DK Guard Ring Polarographic Cell

Size: px
Start display at page:

Download "DK Guard Ring Polarographic Cell"

Transcription

1 REHDER DEVELOPMENT COMPANY 2139 Klondike Rd W Lafayette, IN USA Tel: (765) Fax: (765) Web: t.houck@rehder-dev.com Technical Innovations DK Guard Ring Polarographic Cell Instruction Manual Sept 19, 2013 Permeometer is a trademark of Createch. 1 of 8

2 REHDER POLAROGRAPHIC CELLS Rehder Polarographic Cells are designed to function using standard polarographic measuring procedures as described in the March 1982, March 1984 and October 1987 issues of the International Contact Lens Clinic. Great care must be taken to ensure that the 24 karat gold and pure silver electrodes do not become scratched. The Rehder Guard Ring Polarographic Cell enables polarographic readings to be taken without the error created by edge effect. This eliminates the need to calculate the value of the edge effect and to adjust your data. These cells are designed to function using standard polarographic measuring procedures. During normal use the surface of the silver will become coated with Silver Chloride. Cleaning after every 8 or 10 lenses should be sufficient to keep the cell in proper working condition. The tarnished surface may be removed by gently rubbing with a soft, high quality, facial tissue and fine silver cleaning polish. Caution: After polishing, you must thoroughly remove any "Tarnish Guard", or any substance designed to remain on the surface to prevent tarnishing, from the Cell working surface before it is put into service. Ethanol may be used on the Cell working surface but should be avoided on the Cell base block. After polishing, the cell should be thoroughly rinsed with water before it is used to take polarographic measurements. Make sure that the polarographic amplifier is turned off or that the Cell is disconnected from the amplifier prior to polishing the Cell. This will help to avoid plating the surface of the gold electrode with silver. Excessive polishing will wear away the working surface of the cell and may expose an increased amount of the gold surface. If the cell becomes excessively worn it should be resurfaced or replaced. All Guard Ring Cells are fitted with a solid state temperature sensor embedded within the silver anode of each Cell. The sensors are temperature calibrated to operate with the temperature readout on the Createch 201T Permeometer. Each Cell has a connector attached to mate with the Permeometer and is ready for use. The new cells have a larger base block and will not fit into the old cell mounting fixtures. Older cells have a smaller base block and will require an adapter to fit properly in the new mounting fixtures. Radius Cell Flat Cell Permeometer is a trademark of Createch. 2 of 8

3 The switch found on the calibration box, about mid way along the lead from the cell, is used to select different options for the current produced by the guard ring. If the guard ring switch is ON the meter reading shows the sum of the currents produced by the cathode and the guard ring. (As if the cell has a 5.5mm diameter cathode.) If the switch is in the OFF position, the guard ring is inactive and no current flows through that element. The cell then performs like a standard cell without a guard ring. If the switch is in the GND (Ground) position, the guard ring is active, current flows through the element, but the current is not read by the meter. This is the condition in which the guard ring eliminates the edge effect from the cathode. Permeometer is a trademark of Createch. 3 of 8

4 DETERMINING THE "DARK CURRENT" OF REHDER POLAROGRAPHIC CELLS The "Dark Current" value should be subtracted from current readings taken during Transmissibility measurements. The "Offset" dial on the Permeometer may be used to automatically subtract this value. 1. Set up as illustrated below and put the Cell in the mounting fixture, compressing the "sandwich" with the plunger. 2. Put the Cell and Fixture in your 35 C 3. Plug Cell into the Permeometer. 4. Switch on the Permeometer and make sure the "Offset" dial is set to ZERO. 5. Run the Cell until the oxygen in the cigarette paper has been consumed and the current reading stabilizes. (This may take 45 minutes or longer) A chart recorder plugged into the Permeometer will make it easier to monitor this process. 6. The stabilized current reading is the "Dark Current" for the Cell and each Cell's value should remain constant. Record this value for future use. You should expect a "Dark Current" reading of approximately 0.10 µa to 0.20 µa. Note: Only nonpreserved saline solution should be used on your samples or on the polarographic cell. (Unisol or equivalent is satisfactory). Preserved saline solutions may contain mercury compounds that will interfere with the polarographic electrochemical reduction of oxygen at the cathode. The mercury compound may also attack the silver anode. Aluminum Foil PMMA Lens Nonpreserved Saline Saturated Cigarette Paper Nonpreserved Saline Saturated Cigarette Paper PMMA Lens Flat Cell Glass Microscope Slide Radius Cell Permeometer is a trademark of Createch. 4 of 8

5 REHDER DEVELOPMENT COMPANY 2139 Klondike Rd W Lafayette, IN USA Tel: (765) Fax: (765) Web: rehder - dev. com t.houck@ rehder - dev. Technical Innovations com For Flat Cell Strech nylon mesh (panty hose type) over tube ends and secure with thread in groove. Nylon mesh on tube end holds sample onto Cell For Radius Cell Sample Mounting Position Radius Cell Cell Mounting Fixture Flat Cell Radius Cell Permeometer is a trademark of Createch. 5 of 8

6 Cleaning the Polarographic Cells: During normal operation of the polarographic cells the surface of the silver anode will become coated with silver chloride. The silver chloride coating will appear as a dull gray, tarnished, surface on the anode. If the silver chloride coating is allowed to build up on the anode the cell will not perform properly. Therefore, you must polish off the silver chloride coating to expose the clean surface of the silver. It is important that the cell be disconnected from the Permeometer or that the Permeometer is turned off during the polishing operation. If the current is turned on to the cell while polishing, a coating of silver will be Plated onto the gold cathode. You should continue polishing until the tarnished surface has been completely removed from the anode and any silver deposits on the gold have been removed. When finished, you should see bright silver and bright gold. I use two types of polish when cleaning tarnished cells. The first, is the Boston Lens cleaning polish which is only very, mildly abrasive. The second, which is slightly more abrasive, is a metal polishing product called Simichrome Polish. This polish is manufactured in Germany and can be found in automotive and hardware stores. I think that most fine silver polishes should work OK. You should test the polish on the outside of the cell to make sure there is not an adverse chemical reaction with the insulating material before using the polish on the cell surface. You can also use the silver polishing wadding material. Caution: Do Not use polish that includes any type of "Tarnish Guard" or any substance designed to remain on the surface to prevent tarnishing. Please keep in mind that these cells are consumable items. Each time the silver chloride is removed, a small amount of the surface of the silver disappears. Eventually, the silver surface will be eroded to the point where the cell surface should be remachined. Depending on the eroded condition of the cell, and some other factors, a cell may not be restorable. The silver anode on the polarographic cell is several millimeters thick. However, the temperature sensor is located within the silver anode and is very near the surface of the cell in order to provide quick response and good resolution. The position of the temperature sensor and the thickness of the gold usually allows a cell surface to be remachined one time before it is no longer usable, this, of course, depends on the amount of erosion. If the cell becomes excessively eroded it should be discarded. It should take hundreds of lenses being tested to reach the point where the cell needs to be remachined. I would recommend remachining if the silver on the cell surface becomes eroded by about 150 microns. The gold cathode will not be eroded nearly as quickly as the silver. If you notice that the gold is protruding slightly above the radius of the silver, then it is probably time to have the cell remachined. Permeometer is a trademark of Createch. 6 of 8

7 Set Up Instructions for Dk/L Measurements: Three papers from the International Contact Lens Clinic, March 1982, March 1984 and October 1987, that were authored by Dr. Irving Fatt and others, should help you understand the process of determining Dk/L. Setting up the equipment is not complicated. The operating instructions for the Electronic Thickness Gauge and Permeometer should help you get things set up. You can also refer to information in the web site and information found in the ISO standard. The Guard Ring Cell plugs into the front of the Permeometer in either the "A" or "B" position. The cell has the temperature sensor built in and temperature of the Cell will be displayed on the Permeometer. You must select the A or B channel on the Permeometer. If you have purchased the computer interface and software so you will need to follow the instructions for connecting the computer and installing the software. By using the computer, you will not need to use a chart recorder to track oxygen consumption or track temperature so you will not need the two phone plugs that were supplied. The phone plug connectors will plug into the rear of the Permeometer if you do choose to use a chart recorder. You will need to attach some nylon mesh to the end of the tube on the Cell Mounting Fixture. The mesh holds the sample in place on the Cell. Refer to the diagram in the Cell Operating Instructions that shows how the mesh is installed and how the sample is mounted. The Mounting Fixture with the sample mounted, needs to be placed in an environmental chamber to control temperature and humidity. Prior to mounting, the Sample, the Fixture and Cell should all be stored in the chamber at the test temperature. This will help to avoid long delays in bringing the sample to the desired temperature for testing. Dr. Fatt usually kept all the vials with his samples to be tested in the chamber so everything was at temperature and ready for testing. The Electronic Thickness Gauge is used to measure the thickness of the sample. It is a stand alone instrument, it has been calibrated, and is ready for use. You can determine "Transmissibility" of your sample from the data taken from the Permeometer. To determine "Permeability" of the material you must know the thickness. Permeometer is a trademark of Createch. 7 of 8

8 Set Up for Dk Measurements with Computer Interface Incubator Cell with Sample mounted in Fixture Inside Incubator Permeometer With Computer Interface Computer Permeometer is a trademark of Createch. 8 of 8

Calibration and Gage Block Use

Calibration and Gage Block Use REHDER DEVELOPMENT COMPANY 2139 Klondike Rd W Lafayette, IN 47906 USA Tel: (765) 418-1997 Fax: (765) 463-3779 Web: http://www.rehder-dev.com Email: t.houck@rehder-dev.com Technical Innovations Calibration

More information

Anodising is the process where Aluminum Oxide film is created on the section to avoid corrosion. (The tanks used are nos. 16, 17, 19 and 20)

Anodising is the process where Aluminum Oxide film is created on the section to avoid corrosion. (The tanks used are nos. 16, 17, 19 and 20) ABOUT ANODISING Our New anodizing plant was commissioned in October 2012. Ours is a fully automated plant and the plant is completely designed and commissioned by the Monti Engineering from Italy. The

More information

Termination Procedure

Termination Procedure Connector Piece Parts Contact/Connector Head Twist On Nut MX MX Boot Procedure Chart Procedure Tool Required Tool Part Number Cable Preparation & Fiber Cleaning Jacket Stripper 86710-0004 Cable Preparation

More information

WARNING. Failure to observe these instructions could lead to severe injury or death.

WARNING. Failure to observe these instructions could lead to severe injury or death. INSTALLATION INSTRUCTIONS WINCH MOUNTING KIT Part Number: 80156, 80160 Application: 2008 Ford F150 Your safety, and the safety of others, is very important. To help you make informed decisions about safety,

More information

SUPERSEDED. NOT the LATEST REVISION

SUPERSEDED. NOT the LATEST REVISION Figure 1 OPTIMATE 2.5mm bayonet ceramic connector kits listed in Figure 1 are designed to be applied to fiber optic cable. Coupling Receptacle Kit 501381 1 is used to mate two bayonet connectors in free

More information

INSTALLATION INSTRUCTIONS

INSTALLATION INSTRUCTIONS INSTALLATION INSTRUCTIONS Brush Guard Kit Kit no. 67730 (black), 67731 (stainless) for GMC Trucks Your safety, and the safety of others, is very important. To help you make informed decisions about safety,

More information

SUPERSEDED REVISION. Reasons for reissue of this instruction sheet are provided in Section 7, REVISION SUMMARY.

SUPERSEDED REVISION. Reasons for reissue of this instruction sheet are provided in Section 7, REVISION SUMMARY. PRO BEAM Jr. EB cable plug connectors are designed to be installed onto jacketed fiber optic cable with KEVLAR strength members. The connector must be assembled using a cable plug connector shell kit,

More information

INSTALLATION INSTRUCTIONS

INSTALLATION INSTRUCTIONS INSTALLATION INSTRUCTIONS Trans4mer Gen II Mount System For 2011-2012 Dodge Ram HD 4500/5500 Bracket Kit 90160 Your safety, and the safety of others, is very important. To help you make informed decisions

More information

INSTALLATION INSTRUCTIONS

INSTALLATION INSTRUCTIONS INSTALLATION INSTRUCTIONS Trans4mer Gen II Mount System For 2008 Ford Super Duty Kit 80140 (Large Frame Black) and Kit 80150 (Mid Frame Black) and Kit 80155 (Mid Frame Stainless) Your safety, and the safety

More information

A Guide to Calibration on the BioFlo 120 and BioFlo 320: Dissolved Oxygen Sensors

A Guide to Calibration on the BioFlo 120 and BioFlo 320: Dissolved Oxygen Sensors SHORT PROTOCOL No. 40 I May 2017 A Guide to Calibration on the BioFlo 120 and BioFlo 320: Dissolved Oxygen Sensors Stacey Willard Eppendorf Inc., USA Abstract Dissolved oxygen (DO) is a critical process

More information

Module-3 Lecture-7 Limit guage-2. (Refer Slide Time: 00:12)

Module-3 Lecture-7 Limit guage-2. (Refer Slide Time: 00:12) Metrology Prof. Dr. Kanakuppi Sadashivappa Department of Industrial and Production Engineering Bapuji Institute of Engineering and Technology-Davangere Module-3 Lecture-7 Limit guage-2 (Refer Slide Time:

More information

Line-Following Robot

Line-Following Robot 1 Line-Following Robot Printed Circuit Board Assembly Jeffrey La Favre October 5, 2014 After you have learned to solder, you are ready to start the assembly of your robot. The assembly will be divided

More information

Figure 1. Oil-immersion objectives available for use with the Lionheart FX.

Figure 1. Oil-immersion objectives available for use with the Lionheart FX. Tech Note Oil Objective Introduction The Lionheart FX automated imager is compatible with high numerical aperture oil immersion objectives. These objectives offer magnification up to 100X and significantly

More information

MultiPrep Procedure. Backside Thinning of a Flip-Chip Device G. D. Liechty, C. A. Smith, Allied High Tech Products, Inc.

MultiPrep Procedure. Backside Thinning of a Flip-Chip Device G. D. Liechty, C. A. Smith, Allied High Tech Products, Inc. MultiPrep Procedure Backside Thinning of a Flip-Chip Device G. D. Liechty, C. A. Smith, Allied High Tech Products, Inc., August 2003 Overview When thinning electronic devices for various analyses, including

More information

The Elements of Jewelry By ReadWorks

The Elements of Jewelry By ReadWorks The Elements of Jewelry The Elements of Jewelry By ReadWorks Stroll by a jewelry store on any day, and you ll see cases full of shiny items. Light glints off of gold rings, and silver necklaces sparkle

More information

WEAR SAFETY GLASSES WHEN INSTALLING THIS KIT.

WEAR SAFETY GLASSES WHEN INSTALLING THIS KIT. INSTALLATION INSTRUCTIONS Trans4mer Mounting Systems Part No. 29753 (black) Part No. 65654 (stainless) for full size GM pickups, and Blazer, Yukon, Suburban, Tahoe As you read these instructions, you will

More information

PRL GLASS PANIC HANDLES DOOR MANUAL PL100 PANIC HANDLE PL100-A SERIES. For all glass doors

PRL GLASS PANIC HANDLES DOOR MANUAL PL100 PANIC HANDLE PL100-A SERIES. For all glass doors DOOR MANUAL PL100 PANIC HANDLE SERIES For all glass doors Content of this manual: DOOR FABRICATION ( panic handle installation ) - page 3 Parts list, tools & supplies required Installation of handle to

More information

Signet 2610 Process Optical Dissolved Oxygen Sensor

Signet 2610 Process Optical Dissolved Oxygen Sensor Signet 2610 Process Optical Dissolved Oxygen Sensor *3-2610.090* 3-2610.090 Rev. A 04/12 English Operator's Manual Introduction System Description Your new RDO Pro is a rugged, reliable sensor designed

More information

Non-Contact Capacitance Gauging Instrument & Series 2800 Capacitive Probes

Non-Contact Capacitance Gauging Instrument & Series 2800 Capacitive Probes 4810 Non-Contact Capacitance Gauging Instrument & Series 2800 Capacitive Probes Sub nanometer resolution for ultra-precise measurements Exceptional temperature stability Wide variety of precision capacitive

More information

INSTALLATION INSTRUCTIONS

INSTALLATION INSTRUCTIONS INSTALLATION INSTRUCTIONS Trans4mer Winch Mount For 2007-2010 Chevrolet Silverado HD and GMC Sierra HD Bracket Kit 90140 Your safety, and the safety of others, is very important. To help you make informed

More information

Table of Contents. B. Base Tool Changer...2 MC-16 Manual Tool Changer...2

Table of Contents. B. Base Tool Changer...2 MC-16 Manual Tool Changer...2 Table of Contents B. Base Tool Changer...2 MC-16 Manual Tool Changer...2 1. Product Overview... 2 1.1 Master Plate Assembly... 2 1.1.1 Optional Ratchet Knob... 2 1.2 Tool Plate... 3 1.3 Optional Modules...

More information

Profiles for floors of same height Proclassic Proclassic F

Profiles for floors of same height Proclassic Proclassic F PROFILPAS S.P.A. VIA EINSTEIN, 38 35010 CADONEGHE (PADOVA) ITALY TEL. +39 (0)49 8878411 +39 (0)49 8878412 FAX. +39 (0)49-706692 EMAIL: INFO@PROFILPAS.COM Profiles for floors of same height Proclassic Proclassic

More information

INSTALLATION INSTRUCTIONS

INSTALLATION INSTRUCTIONS INSTALLATION INSTRUCTIONS Trans4mer Mounting System Dodge Ram Kit No. 65220, 73132, 75525, 76253 (black) or 65221, 73133, 75530, 76254 (stainless) As you read these instructions, you will see NOTES, CAUTIONS

More information

Table of Contents. B. Base Tool Changer...2 MC-36 Manual Tool Changer...2

Table of Contents. B. Base Tool Changer...2 MC-36 Manual Tool Changer...2 Table of Contents B. Base Tool Changer...2 MC-36 Manual Tool Changer...2 1. Product Overview... 2 1.1 Master Plate Assembly... 2 1.1.1 Optional Ratchet Knob... 2 1.2 Tool Plate... 3 1.3 Optional Modules...

More information

Table of Contents. B. Base Tool Changer...2 MC-6 Manual Tool Changer...2

Table of Contents. B. Base Tool Changer...2 MC-6 Manual Tool Changer...2 Table of Contents B. Base Tool Changer...2 MC-6 Manual Tool Changer...2 1. Product Overview... 2 1.1 Master Plate Assembly... 2 1.2 Tool Plate Assembly... 3 1.3 Optional Modules... 3 2. Installation...

More information

Electron Microscopy Sciences

Electron Microscopy Sciences Electron Microscopy Sciences INSTRUCTIONAL MANUAL CAT. 50120-10 Model 900 Grinding and Polishing Machine P.O. Box 550 s1560 Industry Road s Hatfield PA 19440 1 Introduction The Model 900 is a small, compact

More information

H2-50 Hydrogen Generator Field Update

H2-50 Hydrogen Generator Field Update This field update is intended to provide extra protection in the event the H2 generation cell fractures or cracks. Please add the additional parts to your H2-50 as soon as possible. Please take a digital

More information

RING CUTTING SYSTEM OPERATOR S MANUAL

RING CUTTING SYSTEM OPERATOR S MANUAL . RING CUTTING SYSTEM OPERATOR S MANUAL INTRODUCTION GEM is not a modified power tool; rather, it s a device that has been designed specifically to cut rings quickly and safely. The abrasive quality of

More information

Ribcage Installation. Part 2 - Assembly. Back-Bone V1.06

Ribcage Installation. Part 2 - Assembly. Back-Bone V1.06 Ribcage Installation Part 2 - Assembly Back-Bone V1.06 Contents Section 1 Before You Get Started... 2 Included With Your Kit:... 2 Figure: A... 3 CAUTION!... 4 Note:... 4 Tools Required... 5 Section 2:

More information

An Adjustable Threading Feed Attachment for a Lathe Without Metric Threading Capability, by Ted Clarke

An Adjustable Threading Feed Attachment for a Lathe Without Metric Threading Capability, by Ted Clarke An Adjustable Threading Feed Attachment for a Lathe Without Metric Threading Capability by Ted Clarke Metric pitch threads, with the exception of the Royal Microscopical Society (RMS) 36 threads per inch

More information

Model R7900. Instruction Manual. Ultrasonic Thickness Gauge. reedinstruments. www. com

Model R7900. Instruction Manual. Ultrasonic Thickness Gauge. reedinstruments. www. com Model R7900 Ultrasonic Thickness Gauge Instruction Manual reedinstruments com Table of Contents Features... 3 Specifications...4-5 Instrument Description... 6 Operating Instructions...7-10 Adjusting the

More information

NNIN Nanotechnology Education

NNIN Nanotechnology Education NNIN Nanotechnology Education Teacher s Guide Hiding Behind the Mask Purpose: This lab is designed to help students understand one aspect of semiconductor manufacture: selective layering. Level : High

More information

FIXED PANEL SLIDER QCI5241

FIXED PANEL SLIDER QCI5241 INSTALLATION INSTRUCTIONS FIXED PANEL SLIDER QCI5241 FRAMELESS PANEL / DOOR / PANEL FRAMELESS DOOR / PANEL QCI5241 REV. 0 Page 1 Certified 06/16/2016 Parts List *Quantities may vary QCI5241 REV. 0 Page

More information

OPERATION MANUAL. Model LET-SW5. April Far West Technology, Inc. 330 South Kellogg Ave, Suite D Goleta, CA 93117

OPERATION MANUAL. Model LET-SW5. April Far West Technology, Inc. 330 South Kellogg Ave, Suite D Goleta, CA 93117 OPERATION MANUAL Model LET-SW5 April 2010 Far West Technology, Inc. 330 South Kellogg Ave, Suite D Goleta, CA 93117 GENERAL INFORMATION This instrument is manufactured in the United States of America by:

More information

CHROME HEARTS EYEWEAR WARRANTY

CHROME HEARTS EYEWEAR WARRANTY CHROME HEARTS EYEWEAR WARRANTY Chrome Hearts Eyewear guarantees your eyewear for a period of one year. Under this guarantee, should any part prove defective in either materials or manufacture, it will

More information

Specifications: Hole Sizes: Selector Pin Hole: 3/8 (0.375 ) Trigger Pin Hole: 5/32 (0.156 ) Hammer Pin Holes: 5/32 (0.156 )

Specifications: Hole Sizes: Selector Pin Hole: 3/8 (0.375 ) Trigger Pin Hole: 5/32 (0.156 ) Hammer Pin Holes: 5/32 (0.156 ) 80% Arms AR-15 Jig manual Thank you for purchasing our AR- 15 jig. This precision self- aligning jig will allow you to finish your 80% Arms and other 80% lowers using only a drill press and a few common

More information

Seapoint Rhodamine Fluorometer

Seapoint Rhodamine Fluorometer Seapoint Rhodamine Fluorometer User Manual Standard Version 6/00 Dimensions Open Configuration (no pump) 2.5" (6.4 cm) 6.6" (16.8 cm) 5.1" (13.0 cm) Pumped Configuration Figure 1. 1/2 Scale Drawing. Page

More information

LPK1550 Hydraulic Crimping Tool 15-ton

LPK1550 Hydraulic Crimping Tool 15-ton SERVICE MANUAL LPK1550 Hydraulic Crimping Tool 15-ton Serial Code FYF Read and understand all of the instructions and safety information in this manual before operating or servicing this tool. Register

More information

Flat Sample Holder Part Number

Flat Sample Holder Part Number Flat Sample Holder Part Number 990-00403 (Patent Pending) Operator s Manual Copyright 2016 Gamry Instruments, Inc. April 8, 2016 988-00043 Rev. 1 If You Have Problems Please visit our service and support

More information

INSTALLATION INSTRUCTIONS

INSTALLATION INSTRUCTIONS INSTALLATION INSTRUCTIONS Trans4mer Grille Guard/Winch Mount For 2007 Chevrolet Silverado 1500 and GMC Sierra 1500 Kit 75230 (Black) and 76007 (Stainless) Your safety, and the safety of others, is very

More information

Model 50 Ground Loop ph Sensor Model 55 Solution Ground ph Sensor

Model 50 Ground Loop ph Sensor Model 55 Solution Ground ph Sensor Model 50 Ground Loop ph Sensor Model 55 Solution Ground ph Sensor Features Patented Plunger 1 ph Electrode Design Patented Porous 2 Teflon Liquid Junctions Double Junction Reference Cell Integral Differential

More information

SC One Piece Connectors Termination Procedure Series: 728-NNN0-NNN0N

SC One Piece Connectors Termination Procedure Series: 728-NNN0-NNN0N SC One Piece Connectors Termination Procedure Series: 728-NNN0-NNN0N C B A Version Update address and fax number Revise Format Initial Release Revision History Summary 10/26/18 12/1/03 1997 Issue Date

More information

Profiles for wooden and laminate floors Unisystem Plus - Unisystem

Profiles for wooden and laminate floors Unisystem Plus - Unisystem PROFILPAS S.P.A. VIA EINSTEIN, 38 35010 CADONEGHE (PADOVA) ITALY TEL. +39 (0)49 8878411 +39 (0)49 8878412 FAX. +39 (0)49-706692 EMAIL: INFO@PROFILPAS.COM s for wooden and laminate floors Unisystem Plus

More information

Power wire and the big three upgrade and why it s important.

Power wire and the big three upgrade and why it s important. Power wire and the big three upgrade and why it s important. We all know that it takes wires to make electronics work. But what it the reason there are different sizes of wires? Does it make any difference?

More information

Band-Master ATS Nano Pneumatic Banding Tool Operating Instructions

Band-Master ATS Nano Pneumatic Banding Tool Operating Instructions Band-Master ATS 601-118 Nano Pneumatic Banding Tool CONTENTS 601-118 Overview... 3 Safety.... 5 Initial Tool Set-up... 5 Regulator assembly mounting... 5 Attach tool head to regulator.... 6 Operating instructions...

More information

Line volt rail & fixtures. Low volt quick connect fixtures with transformers

Line volt rail & fixtures. Low volt quick connect fixtures with transformers INSTRUCTIONS: Line Volt Flexrail2 Safety Instructions: A qualified electrician must install system only. System is intended for installation in accordance with National Electric Code, local and Federal

More information

Introduction. Instructional Objectives. Materials. Procedure. I. Microscope Parts and Function. Honors Biology

Introduction. Instructional Objectives. Materials. Procedure. I. Microscope Parts and Function. Honors Biology Honors Biology Introduction to the Microscope Lab Activity This lab was created by Mr. Buckley from Edward Knox High School. Credit is given for this original activity to Mr. Buckley. Introduction "Micro"

More information

User Manual. Digital Compound Binocular LED Microscope. MicroscopeNet.com

User Manual. Digital Compound Binocular LED Microscope. MicroscopeNet.com User Manual Digital Compound Binocular LED Microscope Model MD82ES10 MicroscopeNet.com Table of Contents i. Caution... 1 ii. Care and Maintenance... 2 1. Components Illustration... 3 2. Installation...

More information

Technical Newsletter

Technical Newsletter Waterbase and Discharge Ink Printing Objectives: To improve production performance, minimize down time, and optimize screen life by preparing durable stencils with quality products and screen making techniques.

More information

Comp-DS Driveshaft. User Manual B

Comp-DS Driveshaft. User Manual B Comp-DS Driveshaft User Manual 2010-1378B Driveshaft Parts List 1.22B 1.21C 1.22C 1.22D 1.21A 1.21B 1.22A 1.1 Figure 1 1.0 Complete Driveshaft 1.1 Tube and Flange Assembly 1.2 Coupling Assembly (2 required

More information

8mm x 25mm "Z" Bolt Plates. (2) Tube Spacers. (2) 12mm Bolt Plates w/ Nut

8mm x 25mm Z Bolt Plates. (2) Tube Spacers. (2) 12mm Bolt Plates w/ Nut PARTS LIST: 1 Grille Guard 10 12mm Lock Washers 1 Driver/Left Side Frame Mounting Bracket 8 12mm Hex Nuts 1 Passenger/Right Side Frame Mounting Bracket 2 10-1.50mm x 25mm Button Head Bolts 1 Driver/Left

More information

Differential ph Design Overcomes Common ph Sensor Challenges

Differential ph Design Overcomes Common ph Sensor Challenges APPLICATION NOTE Differential ph Design Overcomes Common ph Sensor Challenges Conventional ph Measurement Methodology All ph measurement systems operate on the principle of an electrochemical cell; that

More information

TUpH 396RVP and 398RVP Retractable ph/orp Sensors

TUpH 396RVP and 398RVP Retractable ph/orp Sensors Instruction Sheet TUpH Sensors LIQ_MAN_ABR_396RVP-398RVP September 2013 TUpH 396RVP and 398RVP Retractable ph/orp Sensors Specifications SPECIFICATIONS 396RVP 398RVP Measurements and Ranges ph: 0-14; ORP:

More information

Rebuild 19mm & 25mm Sensors

Rebuild 19mm & 25mm Sensors Instruction Manual QUICK GUIDE Changing the Membrane Chapter 3 Polarizing Chapter 4 Replacement Parts Chapter 5 OxyProbe Rebuild 19mm & 25mm Sensors Doc No. L2553 12/2000 OxyProbe is a registered trademark

More information

UTILIBAY UTB2 HIGH BAY WARNINGS

UTILIBAY UTB2 HIGH BAY WARNINGS UTILIBAY HIGH BAY WARNINGS WARNING: This product can expose you to chemical including lead and lead compounds, which are known to the State of California to cause cancer and birth defects or other reproduction

More information

4 - Channel SWIR Light Sensor

4 - Channel SWIR Light Sensor SKR 1870 Skye Instruments Ltd., 21 Ddole Enterprise Park, Llandrindod Wells, Powys LD1 6DF UK Tel: +44 (0) 1597 824811 skyemail@skyeinstruments.com www.skyeinstruments.com Iss. 1.2 Skye Instruments Ltd.

More information

Welding & Fabrication Tools FT-Flat Electrode Welding

Welding & Fabrication Tools FT-Flat Electrode Welding Welding & Fabrication Tools FT-Flat Electrode Welding Instruction Manual The Next Step in Belting Welding & Fabrication FT-Flat Electrode Welding Table of Contents Page How to Use this Manual 3 Symbols

More information

GLASS GLOBE MOBILE LINEAR CHANDELIER

GLASS GLOBE MOBILE LINEAR CHANDELIER PARTS ENCLOSED: (B) Top ceiling plate [1] (C) Bottom ceiling plate [1] (D) Glass globes [7] (E) Screws [6] (F) Nuts [5] (G) Extension rods [15] (H) Transformer [1] (I) Threaded inserts [5] (J) 1.6W LED

More information

Signet ph/orp Electrodes General Purpose

Signet ph/orp Electrodes General Purpose Signet 2724-2726 ph/orp Electrodes General Purpose Compatible with ALL Signet ph/orp instruments and SmartPro transmitters Features Patented reference design for exceptional performance and prolonged life

More information

MPS3 with MGMount User manual

MPS3 with MGMount User manual MPS3 with MGMount User manual 1 PRESENTATION The Aimpoint MPS3 (Multi Purpose Sight) is a passive electronic reflex collimator sight designed for use on medium or heavy weapons. Developed primarily for

More information

INSTALLATION INSTRUCTIONS

INSTALLATION INSTRUCTIONS INSTALLATION INSTRUCTIONS Trans4mer Grille Guard/Winch Mount Kit 6367 For Chevrolet Silverado 1500, Suburban 1500 and Tahoe As you read these instructions, you will see NOTES, CAUTIONS and WARNINGS. Each

More information

INSTALLATION INSTRUCTIONS

INSTALLATION INSTRUCTIONS INSTALLATION INSTRUCTIONS Trans4mer Grille Guard/Winch Mount Kit 7646 & 766 For GMC Sierra 500 As you read these instructions, you will see NOTES, CAUTIONS and WARNINGS. Each message has a specific purpose.

More information

Thomas Scientific Swedesboro, NJ U.S.A.

Thomas Scientific Swedesboro, NJ U.S.A. Thomas Scientific Swedesboro, NJ 08085-0099 U.S.A. Wiley Mini Mill 3383-L10 (115 V, 60 HZ) USE AND CARE OF CATALOG NUMBER: 3383-L10 Wiley Mini Mill (115 V, 60 HZ) PRELIMINARY 1. Mill has been properly

More information

Module-2 Lecture-2 Combination set, Vernier calipers. (Refer Slide Time: 00:16)

Module-2 Lecture-2 Combination set, Vernier calipers. (Refer Slide Time: 00:16) Metrology Prof. Dr. Kanakuppi Sadashivappa Department of Industrial and Production Engineering Bapuji Institute of Engineering and Technology-Davangere Module-2 Lecture-2 Combination set, Vernier calipers

More information

Hand Crimping Tools , , and AUG 12 Rev D

Hand Crimping Tools , , and AUG 12 Rev D Instruction Sheet 408-7332 Hand Tools 90123-2, 90123-5, and 90124-2 29 AUG 12 PROPER USE GUIDELINES Cumulative Trauma Disorders can result from the prolonged use of manually powered hand tools. Hand tools

More information

GRILLE GUARD SPRINTER VAN (EXCLUDES X4) INCLUDES MERCEDES, FREIGHTLINER AND DODGE PARTS LIST:

GRILLE GUARD SPRINTER VAN (EXCLUDES X4) INCLUDES MERCEDES, FREIGHTLINER AND DODGE PARTS LIST: PARTS LIST: 1 Grille Guard 8 12mm Hex Nuts 1 Driver/Left Side Frame Mounting Bracket 2 10-1.50mm x 25mm Button Head Bolts 1 Passenger/Right Side Frame Mounting Bracket 4 10mm x 20mm OD x 2mm Flat Washers

More information

Free Tutorial! Tumbling Magic Introduction. Author: Patti Bullard

Free Tutorial! Tumbling Magic Introduction. Author: Patti Bullard Free Tutorial! Tumbling Magic 101 www.wiredupbeads.com Introduction Author: Patti Bullard Tumbling your handmade metal jewelry with a rotary tumbler is a simple process, yet it produces magical results.

More information

Elcometer Muller Laboratory Grinder

Elcometer Muller Laboratory Grinder English Elcometer 2000 Muller Laboratory Grinder Operating Instructions English is a registered trademark of Elcometer Limited. All other trademarks acknowledged. Copyright Elcometer Limited. 2009. All

More information

FIBEROPTIC DISPLACEMENT SENSOR. with Analog Output

FIBEROPTIC DISPLACEMENT SENSOR. with Analog Output USER MANUAL FOR FIBEROPTIC DISPLACEMENT SENSOR with Analog Output TYPE RC REFLECTANCE COMPENSATED PHILTEC www.philtec.com Fiberoptic Sensors for the Measurement of Distance, Displacement and Vibration

More information

IMP-2A ELECTRODE IMPEDANCE TESTER SINGLE CHANNEL

IMP-2A ELECTRODE IMPEDANCE TESTER SINGLE CHANNEL MicroProbes for Life Science IMP-2A ELECTRODE IMPEDANCE TESTER SINGLE CHANNEL User Instructions Version: October, 2015 18247-D Flower Hill Way Gaithersburg, MD 20879 Phone: +1 301 330-9788 Fax: +1 301

More information

Ox-RAC-08 Ribbon Angled Fiber Cleaver User Manual

Ox-RAC-08 Ribbon Angled Fiber Cleaver User Manual Ox-RAC-08 Ribbon Angled Fiber Cleaver User Manual Issue 2.0 Contents Introduction... 2 Contents of Cleaving Kit & Unpacking... 3 Cleaving Problems... 8 Blade damage:... 9 Cleaver Maintenance... 10 Cleaning

More information

INSTALLATION INSTRUCTIONS

INSTALLATION INSTRUCTIONS INSTALLATION INSTRUCTIONS Trans4mer Grille Guard/Winch Mount Kit 6334 For Chevrolet Silverado 500HD & 3500 This WARN Trans4mer system can be customized to give your Chevy Silverado a wide variety of looks,

More information

Components of the Microscope

Components of the Microscope Swift M3 Microscope The Swift M3 is a versatile microscope designed for both microscopic (high magnification, small field of view) and macroscopic (low magnification, large field of view) applications.

More information

B Operating Instructions V1.00/EN/

B Operating Instructions V1.00/EN/ Conductive conductivity/ultra-pure water sensors with a 2-electrode system, types 202922, 202923, 202924, 202925 Glass conductivity sensors, type 202922/30 Diaphragm tubes, type 201083 Compensation thermometer,

More information

Frameless Fixed Panel Slider

Frameless Fixed Panel Slider INSTALLATION INSTRUCTIONS Frameless Fixed Panel Slider QCI-5279 SINGLE ROLLER WITH ANTI-JUMP DOUBLE ROLLERS QCI5279 Rev Page Certified 08/09/6 Tools: To install your New Shower Enclosure, you may need

More information

MICROSCOPY and CELL STRUCTURE

MICROSCOPY and CELL STRUCTURE MICROSCOPY and CELL STRUCTURE Readings: Review pp. 69-71, and Fig. 4.1 on p. 65 in your text (POHS, 5 th ed.). Introduction: Biologists rely on many different types of microscopic techniques to find out

More information

IGPG Car Wash Round Robin Test Procedure

IGPG Car Wash Round Robin Test Procedure 1. Scope The intension of this round robin test is to investigate whether the car wash test described in ISO 20566 and used to validate the abrasion performance of exterior car body parts is suitable as

More information

Just turn the outer filter ring

Just turn the outer filter ring Just turn the outer filter ring 2 For information of other language than English, please visit our web-site of following URL. http://www.kenkoglobal.com/ndx Contents 4 Newly Developed Unique Design 5 Easy

More information

INSTALLATION INSTRUCTIONS

INSTALLATION INSTRUCTIONS INSTALLATION INSTRUCTIONS Trans4mer Grille Guard/Winch Mount Kit 76139 & 7643 For Chevrolet Silverado 1500 As you read these instructions, you will see NOTES, CAUTIONS and WARNINGS. Each message has a

More information

Grand Slam 370 Installation Instructions

Grand Slam 370 Installation Instructions Grand Slam 370 Installation Instructions Patent #7,114,283 TACO Metals Inc., 50 NE 179 Street Miami, Florida 33162 info@tacomarine.com TACOmarine.com Your Kit Contains 2- Grand Slam 370 mounts 4-5/16"

More information

KPS Turbidity Meter User s Manual

KPS Turbidity Meter User s Manual KPS Turbidity Meter User s Manual Küppers Elektromechanik GmbH Quality system certified to DIN ISO 91 Contents Product description... page 3 Technical data... page 3 Dimensions... page 4 Getting started...

More information

SAUTER TC Version /2017 GB

SAUTER TC Version /2017 GB Sauter GmbH Ziegelei 1 D-72336 Balingen E-Mail: info@sauter.eu Tel: +49-[0]7433-9933-199 Fax: +49-[0]7433-9933-149 Internet: www.sauter.eu Instruction Manual Digital Coating Thickness Gauge SAUTER TC Version

More information

+ - 1a. 1b. c. 1d. 1e. Runway Suspended / Surface Cont. Run Installation. 2.0mm DIL DIL Revision - M 23/11/2018 Page 1

+ - 1a. 1b. c. 1d. 1e. Runway Suspended / Surface Cont. Run Installation. 2.0mm DIL DIL Revision - M 23/11/2018 Page 1 230-240V / 50-60Hz IP20 IK00 + - 2.0mm WARNING - THIS LUMINAIRE MUST BE EARTHED This luminaire has been tested and is manufactured to comply with BS EN 60598 : specification for general requirements and

More information

NOTE: Landscape Forms is not responsible for site preparation, footings, or electrical wiring.

NOTE: Landscape Forms is not responsible for site preparation, footings, or electrical wiring. 11/2009 Stop Embedded Mount Bollard HANDLE WITH CARE! Pangard II Polyester Powder coat is a strong, long-lasting finish. Protect this finish from damage during installation. Place unwrapped powder coated

More information

Series APD4059. For Load Cells/Pressure Transducer Transmitters. Field Rangeable, with Calibration Resistor. Standard Features.

Series APD4059. For Load Cells/Pressure Transducer Transmitters. Field Rangeable, with Calibration Resistor. Standard Features. TECHNICAL MANUAL Series APD4059 For Load Cells/Pressure Transducer Transmitters. Field Rangeable, with Calibration Resistor Standard Features Use Internal or External Calibration Resistor Sense Lead Compensation

More information

RING CUTTING SYSTEM OPERATION MANUAL. The lighter your touch, the faster GEM cuts

RING CUTTING SYSTEM OPERATION MANUAL. The lighter your touch, the faster GEM cuts RING CUTTING SYSTEM OPERATION MANUAL The lighter your touch, the faster GEM cuts ❶ SELECT THE CORRECT GEM CUTTING DISC ❶ REVIEW THE GEM QUICK REFERENCE CHART (BELOW). ❷ ASK THE PATIENT if they know ring

More information

CAUTION! This manual contains important information for the correct installation, operation and maintenance of the equipment described herein.

CAUTION! This manual contains important information for the correct installation, operation and maintenance of the equipment described herein. CAUTION! This manual contains important information for the correct installation, operation and maintenance of the equipment described herein. All persons involved in such installation, operation, and

More information

DIODE / TRANSISTOR TESTER KIT

DIODE / TRANSISTOR TESTER KIT DIODE / TRANSISTOR TESTER KIT MODEL DT-100K Assembly and Instruction Manual Elenco Electronics, Inc. Copyright 1988 Elenco Electronics, Inc. Revised 2002 REV-K 753110 DT-100 PARTS LIST If you are a student,

More information

DEUTSCH Instructional Manual for HDP-400 Power Crimper (TE CONNECTIVITY PN )

DEUTSCH Instructional Manual for HDP-400 Power Crimper (TE CONNECTIVITY PN ) ORIGINAL INSTRUCTIONS DEUTSCH Instructional Manual for HDP-400 Power Crimper (TE CONNECTIVITY PN 1606312-1) Instruction Sheet 0425-034-0000 15 JAN 18 Rev H NOTE All numerical values are in metric units

More information

Many types of anodic coatings exist through an electrolytic formation. The important ones you need to know are:

Many types of anodic coatings exist through an electrolytic formation. The important ones you need to know are: ANODIZING MACHINE Anodizing is a process of treating the surface of an aluminum to convert it to aluminum oxide. Anodizing aluminum takes so many processes to arrive at the state of aluminum oxide. The

More information

OWNER S MANUAL. Safety. Please read this owner s manual before use and keep it at hand for reference. Warranty

OWNER S MANUAL. Safety. Please read this owner s manual before use and keep it at hand for reference. Warranty Please read this owner s manual before use and keep it at hand for reference. OWNER S MANUAL Safety Important safety instructions for using the INCRA Miter5000 Before using the INCRA Miter5000, read and

More information

Round Indicator Lights Mounting Hole Page

Round Indicator Lights Mounting Hole Page PRODUCT SERIES Round Indicator Lights Page 8, &. 9. &..7 - NEMA -R.7 8 & 9. 7. 8. 9 7. 8. - NEMA -R.9.9 Indicator Lights.7,.. 7. Rectangular Indicator Lights Panel Cutout. x.. x. 7 & 7. x.8 8. x. 9 & 7.

More information

INSTALLATION INSTRUCTIONS

INSTALLATION INSTRUCTIONS INSTALLATION INSTRUCTIONS Trans4mer Grille Guard/Winch Mount Kit 63470 For Ford 4x4 F150, F250 Light Duty & Expedition As you read these instructions, you will see NOTES, CAUTIONS and WARNINGS. Each message

More information

The old adage seeing is believing is appropriate when

The old adage seeing is believing is appropriate when 26 Quality Digest/October 2001 The old adage seeing is believing is appropriate when referring to optical comparators. Because these measurement tools display a magnified image of a part, a tremendous

More information

3/8 FRAMELESS BYPASS SLIDING SHOWER ENCLOSURE

3/8 FRAMELESS BYPASS SLIDING SHOWER ENCLOSURE INSTALLATION INSTRUCTIONS 3/8 FRAMELESS BYPASS SLIDING SHOWER ENCLOSURE QCI-5017 QCI5017 Rev. 3 Page 1 of 9 Certified 12/6/17 INSTALLATION NOTES: Unpack your unit carefully and inspect for freight damage.

More information

INSTALLATION INSTRUCTIONS

INSTALLATION INSTRUCTIONS TEL:1-866-XANATOS INSTALLATION INSTRUCTIONS FOR 2017-2018 NISSAN TITAN (EXCLUDES TITAN XD MODELS) PART#AB-NI20 PARTS LIST: 1 Bull Bar 2 8mm Nylon Lock Nuts 1 Driver/Left Side Bull Bar Bracket 2 8-1.25mm

More information

Elimination of Elevator Bounce

Elimination of Elevator Bounce For the Agilent Archon Autosampler Rework Instructions CAUTION This kit is intended for use by Agilent Service personnel only. Elevator Removal 1 Open top cover. 2 Open front lower door. 3 Remove vial

More information

2 Channel Light Sensor

2 Channel Light Sensor SKR 1840 Skye Instruments Ltd., 21 Ddole Enterprise Park, Llandrindod Wells, Powys LD1 6DF UK Tel: +44 (0) 1597 824811 skyemail@skyeinstruments.com www.skyeinstruments.com Iss. 1.1 Skye Instruments Ltd.

More information

eye in hell CD 3002 Sensor Manual

eye in hell CD 3002 Sensor Manual Your eye in hell CD 3002 Sensor Manual Siemens Laser Analytics AB Oct 2001 CD 3002 Sensor manual Document number: CID 3002-1102 Rev. 2A LDS 3000 Sensor manual Content Content 1. CD 3002 1 1.1 General 1

More information

INSTALLATION INSTRUCTIONS

INSTALLATION INSTRUCTIONS INSTALLATION INSTRUCTIONS Trans4mer Winch Mount For 2011-2012 Chevrolet Silverado HD and GMC Sierra HD Bracket Kit 90165 Your safety, and the safety of others, is very important. To help you make informed

More information