LSC 2013 Advances in Liquid Scintillation Spectrometry A prototype of a portable TDCR system at ENEA

Size: px
Start display at page:

Download "LSC 2013 Advances in Liquid Scintillation Spectrometry A prototype of a portable TDCR system at ENEA"

Transcription

1 LSC 2013 Advances in Liquid Scintillation Spectrometry A prototype of a portable TDCR system at ENEA Marco Capogni, Pierino De Felice ENEA National Institute of Ionizing Radiation Metrology (INMRI) Casaccia Research Center Anguillarese street, 301 I Rome (Italy) marco.capogni@enea.it Centre de Cultura Contemporània de Barcelona March 2013

2 OUTLINE TDCR technique and the new ENEA TDCR portable instrument for radioactivity measurements in-situ DPP technology in the field of precise measurements of the radioactivity with Liquid Scintillators TDCR analysis within CERN ROOT framework Application to the 63 Ni and 99 Tc activity measurements Conclusions

3 TDCR technique in Radionuclide Metrology at ENEA TDCR technique was introduced at ENEA-INMRI in February 2011 by the Hidex 300 SL TDCR counter Metro version and used for the first time in the standardisation of a 63 Ni source in the framework of the ESIR comparison organised under the auspices of the BIPM An ENEA portable TDCR system working with the new front-end electronics based on the Digital Pulse Processing (DPP) implemented in a Fast Programmable Gate Array (FPGA) device was built in the framework of the EMRP ENG08 Metrofission project

4 Metrofission: ENEA Research of components and Materials (PMTs, optical chamber, etc) Portability Fast Front End Electronics Efficient and sensitive PMTs Compact counter and user interface WP6 & WP7 task: Portable TDCR-LSC for in-situ measurements Digitizer CAEN DT5720/B Experimental Test Development of the analysis and simulation SW based on CERN ROOT + GEANT4 Measurements between different LSC counters of a same radioactive solution and comparison with the results of the portable TDCR

5 PMTs for the ENEA portable TDCR HAMAMATSU R7600U Series Small dimensions (30x30mm) Bialkali photocathode Spectral range ( nm) Quantum Efficiency (350 nm) 43% Dynods MC/10 HV 900 V Gain Transit Time 9,6 ns High quantum efficiency Relative low voltage 900 V Cathode to the Background for Photon Counting mode 8 th dynode output for spectroscopy

6 ENEA Portable TDCR system Optical chamber made in PTFE with prismatic shape Lift for the LS vial and optical shutter (PMT s always powered) Compact design (HV modules included in the box)

7 HV Power Supply 12 V battery The C PMT HV power supply units are on-board type with a design that aims at providing compactness, high performance and low power consumption. The three HV power supply units are powered by a 12 V battery. Maximum output voltage is V and can be regulated by a potentiometer Dimensions of a single unit: 46x24x12 mm 3 Weight: 31g.

8 Electronics: Digital Approach Digital approach based on CAEN Desktop Digitizer DT5720/B 4 acquisition channels Dimensions: 154 x 50 x 164 mm 3 (W x H x D) Weight: 680 g A/D: 12 bits 250 MS/s ADCs DPP: FPGA based for digital Charge Integration/PSD Upgradeable via USB Link INTERF: Control Software for Windows and Linux Readout USB (30 MB/s) Electrical Power: 12 +/- 10% Vdc

9 Electronics: DPP Firmware Charge Integration (CI) (DT5720) Charge ( 2 Byte) List Mode File TTT Charge Trigger Time Tag ( 4 Byte ; 4 ns res.)

10 The Single Electron Peak Histogram mode channels a.u. Waveform in oscilloscope mode ns

11 CAEN DAQ SW The DAQ SW developed by CAEN after preliminary contact with ENEA to use the Digitizer with TDCR It has been written following the main idea of the paper Bouchard J., Cassette P. ARI 52 (2000) The DAQ SW looks for coincidences comparing the TTTs and makes, like a black-box, the TDCR analysis

12 ENEA TDCR Analysis in ROOT By following the same MAC3 philosophy a script in C++ was developed at ENEA to analyze in the CERN ROOT framework ( the data recorded by the Desktop CAEN digitizer DT5720 The script makes the TDCR analysis, saves the results in an ASCII file and produces a root tree

13 Energy spectrum The tree contains the data recorded events by events Raw data (TTT and Charge) Energy spectrum of Triple Coincidences channels Asymmetry between the 3 channels for coincident events ea0t=(en0t-en1t); ea1t=(en0t-en2t); ea2t=(en1t-en2t);

14 Time Interval Distribution Original events (non dead time imposed) ms Events with a dead time imposed of 50 ms ms By two other trees produced by the same script there is the possibility to plot the time interval distribution for all the 3 TDCR channels

15 63 Ni measurements Source prepared in glass vial filled with UG (10 ml) in our Radiochemistry Lab (Maria Letizia Cozzella) 6 sources with mass of about 10 mg and different aliquot of CCl 4 quenching agent for C-N g-purity check of the master solution performed by HpGE detector (Aldo Fazio) only sources without any quench used for TDCR measurements Dead Time imposed on the data of portable TDCR: 50 ms Coincidence resolving window = 40 ns Hidex kb = cm MeV -1 TDCR parameters: 0.62 (portable TDCR) ; 0.80 (Hidex) portable TDCR C-N Data analyzed by TDCR07c.for code provide by CEA-LNHB (P. Cassette) Method

16 99 Tc measurements performed within an international comparison organised under the auspices of BIPM Source prepared in glass vial filled with UG (10 ml) in our Radiochemistry Lab (Maria Letizia Cozzella) 12 sources with mass raging between 10 and 80 mg; CCl 4 added in different aliquot as quenching agent for C-N measurements g-purity check of the master solution performed by HpGE detector (Aldo Fazio) only sources without any quench used for TDCR measurements Dead Time imposed on the data of portable TDCR: 50 ms Coincidence resolving window = 40 ns TDCR parameters: = (portable TDCR) = (Hidex) Data analyzed by TDCR07c.for code provided by CEA-LNHB (P. Cassette) Hidex kb = cm MeV -1 portable TDCR C-N

17 99 Tc Budget of uncertainties for portable TDCR Quantity % Counting statistics 0,250 Weighing 0,05 background 0,004 Dead time 0,1 Decay data 0.05 Half-life 0,01 Impurities 0,07 Absorption 0,02 TDCR value 0,02 PMT asymmetry 0,2 Ionization quench 0,2 Scintillator stability 0,1 Resolving time 0,1 Combined standard uncertainty 0.43

18 Comparison between the results obtained by the two SW codes (CAEN DAQ & ENEA ROOT) D (%)=100*(1-a_ENEA/a_CAEN) Nuclide D % 63 Ni 0,06 99 Tc -0,37 It is important to check the new digitizer technology with the MAC3 in reference conditions

19 Discussion and Perspectives A mini-tdcr prototype for measurements in-situ was built at ENEA- INMRI in the framework of the EMRP JRP Metrofission 63 Ni and 99 Tc measurements were carried out by this prototype and compared with other LSC counters available at ENEA-IMRI based both on the 2PMTs technique and the TDCR technique (Hidex system Metro version) Two SW codes (CAEN DAQ and ENEA ROOT) for TDCR analysis of the data recorded by a CAEN Digitizer DT5720 are available at ENEA-INMRI New investigation on 3 H and other nuclides ( 14 C, 90 Y,..) are still in progress - a simulator based on GEANT4 is also in construction Pulse shape discrimination and g-channel can be included thanks to the digitizer technology

20 Acknowledgments We re very grateful to Mr Massimo Pagliari (ENEA-INMRI) and Mr Sergio Mancini (ENEA) for their mechanical and electronics competences and for the assistance given during this work Thanks to Stefano Loreti (ENEA-INMRI) for some useful discussions on the CAEN digitizer Many students (Eng Dr Carmine Biondi, Dr Valerio Di Sabatino from University La Sapienza Rome, Mr Paulius Butkus from Vilnius University) who collaborated and are collaborating directly with me in this adventure..

21 Thank very much to all of you for your attention

22

INDEX. Firmware for DPP (Digital Pulse Processing) DPP-PSD Digital Pulse Processing for Pulse Shape Discrimination

INDEX. Firmware for DPP (Digital Pulse Processing) DPP-PSD Digital Pulse Processing for Pulse Shape Discrimination Firmware for DPP (Digital Pulse Processing) Thanks to the powerful FPGAs available nowadays, it is possible to implement Digital Pulse Processing (DPP) algorithms directly on the acquisition boards and

More information

Traditional analog QDC chain and Digital Pulse Processing [1]

Traditional analog QDC chain and Digital Pulse Processing [1] Giuliano Mini Viareggio April 22, 2010 Introduction The aim of this paper is to compare the energy resolution of two gamma ray spectroscopy setups based on two different acquisition chains; the first chain

More information

CAEN. Electronic Instrumentation. CAEN Silicon Photomultiplier Kit

CAEN. Electronic Instrumentation. CAEN Silicon Photomultiplier Kit CAEN Tools for Discovery Electronic Instrumentation CAEN Silicon Photomultiplier Kit CAEN realized a modular development kit dedicated to Silicon Photomultipliers, representing the state-of-the art in

More information

Digital coincidence acquisition applied to portable β liquid scintillation counting device

Digital coincidence acquisition applied to portable β liquid scintillation counting device Nuclear Science and Techniques 24 (2013) 030401 Digital coincidence acquisition applied to portable β liquid scintillation counting device REN Zhongguo 1,2 HU Bitao 1 ZHAO Zhiping 2 LI Dongcang 1,* 1 School

More information

With a little bit of effort, manufacturers could make better LS counters, even suitable for radionuclide metrology!

With a little bit of effort, manufacturers could make better LS counters, even suitable for radionuclide metrology! With a little bit of effort, manufacturers could make better LS counters, even suitable for radionuclide metrology! Background During an international specialists meeting on LS techniques in radionuclide

More information

MuLan Experiment Progress Report

MuLan Experiment Progress Report BV 37 PSI February 16 2006 p. 1 MuLan Experiment Progress Report PSI Experiment R 99-07 Françoise Mulhauser, University of Illinois at Urbana Champaign (USA) The MuLan Collaboration: BERKELEY BOSTON ILLINOIS

More information

4πβ (LS)-γ (HPGe) Digital Coincidence System Based on Synchronous. High-Speed Multichannel Data Acquisition *

4πβ (LS)-γ (HPGe) Digital Coincidence System Based on Synchronous. High-Speed Multichannel Data Acquisition * 4πβ (LS)-γ (HPGe) Digital Coincidence System Based on Synchronous High-Speed Multichannel Data Acquisition * Jifeng Chen( 陈吉锋 ) 1 Kezhu Song( 宋克柱 ) 1 Juncheng Liang( 梁珺成 ) 2 Jiacheng Liu( 柳加成 ) 3 1 State

More information

Gamma Ray Spectroscopy with NaI(Tl) and HPGe Detectors

Gamma Ray Spectroscopy with NaI(Tl) and HPGe Detectors Nuclear Physics #1 Gamma Ray Spectroscopy with NaI(Tl) and HPGe Detectors Introduction: In this experiment you will use both scintillation and semiconductor detectors to study γ- ray energy spectra. The

More information

LaBr 3 :Ce scintillation gamma camera prototype for X and gamma ray imaging

LaBr 3 :Ce scintillation gamma camera prototype for X and gamma ray imaging 8th International Workshop on Radiation Imaging Detectors Pisa 2-6 July 2006 LaBr 3 :Ce scintillation gamma camera prototype for X and gamma ray imaging Roberto Pani On behalf of SCINTIRAD Collaboration

More information

EXPERIMENT 5. SCINTILLATION COUNTING AND QUENCH CORRECTION.

EXPERIMENT 5. SCINTILLATION COUNTING AND QUENCH CORRECTION. 59 EXPERIMENT 5. SCINTILLATION COUNTING AND QUENCH CORRECTION. (The report for this experiment is due 1 week after the completion of the experiment) 5.1 Introduction Liquid scintillation is the method

More information

Digital trigger system for the RED-100 detector based on the unit in VME standard

Digital trigger system for the RED-100 detector based on the unit in VME standard Journal of Physics: Conference Series PAPER OPEN ACCESS Digital trigger system for the RED-100 detector based on the unit in VME standard To cite this article: D Yu Akimov et al 2016 J. Phys.: Conf. Ser.

More information

nanodpp datasheet I. FEATURES

nanodpp datasheet I. FEATURES datasheet nanodpp I. FEATURES Ultra small size high-performance Digital Pulse Processor (DPP). 16k channels utilizing smart spectrum-size technology -- all spectra are recorded and stored as 16k spectra

More information

CAEN Tools for Discovery

CAEN Tools for Discovery Viareggio 5 September 211 Introduction In recent years CAEN has developed a complete family of digitizers that consists of several models differing in sampling frequency, resolution, form factor and other

More information

event physics experiments

event physics experiments Comparison between large area PMTs at cryogenic temperature for neutrino and rare Andrea Falcone University of Pavia INFN Pavia event physics experiments Rare event physics experiment Various detectors

More information

Advancement in development of photomultipliers dedicated to new scintillators studies.

Advancement in development of photomultipliers dedicated to new scintillators studies. Advancement in development of photomultipliers dedicated to new scintillators studies. Maciej Kapusta, Pascal Lavoutea, Florence Lherbet, Cyril Moussant, Paul Hink INTRODUCTION AND OUTLINE In the validation

More information

LaBr 3 :Ce, the latest crystal for nuclear medicine

LaBr 3 :Ce, the latest crystal for nuclear medicine 10th Topical Seminar on Innovative Particle and Radiation Detectors 1-5 October 2006 Siena, Italy LaBr 3 :Ce, the latest crystal for nuclear medicine Roberto Pani On behalf of SCINTIRAD Collaboration INFN

More information

Digitization of PMT signals with FADCs: comparison of simulation and measurement

Digitization of PMT signals with FADCs: comparison of simulation and measurement Digitization of PMT signals with FADCs: comparison of simulation and measurement Arno Gadola General, 10. 12.05.2010 Outline Summary of previous presentations Impact of sampling rate Verification of simulation

More information

nanomca-sp datasheet I. FEATURES

nanomca-sp datasheet I. FEATURES datasheet nanomca-sp 80 MHz HIGH PERFORMANCE, LOW POWER DIGITAL MCA WITH BUILT IN PREAMPLIFIER Model Numbers: SP0534A/B to SP0539A/B Standard Models: SP0536B and SP0536A I. FEATURES Built-in preamplifier

More information

nanomca 80 MHz HIGH PERFORMANCE, LOW POWER DIGITAL MCA Model Numbers: NM0530 and NM0530Z

nanomca 80 MHz HIGH PERFORMANCE, LOW POWER DIGITAL MCA Model Numbers: NM0530 and NM0530Z datasheet nanomca 80 MHz HIGH PERFORMANCE, LOW POWER DIGITAL MCA Model Numbers: NM0530 and NM0530Z I. FEATURES Finger-sized, high performance digital MCA. 16k channels utilizing smart spectrum-size technology

More information

nanomca datasheet I. FEATURES

nanomca datasheet I. FEATURES datasheet nanomca I. FEATURES Finger-sized, high performance digital MCA. 16k channels utilizing smart spectrum-size technology -- all spectra are recorded and stored as 16k spectra with instant, distortion-free

More information

Chemistry 985. Some constants: q e 1.602x10 19 Coul, ɛ x10 12 F/m h 6.626x10 34 J-s, c m/s, 1 atm = 760 Torr = 101,325 Pa

Chemistry 985. Some constants: q e 1.602x10 19 Coul, ɛ x10 12 F/m h 6.626x10 34 J-s, c m/s, 1 atm = 760 Torr = 101,325 Pa Chemistry 985 Fall, 2o17 Distributed: Mon., 17 Oct. 17, 8:30AM Exam # 1 OPEN BOOK Due: 17 Oct. 17, 10:00AM Some constants: q e 1.602x10 19 Coul, ɛ 0 8.854x10 12 F/m h 6.626x10 34 J-s, c 299 792 458 m/s,

More information

Cosmic Ray Detector Hardware

Cosmic Ray Detector Hardware Cosmic Ray Detector Hardware How it detects cosmic rays, what it measures and how to use it Matthew Jones Purdue University 2012 QuarkNet Summer Workshop 1 What are Cosmic Rays? Mostly muons down here

More information

SCINTILLATION FIBER DETECTOR FOR DOSE VERIFICATION IN PROTON THERAPY

SCINTILLATION FIBER DETECTOR FOR DOSE VERIFICATION IN PROTON THERAPY SCINTILLATION FIBER DETECTOR FOR DOSE VERIFICATION IN PROTON THERAPY Suhyun Lee Nuclear Physics Lab. Korea University CONTENTS Introduction Detector - Scintillation-fiber array - Electronics Beam Test

More information

Pixel hybrid photon detectors

Pixel hybrid photon detectors Pixel hybrid photon detectors for the LHCb-RICH system Ken Wyllie On behalf of the LHCb-RICH group CERN, Geneva, Switzerland 1 Outline of the talk Introduction The LHCb detector The RICH 2 counter Overall

More information

GAMMA-GAMMA CORRELATION Latest Revision: August 21, 2007

GAMMA-GAMMA CORRELATION Latest Revision: August 21, 2007 C1-1 GAMMA-GAMMA CORRELATION Latest Revision: August 21, 2007 QUESTION TO BE INVESTIGATED: decay event? What is the angular correlation between two gamma rays emitted by a single INTRODUCTION & THEORY:

More information

NIST EUVL Metrology Programs

NIST EUVL Metrology Programs NIST EUVL Metrology Programs S.Grantham, C. Tarrio, R.E. Vest, Y. Barad, S. Kulin, K. Liu and T.B. Lucatorto National Institute of Standards and Technology (NIST) Gaithersburg, MD USA L. Klebanoff and

More information

Tutors Dominik Dannheim, Thibault Frisson (CERN, Geneva, Switzerland)

Tutors Dominik Dannheim, Thibault Frisson (CERN, Geneva, Switzerland) Danube School on Instrumentation in Elementary Particle & Nuclear Physics University of Novi Sad, Serbia, September 8 th 13 th, 2014 Lab Experiment: Characterization of Silicon Photomultipliers Dominik

More information

Characterisation of SiPM Index :

Characterisation of SiPM Index : Characterisation of SiPM --------------------------------------------------------------------------------------------Index : 1. Basics of SiPM* 2. SiPM module 3. Working principle 4. Experimental setup

More information

CHAPTER 8 PHOTOMULTIPLIER TUBE MODULES

CHAPTER 8 PHOTOMULTIPLIER TUBE MODULES CHAPTER 8 PHOTOMULTIPLIER TUBE MODULES This chapter describes the structure, usage, and characteristics of photomultiplier tube () modules. These modules consist of a photomultiplier tube, a voltage-divider

More information

nanomca-ii-sp datasheet

nanomca-ii-sp datasheet datasheet nanomca-ii-sp 125 MHz ULTRA-HIGH PERFORMANCE DIGITAL MCA WITH BUILT IN PREAMPLIFIER Model Numbers: SP8004 to SP8009 Standard Models: SP8006B and SP8006A I. FEATURES Finger-sized, ultra-high performance

More information

Total Absorption Dual Readout Calorimetry R&D

Total Absorption Dual Readout Calorimetry R&D Available online at www.sciencedirect.com Physics Procedia 37 (2012 ) 309 316 TIPP 2011 - Technology and Instrumentation for Particle Physics 2011 Total Absorption Dual Readout Calorimetry R&D B. Bilki

More information

PMT tests at UMD. Vlasios Vasileiou Version st May 2006

PMT tests at UMD. Vlasios Vasileiou Version st May 2006 PMT tests at UMD Vlasios Vasileiou Version 1.0 1st May 2006 Abstract This memo describes the tests performed on three Milagro PMTs in UMD. Initially, pulse-height distributions of the PMT signals were

More information

The Trigger System of the MEG Experiment

The Trigger System of the MEG Experiment The Trigger System of the MEG Experiment On behalf of D. Nicolò F. Morsani S. Galeotti M. Grassi Marco Grassi INFN - Pisa Lecce - 23 Sep. 2003 1 COBRA magnet Background Rate Evaluation Drift Chambers Target

More information

Mass Spectrometry and the Modern Digitizer

Mass Spectrometry and the Modern Digitizer Mass Spectrometry and the Modern Digitizer The scientific field of Mass Spectrometry (MS) has been under constant research and development for over a hundred years, ever since scientists discovered that

More information

DAQ & Electronics for the CW Beam at Jefferson Lab

DAQ & Electronics for the CW Beam at Jefferson Lab DAQ & Electronics for the CW Beam at Jefferson Lab Benjamin Raydo EIC Detector Workshop @ Jefferson Lab June 4-5, 2010 High Event and Data Rates Goals for EIC Trigger Trigger must be able to handle high

More information

Data Acquisition System for the Angra Project

Data Acquisition System for the Angra Project Angra Neutrino Project AngraNote 012-2009 (Draft) Data Acquisition System for the Angra Project H. P. Lima Jr, A. F. Barbosa, R. G. Gama Centro Brasileiro de Pesquisas Físicas - CBPF L. F. G. Gonzalez

More information

CAEN. Electronic Instrumentation DPP-PSD. Rev July Digital Pulse Processing for Pulse Shape Discrimination. User Manual UM2580

CAEN. Electronic Instrumentation DPP-PSD. Rev July Digital Pulse Processing for Pulse Shape Discrimination. User Manual UM2580 Tools for Discovery n Rev 4-21 July 2014 User Manual UM2580 DPP-PSD Digital Pulse Processing for Pulse Shape Discrimination Rev 8 - September 29th, 2016 Purpose of this Manual This User Manual contains

More information

itop System Overview Kurtis Nishimura University of Hawaii October 12, 2012 US Belle II Firmware Review

itop System Overview Kurtis Nishimura University of Hawaii October 12, 2012 US Belle II Firmware Review itop System Overview Kurtis Nishimura University of Hawaii October 12, 2012 US Belle II Firmware Review Detection of Internally Reflected Cherenkov Light Charged particles of same momentum but different

More information

TB-5 User Manual. Products for Your Imagination

TB-5 User Manual. Products for Your Imagination TB-5 User Manual 1 Introduction... 2 1.1 TB-5 Description... 2 1.2 DP5 Family... 2 1.3 Options and Variations... 3 2 Specifications... 3 2.1 Spectroscopic Performance... 3 2.2 Processing, physical, and

More information

Data Compression and Analysis Methods for High- Throughput Radiation Detector Systems

Data Compression and Analysis Methods for High- Throughput Radiation Detector Systems 1 Data Compression and Analysis Methods for High- Throughput Radiation Detector Systems John Mattingly Associate Professor, Nuclear Engineering North Carolina State University 2 Introduction The capabilities

More information

Development of LYSO detector modules for a charge-particle EDM polarimeter

Development of LYSO detector modules for a charge-particle EDM polarimeter Mitglied der Helmholtz-Gemeinschaft Development of LYSO detector modules for a charge-particle EDM polarimeter on behalf of the JEDI collaboration Dito Shergelashvili, PhD student @ SMART EDM_Lab, TSU,

More information

DETECTORS Important characteristics: 1) Wavelength response 2) Quantum response how light is detected 3) Sensitivity 4) Frequency of response

DETECTORS Important characteristics: 1) Wavelength response 2) Quantum response how light is detected 3) Sensitivity 4) Frequency of response DETECTORS Important characteristics: 1) Wavelength response 2) Quantum response how light is detected 3) Sensitivity 4) Frequency of response (response time) 5) Stability 6) Cost 7) convenience Photoelectric

More information

arxiv: v2 [physics.ins-det] 17 Oct 2015

arxiv: v2 [physics.ins-det] 17 Oct 2015 arxiv:55.9v2 [physics.ins-det] 7 Oct 25 Performance of VUV-sensitive MPPC for Liquid Argon Scintillation Light T.Igarashi, S.Naka, M.Tanaka, T.Washimi, K.Yorita Waseda University, Tokyo, Japan E-mail:

More information

R & D for Aerogel RICH

R & D for Aerogel RICH 1 R & D for Aerogel RICH Ichiro Adachi KEK Proto-Collaboration Meeting March 20, 2008 2 1 st Cherenkov Image detected by 3 hybrid avalanche photon detectors from a beam test About 3:00 AM TODAY Clear image

More information

GAMMA-RAD5 User Manual

GAMMA-RAD5 User Manual GAMMA-RAD5 User Manual 1 Introduction... 2 1.1 Gamma-Rad5 Description... 2 1.2 DP5 Family... 2 1.3 Options and Variations... 3 2 Specifications... 4 2.1 Spectroscopic Performance... 4 2.2 Processing, physical,

More information

ANTICOINCIDENCE LOW LEVEL COUNTING

ANTICOINCIDENCE LOW LEVEL COUNTING Med Phys 4RB3/6R3 LABORATORY EXPERIMENT #7 ANTICOINCIDENCE LOW LEVEL COUNTING Introduction This is the only experiment in this series which involves a multi- system. The low-level electronics used was

More information

Silicon Photomultiplier

Silicon Photomultiplier Silicon Photomultiplier Operation, Performance & Possible Applications Slawomir Piatek Technical Consultant, Hamamatsu Corp. Introduction Very high intrinsic gain together with minimal excess noise make

More information

SUPPLEMENTARY INFORMATION

SUPPLEMENTARY INFORMATION SUPPLEMENTARY INFORMATION doi:10.1038/nature10864 1. Supplementary Methods The three QW samples on which data are reported in the Letter (15 nm) 19 and supplementary materials (18 and 22 nm) 23 were grown

More information

Wavelength-shifting Optical Module (WOM)

Wavelength-shifting Optical Module (WOM) Wavelength-shifting Optical Module (WOM) Vincenzo Di Lorenzo HAP Workshop: Advanced Technologies Mainz 2016-02-02 Cherenkov light Increase the energy resolution decreasing the energy threshold Cherenkov

More information

RAPSODI RAdiation Protection with Silicon Optoelectronic Devices and Instruments

RAPSODI RAdiation Protection with Silicon Optoelectronic Devices and Instruments RAPSODI RAdiation Protection with Silicon Optoelectronic Devices and Instruments Massimo Caccia Universita dell Insubria Como (Italy) on behalf of The RAPSODI collaboration 11th Topical Seminar on Innovative

More information

CATIROC a multichannel front-end ASIC to read out the SPMT system of the JUNO experiment

CATIROC a multichannel front-end ASIC to read out the SPMT system of the JUNO experiment CATIROC a multichannel front-end ASIC to read out the SPMT system of the JUNO experiment Dr. Selma Conforti (OMEGA/IN2P3/CNRS) OMEGA microelectronics group Ecole Polytechnique & CNRS IN2P3 http://omega.in2p3.fr

More information

The HPD DETECTOR. Michele Giunta. VLVnT Workshop "Technical Aspects of a Very Large Volume Neutrino Telescope in the Mediterranean Sea"

The HPD DETECTOR. Michele Giunta. VLVnT Workshop Technical Aspects of a Very Large Volume Neutrino Telescope in the Mediterranean Sea The HPD DETECTOR VLVnT Workshop "Technical Aspects of a Very Large Volume Neutrino Telescope in the Mediterranean Sea" In this presentation: The HPD working principles The HPD production CLUE Experiment

More information

HF Upgrade Studies: Characterization of Photo-Multiplier Tubes

HF Upgrade Studies: Characterization of Photo-Multiplier Tubes HF Upgrade Studies: Characterization of Photo-Multiplier Tubes 1. Introduction Photomultiplier tubes (PMTs) are very sensitive light detectors which are commonly used in high energy physics experiments.

More information

K 223 Angular Correlation

K 223 Angular Correlation K 223 Angular Correlation K 223.1 Aim of the Experiment The aim of the experiment is to measure the angular correlation of a γ γ cascade. K 223.2 Required Knowledge Definition of the angular correlation

More information

A Fast Waveform-Digitizing ASICbased DAQ for a Position & Time Sensing Large-Area Photo-Detector System

A Fast Waveform-Digitizing ASICbased DAQ for a Position & Time Sensing Large-Area Photo-Detector System A Fast Waveform-Digitizing ASICbased DAQ for a Position & Time Sensing Large-Area Photo-Detector System Eric Oberla on behalf of the LAPPD collaboration PHOTODET 2012 12-June-2012 Outline LAPPD overview:

More information

Scintillators as an external trigger for cathode strip chambers

Scintillators as an external trigger for cathode strip chambers Scintillators as an external trigger for cathode strip chambers J. A. Muñoz Department of Physics, Princeton University, Princeton, NJ 08544 An external trigger was set up to test cathode strip chambers

More information

Characterization of Silicon Photomultipliers and their Application to Positron Emission Tomography. Zhiwei Yang. Abstract

Characterization of Silicon Photomultipliers and their Application to Positron Emission Tomography. Zhiwei Yang. Abstract DESY Summer Student Program 2009 Report No. Characterization of Silicon Photomultipliers and their Application to Positron Emission Tomography Zhiwei Yang V. N. Karazin Kharkiv National University E-mail:

More information

PROGRESS in TOF PET timing resolution continues to

PROGRESS in TOF PET timing resolution continues to Combined Analog/Digital Approach to Performance Optimization for the LAPET Whole-Body TOF PET Scanner W. J. Ashmanskas, Member, IEEE, Z. S. Davidson, B. C. LeGeyt, F. M. Newcomer, Member, IEEE, J. V. Panetta,

More information

Contents. The AMADEUS experiment at the DAFNE collider. The AMADEUS trigger. SiPM characterization and lab tests

Contents. The AMADEUS experiment at the DAFNE collider. The AMADEUS trigger. SiPM characterization and lab tests Contents The AMADEUS experiment at the DAFNE collider The AMADEUS trigger SiPM characterization and lab tests First trigger prototype; tests at the DAFNE beam Second prototype and tests at PSI beam Conclusions

More information

FAST DIGITIZING TECHNIQUES APPLIED TO SCINTILLATION DETECTORS

FAST DIGITIZING TECHNIQUES APPLIED TO SCINTILLATION DETECTORS 9 th Topical Seminar on Innovative Particle and Radiation Detectors 23-26 May 2004 Siena, Italy FAST DIGITIZING TECHNIQUES APPLIED TO SCINTILLATION DETECTORS L. Bertalot 1, B. Esposito 1, Y. Kaschuck 2,

More information

The LUCID-2 Detector RICHARD SOLUK, UNIVERSITY OF ALBERTA FOR THE ATLAS- LUCID GROUP

The LUCID-2 Detector RICHARD SOLUK, UNIVERSITY OF ALBERTA FOR THE ATLAS- LUCID GROUP The LUCID-2 Detector RICHARD SOLUK, UNIVERSITY OF ALBERTA FOR THE ATLAS- LUCID GROUP LUCID (LUminosity Cerenkov Integrating Detector) LUCID LUCID LUCID is the only dedicated luminosity monitor in ATLAS

More information

The Argonne 6cm MCP-PMT System. Bob Wagner for Argonne LAPPD Collaboration ANNIE Collaboration Meeting Monday 27 Oct 2014

The Argonne 6cm MCP-PMT System. Bob Wagner for Argonne LAPPD Collaboration ANNIE Collaboration Meeting Monday 27 Oct 2014 The Argonne 6cm MCP-PMT System Bob Wagner for Argonne LAPPD Collaboration ANNIE Collaboration Meeting Monday 27 Oct 2014 Thanks to Argonne Postdocs Junqi Xie (photocathode) & Jingbo Wang (analysis) for

More information

Use of a CAEN digitiser for nuclear safeguards and security applications with a scintillator detector

Use of a CAEN digitiser for nuclear safeguards and security applications with a scintillator detector Use of a CAEN digitiser for nuclear safeguards and security applications with a scintillator detector J. Paepen P. Peerani P. Schillebeeckx A. Tomanin R. Wynants 2013 Report EUR 26480 EN European Commission

More information

Photon Detector with PbWO 4 Crystals and APD Readout

Photon Detector with PbWO 4 Crystals and APD Readout Photon Detector with PbWO 4 Crystals and APD Readout APS April Meeting in Denver, CO on May 4, 2004 presented by Kenta Shigaki (Hiroshima University, Japan) for the ALICE-PHOS Collaboration - Presentation

More information

The role ofcompact PSPMTs for image quality enhancement in nuclear medicine

The role ofcompact PSPMTs for image quality enhancement in nuclear medicine Nuclear Instruments and Methods in Physics Research A 505 (2003) 599 603 The role ofcompact PSPMTs for image quality enhancement in nuclear medicine M.N. Cinti a, R. Pani b, *, R. Pellegrini b, F. Garibaldi

More information

Development of TOP counter for Super B factory

Development of TOP counter for Super B factory 2009/5/11-13 Workshop on fast Cherenkov detectors - Photon detection, DIRC design and DAQ Development of TOP counter for Super B factory - Introduction - Design study - Focusing system - Prototype development

More information

EKA Laboratory Muon Lifetime Experiment Instructions. October 2006

EKA Laboratory Muon Lifetime Experiment Instructions. October 2006 EKA Laboratory Muon Lifetime Experiment Instructions October 2006 0 Lab setup and singles rate. When high-energy cosmic rays encounter the earth's atmosphere, they decay into a shower of elementary particles.

More information

Electronic Instrumentation for Radiation Detection Systems

Electronic Instrumentation for Radiation Detection Systems Electronic Instrumentation for Radiation Detection Systems January 23, 2018 Joshua W. Cates, Ph.D. and Craig S. Levin, Ph.D. Course Outline Lecture Overview Brief Review of Radiation Detectors Detector

More information

Goal of the project. TPC operation. Raw data. Calibration

Goal of the project. TPC operation. Raw data. Calibration Goal of the project The main goal of this project was to realise the reconstruction of α tracks in an optically read out GEM (Gas Electron Multiplier) based Time Projection Chamber (TPC). Secondary goal

More information

Results on the LED Pulser System for the Hall A DVCS Experiment

Results on the LED Pulser System for the Hall A DVCS Experiment Results on the LED Pulser System for the Hall A DVCS Experiment Fernando J. Barbosa, Pierre Bertin Jefferson Lab 28 February 2003 System Description The LED Pulser System Diagram is shown in figure 1.

More information

CMS Beam Condition Monitoring Wim de Boer, Hannes Bol, Alexander Furgeri, Steffen Muller

CMS Beam Condition Monitoring Wim de Boer, Hannes Bol, Alexander Furgeri, Steffen Muller CMS Beam Condition Monitoring Wim de Boer, Hannes Bol, Alexander Furgeri, Steffen Muller BCM2 8diamonds BCM1 8diamonds each BCM2 8diamonds Beam Condition Monitoring at LHC BCM at LHC is done by about 3700

More information

Components of Optical Instruments

Components of Optical Instruments Components of Optical Instruments General Design of Optical Instruments Sources of Radiation Wavelength Selectors (Filters, Monochromators, Interferometers) Sample Containers Radiation Transducers (Detectors)

More information

5. Scintillation counters

5. Scintillation counters 5. Scintillation counters to detect radiation by means of scintillation is among oldest methods of particle detection historical example: particle impinging on ZnS screen -> emission of light flash principle

More information

Product Range Electronic Units

Product Range Electronic Units Pyramid Technical Consultants, Inc. 1050 Waltham Street Suite 200 Lexington, MA 02421 TEL: +1 781 402-1700 TEL (UK): +44 1273 492001 FAX: (781) 402-1750 EMAIL: SUPPORT@PTCUSA.COM Product Range Electronic

More information

CALICE AHCAL overview

CALICE AHCAL overview International Workshop on the High Energy Circular Electron-Positron Collider in 2018 CALICE AHCAL overview Yong Liu (IHEP), on behalf of the CALICE collaboration Nov. 13, 2018 CALICE-AHCAL Progress, CEPC

More information

5. Scintillation counters

5. Scintillation counters 5. Scintillation counters to detect radiation by means of scintillation is among oldest methods of particle detection particle impinging on ZnS screen -> emission of light flash principle of scintillation

More information

Physics Laboratory Scattering of Photons from Electrons: Compton Scattering

Physics Laboratory Scattering of Photons from Electrons: Compton Scattering RR Oct 2001 SS Dec 2001 MJ Oct 2009 Physics 34000 Laboratory Scattering of Photons from Electrons: Compton Scattering Objective: To measure the energy of high energy photons scattered from electrons in

More information

Detectors for microscopy - CCDs, APDs and PMTs. Antonia Göhler. Nov 2014

Detectors for microscopy - CCDs, APDs and PMTs. Antonia Göhler. Nov 2014 Detectors for microscopy - CCDs, APDs and PMTs Antonia Göhler Nov 2014 Detectors/Sensors in general are devices that detect events or changes in quantities (intensities) and provide a corresponding output,

More information

Homework Set 3.5 Sensitive optoelectronic detectors: seeing single photons

Homework Set 3.5 Sensitive optoelectronic detectors: seeing single photons Homework Set 3.5 Sensitive optoelectronic detectors: seeing single photons Due by 12:00 noon (in class) on Tuesday, Nov. 7, 2006. This is another hybrid lab/homework; please see Section 3.4 for what you

More information

CAEN. Electronic Instrumentation. UM2088 DPHA User Manual. Rev June Rev June Tools for Discovery

CAEN. Electronic Instrumentation. UM2088 DPHA User Manual. Rev June Rev June Tools for Discovery Tools for Discovery n Rev. 3-03 June 2013 UM2088 DPHA User Manual Rev. 3-03 June 2013 Purpose of this Manual This User Manual contains the full description of Digital Pulse Height Analyzer for 724 series.

More information

Radiation Detection Instrumentation

Radiation Detection Instrumentation Radiation Detection Instrumentation Principles of Detection and Gas-filled Ionization Chambers Neutron Sensitive Ionization Chambers Detection of radiation is a consequence of radiation interaction with

More information

Performance of the MCP-PMTs of the TOP counter in the first beam operation of the Belle II experiment

Performance of the MCP-PMTs of the TOP counter in the first beam operation of the Belle II experiment Performance of the MCP-PMTs of the TOP counter in the first beam operation of the Belle II experiment K. Matsuoka (KMI, Nagoya Univ.) on behalf of the Belle II TOP group 5th International Workshop on New

More information

Photon Counters SR430 5 ns multichannel scaler/averager

Photon Counters SR430 5 ns multichannel scaler/averager Photon Counters SR430 5 ns multichannel scaler/averager SR430 Multichannel Scaler/Averager 5 ns to 10 ms bin width Count rates up to 100 MHz 1k to 32k bins per record Built-in discriminator No interchannel

More information

Diamond sensors as beam conditions monitors in CMS and LHC

Diamond sensors as beam conditions monitors in CMS and LHC Diamond sensors as beam conditions monitors in CMS and LHC Maria Hempel DESY Zeuthen & BTU Cottbus on behalf of the BRM-CMS and CMS-DESY groups GSI Darmstadt, 11th - 13th December 2011 Outline 1. Description

More information

LHCb Preshower(PS) and Scintillating Pad Detector (SPD): commissioning, calibration, and monitoring

LHCb Preshower(PS) and Scintillating Pad Detector (SPD): commissioning, calibration, and monitoring LHCb Preshower(PS) and Scintillating Pad Detector (SPD): commissioning, calibration, and monitoring Eduardo Picatoste Olloqui on behalf of the LHCb Collaboration Universitat de Barcelona, Facultat de Física,

More information

HPS Upgrade Proposal

HPS Upgrade Proposal HPS Upgrade Proposal HPS collaboration July 20, 2017 Analysis of the HPS engineering run data showed worse than expected reach in both the bump hunt and the vertexing searches. These reach discrepancies

More information

THE increasing interest on pinhole collimation of gamma

THE increasing interest on pinhole collimation of gamma IEEE TRANSACTIONS ON NUCLEAR SCIENCE, VOL. 54, NO. 3, JUNE 2007 469 CsI(Tl) Micro-Pixel Scintillation Array for Ultra-high Resolution Gamma-ray Imaging M. N. Cinti, R. Scafè, R. Pellegrini, C. Trotta,

More information

Measuring Atlas Radiation Backgrounds in the Muon System at Startup: A U.S. ATLAS Upgrade R&D Project

Measuring Atlas Radiation Backgrounds in the Muon System at Startup: A U.S. ATLAS Upgrade R&D Project Slide 1 Measuring Atlas Radiation Backgrounds in the Muon System at Startup: A U.S. ATLAS Upgrade R&D Project, Leif Shaver, Michael Starr, Matt Adams (2007-08, undergraduate) THIS WORK IS AN ATLAS UPGRADE

More information

PET Detectors. William W. Moses Lawrence Berkeley National Laboratory March 26, 2002

PET Detectors. William W. Moses Lawrence Berkeley National Laboratory March 26, 2002 PET Detectors William W. Moses Lawrence Berkeley National Laboratory March 26, 2002 Step 1: Inject Patient with Radioactive Drug Drug is labeled with positron (β + ) emitting radionuclide. Drug localizes

More information

ORTEC Experiment 13. Gamma-Gamma Coincidence with Angular Correlation. Equipment Required

ORTEC Experiment 13. Gamma-Gamma Coincidence with Angular Correlation. Equipment Required ORTEC Experiment 13 Equipment Required Two 905-3 2-in. x 2-in. NaI(Tl) Scintillation Detector Assemblies. Two 266 Photomultiplier Tube Bases. Two 113 Scintillation Preamplifiers. Two 556 High Voltage Power

More information

Scintillation Counters

Scintillation Counters PHY311/312 Detectors for Nuclear and Particle Physics Dr. C.N. Booth Scintillation Counters Unlike many other particle detectors, which exploit the ionisation produced by the passage of a charged particle,

More information

PHYSICS ADVANCED LABORATORY I COMPTON SCATTERING Spring 2002

PHYSICS ADVANCED LABORATORY I COMPTON SCATTERING Spring 2002 PHYSICS 334 - ADVANCED LABORATORY I COMPTON SCATTERING Spring 00 Purposes: Demonstrate the phenomena associated with Compton scattering and the Klein-Nishina formula. Determine the mass of the electron.

More information

Lecture 12 OPTICAL DETECTORS

Lecture 12 OPTICAL DETECTORS Lecture 12 OPTICL DETECTOS (eference: Optical Electronics in Modern Communications,. Yariv, Oxford, 1977, Ch. 11.) Photomultiplier Tube (PMT) Highly sensitive detector for light from near infrared ultraviolet

More information

The MUSE experiment. Technical Overview. Guy Ron (for the MUSE collaboration) Hebrew University of Jerusalem

The MUSE experiment. Technical Overview. Guy Ron (for the MUSE collaboration) Hebrew University of Jerusalem The MUSE experiment Technical Overview Guy Ron (for the MUSE collaboration) Hebrew University of Jerusalem MUSE is not your garden variety scattering experiment Low beam flux Large angle, non-magnetic

More information

Spontaneous Fission Spectrum of Neutrons from 252 Cf with Kinetic Energies Less than 1 MeV. Suraj Bastola. A senior thesis submitted to the faculty of

Spontaneous Fission Spectrum of Neutrons from 252 Cf with Kinetic Energies Less than 1 MeV. Suraj Bastola. A senior thesis submitted to the faculty of Spontaneous Fission Spectrum of Neutrons from 252 Cf with Kinetic Energies Less than 1 MeV Suraj Bastola A senior thesis submitted to the faculty of Brigham Young University in partial fulfillment of the

More information

ARTICLE IN PRESS. Nuclear Instruments and Methods in Physics Research A

ARTICLE IN PRESS. Nuclear Instruments and Methods in Physics Research A Nuclear Instruments and Methods in Physics Research A 614 (2010) 308 312 Contents lists available at ScienceDirect Nuclear Instruments and Methods in Physics Research A journal homepage: www.elsevier.com/locate/nima

More information

The KM3NeT Digital Optical Module NNN16 IHEP,Beijing. Ronald Bruijn Universiteit van Amsterdam/Nikhef

The KM3NeT Digital Optical Module NNN16 IHEP,Beijing. Ronald Bruijn Universiteit van Amsterdam/Nikhef The KM3NeT Digital Optical Module NNN16 IHEP,Beijing Ronald Bruijn Universiteit van Amsterdam/Nikhef 1 Large Volume Neutrino Telescopes Cherenkov light from the charged products of neutrino interactions

More information

Electron-Bombarded CMOS

Electron-Bombarded CMOS New Megapixel Single Photon Position Sensitive HPD: Electron-Bombarded CMOS University of Lyon / CNRS-IN2P3 in collaboration with J. Baudot, E. Chabanat, P. Depasse, W. Dulinski, N. Estre, M. Winter N56:

More information

Measurement of the FD camera light collection efficiency and uniformity

Measurement of the FD camera light collection efficiency and uniformity GAP - 2000-010 Roma, 1 March 2000 Measurement of the FD camera light collection efficiency and uniformity P. Facal San Luis Sezione INFN di Roma II, Roma, Italy and Universidad de Santiago de Compostela,

More information

Seminar. BELLE II Particle Identification Detector and readout system. Andrej Seljak advisor: Prof. Samo Korpar October 2010

Seminar. BELLE II Particle Identification Detector and readout system. Andrej Seljak advisor: Prof. Samo Korpar October 2010 Seminar BELLE II Particle Identification Detector and readout system Andrej Seljak advisor: Prof. Samo Korpar October 2010 Outline Motivation BELLE experiment and future upgrade plans RICH proximity focusing

More information