Software and Hardware in CCTA. Elly Castellano PhD
|
|
- Harry Dickerson
- 5 years ago
- Views:
Transcription
1 Software and Hardware in CCTA Elly Castellano PhD
2 Outline technical requirements for coronary CTA the modern cardiac CT scanner ECG-gating technology image reconstruction algorithms 2
3 Technical requirements for coronary CTA 3
4 Technical requirements for coronary CTA ECG-gating capability freezes cardiac motion high temporal resolution reduces blurring due to cardiac motion high contrast and spatial resolution enhances conspicuity of narrow vessels large detector coverage per rotation facilitates short scan times without ECG gating images courtesy of Willi Kalender with ECG gating 4
5 ECG gating two physiological windows of opportunity HR < 65 HR > 65 R-R interval 100% V V t t ~ 65-75% ~ 25-35% end diastole end systole choice in timing and width of acquisition window to capture one / several / all phases of cardiac cycle 5
6 Temporal resolution determined by time taken to acquire projection data rotation time 180º+ or 360º scan conventional CT or DSCT at 60 bpm require < 200 ms possible with 0.4 s rotation time and 180º+ scan 180º+ scan 360º scan 6
7 determined by Spatial resolution in-plane: reconstruction kernel out-of-plane: slice thickness require sharp reconstruction kernels increase spatial resolution, but increase image noise < 1 mm slices increase spatial resolution, but increase image noise Corresponding CT and conventional angiogram images of RCA 7
8 Contrast resolution determined by difference in HU between arteries and background kv setting partial volume effects image noise require contrast enhancement low kv to increase sensitivity to iodine, but image noise and likelihood of photon starvation increase high maximum tube ma for use with lower kv settings < 1 mm slices to reduce partial volume effects, but image noise increases 8
9 determined by Coverage per rotation width of detector array with thin detector elements require > 30 mm of sub-mm detector rows 4-5 rotations to cover heart all data acquisition during optimal contrast enhancement image-stitching artefacts reduced 9
10 The modern cardiac CT scanner 10
11 The basic model: cardiac-enabled 64-DCT scanners* supplier example scanner model detector array rows x element mm z-coverage mm slices per rotation x thinnest slice mm minimum rot s GE Optima CT x x Siemens Definition Edge 64 x x Philips Brilliance x x Toshiba Aquilion CXL 64 x x NB manufacturers with z-ffs or detector resampling double number of slices per rotation * NB scanner specifications are continually changing 11
12 Top MDCT scanners 12
13 Revolution series: Rev CT 512-slice scanner 256 detector rows with resampling improved anatomical coverage 160 mm detector covers heart improved spatial resolution new x-ray tube HD mode using xy-ffs improved contrast resolution Gemstone Clarity detector in 3 sections increased detector efficiency lower detector afterglow image courtesy of GE 13
14 Brilliance ict / ict Elite 256-slice scanner 128 detector rows with z-ffs improved anatomical coverage 80 mm detector faster rotation time 0.27 s with AirGlide technology and imrc tube images courtesy of Philips 14
15 Definition Force 2 imaging chains mounted at ~94º 50 and 35 cm FOV improved temporal resolution faster acquisition of projection data set improved anatomical coverage 96 detector rows, 57.6 mm Turbo Flash mode high pitch helical mode image courtesy of Siemens 15
16 Aquilion One ViSION 640-slice scanner 320 detector rows with double sampling improved anatomical coverage 160 mm detector covers heart volume mode no table feed image courtesy of Toshiba 16
17 Top MDCT scanners: summary* supplier scanner model detector design rows x element mm z-coverage mm slices per rotation x thinnest slice mm minimum rot s GE Siemens Revolution CT Definition Force 256 x x x 0.6 x x Philips ict / Elite 128 x x Toshiba Aquilion One Vision 320 x x NB manufacturers with z-ffs or detector resampling double number of slices per rotation *from manufacturers information. NB scanner specifications are continually changing 17
18 ECG gating technology 18
19 ECG gating of acquisition essentially two types: prospective retrospective different image sets generated different radiation burden to patient 19
20 Prospective gating acquisition triggered by ECG set time after R wave acquisition window matched to minimum acquisition time no padding acquisition image sets in selected phase acquisition window extended beyond minimum acquisition time with padding acquisition at full or reduced ma image sets in 1+ selected phases can adjust for heart rate changes ma ma prescribed mas prescribed mas t t padding mas 20
21 Prospective gating choice of table motions step-and-shoot slow helical high-pitch helical (Flash) none axial images reconstructed from partial projection data sets MPRs, VR image sets patient dose 1 5 msv CXRs partial projection data set 21
22 Prospective gating Flash helical mode LAD with stent Image courtesy of Universität Ulm, Germany, and Philips images courtesy of Institute of Medical Physics / Erlangen, Germany, and Siemens 22
23 Retrospective gating long acquisition fast rotation time choice of table motion slow helical none pitch typically projections tagged with ECG signal ECG-gated ma pulsing selectable temporal window position and width restricts reconstruction to selected cardiac phase(s) 23
24 Retrospective gating images reconstructed from partial projection data set to improve temporal resolution single or multi-segment recon determines temporal resolution reconstruction window can be adjusted essential for arrhythmic patients typical patient dose msv CXRs single-segment recon multi-segment recon 24
25 Retrospective gating coronary CTA from images reconstructed at one cardiac phase cardiac function from images reconstructed at several cardiac phases Functional Analysis, EF 80% images courtesy of Dr Rybicki, BWH, and Toshiba 25
26 Dealing with irregular heart rates arrhythmia detection algorithms acquisition delayed until rhythm re-established compensation for heart rate changes acquisition window automatically widened to capture required phase(s) may impact on radiation dose to patient 26
27 Image reconstruction algorithms 27
28 Filtered back-projection choice of convolution kernels images courtesy of Willi Kalender 28
29 Iterative reconstruction from Beister et al
30 IR: vendor solutions supplier IR algorithms GE ASiR Veo ASiR-V Siemens SAFIRE ADMIRE Philips idose4 IMR Toshiba AIDR 3D proprietary black boxes! 30
31 IR solutions in raw data and image domains used by Philips (idose), Siemens and Toshiba based on statistical modelling of x-ray detection from Willemink et al
32 IR solutions in raw data domain used by GE, Philips (IMR) based on statistical modelling of x-ray detection geometric modelling of imaging chain (Veo) physics modelling (ASiR-V) from Willemink et al
33 Iterative reconstruction images presented different IR strengths mixed with FBP lower image noise than FBP opportunity for dose reduction FBP SAFIRE images courtesy of Siemens 33
34 Iterative reconstruction slower recon than FBP IR-specific artefacts may be introduced disappearance of hypodense seam from Beister et al
35 Conclusions coronary CT angiography requires cardiacenabled scanners 64+ DCT choice of ECG gating techniques iterative reconstruction algorithms should be carefully validated 35
36 Acknowledgements Dee Mears, RMH Nina Arcuri, RBH Willi Kalender ImPACT group Philips Siemens Toshiba GE 36
Iterative Reconstruction
RECENT ADVANCES IN CT RADIATION DOSE REDUCTION TECHNIQUES Iterative Reconstruction Kalpana Kanal, PhD, FSCBTMR, FACR, FAAPM Professor and Director, Diagnostic Physics Section University of Washington Seattle,
More informationResearch Support. Dual-Source CT: What is it and How Do I Test it? Cynthia H. McCollough, Ph.D.
Dual-Source CT: What is it and How Do I Test it? Cynthia H. McCollough, Ph.D. CT Clinical Innovation Center Department of Radiology Mayo Clinic College of Medicine Rochester, MN Research Support National
More informationPitfalls and Remedies of MDCT Scanners as Quantitative Instruments
intensity m(e) m (/cm) 000 00 0 0. 0 50 0 50 Pitfalls and Remedies of MDCT Scanners as Jiang Hsieh, PhD GE Healthcare Technology University of Wisconsin-Madison Root-Causes of CT Number Inaccuracies Nature
More informationImage Quality and Dose. Image Quality and Dose. Image Quality and Dose Issues in MSCT. Scanner parameters affecting IQ and Dose
Image Quality and Dose Issues in MSCT Image Quality and Dose Image quality Image noise Spatial resolution Contrast Artefacts Speckle and sharpness S. Edyvean St. George s Hospital London SW17 0QT Radiation
More informationIterative Reconstruction in Image Space. Answers for life.
Iterative Reconstruction in Image Space Answers for life. Iterative Reconstruction in Image Space * (IRIS) * Please note: IRIS is used as an abbreviation for Iterative Reconstruction in Image Space throughout
More information1. Patient size AEC. Large Patient High ma. Small Patient Low ma
Comparison of the function and performance of CT AEC systems CTUG meeting by Emily Field Trainee clinical scientist 14 th th Breakdown CT Automatic Exposure Control (AEC) Background Project Description
More information12/21/2016. Siemens Medical Systems Research Agreement Philips Healthcare Research Agreement AAN and ASN Committees
Joseph V. Fritz, PhD Nandor Pintor, MD Dent Neurologic Institute ASN 2017 Friday, January 20, 2017 Siemens Medical Systems Research Agreement Philips Healthcare Research Agreement AAN and ASN Committees
More informationWide-Detector CT for TAVR Planning:
Wide-Detector CT for TAVR Planning: Impact on Iodine Dose, Radiation Dose, and Image Quality SCBTMR 2015 Annual Course Thursday, October 8 William P. Shuman MD FSCBTMR Department of Radiology University
More information160-slice CT SCANNER / New Standard for the Future
TECHNOLOGY HISTORY For over 130 years, Toshiba has been a world leader in developing technology to improve the quality of life. Our 50,000 global patents demonstrate a long, rich history of leading innovation.
More informationMultislice CT: Current Technology and Future Developments
Multislice CT: Current Technology and Future Developments 1 Stefan Ulzheimer and Thomas Flohr Contents 1.1 Introduction 3 1.2 System Design 5 1.2.1 Gantry 6 1.2.2 X-Ray Tube and Generator 6 1.2.3 MDCT
More informationSAFIRE. Sinogram Affirmed Iterative Reconstruction. Answers for life.
Neuro Thoracic Abdominal Abdominal Cardiovascular Pediatric SAFIRE Sinogram Affirmed Iterative Reconstruction Answers for life. SAFIRE * (Sinogram Affirmed Iterative Reconstruction) * The information
More informationHISTORY. CT Physics with an Emphasis on Application in Thoracic and Cardiac Imaging SUNDAY. Shawn D. Teague, MD
CT Physics with an Emphasis on Application in Thoracic and Cardiac Imaging Shawn D. Teague, MD DISCLOSURES 3DR- advisory committee CT PHYSICS WITH AN EMPHASIS ON APPLICATION IN THORACIC AND CARDIAC IMAGING
More informationAutomated dose control in multi-slice CT. Nicholas Keat Formerly ImPACT, St George's Hospital, London
Automated dose control in multi-slice CT Nicholas Keat Formerly ImPACT, St George's Hospital, London Introduction to presentation CT contributes ~50+ % of all medical radiation dose Ideally all patients
More informationDetector technology in simultaneous spectral imaging
Computed tomography Detector technology in simultaneous spectral imaging Philips IQon Spectral CT Z. Romman, I. Uman, Y. Yagil, D. Finzi, N. Wainer, D. Milstein; Philips Healthcare While CT has become
More informationInfluence of different iteration levels in fourth generation iterative reconstruction technique on image noise in CT examinations of the neck
Influence of different iteration levels in fourth generation iterative reconstruction technique on image noise in CT examinations of the neck Poster No.: C-2205 Congress: ECR 2012 Type: Scientific Paper
More informationTOPICS: CT Protocol Optimization over the Range of Patient Age & Size and for Different CT Scanner Types: Recommendations & Misconceptions
CT Protocol Optimization over the Range of Patient Age & Size and for Different CT Scanner Types: Recommendations & Misconceptions TOPICS: Computed Tomography Quick Overview CT Dosimetry Effects of CT
More informationInside Biograph mct.
Inside Biograph mct The technologies behind the world s first molecular CT. www.siemens.com/mi Large 78 cm bore helps reduce claustrophobia and provides more room for RTP positioning devices. 227 kg (500
More informationMaximum Performance, Minimum Space
TECHNOLOGY HISTORY For over 130 years, Toshiba has been a world leader in developing technology to improve the quality of life. Our 50,000 global patents demonstrate a long, rich history of leading innovation.
More informationThe disclaimer on page 1 is an integral part of this document. Copyright November 30, 2017 by AAPM. All rights reserved.
DISCLAIMER: TO THE EXTENT ALLOWED BY LOCAL LAW, THIS INFORMATION IS PROVIDED TO YOU BY THE AMERICAN ASSOCIATION OF PHYSICISTS IN MEDICINE, A NON-PROFIT ORGANIZATION ORGANIZED TO PROMOTE THE APPLICATION
More informationabc MHRA Philips Mx8000 IDT CT scanner technical evaluation September 2004 Best choice best practice nww.medical-devices.nhs.
abc September 2004 MHRA 04099 Philips Mx8000 IDT CT scanner technical evaluation Best choice best practice www.mhra.gov.uk nww.medical-devices.nhs.uk About MHRA evaluation reports. What you can expect.
More informationEnhanced Functionality of High-Speed Image Processing Engine SUREengine PRO. Sharpness (spatial resolution) Graininess (noise intensity)
Vascular Enhanced Functionality of High-Speed Image Processing Engine SUREengine PRO Medical Systems Division, Shimadzu Corporation Yoshiaki Miura 1. Introduction In recent years, digital cardiovascular
More informationBreast Tomosynthesis. Bob Liu, Ph.D. Department of Radiology Massachusetts General Hospital And Harvard Medical School
Breast Tomosynthesis Bob Liu, Ph.D. Department of Radiology Massachusetts General Hospital And Harvard Medical School Outline Physics aspects of breast tomosynthesis Quality control of breast tomosynthesis
More informationCOMPUTED TOMOGRAPHY 1
COMPUTED TOMOGRAPHY 1 Why CT? Conventional X ray picture of a chest 2 Introduction Why CT? In a normal X-ray picture, most soft tissue doesn't show up clearly. To focus in on organs, or to examine the
More informationTORNIER BLUEPRINT. 3D Planning + PSI SCAN PROTOCOL
TORNIER BLUEPRINT 3D Planning + PSI SCAN PROTOCOL Contents 3 Introduction 3 Patient preparation 3 Scanning instructions 4 Image instructions 5 Scanning parameters 6 Technical instructions 2 BLUEPRINT 3D
More informationImprovement of CT image quality with iterative reconstruction idose4
Improvement of CT image quality with iterative reconstruction idose4 Poster No.: C-0387 Congress: ECR 2014 Type: Scientific Exhibit Authors: M.-L. Olsson, K. Norrgren, M. Söderberg; Malmö/SE Keywords:
More informationClinical Experience Using the Open Bore Multislice CT System Supria (16 slice CT) MEDIX VOL. 61 P.8 P.11
Clinical Experience Using the Open Bore Multislice CT System Supria (16 slice CT) Hiroki Kadoya Yukiko Kitagawa MEDIX VOL. 61 P.8 P.11 Clinical Experience Using the Open Bore Multislice CT System Supria
More informationAquilion Precision Ultra-High Resolution CT: Quantifying diagnostic image quality
Aquilion Precision Ultra-High CT: Quantifying diagnostic image quality Kirsten Boedeker, PhD, DABR Senior Manager, Quantitative Image Quality Canon Medical Systems Corporation Introduction Over the last
More informationPET: New Technologies & Applications, Including Oncology
PET: New Technologies & Applications, Including Oncology, PhD, FIEEE Imaging Research Laboratory Department of Radiology University of Washington, Seattle, WA Disclosures Research Contract, GE Healthcare
More informationNeuViz 16 Computed Tomography. Elevating routine imaging for exceptional results
NeuViz 16 Computed Tomography Elevating routine imaging for exceptional results Essence NeuViz 16 Raising the bar on clinical utility in routine imaging. Get more. More clinical information for patients.
More informationPreclinical X-Ray Computed Tomography
Preclinical X-Ray Computed Tomography Marc Kachelrieß German Cancer Research Center (DKFZ) Heidelberg, Germany www.dkfz.de/ct Is CT a Molecular Imaging Modality? Imaging Modality Molecular Sensitivity
More informationBrilliance in everything Philips CT products and services
Brilliance in everything Philips CT products and services Ready for anything No one does more than Philips to help you gain the productivity you need with a comprehensive approach to CT that marries significant
More informationIntroduction. Chapter 16 Diagnostic Radiology. Primary radiological image. Primary radiological image
Introduction Chapter 16 Diagnostic Radiology Radiation Dosimetry I Text: H.E Johns and J.R. Cunningham, The physics of radiology, 4 th ed. http://www.utoledo.edu/med/depts/radther In diagnostic radiology
More informationCardiac MR. Dr John Ridgway. Leeds Teaching Hospitals NHS Trust, UK
Cardiac MR Dr John Ridgway Leeds Teaching Hospitals NHS Trust, UK Cardiac MR Physics for clinicians: Part I Journal of Cardiovascular Magnetic Resonance 2010, 12:71 http://jcmr-online.com/content/12/1/71
More informationPET/CT Instrumentation Basics
/ Instrumentation Basics 1. Motivations for / imaging 2. What is a / Scanner 3. Typical Protocols 4. Attenuation Correction 5. Problems and Challenges with / 6. Examples Motivations for / Imaging Desire
More informationModel Based Iterative Reconstructions represent a paradigm shift - Imaging with almost no noise
Model Based Iterative Reconstructions represent a paradigm shift - Imaging with almost no noise Jonas Rydberg, M.D. Professor of Radiology Indiana University School of Medicine Indianapolis, Indiana Medical
More informationdiagnostic examination
RADIOLOGICAL PHYSICS 2011 Raphex diagnostic examination Adel A. Mustafa, Ph.D., Editor PUBLISHED FOR: RAMPS (Radiological and Medical Physics Society of New York) preface The RAPHEX Diagnostic exam 2011
More informationQC Testing for Computed Tomography (CT) Scanner
QC Testing for Computed Tomography (CT) Scanner QA - Quality Assurance All planned and systematic actions needed to provide confidence on a structure, system or component. all-encompassing program, including
More informationSlide 1. Slide 2. Slide 3 ACR CT Accreditation. Multi-Slice CT Artifacts and Quality Control. What are the rules or recommendations for CT QC?
Slide 1 Multi-Slice CT Artifacts and Quality Control Dianna Cody, Ph.D. Chief, Radiologic Physics UT MD Anderson Cancer Center Houston, TX Slide 2 What are the rules or recommendations for CT QC? AAPM
More informationTranslating Protocols Between Scanner Manufacturer and Model
Translating Protocols Between Scanner Manufacturer and Model Robert J. Pizzutiello, MS, FAAPM, FACMP Sr. Vice-President, Global Physics Solutions President, Upstate Medical Physics Objectives Understand
More informationComputed Tomography. The Fundamentals of... THE FUNDAMENTALS OF... Jason H. Launders, MSc. Current Technology
The Fundamentals of... Computed Tomography Computed Tomography (CT) systems use x-rays to produce images of slices through a patient s anatomy. Despite having lower spatial resolution than other x-ray
More information2D, 3D CT Intervention, and CT Fluoroscopy
2D, 3D CT Intervention, and CT Fluoroscopy SOMATOM Definition, Definition AS, Definition Flash Answers for life. Siemens CT Vision Siemens CT Vision The justification for the existence of the entire medical
More informationData. microcat +SPECT
Data microcat +SPECT microcat at a Glance Designed to meet the throughput, resolution and image quality requirements of academic and pharmaceutical research, the Siemens microcat sets the standard for
More informationSECTION I - CHAPTER 2 DIGITAL IMAGING PROCESSING CONCEPTS
RADT 3463 - COMPUTERIZED IMAGING Section I: Chapter 2 RADT 3463 Computerized Imaging 1 SECTION I - CHAPTER 2 DIGITAL IMAGING PROCESSING CONCEPTS RADT 3463 COMPUTERIZED IMAGING Section I: Chapter 2 RADT
More informationRadiology Physics Lectures: Digital Radiography. Digital Radiography. D. J. Hall, Ph.D. x20893
Digital Radiography D. J. Hall, Ph.D. x20893 djhall@ucsd.edu Background Common Digital Modalities Digital Chest Radiograph - 4096 x 4096 x 12 bit CT - 512 x 512 x 12 bit SPECT - 128 x 128 x 8 bit MRI -
More informationDose Reduction in Helical CT: Dynamically Adjustable z-axis X-Ray Beam Collimation
Medical Physics and Informatics Original Research Christner et al. CT Dose Reduction Medical Physics and Informatics Original Research Downloaded from www.ajronline.org by 8.243.133.8 on 2/26/18 from IP
More informationQC by the MPE in Belgium
Acceptance testing of state-of-the-art CT scanners using a new national protocol: first experience on a large number of scanners of different make and model the working group Radiology of the Belgian Hospital
More informationPARTIAL-DATA INTERPOLATION DURING ARCING OF AN X-RAY TUBE IN A COMPUTED TOMOGRAPHY SCANNER JAISINGH RAJWADE
PARTIAL-DATA INTERPOLATION DURING ARCING OF AN X-RAY TUBE IN A COMPUTED TOMOGRAPHY SCANNER JAISINGH RAJWADE Bachelor of Engineering (B.E.) Industrial Electronics University of Pune, India November 1997
More informationCT Basics: Image Quality Module 6
Module 6 For educational and institutional use. This transcript is licensed for noncommercial, educational inhouse or online educational course use only in educational and corporate institutions. Any broadcast,
More informationSOMATOM Esprit A Bundle of Energy
SOMATOM Esprit A Bundle of Energy DATA SOMATOM Esprit An economical CT scanner designed for...... Excellent spiral image quality... A wide range of clinical applications... Value performance and reliabilty
More informationRadionuclide Imaging MII Single Photon Emission Computed Tomography (SPECT)
Radionuclide Imaging MII 3073 Single Photon Emission Computed Tomography (SPECT) Single Photon Emission Computed Tomography (SPECT) The successful application of computer algorithms to x-ray imaging in
More informationDOUBLE gated in vivo small animal cone beam micro
Low Dose Phase Correlated Cone Beam Micro CT of Small Animals Stefan Sawall, Frank Bergner, Robert Lapp, Markus Mronz, Marek Karolczak, Andreas Hess, and Marc Kachelrieß, Member, IEEE I. INTRODUCTION DOUBLE
More informationSPECT Reconstruction & Filtering
SPECT Reconstruction & Filtering Goals Understand the basics of SPECT Reconstruction Filtered Backprojection Iterative Reconstruction Make informed choices on filter selection and settings Pre vs. Post
More informationI. PERFORMANCE OF X-RAY PRODUCTION COMPONENTS FLUOROSCOPIC ACCEPTANCE TESTING: TEST PROCEDURES & PERFORMANCE CRITERIA
FLUOROSCOPIC ACCEPTANCE TESTING: TEST PROCEDURES & PERFORMANCE CRITERIA EDWARD L. NICKOLOFF DEPARTMENT OF RADIOLOGY COLUMBIA UNIVERSITY NEW YORK, NY ACCEPTANCE TESTING GOALS PRIOR TO 1st CLINICAL USAGE
More informationInterventional X-ray quality measure based on a psychovisual detectability model
Interventional X-ray quality measure based on a psychovisual detectability model Asli Kumcu, Benhur Ortiz-Jaramillo, Ljiljana Platisa, Bart Goossens, Wilfried Philips iminds-telin-ipi, Ghent University,
More informationWE MAKE RELATIONSHIP FOR LIFE.
2018 WE MAKE RELATIONSHIP FOR LIFE. www.sanrad.in Contents Corporate Profile 02 Company Profile Rich in heritage and synonymous with healthcare, the field of medicine is deeply ingrained in Sanrad s roots
More informationArchitecture of Quality Imaging Mary K. Henne, MS, CNMT, RDMS, RVT Ultrasound Education Specialist GE Healthcare
Architecture of Quality Imaging Mary K. Henne, MS, CNMT, RDMS, RVT Ultrasound Education Specialist GE Healthcare 2 DOC1292532 Architecture of Quality Imaging Agile Acoustic Architecture E-Series and XDclear
More informationTruly flexible to meet your clinical needs
Truly flexible to meet your clinical needs 2 Adapting to meet your needs Flexible Fast and responsive Excellent image quality Designed with ergonomic efficiency Equipped with dose management tools 3 Three
More informationNIH Public Access Author Manuscript Int J Cardiovasc Imaging. Author manuscript; available in PMC 2008 May 26.
NIH Public Access Author Manuscript Published in final edited form as: Int J Cardiovasc Imaging. 2001 August ; 17(4): 287 296. A comparison of prospective and retrospective respiratory navigator gating
More informationCHAPTER 2 COMMISSIONING OF KILO-VOLTAGE CONE BEAM COMPUTED TOMOGRAPHY FOR IMAGE-GUIDED RADIOTHERAPY
14 CHAPTER 2 COMMISSIONING OF KILO-VOLTAGE CONE BEAM COMPUTED TOMOGRAPHY FOR IMAGE-GUIDED RADIOTHERAPY 2.1 INTRODUCTION kv-cbct integrated with linear accelerators as a tool for IGRT, was developed to
More informationINNOVATION BY DESIGN. Toshiba A History of Leadership REMOTE CONTROL R/F SYSTEM
INNOVATION BY DESIGN For over 130 years, Toshiba has led the world in developing technology to improve the quality of life. This Made for Life TM commitment is reflected in our family of leading-edge imaging
More informationCT Basics: Data Acquisition Module 3
Module 3 Transcript For educational and institutional use. This transcript is licensed for noncommercial, educational inhouse or online educational course use only in educational and corporate institutions.
More informationDigital Radiographic Inspection replacing traditional RT and 3D RT Development
Digital Radiographic Inspection replacing traditional RT and 3D RT Development Iploca Novel Construction Meeting 27&28 March 2014 Geneva By Jan van der Ent Technical Authority International Contents Introduction
More information7/24/2014. Image Quality for the Radiation Oncology Physicist: Review of the Fundamentals and Implementation. Disclosures. Outline
Image Quality for the Radiation Oncology Physicist: Review of the Fundamentals and Implementation Image Quality Review I: Basics and Image Quality TH-A-16A-1 Thursday 7:30AM - 9:30AM Room: 16A J. Anthony
More informationFocal Spot Blooming in CT: We Didn t Know We Had a Problem Until We Had a Solution
Focal Spot Blooming in CT: We Didn t Know We Had a Problem Until We Had a Solution Cynthia H. McCollough, PhD, DABR, FAAPM, FACR Director, CT Clinical Innovation Center Professor of Medical Physics and
More informationOptimized CT metal artifact reduction using the Metal Deletion Technique (MDT)
Optimized CT metal artifact reduction using the Metal Deletion Technique (MDT) F Edward Boas, Roland Bammer, and Dominik Fleischmann Extended abstract for RSNA 2012 Purpose CT metal streak artifacts are
More informationMulti-Access Biplane Lab
Multi-Access Biplane Lab Advanced technolo gies deliver optimized biplane imaging Designed in concert with leading physicians, the Infinix VF-i/BP provides advanced, versatile patient access to meet the
More informationPhotomultiplier Tube
Nuclear Medicine Uses a device known as a Gamma Camera. Also known as a Scintillation or Anger Camera. Detects the release of gamma rays from Radionuclide. The radionuclide can be injected, inhaled or
More informationof sufficient quality and quantity
of sufficient quality and quantity The patient s body attenuates the beam as it passes though the body More energy is deposited in organs located near the entry of the beam than near the exit of the beam
More informationMR Advance Techniques. Flow Phenomena. Class II
MR Advance Techniques Flow Phenomena Class II Flow Phenomena In this class we will explore different phenomenona produced from nuclei that move during the acquisition of data. Flowing nuclei exhibit different
More informationSimultaneous Multi-Slice (Slice Accelerated) Diffusion EPI
Simultaneous Multi-Slice (Slice Accelerated) Diffusion EPI Val M. Runge, MD Institute for Diagnostic and Interventional Radiology Clinics for Neuroradiology and Nuclear Medicine University Hospital Zurich
More informationNew Technology in Nuclear Medicine
New Technology in Nuclear Medicine Reed G. Selwyn, PhD, DABR Vice Chair of Research & Imaging Sciences Associate Professor and Chief, Medical Physics Dept. of Radiology, University of New Mexico Objectives
More informationHigh Field MRI: Technology, Applications, Safety, and Limitations
High Field MRI: Technology, Applications, Safety, and Limitations R. Jason Stafford, Ph.D. The University of Texas M. D. Anderson Cancer Center, Houston, TX Introduction The amount of available signal
More informationImproved Tomosynthesis Reconstruction using Super-resolution and Iterative Techniques
Improved Tomosynthesis Reconstruction using Super-resolution and Iterative Techniques Wataru FUKUDA* Junya MORITA* and Masahiko YAMADA* Abstract Tomosynthesis is a three-dimensional imaging technology
More informationJanuary 2012 February 1, Release Date: Expiration Date:
CT for Technologists is a training program designed to meet the needs of radiologic technologists entering or working in the field of computed tomography (CT). This series is designed to augment classroom
More informationWide beam CT dosimetry. Elly Castellano
Wide beam CT dosimetry Elly Castellano Outline revision: CT dose indices wide-beam CT: the end of the road for CTDI? the IEC rescue plan for CTDI 100 the american way AAPM report 111 better estimates of
More informationCT Basics: Computed Tomography Fundamentals Module 1
Module 1 Transcript For educational and institutional use. This transcript is licensed for noncommercial, educational inhouse or online educational course use only in educational and corporate institutions.
More informationDistributed source x-ray tube technology for tomosynthesis imaging
Distributed source x-ray tube technology for tomosynthesis imaging Authors: F. Sprenger a*, X. Calderon-Colon b, Y. Cheng a, K. Englestad a, J. Lu b, J. Maltz c, A. Paidi c, X. Qian b, D. Spronk a, S.
More informationQuantitation of clinical feedback on image quality differences between two CT scanner models
Received: 4 August 2016 Revised: 4 November 2016 Accepted: 12 December 2016 DOI: 10.1002/acm2.12050 MEDICAL IMAGING Quantitation of clinical feedback on image quality differences between two CT scanner
More informationEvaluation Report. Single Slice CT Scanner Comparison Report Version (Free to the NHS) ImPACT report. MDA Evaluation Report MDA 03022
February 2003 Evaluation Report NUMBER MDA 03022 Single Slice CT Scanner Comparison Report Version 8 ImPACT report MDA Evaluation Report MDA 03022 Crown Copyright 100 (Free to the NHS) WHAT YOU CAN EXPECT
More informationP a n o r a m i c a n d C e p h a l o m e t r i c X - r a y s
AN ALL AROUND PERFECT PICTURE. The perfect combination of image quality, efficiency and design. P a n o r a m i c a n d C e p h a l o m e t r i c X - r a y s Air Techniques experts in digital imaging ProVecta
More informationSOMATOM Sensation 4 Computed Tomography System for Multislice Spiral Scanning
SOMATOM Sensation 4 Computed Tomography System for Multislice Spiral Scanning SOMATOM Sensation 4 Computed Tomography System for Multislice Spiral Scanning Siemens has been a global player in CT for more
More informationX-RAY COMPUTED TOMOGRAPHY
X-RAY COMPUTED TOMOGRAPHY Bc. Jan Kratochvíla Czech Technical University in Prague Faculty of Nuclear Sciences and Physical Engineering Abstract Computed tomography is a powerful tool for imaging the inner
More informationX-ray detectors in healthcare and their applications
X-ray detectors in healthcare and their applications Pixel 2012, Inawashiro September 4th, 2012 Martin Spahn, PhD Clinical applications of X-ray imaging Current X-ray detector technology (case study radiography
More informationP a n o r a m i c a n d C e p h a l o m e t r i c X - r a y s
AN ALL AROUND PERFECT PICTURE. The perfect combination of image quality, efficiency and design. P a n o r a m i c a n d C e p h a l o m e t r i c X - r a y s Air Techniques experts in digital imaging ProVecta
More informationCT Basics: Equipment and Instrumentation Module 2
Module 2 Transcript For educational and institutional use. This transcript is licensed for noncommercial, educational in-house or online educational course use only in educational and corporate institutions.
More informationThe SENSE Ghost: Field-of-View Restrictions for SENSE Imaging
JOURNAL OF MAGNETIC RESONANCE IMAGING 20:1046 1051 (2004) Technical Note The SENSE Ghost: Field-of-View Restrictions for SENSE Imaging James W. Goldfarb, PhD* Purpose: To describe a known (but undocumented)
More informationMeasurement of table feed speed in modern CT
JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 15, NUMBER 3, 2014 Measurement of table feed speed in modern CT Atsushi Fukuda, 1,2a Pei-Jan P. Lin, 3 Kosuke Matsubara, 2 Tosiaki Miyati 2 Department
More informationIndustry Breakthrough
Industry Breakthrough Dynamic SPECT Acquisition Quantifying Myocardial Blood Flow Nuclear Cardiology in the 21st Century In the 21st century, most nuclear cameras are still relying on a technology invented
More informationPhilips XPER FD10C R7.0.4
Philips XPER FD10C R7.0.4 Reconditioned 2005 System- Upgraded to R7 in Oct 2010 The Allura Xper FD10 (Ceiling) single-plane cardiovascular system is comprised of a ceiling mounted C-arm stand and digital
More information1. Queries are issued to the image archive for information about computed tomographic (CT)
Appendix E1 Exposure Extraction Method examinations. 1. Queries are issued to the image archive for information about computed tomographic (CT) 2. Potential dose report screen captures (hereafter, dose
More informationCurrent technology in digital image production (CR/DR and other modalities) Jaroonroj Wongnil 25 Mar 2016
Current technology in digital image production (CR/DR and other modalities) Jaroonroj Wongnil 25 Mar 2016 Current technology in digital image production (CR/DR and other modalities) 2/ Overview Digital
More informationAuthors: Cabral, Ricardo 1 ; Carvoeiras, Pedro 2 ; Fatana, João 2, ; Alves, Rita 1. 1 Centro Hospitalar Lisboa Norte - Hospital de Santa Maria; 2
Authors: Cabral, Ricardo 1 ; Carvoeiras, Pedro 2 ; Fatana, João 2, ; Alves, Rita 1. 1 Centro Hospitalar Lisboa Norte - Hospital de Santa Maria; 2 Medical Consult, SA; Establish a method to correlate image
More informationResona 6 Premium Ultrasound System
The system Resona 6 is the newly developed, unique result of the mergence of leading companies Mindray Bio-medical Electronics Co. Ltd. and ZONARE Medical Systems, Inc.. By additions to the core competencies
More informationWilliam Hallet - PRIMA IV 1
Quantitative and application specific imaging PET: the measurement process,reconstruction, calibration, quantification Dr William Hallett Centre for Imaging Sciences Imperial College Hammersmith Hospital
More informationAdvanced Noise Reduction Processing for X-ray CT System with Iterative Processing. Koichi Hirokawa MEDIX VOL. 56 P.43 P.46
Advanced Noise Reduction Processing for X-ray CT System with Iterative Processing Taiga Goto Koichi Hirokawa Hisashi Takahashi MEDIX VOL. 56 P.43 P.46 Advanced Noise Reduction Processing for X-ray CT System
More information23 CP Clarify Enhanced US Volume Image and Frame Type Values 3 and 4
23 CP-1463 - Clarify Enhanced US Volume Image and Frame Type Values 3 and 4 Page 1 1 Status Finale Text 2 Date of Last Update 2015/11/10 3 Person Assigned David Clunie 4 mailto:dclunie@dclunie.com 5 Submitter
More informationComputerized Medical Imaging and Graphics
Computerized Medical Imaging and Graphics 36 (2012) 387 395 Contents lists available at SciVerse ScienceDirect Computerized Medical Imaging and Graphics jo ur n al homep age : www.elsevier.com/locate/compmedimag
More information23 CP Clarify Enhanced US Volume Image and Frame Type Values 3 and 4
23 CP-1463 - Clarify Enhanced US Volume Image and Frame Type Values 3 and 4 Page 1 1 Status Letter Ballot 2 Date of Last Update 2015/09/16 3 Person Assigned David Clunie 4 mailto:dclunie@dclunie.com 5
More informationGE Healthcare. Discovery MI. with LightBurst Digital 4-Ring Detector
GE Healthcare Discovery MI with LightBurst Digital 4-Ring Detector Discovery TM MI is a PET/CT system conceptualized with lofty goals, equal only to your own. It was created to help you diagnose and stage
More informationThe future of nuclear imaging is clear
Cardius X-ACT The future of nuclear imaging is clear Increased regulations, growing competition, and concerns about radiation exposure are just a sampling of the current challenges facing the nuclear medicine
More information