INTERNATIONAL OIML R 89 RECOMMENDATION

Size: px
Start display at page:

Download "INTERNATIONAL OIML R 89 RECOMMENDATION"

Transcription

1 INTERNATIONAL OIML R 89 RECOMMENDATION Edition 1990 (E) Electroencephalographs - Metrological characteristics - Methods and equipment for verification Electroencéphalographes - Caractéristiques métrologiques - Méthodes et moyens de vérification OIML R 89 Edition 1990 (E) ORGANISATION INTERNATIONALE DE MÉTROLOGIE LÉGALE INTERNATIONAL ORGANIZATION OF LEGAL METROLOGY

2 Foreword The International Organization of Legal Metrology (OIML) is a worldwide, intergovernmental organization whose primary aim is to harmonize the regulations and metrological controls applied by the national metrological services, or related organizations, of its Member States. The two main categories of OIML publications are: International Recommendations (OIML R), which are model regulations that establish the metrological characteristics required of certain measuring instruments and which specify methods and equipment for checking their conformity; the OIML Member States shall implement these Recommendations to the greatest possible extent; International Documents (OIML D), which are informative in nature and intended to improve the work of the metrological services. OIML Draft Recommendations and Documents are developed by technical committees or subcommittees which are formed by the Member States. Certain international and regional institutions also participate on a consultation basis. Cooperative agreements are established between OIML and certain institutions, such as ISO and IEC, with the objective of avoiding contradictory requirements; consequently, manufacturers and users of measuring instruments, test laboratories, etc. may apply simultaneously OIML publications and those of other institutions. International Recommendations and International Documents are published in French (F) and English (E) and are subject to periodic revision. This publication reference OIML R 89 (E), edition 1990 which is under the responsibility of TC 18/SC 4 Bio-electrical instruments, was sanctionned by the International Conference of Legal Metrology in OIML publications may be obtained from the Organization s headquarters: Bureau International de Métrologie Légale 11, rue Turgot Paris - France Telephone: 33 (0) and Fax: 33 (0) biml@oiml.org Internet: 2

3 ELECTROENCEPHALOGRAPHS METROLOGICAL CHARACTERISTICS METHODS and EQUIPMENT for VERIFICATION 1. Scope 1.1. This Recommendation concerns analogue electroencephalographs, which are defined as diagnostic medical instruments having at least eight channels. Such instruments are intended to accurately measure and record bioelectrical potentials and their time sequences as detected by electrodes attached to the surface of the human scalp This Recommendation does not apply to information-processing electroencephalographs, nor instruments that digitize the signals, nor other special-purpose instruments This Recommendation is intended for the use of metrological services and other responsible governmental agencies. It specifies the metrological characteristics to be determined, methods of and equipment for testing, and initial and subsequent verifications of electroencephalographs that are regulated by laws. Metrological characteristics include errors in measuring voltage signals and time intervals as well as a number of other characteristics that affect measurement accuracy. Procedures are given for determining the relative error of measurement of sixteen instrument characteristics, and formulae are provided for calculating the associated permitted limits of these characteristics. These testing and control procedures provide a basis for ensuring that electroencephalographs of adequate accuracy are available for clinical measurements. The characteristics of the instrument covered are those that have been deemed to be important in order to achieve the necessary measurement accuracy in clinical practice; therefore, any changes in the requirements prescribed should be based on clinical experience and the needs of electroencephalography This Recommendation does not address pattern approval, terminology, technical requirements, electrical safety and the associated test methods. Some of these matters are being considered by the International Electrotechnical Commission (IEC) in the (draft) Publication xx "Medical electrical equipment Part 2: Particular requirements for the safety of electroencephalographs" under the auspices of SC 62D on "Electromedical equipment". A subject that is also under consideration by SC 62D is " Particular requirements for the performance of electroencephalographs". The requirements of this Recommendation are also compatible with "Recommended specifications for electroencephalographs" developed by the International Federation of Societies for Electroencephalography and Clinical Neurophysiology (IFSECN). Those specifications can be found in "Recommendations for the Practice of Clinical Neurophysiology", IFSECN, Elsevier, Amsterdam- New York-Oxford (1983). 3

4 2. Characteristics to be verified The characteristics shown in Table 1 should be measured during verification of electroencephalographs. Table 1 Metrological characteristics to be verified. Metrological characteristics Point Verified during Initial Subsequent verification Relative voltage-measurement error Relative sensitivity-setting error Relative time-interval measurement error Relative error of the recording speed Recording hysterisis Relative errors of internal calibrator and time marker Overshoot Time constant Amplitude-frequency response curve Input impedance Common-mode rejection ratio Baseline width Baseline drift Intrinsic noise level referred to the input Interchannel crosstalk coefficient Relative error of the interelectrode-resistance measurement Verification of current in the patient circuit is also required at initial verification (see point 5.4). 4

5 3. Measuring instruments used for verification The measuring instruments listed in Table 2 are recommended for the verification of electroencephalographs. Table 2 Measuring instruments used for verification Measuring instrument Symbol Main characteristics Sine-wave signal generator (*) G1 Frequency range 0.01 Hz 100 Hz Max.frequency error ± 1 % Voltage range 10 mv 10 V RMS Max. voltage error ± 1 % Double-ended output Square-wave signal generator (*) G2 Frequency range 0.01 Hz Hz Max. frequency error ± 1 % Voltage range 1 mv 10 V Max. voltage error ± 0.5 % Double-ended output Voltage divider (**) D1 Division factor (R1 = 100 kω; R2 = 10 Ω; max.error ± 1 %) Max.error of division ± 2 % Simulated skin-electrode Z1 R3 = 4.7 kω ± 1 % impedance (***) DC voltage source U1 Voltage 1.5 V ± 5 % Length-measuring device Measurement length 1 mm 100 mm Max.error ± 0.1 mm on measuring length from 0 mm to 10 mm, and ± 1 % on measuring length from 10 mm to 100 mm. Magnifying glass Magnification 5 Resistors R4-R12 R4 = 200 Ω ; R5 = 50 Ω ; R6 = 100 kω ; R7 = 620 kω, R8 = 5 kω, R9 = 1 kω ; R10 = 5 kω; R11 = 20 kω ; R12 = 50 kω Max. error ± 1 % Capacitors C1, C2 C1 = 0.5 µf; C2 = 4.7 nf Max.error ± 5 % Impedance Z2 R7 in parallel with C2 (*) If the voltage or frequency of generator G1 or G2 do not meet the specifications of Table 2, a voltmeter and a frequency meter with the accuracy specified for the generator output voltage and frequency should be provided. (**) (***) See Figure 1. Impedance Z1 (called in the text "patient skin-electrode impedance" is intended to simulate the electrical characteristics of the electrode-skin interface. 5

6 4. Verification conditions and preparation for verification 4.1. The recommended conditions for verification are as follows: ambient temperature: 15 C to 25 C, atmospheric pressure: 96 kpa to 104 kpa, relative humidity of air: 50 % to 80 % (no condensation), mains voltage fluctuations: ± 2 % of nominal voltage, mains frequency: (60 Hz or 50 Hz) ± 2 %. The ranges of atmospheric conditions may be extended for regions with extreme temperatures (or other climatic conditions) or high altitude Deviations of the power-supply voltage should not exceed the values indicated in the manufacturer's manual Before the electroencephalographs are verified, they and the measuring instruments required for verification should be assembled according to the requirements of the national metrology service or other officially authorized organizations. Measuring instruments used during verification should be checked by authorized personnel. 5. Verification 5.1. External examination External examination includes a check that the manufacturer's manual, which should provide the necessary information on the electroencephalograph, is available and includes the following: values of commonly accepted characteristics, their tolerance limits and procedures for their determination, diagrams and construction details necessary for carrying out verification procedures, operating and maintenance instructions, instructions for special medical applications. External examination includes checks for: absence of corrosion and mechanical damage, and freedom from any trace of deterioration of the lead cables Testing The electroencephalograph shall be tested after the warm-up period specified by the manufacturer. The testing shall include checks for the presence and deflection of the recording trace, the pen pressure, either the ink feed or the adjustment of the heating system for thermal pens, the movement of the record at different speeds and the presence of calibration signals and sensitivity controls (step and smooth) Determination of metrological characteristics To determine the metrological characteristics each measurement shall be repeated at least three times and each of the measured values shall lie within specified limits. Unless otherwise stated, the electrode selector shall be set so as to apply an input test signal to each channel simultaneously. The recorded output shall be measured in a way that excludes the trace line width. 6

7 The limiting values of error include both the errors of the EUT (*) and of the reference instruments. If the settings for filters or sensitivities indicated by the manufacturer differ from those specified in this Recommendation the procedure shall be modified to take these settings into account (see footnote, Table 3). The associated diagrams in this Recommendation are given for a 16-channel instrument as an example. The diagrams for other instruments having at least 8 channels are similar Determination of relative voltage-measurement error Definition: The relative voltage-measurement error is the difference between the voltage recorded by an electroencephalograph and the voltage applied to its input (whose reference value is taken as a conventional true value), divided by the input voltage. Method of measurement: The relative voltage-measurement error shall be determined directly by measuring the amplitude of the square-wave signal recorded, referring it to the input using the sensitivitysetting, and comparing it with the amplitude of the input voltage as determined by the reference voltmeter (conventional true value). Measurement circuit: The measurement set-up is shown in Figure 1. The appendix presents the table of correspondence between the system of numerical definitions of electrodes and the International "10-20" System. Measurement procedure: The switches of the low-frequency (time constant) and high-frequency filters are set to cover the necessary EEG band width (e.g. 1 s and 70 Hz respectively). The electrode selector (s) is (are) set as shown in Figure 1. The recording speed is set at 30 mm/s. Switch S1 is set at position 2 (Z1 is switched on) and switch S2 at position 3. Generator G2 is set at a frequency of 10 Hz. The sensitivity of all channels is set as specified in Table 3. A square-wave signal from generator G2 and voltage divider D1 with a peak-to-peak amplitude shown in Table 3 is applied to the electroencephalograph's input through the simulated skin-electrode impedance Z1. For each voltage of the input signal and corresponding sensitivity the dimension of the recorded signal is measured. The measurement is repeated in the presence of a constant voltage that simulates the maximum value of electrode potential polarization equal to ± 300 mv, switch S2 being set at positions 1 and 2, in turn. Table 3 Sensitivity-setting and input signal voltage for determining the relative voltage measurement error (**) Input voltage, µv (peak-to-peak) Sensitivity-setting, µv/mm (*) (**) EUT: Equipment under test. If a listed sensitivity is not available on the instrument, the next lower available value may be used with the input voltage scaled down proportionately. See also note (*) page 9. 7

8 8

9 Calculation: The relative voltage measurement error, in percent, shall be calculated by the following formula: U m Uin δu = 100 (1) Uin where: U m = h m S n * is the recorded voltage peak-to-peak amplitude, in µv, h m is the peak-to-peak amplitude of the recorded output signal, in mm, S n * is the nominal value of the sensitivity setting, in µv/mm, (*) U in is the peak-to-peak amplitude of the input voltage, in µv. Requirement: For the signals recorded with and without a constant D.C. polarization voltage of ± 300 mv, the error given by formula (1) shall not exceed the value: 10 (1 + U 1 / U in ) (2) where: U 1 is the lowest value of the voltage measurement range, e.g. 5 µv. The commonly accepted polarity shall be checked by applying a negative input signal to input 1 (G1) which shall cause the pens to deflect upwards Determination of relative sensitivity-setting error Definition: The relative sensitivity-setting error is the difference between the nominal and measured values of sensitivity, divided by the nominal value. Method of measurement: The relative sensitivity-setting error shall be determined directly by measuring the peak-to-peak amplitude of the recorded sine-wave signal and of the stimulus, calculating the value of sensitivity and comparing it with the nominal value. Measurement circuit: The measurement set-up is shown in Figure 2. Measurement procedure: The filter switches are set as specified in point The sensitivity is set at 1 µv/mm and the recording speed at 30 mm/s. A sine-wave signal from generator G1 and voltage divider D1 with a peak-to-peak amplitude of 20 µv and a frequency of 10 Hz is applied to the electroencephalograph's input. The amplitude of the recorded signal is measured. The measurements are repeated for sensitivity equal to 5, 10, 20, 50 and 100 µv/mm and for input signals with peak-to-peak amplitude of 100, 200, 400, and µv respectively. The footnote of Table 3 applies. Calculation: The relative sensitivity-setting error, in percent, shall be calculated by the following formula: * * Sn Sm δ s = 100 (3) * S where: S m * = U in / h m is the measured sensitivity value, in µv/mm, h m is the recorded signal peak-to-peak amplitude, in mm, U m is the input voltage peak-to-peak amplitude, in uv, S n * is the nominal sensitivity value, in µv/mm. n (*) In this Recommendation, the term "sensitivity" is defined by the ratio of the input to the output amplitude, is expressed in µv/mm and is represented by the symbol S *. This definition of sensitivity, commonly used in the field of electroencephalography, is opposite to the definition of sensitivity S in all the other fields (S * = 1/S). 9

10 10

11 Requirement: The relative sensitivity-setting error as determined by formula (3) shall not exceed ± 5 % Determination of relative time-interval measurement error Definition: The relative time-interval measurement error is the difference between the measured and conventional true values of the input signal period, divided by the latter. Method of measurement: The relative time-interval measurement error shall be determined directly by measuring the period (linear length) of the recorded square-wave signal, dividing it by the nominal recording speed and comparing it with the reciprocal of the input frequency. Measurement circuit: The measurement set-up is shown in Figure 1. Measurement procedure: The switches S1 and S2 are set at positions 2 and 3 respectively. The filter switches are set as specified in point The sensitivity is set at 10 µv/mm. A square-wave signal from generator G2 and voltage divider D1 with a peak-to-peak amplitude of 200 µv is applied to the electroencephalograph's input. The values of frequency of generator G2 and the recording speed values are taken as specified in Table 4, for different intervals of measurement. In each case no fewer than 3 cycles of the input signal are recorded. It is sufficient to measure top and bottom channels only. Table 4 Generator G2 frequencies to be adjusted and time intervals to be measured Time intervals to be measured, s Generator G2 frequency, Hz Recording speed, mm/s Calculation: The relative time-interval measurement error, in percent, shall be calculated by the following formula: Tm Tin δt = 100 (4) Tin where: T m = L m /V n is the recorded time interval, in s, L m is the length of 3 cycles, in mm, V n is the nominal recording speed, in mm/s, T in is the time interval corresponding to 3 cycles of the input signal, in s. Requirement: The relative time-interval measurement error determined by formula (4) shall not exceed the value: 5 (1 + T 1 /T in ) (5) where: T 1 is the lower limit of the time-interval measurement range, e.g s. 11

12 Determination of the relative error of the recording speed Definition: The relative error of the recording speed is the difference between the measured value of the recording speed and the nominal value, divided by the nominal value. Method of measurement: The relative error of the recording speed shall be determined directly by measuring the recorded sine-wave signal period, calculating the value of the recording speed from the generator frequency, and comparing it with the nominal value. Measurement circuit: The measurement set-up is shown in Figure 2. Measurement procedure: The filter switches are set as specified in point The sensitivity is set at 10 µv/mm, and the recording speed is set at the value to be determined. A sine-wave (or square wave) signal from generator G1 (or G2) and voltage divider D1 with a peak-to-peak amplitude of 180 µv is applied to the electro-encephalograph's input. Measurements are made at speeds of 15, 30 and 60 mm/s. The generator frequency is chosen so that the length of the period of the recorded signal is at least 2 mm. At least 10 cycles are recorded at each speed. Calculation: The relative error of the recording speed, in percent, shall be calculated by the following formula: where: V m = L m /nt e is the measured value of the recording speed, in mm/s, L m is the length of the section recorded for n cycles (n 10), in mm, T e is the period of the input signal, in s, V n is the nominal value of the recording speed, in mm/s. δ v Vm Vn = 100 (6) V Requirement: The relative error of the recording speed as determined by formula (6) shall not exceed ± 5 % Determination of recording hysteresis Definition: The recording hysteresis is the distance between the base lines of a trace obtained, respectively, after a positive and a negative input signal is returned to zero (see Figure 3). n Figure 3 Determination of recording hysteresis 12

13 Method of measurement: The recording hysteresis shall be determined directly by measuring the distance between the baselines, respectively after a positive and a negative input signal is returned to zero. Measurement circuit: The measurement set-up is shown in Figure 4. Measurement procedure: For determination of recording hysteresis a differentiating circuit with the time constant equal to 50 ms (e.g. R6 = 100 kω, C1 = 0.5 µf), is connected between voltage divider D1 and the input. The filter switches are set as specified in point The sensitivity is set at 50 µv/mm and the recording speed is set 30 mm/s. A differentiated signal of 500 µv and 1 Hz is applied to the electroencephalograph's input, and the recording hysteresis is measured. Requirement: The recording hysteresis hi shall not exceed 0.5 mm Determination of relative errors of internal calibrator and time marker Definition: The relative error of the internal calibrator or of the time marker is the difference between the measured and nominal values of voltage output of the internal calibrator or of the time interval of the time marker, divided by their respective nominal values. Method of measurement: The relative errors of the internal calibrator and of the time marker shall be determined by comparing the nominal values of the voltage and time interval of the internal signals of the instrument with the voltage and time interval of a signal applied to the input, the recorded value of the signal being made equal to the recorded internal value. Measurement circuit: The measurement set-up is shown in Figure 1. Measurement procedure: Switches S1 and S2 are set at positions 1 and 3 respectively and the internal calibrator is selected. The filter switches are set as specified in point with the recording speed of 15 mm/s. Signals are recorded from the internal calibrator and the time marker. Then the electrodes are selected. A square-wave signal from generator G2 and voltage divider D1 with a frequency of 1 Hz and an amplitude equal in value to the internal calibrator voltage, is applied to the input of the device to be verified. Generator G2 voltage and frequency is adjusted so that the dimensions of the recorded signals induced respectively by the internal and external stimuli coincide. The tests are performed at the values of the internal-calibrator voltages and the sensitivities given in Table 5. Table 5 The values of sensitivity to be selected and internal-calibrator voltages to be measured Internal calibrator voltage µv Sensitivity µv/m Calculation: The relative internal-calibrator error, in percent, shall be calculated by the following formula: δ c U cm U cn = 100 (7) U cn 13

14 14

15 where: U cm is the measured peak-to-peak amplitude of the internal-calibrator voltage, in µv, U cn is the nominal value of peak-to-peak amplitude of the internal-calibrator voltage, in µv. The relative time-marker error, in percent, shall be calculated by the following formula: where: T m is the recorded value of the time-marker interval, in s, T n is the nominal value of the time-marker interval, in s. δ tm Tm Tn = 100 (8) T Requirement: Neither the relative internal-calibrator error nor the relative time-marker error as determined respectively by formulae (7) and (8) shall exceed ± 2 %. If there is no time marker on the EUT, the EUT shall be considered to have met the relative time-marker error Determination of overshoot Definition: The overshoot is the difference between the maximum peak-to-peak amplitude of a recorded square-wave signal and its minimum value, divided by twice the minimum value (see Figure 5). n Figure 5 Determination of overshoot. Method of measurement: The overshoot shall be determined directly by measuring the peak-to-peak amplitude of the recorded square-wave signal. Measurement circuit: The measurement set-up is shown in Figure 1. Measurement procedure: Switches S1 and S2 are set at positions 1 and 3 respectively. The filter switches are set as specified in point The sensitivity is set at 10 µv/mm and the recording speed is 30 mm/s. A square-wave signal from generator G2 and voltage divider D1 with a peak-to-peak amplitude of 180 µv and a frequency of 10 Hz is applied to the electroencephalograph's input. At least 3 cycles are recorded, and the maximum and minimum peak-to-peak amplitude of each cycle is measured. Calculation: The overshoot, in percent, shall be calculated by the following formula: δ o h max h min = 100 (9) 2h where: h max and h min are the measured values of the linear dimensions of the maximum and minimum peak-to-peak amplitudes respectively of each period recorded, in mm. Requirement: The overshoot as determined by formula (9) shall not exceed 10 %. min 15

16 Determination of time constant Definition: The electroencephalograph time constant is defined as the time interval required for a recorded square-wave signal amplitude to decay to 1/e (37 %) of an initial value (see Figure 6). Figure 6 Determination of time constant. Method of measurement: The time constant shall be determined directly by measuring the linear dimensions of the decay of the recorded square-wave signal following the overshoot. Measurement circuit: The measurement set-up is shown in Figure 1. Measurement procedure: Switches S1 and S3 are set at positions 1 and 3 respectively. The highfrequency filter switch is set as specified in point and the time-constant and low-frequency switches are set at a value to be determined. The sensitivity is set at 20 µv/mm and the recording speed is set at 30 mm/s. A square-wave signal from generator G2 and voltage divider D1 with a peak-to-peak amplitude of 180 µ.v and a frequency of 0.1 Hz is applied to the electroencephalograph's input. The length of the segment recorded between the moment of maximum overshoot and the moment when the amplitude of the recorded signal decreases to 37 % of the initial value is measured. The measurements are repeated for all positions of the time-constant switch. Calculation: The time constant, in s, for each of the recorded signals shall be calculated by the following formula: T c = L m /V n (10) where: L m is the length of the segment in which the signal amplitude decreases to 37 % of the initial value, in mm, V n is the nominal value of the recording speed, in mm/s. Requirement: The time constant as determined by formula (10) shall be within ± 10 % of the values specified by the manufacturer Determination of amplitude-frequency response curve Definition: The amplitude-frequency response curve is the variation with frequency of the recorded signal amplitude, the input signal amplitude being constant. Method of measurement: The amplitude-frequency response curve shall be determined by comparing the peak-to-peak amplitude of the sine-wave signal recorded at different frequencies with the peak-to-peak amplitude of the signal recorded at 10 Hz. Measurement circuit: The measurement set-up is shown in Figure 2. 16

17 17

18 18

19 Measurement of the pass band: The filter switches are set as specified in point and the recording speed is set at 15 mm/s for frequencies less than 10 Hz and 30 mm/s for all other frequencies. A sine-wave signal from generator G1 and voltage divider D1 with peak-to-peak amplitude of 180 µv (held constant) is applied to the input, successively at frequencies 0.16, 0.5, 1.5, 5, 10, 15, 30, 60, 75 Hz. In each case at least 5 cycles are recorded and the peak-to-peak amplitudes of the recorded signals are measured. Requirement: The peak-to-peak amplitude of the signal recorded at frequencies between 1 and 60 Hz relative to the peak-to-peak amplitude of a signal recorded at 10 Hz shall be between 90 % and 110 %. Measurement of the cut-off frequencies of the filters: The procedures are the same as those specified above except that the filter switches are set at values to be determined. The generator G1 frequency is set, in turn, at 0.9 F c and 1.1 F c, where F c is the cutt-off frequency value to be determined. The linear dimension of the peak-to-peak amplitude of the recorded signal is measured. The measurements are made for all values of cut-off frequencies of the high-pass and low-pass filters. The entire amplitude-frequency characteristic for all filters should be provided in the manufacturer's manual. Requirement: The linear dimension of the recorded signals shall meet the following conditions: A0.9F c 0.7 A10 A1.1F c A0.9F c 0.7 A10 A1.1F c where: A0.9F c A10 Al.lF c for low frequencies for high frequencies is the linear dimension of the signal peak-to-peak amplitude recorded at a frequency of 0.9 F c, in mm, is the linear dimension of the signal peak-to-peak amplitude recorded at a frequency of 10 Hz, in mm, is the linear dimension of the signal peak-to-peak amplitude recorded at a frequency of 1.1 F c, in mm Determination of input impedance Definition: The input impedance is an impedance measured between any two inputs of the device. Method of measurement: The input impedance shall be determined by comparing the linear dimensions of peak-to-peak amplitudes of the sine-wave signals recorded with and without supplementary impedance connected in series. Measurement circuit: The measurement set-up is shown in Figure 7. Measurement procedure: The filter switches are set as specified in point The sensitivity is set at 10 µv/mm and the recording speed is set at 30 mm/s. Switches S1 and S2 are set at 1 and 3 respectively. A sine-wave signal from generator G1 and voltage divider D1 with a peak-to-peak amplitude of 180 µv and a frequency of 10 Hz is applied to the input with the recorded length of 25 mm. The recorded signal peakto-peak amplitude is measured. Then the measurement is repeated with S1 at position 2. The measurements are repeated in presence of a DC voltage of ± 300 mv, with S2 being set at 1 and 2, in turn. Calculation: The input impedance, in MΩ, shall be calculated by the following formula: Z H 2 in = Z (11) 2 H1 H 2 19

20 where: H 1 is the linear dimension of the signal recorded with S1 at position 1, in mm, H 2 is the linear dimension of the signal recorded with S1 at position 2, in mm, Z 2 is the value of the impedance Z2 connected in series, in MΩ. Requirement: The input impedance as determined by formula (11) shall not be less than 1 MΩ when the same pair of input terminals is selected by up to 8 channels Determination of common-mode rejection ratio Definition: The common-mode rejection ratio is the ratio of the peak-to-peak amplitude of an in-phase signal applied at the input of the EUT to the peak-to-peak amplitude of an out-of-phase signal that results in the same peak-to-peak amplitude of the recorded signal. Method of measurement: The common-mode rejection ratio shall be determined indirectly by measuring the linear dimension of the peak-to-peak amplitude of the signal recorded by the EUT, when a sine-wave signal with the frequency 50 Hz or 60 Hz of a given amplitude is applied in common mode (between input and ground or earth). Measurement circuit: The measurement set-up is shown in Figure 8. Measurement procedure: The filter switches are set as specified in point and switch S is set at 1. The sensitivity is set at 5 µv/mm and the recording speed is set at 15 mm/s. The peak-to-peak amplitude of the voltage of generator G1 is set at 400 mv with a frequency of 50 Hz or 60 Hz. The peak-to-peak amplitude of the recorded signal is measured. Then switch S is set to position 2 (unbalanced) and the measurements are repeated with a sensitivity of 100 µv/mm. Calculation: The balanced and unbalanced common-mode rejection ratio shall be calculated by the following formula: K U 3 = 10 (12) * h Sn where: h is the linear dimension of the recorded signal peak-to-peak amplitude, in mm, S n * is the nominal value of the sensitivity setting, in µv/mm, U is the peak-to-peak amplitude of the input voltage, in mv (in this case equal to 400 mv). Requirement: The common-mode rejection ratio as determined by the formula (12) shall not be less than 10 4 for the balanced condition and 200 for the unbalanced condition for each channel Determination of baseline width Definition: The base line width is the width of the line drawn on the paper, the input of the EUT being grounded. Method of measurement: The baseline width shall be determined directly by measuring it perpendicular to the trace (Figure 9). Figure 9 Determination of baseline width. 20

21 Measurement procedure: All input terminals are connected to ground (earth) and the filters are set to provide the narrowest band width. The sensitivity is set at 100 µv/mm and the recording speed is 15 mm/s. The width of the recorded line is measured. Requirement: The baseline width shall not exceed 0.5 mm Determination of baseline drift Definition: The baseline drift is the deviation of the baseline during a given time interval, the input of the EUT being grounded (earthed) (see Figure 10). Figure 10 Determination of the baseline drift. Method of measurement: The baseline drift shall be determined directly by measuring the baseline deviation during the time interval of 10 min. Measurement procedure: The sensitivity and recording speed are set as in point and the filter switches are set as in point The deviation of the baseline in 10 min is measured. Requirement: The baseline drift in 10 min shall not exceed 1.0 mm Determination of intrinsic noise level referred to the input Definition: The intrinsic noise level referred to the input is the maximum peak-to-peak amplitude of the signal recorded when the electroencephalograph's input is connected to the skin-electrode impedance Z1; it is evaluated for a given time interval and referred to the electroencephalograph's input (see Figure 11). Method of measurement: The intrinsic noise level referred to the input shall be determined directly by measuring the maximum peak-to-peak amplitude of the recorded signal for a time interval of 60 s and referred to the input through the sensitivity-setting. Measurement circuit: The measurement set-up is shown in Figure 12. Measurement procedure: The inputs are connected to simulated skin-electrode impedances Z1. The filter switches are set as specified in point The sensitivity is set at 1 µv/mm and the recording speed is set at 15 mm/s. The recording lasts for 60 s. The linear dimensions of the maximum peak-to-peak amplitude of the recorded signal are measured for the entire record and also for 3 (15 mm long) segments i.e. for a time interval of 1 s. In addition the deviation of the medium value of the baseline (noise component less than 0.5 Hz) on the segment of 90 mm (i.e. for 6 s) is measured. 21

22 Note: In order to avoid external noise pickup, it is recommended that the resistors and switches of the measurement set-up (shown in Figure 12) should be placed in a shielded metal box with its case connected to ground (earth) (pin No. 24 of the terminal box). The resistors (or box containing the resistors and switches) should be connected to the terminal box by unshielded, untwisted leads that are as short as possible but no longer than 1.5 m. Calculation: The noise level referred to the input, in µv, shall be calculated by the following formula: * U N = h n S n (13) where: h n is the linear dimension of the maximum peak-to-peak amplitude of the noise measured on a recording, in mm, S n * is the nominal value of the sensitivity-setting, in µv/mm. Requirement: The intrinsic noise level referred to the input shall satisfy the following conditions: no more than 1 deflection greater than 4 µv in 60 s, no more than 1 deflection greater than 2 µv in 1 s, and the remaining deflections shall not exceed 1.5 µv in 60 s. Figure 11 Determination of the intrinsic noise level referred to the input Determination of interchannel crosstalk coefficient Definition: The interchannel crosstalk coefficient is the ratio of the peak-to-peak amplitude of the voltage induced on the channel under test to the peak-to-peak amplitude of the voltage applied at all the other channels. Method of measurement: The interchannel crosstalk coefficient shall be determined by comparing the peak-to-peak amplitude of signals recorded on the channel under test to the peak-to-peak amplitude of signals recorded on all other channels. Measurement circuit: The measurement set-up is shown in Figure 13. Measurement procedure: The filter switches are set as specified in point The channel under test is connected through simulated skin-electrode impedance Z1 to ground and the sensitivity of this channel is set at 1µV/mm. The sensitivity of other channels is set at 200 µv/mm. A sine-wave signal from generator G1 and 22

23 23

24 24

25 voltage divider D1 with a peak-to-peak amplitude of 1 mv and frequencies manually swept from 10 Hz to 40 Hz are applied to these channels. Linear dimensions of the signal peak-to-peak amplitude recorded at the channel under test are measured. The measurements are repeated successively for each channel. Calculation: The interchannel crosstalk coefficient, in percent, shall be calculated by the following formula: where: i = 1,2,..n is the number of the channel under test, h i Si Wi = 100 (14) U n is the number of channels, h i is the peak-to-peak amplitude of the signal recorded on the channel influenced by the noise, in mm, U in is the peak-to-peak amplitude of the voltage applied to the inputs of the remaining channels, in µv, S i * is the nominal value of the sensitivity of the channel under test, in µv/mm. in Requirement: The interchannel crosstalk coefficient at frequencies from 10 Hz to 40 Hz shall not exceed 2 % Determination of relative error of the interelectrode-resistance measurement Definition: The relative error of the interelectrode-resistance measurement is the difference between the indicated resistance value and the conventional true value, divided by the latter. Method of measurement: The relative error of the interelectrode-resistance measurement shall be determined by comparing the data of the incorporated resistance measuring device with those of the resistor connected to the input of the EUT. Measurement procedure: The device under test is set in the mode of measurement of the interelectrode resistance. Resistances of 1, 5, 20 and 50 kω (R9-R12 in Table 2) are applied to the input of the device. Calculation: The relative error of the interelectrode resistance measurement, in percent, shall be calculated by the following formula: Rm R o δ R = 100 (15) R o where: R o is the resistance connected to the input of the device to be verified, in kω, R m is the indicated resistance value, in kω. Requirement: The relative error of the interelectrode-resistance measurement as determined by formula (15) shall not exceed ± 10 % Determination of current in the patient circuit Definition: The current in the patient circuit is defined as the current passing through a patient while connected to an electroencephalograph. Method of measurement: The current in the patient circuit shall be determined indirectly by measuring the voltage recorded by the electroencephalograph with a resistance connected to the input of the EUT and calculating the value of the current. Measurement circuit: The measurement set-up is shown in Figure 14. * 25

26 26

27 where: Measurement procedure: The sensitivity is set at 100 µv/mm and the recording speed is set at 15 mm/s. The filter switches are set as specified in point The base line is recorded, then the linear dimension of the step signal recorded peak-to-peak is measured with S open. Calculation: The current in the patient circuit, in µa, shall be calculated by the following formula: I h r * 3 = (16) S n 10 h is the linear dimension of the step voltage recorded, in mm, S n * is the sensitivity, in µv/mm, r is the value of the resistance R8 connected to the input, in kω. Requirement: The current in the patient circuit as determined by formula (16) shall not exceed 0.1 µa. 6. Verification report and certificate 6.1. The verification report shall contain the following information: (a) the designation and serial number of the instrument verified, (b) the country of origin and manufacturer, (c) the equipment used for verification, (d) the date of verification, (e) the references to the applicable regulations and the procedures used, (f) the measurement results for the various characteristics, (g) the organizations and persons responsible for verification Electroencephalographs verified in accordance with this Recommendation may be granted a certificate of verification in accordance with the established laws and regulations of a given country. The certification shall specify the expiry date of the validity of the verification. 27

28 APPENDIX Table of correspondence between the system of numerical definitions of electrodes and the international System Fp 1 Fp 2 F 3 F 4 C 3 C 4 P 3 P 4 O 1 O 2 F 7 F 8 T 3 T T 5 T 6 A 1 A 2 F Z C z P z N Figure 15 Electrode lead terminal box. 28

29 Contents Foreword Scope Characteristics to be verified Measuring instruments used for verification Verification conditions and preparation for verification Verification External examination Testing Determination of metrological characteristics Determination of relative voltage-measurement error Determination of relative sensitivity-setting error Determination of relative time-interval measurement error Determination of the relative error of the recording speed Determination of recording hysteresis Determination of relative errors of internal calivrator and time marker Determination of overshoot Determination of time constant Determination of amplitude-frequency response curve Determination of input impedance Determination of common-mode rejection ratio Determination of baseline width Determination of baseline drift Determination of intrinsic noise level referred to the input Determination of interchannel crosstalk coefficient Determination of relative error of the interelectrode-resistance measurement Determination of current in the patient circuit Verification report and certificate Appendix: Table of correspondence between the system of numerical definitions of electrodes and the international System

INTERNATIONAL OIML R 103 RECOMMENDATION

INTERNATIONAL OIML R 103 RECOMMENDATION INTERNATIONAL OIML R 103 RECOMMENDATION Edition 1992 (E) Measuring instrumentation for human response to vibration (with reference to International Standards 8041 and 5347 of the International Organization

More information

Organisation Internationale de Métrologie Légale

Organisation Internationale de Métrologie Légale Organisation Internationale de Métrologie Légale INTERNATIONAL RECOMMENDATION Sound level meters Sonomètres OIML R 58 Edition 1998 (E) CONTENTS Foreword... 3 1 Scope... 4 2 Construction and maximum permissible

More information

OIML R 130 RECOMMENDATION. Edition 2001 (E) ORGANISATION INTERNATIONALE INTERNATIONAL ORGANIZATION. Octave-band and one-third-octave-band filters

OIML R 130 RECOMMENDATION. Edition 2001 (E) ORGANISATION INTERNATIONALE INTERNATIONAL ORGANIZATION. Octave-band and one-third-octave-band filters INTERNATIONAL RECOMMENDATION OIML R 130 Edition 2001 (E) Octave-band and one-third-octave-band filters Filtres à bande d octave et de tiers d octave OIML R 130 Edition 2001 (E) ORGANISATION INTERNATIONALE

More information

Organisation Internationale de Métrologie Légale

Organisation Internationale de Métrologie Légale Organisation Internationale de Métrologie Légale INTERNATIONAL RECOMMENDATION Integrating-averaging sound level meters Sonomètres intégrateurs-moyenneurs OIML R 88 Edition 1998 (E) CONTENTS Foreword...

More information

Westergren tubes for measurement of erythrocyte sedimentation rate

Westergren tubes for measurement of erythrocyte sedimentation rate INTERNATIONAL OIML R 78 RECOMMENDATION Edition 1989 (E) Westergren tubes for measurement of erythrocyte sedimentation rate Pipettes Westergren pour la mesure de la vitesse de sédimentation des hématies

More information

Metal stamps for verification officers

Metal stamps for verification officers INTERNATIONAL OIML R 42 RECOMMENDATION Edition 1981 (E) Metal stamps for verification officers Poinçons de métal pour agents de vérification OIML R 42 Edition 1981 (E) ORGANISATION INTERNATIONALE DE MÉTROLOGIE

More information

OIML E 5 EXPERT REPORT. 6th Edition 2015 (E) ORGANISATION INTERNATIONALE INTERNATIONAL ORGANIZATION

OIML E 5 EXPERT REPORT. 6th Edition 2015 (E) ORGANISATION INTERNATIONALE INTERNATIONAL ORGANIZATION EXPERT REPORT OIML E 5 6th Edition 2015 (E) Overview of the present status of the standards referred to in OIML D 11:2013 General requirements for measuring instruments Environmental conditions Revue du

More information

Material measures of length for general use

Material measures of length for general use INTERNATIONAL OIML R 35 RECOMMENDATION Edition 1985 (E) Material measures of length for general use Material measures of length for general use OIML R 35 Edition 1985 (E) ORGANISATION INTERNATIONALE DE

More information

INTERNATIONAL OIML R 45 RECOMMENDATION. Casks and barrels ORGANISATION INTERNATIONALE INTERNATIONAL ORGANIZATION. Edition 1980 (E)

INTERNATIONAL OIML R 45 RECOMMENDATION. Casks and barrels ORGANISATION INTERNATIONALE INTERNATIONAL ORGANIZATION. Edition 1980 (E) INTERNATIONAL OIML R 45 RECOMMENDATION Edition 1980 (E) Casks and barrels Tonneaux et futailles OIML R 45 Edition 1980 (E) ORGANISATION INTERNATIONALE DE MÉTROLOGIE LÉGALE INTERNATIONAL ORGANIZATION OF

More information

OIML R 75-2 RECOMMENDATION. Edition 2002 (E) ORGANISATION INTERNATIONALE INTERNATIONAL ORGANIZATION

OIML R 75-2 RECOMMENDATION. Edition 2002 (E) ORGANISATION INTERNATIONALE INTERNATIONAL ORGANIZATION INTERNATIONAL RECOMMENDATION OIML R 75-2 Edition 2002 (E) Heat meters Part 2: Type approval tests and initial verification tests Compteurs d énergie thermique Partie 2: Essais d approbation de type et

More information

Non-destructive testing Equipment for eddy current examination. Part 1: Instrument characteristics and verification

Non-destructive testing Equipment for eddy current examination. Part 1: Instrument characteristics and verification Provläsningsexemplar / Preview INTERNATIONAL STANDARD ISO 15548-1 Second edition 2013-12-01 Non-destructive testing Equipment for eddy current examination Part 1: Instrument characteristics and verification

More information

Non-destructive testing Equipment for eddy current examination Array probe characteristics and verification

Non-destructive testing Equipment for eddy current examination Array probe characteristics and verification Provläsningsexemplar / Preview INTERNATIONAL STANDARD ISO 20339 First edition 2017-03 Non-destructive testing Equipment for eddy current examination Array probe characteristics and verification Essais

More information

5500A. Multi-Product Calibrator. Extended Specifications 2005

5500A. Multi-Product Calibrator. Extended Specifications 2005 5500A Multi-Product Calibrator Extended Specifications 2005 5500A Specifications The following paragraphs detail specifications for the 5500A Calibrator. The specifications are valid after allowing a warm-up

More information

Calibration Laboratory Assessment Service CLAS Certificate Number Page 1 of 10

Calibration Laboratory Assessment Service CLAS Certificate Number Page 1 of 10 Calibration Laboratory Assessment Service CLAS Certificate Number 95-02 Page 1 of 10 400 Britannia Road East, Unit #1 Mississauga, Ontario L4Z 1X9 Contact: Mr. Vince Casali Tel (905) 890-7600, (800) 36FLUKE

More information

5520A. Multi-Product Calibrator. Extended Specifications 2005

5520A. Multi-Product Calibrator. Extended Specifications 2005 5520A Multi-Product Calibrator Extended Specifications 2005 5520A Specifications The following tables list the 5520A specifications. All specifications are valid after allowing a warm-up period of 30 minutes,

More information

IDEAL INDUSTRIES, INC. TECHNICAL MANUAL MODEL: MODEL: Multimeter Service Information

IDEAL INDUSTRIES, INC. TECHNICAL MANUAL MODEL: MODEL: Multimeter Service Information IDEAL INDUSTRIES, INC. TECHNICAL MANUAL MODEL: 61-340 MODEL: 61-342 Multimeter Service Information The Service Information provides the following information: Precautions and safety information Specifications

More information

Fluke 5820A Oscilloscope Calibrator Specifications

Fluke 5820A Oscilloscope Calibrator Specifications Fluke 5820A Oscilloscope Calibrator Specifications Voltage Function Volt Function DC Signal Square Wave Signal 1 Load into 50Ω into 1 MΩ into 50Ω into 1 MΩ Amplitude Range 0V to ±6.6V 0V to ±130V ±1 mv

More information

TITLE OF THE CD (English): Revision: OIML B 10-1 and B 10-2 Framework for a Mutual Acceptance Arrangement on OIML Type Evaluations

TITLE OF THE CD (English): Revision: OIML B 10-1 and B 10-2 Framework for a Mutual Acceptance Arrangement on OIML Type Evaluations COMMITTEE DRAFT OIML CD 2 Date: 12 November 2010 Reference number: B 10/2010/CD 2 Supersedes document: B 10/2009/CD 1 OIML TC 3/SC 5 Circulated to P- and O-members and liaison international bodies and

More information

79/26 Series III Multimeter

79/26 Series III Multimeter 79/26 Series III Multimeter Instruction Sheet W Read First: Safety Information Never use the meter if the meter or test leads look damaged. Be sure the test leads and switch are in the correct position

More information

TECHNICAL SPECIFICATION

TECHNICAL SPECIFICATION TECHNICAL SPECIFICATION IEC/TS 61800-8 Edition 1.0 2010-05 colour inside Adjustable speed electrical power drive systems Part 8: Specification of voltage on the power interface INTERNATIONAL ELECTROTECHNICAL

More information

Manual Supplement. This supplement contains information necessary to ensure the accuracy of the above manual.

Manual Supplement. This supplement contains information necessary to ensure the accuracy of the above manual. Manual Title: 550A Getting Started Supplement Issue: Part Number: 415509 Issue Date: 9/18 Print Date: November 01 Page Count: 19 Revision/Date: This supplement contains information necessary to ensure

More information

5520A. Multi-Product Calibrator. Extended Specifications

5520A. Multi-Product Calibrator. Extended Specifications 5520A Multi-Product Calibrator Extended Specifications Specifications The following tables list the 5520A specifications. All specifications are valid after allowing a warm-up period of 30 minutes, or

More information

IDEAL INDUSTRIES, INC. TECHNICAL MANUAL MODEL:

IDEAL INDUSTRIES, INC. TECHNICAL MANUAL MODEL: IDEAL INDUSTRIES, INC. TECHNICAL MANUAL MODEL: 61-352 The Service Information provides the following information: Precautions and safety information Specifications Basic maintenance (cleaning, replacing

More information

Manual Supplement. This supplement contains information necessary to ensure the accuracy of the above manual.

Manual Supplement. This supplement contains information necessary to ensure the accuracy of the above manual. Manual Title: 5502E Getting Started Supplement Issue: 3 Part Number: 4155211 Issue Date: 9/18 Print Date: November 2012 Page Count: 12 Revision/Date: This supplement contains information necessary to ensure

More information

Fluke 83V and 87V Digital Multimeters

Fluke 83V and 87V Digital Multimeters Fluke 83V and 87V Digital Multimeters Detailed Specifications For all detailed specifications: is given as ±([% of reading] + [number of least significant digits]) at 18 C to 28 C, with relative humidity

More information

SIMEAS-T. Operating Instructions Transducer without auxiliary power. 7KG6111 and 7KG6101. Operating Instructions

SIMEAS-T. Operating Instructions Transducer without auxiliary power. 7KG6111 and 7KG6101. Operating Instructions Operating Instructions SIMEAS-T s Operating Instructions Transducer without auxiliary power for alternating current for alternating voltage for alternating voltage with expanded end range 7KG6111 and 7KG6101

More information

OIML R 46-3 RECOMMENDATION. Edition 2013 (E) ORGANISATION INTERNATIONALE INTERNATIONAL ORGANIZATION

OIML R 46-3 RECOMMENDATION. Edition 2013 (E) ORGANISATION INTERNATIONALE INTERNATIONAL ORGANIZATION INTERNATIONAL RECOMMENDATION OIML R 46-3 Edition 2013 (E) Active electrical energy meters. Part 3: Test report format Compteurs actifs d'énergie électrique. Partie 3: Format du rapport d essais OIML R

More information

INTERNATIONAL STANDARD

INTERNATIONAL STANDARD INTERNATIONAL STANDARD IEC 60603-7-7 Second edition 2006-06 Connectors for electronic equipment Part 7-7: Detail specification for 8-way, shielded, free and fixed connectors, for data transmissions with

More information

Part 1: General principles

Part 1: General principles Provläsningsexemplar / Preview INTERNATIONAL STANDARD ISO 9934-1 Second edition 2015-09-01 Non-destructive testing Magnetic particle testing Part 1: General principles Essais non destructifs Magnétoscopie

More information

INTERNATIONAL STANDARD

INTERNATIONAL STANDARD INTERNATIONAL STANDARD IEC 60603-7-7 First edition 2002-04 Connectors for electronic equipment Part 7-7: Detail specification for 8-way, shielded, free and fixed connectors, for data transmission with

More information

Prepared by Mick Maytum

Prepared by Mick Maytum IEC Technical Committee 109: Standards on insulation co-ordination for low-voltage equipment Warning Prepared by Mick Maytum mjmaytum@gmail.com The document content is of a general nature only and is not

More information

Human response to vibration Measuring instrumentation. Part 1: General purpose vibration meters

Human response to vibration Measuring instrumentation. Part 1: General purpose vibration meters INTERNATIONAL STANDARD ISO 8041-1 First edition 2017-05 Human response to vibration Measuring instrumentation Part 1: General purpose vibration meters Réponse des individus aux vibrations Appareillage

More information

INTERNATIONAL STANDARD

INTERNATIONAL STANDARD INTERNATIONAL STANDARD ISO 2954 Second edition 2012-04-01 Mechanical vibration of rotating and reciprocating machinery Requirements for instruments for measuring vibration severity Vibrations mécaniques

More information

INTERNATIONAL STANDARD

INTERNATIONAL STANDARD INTERNATIONAL STANDARD IEC 61000-4-5 Second edition 2005-11 BASIC EMC PUBLICATION Electromagnetic compatibility (EMC) Part 4-5: Testing and measurement techniques Surge immunity test This English-language

More information

OIML R 46-1/-2 RECOMMENDATION. Edition 2012 (E) ORGANISATION INTERNATIONALE INTERNATIONAL ORGANIZATION

OIML R 46-1/-2 RECOMMENDATION. Edition 2012 (E) ORGANISATION INTERNATIONALE INTERNATIONAL ORGANIZATION INTERNATIONAL RECOMMENDATION OIML R 46-1/-2 Edition 2012 (E) Active electrical energy meters. Part 1: Metrological and technical requirements Part 2: Metrological controls and performance tests Compteurs

More information

This document is a preview generated by EVS

This document is a preview generated by EVS INTERNATIONAL STANDARD ISO 15848-2 Second edition 2015-06-01 Industrial valves Measurement, test and qualification procedures for fugitive emissions Part 2: Production acceptance test of valves Robinetterie

More information

TR CRITERIA FOR LABORATORY ACCREDITATION IN THE FIELD OF ELECTRICAL DCLF METROLOGY. Approved By: Senior Manager: Mpho Phaloane Revised By:

TR CRITERIA FOR LABORATORY ACCREDITATION IN THE FIELD OF ELECTRICAL DCLF METROLOGY. Approved By: Senior Manager: Mpho Phaloane Revised By: CRITERIA FOR LABORATORY ACCREDITATION IN THE FIELD OF ELECTRICAL LF METROLOGY Approved By: Senior Manager: Mpho Phaloane Revised By: Specialist Technical Committee Date of Approval: 2015-08-26 Date of

More information

Protein Measuring Instruments for Cereal Grain and Oil Seeds

Protein Measuring Instruments for Cereal Grain and Oil Seeds OIML TC17 / SC 8 / N 4 May 2006 Supersedes document: N 3 INTERNATIONAL ORGANIZATION OF LEGAL METROLOGY First Committee Draft of a Recommendation on Protein Measuring Instruments for Cereal Grain and Oil

More information

ISO INTERNATIONAL STANDARD. Non-destructive testing Ultrasonic thickness measurement

ISO INTERNATIONAL STANDARD. Non-destructive testing Ultrasonic thickness measurement INTERNATIONAL STANDARD ISO 16809 First edition 2012-11-15 Non-destructive testing Ultrasonic thickness measurement Essais non destructifs Mesurage de l'épaisseur par ultrasons Reference number ISO 2012

More information

SERIES K: PROTECTION AGAINST INTERFERENCE

SERIES K: PROTECTION AGAINST INTERFERENCE International Telecommunication Union ITU-T K.49 TELECOMMUNICATION STANDARDIZATION SECTOR OF ITU (12/2005) SERIES K: PROTECTION AGAINST INTERFERENCE Test requirements and performance criteria for voice

More information

ISO INTERNATIONAL STANDARD. Non-destructive testing Ultrasonic inspection Evaluating electronic characteristics of ultrasonic test instruments

ISO INTERNATIONAL STANDARD. Non-destructive testing Ultrasonic inspection Evaluating electronic characteristics of ultrasonic test instruments INTERNATIONAL STANDARD ISO 12710 First edition 2002-09-15 Non-destructive testing Ultrasonic inspection Evaluating electronic characteristics of ultrasonic test instruments Essais non destructifs Contrôle

More information

1587/1577. Insulation Multimeters. Technical Data. Two powerful tools in one.

1587/1577. Insulation Multimeters. Technical Data. Two powerful tools in one. Test Equipment Depot - 800.517.8431-99 Washington Street Melrose, MA 02176 - TestEquipmentDepot.com 1587/1577 Insulation Multimeters Technical Data Two powerful tools in one. The Fluke 1587 and 1577 Insulation

More information

RECOMMENDATION ITU-R BT *

RECOMMENDATION ITU-R BT * Rec. ITU-R BT.656-4 1 RECOMMENDATION ITU-R BT.656-4 * Interfaces for digital component video signals in 525-line and 625-line television systems operating at the 4:2:2 level of Recommendation ITU-R BT.601

More information

RECOMMENDATION ITU-R BT.1302 *

RECOMMENDATION ITU-R BT.1302 * Rec. ITU-R BT.1302 1 RECOMMENDATION ITU-R BT.1302 * Interfaces for digital component video signals in 525-line and 625-line television systems operating at the 4:2:2 level of Recommendation ITU-R BT.601

More information

Four-Channel Differential AC Amplifier

Four-Channel Differential AC Amplifier Four-Channel Differential AC Amplifier INSTRUCTION MANUAL FOR HIGH-GAIN DIFFERENTIAL AMPLIFIER MODEL 1700 Serial # Date A-M Systems, Inc. PO Box 850 Carlsborg, WA 98324 U.S.A. 360-683-8300 800-426-1306

More information

INTERNATIONAL STANDARD

INTERNATIONAL STANDARD INTERNATIONAL STANDARD IEC 60728-11 Second edition 2005-01 Cable networks for television signals, sound signals and interactive services Part 11: Safety IEC 2005 Copyright - all rights reserved No part

More information

Initial and subsequent verification of measuring instruments and processes

Initial and subsequent verification of measuring instruments and processes INTERNATIONAL OIML D 20 DOCUMENT Edition 1988 (E) Initial and subsequent verification of measuring instruments and processes Vérifications primitive et ultérieure des instruments et processus de mesure

More information

)454 / 03/0(/-%4%2 &/2 53% /. 4%,%0(/.%490% #)2#5)43 30%#)&)#!4)/.3 &/2 -%!352).' %15)0-%.4 %15)0-%.4 &/2 4(% -%!352%-%.4 /&!.!,/'5% 0!2!

)454 / 03/0(/-%4%2 &/2 53% /. 4%,%0(/.%490% #)2#5)43 30%#)&)#!4)/.3 &/2 -%!352).' %15)0-%.4 %15)0-%.4 &/2 4(% -%!352%-%.4 /&!.!,/'5% 0!2! INTERNATIONAL TELECOMMUNICATION UNION )454 / TELECOMMUNICATION (10/94) STANDARDIZATION SECTOR OF ITU 30%#)&)#!4)/.3 &/2 -%!352).' %15)0-%.4 %15)0-%.4 &/2 4(% -%!352%-%.4 /&!.!,/'5% 0!2!-%4%23 03/0(/-%4%2

More information

1587/1577 Insulation Multimeters

1587/1577 Insulation Multimeters 1587/1577 Insulation Multimeters Technical Data Two powerful tools in one. The Fluke 1587 and 1577 Insulation Multimeters combine a digital insulation tester with a full-featured, true-rms digital multimeter

More information

INTERNATIONAL STANDARD

INTERNATIONAL STANDARD INTERNATIONAL STANDARD IEC 61935-1 Edition 3.0 2009-07 Specification for the testing of balanced and coaxial information technology cabling Part 1: Installed balanced cabling as specified in ISO/IEC 11801

More information

SDM3045X Digital Multimeter. DataSheet

SDM3045X Digital Multimeter. DataSheet SDM3045X Digital Multimeter DataSheet-2016.06 User-friendly Design Product Overview SDM3045X is a 4½ digit digital (60000 count) multimeter incorporating a dual-display and is especially well suited for

More information

MGM 3000X Q67000-A5179 P-DSO-20-1 (SMD) MGM 3000X Q67006-A5179 P-DSO-20-1 Tape & Reel (SMD)

MGM 3000X Q67000-A5179 P-DSO-20-1 (SMD) MGM 3000X Q67006-A5179 P-DSO-20-1 Tape & Reel (SMD) Video Modulator for FM/AM-Audio MGM 3000X Bipolar IC Features FM- and AM-audio modulator Audio carrier output for suppression of harmonics Sync level clamping of video input signal Controlling of peak

More information

NI PXI-4461 Specifications

NI PXI-4461 Specifications NI PXI-446 Specifications Analog Input Input Characteristics This document lists specifications for the NI PXI-446 Dynamic Signal Acquisition (DSA) device. These specifications are typical at 5 C unless

More information

INAB Policy Statement on Scope Format for Calibration Laboratories PS27 1. INTRODUCTION

INAB Policy Statement on Scope Format for Calibration Laboratories PS27 1. INTRODUCTION INAB Policy Statement on Scope Format for Calibration Laboratories PS27 1. INTRODUCTION 1.1. The definitive statement of the accreditation status of a calibration laboratory is the accreditation certificate

More information

34134A AC/DC DMM Current Probe. User s Guide. Publication number April 2009

34134A AC/DC DMM Current Probe. User s Guide. Publication number April 2009 User s Guide Publication number 34134-90001 April 2009 For Safety information, Warranties, Regulatory information, and publishing information, see the pages at the back of this book. Copyright Agilent

More information

Optics and optical instruments Field procedures for testing geodetic and surveying instruments. Part 8:

Optics and optical instruments Field procedures for testing geodetic and surveying instruments. Part 8: Provläsningsexemplar / Preview INTERNATIONAL STANDARD ISO 17123-8 Second edition 2015-06-15 Optics and optical instruments Field procedures for testing geodetic and surveying instruments Part 8: GNSS field

More information

DM-45 Digital Multimeter

DM-45 Digital Multimeter INSTRUCTION MANUAL DM-45 Digital Multimeter Read and understand all of the instructions and safety information in this manual before operating or servicing this tool. Description The Greenlee DM-45 Digital

More information

INTERNATIONAL. High-voltage test techniques Partial discharge measurements

INTERNATIONAL. High-voltage test techniques Partial discharge measurements INTERNATIONAL STANDARD IEC 60270 Third edition 2000-12 High-voltage test techniques Partial discharge measurements This English-language version is derived from the original bilingual publication by leaving

More information

Declaration of Conformity to the DeviceNet Specification

Declaration of Conformity to the DeviceNet Specification Declaration of Conformity to the DeviceNet Specification ODVA hereby issues this Declaration of Conformity to the DeviceNet Specification for the product(s) described below. The Vendor listed below (the

More information

INTERNATIONAL STANDARD

INTERNATIONAL STANDARD IEC 61606-3 INTERNATIONAL STANDARD Edition 1.0 2008-10 Audio and audiovisual equipment Digital audio parts Basic measurement methods of audio characteristics Part 3: Professional use IEC 61606-3:2008(E)

More information

TECHNICAL REQUIREMENTS FOR ELECTROMAGNETIC DISTURBANCES EMITTED FROM LIGHTING EQUIPMENT INSTALLED IN TELECOMMUNICATION CENTERS

TECHNICAL REQUIREMENTS FOR ELECTROMAGNETIC DISTURBANCES EMITTED FROM LIGHTING EQUIPMENT INSTALLED IN TELECOMMUNICATION CENTERS TR550004 TECHNICAL REQUIREMENTS FOR ELECTROMAGNETIC DISTURBANCES EMITTED FROM LIGHTING EQUIPMENT INSTALLED IN TELECOMMUNICATION CENTERS TR NO. 174001 EDITION 2.1 September 3 rd, 2018 Nippon Telegraph and

More information

Type Ordering Code Package TDA Q67000-A5168 P-DIP-18-5

Type Ordering Code Package TDA Q67000-A5168 P-DIP-18-5 Video Modulator for FM-Audio TDA 5666-5 Preliminary Data Bipolar IC Features FM-audio modulator Sync level clamping of video input signal Controlling of peak white value Continuous adjustment of modulation

More information

This is a preview - click here to buy the full publication TECHNICAL REPORT. Short-circuit currents in three-phase a.c. systems

This is a preview - click here to buy the full publication TECHNICAL REPORT. Short-circuit currents in three-phase a.c. systems TECHNICAL REPORT IEC TR 60909-1 Second edition 2002-07 Short-circuit currents in three-phase a.c. systems Part 1: Factors for the calculation of short-circuit currents according to IEC 60909-0 This English-language

More information

OIML D 11 DOCUMENT. Edition 2013 (E) ORGANISATION INTERNATIONALE INTERNATIONAL ORGANIZATION

OIML D 11 DOCUMENT. Edition 2013 (E) ORGANISATION INTERNATIONALE INTERNATIONAL ORGANIZATION INTERNATIONAL DOCUMENT OIML D 11 Edition 2013 (E) General requirements for measuring instruments - Environmental conditions Exigences générales pour les instruments de mesure - Conditions environnementales

More information

PXIe Contents. Required Software CALIBRATION PROCEDURE

PXIe Contents. Required Software CALIBRATION PROCEDURE CALIBRATION PROCEDURE PXIe-5160 This document contains the verification and adjustment procedures for the PXIe-5160. Refer to ni.com/calibration for more information about calibration solutions. Contents

More information

Guide for Conducting a Stray Voltage Evaluation Under the MPSC Stray Voltage Rule 1

Guide for Conducting a Stray Voltage Evaluation Under the MPSC Stray Voltage Rule 1 Guide for Conducting a Stray Voltage Evaluation Under the MPSC Stray Voltage Rule 1 Introduction and Overall Purpose: This guide is intended to be used in conjunction with the administrative rules addressing

More information

This document is a preview generated by EVS

This document is a preview generated by EVS INTERNATIONAL STANDARD ISO 19235 First edition 2015-10-15 Analogue quartz clocks Timing accuracy Horloges analogiques à quartz Précision du temps Reference number ISO 19235:2015(E) ISO 2015 ISO 19235:2015(E)

More information

Calibration Specialists Ltd

Calibration Specialists Ltd The National Technological Park, Castletroy, Limerick Calibration Laboratory Registration number: 001C is accredited by the Irish National Board (INAB) to undertake calibration as detailed in the Schedule

More information

Schedule of Accreditation issued by United Kingdom Accreditation Service 2 Pine Trees, Chertsey Lane, Staines-upon-Thames, TW18 3HR, UK

Schedule of Accreditation issued by United Kingdom Accreditation Service 2 Pine Trees, Chertsey Lane, Staines-upon-Thames, TW18 3HR, UK 2 Pine Trees, Chertsey Lane, Staines-upon-Thames, TW18 3HR, Unit 8 The Centre Holywell Business Park Northfield Road Southam Warwickshire CV47 0FP Contact: Mr N Morgan Tel: +44 (0)1926 812066 Fax: +44

More information

NI 6143 Specifications

NI 6143 Specifications NI 6143 Specifications This document lists the I/O terminal summary and specifications for the NI PCI/PXI-6143. For the most current edition of this document, refer to ni.com/manuals. Refer to the DAQ

More information

SPECIFICATION AND PERFORMANCE CHARACTERISTICS SERIAL ATA CABLE ASSEMBLIES

SPECIFICATION AND PERFORMANCE CHARACTERISTICS SERIAL ATA CABLE ASSEMBLIES SPECIFICATION AND PERFORMANCE CHARACTERISTICS OF SERIAL ATA CABLE ASSEMBLIES CIRCUIT ASSEMBLY CORP. 18 THOMAS STREET, IRVINE, CA 92618-2777 Page No. 1 CONTENTS: 1.0 SCOPE.. 3 2.0 APPLICABLE DOCUMENTS 3

More information

RS Pro DIGITAL MULTI-METER SDM3055 SERIES

RS Pro DIGITAL MULTI-METER SDM3055 SERIES ENGLISH Product Datasheet Stock No: 1236465 1236466 RS Pro DIGITAL MULTI-METER SDM3055 SERIES Product Overview RSDM3055 is a digital multimeter designed with 5 ½ digits readings resolution and dual-display,

More information

Microscopes Imaging distances related to mechanical reference planes. Part 2: Infinity-corrected optical systems

Microscopes Imaging distances related to mechanical reference planes. Part 2: Infinity-corrected optical systems Provläsningsexemplar / Preview INTERNATIONAL STANDARD ISO 9345-2 Second edition 2014-09-15 Microscopes Imaging distances related to mechanical reference planes Part 2: Infinity-corrected optical systems

More information

Agilent 4072A Advanced Parametric Test System with Agilent SPECS

Agilent 4072A Advanced Parametric Test System with Agilent SPECS Agilent 4072A Advanced Parametric Test System with Agilent SPECS Technical Data 1. General Description The Agilent 4072A Advanced Parametric Test System is designed to perform precision DC measurement,

More information

Flight Model (FM) Lot Acceptance Test Model (LAT) Quartz Crystal. Synthetic HiQ Quartz, SC-cut, HC-35/U 4-point

Flight Model (FM) Lot Acceptance Test Model (LAT) Quartz Crystal. Synthetic HiQ Quartz, SC-cut, HC-35/U 4-point in Specification AXIOM75SH Rev.: 1 Date: 2018-02-16 Oscillator type: Low Phase Noise OCXO for Space Application (COTS version) Features: Lower cost Commercial Off-The-Shelf version (COTS) Qualified according

More information

INTERNATIONAL STANDARD

INTERNATIONAL STANDARD INTERNATIONAL STANDARD IEC 60748-4-3 First edition 2006-08 Semiconductor devices Integrated circuits Part 4-3: Interface integrated circuits Dynamic criteria for analogue-digital converters (ADC) IEC 2006

More information

ISO INTERNATIONAL STANDARD. Non-destructive testing of welds Radiographic testing Part 1: X- and gamma-ray techniques with film

ISO INTERNATIONAL STANDARD. Non-destructive testing of welds Radiographic testing Part 1: X- and gamma-ray techniques with film INTERNATIONAL STANDARD ISO 17636-1 First edition 2013-01-15 Non-destructive testing of welds Radiographic testing Part 1: X- and gamma-ray techniques with film Contrôle non destructif des assemblages soudés

More information

Rakon Product Proposal

Rakon Product Proposal RTX5032A -- SMD Temperature Compensated Crystal Oscillator -- -- High performance TCXO offering excellent Phase Noise, Frequency Stability and VCO tilt compensation. -- Product description -- The RTX5032A

More information

Low Power, Precision FET-INPUT OPERATIONAL AMPLIFIERS

Low Power, Precision FET-INPUT OPERATIONAL AMPLIFIERS OPA3 OPA3 OPA3 OPA3 OPA3 OPA3 OPA3 OPA3 OPA3 Low Power, Precision FET-INPUT OPERATIONAL AMPLIFIERS FEATURES LOW QUIESCENT CURRENT: 3µA/amp OPA3 LOW OFFSET VOLTAGE: mv max HIGH OPEN-LOOP GAIN: db min HIGH

More information

Schedule of Accreditation issued by United Kingdom Accreditation Service 2 Pine Trees, Chertsey Lane, Staines-upon-Thames, TW18 3HR, UK

Schedule of Accreditation issued by United Kingdom Accreditation Service 2 Pine Trees, Chertsey Lane, Staines-upon-Thames, TW18 3HR, UK code Location code Customers Sites 2 Pine Trees, Chertsey Lane, Staines-upon-Thames, TW18 3HR, UK Calibration Centre Bolkiah Garrison BB3510 Negara Brunei Darussalam Contact: Mr Yussof Taha Tel: +673-2-386475

More information

DVM98. True RMS Digital Multimeter. 1 Safety information. 1.1 Preliminary. 1.2 During use

DVM98. True RMS Digital Multimeter. 1 Safety information. 1.1 Preliminary. 1.2 During use True RMS Digital Multimeter DVM98 1 Safety information This multimeter has been designed according to IEC - 1010 concerning electronic measuring instruments with an overvoltage category (CAT II) and pollution

More information

SEM210 SERIES. Programmable In-head Universal Temperature Transmitter INDEX SECTION CONTENTS PAGE NO.

SEM210 SERIES. Programmable In-head Universal Temperature Transmitter INDEX SECTION CONTENTS PAGE NO. INDEX SECTION CONTENTS PAGE NO. SEM210 SERIES 1.0 DESCRIPTION 1 2.0 SPECIFICATION 15 3.0 INSTALLATION 67 Programmable Inhead Universal Temperature Transmitter Status Instruments Ltd, Green Lane Business

More information

supplies and frequency converters. BumbleBee is also very effective in fast transient measurements

supplies and frequency converters. BumbleBee is also very effective in fast transient measurements High Voltage Differential Probe Order-No: 880-102-501 Features: 1000 V CAT III 400 MHz Bandwidth < 1 % Deviation within Specified Operating Temperatures (0 C to +50 C) High CMRR 4 Mode Attenuation Useable

More information

This document is a preview generated by EVS

This document is a preview generated by EVS INTERNATIONAL STANDARD ISO 3070-2 Fourth edition 2016-08-01 Corrected version 2016-12-15 Machine tools Test conditions for testing the accuracy of boring and milling machines with horizontal spindle Part

More information

BIOMEDICAL INSTRUMENTATION PROBLEM SHEET 1

BIOMEDICAL INSTRUMENTATION PROBLEM SHEET 1 BIOMEDICAL INSTRUMENTATION PROBLEM SHEET 1 Dr. Gari Clifford Hilary Term 2013 1. (Exemplar Finals Question) a) List the five vital signs which are most commonly recorded from patient monitors in high-risk

More information

Schedule of Accreditation issued by United Kingdom Accreditation Service 2 Pine Trees, Chertsey Lane, Staines-upon-Thames, TW18 3HR, UK

Schedule of Accreditation issued by United Kingdom Accreditation Service 2 Pine Trees, Chertsey Lane, Staines-upon-Thames, TW18 3HR, UK 1 Wilson Street Contact: Mr Ibrahim Ibrahim Thornaby Tel: +44 (0) 1642 626148 Stockton-On-Tees Fax: +44 (0) 870 143 1869 TS17 7AR E-Mail: ibrahim@calibrate.co.uk United Kingdom Website: www.calibrate.co.uk

More information

Part 3: Image quality classes

Part 3: Image quality classes INTERNATIONAL STANDARD ISO 19232-3 Second edition 2013-06-15 Non-destructive testing Image quality of radiographs Part 3: Image quality classes Essais non destructifs Qualité d image des radiogrammes Partie

More information

IAEA Coordinated Research Project on Development of Harmonized QA/QC Procedures for Maintenance and Repair of Nuclear Instruments

IAEA Coordinated Research Project on Development of Harmonized QA/QC Procedures for Maintenance and Repair of Nuclear Instruments PROCEDURE: TEST PROCEDURE FOR GEIGER-MUELLER RADIATION DETECTORS Nº: MRNI-501 DECEMBER 2008 PAGE: 1 OF: 17 IAEA Coordinated Research Project on Development of Harmonized QA/QC Procedures for Maintenance

More information

ISO INTERNATIONAL STANDARD

ISO INTERNATIONAL STANDARD INTERNATIONAL STANDARD ISO 10791-3 First edition 1998-12-15 Test conditions for machining centres Part 3: Geometric tests for machines with integral indexable or continuous universal heads (vertical Z-axis)

More information

Appendix A: Specifications

Appendix A: Specifications Established 1981 Advanced Test Equipment Rentals www.atecorp.com 800-404-ATEC (2832) Appendix A: Specifications This section provides a complete description of the video measurement set specifications.

More information

This is a preview - click here to buy the full publication INTERNATIONAL ELECTROTECHNICAL COMMISSION

This is a preview - click here to buy the full publication INTERNATIONAL ELECTROTECHNICAL COMMISSION INTERNATIONAL ELECTROTECHNICAL COMMISSION CISPR 22 Fifth edition 2005-04 INTERNATIONAL SPECIAL COMMITTEE ON RADIO INTERFERENCE Information technology equipment Radio disturbance characteristics Limits

More information

Power Amplifier. Input Characteristics. SPECIFICATIONS NI PXIe-4610 Power Amplifier. Signal Range

Power Amplifier. Input Characteristics. SPECIFICATIONS NI PXIe-4610 Power Amplifier. Signal Range SPECIFICATIONS NI PXIe-4610 Power Amplifier This document lists specifications for the NI PXIe-4610 Power Amplifier. These specifications are typical at 25 ºC unless otherwise stated. The operating range

More information

SERIES K: PROTECTION AGAINST INTERFERENCE

SERIES K: PROTECTION AGAINST INTERFERENCE I n t e r n a t i o n a l T e l e c o m m u n i c a t i o n U n i o n ITU-T K.132 TELECOMMUNICATION STANDARDIZATION SECTOR OF ITU (01/2018) SERIES K: PROTECTION AGAINST INTERFERENCE Electromagnetic compatibility

More information

EMC standards. Presented by: Karim Loukil & Kaïs Siala

EMC standards. Presented by: Karim Loukil & Kaïs Siala Training Course on Conformity and Interoperability on Type Approval testing for Mobile Terminals, Homologation Procedures and Market Surveillance, Tunis-Tunisia, from 20 to 24 April 2015 EMC standards

More information

ISO INTERNATIONAL STANDARD

ISO INTERNATIONAL STANDARD INTERNATIONAL STANDARD ISO 13472-2 First edition 2010-05-15 Acoustics Measurement of sound absorption properties of road surfaces in situ Part 2: Spot method for reflective surfaces Acoustique Mesurage

More information

BENNING. Service Information MODEL: MM1-1 / MM1-2 / MM1-3

BENNING. Service Information MODEL: MM1-1 / MM1-2 / MM1-3 BENNING MODEL: MM1-1 / MM1-2 / MM1-3 Service Information The Service Information provides the following information: Precautions and safety information Specifications Basic maintenance (cleaning, replacing

More information

INTERNATIONAL STANDARD

INTERNATIONAL STANDARD INTERNATIONAL STANDARD IEC 62092 First edition 2001-08 Utrasonics Hydrophones Characteristics and calibration in the frequency range from 15 MHz to 40 MHz Ultrasons Hydrophones Caractéristiques et étalonnage

More information

INTEGRATED CIRCUITS DATA SHEET. TDA1596 IF amplifier/demodulator for FM radio receivers. Product specification File under Integrated Circuits, IC01

INTEGRATED CIRCUITS DATA SHEET. TDA1596 IF amplifier/demodulator for FM radio receivers. Product specification File under Integrated Circuits, IC01 INTEGRATED CIRCUITS DATA SHEET File under Integrated Circuits, IC01 April 1991 GENERAL DESCRIPTION The provides IF amplification, symmetrical quadrature demodulation and level detection for quality home

More information

AES standard method for digital audio engineering Measurement of digital audio equipment

AES standard method for digital audio engineering Measurement of digital audio equipment Revision of AES17-1991 AES standard method for digital audio engineering Measurement of digital audio equipment Published by Audio Engineering Society, Inc. Copyright 1998 by the Audio Engineering Society

More information

n DC Voltage Measurement (DCV) 7461A n AC Voltage Measurement (ACV, ACV (AC+DC)) 7451A 7461A/7451A Additional error depending on the integration time

n DC Voltage Measurement (DCV) 7461A n AC Voltage Measurement (ACV, ACV (AC+DC)) 7451A 7461A/7451A Additional error depending on the integration time Specifications Unless otherwise specified, the measurement accuracy is guaranteed for one year under the following conditions: Temperature; 23±5 C, relative humidity; 85% or less (75% or less in resistance

More information