Attenuation Correction in Hybrid MR-BrainPET Imaging

Size: px
Start display at page:

Download "Attenuation Correction in Hybrid MR-BrainPET Imaging"

Transcription

1 Mitglied der Helmholtz-Gemeinschaft Attenuation Correction in Hybrid MR-BrainPET Imaging Elena Rota Kops Institute of Neuroscience and Biophysics Medicine Brain Imaging Physics

2 Interactions of 511 kev Photons in Matter Photoelectric effect Compton scattering I = I 0 exp( - µ dx) µ µ compton + µ photoelectric I 0 = initial photon flux I = photon flux without interaction in medium x = thickness of the medium µ = linear attenuation coefficient (it depends on the atomic number Z of the material and on the photon energy) µ compton µ photoelectric µ Half-values thks (1/cm) (1/cm) (1/cm) (cm) Brain tissue ~0.096 ~ ~ Bone ~0.169 ~0.001 ~ Oct. 28, 2008 INB Medicine (Brain Imaging Physics) Folie 2

3 Linear attenuation coefficient µ in PET scanner Emission The detector pair measures: The measured counts: Counts m = A dx exp( - µ dx) Counts corr = Counts m exp( + µ dx) = A dx Line of Response (LOR) = Counts m ACF = A dx ACF = attenuation correction factor Oct. 28, 2008 INB Medicine (Brain Imaging Physics) Folie 3

4 In case of the good old dedicated PET scanners Blank Transmission I o I o rotating line source with positron emitter Ge-68 I during the blank scan I = I o during the transmission scan I = I o * exp( - µ dx I 0 = exp(+ µ dx) = ACF I Oct. 28, 2008 INB Medicine (Brain Imaging Physics) Folie 4

5 Attenuation image I 0 ln = µ dx) I which is exactly the Radon transform to get an image of the attenuation coefficients µ Line integrals of the transmission image µ dx => forward projection I 0 = ACF for each LOR I Oct. 28, 2008 INB Medicine (Brain Imaging Physics) Folie 5

6 Attenuation Measurement in PET/CT CT (HU) smoothed CT (HU) µ-map (1/cm) 68 Ge measured µ-map (1/cm) CT 0HU CT > 0HU µ µ PET PET = µ = µ PET H 2 O PETO H 2 PET 1 PET 1 with µ H 2 O = cm and µ Bone = cm CT 1 CT 1 µ H 2 O = cm and µ Bone = cm ( CT ) / 1000 CT 2 ( PET PET µ 2 + CT H O µ Bone µ H O ) CT 1000( µ µ ) Bone CT H 2 O cm cm 1 1 Burger C et al. PET attenuation coefficients from CT images: experimental evaluation of the transformation of CT into PET 511-keV attenuation coefficients. Eur J Nucl Med (2002) 29: Oct. 28, 2008 INB Medicine (Brain Imaging Physics) Folie 6

7 Attenuation Measurement in PET/CT CT (HU) smoothed CT (HU) µ-map (1/cm) 68 Ge measured µ-map (1/cm) CT BreakPoint µ PET = 9. 6x10 5 ( HU ) cm 1 CT > BreakPoint µ PET = a ( HU ) + b cm 1 with a and b functions of kvp Carney JPJ et al. Method for transforming CT images for attenuation correction in PET/CT imaging. Med Phys (2006) 33: Oct. 28, 2008 INB Medicine (Brain Imaging Physics) Folie 7

8 From PET/CT to MR-PET Oct. 28, 2008 INB Medicine (Brain Imaging Physics) Folie 8

9 Attempts to get a segmented MRI Two-Compartment Method ACF = exp(+ µ (x,y)dl') LOR brain thks skull thks ACF (brain+skull) ACF (only brain) rel. difference (cm) (cm) (1/cm) (1/cm) (%) LOR 1 LOR 1 LOR 2 LOR LOR 4 LOR 4 LOR 3 LOR LOR 5 LOR 5 Oct. 28, 2008 INB Medicine (Brain Imaging Physics) Folie 9

10 Attempts to get a segmented MRI Morphological Operators MRI CT segmented MRI segmented CT Dogdas B, Shattuck DW, and Leahy RM Segmentation of Skull and Scalp in 3-D human MRI Using Mathematical Morphology Human Brain Mapping (2005) 26: Oct. 28, 2008 INB Medicine (Brain Imaging Physics) Folie 10

11 Attempts to get a segmented MRI Neural network- and Knowledge-based segmentation First results T1-weighted MPRAGE data set Classification and Segmentation based on - Neural network and - Knowledge on relative positions into GM, WM, CSF, extracerebral region Wagenknecht et al., Methods Inf Med 43(2004), Further segmentation ( ongoing work) Brain tissue, different cavities, skull, scalp Wagenknecht G Central Institute for Electronics (ZEL), Jülich In collaboration with Rota Kops E, Herzog H, INB3, Jülich Oct. 28, 2008 INB Medicine (Brain Imaging Physics) Folie 11

12 Attempts to get a segmented MRI Two-points Dixon MRI. Whole Body Water Fat Martinez-Möller A, Souvatzoglou M, Navab N, Schwaiger M, and Nekolla S MR-based attenuation correction for whole-body MR/PET J Nucl Med (2008) 49 (Supplemet):65P Segmented image CT Lesion SUV = - Oct. 28, 2008 INB Medicine (Brain Imaging Physics) Folie 12

13 Attempts to get a segmented MRI ultra short TE (UTE) Normal tibia Difference UTE image: IR=500/0.08 minus 4.5/250 ms Frontal bone Difference UTE image: IR=500/0.08 minus 4.5/200 msec at 1.5T Tyler DJ, Robson MD, Henkelman RM, Young IR, and Bydder GM Magnetic Resonance Imaging with Ultrashort TE (UTE) Pulse Sequences: Technical Considerations J Magn Reson Imaging (2007) 25: Oct. 28, 2008 INB Medicine (Brain Imaging Physics) Folie 13

14 Attempts to get a segmented MRI ultra short TE (UTE): New Results Keereman V, Vandenberghe S, De Deene Y, Luipaert R, and Broux T MR-based Attenuation Correction for PET Using an Ultrashort Echo Time (UTE) Sequence Presented at the IEEE Medical Imaging Conference 2008, Dresden Oct. 28, 2008 INB Medicine (Brain Imaging Physics) Folie 14

15 Alternative methods to get attenuation maps Template Based. Theory + = T1-MR template (e.g. a self-build one) individual T1-MR Transmission template (e.g. a self-build one) individualized attenuation map complete individualized attenuation map reformatting into ECAT7 image and voxel dimensions forward projection reconstruction with with attenuation correction and ROI analysis complete individualized attenuation map image: 128x128x65 voxel: 5.15 x 5.15 x 2.42 mm 3 sinogram of ACFs Rota Kops E, and Herzog H Alternative Methods for Attenuation Correction for PET Images in MR-PET Scanners IEEE NSS/MIC Conference record (2007) pp & Template-basierte Schwächungskorrektur für PET-Bilder in MR-PET-Geräten Nuklearmedizin (2008) 47:A63 Oct. 28, 2008 INB Medicine (Brain Imaging Physics) Folie 15

16 Alternative methods to get attenuation maps Template Based. Results Measured attenuation image Female and Male Data 8.00 Male subject relative Differenzen (%) Frontal cortex Parietal cortex Occipital cortex Caudate nucleus Putamen Thalamus Female subject Template based on SPM2 with template based fp attenuation factors y = 0.93x R 2 = 0.95 Female subjects y = 1.00x R 2 = with measured fp attenuation factors w ith measured attenuation factors with tamplate based fp attenuation factors y = 0.89x R 2 = 0.91 Male subjects y = 0.99x R 2 = with measured fp attenuation factors with measured attenuation factors Oct. 28, 2008 INB Medicine (Brain Imaging Physics) Folie 16

17 Alternative methods to get attenuation maps Template based. Ongoing Work. Brain Ongoing work: 1) Optimization of templates: possible gender specific differences and 2) Use of UTE sequences on 3T. Oct. 28, 2008 INB Medicine (Brain Imaging Physics) Folie 17

18 Alternative methods to get attenuation maps MR-CT Database. Atlas Database of corresponding MR-CT Pairs {MR i } {CT i } {MR new } Registration for every MR i to the patient MR new and apply to CT i Find all neighbouring MR patches Perform Gaussian Regression on patch and position CT estimates for every voxel Scale CT values (HU) to PET attenuation values Estimate PET Attenuation Image Hofmann M et al, MR-based PET Attenuation Correction Method and Validation Presented at IEEE NSS/MIC Conference 2007 & MRI-Based Attenuation Correction for PET/MRI: A Novel Approach Combining Pattern Recognition and Atlas Registration J Nucl Med (2008) 49: Oct. 28, 2008 INB Medicine (Brain Imaging Physics) Folie 18

19 Alternative methods to get attenuation maps MR-CT Database. Results x 10 7 Relative difference between PET (MR-AC) and PET (CT-AC) Activity in PET (MR-AC) Nr of segmented VOI Activity in PET (MR-AC) x 10 7 Fig. 3: Left: Relative differences between activity concentrations in the PET image obtained using MR-based AC in comparison to CT-based AC. Right: Scatter plot, with the ideal diagonal in green. Hofmann M et al, MR-based PET Attenuation Correction Method and Validation Presented at IEEE NSS/MIC Conference 2007 & MRI-Based Attenuation Correction for PET/MRI: A Novel Approach Combining Pattern Recognition and Atlas Registration J Nucl Med (2008) 49: Oct. 28, 2008 INB Medicine (Brain Imaging Physics) Folie 19

20 Alternative methods to get attenuation maps Ongoing Work. Brain Extensive evaluation on 17 MR-CT pairs, 3 PET/CT-MR triplets with standard T1SE sequence on 1.5T. Ongoing work: 1) Optimization of Flash 3D and 2) Use of UTE sequences on 3T. MR Image Predicted PseudoCT Real CT PseudoCT vs CT prediction error of on average 84.7 HU per voxel. On a dataset of 17 MR-CT pairs, the average PseudoCT prediction error was HU. Whole Body Ongoing work: 1) Development on a probabilistic framework. 2) Optimization of dual echo protocol with less than 30s acquisition time per bed position, i.e. breath hold acquisition possible. Hofmann M et al, MR-based PET Attenuation Correction Initial Results for Whole Body Presented at IEEE NSS/MIC Conference 2008 in Dresden Oct. 28, 2008 INB Medicine (Brain Imaging Physics) Folie 20

21 Summary Two possible ways to get ACF: - Segmentation of the clinical MR images -- Differentiation between skull and cavities --- Knowledge based segmentation --- Usage of different MR pulse sequences (UTE) - Alternative methods -- based on measured transmission images (only brain) -- based on CT images (brain and whole body) Oct. 28, 2008 INB Medicine (Brain Imaging Physics) Folie 21

LSO PET/CT Pico Performance Improvements with Ultra Hi-Rez Option

LSO PET/CT Pico Performance Improvements with Ultra Hi-Rez Option LSO PET/CT Pico Performance Improvements with Ultra Hi-Rez Option Y. Bercier, Member, IEEE, M. Casey, Member, IEEE, J. Young, Member, IEEE, T. Wheelock, Member, IEEE, T. Gremillion Abstract-- Factors which

More information

MRI Summer Course Lab 2: Gradient Echo T1 & T2* Curves

MRI Summer Course Lab 2: Gradient Echo T1 & T2* Curves MRI Summer Course Lab 2: Gradient Echo T1 & T2* Curves Experiment 1 Goal: Examine the effect caused by changing flip angle on image contrast in a simple gradient echo sequence and derive T1-curves. Image

More information

Simultaneous Reconstruction of the Activity Image and Registration of the CT image in TOF-PET. Ahmadreza Rezaei, Johan Nuyts

Simultaneous Reconstruction of the Activity Image and Registration of the CT image in TOF-PET. Ahmadreza Rezaei, Johan Nuyts Simultaneous Reconstruction of the Activity Image and Registration of the CT image in TOF-PET Ahmadreza Rezaei, Johan Nuyts Activity Reconstruction & Attenuation Registration Attenuation Correction, Background

More information

Pitfalls and Remedies of MDCT Scanners as Quantitative Instruments

Pitfalls and Remedies of MDCT Scanners as Quantitative Instruments intensity m(e) m (/cm) 000 00 0 0. 0 50 0 50 Pitfalls and Remedies of MDCT Scanners as Jiang Hsieh, PhD GE Healthcare Technology University of Wisconsin-Madison Root-Causes of CT Number Inaccuracies Nature

More information

Technical and methodological aspects of PET/MR

Technical and methodological aspects of PET/MR Clin Transl Imaging (2013) 1:11 16 DOI 10.1007/s40336-013-0011-5 REVIEW ARTICLE Technical and methodological aspects of PET/MR Sibylle I. Ziegler Gaspar Delso Received: 19 December 2012 / Accepted: 1 February

More information

Chiara Secco. PET Performance measurements of the new LSO-Based Whole Body PET/CT. Scanner biograph 16 HI-REZ using the NEMA NU Standard.

Chiara Secco. PET Performance measurements of the new LSO-Based Whole Body PET/CT. Scanner biograph 16 HI-REZ using the NEMA NU Standard. Chiara Secco PET Performance measurements of the new LSO-Based Whole Body PET/CT Scanner biograph 16 HI-REZ using the NEMA NU 2-2001 Standard. INTRODUCTION Since its introduction, CT has become a fundamental

More information

PET: New Technologies & Applications, Including Oncology

PET: New Technologies & Applications, Including Oncology PET: New Technologies & Applications, Including Oncology, PhD, FIEEE Imaging Research Laboratory Department of Radiology University of Washington, Seattle, WA Disclosures Research Contract, GE Healthcare

More information

PET Performance Evaluation of MADPET4: A Small Animal PET Insert for a 7-T MRI Scanner

PET Performance Evaluation of MADPET4: A Small Animal PET Insert for a 7-T MRI Scanner PET Performance Evaluation of MADPET4: A Small Animal PET Insert for a 7-T MRI Scanner September, 2017 Results submitted to Physics in Medicine & Biology Negar Omidvari 1, Jorge Cabello 1, Geoffrey Topping

More information

12/21/2016. Siemens Medical Systems Research Agreement Philips Healthcare Research Agreement AAN and ASN Committees

12/21/2016. Siemens Medical Systems Research Agreement Philips Healthcare Research Agreement AAN and ASN Committees Joseph V. Fritz, PhD Nandor Pintor, MD Dent Neurologic Institute ASN 2017 Friday, January 20, 2017 Siemens Medical Systems Research Agreement Philips Healthcare Research Agreement AAN and ASN Committees

More information

Positron Emission Tomography - PET

Positron Emission Tomography - PET Positron Emission Tomography - PET Positron Emission Tomography Positron Emission Tomography (PET): Coincidence detection of annihilation radiation from positron-emitting isotopes followed by tomographic

More information

HISTORY. CT Physics with an Emphasis on Application in Thoracic and Cardiac Imaging SUNDAY. Shawn D. Teague, MD

HISTORY. CT Physics with an Emphasis on Application in Thoracic and Cardiac Imaging SUNDAY. Shawn D. Teague, MD CT Physics with an Emphasis on Application in Thoracic and Cardiac Imaging Shawn D. Teague, MD DISCLOSURES 3DR- advisory committee CT PHYSICS WITH AN EMPHASIS ON APPLICATION IN THORACIC AND CARDIAC IMAGING

More information

An innovative detector concept for hybrid 4D-PET/MRI Imaging

An innovative detector concept for hybrid 4D-PET/MRI Imaging Piergiorgio Cerello (INFN - Torino) on behalf of the 4D-MPET* project *4 Dimensions Magnetic compatible module for Positron Emission Tomography INFN Perugia, Pisa, Torino; Polytechnic of Bari; University

More information

PET/CT Instrumentation Basics

PET/CT Instrumentation Basics / Instrumentation Basics 1. Motivations for / imaging 2. What is a / Scanner 3. Typical Protocols 4. Attenuation Correction 5. Problems and Challenges with / 6. Examples Motivations for / Imaging Desire

More information

Using GATE to understand performance of a full-torso PET scanner

Using GATE to understand performance of a full-torso PET scanner University of Pennsylvania Using GATE to understand performance of a full-torso PET scanner Varsha Viswanath 1 Margaret E. Daube-Witherspoon 1, Matthew E. Werner 1, Suleman Surti 1, Andreia Trindade 2,

More information

Pulse Sequence Design and Image Procedures

Pulse Sequence Design and Image Procedures Pulse Sequence Design and Image Procedures 1 Gregory L. Wheeler, BSRT(R)(MR) MRI Consultant 2 A pulse sequence is a timing diagram designed with a series of RF pulses, gradients switching, and signal readout

More information

Medical Images Analysis and Processing

Medical Images Analysis and Processing Medical Images Analysis and Processing - 25642 Emad Course Introduction Course Information: Type: Graduated Credits: 3 Prerequisites: Digital Image Processing Course Introduction Reference(s): Insight

More information

Medical Imaging. X-rays, CT/CAT scans, Ultrasound, Magnetic Resonance Imaging

Medical Imaging. X-rays, CT/CAT scans, Ultrasound, Magnetic Resonance Imaging Medical Imaging X-rays, CT/CAT scans, Ultrasound, Magnetic Resonance Imaging From: Physics for the IB Diploma Coursebook 6th Edition by Tsokos, Hoeben and Headlee And Higher Level Physics 2 nd Edition

More information

Pulse Sequence Design Made Easier

Pulse Sequence Design Made Easier Pulse Sequence Design Made Easier Gregory L. Wheeler, BSRT(R)(MR) MRI Consultant gurumri@gmail.com 1 2 Pulse Sequences generally have the following characteristics: An RF line characterizing RF Pulse applications

More information

Image Quality/Artifacts Frequency (MHz)

Image Quality/Artifacts Frequency (MHz) The Larmor Relation 84 Image Quality/Artifacts (MHz) 42 ω = γ X B = 2πf 84 0.0 1.0 2.0 Magnetic Field (Tesla) 1 A 1D Image Magnetic Field Gradients Magnet Field Strength Field Strength / Gradient Coil

More information

Monte Carlo Simulation Study of a Dual-Plate PET Camera Dedicated to Breast Cancer Imaging

Monte Carlo Simulation Study of a Dual-Plate PET Camera Dedicated to Breast Cancer Imaging IEEE Nuclear Science Symposium Conference Record M-9 Monte Carlo Simulation Study of a Dual-Plate PET Camera Dedicated to Breast Cancer Imaging Jin Zhang, Member, IEEE, Peter D. Olcott, Member, IEEE, Angela

More information

UCLA UCLA Previously Published Works

UCLA UCLA Previously Published Works UCLA UCLA Previously Published Works Title Attenuation correction for small animal PET tomographs Permalink https://escholarship.org/uc/item/41n377p3 Journal Physics in Medicine and Biology, 5(8) ISSN

More information

X-rays in medical diagnostics

X-rays in medical diagnostics X-rays in medical diagnostics S.Dolanski Babić 2017/18. History W.C.Röntgen (1845-1923) discovered a new type of radiation Nature, Jan. 23. 1896.; Science, Feb.14. 1896. X- rays: Induced the ionization

More information

X-Ray-Based Attenuation Correction for Positron Emission Tomography/Computed Tomography Scanners

X-Ray-Based Attenuation Correction for Positron Emission Tomography/Computed Tomography Scanners X-Ray-Based Attenuation Correction for Positron Emission Tomography/Computed Tomography Scanners Paul E. Kinahan, Bruce H. Hasegawa, and Thomas Beyer A synergy of positron emission tomography (PET)/computed

More information

A NOVEL CONCEPT FOR A POSITRON EMISSION TOMOGRAPHY SCANNER

A NOVEL CONCEPT FOR A POSITRON EMISSION TOMOGRAPHY SCANNER A NOVEL CONCEPT FOR A POSITRON EMISSION TOMOGRAPHY SCANNER An Undergraduate Research Scholars Thesis by BRIAN KELLY, MATTHEW LEE ELLIOT LEVIN and JEENA KHATRI Submitted to Honors and Undergraduate Research

More information

Performance characterization of a novel thin position-sensitive avalanche photodiode-based detector for high resolution PET

Performance characterization of a novel thin position-sensitive avalanche photodiode-based detector for high resolution PET 2005 IEEE Nuclear Science Symposium Conference Record M11-126 Performance characterization of a novel thin position-sensitive avalanche photodiode-based detector for high resolution PET Jin Zhang, Member,

More information

Performance evaluation of a new highsensitivity time-of-flight clinical PET/CT system

Performance evaluation of a new highsensitivity time-of-flight clinical PET/CT system Huo et al. EJNMMI Physics (2018) 5:29 https://doi.org/10.1186/s40658-018-0229-4 EJNMMI Physics ORIGINAL RESEARCH Open Access Performance evaluation of a new highsensitivity time-of-flight clinical PET/CT

More information

Supplementary Figure 1

Supplementary Figure 1 Supplementary Figure 1 Left aspl Right aspl Detailed description of the fmri activation during allocentric action observation in the aspl. Averaged activation (N=13) during observation of the allocentric

More information

ISSN X CODEN (USA): PCHHAX. The role of dual spin echo in increasing resolution in diffusion weighted imaging of brain

ISSN X CODEN (USA): PCHHAX. The role of dual spin echo in increasing resolution in diffusion weighted imaging of brain Available online at www.derpharmachemica.com ISSN 0975-413X CODEN (USA): PCHHAX Der Pharma Chemica, 2016, 8(17):15-20 (http://derpharmachemica.com/archive.html) The role of in increasing resolution in

More information

Introduction. Chapter 16 Diagnostic Radiology. Primary radiological image. Primary radiological image

Introduction. Chapter 16 Diagnostic Radiology. Primary radiological image. Primary radiological image Introduction Chapter 16 Diagnostic Radiology Radiation Dosimetry I Text: H.E Johns and J.R. Cunningham, The physics of radiology, 4 th ed. http://www.utoledo.edu/med/depts/radther In diagnostic radiology

More information

Gradient Spoiling. Average balanced SSFP magnetization Reduce sensitivity to off-resonance. FFE, FISP, GRASS, GRE, FAST, Field Echo

Gradient Spoiling. Average balanced SSFP magnetization Reduce sensitivity to off-resonance. FFE, FISP, GRASS, GRE, FAST, Field Echo Gradient Spoiling Average balanced SSFP magnetization Reduce sensitivity to off-resonance FFE, FISP, GRASS, GRE, FAST, Field Echo 1 Gradient-Spoiled Sequence (GRE, FFE, FISP, GRASS) RF TR G z G y G x Signal

More information

2014 M.S. Cohen all rights reserved

2014 M.S. Cohen all rights reserved 2014 M.S. Cohen all rights reserved mscohen@g.ucla.edu IMAGE QUALITY / ARTIFACTS SYRINGOMYELIA Source http://gait.aidi.udel.edu/res695/homepage/pd_ortho/educate/clincase/syrsco.htm Surgery is usually recommended

More information

Yinsheng Li 1, Peter Bannas 2, M.D., Perry Pickhardt M.D. 2, Meghan Lubner M.D. 2, Ke Li Ph.D. 1,2, and Guang-Hong Chen Ph.D. 1,2

Yinsheng Li 1, Peter Bannas 2, M.D., Perry Pickhardt M.D. 2, Meghan Lubner M.D. 2, Ke Li Ph.D. 1,2, and Guang-Hong Chen Ph.D. 1,2 Yinsheng Li 1, Peter Bannas 2, M.D., Perry Pickhardt M.D. 2, Meghan Lubner M.D. 2, Ke Li Ph.D. 1,2, and Guang-Hong Chen Ph.D. 1,2 1. Department of Medical Physics, University of Wisconsin-Madison 2. Department

More information

LaBr 3 :Ce, the latest crystal for nuclear medicine

LaBr 3 :Ce, the latest crystal for nuclear medicine 10th Topical Seminar on Innovative Particle and Radiation Detectors 1-5 October 2006 Siena, Italy LaBr 3 :Ce, the latest crystal for nuclear medicine Roberto Pani On behalf of SCINTIRAD Collaboration INFN

More information

Slide 1. Slide 2. Slide 3 ACR CT Accreditation. Multi-Slice CT Artifacts and Quality Control. What are the rules or recommendations for CT QC?

Slide 1. Slide 2. Slide 3 ACR CT Accreditation. Multi-Slice CT Artifacts and Quality Control. What are the rules or recommendations for CT QC? Slide 1 Multi-Slice CT Artifacts and Quality Control Dianna Cody, Ph.D. Chief, Radiologic Physics UT MD Anderson Cancer Center Houston, TX Slide 2 What are the rules or recommendations for CT QC? AAPM

More information

SPECT Reconstruction & Filtering

SPECT Reconstruction & Filtering SPECT Reconstruction & Filtering Goals Understand the basics of SPECT Reconstruction Filtered Backprojection Iterative Reconstruction Make informed choices on filter selection and settings Pre vs. Post

More information

diagnostic examination

diagnostic examination RADIOLOGICAL PHYSICS 2011 Raphex diagnostic examination Adel A. Mustafa, Ph.D., Editor PUBLISHED FOR: RAMPS (Radiological and Medical Physics Society of New York) preface The RAPHEX Diagnostic exam 2011

More information

Combined micropet /MR System: Performance Assessment of the Full PET Ring with Split Gradients 4.8

Combined micropet /MR System: Performance Assessment of the Full PET Ring with Split Gradients 4.8 Combined micropet /MR System: Performance Assessment of the Full PET Ring with Split Gradients 4.8 UNIVERSITY OF CAMBRIDGE 1.2 Rob C. Hawkes 1, Tim D. Fryer 1, Alun J. Lucas 1,2, Stefan B. Siegel 3, Richard

More information

PD233: Design of Biomedical Devices and Systems

PD233: Design of Biomedical Devices and Systems PD233: Design of Biomedical Devices and Systems (Lecture-8 Medical Imaging Systems) (Imaging Systems Basics, X-ray and CT) Dr. Manish Arora CPDM, IISc Course Website: http://cpdm.iisc.ac.in/utsaah/courses/

More information

Explain what is meant by a photon and state one of its main properties [2]

Explain what is meant by a photon and state one of its main properties [2] 1 (a) A patient has an X-ray scan taken in hospital. The high-energy X-ray photons interact with the atoms inside the body of the patient. Explain what is meant by a photon and state one of its main properties....

More information

2/14/2019. Nuclear Medicine Artifacts. Symmetric energy windows

2/14/2019. Nuclear Medicine Artifacts. Symmetric energy windows Nuclear Medicine Artifacts SCPMG Medical Imaging Technology & Informatics Medical Physics Group Brian Helbig, MS, DABR 1 2 Symmetric energy windows 3 1 Dynamic clinical study Energy peak shift Electrical

More information

Celesteion Time-of-Flight Technology

Celesteion Time-of-Flight Technology Celesteion Time-of-Flight Technology Bing Bai, PhD Clinical Sciences Manager, PET/CT Canon Medical Systems USA Introduction Improving the care for every patient while providing a high standard care to

More information

MRI Metal Artifact Reduction

MRI Metal Artifact Reduction MRI Metal Artifact Reduction PD Dr. med. Reto Sutter University Hospital Balgrist Zurich University of Zurich OUTLINE Is this Patient suitable for MR Imaging? Metal artifact reduction Is this Patient suitable

More information

Assessment of Image Quality of a PET/CT scanner for a Standarized Image situation Using a NEMA Body Phantom

Assessment of Image Quality of a PET/CT scanner for a Standarized Image situation Using a NEMA Body Phantom Assessment of Image Quality of a PET/CT scanner for a Standarized Image situation Using a NEMA Body Phantom The impact of Different Image Reconstruction Parameters on Image quality by QUAYE MICHAEL This

More information

Usefulness of noise adaptive non-linear Gaussian filter in FDG-PET study

Usefulness of noise adaptive non-linear Gaussian filter in FDG-PET study ORIGINAL ARTICLE Annals of Nuclear Medicine Vol. 19, No. 6, 469 477, 2005 Usefulness of noise adaptive non-linear Gaussian filter in FDG-PET study Makoto NAGAYOSHI,*, ** Kenya MURASE,* Kouichi FUJINO,**

More information

Suppression of metal artifacts using image-based monoenergetic DECT imaging

Suppression of metal artifacts using image-based monoenergetic DECT imaging Suppression of metal artifacts using image-based monoenergetic DECT imaging Poster No.: C-0519 Congress: ECR 2011 Type: Scientific Paper Authors: B. Krauss, B. Schmidt, M. Sedlmair, T. Flohr; Forchheim/DE

More information

Standards for Imaging Endpoints in Clinical Trials: Standardization and Optimization of Image Acquisitions: Magnetic Resonance

Standards for Imaging Endpoints in Clinical Trials: Standardization and Optimization of Image Acquisitions: Magnetic Resonance FDA Workshop April 13, 2010 Standards for Imaging Endpoints in Clinical Trials: Standardization and Optimization of Image Acquisitions: Magnetic Resonance Edward F. Jackson, PhD Professor and Chief, Section

More information

Imaging the brain at ultra-high resolution using 3D FatNavs

Imaging the brain at ultra-high resolution using 3D FatNavs Imaging the brain at ultra-high resolution using 3D FatNavs Daniel Gallichan Centre d Imagerie BioMédicale EPFL, Lausanne, Switzerland Overview Introduction How motion affects MRI scans Ways we can track

More information

Fundamentals of Positron Emission Tomography (PET)

Fundamentals of Positron Emission Tomography (PET) Fundamentals of Positron Emission Tomography (PET) NPRE 435, Principles of Imaging with Ionizing Radiation, Fall 2017 Content Fundamentals of PET Camera & Detector Design Real World Considerations Performance

More information

Changing the Shape of Nuclear Medicine

Changing the Shape of Nuclear Medicine TRUTH IN IMAGING Changing the Shape of Nuclear Medicine Multi-Purpose SPECT Scanner Nothing Gets Closer Introducing 360 Body Contour Scanning With 360 degree detector coverage, and unique proximity sensors

More information

Research Support. Dual-Source CT: What is it and How Do I Test it? Cynthia H. McCollough, Ph.D.

Research Support. Dual-Source CT: What is it and How Do I Test it? Cynthia H. McCollough, Ph.D. Dual-Source CT: What is it and How Do I Test it? Cynthia H. McCollough, Ph.D. CT Clinical Innovation Center Department of Radiology Mayo Clinic College of Medicine Rochester, MN Research Support National

More information

Fig. 1

Fig. 1 PhysicsAndMathsTutor.com 1 1. Fig. 1 shows data for the intensity of a parallel beam of X-rays after penetration through varying thicknesses of a material. intensity / MW m 2 thickness / mm 0.91 0.40 0.69

More information

Detector technology challenges for nuclear medicine and PET

Detector technology challenges for nuclear medicine and PET Nuclear Instruments and Methods in Physics Research A 513 (2003) 1 7 Detector technology challenges for nuclear medicine and PET Paul K. Marsden Guy s and St. Thomas Clinical PET Centre, King s College

More information

Initial Certification

Initial Certification Initial Certification Nuclear Medical Physics (NMP) Study Guide Part 2 Content Guide and Sample Questions The content of all ABR exams is determined by a panel of experts who select the items based on

More information

Simultaneous Multi-Slice (Slice Accelerated) Diffusion EPI

Simultaneous Multi-Slice (Slice Accelerated) Diffusion EPI Simultaneous Multi-Slice (Slice Accelerated) Diffusion EPI Val M. Runge, MD Institute for Diagnostic and Interventional Radiology Clinics for Neuroradiology and Nuclear Medicine University Hospital Zurich

More information

Lunar Technology Advantages

Lunar Technology Advantages Lunar Technology Advantages DXA stands for Dual-Energy X-ray Absorptiometry. It is a measurement method that uses the differences in the absorption of high energy and low energy X-ray photons by different

More information

... In vivo imaging in Nuclear Medicine. 1957: Anger camera (X;Y) X Y

... In vivo imaging in Nuclear Medicine. 1957: Anger camera (X;Y) X Y József Varga, PhD EMISSION IMAGING BASICS OF QUANTIFICATION Imaging devices Aims of image processing Reconstruction University of Debrecen Department of Nuclear Medicine. In vivo imaging in Nuclear Medicine

More information

Improve Image Quality of Transversal Relaxation Time PROPELLER and FLAIR on Magnetic Resonance Imaging

Improve Image Quality of Transversal Relaxation Time PROPELLER and FLAIR on Magnetic Resonance Imaging Journal of Physics: Conference Series PAPER OPEN ACCESS Improve Image Quality of Transversal Relaxation Time PROPELLER and FLAIR on Magnetic Resonance Imaging To cite this article: N Rauf et al 2018 J.

More information

NEMA and clinical evaluation of a novel brain PET-CT scanner

NEMA and clinical evaluation of a novel brain PET-CT scanner Journal of Nuclear Medicine, published on December 23, 2015 as doi:10.2967/jnumed.115.159723 NEMA and clinical evaluation of a novel brain PET-CT scanner Kira S. Grogg 1, Terrence Toole 2, Jinsong Ouyang

More information

Study of Silicon Photomultipliers for Positron Emission Tomography (PET) Application

Study of Silicon Photomultipliers for Positron Emission Tomography (PET) Application Study of Silicon Photomultipliers for Positron Emission Tomography (PET) Application Eric Oberla 5 June 29 Abstract A relatively new photodetector, the silicon photomultiplier (SiPM), is well suited for

More information

H 2 O and fat imaging

H 2 O and fat imaging H 2 O and fat imaging Xu Feng Outline Introduction benefit from the separation of water and fat imaging Chemical Shift definition of chemical shift origin of chemical shift equations of chemical shift

More information

Image Analysis Using R

Image Analysis Using R Image Analysis Using R Chris Campbell LondonR - 13th July 2010 Steps to image analysis Image capture Clean image/reduce noise Extract information Analyze information Image Capture Light Photography Light

More information

Unit thickness. Unit area. σ = NΔX = ΔI / I 0

Unit thickness. Unit area. σ = NΔX = ΔI / I 0 Unit thickness I 0 ΔI I σ = ΔI I 0 NΔX = ΔI / I 0 NΔX Unit area Δx Average probability of reaction with atom for the incident photons at unit area with the thickness of Delta-X Atom number at unit area

More information

Biomedical Imaging Informatics

Biomedical Imaging Informatics Biomedical Imaging Informatics Daniel L. Rubin, Hayit Greenspan, and James F. Brinkley 9 After reading this chapter, you should know the answers to these questions: What makes images a challenging type

More information

Methods. Experimental Stimuli: We selected 24 animals, 24 tools, and 24

Methods. Experimental Stimuli: We selected 24 animals, 24 tools, and 24 Methods Experimental Stimuli: We selected 24 animals, 24 tools, and 24 nonmanipulable object concepts following the criteria described in a previous study. For each item, a black and white grayscale photo

More information

The SENSE Ghost: Field-of-View Restrictions for SENSE Imaging

The SENSE Ghost: Field-of-View Restrictions for SENSE Imaging JOURNAL OF MAGNETIC RESONANCE IMAGING 20:1046 1051 (2004) Technical Note The SENSE Ghost: Field-of-View Restrictions for SENSE Imaging James W. Goldfarb, PhD* Purpose: To describe a known (but undocumented)

More information

Investigation of low noise, low cost readout electronics for high sensitivity PET systems based on Avalanche Photodiode arrays

Investigation of low noise, low cost readout electronics for high sensitivity PET systems based on Avalanche Photodiode arrays Investigation of low noise, low cost readout electronics for high sensitivity PET systems based on Avalanche Photodiode arrays Frezghi Habte, Member, IEEE and Craig S.Levin, Member, IEEE Abstract A compact,

More information

MC SIMULATION OF SCATTER INTENSITIES IN A CONE-BEAM CT SYSTEM EMPLOYING A 450 kv X-RAY TUBE

MC SIMULATION OF SCATTER INTENSITIES IN A CONE-BEAM CT SYSTEM EMPLOYING A 450 kv X-RAY TUBE MC SIMULATION OF SCATTER INTENSITIES IN A CONE-BEAM CT SYSTEM EMPLOYING A 450 kv X-RAY TUBE A. Miceli ab, R. Thierry a, A. Flisch a, U. Sennhauser a, F. Casali b a Empa - Swiss Federal Laboratories for

More information

Photomultiplier Tube

Photomultiplier Tube Nuclear Medicine Uses a device known as a Gamma Camera. Also known as a Scintillation or Anger Camera. Detects the release of gamma rays from Radionuclide. The radionuclide can be injected, inhaled or

More information

Noise Characteristics of the FORE+OSEM(DB) Reconstruction Method for the MiCES PET Scanner

Noise Characteristics of the FORE+OSEM(DB) Reconstruction Method for the MiCES PET Scanner Noise Characteristics of the FORE+OSEM(DB) Reconstruction Method for the MiCES PET Scanner Kisung Lee, Member, IEEE, Paul E. Kinahan, Senior Member, Robert S. Miyaoka, Member, IEEE, Jeffrey A. Fessler,

More information

Clear delineation of optic radiation and very small vessels using phase difference enhanced imaging (PADRE)

Clear delineation of optic radiation and very small vessels using phase difference enhanced imaging (PADRE) Clear delineation of optic radiation and very small vessels using phase difference enhanced imaging (PADRE) Poster No.: C-2459 Congress: ECR 2010 Type: Scientific Exhibit Topic: Neuro Authors: T. Yoneda,

More information

Performance Evaluation of SiPM Detectors for PET Imaging in the Presence of Magnetic Fields

Performance Evaluation of SiPM Detectors for PET Imaging in the Presence of Magnetic Fields 2008 IEEE Nuclear Science Symposium Conference Record M02-4 Performance Evaluation of SiPM Detectors for PET Imaging in the Presence of Magnetic Fields Samuel España, Student Member, IEEE, Gustavo Tapias,

More information

2010 Philips BrightView XCT SPECT/CT

2010 Philips BrightView XCT SPECT/CT 2010 Philips BrightView XCT SPECT/CT Unit was purchased from Philips training center in 2015. Installed but never been used by the current facility. (Scroll for pictures) BrightView XCT Camera with PinPoint

More information

2594 IEEE TRANSACTIONS ON NUCLEAR SCIENCE, VOL. 56, NO. 5, OCTOBER /$ IEEE

2594 IEEE TRANSACTIONS ON NUCLEAR SCIENCE, VOL. 56, NO. 5, OCTOBER /$ IEEE 2594 IEEE TRANSACTIONS ON NUCLEAR SCIENCE, VOL. 56, NO. 5, OCTOBER 2009 Investigation of Depth of Interaction Encoding for a Pixelated LSO Array With a Single Multi-Channel PMT Yongfeng Yang, Member, IEEE,

More information

1. Patient size AEC. Large Patient High ma. Small Patient Low ma

1. Patient size AEC. Large Patient High ma. Small Patient Low ma Comparison of the function and performance of CT AEC systems CTUG meeting by Emily Field Trainee clinical scientist 14 th th Breakdown CT Automatic Exposure Control (AEC) Background Project Description

More information

SIGNA Explorer Lift revives our MR

SIGNA Explorer Lift revives our MR Seiji Shiotani, MD, PhD Seirei Fuji Hospital in Fuji City, Shizuoka, Japan Masayoshi Sugimura Seirei Fuji Hospital in Fuji City, Shizuoka, Japan SIGN Explorer Lift revives our MR The clinical usefulness

More information

Optimization of Energy Modulation Filter for Dual Energy CBCT Using Geant4 Monte-Carlo Simulation

Optimization of Energy Modulation Filter for Dual Energy CBCT Using Geant4 Monte-Carlo Simulation Original Article PROGRESS in MEDICAL PHYSICS 27(3), Sept. 2016 http://dx.doi.org/10.14316/pmp.2016.27.3.125 pissn 2508-4445, eissn 2508-4453 Optimization of Energy Modulation Filter for Dual Energy CBCT

More information

MRI imaging in neuroscience Dr. Thom Oostendorp Lab class: 2 hrs

MRI imaging in neuroscience Dr. Thom Oostendorp Lab class: 2 hrs MRI imaging in neuroscience Dr. Thom Oostendorp Lab class: 2 hrs 1 Introduction In tomographic imaging techniques, such as MRI, a certain tissue property within a slice is imaged. For each voxel (volume

More information

MPPC and Liquid Xenon technologies from particle physics to medical imaging

MPPC and Liquid Xenon technologies from particle physics to medical imaging CANADA S NATIONAL LABORATORY FOR PARTICLE AND NUCLEAR PHYSICS Owned and operated as a joint venture by a consortium of Canadian universities via a contribution through the National Research Council Canada

More information

Primer on molecular imaging technology

Primer on molecular imaging technology Primer on molecular imaging technology Craig S. Levin Division of Nuclear Medicine, Department of Radiology and Molecular Imaging Program at Stanford (MIPS), Stanford University School of Medicine, 300

More information

Multi-channel SQUID-based Ultra-Low Field Magnetic Resonance Imaging in Unshielded Environment

Multi-channel SQUID-based Ultra-Low Field Magnetic Resonance Imaging in Unshielded Environment Multi-channel SQUID-based Ultra-Low Field Magnetic Resonance Imaging in Unshielded Environment Andrei Matlashov, Per Magnelind, Shaun Newman, Henrik Sandin, Algis Urbaitis, Petr Volegov, Michelle Espy

More information

Digital Imaging CT & MR

Digital Imaging CT & MR Digital Imaging CT & MR January 22, 2008 Digital Radiography, CT and MRI generate images in a digital format What is a Digital Image? A digital image is made up of picture elements, pixels row by column

More information

New Technology in Nuclear Medicine

New Technology in Nuclear Medicine New Technology in Nuclear Medicine Reed G. Selwyn, PhD, DABR Vice Chair of Research & Imaging Sciences Associate Professor and Chief, Medical Physics Dept. of Radiology, University of New Mexico Objectives

More information

K-edge subtraction X-ray imaging with a pixellated spectroscopic detector

K-edge subtraction X-ray imaging with a pixellated spectroscopic detector K-edge subtraction X-ray imaging with a pixellated spectroscopic detector Silvia Pani Department of Physics, University of Surrey Summary Hyperspectral imaging K-edge subtraction X-ray imaging for mammography

More information

Investigation of Solid-State Photomultipliers for Positron Emission Tomography Scanners

Investigation of Solid-State Photomultipliers for Positron Emission Tomography Scanners Journal of the Korean Physical Society, Vol. 50, No. 5, May 2007, pp. 1332 1339 Investigation of Solid-State Photomultipliers for Positron Emission Tomography Scanners Jae Sung Lee Department of Nuclear

More information

NM Module Section 2 6 th Edition Christian, Ch. 3

NM Module Section 2 6 th Edition Christian, Ch. 3 NM 4303 Module Section 2 6 th Edition Christian, Ch. 3 Gas Filled Chamber Voltage Gas filled chamber uses Hand held detectors cutie pie Geiger counter Dose calibrators Cutie pie Chamber voltage in Ionization

More information

Thermionic x-ray. Alternative technologies. Electron Field Emission. CNT Based Field Emission X-Ray Source

Thermionic x-ray. Alternative technologies. Electron Field Emission. CNT Based Field Emission X-Ray Source Energy Level (ev) Multi-beam x-ray source array based on carbon nanotube field emission O. Zhou, JP Lu, X. Calderon-Colon, X. Qian, G. Yang, G. Cao, E. Gidcumb, A. Tucker, J. Shan University of North Carolina

More information

Design of a Static Full-Ring Multi-Pinhole Collimator for Brain SPECT

Design of a Static Full-Ring Multi-Pinhole Collimator for Brain SPECT Design of a Static Full-Ring Multi-Pinhole Collimator for Brain SPECT Karen Van Audenhaege, Student Member, IEEE, Roel Van Holen, Member, IEEE, Karel Deprez, Joel S. Karp, Senior Member, IEEE, Scott Metzler,

More information

Factors Affecting the resolution of SPECT Imaging. h.

Factors Affecting the resolution of SPECT Imaging. h. Factors Affecting the resolution of SPECT Imaging H. E. Mostafa *1, H. A. Ayoub 2 and Sh.Magraby 1 1 Kasr El-Ini Center for Oncology, Cairo University, 2 Faculty of Science, Suez Canal University hayamayoub@yahoo.com

More information

COMPUTED TOMOGRAPHY 1

COMPUTED TOMOGRAPHY 1 COMPUTED TOMOGRAPHY 1 Why CT? Conventional X ray picture of a chest 2 Introduction Why CT? In a normal X-ray picture, most soft tissue doesn't show up clearly. To focus in on organs, or to examine the

More information

Development of PET using 4 4 Array of Large Size Geiger-mode Avalanche Photodiode

Development of PET using 4 4 Array of Large Size Geiger-mode Avalanche Photodiode 2009 IEEE Nuclear Science Symposium Conference Record M09-8 Development of PET using 4 4 Array of Large Size Geiger-mode Avalanche Photodiode K. J. Hong, Y. Choi, J. H. Kang, W. Hu, J. H. Jung, B. J. Min,

More information

Radionuclide Imaging MII Single Photon Emission Computed Tomography (SPECT)

Radionuclide Imaging MII Single Photon Emission Computed Tomography (SPECT) Radionuclide Imaging MII 3073 Single Photon Emission Computed Tomography (SPECT) Single Photon Emission Computed Tomography (SPECT) The successful application of computer algorithms to x-ray imaging in

More information

Performance Characteristics of a State of the Art Preclinical PET/SPECT/CT Scanner

Performance Characteristics of a State of the Art Preclinical PET/SPECT/CT Scanner Performance Characteristics of a State of the Art Preclinical PET/SPECT/CT Scanner Nya Mehnwolo Boayue 1 Samuel Kuttner 1 1 Center for Diagnostic Physics University Hospital of North-Norway Medfys, 2016

More information

Importance of Precise Timing for Medical Diagnostic Devices

Importance of Precise Timing for Medical Diagnostic Devices Importance of Precise Timing for Medical Diagnostic Devices M.C.S. Williams a,b, A. Zichichi a,b,c and the CERN-Bologna group. a CERN Geneva, Switzerland b INFN and Dipartimento di Fisica e Astronomia,

More information

Designing an MR compatible Time of Flight PET Detector Floris Jansen, PhD, Chief Engineer GE Healthcare

Designing an MR compatible Time of Flight PET Detector Floris Jansen, PhD, Chief Engineer GE Healthcare GE Healthcare Designing an MR compatible Time of Flight PET Detector Floris Jansen, PhD, Chief Engineer GE Healthcare There is excitement across the industry regarding the clinical potential of a hybrid

More information

Imaging with FDG PET is a valuable technique for tumor

Imaging with FDG PET is a valuable technique for tumor Noise Reduction in Oncology FDG PET Images by Iterative Reconstruction: A Quantitative Assessment Cyril Riddell, Richard E. Carson, Jorge A. Carrasquillo, Steven K. Libutti, David N. Danforth, Millie Whatley,

More information

Simulation and evaluation of a cost-effective high-performance brain PET scanner.

Simulation and evaluation of a cost-effective high-performance brain PET scanner. Research Article http://www.alliedacademies.org/biomedical-imaging-and-bioengineering/ Simulation and evaluation of a cost-effective high-performance brain PET scanner. Musa S Musa *, Dilber U Ozsahin,

More information

An Introduction: Radon Transform, X-ray Transform, Inverse Problems

An Introduction: Radon Transform, X-ray Transform, Inverse Problems Other applications: SPECT and Attenuated An Introduction:, X-ray Transform, TING ZHOU Northeastern University January 9, 2018 Other applications: SPECT and Attenuated Outline 1 Course Information (syllabus,

More information

(C) 2018 BrainAnalyze S.A.S 25 rue du Maréchal Foch, Versailles, France Mob:

(C) 2018 BrainAnalyze S.A.S 25 rue du Maréchal Foch, Versailles, France Mob: BrainAnalyst Innovative Solution for Neuro-Image Processings Multi-Environment, Multi-Parametric, Fast, Accurate, Automated, Manufacturer Independent BrainAnalyst is the Complete and Innovative Neuro-Image

More information

25 CP Generalize Concepts in Abstract Multi-dimensional Image Model Component Semantics Page 1

25 CP Generalize Concepts in Abstract Multi-dimensional Image Model Component Semantics Page 1 25 CP-1390 - Generalize Concepts in Abstract Multi-dimensional Image Model Component Semantics Page 1 1 STATUS Letter Ballot 2 Date of Last Update 2014/09/08 3 Person Assigned David Clunie 4 mailto:dclunie@dclunie.com

More information

160-slice CT SCANNER / New Standard for the Future

160-slice CT SCANNER / New Standard for the Future TECHNOLOGY HISTORY For over 130 years, Toshiba has been a world leader in developing technology to improve the quality of life. Our 50,000 global patents demonstrate a long, rich history of leading innovation.

More information