Correction None Auto Accessories page Sample Transport On Sample Transport: Start (mm) 15.0 Sample Transport: Stop (mm) 95.0 Sample Transport: Interval 10.0 Auto Store page Storage on (prompt at start) On 8. Click the Start button to commence reading. Spare parts These spare parts are available from Agilent. Use only Agilent-supplied parts, unless otherwise specified. Kits containing spare parts are available for the Film Holder. The kits contain magnetic strip, V-plate, gel boat holder, resting plate, 5 mm magnetic mask, 10 mm magnetic mask and plastic clamp. Model Part number Cary 1/3/100/300 9910064500 Cary 4/5/400/500/500i/4000/5000/6000i 9910064600