/02/$ IEEE 1109
|
|
- Derick White
- 5 years ago
- Views:
Transcription
1 Performance Measurements for the GSO-based Brain PET Camera (G-PET) S. Surtil Student Member, IEEE) J.S. Karpl Muchllchncr Senior Member, IEEE) L.-E. Adam1 * Senior Member. IEEE AbstractPerformance measurements on the high sensitivity, high resolution G-PET scanner have been completed. This scanner a diameter of 42.cm and axial field-ofview of 25.6cm was designed for brain receptor imaging, as well as regular clinical and blood perfusion studies in the brain. The transverse and axial resolution near the center are 4. and 5. mm (fwtm of 8. and 1. mm), respectively. At a radial offset of 1Ocm these numbers deteriorate by less than 12%. The absolute sensitivity of this scanner measured a 7 cm long line souxe is 4.57 cps/kbq. Scatter fraction measured a line source in a 2cm diameter by 19cm long cylinder is 34.5%. For the same cylinder, the peak NEC rate is measured to be 75 kcps at an activity concentration of kbq/ml (.35 pci/cc), while the peak true coincidence rate is 2 kcps. Image contrast measured six s placed in the cylinder an activity concentration ratio of 8:l is 21-24% for the smallest (diameter of 1mm). However, due to the loss of counts near the edges, the image contrast for this is as high as 36% when using a region half the size of the diameter and a 3 minutes acquisition time. We also show results from the 3D Hoffman brain phantom as well as 18F-FDG patient scans. These images illustrate the high visual quality of images acquired on the G-PET scanner. I. INTRODTJCTION A high sensitivity. high resolution PET scarum using a gatlolir~ium orthosilicatc (GSO) crystal basctl Anger-logic tlctcctor [I] has bccri tlcvclopctl for brain imaging (G-PET) [a]. The irltcrdctl applications inclutlc rcscarch stutlics using llc-taggctl radio-ligaritls for ricuro-rcccptor irnagirlg ~~rotocols. as well as 18F-taggctl ligards and O-water brain stutlics (Figure 1). This scarmcr was tlcsigrlctl to hvc irnprovctl spatial rcsolutiori. high scrisitivity. arid high count-rate capability cornparctl to the curmnt NaI(Tl)l-~~tl HEAD PENN-PET (H-PET) scarmcr [3]. The GPET scmrm uses 4 x 4 x 1Omm GSO crystals. which have an cncrgy rcsolutior~ of 9-1X at 511 kcv. rclativcly high tlcrisity (or liricar attcmmtiori cocfficicrit j1). arid short tlccay tirnc. Thcsc physical ~mrarnctcrs of GSO cornparc favorably those of NaI(T1) and LSO. The scarmcr has a tliarnctcr of 42 cm. an axial ficltl-of-view (FOV) of 25.6 cm. and it opcratcs only in the 3D or volume imaging rnotlc (no scpta). The G-PET scmrm will bc critical for study of structures which fall blow tlic tlirmlioltl of most corivcritiorml PET tcchology. while its high count-rate capability will bc irriportarit for study of physiological cffccts fast temporal tlyriamics. The authors arc the ldcpartrr~cr~t Pennsylvania, Philadelphia, PA 1914 phia, PA surti~rad.upcnn.cdu and of Radiology, Lnivcrsity of ADAC ITGiLl, Philadcl- Fig. 1. The Hospital G-PET of the Member, IEEE) G. brain scanner installed in the Lnivcrsity of Pennsylvania. II. SCANNER PET Ccntcr at the DIXSIGN A schematic tliagrarn of the scarmcr s $mton tlctcctiorl systcrn is shown in Figure 2. A total of 1856 crystals (32 colurrms by 58 rows) and rnrn tliarnctcr PMTs (36 colurriris by 8 rows in a licxagorial lattice) couplctl to a siriglc curvctl lightguitlc arc usctl. The lightguitlc thickness is 1.9 cm 5 rnrn tlccp slots cut in both tlic trarisvcrsc arid axial tlirncrlsiorls. The optimization of the lightguitlc emthccs tlic best crystal tliscrirniriatiori using a local wciglitctl ccntmitl calculatior~ [I]. Imrt of the scarmcr is 3 cm in tliarnctcr. a 2.5 cm thick lcatl sitlc shicltl usctl to rcjcct cvcnts from activity outsitlc the FOV. An q-amp basctl ydsc shapin, D circuit has bccri tlcvclopctl to rnakc the GSO signal symmetric. resulting in irn~mnml samplirig (digitization) of tlic signal by tlic flash analog-to-tligital-convcrtc~s (ADCS) running at a clock spcctl of 5 MHz. Optimal sampling of the signal is rquird to acliicvc good cricrgy rcsolutiori at iritcgmtiori tirncs of ris. Wc use a triggcririg sclicrric to first tlctcrmiric if an iritcractiori is a coiricitlcrit cvcrit. bcforc ~~crformirig any Imsition calculatior~s. To keep the clcctronics sirnplc. PMT signals arc groupctl into small trigger charmcls which liavc bccri optirnizctl to rcthcc tlic tri ggcririg tlcatltirnc by using a high trigger cncrgy tlmdmltl [4]. Each trigger elmml currently cornpriscs of three acljaccrlt colurrms of PMTs a oric colurriri ovcrlap bctwccri acljaccrit trigger charincls. Overall. the G-PET scarmcr has 18 trigger charmcls 24 PMTs ~)cr trigger charmcl. With this ammgcmcr~t /2/$ IEEE 119
2 Fig. 2. Schematic of G-PET scanner (transaxial cross-section), detector diameter of 42 cm and axial field-of-view of 25.6 cm. we can set the trigger channel threshold at 4 kev, allowing a significant reduction in the number of non-photopeak events triggering the electronics. Additionally, we have built in the capability to divide each trigger channel axially into two equal halves (five rows each an overlap of two rows) which would result in a doubling of the total number of trigger channels to 36. Once implemented in the future, this could potentially reduce the scanner deadtime even further. B. Data Processing and Image Reconstruction Three types of corrections are performed before the data are binned into a sinogram. These are: PMT gain matching, energy correction, and distortion removal. Before energy and position calculations, the digitally integrated PMT signals are corrected for PMT gain variations using a lookup table. The PMT gain matching is essential in order to achieve optimum crystal separation as well as good energy resolution. Once the position and deposited energy of an interaction have been calculated, a position-dependent energy correction is performed via a lookup table. Energy correction compensates mainly for systematic variations in the light collection from crystals near a PMT center compared to those crystals near the PMT edge. In principle, this correction also accounts for variations in the light output of individual crystals. However, our measurements have shown that the standard deviation of the light output of the 1856 crystals used in G-PET is only 6.8%, and during assembly, care was taken to use crystals of similar light outputs next to each other. The final data correction involves removal of distortion in the calculated position due to the use of a local centroid algorithm, and the eventual placement of the event in a real crystal position on the scanner. For this purpose, we developed an automated search algorithm which detects the minima in a flood histogram, defines the crystal boundaries in it, and eventually assigns a real position to all events which occur in each boundary region. After all these corrections have been performed, the acquired data from the scanner are rebinned into 32 angles (32 crystals/row), rays, upto 15 out-of-plane tilts, and 115 axial slices. The radial samples are 4.4 mm apart in the center of the scanner and closer together towards the edge (4.4 mm is the crystal-to-crystal spacing along the scanner circumference). The axial pitch of the 58 crystal rows is also 4.4 mm; thus the slice separation is 2.2 mm. Transverse interleaving is performed to improve the radial sampling; the 32 angles rays are resorted into 16 angles 125 rays. The radial bins are now 2.2 mm apart in the center of the scanner (closer together at the edge). In order to remove the non-uniform radial sampling and to generate a sinogram compatible existing software, data interpolation is performed prior to further data corrections (normalization, scatter, and attenuation) and reconstruction. During interpolation, the 125 radial bins (unevenly sampled) are interpolated into 256, 1 mm (evenly sampled) bins; 16 angles are interpolated into 192 angles and 115 slices are interpolated into 128, 2 mm slices. This step may be eliminated in the future to avoid loss of resolution associated interpolation. III. Scanner Performance Measurements Performance measurements were done on the G-PET scanner following the procedure outlined in the new NEMA NU2-21 standards. A. Spatial Resolution Spatial resolution measurements were performed using a point source of 18 F activity in a thin glass capillary tube an inner diameter of less than 1 mm. The axial length of the point source was also kept to less than 1 mm. According to NEMA requirements, measurements were performed at radial positions of 1 and 1 cm, as well as an in-between radial position of 7 cm. The acquired sinograms were reconstructed using Fourier rebinning followed by filtered back projection a ramp filter out any smoothing. The image was reconstructed into a field-ofview array where the pixel size is 1 mm in the transverse and 2 mm in the axial directions. For comparison purposes we also reconstructed the image into a field-of-view (pixel size of 2 mm) using 3D RAMLA iterative reconstruction [5]. Figure 3 summarizes the resolution results. From Figure 3 we clearly see that the spatial resolution of the G-PET scanner is 4 mm in the transverse direction after Fourier rebinning and FBP. This resolution worsens by about 12% due to depth-of-interaction effects at radial distance of 1 cm. The results from 3D RAMLA reconstruction are generally worse by about 1 mm mainly due to the optimization of the reconstruction parameters for clinical imaging situations (and not point sources in air), where reduced noise in the image is desired. Only one iteration was performed in the RAMLA reconstruction a lambda value of.24 and a blob radius of 2.5. The RAMLA results clearly show that the significant deterioration of axial resolution at radial distance of 1 cm (after FBP), is an artifact arising due to the Fourier rebinning 111
3 FWHM [mm] FWTM [mm] Transverse (FBP) Transverse (3D RAMLA) Axial (FBP) Axial (3D RAMLA) Radial position of source [cm] Transverse (FBP) Transverse (3D RAMLA) Axial (FBP) Axial (3D RAMLA) Radial position of source [cm] Fig. 3. Spatial Resolution for a point source in the G-PET scanner. The Top and Bottom figures show the measured fwhm and fwtm values, respectively, in the reconstructed images. The 2D FBP projection was performed a ramp filter out any smoothing after Fourier rebinning of the data. process and does not represent the true performance of the scanner. B. Scatter The scatter fraction for the G-PET scanner was measured using a line source filled 18 F and placed at three radial positions (, 45, and 9 mm) in a water-filled 2 cm diameter and 19 cm long cylindrical phantom. The data were acquired at low count-rates and rebinned using single slice rebinning (SSRB). The energy gates for these measurements were set at EW1 ([435, 665]keV) and EW2 ([41, 665]keV). The measurement was performed the line source at three different radial positions (, 45, and 9 mm) and data analyzed according to the NEMA standard. The maximum scatter fraction (SF) value in the central slices varies from 44% for source radial position of mm to about 42% and 3% when the source is placed 45 and 9 mm radially respectively. The average scatter fraction for the scanner after using the NEMA weighting scheme is measured to be 34.5% (over central 17 cm of the scanner axial length) for EW1. Using a lower energy threshold (EW2) the average scatter fraction increases to about 39%, thus showing the importance of using a high threshold. C. Sensitivity As described in the NU2-21 standard, we measured the absolute sensitivity of the G-PET scanner using a 7 cm long line source at the center of the scanner, out and four different metal sleeves representing varying attenuation coefficients. This measurement technique is based on work described previously by Bailey, et. al. [6]. The measured count rate each metal sleeve was corrected for activity decay and the log of the results plotted as a function of sleeve thickness. Linear regression was employed to fit the data and obtain an extrapolated value for the absolute sensitivity of the scanner. We performed this measurement for two different energy windows: EW1 ([435, 665]keV) and EW2 ([41, 665]keV). The absolute sensitivity was measured to be 4.57 and 4.79 cps/kbq, respectively, showing only a 4.8% difference between the two energy windows. Figure 4 shows a plot of the axial sensitivity profile obtained the EW1 energy window. The axial sensitivity profile peaks at.8 cps/kbq (3 kcps/mci). However, this sensitivity measurement is normalized to the total activity present in the 7 cm long line source. Since the scanner axial FOV is 25.6 cm, the absolute sensitivity for activity in the field-of-view will be about three times higher than the value obtained by the NEMA measurement. Sensitivity [cps/kbq] Fig Slice number NU2-21 axial sensitivity profile for the G-PET scanner. The scanner sensitivity was also measured using the 2 cm diameter by 19 cm long NEMA cylinder. This cylinder was uniformly filled water containing a small amount of 18 F activity dissolved in it. Data were collected at a low count-rate so that random coincidences are negligible. The scatter fraction as measured in Section III B was used to correct for the scattered events in the collected sinograms, and thus estimate the true coincidences (Trues = Total (1-SF)). Measured this way, we obtained a sensitivity of about 2 kcps/kbq/ml (68 kcps/µci/cc). Since the length of this phantom (19 cm) is shorter than the axial FOV of the G-PET scanner (25.6 cm), the measured sensitivity is lower than the true value. The slice sensitivity (every 2. mm) for this phantom in G-PET is about.4 kcps/kbq/ml/2 mm (16 kcps/µci/cc/2 mm) near the center of the FOV. D. Count-rate tests For count-rate measurements, as recommended in the appendix of the NU2-21 standards, we used the 2 cm 1111
4 Count-rate [kcps] Count-rate [kcps] [kbq/ml] True Scatter Random Total Activity concentration [µci/cc] [kbq/ml] True NEC Activity concentration [µci/cc] Fig. 5. Count-rate curves for the G-PET scanner as measured a water-filled 2 cm NEMA cylinder. NEMA cylinder uniformly filled water and 18 F activity dissolved in it. We started about 111 MBq (3 mci) of activity and acquired dynamic scans over several halflives of the 18 F isotope. The tail fitting method using a parabola was used to estimate the background (scatter+random) fraction (background counts/total counts) for each acquisition frame. The background fraction at low activity levels gives the scatter fraction, while at higher activity concentrations it can be used to estimate the Random counts. The Noise Equivalent Count (NEC) rate was then calculated using the formula NEC = T T T+Sc+R, where T is the True, Sc is Scatter, and R is Random count-rate. Figure 5 summarizes the results of this countrate measurement. The energy gates for this measurement were set at EW2 ([41, 665]keV) and the NEC rate peaks at 75 kcps for an activity concentration of kbq/ml (.35 µci/cc). Better results could potentially be achieved the smaller energy window EW1, which reduces scatter in the collected events as well reduces deadtime. E. Image Contrast Image contrast measurements were performed using six small s in the 2 cm NEMA cylinder. The s had internal diameters (d o ) of 37, 28, 22, 17, 13, and 1 mm and were placed so that their centers lie in the same axial plane close to the central slice in the scanner. This measurement was performed in analogy to the prescribed image quality measurement in the NU2-21 standards for whole-body scanners the IEC recovery phantom. A total starting activity of MBq (1.5 mci) of 18 Fwas Fig. 6. A central slice from the reconstructed image for the contrast measurements. All s, filled an 8:1 activity concentration respect to the background, are clearly visible. used the six s containing a similar activity concentration of eight times that of the background cylinder. We imaged this phantom for 5 and 3 min. and used 3D RAMLA reconstruction of fully corrected data. TABLE I Summary of results obtained the G-PET scanner for the contrast measurements a 2 cm NEMA cylinder and six hot s placed in it. The ratio for the to background activity concentrations is 8:1. Sphere dia. Percent contrast (%) Background d o (mm) D=d o D=d o /2 Variability (%) 5 minutes acquisition 5 Mcts minutes acquisition 26 Mcts For analysis, we measured the count density for circular regions diameters (D) similar to the physical dimensions, and centered over the six s in the central slice. Counts in the volume of the are lost due to the partial volume effect. Therefore, we repeated this analysis by reducing the diameter of the circular regions to half that of the size. The background count density was obtained by drawing 12 regions in each of the central and ±1 and ±2 cm slices, leading to an average over 6 background regions for each size. As described in the NU2-21 contrast measurement, the percent contrast for each is defined as 1 C H,j C B,j 1 a H ab 1, where C H,j is the count density in the region-of-interest (ROI) for j, C B,j is the average count density in the background ROI s for j, and a H and a b are the activity concentrations in the s and background respectively. Figure 6 shows the central slice through one of these acquired images which had a total of 26 Mcts. Table I summarizes these results. 1112
5 Fig. 7. Selected slices from the reconstructed image of a 3D Hoffman Brain Phantom measured the G-PET scanner. Images are reconstructed fully 3D RAMLA iterative reconstruction algorithm. F. 3D Hoffman Brain Phantom Figure 7 shows representative slices acquired for a 3D Hoffman Brain Phantom the G-PET scanner. The data were acquired over 45 minutes about 5 million collected events. The image reconstruction was performed using 3D RAMLA reconstruction algorithm. These images show the high spatial resolution and image quality attained by this scanner a good delineation of substructures in the brain phantom. G. Patient Studies Figure 8 shows transverse, sagittal, and coronal views of selected slices from a patient study after an 18 F-FDG injection and scan time of about 3 minutes. High image contrast and spatial resolution lead to good visual quality in routine clinical scans. IV. Discussion and Conclusion A high resolution, high sensitivity, and high count-rate brain scanner using the GSO Anger-logic detector has been developed. Our performance measurements show that this scanner has a transverse resolution of 4 mm (8 mm fwtm) at the center out significant deterioration at a radial distance of 1 cm. The 55% solid angle coverage achieved by this scanner in 3D imaging mode, together the use of 1 mm long GSO crystals, results in an absolute sensitivity of 4.57 kcps/kbq. Good spatial resolution and high sensitivity are essential for imaging certain receptor systems in the brain which may also have low specific activity. In addition, brain perfusion studies 15 O-water, as well as studies using other short-lived isotopes such as 11 C, will require this scanner to operate at high count-rates in order to achieve good signal-to-noise ratio for imaging rapidly changing processes. With the fast signal decay time of GSO, together the restricted light spread in the detector design, the G-PET scanner is capable of operating at significantly higher count-rates little deadtime. The NEC rate for this scanner peaks at 75 kcps (for a 2 cm diameter by 19 cm long cylinder) for an activity concentration of about kbq/ml (.35 µci/cc). The true coincidence rate at this activity concentration was measured to be about 2 kcps. The count-rate results presented here were measured a low energy gate of 41 kev, and could potentially increase when the energy gate is raised to 435 kev. We have also measured the percent contrast achieved by this scanner by imaging six hot s in a uniform warm background. These measurements show that the small s are easily detectable, but even high spatial resolution, the partial volume effect is significant. For clinical 18 F-FDG imaging we currently scan for 3 minutes which provides high image quality. The results we show for the 3D Hoffman brain phantom, as well as the patient studies, illustrate the degree of structure which is visible in these scans. 1113
6 Fig. 8. Transverse, sagittal, and coronal views of selected slices from a patient study. The patient was scanned for 3 minutes after injection of 18 F-FDG. Acknowledgments We would like to thank Dr. Richard Freifelder for help the scanner construction, Dr. Margaret Daube- Witherspoon for data rebinning and reconstruction, and Mr. Chris Cardi for software and data acquisition support. The support we received from the ADAC UGM engineering staff during the completion of this scanner is greatly appreciated. This work was supported by the Counter Drug Technology Assessment Center (CTAC), an office in the Office of National Drug Control Policy, and by the U.S. Dept. of Energy grant No. DE-FG2-88ER6642. References [1] S. Surti, J. S. Karp, R. Freifelder, and F. Liu, Optimizing the performance of a PET detector using discrete GSO crystals on a continuous lightguide, IEEE Trans. Nucl. Sci., vol. 47, no. 3, pp , 2. [2] J. S. Karp, S. Surti, R. Freifelder, M. E. Daube-Witherspoon, C. Cardi, L.-E. Adam, B. Chase, P. Vaska, and G. Muehllehner, Performance of a GSO Brain PET Camera, in IEEE MIC Conference Record, 2. [3] Joel S. Karp, Richard Freifelder, Michael Geagan, Gerd Muehllehner, Paul E. Kinahan, Robert Lewitt, and Lingxiong Shao, Three-Dimensional imaging characteristics of the HEAD PENN-PET scanner, J. Nucl. Med., vol. 38, pp , [4] S. Surti, A Model of Scintillation Detector Performance for Positron Emission Tomography, Ph.D. thesis, University of Pennsylvania, December 2. [5] M. E. Daube-Witherspoon, S. Matej, J. S. Karp, and R. M. Lewitt, Application of the row action maximum likelihood algorithm spherical basis functions to clinical PET imaging, IEEE Trans. Nucl. Sci., vol. 48, no. 1, pp. 24 3, 21. [6] D. L. Bailey, T. Jones, and T. J. Spinks, A method for measuring the absolute sensitivity of Positron Emission Tomographic scanners, European Journal of Nuclear Medicine, vol. 18, pp ,
Chiara Secco. PET Performance measurements of the new LSO-Based Whole Body PET/CT. Scanner biograph 16 HI-REZ using the NEMA NU Standard.
Chiara Secco PET Performance measurements of the new LSO-Based Whole Body PET/CT Scanner biograph 16 HI-REZ using the NEMA NU 2-2001 Standard. INTRODUCTION Since its introduction, CT has become a fundamental
More informationLSO PET/CT Pico Performance Improvements with Ultra Hi-Rez Option
LSO PET/CT Pico Performance Improvements with Ultra Hi-Rez Option Y. Bercier, Member, IEEE, M. Casey, Member, IEEE, J. Young, Member, IEEE, T. Wheelock, Member, IEEE, T. Gremillion Abstract-- Factors which
More informationPerformance Assessment of Pixelated LaBr 3 Detector Modules for TOF PET
Performance Assessment of Pixelated LaBr 3 Detector Modules for TOF PET A. Kuhn, S. Surti, Member, IEEE, J. S. Karp, Senior Member, IEEE, G. Muehllehner, Fellow, IEEE, F.M. Newcomer, R. VanBerg Abstract--
More informationDesign Evaluation of A-PET: A High Sensitivity Animal PET Camera
IEEE TRANSACTIONS ON NUCLEAR SCIENCE, VOL. 50, NO. 5, OCTOBER 2003 1357 Design Evaluation of A-PET: A High Sensitivity Animal PET Camera S. Surti, Member, IEEE, J. S. Karp, Senior Member, IEEE, A. E. Perkins,
More informationA High-Resolution GSO-based Brain PET Camera
A High-Resolution GSO-based Brain PET Camera J.S. Karp', Senior Member IEEE, L.E. Adam', R.Freifelder', Member IEEE, G. Muehllehner3 Senior Member IEEE, F. Liu"', Student Member IEEE, S. Surti"', Student
More informationCHAPTER 8 GENERIC PERFORMANCE MEASURES
GENERIC PERFORMANCE MEASURES M.E. DAUBE-WITHERSPOON Department of Radiology, University of Pennsylvania, Philadelphia, Pennsylvania, United States of America 8.1. INTRINSIC AND EXTRINSIC MEASURES 8.1.1.
More informationPET Performance Evaluation of MADPET4: A Small Animal PET Insert for a 7-T MRI Scanner
PET Performance Evaluation of MADPET4: A Small Animal PET Insert for a 7-T MRI Scanner September, 2017 Results submitted to Physics in Medicine & Biology Negar Omidvari 1, Jorge Cabello 1, Geoffrey Topping
More informationPerformance evaluation of a new highsensitivity time-of-flight clinical PET/CT system
Huo et al. EJNMMI Physics (2018) 5:29 https://doi.org/10.1186/s40658-018-0229-4 EJNMMI Physics ORIGINAL RESEARCH Open Access Performance evaluation of a new highsensitivity time-of-flight clinical PET/CT
More informationPET Performance Measurements for an LSO- Based Combined PET/CT Scanner Using the National Electrical Manufacturers Association NU Standard
PET Performance Measurements for an LSO- Based Combined PET/CT Scanner Using the National Electrical Manufacturers Association NU 2-2001 Standard Yusuf E. Erdi, DSc 1 ; Sadek A. Nehmeh, PhD 1 ; Tim Mulnix,
More informationInitial evaluation of the Indiana small animal PET scanner
Initial evaluation of the Indiana small animal PET scanner Ned C. Rouze, Member, IEEE, Victor C. Soon, John W. Young, Member, IEEE, Stefan Siegel, Member, IEEE, and Gary D. Hutchins, Member, IEEE Abstract
More informationCelesteion Time-of-Flight Technology
Celesteion Time-of-Flight Technology Bing Bai, PhD Clinical Sciences Manager, PET/CT Canon Medical Systems USA Introduction Improving the care for every patient while providing a high standard care to
More informationFocusing on high performance
Advanced Molecular Imaging Vereos PET/CT Focusing on high performance Michael A. Miller, PhD, Philips, Advanced Molecular Imaging Physics This white paper presents a description of the Vereos digital PET/CT
More informationNoise Characteristics of the FORE+OSEM(DB) Reconstruction Method for the MiCES PET Scanner
Noise Characteristics of the FORE+OSEM(DB) Reconstruction Method for the MiCES PET Scanner Kisung Lee, Member, IEEE, Paul E. Kinahan, Senior Member, Robert S. Miyaoka, Member, IEEE, Jeffrey A. Fessler,
More informationCombined micropet /MR System: Performance Assessment of the Full PET Ring with Split Gradients 4.8
Combined micropet /MR System: Performance Assessment of the Full PET Ring with Split Gradients 4.8 UNIVERSITY OF CAMBRIDGE 1.2 Rob C. Hawkes 1, Tim D. Fryer 1, Alun J. Lucas 1,2, Stefan B. Siegel 3, Richard
More informationPET Detectors. William W. Moses Lawrence Berkeley National Laboratory March 26, 2002
PET Detectors William W. Moses Lawrence Berkeley National Laboratory March 26, 2002 Step 1: Inject Patient with Radioactive Drug Drug is labeled with positron (β + ) emitting radionuclide. Drug localizes
More informationThe PennPET Explorer Scanner for Total Body Applications
The PennPET Explorer Scanner for Total Body Applications JS Karp, MJ Geagan, G Muehllehner, ME Werner, T McDermott, JP Schmall, V Viswanath, University of Pennsylvania, Philadelphia, PA AE Perkins, C-H
More informationConceptual Study of Brain Dedicated PET Improving Sensitivity
Original Article PROGRESS in MEDICAL PHYSICS 27(4), Dec. 2016 https://doi.org/10.14316/pmp.2016.27.4.236 pissn 2508-4445, eissn 2508-4453 Conceptual Study of Brain Dedicated PET Improving Sensitivity Han-Back
More informationPET: New Technologies & Applications, Including Oncology
PET: New Technologies & Applications, Including Oncology, PhD, FIEEE Imaging Research Laboratory Department of Radiology University of Washington, Seattle, WA Disclosures Research Contract, GE Healthcare
More informationPerformance characterization of a novel thin position-sensitive avalanche photodiode-based detector for high resolution PET
2005 IEEE Nuclear Science Symposium Conference Record M11-126 Performance characterization of a novel thin position-sensitive avalanche photodiode-based detector for high resolution PET Jin Zhang, Member,
More informationFirst Applications of the YAPPET Small Animal Scanner
First Applications of the YAPPET Small Animal Scanner Guido Zavattini Università di Ferrara CALOR2 Congress, Annecy - FRANCE YAP-PET scanner Scintillator: YAP:Ce Size: matrix of 2x2 match like crystals
More informationSimulation and evaluation of a cost-effective high-performance brain PET scanner.
Research Article http://www.alliedacademies.org/biomedical-imaging-and-bioengineering/ Simulation and evaluation of a cost-effective high-performance brain PET scanner. Musa S Musa *, Dilber U Ozsahin,
More informationRecovery and normalization of triple coincidences in PET
Universidad Carlos III de Madrid Repositorio institucional e-archivo Área de Imagen e Instrumentación (BiiG) http://e-archivo.uc3m.es DBIAB - BIIG - Journal Articles 2015-03 Recovery and normalization
More informationPhotomultiplier Tube
Nuclear Medicine Uses a device known as a Gamma Camera. Also known as a Scintillation or Anger Camera. Detects the release of gamma rays from Radionuclide. The radionuclide can be injected, inhaled or
More informationSimulation of Algorithms for Pulse Timing in FPGAs
2007 IEEE Nuclear Science Symposium Conference Record M13-369 Simulation of Algorithms for Pulse Timing in FPGAs Michael D. Haselman, Member IEEE, Scott Hauck, Senior Member IEEE, Thomas K. Lewellen, Senior
More informationQuality control of Gamma Camera. By Dr/ Ibrahim Elsayed Saad 242 NMT
Quality control of Gamma Camera By Dr/ Ibrahim Elsayed Saad 242 NMT WHAT IS QUALITY? The quality of a practice is to fulfill the expectations and demands from: Patient Clinicain Your self Quality assurance
More informationPitfalls and Remedies of MDCT Scanners as Quantitative Instruments
intensity m(e) m (/cm) 000 00 0 0. 0 50 0 50 Pitfalls and Remedies of MDCT Scanners as Jiang Hsieh, PhD GE Healthcare Technology University of Wisconsin-Madison Root-Causes of CT Number Inaccuracies Nature
More informationMonte Carlo Simulation Study of a Dual-Plate PET Camera Dedicated to Breast Cancer Imaging
IEEE Nuclear Science Symposium Conference Record M-9 Monte Carlo Simulation Study of a Dual-Plate PET Camera Dedicated to Breast Cancer Imaging Jin Zhang, Member, IEEE, Peter D. Olcott, Member, IEEE, Angela
More information... In vivo imaging in Nuclear Medicine. 1957: Anger camera (X;Y) X Y
József Varga, PhD EMISSION IMAGING BASICS OF QUANTIFICATION Imaging devices Aims of image processing Reconstruction University of Debrecen Department of Nuclear Medicine. In vivo imaging in Nuclear Medicine
More informationDISCRETE crystal detector modules have traditionally been
IEEE TRANSACTIONS ON NUCLEAR SCIENCE, VOL. 53, NO. 5, OCTOBER 2006 2513 Performance Comparisons of Continuous Miniature Crystal Element (cmice) Detectors Tao Ling, Student Member, IEEE, Kisung Lee, and
More informationDevelopment of the LBNL Positron Emission Mammography Camera
Development of the LBNL Positron Emission Mammography Camera J.S. Huber, Member, IEEE, W.S. Choong, Member, IEEE, J. Wang, Member, IEEE, J.S. Maltz, Member, IEEE, J. Qi, Member, IEEE, E. Mandelli, Member,
More informationDesign of a Static Full-Ring Multi-Pinhole Collimator for Brain SPECT
Design of a Static Full-Ring Multi-Pinhole Collimator for Brain SPECT Karen Van Audenhaege, Student Member, IEEE, Roel Van Holen, Member, IEEE, Karel Deprez, Joel S. Karp, Senior Member, IEEE, Scott Metzler,
More informationNEMA and clinical evaluation of a novel brain PET-CT scanner
Journal of Nuclear Medicine, published on December 23, 2015 as doi:10.2967/jnumed.115.159723 NEMA and clinical evaluation of a novel brain PET-CT scanner Kira S. Grogg 1, Terrence Toole 2, Jinsong Ouyang
More informationDevelopment of PET using 4 4 Array of Large Size Geiger-mode Avalanche Photodiode
2009 IEEE Nuclear Science Symposium Conference Record M09-8 Development of PET using 4 4 Array of Large Size Geiger-mode Avalanche Photodiode K. J. Hong, Y. Choi, J. H. Kang, W. Hu, J. H. Jung, B. J. Min,
More informationRadionuclide Imaging MII Single Photon Emission Computed Tomography (SPECT)
Radionuclide Imaging MII 3073 Single Photon Emission Computed Tomography (SPECT) Single Photon Emission Computed Tomography (SPECT) The successful application of computer algorithms to x-ray imaging in
More informationNew Technology in Nuclear Medicine
New Technology in Nuclear Medicine Reed G. Selwyn, PhD, DABR Vice Chair of Research & Imaging Sciences Associate Professor and Chief, Medical Physics Dept. of Radiology, University of New Mexico Objectives
More informationFundamentals of Positron Emission Tomography (PET)
Fundamentals of Positron Emission Tomography (PET) NPRE 435, Principles of Imaging with Ionizing Radiation, Fall 2017 Content Fundamentals of PET Camera & Detector Design Real World Considerations Performance
More informationPROGRESS in TOF PET timing resolution continues to
Combined Analog/Digital Approach to Performance Optimization for the LAPET Whole-Body TOF PET Scanner W. J. Ashmanskas, Member, IEEE, Z. S. Davidson, B. C. LeGeyt, F. M. Newcomer, Member, IEEE, J. V. Panetta,
More informationTime-of-flight PET with SiPM sensors on monolithic scintillation crystals Vinke, Ruud
University of Groningen Time-of-flight PET with SiPM sensors on monolithic scintillation crystals Vinke, Ruud IMPORTANT NOTE: You are advised to consult the publisher's version (publisher's PDF) if you
More informationPET is a noninvasive, diagnostic imaging technique that
Performance Measurement of the micropet Focus 120 Scanner Jin Su Kim 1,2, Jae Sung Lee 1,2, Ki Chun Im 3, Su Jin Kim 1,2, Seog-Young Kim 3, Dong Soo Lee 1,2, and Dae Hyuk Moon 3 1 Department of Nuclear
More informationPositron Emission Tomography - PET
Positron Emission Tomography - PET Positron Emission Tomography Positron Emission Tomography (PET): Coincidence detection of annihilation radiation from positron-emitting isotopes followed by tomographic
More informationDiscovery ST. An Oncology System Designed For PET/CT. Revision: B Date: 30 Jan Page 1 of 47
Discovery ST An Oncology System Designed For PET/CT Revision: B Date: 30 Jan 2003 Page 1 of 47 TABLE OF CONTENTS 1 Introduction...3 2 Design Requirements...4 2.1 The Design Objective...4 2.2 Design Philosophy...5
More informationAs the role of gamma cameras expands in positron
BASIC SCIENCE INVESTIGATIONS Feasibility of a High-Speed Gamma-Camera Design Using the High-Yield-Pileup-Event- Recovery Method Wai-Hoi Wong, Hongdi Li, Jorge Uribe, Hossain Baghaei, Yu Wang, and Shigeru
More informationarxiv: v1 [physics.med-ph] 29 Nov 2018
Expected performance of the TT-PET scanner E. Ripiccini, a,b,1 D. Hayakawa, a,b G. Iacobucci, a M. Nessi, a,c E. Nowak, c L. Paolozzi, a O. Ratib, b P. Valerio a and D. Vitturini a a University of Geneva,
More informationSPECT Reconstruction & Filtering
SPECT Reconstruction & Filtering Goals Understand the basics of SPECT Reconstruction Filtered Backprojection Iterative Reconstruction Make informed choices on filter selection and settings Pre vs. Post
More informationHow Gamma Camera s Head-Tilts Affect Image Quality of a Nuclear Scintigram?
November 2014, Volume 1, Number 4 How Gamma Camera s Head-Tilts Affect Image Quality of a Nuclear Scintigram? Hojjat Mahani 1,2, Alireza Kamali-Asl 3, *, Mohammad Reza Ay 2, 4 1. Radiation Application
More informationPrimer on molecular imaging technology
Primer on molecular imaging technology Craig S. Levin Division of Nuclear Medicine, Department of Radiology and Molecular Imaging Program at Stanford (MIPS), Stanford University School of Medicine, 300
More information976 IEEE TRANSACTIONS ON NUCLEAR SCIENCE, VOL. 57, NO. 3, JUNE /$ IEEE
976 IEEE TRANSACTIONS ON NUCLEAR SCIENCE, VOL. 57, NO. 3, JUNE 2010 A Four-Layer DOI Detector With a Relative Offset for Use in an Animal PET System Mikiko Ito, Jae Sung Lee, Sun Il Kwon, Geon Song Lee,
More informationPET/CT Instrumentation Basics
/ Instrumentation Basics 1. Motivations for / imaging 2. What is a / Scanner 3. Typical Protocols 4. Attenuation Correction 5. Problems and Challenges with / 6. Examples Motivations for / Imaging Desire
More informationT h e P h a n t o m L a b o r a t o r y
T h e P h a n t o m L a b o r a t o r y 1 ECTphan Phantom SMR330 M a n u a l Copyright 2015 WARNING The use of this phantom requires radioactive fill solutions. Only people trained in the safe handling
More informationHigh-resolution PET scanners dedicated to small-animal
Micro Insert: A Prototype Full-Ring PET Device for Improving the Image Resolution of a Small- Animal PET Scanner Heyu Wu 1,2, Debashish Pal 3, Tae Yong Song 1, Joseph A. O Sullivan 4, and Yuan-Chuan Tai
More informationInvestigation of Multiple Head Registration / Center of Rotation for SPECT Gamma Cameras
Egyptian J. Nucl. Med., Vol 2, No. 2, Dec. 2009 82 PHYSICS, Original Artical Investigation of Multiple Head Registration / Center of Rotation for SPECT Gamma Cameras Abdelsattar, M.B. Ph.D.; BuHumaid,
More informationDevelopment of an innovative LSO-SiPM detector module for high-performance Positron Emission Tomography
Development of an innovative LSO-SiPM detector module for high-performance Positron Emission Tomography Maria Leonor Trigo Franco Frazão leonorfrazao@ist.utl.pt Instituto Superior Técnico, Lisboa, Portugal
More informationCHAPTER 15 DEVICES FOR EVALUATING IMAGING SYSTEMS
DEVICES FOR EVALUATING IMAGING SYSTEMS O. DEMIRKAYA, R. AL-MAZROU Department of Biomedical Physics, King Faisal Specialist Hospital and Research Centre, Riyadh, Saudi Arabia 15.1. DEVELOPING A QUALITY
More informationDevelopment of a High-Resolution and Depth-of- Interaction Capable Detector for Time-of-Flight PET
Development of a High-Resolution and Depth-of- Interaction Capable Detector for Time-of-Flight PET Srilalan Krishnamoorthy, Member, IEEE, Rony I. Wiener, Madhuri Kaul, Joseph Panetta, Joel S. Karp, Senior
More information2594 IEEE TRANSACTIONS ON NUCLEAR SCIENCE, VOL. 56, NO. 5, OCTOBER /$ IEEE
2594 IEEE TRANSACTIONS ON NUCLEAR SCIENCE, VOL. 56, NO. 5, OCTOBER 2009 Investigation of Depth of Interaction Encoding for a Pixelated LSO Array With a Single Multi-Channel PMT Yongfeng Yang, Member, IEEE,
More informationUCLA UCLA Previously Published Works
UCLA UCLA Previously Published Works Title Attenuation correction for small animal PET tomographs Permalink https://escholarship.org/uc/item/41n377p3 Journal Physics in Medicine and Biology, 5(8) ISSN
More information2/14/2019. Nuclear Medicine Artifacts. Symmetric energy windows
Nuclear Medicine Artifacts SCPMG Medical Imaging Technology & Informatics Medical Physics Group Brian Helbig, MS, DABR 1 2 Symmetric energy windows 3 1 Dynamic clinical study Energy peak shift Electrical
More informationParameters Affecting on Intrinsic Uniformity Test For MEDISO
ISPUB.COM The Internet Journal of Nuclear Medicine Volume 5 Number 2 Parameters Affecting on Intrinsic Uniformity Test For MEDISO S Zobly, A Osman Citation S Zobly, A Osman. Parameters Affecting on Intrinsic
More informationARTICLE IN PRESS. Nuclear Instruments and Methods in Physics Research A
Nuclear Instruments and Methods in Physics Research A 614 (2010) 308 312 Contents lists available at ScienceDirect Nuclear Instruments and Methods in Physics Research A journal homepage: www.elsevier.com/locate/nima
More informationDetector technology challenges for nuclear medicine and PET
Nuclear Instruments and Methods in Physics Research A 513 (2003) 1 7 Detector technology challenges for nuclear medicine and PET Paul K. Marsden Guy s and St. Thomas Clinical PET Centre, King s College
More informationA NOVEL CONCEPT FOR A POSITRON EMISSION TOMOGRAPHY SCANNER
A NOVEL CONCEPT FOR A POSITRON EMISSION TOMOGRAPHY SCANNER An Undergraduate Research Scholars Thesis by BRIAN KELLY, MATTHEW LEE ELLIOT LEVIN and JEENA KHATRI Submitted to Honors and Undergraduate Research
More informationStudy of a Prototype VP-PET Imaging System Based on highly. Pixelated CdZnTe Detectors
Study of a Prototype VP-PET Imaging System Based on highly Pixelated CdZnTe Detectors Zheng-Qian Ye 1, Ying-Guo Li 1, Tian-Quan Wang 1, Ya-Ming Fan 1, Yong-Zhi Yin 1,*, Xi-Meng Chen 1 Affiliations: 1 School
More informationRobert Pagnanelli BSRT(R)(N), CNMT, NCT, FASNC Chief Technologist, Nuclear Imaging Duke University Medical Center. Thursday September 8, 2011
Robert Pagnanelli BSRT(R)(N), CNMT, NCT, FASNC Chief Technologist, Nuclear Imaging Duke University Medical Center Thursday September 8, 2011 Quality Control Quality control should be performed because:
More informationEvaluation of Scatter Fraction and Count Rate Performance of Two Smallanimal PET scanners using dedicated phantoms
2011 IEEE Nuclear Science Symposium Conference Record MIC18.M-36 Evaluation of Scatter Fraction and Count Rate Performance of Two Smallanimal PET scanners using dedicated phantoms Rameshwar Prasad, Student
More informationA new operative gamma camera for Sentinel Lymph Node procedure
A new operative gamma camera for Sentinel Lymph Node procedure A physicist device for physicians Samuel Salvador, Virgile Bekaert, Carole Mathelin and Jean-Louis Guyonnet 12/06/2007 e-mail: samuel.salvador@ires.in2p3.fr
More informationPerformance measurements of a depth-encoding PET detector module based on positionsensitive
Home Search Collections Journals About Contact us My IOPscience Performance measurements of a depth-encoding PET detector module based on positionsensitive avalanche photodiode read-out This article has
More informationThe development of high-resolution PET systems has
Journal of Nuclear Medicine, published on December 12, 2007 as doi:10.2967/jnumed.107.044149 A Feasibility Study of a Prototype PET Insert Device to Convert a General-Purpose Animal PET Scanner to Higher
More informationAttenuation length in strip scintillators. Jonathan Button, William McGrew, Y.-W. Lui, D. H. Youngblood
Attenuation length in strip scintillators Jonathan Button, William McGrew, Y.-W. Lui, D. H. Youngblood I. Introduction The ΔE-ΔE-E decay detector as described in [1] is composed of thin strip scintillators,
More informationACR Update in Nuclear Medicine Accreditation
Disclaimer ACR Update in Nuclear Medicine Accreditation Beth A. Harkness, MS, DABR, FACR Henry Ford Health System Detroit, MI ACR physics subcommittee for nuclear medicine accreditation. My facility is
More informationQC Testing for Computed Tomography (CT) Scanner
QC Testing for Computed Tomography (CT) Scanner QA - Quality Assurance All planned and systematic actions needed to provide confidence on a structure, system or component. all-encompassing program, including
More informationUsing GATE to understand performance of a full-torso PET scanner
University of Pennsylvania Using GATE to understand performance of a full-torso PET scanner Varsha Viswanath 1 Margaret E. Daube-Witherspoon 1, Matthew E. Werner 1, Suleman Surti 1, Andreia Trindade 2,
More information16 Instrumentation and Data Acquisition
Instrumentation and Data Acquisition 275 16 Instrumentation and Data Acquisition Sibylle I. Ziegler and Magnus Dahlbom CONTENTS 16.1 Detectors and Imaging Systems 275 16.1.1 Principles of Scintillation
More informationWilliam Hallet - PRIMA IV 1
Quantitative and application specific imaging PET: the measurement process,reconstruction, calibration, quantification Dr William Hallett Centre for Imaging Sciences Imperial College Hammersmith Hospital
More informationLightburst Digital Detector
GE Healthcare Lightburst Digital Detector INTRODUCTION In clinical practice, PET/CT imaging helps clinicians visualize disease at an early stage, before it metastasizes and involves other organs, tissues
More informationFactors Affecting the resolution of SPECT Imaging. h.
Factors Affecting the resolution of SPECT Imaging H. E. Mostafa *1, H. A. Ayoub 2 and Sh.Magraby 1 1 Kasr El-Ini Center for Oncology, Cairo University, 2 Faculty of Science, Suez Canal University hayamayoub@yahoo.com
More informationFPGA-Based Pulse Pile-up Correction
FPGA-Based Pulse Pile-up Correction M.D. Haselman 1, J. Pasko 1, S. Hauck 1, Senior Member IEEE, T.K. Lewellen 2, Fellow IEEE, R.S. Miyaoka 2, Member IEEE, 1 University of Washington Department of Electrical
More informationNM Module Section 2 6 th Edition Christian, Ch. 3
NM 4303 Module Section 2 6 th Edition Christian, Ch. 3 Gas Filled Chamber Voltage Gas filled chamber uses Hand held detectors cutie pie Geiger counter Dose calibrators Cutie pie Chamber voltage in Ionization
More informationDefense Technical Information Center Compilation Part Notice
UNCLASSIFIED Defense Technical Information Center Compilation Part Notice ADPO 11245 TITLE: An Analysis System for PET Detector DISTRIBUTION: Approved for public release, distribution unlimited This paper
More informationHIGH RESOLUTION COMPUTERIZED TOMOGRAPHY SYSTEM USING AN IMAGING PLATE
HIGH RESOLUTION COMPUTERIZED TOMOGRAPHY SYSTEM USING AN IMAGING PLATE Takeyuki Hashimoto 1), Morio Onoe 2), Hiroshi Nakamura 3), Tamon Inouye 4), Hiromichi Jumonji 5), Iwao Takahashi 6); 1)Yokohama Soei
More informationNIH Public Access Author Manuscript Nucl Instrum Methods Phys Res A. Author manuscript; available in PMC 2007 December 14.
NIH Public Access Author Manuscript Published in final edited form as: Nucl Instrum Methods Phys Res A. 2007 January 21; 570(3): 543 555. A prototype of very high resolution small animal PET scanner using
More informationInvestigation of low noise, low cost readout electronics for high sensitivity PET systems based on Avalanche Photodiode arrays
Investigation of low noise, low cost readout electronics for high sensitivity PET systems based on Avalanche Photodiode arrays Frezghi Habte, Member, IEEE and Craig S.Levin, Member, IEEE Abstract A compact,
More informationExplain what is meant by a photon and state one of its main properties [2]
1 (a) A patient has an X-ray scan taken in hospital. The high-energy X-ray photons interact with the atoms inside the body of the patient. Explain what is meant by a photon and state one of its main properties....
More informationImage Quality and Dose. Image Quality and Dose. Image Quality and Dose Issues in MSCT. Scanner parameters affecting IQ and Dose
Image Quality and Dose Issues in MSCT Image Quality and Dose Image quality Image noise Spatial resolution Contrast Artefacts Speckle and sharpness S. Edyvean St. George s Hospital London SW17 0QT Radiation
More informationNuclear Associates , , CT Head and Body Dose Phantom
Nuclear Associates 76-414,76-414-4150,76-415 CT Head and Body Dose Phantom Users Manual March 2005 Manual No. 76-414-1 Rev. 2 2004, 2005 Fluke Corporation, All rights reserved. Printed in U.S.A. All product
More informationHISTORY. CT Physics with an Emphasis on Application in Thoracic and Cardiac Imaging SUNDAY. Shawn D. Teague, MD
CT Physics with an Emphasis on Application in Thoracic and Cardiac Imaging Shawn D. Teague, MD DISCLOSURES 3DR- advisory committee CT PHYSICS WITH AN EMPHASIS ON APPLICATION IN THORACIC AND CARDIAC IMAGING
More informationPositron Emission Tomography
Positron Emission Tomography UBC Physics & Astronomy / PHYS 409 1 Introduction Positron emission tomography (PET) is a non-invasive way to produce the functional 1 image of a patient. It works by injecting
More informationIntroduction. Chapter 16 Diagnostic Radiology. Primary radiological image. Primary radiological image
Introduction Chapter 16 Diagnostic Radiology Radiation Dosimetry I Text: H.E Johns and J.R. Cunningham, The physics of radiology, 4 th ed. http://www.utoledo.edu/med/depts/radther In diagnostic radiology
More informationNuclear Associates , &
Nuclear Associates 76-823, 76-824 & 76-825 PET/SPECT Phantom Source Tank, Phantom Inserts and Cardiac Insert Users Manual March 2005 Manual No. 76-823-1 Rev. 2 2004, 2005 Fluke Corporation, All rights
More informationAn innovative detector concept for hybrid 4D-PET/MRI Imaging
Piergiorgio Cerello (INFN - Torino) on behalf of the 4D-MPET* project *4 Dimensions Magnetic compatible module for Positron Emission Tomography INFN Perugia, Pisa, Torino; Polytechnic of Bari; University
More informationDetector technology in simultaneous spectral imaging
Computed tomography Detector technology in simultaneous spectral imaging Philips IQon Spectral CT Z. Romman, I. Uman, Y. Yagil, D. Finzi, N. Wainer, D. Milstein; Philips Healthcare While CT has become
More informationImaging with FDG PET is a valuable technique for tumor
Noise Reduction in Oncology FDG PET Images by Iterative Reconstruction: A Quantitative Assessment Cyril Riddell, Richard E. Carson, Jorge A. Carrasquillo, Steven K. Libutti, David N. Danforth, Millie Whatley,
More informationLaBr 3 :Ce, the latest crystal for nuclear medicine
10th Topical Seminar on Innovative Particle and Radiation Detectors 1-5 October 2006 Siena, Italy LaBr 3 :Ce, the latest crystal for nuclear medicine Roberto Pani On behalf of SCINTIRAD Collaboration INFN
More informationInternational Journal of Scientific & Engineering Research, Volume 4, Issue 9, September ISSN
International Journal of Scientific & Engineering Research, Volume 4, Issue 9, September-013 06 Evaluating the effect of acquisition parameters on image quality and acquisition time with SPECT using collimator
More informationX-RAY COMPUTED TOMOGRAPHY
X-RAY COMPUTED TOMOGRAPHY Bc. Jan Kratochvíla Czech Technical University in Prague Faculty of Nuclear Sciences and Physical Engineering Abstract Computed tomography is a powerful tool for imaging the inner
More informationFeatures and Weaknesses of Phantoms for CR/DR System Testing
Physics testing of image detectors Parameters to test Features and Weaknesses of Phantoms for CR/DR System Testing Spatial resolution Contrast resolution Uniformity/geometric distortion Dose response/signal
More informationDesign Studies of A High-Performance Onboard Positron Emission Tomography For Integrated Small Animal PET/CT/RT Radiation Research Systems
Proceedings of the International MultiConference of Engineers and Computer Scientists 2018 Vol II Design Studies of A High-Performance Onboard Positron Emission Tomography For Integrated Small Animal PET/CT/RT
More informationDesign of a High Resolution and High Sensitivity Scintillation Crystal Array with Nearly Perfect Light Collection
Design of a High Resolution and High Sensitivity Scintillation Crystal Array with Nearly Perfect Light Collection Craig S. Levin, Member, IEEE Abstract-- Spatial resolution improvements in Positron Emission
More informationFirst Results From the High-Resolution mousespect Annular Scintillation Camera
First Results From the High-Resolution mousespect Annular Scintillation Camera Andrew L. Goertzen, Douglas W. Jones, Jurgen Seidel, King Li, and Michael V. Green Abstract High resolution SPECT imaging
More informationDesign of a High-Resolution and High-Sensitivity Scintillation Crystal Array for PET With Nearly Complete Light Collection
2236 IEEE TRANSACTIONS ON NUCLEAR SCIENCE, VOL. 49, NO. 5, OCTOBER 2002 Design of a High-Resolution and High-Sensitivity Scintillation Crystal Array for PET With Nearly Complete Light Collection Craig
More informationSoftware and Hardware in CCTA. Elly Castellano PhD
Software and Hardware in CCTA Elly Castellano PhD Outline technical requirements for coronary CTA the modern cardiac CT scanner ECG-gating technology image reconstruction algorithms 2 Technical requirements
More information