DESIGN AND RELIZATION MICROCONTROLLER BASED ELECTRONIC STETHOSCOPE DOR HEART AND SOUND ANALYSIS TO SUPPORT TELEMEDICINE SYSTEM

Size: px
Start display at page:

Download "DESIGN AND RELIZATION MICROCONTROLLER BASED ELECTRONIC STETHOSCOPE DOR HEART AND SOUND ANALYSIS TO SUPPORT TELEMEDICINE SYSTEM"

Transcription

1 DESIGN AND RELIZATION MICROCONTROLLER BASED ELECTRONIC STETHOSCOPE DOR HEART AND SOUND ANALYSIS TO SUPPORT TELEMEDICINE SYSTEM Desy Nurhayati 1), Soegijardjo Soegijoko 2) 1) Biomedical Engineering, School of Electrical Engineering and Informatics Institut Teknologi Bandung Jl. Ganesha 10 Labtek VIII lantai III Bandung Indonesia, 2) Biomedical Engineering, School of Electrical Engineering and Informatics Institut Teknologi Bandung Jl. Ganesha 10 Labtek VIII lantai III Bandung Indonesia, Abstract - This thesis describes the completed design and realization of a portable microcontroller-based electronic stethoscope prototype, for heart and lung sound analysis. The prototype has been designed to cover three auscultation modes: cardiac mode (in the range of 20 Hz to 660 Hz), lung mode (in the range of 150 Hz to 1600 Hz), and extended mode (20 Hz to 8000 Hz). It is expected that the prototype will have improved sound quality with respect to an ordinary acoustic stethoscope. Moreover, the electronic stethoscope has the following additional specifications: 40 times of voltage amplification, sound recording capability (in external EEPROM), heart rate computation and display (in beat per minute), volume (gain) control, and data transmit to PC. The objective of this thesis is to develop a relatively low cost and portable microcontroller-based electronic stethoscope that is expected to significantly contribute to the development telemedicine system and production of future biomedical engineering instruments in Indonesia. It is expected that the electronic stethoscope will be served as a diagnostic instrument and to support telemedicine system. Telemedicine, which uses computer technology to transmit information from primary health care services to referral hospital, is being explored to permit access to health care for patients in remote locations. A specialist in referral hospital can use telemedicine to receive information on a patient in a rural area and send back the necessary response for the specific patient. Since the electronic stethoscope has a sound recording capability, the recorded data can be sent to a PC for possible tele-consultation and tele-diagnosis applications. A series of hardware and software laboratory tests have been conducted and the encouraging results have been used for some hardware and software modification and improvements. Further laboratory measurements and tests using a signal generator, electronic measuring instrument and involving some limited patients have also been completed. The results obtained have indicated that the microcontrollerbased electronic stethoscope prototype has successfully performed its intended specifications. The prototype has also shown its performance during a number of laboratory tests with patients i.e.: body sound measurements, recording and retrieving patient data to/from the external EEPROM, heart rate measurement and display on LCD, and transmit data to PC. Battery life time is approximately 3 hours for continuous using. All of those functional tests have been successfully conducted in Biomedical Engineering Laboratory - ITB. In conclusion, a portable microcontrollerbased electronic stethoscope has been designed, realized and successfully tested. For further development, a number of improvements are still required. Microcontroller and/or PC sofware modules may be developed for further specific sound analysis, system integration, and telemedicine applications. Key words : electronic stethoscope, microcontroller, auscultation, heart sound, lung sound, heart rate, PC, telemedicine. 1. INTRODUCTION A medical doctor made patient diagnosis based on interview about patient disease, physical examination, and other supporting examination (laboratory, radiography, USG, etc). Physical examination consists of inspection, palpation, auscultation, and percussion. Auscultation is patient examination with stethoscope to hear body sound. [1] 1

2 Sound quality of conventional acoustic stethoscope is not clear, especially low intensity sound. The number of sounds could not detect with acoustic stethoscope. Its can make diagnosis error. 2.ELECTRONIC STETHOSCOPE Electronic stethoscope is like acoustic stethoscope, but it has electronic instrument for amplified body sound and make a better quality of body sound. Others function are record sound, mode sound, display heart rate, transmit data to PC and volume control. 3. SYSTEM DESIGN Electronic stethoscope specifications show on table Placement Design of Electret Condenser Microphone on Rubber Tube Electronic Stethoscope The electret condenser microphone is placed in a rubber tube connected to the electronic stethoscope prototype. Distance microphone from electronic stethoscope chest piece is 21,79 centimeters, like on the picture 1. A specific distance of microphone based on equation bellow : [2] L = V/4f L = Distance from microphone to chest piece. V = Sound speed on the air f = Sound resonance frequency Table 1. Technical and Functional Spesifications of Microcontroller Baseds Electronic Stethoscope Tecnical Spesifications Microcontroller ATMEGA8535 Power supply : 9 V battery Electret condenser microphone Preamplifier (TL082) Demultiplexer (4052) Wide band pass filter (TL082) Audio amplifier (LM386) External EEPROM (AT24C256) LCD MATRIX (M1632) Buzzer Earphone (Telebit) Serial communication (RS232) Fungtional Specifications Electronic stethoscope have 3 mode : Heart mode ( Hz) Lung mode ( Hz) Extended range mode ( Hz) Amplified sound 40 x Sound can be hear via earphone. Heart counting and display Transmit sound data to PC and display on PC The completed microcontrollerbased electronic stethoscope prototype consists of both hardware and software parts. The hardware part of the system consist of an electret condenser microphone, a pre-amplifier, a demultiplexer, three different band pass filters, an audio amplifier, and an earphone. Additionally, the system also includes: a microcontroller system, a 2 x 16 matrix LCD, a buzzer, 5 push-buttons, an external EEPROM, an RS- 232 converter, a volume control potentiometer, and an In-System Programming (ISP) port. Picture 1. Placement design of microphone on rubber tube electronic stethoscope This design uses electret condenser microphone with small size, it has 5 mm diameters and 3 mm thick, therefore can placed on rubber tube of electronic stethoscope which has 5 mm diameters. Microphone take from Geius headset HS- 02F. Technical specification of microphone are : sensitivity 58 db + 4 db, omnidirectional, 2,2 Kohm impendances, frequency Hz. [3] 3.2 Microcontroller ATMEGA 8535 Electronic stethoscope design uses microcontroller ATMEGA8535. [4] Function of microcontroller is to control electronic stethoscope operation. Picture 2 display block diagram of microcontroller based electronic stethoscope. Picture 3 display circuit diagram of microcontroller based electronic stethoscope. The microcontroller software has been especially developed to provide the following functions: system s initialization, menu control, demultiplexer control to select one of the three existing filters (modes), ADC control, LCD display, data transmit control, record data control, buzzer control and functional control of the five push buttons 2

3 The heart sound or lung sound received by the microphone, will be first amplified two times by the pre-amplifier, and filtered by the selected filter. The three wide band pass filters have three different frequency bands: heart mode (in the range of 20 Hz to 660 Hz), lung mode (in the range of 150 Hz to 1600 Hz), and extended mode (20 Hz to 8000 Hz). Each of the wide band pass filter uses a TL082 chip consisting of two operational amplifiers. The next stage is the LM386-based audio amplifier, having an amplification of 20 times. The earphone serves as the transducer to produce the audible signal required by the user... menu filter, transmit data, save data or heart rate. Flow chart of program at picture 4. Picture 4. Flow chart of microcontroller software Picture 2. Block Diagram of Microcontroller Based Electronic Stethoscope. Picture 5 show flow chart of mode heart rate. When select this mode, software flow chart start with initialization selecting filter based on input from push button. Heart band pass filter must be activated first when heart mode selected. After band pass filter activated, analog to digital conversion started. Next process is to save data in external EEPROM or transmit data to PC. If selected heart rate mode, heart counting started from digital data heart sound. Heart rate counting can be displayed on LCD (beat per minute). Picture 3. Circuit Diagram of Microcontroller Based Electronic Stethoscope. 3.3 Software Module First step when microcontroller program started is variable initialization, LCD, RS-232, interrupt and menu default. Menu procedure is a menu which operation depend on input from push button. Input from push button to select active menu : 3

4 4.1 Result of band pass filter circuit testing (frequency Hz) Testing of circuit block, to examine is the circuit operates like technical specification of electronic stethoscope. Frequency cut off : V out = V out max testing : 2 = 1,47 V : 2 = 1,039 V Testing result : Frequency cut off V out = 1,04 V : Frequency cutoff low = 18,55 Hz Frequency cutoff high = 623,4 Hz Frequency response of band pass filter circuit show at picture 7. R e s p o n p e n gu a t a n ( k a li) t e r h a da p f r e k u e n s i r a n g k a ia n ba n dp a s s f ilt e r ( f r e k u e n s i H z ) F re k u e n si (H z) Picture 5. Flow Chart of Microcontroler Sofware 4. TESTING AND ANALYSIS After instrument realization, a number of testing has be done, testing of hardware mad software. Picture 6 show microcontroller based electronic stethoscope realization. Picture 6. Microcontroller Based Electronic Stethoscope Realization. Picture 7. Frequency response of band pass filter circuit ( Hz) 4.2 Testing and Analysis of battery life time This testing to measure battery life time. Battery energy (E bat ) counting from voltage of battery and ampere hour of battery. The equation can find bellow : E bat = V Ah = V I t E bat = battery energy (Joule) V bat = battery voltage (Volt) Ah = Ampere-hour V = voltage circuit (Volt) I = current (ampere) t = battery half life (hours) Technical specification of battery : 9 volts 175 mah, energy of this battery is 5670 joule. Measurement of voltage and current electronic stethoscope circuit operates with 9V batteries show on table 2. Table 2. Measurement of voltage and current electronic stethoscope circuit Tegangan Rangkaian (Vdc rms) Arus Rangkaian (Idc rms) Catu (+) Catu (-) Catu (+) Catu (-) + 8,3 V - 8,7 V + 56 ma - 20,2 ma 4

5 Battery life time approximately : For positive supply : Ebat 5670Joule t = = = V I 8.3V 56mA 12198,95 seconds = 3,39 hours For negative supply: Ebat 5670Joule t = = = V I 8,7V 20,2mA 32263,57 seconds = 8,96 hours Based that equation battery life time approximately 3 hours for continuous using. Table 3. Comparing heart rate counted by electronic stethoscope and manual (Patient B) 4.3 Testing results of heart rate counting This testing to compare between manual and electronic stethoscope heart rate counting. Manual heart rate counting uses acoustic stethoscope to counting heart rate. Measurement heart rate to two patients, with assumption this patient on the rest. The results displayed on LCD show at picture 8. Patient A (Woman, 58 years old, with arhythmia cordis) Arithmia cordis is heart rhythm disorders. Heart rate counting with electronic stethoscope is beat per minute, heart rate average 89,2 beat per minute, and deviation standard 2,394. With manual counting, heart rate beat per minute, heart rate average 83,2 beat per minute, and deviation standard 6,746. Difference between measurement with electronic stethoscope and manual because patient with arhythmia cordis has different cardiac rhythm and different amplitude. Table 4. Comparing heart rate counted by electronic stethoscope and manual (Patient A) Picture 8. Heart rate displayed on LCD Patient B (Woman, 30 years old, without heart diesease). Result of heart rate counting with electronic stethoscope is beat per minute, average heart rate 82,9 beat per minute, and deviation standard 2,44. Heart rate counting manual is beat per minute, average heart rate 84 beat per minute, and deviation standard 5,66. Result of heart rate testing show on table 3. Testing show deviation standard of patient with arhytmia cordis bigger normal patient. Manual heart counting is subjective and depends on sensitivity and accuracy of a medical doctor. Electronic stethoscope heart rate counting is objective and accurate. 5

6 4.4 Testing Data Transmit to PC via Sound Card PC Transmit data to PC via sound card, and display sound signal on PC with Cool Edit 2.0 software. Sound signal of heart sound show at picture 9. Picture 11. Signal of heart sound (Mitral Valve) Picture 9. Sound signal of heart valve (mitral valve and tricuspidal valve) Picture 11 show that heart sound signal BJ I result from electrical activity and mechanical BJ IV of the heart on cardiac cycle. At that picture show four of BJ the II heart sound. Picture 12 show relationship between electrical and mechanical even BJ on III cardiac cycle 4.5 Testing Data Transmit to PC via Serial Communication RS232 Data transmit to PC via serial communication RS232, sound signal display on PC with Delphi software. Testing conducted to male patient 24 years old without heart disease. Picture 10 show heart sound data which transmit via serial communication RS232. Picture 12. Relationship Four Heart Sound and Electrical and Mechanical Even on Cardiac Cycle [21] Picture 10. Heart sound signal (mitral valve) which transmit via serial communication RS CONCLUSIONS 1. The results obtained have indicated that the microcontroller-based electronic stethoscope prototype has successfully performed its intended specifications. The prototype has also shown its performance during a number of laboratory tests with patients (six patients) 2. The microcontroller-based electronic stethoscope prototype can perform the following function: body sound measurements, recording and retrieving patient data to/from external EEPROM, 6

7 heart rate measurements, and transmit patient data to PC for further process. 3. Sound data transmit via serial communication or sound card PC, record this data file *.wav. 4. Data transmit via serial communication RS232 and display on PC, heart sound signal show first heart sound, second heart sound, third heart sound and fourth heart sound. 5. Sound signal transmit via sound card PC and display with Cool Edit 2.0 software. 6. Battery life time approximately 3 hours for continuous using 7. This electronic stethoscope specialty compare with another methods : recorded sound, transmit sound data to PC to support telemedicine system. Additional functions to select auscultation mode, display heart rate, portable, display sound signal on PC, and inexpensive. Easily operation and operator select menu from LCD. For further development, a number of improvements are still required. Microcontroller and/or PC sofware modules may be developed for further specific sound analysis, system integration, and telemedicine applications. REFERENCES [1] Bates Barbara, A Guide to Physical Examination and History Taking, JB Lipincot Company Philadelphian, 1991, Fifth Edition [2] Jocelyn Durand, Louis-Gilles Durand, Marie-Claude Grenien, Electronic Stethoscope, United States Patent No , 11 Februari 1997, (9 Maret 2006 jam 21.00) [3], Spesifikasi Teknis Headphone Genius Tipe HS-02F Foldable Headband Headset, [4], 8-bit AVR Microcontroller with 8 K bytes In- System Programmable Flash, Atmega 8535, (tanggal 2 Januari 2007 jam 20.00) [5] Boylestad, Robert L., Introduction Circuit Analysis 10th Edition, Prentice Hall, New York, 2003 [6] Brown et al, Introduction to Biomedical Equipment Technology, Prentice Hall, 2001, Fourth edition [7] Coughlin Robert, Frederick F. Driscott, Operational Amplifiers & Linear Integrated Circuits, 5 th International Edition, Prentice-Hall International Inc., 1998 [8] Erickson Barbara, Auskultasi Bunyi dan Bising Jantung Pedoman Praktis,, Binarupa Aksara, 1988 [9] I Made Joni, Budi Raharjo, Pemrograman C dan Implementasinya, Penerbit Informatika, 2006 [10] Jogiyanto Hartono, Konsep Dasar Pemrogramanan Bahasa C, Penerbit Andi, 2003 [11] Ladjamudin Al Bahra, Rekayasa Perangkat Lunak, Penerbit Graha Ilmu, 2006 [12] Lingga Wardana, Mikrokontroler AVR Seri ATMega 8535, Penerbit Andi, 2006 [13]., TL082 Wide Bandwidth Dual JFET Input Operational Amplifier, April 1998, National Semiconductor Datasheet CD-ROM [14], RS 232 Interface with MAX 232, (7 Januari 2007 jam 04.00) [15], Two-wire Serial EEPROM AT24512,, (12 Januari 2007 jam 21.00) [13], LM386 Low Voltage Audio Power Amplifier, Januari 2000, National Semiconductor Datasheet CD-ROM [16], HD44780U (LCD II), (1 Maret 2007 jam 05.00) [17], The Extended Concise LCD Data Sheet, (1 Maret 2007 jam 06.00) [18] Audioguru, Electronic Stethoscope 2, (24 April 2005 jam 15.00) [19] DesLauries et al, Combined Electronic Acoustical Stethoscope, (!% April 2005 jam 20.00) [20] Duran et al, Electronic Stethoscope, (11 April 2005 jam 20.00) [21] Webster, Medical Instrumentation Aplication and Design, John Wiley & Sons, Inc, 1988, Third Edition. 7

WIRELESS ELECTRONIC STETHOSCOPE USING ZIGBEE

WIRELESS ELECTRONIC STETHOSCOPE USING ZIGBEE WIRELESS ELECTRONIC STETHOSCOPE USING ZIGBEE Ms. Ashlesha Khond, Ms. Priyanka Das, Ms. Rani Kumari 1 Student, Electronics and Communication Engineering, SRM IST, Tamil Nadu, India 2 Student, Electronics

More information

Design and Implementation of Digital Stethoscope using TFT Module and Matlab Visualisation Tool

Design and Implementation of Digital Stethoscope using TFT Module and Matlab Visualisation Tool World Journal of Technology, Engineering and Research, Volume 3, Issue 1 (2018) 297-304 Contents available at WJTER World Journal of Technology, Engineering and Research Journal Homepage: www.wjter.com

More information

Internal Sound Denoising for Traditional Stethoscope Using Inverse Chebyshev IIR Bandstop Filter

Internal Sound Denoising for Traditional Stethoscope Using Inverse Chebyshev IIR Bandstop Filter Internal Sound Denoising for Traditional Stethoscope Using Inverse Chebyshev IIR Bandstop Filter Alonzo Alterado 1, Adrian Vergel Viar 1 and Reynaldo Ted Peñas II, MScEngg 2,* 1 Bachelor of Science in

More information

Obstacle Avoidance Mobile Robot With Ultrasonic Sensors

Obstacle Avoidance Mobile Robot With Ultrasonic Sensors JURNAL TEKNOLOGI TERPADU Vol. 5 No. 1 April 2017 ISSN 2338-6649 Received: February 2017 Accepted: March 2017 Published: April 2017 Obstacle Avoidance Mobile Robot With Ultrasonic Sensors Qory Hidayati

More information

CATALOG. ANALOG COMMUNICATION SYSTEMS DIGITAL COMMUNICATION SYSTEMS Microcontroller kits Arm controller kits PLC Trainer KITS Regulated Power supplies

CATALOG. ANALOG COMMUNICATION SYSTEMS DIGITAL COMMUNICATION SYSTEMS Microcontroller kits Arm controller kits PLC Trainer KITS Regulated Power supplies CATALOG ANALOG COMMUNICATION SYSTEMS DIGITAL COMMUNICATION SYSTEMS Microcontroller kits Arm controller kits PLC Trainer KITS Regulated Power supplies UNION INTRUMENTS #17 & 18, 4 th floor, Hanumathra Arcade

More information

NEXT-LAB SENSOR Scritto da Administrator Martedì 24 Settembre :01 - Ultimo aggiornamento Mercoledì 20 Luglio :44 NEXT-LAB SENSOR 1 / 7

NEXT-LAB SENSOR Scritto da Administrator Martedì 24 Settembre :01 - Ultimo aggiornamento Mercoledì 20 Luglio :44 NEXT-LAB SENSOR 1 / 7 NEXT-LAB SENSOR 1 / 7 3 Diversi strumenti per i moduli NEXT Next-Lab SPL Sensor A high precision sensor for measuring sound pressure - Lab-Bus digital bus - Internal precision ADC 16 Bit - Frequency range

More information

School of Electronic Science and Engineering, Nanjing University of Posts and Telecommunications, Nanjing, , China

School of Electronic Science and Engineering, Nanjing University of Posts and Telecommunications, Nanjing, , China 4th International Conference on Mechatronics, Materials, Chemistry and Computer Engineering (ICMMCCE 2015) A design and implementation of Pulse-Measure instrument based on Microcontroller Zhu Siqing1,

More information

Wireless Stethoscope for Anesthesiologists

Wireless Stethoscope for Anesthesiologists Wireless Stethoscope for Anesthesiologists Spring 2013 Client: Dr. Scott Springman Advisor: John Webster Team Leader: Charlie Rodenkirk Communicator: Alex Eaton BSAC: Lucas Haug BWIG: Yue Yin 1 Table of

More information

GSM based Patient monitoring system

GSM based Patient monitoring system For more Project details visit: http://www.projectsof8051.com/patient-monitoring-through-gsm-modem/ Code Project Title 1615 GSM based Patient monitoring system Synopsis for GSM based Patient monitoring

More information

Screening Audiometer

Screening Audiometer EE89 Electronic Design Lab (EDL) Report, EE Dept, IIT Bombay, December, 00 Screening Audiometer Group No. D0 Mahim Agrawal (0D000) < mahim@ee.iitb.ac.in > Ashok Kumar Bhardwaj (0D00) < ashokkb@ee.iitb.ac.in

More information

International Journal of Electronics and Communication Engineering & Technology (IJECET), ISSN COMMUNICATION ENGINEERING & TECHNOLOGY (IJECET)

International Journal of Electronics and Communication Engineering & Technology (IJECET), ISSN COMMUNICATION ENGINEERING & TECHNOLOGY (IJECET) 0976 INTERNATIONAL 6464(Print), ISSN 0976 6472(Online) JOURNAL Volume OF 4, Issue ELECTRONICS 1, January- February (2013), AND IAEME COMMUNICATION ENGINEERING & TECHNOLOGY (IJECET) ISSN 0976 6464(Print)

More information

Low Cost Screening Audiometer

Low Cost Screening Audiometer Abstract EE 389 EDL Report, EE Dept. IIT Bombay, submitted on Nov.2004 Low Cost Screening Audiometer Group No.: D3 Chirag Jain 01d07018 Prashant Yadav 01d07024 Puneet Parakh 01d07007 Supervisor: Prof.

More information

A Simple Microcontroller-Based 4-20 ma Current Loop Receiver for Sensors with Current Transmitters

A Simple Microcontroller-Based 4-20 ma Current Loop Receiver for Sensors with Current Transmitters A Simple Microcontroller-Based 4-20 ma Current Loop Receiver for Sensors with Current Transmitters A. Surachman, A. Suhendi, M. Budiman, M. Abdullah, and Khairurrijal *) Physics of Electronic Materials

More information

IDA 4 XM V 1.X. Installation and configuration of IDA 4 XM User Manual

IDA 4 XM V 1.X. Installation and configuration of IDA 4 XM User Manual IDA 4 XM V 1.X Installation and configuration of IDA 4 XM User Manual IMPORTANT SAFETY INSTRUCTIONS - Switch the device s power off before any maintenance operation (changing the CU card, etc.) - The 24V

More information

GSM BASED PATIENT MONITORING SYSTEM

GSM BASED PATIENT MONITORING SYSTEM GSM BASED PATIENT MONITORING SYSTEM ABSTRACT This project deals with the monitoring of the patient parameters such as humidity, temperature and heartbeat. Here we have designed a microcontroller based

More information

HEART & BREATH SOUNDS AMPLIFIER

HEART & BREATH SOUNDS AMPLIFIER HEART & BREATH SOUNDS AMPLIFIER BME 201 Department of Biomedical Engineering University of Wisconsin-Madison March 9, 2011 Drew Birrenkott, Team Leader Bradley Wendorff, Communicator Jared Ness, BWIG Caleb

More information

PHYSIOLOGICAL SIGNALS AND VEHICLE PARAMETERS MONITORING SYSTEM FOR EMERGENCY PATIENT TRANSPORTATION

PHYSIOLOGICAL SIGNALS AND VEHICLE PARAMETERS MONITORING SYSTEM FOR EMERGENCY PATIENT TRANSPORTATION PHYSIOLOGICAL SIGNALS AND VEHICLE PARAMETERS MONITORING SYSTEM FOR EMERGENCY PATIENT TRANSPORTATION Dhiraj Sunehra 1, Thirupathi Samudrala 2, K. Satyanarayana 3, M. Malini 4 1 JNTUH College of Engineering,

More information

CEEN Bot Lab Design A SENIOR THESIS PROPOSAL

CEEN Bot Lab Design A SENIOR THESIS PROPOSAL CEEN Bot Lab Design by Deborah Duran (EENG) Kenneth Townsend (EENG) A SENIOR THESIS PROPOSAL Presented to the Faculty of The Computer and Electronics Engineering Department In Partial Fulfillment of Requirements

More information

STM32 microcontroller core ECG acquisition Conditioning System. LIU Jia-ming, LI Zhi

STM32 microcontroller core ECG acquisition Conditioning System. LIU Jia-ming, LI Zhi International Conference on Computer and Information Technology Application (ICCITA 2016) STM32 microcontroller core ECG acquisition Conditioning System LIU Jia-ming, LI Zhi College of electronic information,

More information

A DESIGN OF PORTABLE HEART-RATE MONITORING SYSTEM

A DESIGN OF PORTABLE HEART-RATE MONITORING SYSTEM M.ENGİN et al. / IU-JEEE Vol. 10(2), (2010), 1201-1205 A DESIGN OF PORTABLE HEART-RATE MONITORING SYSTEM Mehmet ENGİN 1, Tayfun DALBASTI 2, Saygın BILDIK 1, Turan KARIPÇIN 1, Erkan Zeki ENGIN 1, Candan

More information

Portable, Low Cost, Low Power Cardiac Interpreter

Portable, Low Cost, Low Power Cardiac Interpreter Portable, Low Cost, Low Power Cardiac Interpreter Avishek Paul Department of Applied Electronics and Instrumentation Engineering RCC Institute of Information Technology, Kolkata, West Bengal, India Jahnavi

More information

3M Littmann Electronic Stethoscope Model 3000 with Ambient Noise Reduction

3M Littmann Electronic Stethoscope Model 3000 with Ambient Noise Reduction 3M Littmann Electronic Stethoscope Model 3000 with Ambient Noise Reduction Commonly Asked Questions Every time I switch on my Model 3000, it starts on a very loud (or very low) level, it was not so when

More information

DESIGN AND LAYOUT SHORT CIRCUIT DETECTT NETWORK ONE PHASE

DESIGN AND LAYOUT SHORT CIRCUIT DETECTT NETWORK ONE PHASE International Journal of Civil Engineering and Technology (IJCIET) Volume 9, Issue 13, December 2018, pp.899 909, Article ID: IJCIET_09_13_0900 Available online at http://www.ia aeme.com/ijciet/issues.asp?jtypeijciet&vtype

More information

EG medlab. Three Lead ECG OEM board. Version Technical Manual. Medlab GmbH Three Lead ECG OEM Module EG01010 User Manual

EG medlab. Three Lead ECG OEM board. Version Technical Manual. Medlab GmbH Three Lead ECG OEM Module EG01010 User Manual Medlab GmbH Three Lead ECG OEM Module EG01010 User Manual medlab Three Lead ECG OEM board EG01010 Technical Manual Copyright Medlab 2008-2016 Version 1.03 1 Version 1.03 28.04.2016 Medlab GmbH Three Lead

More information

Wireless Stethoscope for Anesthesiologists

Wireless Stethoscope for Anesthesiologists Wireless Stethoscope for Anesthesiologists Spring 2013 Client: Dr. Scott Springman Advisor: John Webster Team Leader: Charlie Rodenkirk Communicator: Alex Eaton BSAC: Lucas Haug BWIG: Yue Yin 1 Table of

More information

Deepali Shukla 1 (Asst.Professor), Vandana Pandya 2 (Asst.Professor) Medicaps Institute of Technology & Management, Indore (M.P.

Deepali Shukla 1 (Asst.Professor), Vandana Pandya 2 (Asst.Professor) Medicaps Institute of Technology & Management, Indore (M.P. Open Hardware Platform For Reconstruction Of ECG Signal Deepali Shukla 1 (Asst.Professor), Vandana Pandya 2 (Asst.Professor) Medicaps Institute of Technology & Management, Indore (M.P.), India Abstract

More information

Why a push-pull amplifier?

Why a push-pull amplifier? Unison Research P70 Fully balanced dual mono valves Amplifier in Push- Pull topology Why a push-pull amplifier? A push-pull topology amplifier is characterised by a power stage in which active devices

More information

Design and Implementation of AT Mega 328 microcontroller based firing control for a tri-phase thyristor control rectifier

Design and Implementation of AT Mega 328 microcontroller based firing control for a tri-phase thyristor control rectifier Design and Implementation of AT Mega 328 microcontroller based firing control for a tri-phase thyristor control rectifier 1 Mr. Gangul M.R PG Student WIT, Solapur 2 Mr. G.P Jain Assistant Professor WIT,

More information

Control System and Monitoring of ph, Temperature and Water Levels for a Hatchery Pond of Shrimps

Control System and Monitoring of ph, Temperature and Water Levels for a Hatchery Pond of Shrimps International Workshop on Modern Research Methods in Electrical Engineering (IWORMEE 2013) 35 Control System and Monitoring of ph, Temperature and Water Levels for a Hatchery Pond of Shrimps Alimuddin

More information

International Journal of Advances in Science and Technology (IJAST)

International Journal of Advances in Science and Technology (IJAST) Signal detection and FFT calculation using ATmega644 microcontroller D. Sarkar 1, A.Chowdhury 2 1,2 Department of Electronics & Communication Engineering, NIT Agartala, India ABSTRACT: Detection of a signal

More information

Digital Signal Processing of Speech for the Hearing Impaired

Digital Signal Processing of Speech for the Hearing Impaired Digital Signal Processing of Speech for the Hearing Impaired N. Magotra, F. Livingston, S. Savadatti, S. Kamath Texas Instruments Incorporated 12203 Southwest Freeway Stafford TX 77477 Abstract This paper

More information

Colour Detector Tool Using TCS3200 and Arduino Uno for Blind and Child

Colour Detector Tool Using TCS3200 and Arduino Uno for Blind and Child Journal of Telematics and Informatics (JTI) Vol.6, No.1, March 2018, pp. 37~44 ISSN: 2303-3703 37 Colour Detector Tool Using TCS3200 and Arduino Uno for Blind and Child Mohammad Habib Hasan, Arief Marwanto,

More information

Microphones & Accessories

Microphones & Accessories Accessories of the KLIPPEL R&D and QC SYSTEMs (Document Revision1.15) FEATURES Microphones optimal for research, development and manufacturing of transducers Condenser and electret microphones High performance/cost

More information

Heart & Breath Sound Amplifier

Heart & Breath Sound Amplifier Heart & Breath Sound Amplifier Drew Birrenkott, Caleb Durante, Jared Ness, Bradley Wendorff University of Wisconsin Madison, BME 201 ----------------------------------------- Scott Springman, M.D. Client

More information

MICROPROCESSOR BASED CONTROLLERS

MICROPROCESSOR BASED CONTROLLERS MICROPROCESSOR BASED CONTROLLERS INPUTS Digital Analog TTL Pulse Keyboard Serial Microprocessor Based Controller OUTPUTS On/Off Analog PWM Serial Graphical Text RS232 Abstract: A controller is a system

More information

Laboratory Activities Handbook

Laboratory Activities Handbook Laboratory Activities Handbook Answer Key 0 P a g e Contents Introduction... 2 Optical Heart Rate Monitor Overview... 2 Bare Board Preparation... 3 Light Indicator... 5 Low Pass Filter... 7 Amplifier...

More information

BME 405 BIOMEDICAL ENGINEERING SENIOR DESIGN 1 Fall 2005 BME Design Mini-Project Project Title

BME 405 BIOMEDICAL ENGINEERING SENIOR DESIGN 1 Fall 2005 BME Design Mini-Project Project Title BME 405 BIOMEDICAL ENGINEERING SENIOR DESIGN 1 Fall 2005 BME Design Mini-Project Project Title Basic system for Electrocardiography Customer/Clinical need A recent health care analysis have demonstrated

More information

TELIKOU Intercom System

TELIKOU Intercom System TELIKOU Intercom System DT-100 Desktop Station Instruction Manual 2006 TELIKOU Systems All Rights Reserved www.telikou.com While TELIKOU makes every attempt to maintain the accuracy of the information

More information

Massachusetts Institute of Technology MIT

Massachusetts Institute of Technology MIT Massachusetts Institute of Technology MIT Real Time Wireless Electrocardiogram (ECG) Monitoring System Introductory Analog Electronics Laboratory Guilherme K. Kolotelo, Rogers G. Reichert Cambridge, MA

More information

TM-800 Main Station. Instruction Manual. TELIKOU Systems All Rights Reserved

TM-800 Main Station. Instruction Manual. TELIKOU Systems All Rights Reserved Intercom System TM-800 Main Station Instruction Manual TELIKOU Systems All Rights Reserved I. Introduction Thank you for choosing TELIKOU intercom product. TM-800 main station is suitable for television

More information

Programming and Interfacing

Programming and Interfacing AtmelAVR Microcontroller Primer: Programming and Interfacing Second Edition f^r**t>*-**n*c contents Preface xv AtmelAVRArchitecture Overview 1 1.1 ATmegal64 Architecture Overview 1 1.1.1 Reduced Instruction

More information

MCU-Based Multi Parameter Patient Monitor

MCU-Based Multi Parameter Patient Monitor International Journal of Sciences: Basic and Applied Research (IJSBAR) ISSN 2307-4531 (Print & Online) http://gssrr.org/index.php?journal=journalofbasicandapplied ---------------------------------------------------------------------------------------------------------------------------

More information

List of Items Available in the Laboratory the Lab

List of Items Available in the Laboratory the Lab List of Items Available in the Laboratory the Lab Category Component 555 Timer $0.30 5V Relay $3.50 74xxx Series IC Chip $0.30 Battery - 12V (rechargeable Lead-acid type) $16.00 Battery - 6V (rechargeable

More information

A Super trainer with advanced hardware and software features only found in very expensive equipment.

A Super trainer with advanced hardware and software features only found in very expensive equipment. PLC Trainer PTS T100 LAB EXPERIMENTS A Super trainer with advanced hardware and software features only found in very expensive equipment. You won t find any similar equipment among our competitors at such

More information

Wireless Cardiac Rhythm Monitoring System

Wireless Cardiac Rhythm Monitoring System Wireless Cardiac Rhythm Monitoring System Darshana Dineshkumar Darji #1, Surbhi Prajapati *2, Prof. Neelam Modi #3 # Biomedical Engineering, Government Engineering College, Sector-28, Gandhinagar 1 darshana20994@gmail.com

More information

Development of a MATLAB Data Acquisition and Control Toolbox for BASIC Stamp Microcontrollers

Development of a MATLAB Data Acquisition and Control Toolbox for BASIC Stamp Microcontrollers Chapter 4 Development of a MATLAB Data Acquisition and Control Toolbox for BASIC Stamp Microcontrollers 4.1. Introduction Data acquisition and control boards, also known as DAC boards, are used in virtually

More information

Florida Atlantic University Biomedical Signal Processing Lab Experiment 2 Signal Transduction: Building an analog Electrocardiogram (ECG)

Florida Atlantic University Biomedical Signal Processing Lab Experiment 2 Signal Transduction: Building an analog Electrocardiogram (ECG) Florida Atlantic University Biomedical Signal Processing Lab Experiment 2 Signal Transduction: Building an analog Electrocardiogram (ECG) 1. Introduction: The Electrocardiogram (ECG) is a technique of

More information

WAVELET-BASED ADAPTIVE DENOISING OF PHONOCARDIOGRAPHIC RECORDS P. Várady 1 1 Department of Control Engineering and Information Technology

WAVELET-BASED ADAPTIVE DENOISING OF PHONOCARDIOGRAPHIC RECORDS P. Várady 1 1 Department of Control Engineering and Information Technology Proceedings 3rd Annual Conference IEEE/EMBS Oct.5-8, 00, Istanbul, TURKEY WAVELET-BASED ADAPTIVE DENOISING OF PHONOCARDIOGRAPHIC RECORDS P. Várady Department of Control Engineering and Information Technology

More information

Low-cost photoplethysmograph solutions using the Raspberry Pi

Low-cost photoplethysmograph solutions using the Raspberry Pi Low-cost photoplethysmograph solutions using the Raspberry Pi Tamás Nagy *, Zoltan Gingl * * Department of Technical Informatics, University of Szeged, Hungary nag.tams@gmail.com, gingl@inf.u-szeged.hu

More information

3 AXIS RF ELECTROMAGNETIC FIELD METER

3 AXIS RF ELECTROMAGNETIC FIELD METER 50 MHz to 3 GHz Radio Frequency Radiation Meters Electromagnetic Field strength measurement 3 AXIS RF ELECTROMAGNETIC FIELD METER Model : EMF-819 Your purchase of this RF EMF METER marks a step forward

More information

DECENTRALIZED INTERCOMMUNICATION CONCEPT

DECENTRALIZED INTERCOMMUNICATION CONCEPT SATEC 8.10 RUE JB SEMANAZ 93310 LE PRE SAINT GERVAIS 01.48.45.63.62 Fax : 01.48.45.02.11 Email : satec@satec-electronique.fr DECENTRALIZED INTERCOMMUNICATION CONCEPT V30/06/2006 Printed 30/06/2006 I. GENERAL

More information

LT-800 Stationary Transmitter

LT-800 Stationary Transmitter LT-800 Stationary Transmitter Configuration Stationary FM Transmitter (72 MHz) Stationary FM Transmitter (216 MHz) Stationary FM Transmitter (863 MHz) Thanks to its outstanding audio quality, the Listen

More information

common type of cardiac diseases and may indicate an increased risk of stroke or sudden cardiac death. ECG is the most

common type of cardiac diseases and may indicate an increased risk of stroke or sudden cardiac death. ECG is the most ISSN: 0975-766X CODEN: IJPTFI Available Online through Research Article www.ijptonline.com DESIGNING OF ELECTRONIC CARDIAC EVENTS RECORDER *Dr. R. Jagannathan, K.Venkatraman, R. Vasuki and Sundaresan Department

More information

DEC-001 Installation Instructions

DEC-001 Installation Instructions DEC-001 Installation Instructions Skill Level: The installation of this assembly requires a medium level of expertise in working with modern electronic equipment. The use of appropriate tools, correct

More information

HUMAN BODY MONITORING SYSTEM USING WSN WITH GSM AND GPS

HUMAN BODY MONITORING SYSTEM USING WSN WITH GSM AND GPS HUMAN BODY MONITORING SYSTEM USING WSN WITH GSM AND GPS Mr. Sunil L. Rahane Department of E & TC Amrutvahini College of Engineering Sangmaner, India Prof. Ramesh S. Pawase Department of E & TC Amrutvahini

More information

Digital-to-Analog Converter. Lab 3 Final Report

Digital-to-Analog Converter. Lab 3 Final Report Digital-to-Analog Converter Lab 3 Final Report The Ion Cannons: Shrinand Aggarwal Cameron Francis Nicholas Polito Section 2 May 1, 2017 1 Table of Contents Introduction..3 Rationale..3 Theory of Operation.3

More information

The report presents the functionality of our project, the problems we encountered, the incurred costs and timeline for the project development.

The report presents the functionality of our project, the problems we encountered, the incurred costs and timeline for the project development. April 30, 2010 Dr. Andrew Rawicz School of Engineering Science Simon Fraser University Burnaby, BC V5A 1S6 Re: ENSC 440 Post Mortem for Biomedical Monitoring System Dear Dr. Rawicz: Please see attached

More information

Design of Virtual Sphygmomanometer Based on LABVIEWComparison, Reflection, Biological assets, Accounting standard.

Design of Virtual Sphygmomanometer Based on LABVIEWComparison, Reflection, Biological assets, Accounting standard. Design of Virtual Sphygmomanometer Based on LABVIEWComparison, Reflection, Biological assets, Accounting standard. Li Su a, Boxin Zhang b School of electronic engineering, Xi'an Aeronautical University,

More information

REAL-TIME WIRELESS ECG AND ITS SIGNAL DISPLAY ON LABVIEW

REAL-TIME WIRELESS ECG AND ITS SIGNAL DISPLAY ON LABVIEW REAL-TIME WIRELESS ECG AND ITS SIGNAL DISPLAY ON LABVIEW 1 POOJA AIYAPPA K, 2 SEETHAMMA M.G, 3 BHAUSHI AIYAPPA C 1,2 Dept. of ECE,CIT, Ponnampet, Karnataka, 3 Assistant Professor, Dept. of ECE, CIT, Ponnampet,

More information

Week 8 AM Modulation and the AM Receiver

Week 8 AM Modulation and the AM Receiver Week 8 AM Modulation and the AM Receiver The concept of modulation and radio transmission is introduced. An AM receiver is studied and the constructed on the prototyping board. The operation of the AM

More information

Initial Project Document Vacuum Tube Stereo Amplifier

Initial Project Document Vacuum Tube Stereo Amplifier Initial Project Document Vacuum Tube Stereo Amplifier By Stephen Nichols, Jason Lambert, Rafael Enriquez Group 4 January 30, 2013 1 Narrative Description Motivation The motivation for this idea is nostalgia.

More information

DESIGN OF A DEVICE FOR CHECKING THE CONTINUITY IN ELECTRICAL CIRCUIT

DESIGN OF A DEVICE FOR CHECKING THE CONTINUITY IN ELECTRICAL CIRCUIT DESIGN OF A DEVICE FOR CHECKING THE CONTINUITY IN ELECTRICAL CIRCUIT FA IZAH BINTI YA ACOB POLITEKNIK SULTAN SALAHUDDIN ABDUL AZIZ SHAH (yaacob_faiza@yahoo.com ) MASLIZAH BINTI MUNAHDAR POLITEKNIK SULTAN

More information

Application Note 160 Using the DS1808 in Audio Applications

Application Note 160 Using the DS1808 in Audio Applications www.maxim-ic.com Application Note 160 Using the DS1808 in Audio Applications Introduction The DS1808 Dual Log Audio Potentiometer was designed to provide superior audio performance in applications that

More information

DESIGN A MEDICINE DEVICE FOR BLOOD OXYGEN CONCENTRATION AND HEART BEAT RATE

DESIGN A MEDICINE DEVICE FOR BLOOD OXYGEN CONCENTRATION AND HEART BEAT RATE Transaction in Healthcare and Biomedical Signal Processing ISSN: 1985-9406 Online Publication, June 2010 www.pcoglobal.com/gjto.htm HS-T13/GJTO DESIGN A MEDICINE DEVICE FOR BLOOD OXYGEN CONCENTRATION AND

More information

ENGR 499: Wireless ECG

ENGR 499: Wireless ECG ENGR 499: Wireless ECG Introduction and Project History Michael Atkinson Patrick Cousineau James Hollinger Chris Rennie Brian Richter Our 499 project is to design and build the hardware and software for

More information

MUSIC RESPONSIVE LIGHT SYSTEM

MUSIC RESPONSIVE LIGHT SYSTEM MUSIC RESPONSIVE LIGHT SYSTEM By Andrew John Groesch Final Report for ECE 445, Senior Design, Spring 2013 TA: Lydia Majure 1 May 2013 Project 49 Abstract The system takes in a musical signal as an acoustic

More information

Programming Parameter Guide

Programming Parameter Guide Secure Wireless Microphone ELITE PRO Programming Parameter Guide rev:1 How to use Programmer: Start Programming application Runs On PC or Mac running Windows 7/10. To put Handset into programming mode,

More information

BS 17 SINGLE CHANNEL BELTPACK. User Manual. January 2017 V1.0

BS 17 SINGLE CHANNEL BELTPACK. User Manual. January 2017 V1.0 BS 17 SINGLE CHANNEL BELTPACK User Manual January 2017 V1.0 Table of contents 1.0 GENERAL DESCRIPTION... 3 2.0 INSTALLATION... 3 3.0 TOP PANEL CONTROLS... 4 4.0 SIDE PANEL CONTROLS... 4 5.0 BOTTOM PANEL

More information

Manuel Campus S FM Transmitter (TX2S)

Manuel Campus S FM Transmitter (TX2S) 1 of 5 Distribution: Montena EMC SA, CH-1728 Rossens, Mr Manfred Portman Path: Marquis:> D:/data/phonak/T_memos/Englisch/ Name Date Visum Produced by F. Marquis FMa Approved F. Marquis FMa Released F.

More information

Sampling and Reconstruction

Sampling and Reconstruction Experiment 10 Sampling and Reconstruction In this experiment we shall learn how an analog signal can be sampled in the time domain and then how the same samples can be used to reconstruct the original

More information

Lynx. RoIP Gateway DISPATCH LYNX MOBILE. Optional serial ports provide remote control of radio configuration over the VoIP network.

Lynx. RoIP Gateway DISPATCH LYNX MOBILE. Optional serial ports provide remote control of radio configuration over the VoIP network. LYNX MOBILE Lynx DISPATCH RoIP TM SYSTEM Gateway LYNX TM ROIP GATEWAY OVERVIEW The Lynx system provides Radio over IP (RoIP) communications interoperability between radio base stations and VoIP networked

More information

MONITORING PHYSIOLOGICAL SIGNALS DURING RUNNING EXERCISE

MONITORING PHYSIOLOGICAL SIGNALS DURING RUNNING EXERCISE MONITORING PHYSIOLOGICAL SIGNALS DURING RUNNING EXERCISE Daoming Zhang and Branko Celler Biomedical Systems Laboratory, School of Electrical Engineering and Telecommunications University of New South Wales,

More information

Design of a prototype to detect mobile phone usage in restricted areas

Design of a prototype to detect mobile phone usage in restricted areas Design of a prototype to detect mobile phone usage in restricted areas Shruthi.K #1, Ramaprasad.P *2 # Assistant Professor Senior Scale, Dept. of E&C, Manipal Institute of Technology, Manipal University,

More information

Honeywell. W7620 Zone Controller Water Source Heat Pump A. GENERAL RESIDENTIAL AND BUILDING CONTROL GUIDE SPECIFICATION FOR THE ENGINEER

Honeywell. W7620 Zone Controller Water Source Heat Pump A. GENERAL RESIDENTIAL AND BUILDING CONTROL GUIDE SPECIFICATION FOR THE ENGINEER Honeywell RESIDENTIAL AND BUILDING CONTROL GUIDE SPECIFICATION FOR THE ENGINEER W7620 Zone Controller Water Source Heat Pump - -.> The W7620 Zone Controller provides customized control of HVAC zone equipment.

More information

Monitoring Temperature using LM35 and Arduino UNO

Monitoring Temperature using LM35 and Arduino UNO Sharif University of Technology Microprocessor Arduino UNO Project Monitoring Temperature using LM35 and Arduino UNO Authors: Sadegh Saberian 92106226 Armin Vakil 92110419 Ainaz Hajimoradlou 92106142 Supervisor:

More information

Associate In Applied Science In Electronics Engineering Technology Expiration Date:

Associate In Applied Science In Electronics Engineering Technology Expiration Date: PROGRESS RECORD Study your lessons in the order listed below. Associate In Applied Science In Electronics Engineering Technology Expiration Date: 1 2330A Current and Voltage 2 2330B Controlling Current

More information

Programming Parameters - Feature Comparison : Rev 1

Programming Parameters - Feature Comparison : Rev 1 Secure Wireless Microphone ELITE PRO CLASSIC Programming Parameters - Feature Comparison : Rev 1 How to use Programmer: Start Programming application Runs On PC or Mac running Windows 7/8. To put Handset

More information

Hearing Aid Redesign: Test Plans ELECTRICAL TESTING

Hearing Aid Redesign: Test Plans ELECTRICAL TESTING ELECTRICAL TESTING Table of Contents: Number Test Page EE 1 Switch 2 EE 2 Speaker 7 EE 3 Sound Processing 11 EE 4 Microphone 14 EE 5 Battery Charger 18 EE 6 Bandpass and Pre-Amplification 20 EE 7 System

More information

AC : PERSONAL LAB HARDWARE: A SINE WAVE GENERATOR, LOGIC PULSE SIGNAL, AND PROGRAMMABLE SYNCHRONOUS SERIAL INTERFACE FOR ENHANCING EDUCATION

AC : PERSONAL LAB HARDWARE: A SINE WAVE GENERATOR, LOGIC PULSE SIGNAL, AND PROGRAMMABLE SYNCHRONOUS SERIAL INTERFACE FOR ENHANCING EDUCATION AC 2010-1527: PERSONAL LAB HARDWARE: A SINE WAVE GENERATOR, LOGIC PULSE SIGNAL, AND PROGRAMMABLE SYNCHRONOUS SERIAL INTERFACE FOR ENHANCING EDUCATION Jeffrey Richardson, Purdue University James Jacob,

More information

EFFECTS OF PHYSICAL CONFIGURATIONS ON ANC HEADPHONE PERFORMANCE

EFFECTS OF PHYSICAL CONFIGURATIONS ON ANC HEADPHONE PERFORMANCE EFFECTS OF PHYSICAL CONFIGURATIONS ON ANC HEADPHONE PERFORMANCE Lifu Wu Nanjing University of Information Science and Technology, School of Electronic & Information Engineering, CICAEET, Nanjing, 210044,

More information

EEL4914 Senior Design. Final Design Report

EEL4914 Senior Design. Final Design Report EEL4914 Senior Design Final Design Report Electric Super Bike The Best Team in the World Matt Fisher madfish@ufl.edu Richard Orr gautama@ufl.edu 21 April 2008 1 Contents Contents...2 Abstract...3 Project

More information

Laboratory Assignment 1 Sampling Phenomena

Laboratory Assignment 1 Sampling Phenomena 1 Main Topics Signal Acquisition Audio Processing Aliasing, Anti-Aliasing Filters Laboratory Assignment 1 Sampling Phenomena 2.171 Analysis and Design of Digital Control Systems Digital Filter Design and

More information

Model Microphone Preamplifier Power Supply Manual

Model Microphone Preamplifier Power Supply Manual Model 2221 Microphone Preamplifier Power Supply Manual Larson Davis Model 2221 Microphone Preamplifier Power Supply Manual I2221.01 Rev H Copyright Copyright 2009-2015 by PCB Piezotronics, Inc. This manual

More information

SonoLab Echo-I User Manual

SonoLab Echo-I User Manual SonoLab Echo-I User Manual Overview: SonoLab Echo-I is a single board digital ultrasound pulse-echo solution. The system has a built in 50 volt high voltage generation circuit, a bipolar pulser, a transmit/receive

More information

Pulse Doppler Flow-Dop

Pulse Doppler Flow-Dop EDUCATION Pulse Doppler Flow-Dop GAMPT-50100 User Manual Fon: +49 (0) 3461-278 691-0 Fax: +49 (0) 3461-278 691-101 email: info@gampt.de Gesellschaft für Angewandte Medizinische Physik und Technik mbh (GAMPT

More information

WS-29 DUAL CHANNEL WIRELESS BELTPACK

WS-29 DUAL CHANNEL WIRELESS BELTPACK WS-29 DUAL CHANNEL WIRELESS BELTPACK USER MANUAL Issue March 2011 ASL Intercom BV DESIGNED AND MANUFACTURED BY: ASL INTERCOM BV ZONNEBAAN 42 3542 EG UTRECHT THE NETHERLANDS PHONE: +31 (0)30 2411901 FAX:

More information

DR-TRC105-EV Evaluation Kit. User s Guide

DR-TRC105-EV Evaluation Kit. User s Guide DR-TRC105-EV Evaluation Kit User s Guide DR-TRC105-304-EV DR-TRC105-315-EV DR-TRC105-345-EV DR-TRC105-372-EV DR-TRC105-390-EV DR-TRC105-403-EV DR-TRC105-434-EV DR-TRC105-450-EV 2010-2015 by Murata Electronics

More information

To Measure Dielectric Constant of Liquid By Using PIC Microcontroller

To Measure Dielectric Constant of Liquid By Using PIC Microcontroller To Measure Dielectric Constant of Liquid By Using PIC Microcontroller Prof.P.K.Ghuge 1, Prof.N.R.Kolhare 2 1 Department of ETC, P.E.S.college of Engineering Aurangabad,Maharashtra,(India) 2 Department

More information

6.101 Introductory Analog Electronics Laboratory

6.101 Introductory Analog Electronics Laboratory 6.101 Introductory Analog Electronics Laboratory Spring 2015, Instructor Gim Hom Project Proposal Transmitting, Receiving, and Interpreting ECG Waveforms Daniel Moon (dhmoon@mit.edu) Thipok (Ben) Rak-amnouykit

More information

Control of Electrical Lights and Fans using TV Remote

Control of Electrical Lights and Fans using TV Remote EE 389 Electronic Design Lab -II, Project Report, EE Dept., IIT Bombay, October 2005 Control of Electrical Lights and Fans using TV Remote Group No. D10 Liji Jayaprakash (02d07021)

More information

JEPPIAAR SRR Engineering College Padur, Ch

JEPPIAAR SRR Engineering College Padur, Ch An Automated Non-Invasive Blood Glucose Estimator and Infiltrator M. Florence Silvia 1, K. Saran 2, G. Venkata Prasad 3, John Fermin 4 1 Asst. Prof, 2, 3, 4 Student, Department of Electronics and Communication

More information

LT-800 Stationary FM Transmitter

LT-800 Stationary FM Transmitter LT-800 Stationary FM Transmitter Thanks to its outstanding audio quality, the Listen LT-800 Stationary Transmitter can be used in a variety of applications. The LT-800 is connected to your main audio system,

More information

EE-241. Introductory Electronics Laboratory Project Ideas Fall 2009

EE-241. Introductory Electronics Laboratory Project Ideas Fall 2009 EE-241. Introductory Electronics Laboratory Project Ideas Fall 2009 EASY TO MODERATE 1. Musical notes display In this project students would build a display unit that will show high and low frequency sounds

More information

CHAPTER 12 NORTHERN ILLINOIS UNIVERSITY

CHAPTER 12 NORTHERN ILLINOIS UNIVERSITY CHAPTER 12 NORTHERN ILLINOIS UNIVERSITY Department of Electrical Engineering DeKalb, IL 60115 Principal Investigators: Mansour Tahernezhadi (815)-753-8568 Xuan Kong (815)-753-9942 127 128 NSF 1999 Engineering

More information

Design of LVDT Based Digital Weighing System

Design of LVDT Based Digital Weighing System International Journal of Electronics and Computer Science Engineering 2100 Available Online at www.ijecse.org ISSN- 2277-1956 Pratiksha Sarma 1, P. K. Bordoloi 2 1,2 Department of Applied Electronics and

More information

A Duty-Cycle Controlled Variable-Gain Instrumentation Amplifier Applied For Two-electrode ECG Measurement

A Duty-Cycle Controlled Variable-Gain Instrumentation Amplifier Applied For Two-electrode ECG Measurement A Duty-Cycle Controlled Variable-Gain Instrumentation Amplifier Applied For Two-electrode ECG Measurement R. Romero Antayhua, G. Manoel Da Silva, F. Rangel de Sousa Integrated Circuits Laboratory-LCI Federal

More information

ELECTRONIC STETHOSCOPE AND HEART RATE MONITOR A GRADUATION PROJECT SUBMITTED TO BIOMEDICAL ENGINEERING DEPARTMENT OF NEAR EAST UNIVERSITY

ELECTRONIC STETHOSCOPE AND HEART RATE MONITOR A GRADUATION PROJECT SUBMITTED TO BIOMEDICAL ENGINEERING DEPARTMENT OF NEAR EAST UNIVERSITY ELECTRONIC STETHOSCOPE AND HEART RATE MONITOR NICOSIA-2015 ELECTRONIC STETHOSCOPE AND HEART RATE MONITOR A GRADUATION PROJECT SUBMITTED TO BIOMEDICAL ENGINEERING DEPARTMENT OF NEAR EAST UNIVERSITY BY MOHAMAD

More information

Design of Low-Cost General Purpose Microcontroller Based Neuromuscular Stimulator

Design of Low-Cost General Purpose Microcontroller Based Neuromuscular Stimulator Journal of Medical Systems, Vol. 24, No. 2, 2000 Design of Low-Cost General Purpose Microcontroller Based Neuromuscular Stimulator Sabri Koçer, 1 M. Rahmi Canal, 1 and İnan Güler 1 In this study, a general

More information

FreeSpeak. Digital Clarity and Versatile Operations. For Unsurpassed Wireless Performance. FreeSpeak is not available in the U.S. and Canada.

FreeSpeak. Digital Clarity and Versatile Operations. For Unsurpassed Wireless Performance. FreeSpeak is not available in the U.S. and Canada. 1.88-1.93GHz Digital Wireless Intercom System FreeSpeak Digital Clarity and Versatile Operations For Unsurpassed Wireless Performance FreeSpeak is not available in the U.S. and Canada. About Clear-Com

More information

Design and study of frequency response of band pass and band reject filters using operational amplifiers

Design and study of frequency response of band pass and band reject filters using operational amplifiers International Journal of Advanced Educational Research ISSN: 2455-6157 Impact Factor: RJIF 5.12 www.educationjournal.org Volume 2; Issue 6; November 2017; Page No. 22-26 Design and study of frequency response

More information