TECHNICAL PROTOCOL FOR KEY COMPARISON CCAUV.A-K4

Size: px
Start display at page:

Download "TECHNICAL PROTOCOL FOR KEY COMPARISON CCAUV.A-K4"

Transcription

1 EL CENTRO NACIONAL DE METROLOGÍA & THE DANISH PRIMARY LABORATORY FOR ACOUSTICS TECHNICAL PROTOCOL FOR KEY COMPARISON CCAUV.A-K4 OCTOBER 2006 DRAFT Knud Rasmussen & Salvador Barrera-Figueroa Danish Primary Laboratory for Acoustics Kgs. Lyngby Denmark Tel Fax

2 DRAFT TECHNICAL PROTOCOL FOR KEY COMPARISON CCAUV.A-K4 BACKGROUND It was agreed during the 2 nd meeting of the CCAUV in Sèvres, France on October 2001 that after finishing the first Key Comparison organised by the CCAUV, denominated as CCAUV.A-K1, concerned with pressure calibration of laboratory standard microphones LS1P, a free-field calibration of laboratory standard microphones type LS2aP should be the next step. The CCAUV defined an outline for a key comparison and it was agreed that CENAM would be the pilot laboratory with technical assistance given by DPLA. This document defines the protocol for the comparison. It should be read in conjunction with the Guidelines for key comparisons carried out by Consultative Committees by T J Quinn which includes more details on the purpose and conduct of key comparisons in general. The purpose of this document is to specify the procedures necessary for the comparison, but not the procedures used for the realization of the standards being compared. PARTICIPANTS The following laboratories will participate: BNM-LNE, France CENAM, Mexico DPLA, Denmark INMETRO, Brazil NIST, USA NMIJ, Japan NPL, UK PTB, Germany KRISS, Korea Contact details are attached. MICROPHONES TO BE CIRCULATED Two LS2aP microphones have been selected for this comparison. They are Brüel & Kjær type 4180 microphones serial numbers yyyyyyy and zzzzzzz. Each participant is responsible for the collection and return of the microphones to DPLA. Local customs formalities must be observed and if the participating laboratory requires DPLA to supply an ATA carnet for this purpose (or any other documentation) then they must give sufficient notice for DPLA to obtain that documentation. The timetable must be followed regardless of any delays caused by customs irregularities and this could cause a laboratory to loose the opportunity to participate in the comparison.

3 The microphones may be transported either by a courier company like DHL or by hand. If a courier company is chosen the microphones will be put in a thermo-jug packed in an isolated box delivered by DPLA. The microphones shall be returned using the same packing. If the microphones are transported by hand the microphones will be packed into the small box as provided by the manufacturer in order to avoid check-in problems during travelling. In this situation the microphones shall be carried only in the passenger cabin and never in the hold of an aircraft. No packaging will be supplied beyond that provided by the manufacturer of the microphones as it is thought that the best protection is the vigilance of the person carrying the microphones and this is easier to achieve when the package is small. Large temperature changes and sudden shocks must be avoided as these could cause an irreversible change in sensitivity or for the microphones to become less stable. Sudden shocks can be caused by applying sound calibrators, pistonphones or dehumidifiers to the microphones and these actions should also be avoided or performed with great care. The microphones must not be used for any purposes other than those associated with their calibration for this comparison. The microphones will return to DPLA for a check calibration after each calibration in a participating laboratory. This is so that the stability of the devices can be monitored and so that results from different laboratories can still be compared should a change occur. MEASUREMENTS This key comparison is only concerned with primary methods of calibration and will only consider results from such methods. The microphones require a polarising voltage of 200 V. Each laboratory is to report the open-circuit free-field sensitivity level of both microphones at the nominal preferred 1/3 rd -octave frequencies from 1 khz to 31,5 khz and optionally also 40 khz using their normal primary calibration method. The nominal frequencies are defined by the R10 series given in ISO 266. These frequencies are chosen in order to comply with the results of the CCAUV.A-K3 key comparison. It is recognized that the calibration technique used by the participants are quite different and may not allow calibrations to be performed at the nominal preferred frequencies but at other frequencies close to those. It is accepted that an interpolation technique is used to report the results at the preferred nominal frequencies. The interpolation technique used to derive the result at the required frequency shall be reported together with the results at the actual measurement frequencies in a separate report or as an annex to the main report. Also it may not be feasible to perform a free-field reciprocity calibration at frequencies down to 1 khz. It is accepted that the reported free-field sensitivity at the lowest frequencies are based on estimations derived by calculations or from a pressure calibration. The method used to estimate the sensitivities at the lowest frequencies shall be reported. It is recommended not to use the standard corrections given in IEC/TS since a side effect of the key comparison is to collect information to support or challenge this document. The initial free-field reciprocity calibrations at the pilot laboratory will be conducted in linear frequency steps of 120 Hz from 1-51 khz while the following performance checks will be reduced to the frequency range from 3 51 khz in addition to a pressure reciprocity calibration in the full frequency range. The microphones submitted shall only be used as receivers. Results shall be corrected to the reference environmental conditions given in IEC

4 REPORTING RESULTS Each laboratory shall report their results using the standard certificate that they would issue to a customer. This certificate should include: The environmental conditions during the calibrations. The actual distance(s) used between the microphones during the calibrations. Values of the temperature and pressure coefficients of the microphones used in the calculations. A summary of the uncertainty calculation and the overall uncertainties calculated for a coverage factor of k=2. It must be noticed that the CIPM Guidelines require uncertainty budgets to be submitted in advance of participation. Each laboratory shall submit a sufficiently detailed uncertainty budget and failure on doing so may result in the exclusion of the laboratory. The (preliminary) budget shall be submitted before the key comparison starts. However, any changes will be accepted, provided they are submitted before the laboratory receives the microphones and carries out the measurements. An additional covering letter or report should include (where necessary): Details of any deviations from the recommendations in IEC and an estimate of the effect on the reported results. The values of the acoustic centres used in the calculation. Calibration results using three decimals (if not already given in the certificate). The interpolation technique used if the actual measurements are not performed at the required frequencies. The technique used to estimate the free-field sensitivity at low frequencies if a direct calibration is not available. A recalculation of the results using a common set of acoustic centre values (to be circulated by the pilot laboratory). The results should be posted to DPLA within a maximum of four weeks of the return of the microphones. This deadline shall be strictly enforced. Failure on submitting the results within the deadline may lead to the exclusion of the participant laboratory. An or fax should be sent to announce that the results have been despatched. When all participants have finished the measurements, the reports together with the results from DPLA will be submitted to the pilot laboratory (CENAM) for analysis. If a result is found anomalous the laboratory in question will be notified and given 3 weeks to respond. The results from DPLA will then be circulated to the participants and the pilot laboratory will prepare a first draft report. FINANCE Participants are responsible for their own costs, the collection and return of the microphones and for any damage to the microphones while they are away from DPLA. When the microphones are transported by a courier company the participant shall contact the courier company and arrange for the microphones to be picked up at DPLA. In case DHL or UPS are used the laboratory may inform DPLA on their relevant import account number and full receiver address of the laboratory. DPLA will then contact the company and ask for a pick-up.

5 KEY COMPARISON REFERENCE VALUE The determination of the Key Comparison Reference Values is an important outcome of this project. It is expected that the analysis of the results can be conducted in the same way as used for the previous key comparison CCAUV.A-K3. TIMETABLE The timetable is shown below. It is essential that the microphones be returned to DPLA on time even if measurements are not complete at the participating laboratory. If a problem arises with the timetable then it may be possible for two participants to exchange their place in the timetable. If measurements cannot be completed at a laboratory then it may be possible for the microphones to go to that laboratory after the termination of the measurement round (in 2008). However DPLA cannot guarantee to perform check measurements after this date. In the event of one of the microphones failing then DPLA will attempt to find a substitute though this may make the analysis of the results more complicated. No Calibration laboratory Starting date Finishing date 1 DPLA 12 February March CENAM 5 March March DPLA 2 April April KRISS 23 April May DPLA 21 May June NPL 11 June July DPLA 9 July July NMIJ 30 July August DPLA 27 August September BNM-LNE 17 September October DPLA 15 October November PTB 5 November November DPLA 3 December December INMETRO 4 January February DPLA 4 February February NIST 26 February March DPLA 24 March April 2008 List of contact persons

6 BNM-LNE: Jean-Noel Durocher Laboratoire national de métrologie et d essais 29, avenue Roger Hennequin Trappes Cedex France Tel: Fax: jean-noel.durocher@lne.fr CENAM: Andrés Esteban Pérez Matzumoto Centro Nacional de Metrologia Vibration and Acoustics Division Km 4,5 Carretera a Los Cués, Municipio El Marqués C.P Querétaro Mexico Tel: Fax: eperez@cenam.mx DPLA: Knud Rasmussen, Salvador Barrera-Figueroa Danish Primary Laboratory of Acoustics, DFM Matematiktorvet Building 307, DTU 2800 Kgs. Lyngby Denmark Tel: ; Fax: kra@dfm.dtu.dk; sbf@dfm.dtu.dk INMETRO: Walter E. Hoffmann INMETRO - Laboratório de Eletroacústica Av. N. Sra Das Gracas, 50 Xerém CEP Duque de Caxias, RJ Brazil Tel: Fax: laeta@inmetro.gov.br KRISS: Sang Joon Suh Korea Research Institute of Standards and Science Acoustics and Vibration Group Division of Physical Metrology 1 Doryong-Dong, Yoseong Daejeon Korea, Republic of Tel: Fax: sjs@kriss.re.kr NIST: Victor Nedzelnitsky National Institute of Standards and Technology Acoustical Metrology 100 Bureau Drive, Stop 8221 Building 233, Room B140 MD , Gaithersburg United States Tel: Fax: victor.nedzelnitsky@nist.gov NMIJ: Ryuzo Horiuchi National Metrology Institute of Japan, AIST Acoustics and Vibration Metrology Division Tsukuba Central 2 Tsubuka Japan Tel: Fax: ryuzo.horiuchi@aist.go.jp NPL: Richard Barham National Physical Laboratory Quality of Life Division Hampton Road Teddington, Middlesex TW11 0LW United Kingdom Tel: Fax: Richard.barham@npl.co.uk PTB: Christian Koch Physikalisch-Technische Bundesanstalt FB 1.6 Bundesalle 100 D Braunschweig Germany Tel: Fax: christian.koch@ptb.de

TECHNICAL PROTOCOL FOR AFRIMETS RMO COMPARISON AFRIMETS.AUV.A-S1. Version 4

TECHNICAL PROTOCOL FOR AFRIMETS RMO COMPARISON AFRIMETS.AUV.A-S1. Version 4 TECHNICAL PROTOCOL FOR AFRIMETS RMO COMPARISON AFRIMETS.AUV.A-S1 Version 4 30 May 2013 INTRODUCTION The first CIPM CCAUV key comparison, where the artefacts were LS2aP microphones, was the CCAUV.A-K3 as

More information

Primary calibration of measurement microphones in the world: state of art

Primary calibration of measurement microphones in the world: state of art Primary calibration of measurement microphones in the world: state of art Thiago Antônio Bacelar Milhomem, Zemar Martins Defilippo Soares Electroacoustics Laboratory / Acoustics and Vibration Metrology

More information

TECHNICAL PROTOCOL FOR REGIONAL SUPPLEMENTARY COMPARISON APMP.AUV.A-S1

TECHNICAL PROTOCOL FOR REGIONAL SUPPLEMENTARY COMPARISON APMP.AUV.A-S1 TECHNICAL PROTOCOL FOR REGIONAL SUPPLEMENTARY COMPARISON APMP.AUV.A-S1 August 14, 2008 National Institute of Metrology (Thailand) (NIMT) Page 1 of 9 Preface It was agreed during the 8 th APMP TCAUV meeting

More information

Report on key comparison CCAUV.A-K5: Pressure calibration of laboratory standard microphones in the frequency range 2 Hz to 10 khz

Report on key comparison CCAUV.A-K5: Pressure calibration of laboratory standard microphones in the frequency range 2 Hz to 10 khz NPL REPORT AC 11 Report on key comparison CCAUV.A-K5: Pressure calibration of laboratory standard microphones in the frequency range 2 Hz to 1 khz JANINE AVISON RICHARD BARHAM SEPTEMBER 214 Report on key

More information

REPORT ON KEY COMPARISON APMP.AUV.A K3

REPORT ON KEY COMPARISON APMP.AUV.A K3 REPORT ON KEY COMPARISON APMP.AUV.A K3 1 December, 2008 Dr Hyu-sang Kwon Korea Research Institute of Standards and Science (KRISS) 1 Abstract A regional key-comparison APMP.AUV.A K3 has been carried out

More information

Physikalisch Technische Bundesanstalt

Physikalisch Technische Bundesanstalt EURAMET Intercomparison: Involute Gear Artifacts 1 Physikalisch Technische Bundesanstalt EURAMET Intercomparison Involute Gear Artifacts Technical Protocol Rev 3 EURAMET Intercomparison: Involute Gear

More information

EURO-ASIAN COOPERATION OF NATIONAL METROLOGICAL INSTITUTIONS (COOMET)

EURO-ASIAN COOPERATION OF NATIONAL METROLOGICAL INSTITUTIONS (COOMET) ЕВРО-АЗИАТСКОЕ СОТРУДНИЧЕСТВО ГОСУДАРСТВЕННЫХ МЕТРОЛОГИЧЕСКИХ УЧРЕЖДЕНИЙ (KOOMET) EURO-ASIAN COOPERATION OF NATIONAL METROLOGICAL INSTITUTIONS (COOMET) Federal State Unitary Enterprise «Russian Metrological

More information

Pilot Study EURAMET.AUV.V-P1: Bilateral comparison on magnitude of the complex charge sensitivity of accelerometers from 10 Hz to 10 khz

Pilot Study EURAMET.AUV.V-P1: Bilateral comparison on magnitude of the complex charge sensitivity of accelerometers from 10 Hz to 10 khz Pilot Study EURAMET.AUV.V-P1: Bilateral comparison on magnitude of the complex charge sensitivity of accelerometers from 10 Hz to 10 khz 1) Pilot laboratory: Laboratoire national de métrologie et d'essais

More information

INTERNATIONAL COMPARISON OF ACCELERATION STANDARDS - PRELIMINARY RESULTS OF SIM PROJECT AV-1

INTERNATIONAL COMPARISON OF ACCELERATION STANDARDS - PRELIMINARY RESULTS OF SIM PROJECT AV-1 INTERNATIONAL COMPARISON OF ACCELERATION STANDARDS - PRELIMINARY RESULTS OF SIM PROJECT AV-1 Gustavo Palmeira Ripper INMETRO, Divisão de Acústica e Vibrações, Laboratório de Vibrações Av. N. S. das Graças,

More information

An automatic method to detect defaults in the measurement chain of a sound level meter, used for unattended noise measurements

An automatic method to detect defaults in the measurement chain of a sound level meter, used for unattended noise measurements An automatic method to detect defaults in the measurement chain of a sound level meter, used for unattended noise measurements E. Aflalo a, F. Dupont a, P. Cellard b and J.-N. Durocher b a 01dB-Metravib,

More information

CCEM-K9 COMPARISON OF AC-DC HIGH VOLTAGE STANDARDS

CCEM-K9 COMPARISON OF AC-DC HIGH VOLTAGE STANDARDS CCEM-K9 COMPARISON OF AC-DC HIGH VOLTAGE STANDARDS Key comparison final report Andre POLETAEFF LNE 29, avenue Roger Hennequin 78197 TRAPPES FRANCE Report on the CCEM-K9 comparison of AC-DC high voltage

More information

Bilateral Comparison EURAMET.AUV.V-K1.2. (vibration acceleration) Swiss Federal Office of Metrology METAS Christian Hof

Bilateral Comparison EURAMET.AUV.V-K1.2. (vibration acceleration) Swiss Federal Office of Metrology METAS Christian Hof Final report 2011 04 20 Bilateral Comparison EURAMET.AUV.V-K1.2 (vibration acceleration) Swiss Federal Office of Metrology METAS Christian Hof Abstract This report describes the results obtained in a bilateral

More information

CCAUV.A-K5, COOMET.AUV.A-K5 and AFRIMETS.AUV.A-K5. Key comparison CCAUV.A-K5

CCAUV.A-K5, COOMET.AUV.A-K5 and AFRIMETS.AUV.A-K5. Key comparison CCAUV.A-K5 CCAUV.A-K5, COOMET.AUV.A-K5 and AFRIMETS.AUV.A-K5 Key comparison CCAUV.A-K5 MEASURANDS : Pressure sensitivity levels and pressure sensitivity phases of two standard microphones IEC type LS1P FREQUENCIES

More information

Microphone calibration service for airborne ultrasound

Microphone calibration service for airborne ultrasound Microphone calibration service for airborne ultrasound Christoph KLING Physikalisch-Technische Bundesanstalt (PTB), Germany ABSTRACT The application of ultrasound techniques is wide-spread in many fields

More information

Physikalisch-Technische Bundesanstalt EUROMET supplementary comparison EUROMET.EM-S38 Comparison of small current sources

Physikalisch-Technische Bundesanstalt EUROMET supplementary comparison EUROMET.EM-S38 Comparison of small current sources Physikalisch-Technische Bundesanstalt 2010-01-12 EUROMET supplementary comparison EUROMET.EM-S38 Comparison of small current sources Technical protocol Content 1 Introduction 2 The travelling standards

More information

Technical Protocol. SIM Comparison of Calibration Coefficients at Radiotherapy Level for Orthovoltage X-ray Beams. Pilot laboratory: NIST

Technical Protocol. SIM Comparison of Calibration Coefficients at Radiotherapy Level for Orthovoltage X-ray Beams. Pilot laboratory: NIST SIM Comparison of Calibration Coefficients at Radiotherapy Level for Orthovoltage X-ray Beams Technical Protocol Pilot laboratory: NIST Contact person: Michelle O Brien National Institute of Standards

More information

EUROMET PROJECT 401 Harmonisation of Audiometry Measurements within Europe

EUROMET PROJECT 401 Harmonisation of Audiometry Measurements within Europe EUROMET PROJECT 401 Harmonisation of Audiometry Measurements within Europe Timothy Sherwood June 2004 EUROMET PROJECT 401 - HARMONISATION OF AUDIOMETRY MEASUREMENTS WITHIN EUROPE T R Sherwood Quality of

More information

Bi-lateral comparison of pistonphone calibration between INMETRO and INTI

Bi-lateral comparison of pistonphone calibration between INMETRO and INTI PROCEEDINGS of the 22 nd International Congress on Acoustics Acoustical Measurement and Instrumentation (others): Paper ICA2016-190 Bi-lateral comparison of pistonphone calibration between INMETRO and

More information

FINAL REPORT: KEY COMPARISON EURAMET.AUV.A-K5 (DRAFT B REPORT)

FINAL REPORT: KEY COMPARISON EURAMET.AUV.A-K5 (DRAFT B REPORT) NPL REPORT AC 18 FINAL REPORT: KEY COMPARISON EURAMET.AUV.A-K5 (DRAFT B REPORT) Authors: JANINE AVISON, PETER HARRIS and STEPHEN ROBINSON (NPL) Co-authors: DOMINIQUE RODRIGUES CLAUDIO GUGLIELMONE CHRISTIAN

More information

14 th CCM Meeting. Report of the: Working Group on High Pressure. 20 February 2013, BIPM Dr. Jorge C. Torres Guzmán, CENAM, Mexico

14 th CCM Meeting. Report of the: Working Group on High Pressure. 20 February 2013, BIPM Dr. Jorge C. Torres Guzmán, CENAM, Mexico 14 th CCM Meeting Report of the: Working Group on High Pressure 20 February 2013, BIPM Dr. Jorge C. Torres Guzmán, CENAM, Mexico Contents Terms of Reference Program of work Membership Proposed Changes

More information

TPS 49 EDITION 2 JUNE 2009

TPS 49 EDITION 2 JUNE 2009 TPS 49 EDITION 2 JUNE 2009 Interim arrangements & guidance on the interpretation of IEC 61672 Sound Level Meters - Periodic Tests CONTENTS SECTION PAGE 1 Purpose and duration 2 2 Statement 2 3 Specific

More information

Reciprocity calibration of Laboratory. Standard Microphones. There is simply no substitute for knowing what you re doing

Reciprocity calibration of Laboratory. Standard Microphones. There is simply no substitute for knowing what you re doing Reciprocity calibration of Laboratory Standard Microphones Maria Enge, Philipp Begoff, Frank Schulz There is simply no substitute for knowing what you re doing Jeff Case www.spektra-dresden.com Topics

More information

Continuous development of the national standard for vibration

Continuous development of the national standard for vibration Continuous development of the national standard for vibration by Ian Veldman, National Metrology Laboratory, CSIR In this fast technologically-advancing world, higher measurement accuracy is part of the

More information

Bureau International des Poids et Mesures. International Recognition of NMI Calibration and Measurement Capabilities: The CIPM MRA

Bureau International des Poids et Mesures. International Recognition of NMI Calibration and Measurement Capabilities: The CIPM MRA Bureau International des Poids et Mesures International Recognition of NMI Calibration and Measurement Capabilities: The CIPM MRA Prof. Michael Kühne International School of Physics Enrico Fermi Metrology

More information

CCAUV/ CCAUV. Activities in KRISS. Wan-Cho Cho

CCAUV/ CCAUV. Activities in KRISS. Wan-Cho Cho CCAUV/17-57 2017 CCAUV Activities in Wan-Cho Cho General Organization change Acoustics Vibration Ultrasound 1 General CMC updates (approved at April 2017) Acoustics Whole previously existing items are

More information

Progress towards the revision of the SI. Estefanía de Mirandés CCU executive secretary BIPM

Progress towards the revision of the SI. Estefanía de Mirandés CCU executive secretary BIPM Progress towards the revision of the SI Estefanía de Mirandés CCU executive secretary BIPM Special 2017 CODATA adjustment of h, e, k and N A Resolution 1 of the 24th CGPM (2011) : The General Conference

More information

Certificate of Calibration

Certificate of Calibration Certificate of Calibration ISO/IEC 17025:2005 Certificate Number 1-5607144213-1 Model Number Manufacturer Description Serial Number 34401A Keysight Technologies Inc Digital multimeter, 6.5 digit MY41007563

More information

INTERNATIONAL OIML R 103 RECOMMENDATION

INTERNATIONAL OIML R 103 RECOMMENDATION INTERNATIONAL OIML R 103 RECOMMENDATION Edition 1992 (E) Measuring instrumentation for human response to vibration (with reference to International Standards 8041 and 5347 of the International Organization

More information

Organisation Internationale de Métrologie Légale

Organisation Internationale de Métrologie Légale Organisation Internationale de Métrologie Légale INTERNATIONAL RECOMMENDATION Sound level meters Sonomètres OIML R 58 Edition 1998 (E) CONTENTS Foreword... 3 1 Scope... 4 2 Construction and maximum permissible

More information

PRODUCT DATA USES. BENEFITS Normal incidence parameters are determined Fast and accurate measurements. Type 4206A. Type Type 4206T 50 Hz 1.

PRODUCT DATA USES. BENEFITS Normal incidence parameters are determined Fast and accurate measurements. Type 4206A. Type Type 4206T 50 Hz 1. PRODUCT DATA Impedance Tube Kit (50 Hz 6.4 khz) Type 4206 Impedance Tube Kit (100 Hz 3.2 khz) Type 4206 A Transmission Loss Tube Kit (50 Hz 6.4 khz) Type 4206 T Brüel & Kjær offers a complete range of

More information

CHARACTERIZING ROCKWELL DIAMOND INDENTERS USING DEPTH OF PENETRATION

CHARACTERIZING ROCKWELL DIAMOND INDENTERS USING DEPTH OF PENETRATION HARDMEKO 2004 Hardness Measurements Theory and Application in Laboratories and Industries 11-12 November, 2004, Washington, D.C., USA CHARACTERIZING ROCKWELL DIAMOND INDENTERS USING DEPTH OF PENETRATION

More information

SIM.EM S9.b, 1 Ω and 10 kω

SIM.EM S9.b, 1 Ω and 10 kω SIM.EM S9.b, 1 Ω and 10 kω 2012 Resistance Bilateral Comparison between SIM/COOMET Laboratories TECHNICAL PROTOCOL Laboratories: Pilot: Instituto Nacional de Tecnología Industrial, INTI - Buenos Aires,

More information

COOMET Pilot Comparison 473/RU-a/09: Comparison of hydrophone calibrations in the frequency range 250 Hz to 200 khz

COOMET Pilot Comparison 473/RU-a/09: Comparison of hydrophone calibrations in the frequency range 250 Hz to 200 khz COOMET Pilot Comparison 473/RU-a/09: Comparison of hydrophone calibrations in the frequency range 250 Hz to 200 khz Chen Yi 1, A E Isaev 2, Wang Yuebing 1, A M Enyakov 2, Fei Teng 1 and A N Matveev 2 1

More information

PRODUCT DATA. Sound Intensity Calibrator Type 4297

PRODUCT DATA. Sound Intensity Calibrator Type 4297 PRODUCT DATA Sound Intensity Calibrator Type 4297 Sound Intensity Calibrator Type 4297 is used for on-site sound-pressure calibration and pressure-residual intensity-index verification. Its most important

More information

UPGRADE OF THE MEDIUM AND HIGH FREQUENCY VIBRATION CALIBRATION REFERENCE EQUIPMENT AND EXTENSION TO LOW FREQUENCIES

UPGRADE OF THE MEDIUM AND HIGH FREQUENCY VIBRATION CALIBRATION REFERENCE EQUIPMENT AND EXTENSION TO LOW FREQUENCIES XIX IMEKO World Congress Fundamental and Applied Metrology September 6 11, 9, Lisbon, Portugal UPGRADE OF THE MEDIUM AND HIGH FREQUENCY VIBRATION CALIBRATION REFERENCE EQUIPMENT AND EXTENSION TO LOW FREQUENCIES

More information

EUROMET 557 COMPARISON OF AC-DC HIGH VOLTAGE STANDARDS

EUROMET 557 COMPARISON OF AC-DC HIGH VOLTAGE STANDARDS EUROMET 557 COMPARISON OF AC-DC HIGH VOLTAGE STANDARDS Key comparison Final Report Andre POLETAEFF LNE 29, avenue Roger Hennequin 78197 TRAPPES FRANCE Report on the EUROMET comparison of AC-DC high voltage

More information

Ears Project Newsletter No 3. Welcome. C o n t e n t s. Welcome Preface of the coordinator. News & facts Latest news

Ears Project Newsletter No 3. Welcome. C o n t e n t s. Welcome Preface of the coordinator. News & facts Latest news C o n t e n t s Welcome Preface of the coordinator News & facts Latest news Highlights Infrasound source Ultrasound source Transducer compatibility Ear simulator design Dissemination How we re seen Business

More information

Dr. Takashi Usuda (NMIJ), CCAUV President Dr. Gianna Panfilo (BIPM), Secretary

Dr. Takashi Usuda (NMIJ), CCAUV President Dr. Gianna Panfilo (BIPM), Secretary CCAUV/17-46 Dr. Takashi Usuda (NMIJ), CCAUV President Dr. Gianna Panfilo (BIPM), Secretary Strategic Planning Working Group (SPWG) Dr. Michael Gaitan (NIST), Chairperson Dr. Salvador Figueroa (DFM), Co-Chair

More information

PRODUCT DATA. Sound Intensity Probe Kit Type 3599 including Sound Intensity Microphone Pair Type 4197 and Dual Preamplifier Type 2683

PRODUCT DATA. Sound Intensity Probe Kit Type 3599 including Sound Intensity Microphone Pair Type 4197 and Dual Preamplifier Type 2683 PRODUCT DATA Sound Intensity Probe Kit Type 3599 including Sound Intensity Microphone Pair Type 4197 and Dual Preamplifier Type 2683 Type 3599 is a two-microphone probe kit for measuring sound intensity.

More information

FINAL REPORT. CCEM.RF-K4.CL COMPARISON RF-Voltage measurements up to 1 GHz * Jan P.M. de Vreede 1 (Completed by James Randa 2 )

FINAL REPORT. CCEM.RF-K4.CL COMPARISON RF-Voltage measurements up to 1 GHz * Jan P.M. de Vreede 1 (Completed by James Randa 2 ) FINAL REPORT CCEM.RF-K4.CL COMPARISON RF-Voltage measurements up to 1 GHz * Jan P.M. de Vreede 1 (Completed by James Randa 2 ) 1 Department of Electricity, Radiation and Length Van Swinden Laboratorium

More information

SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005 & ANSI/NCSL Z

SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005 & ANSI/NCSL Z SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005 & ANSI/NCSL Z540-1-1994 BRUEL AND KJAER North American Service Center Calibration Laboratory 3079 Premiere Parkway Suite 120 Duluth, GA 30097 William Shipman

More information

APMP KEY COMPARISON APMP.EM.RF-K3.F. Bilateral Comparison of horn antenna gain

APMP KEY COMPARISON APMP.EM.RF-K3.F. Bilateral Comparison of horn antenna gain APMP KEY COMPARISON APMP.EM.RF-K3.F Bilateral Comparison of horn antenna gain Final Report Jin-Seob Kang Korea Research Institute of Standards and Science (KRISS) 67 Gajeong-ro, Yuseong-gu, Daejeon 305-340,

More information

International comparisons of He-Ne lasers stabilized with 127 I 2 at λ 633 nm (July 2000)

International comparisons of He-Ne lasers stabilized with 127 I 2 at λ 633 nm (July 2000) International Comparison metrologia International comparisons of He-Ne lasers stabilized with 127 I 2 at λ 633 nm (July 2000) Part X: Comparison of INMETRO (Brazil), INTI (Argentina), NRC-INMS (Canada),

More information

The cost and benefits of the CIPM MRA to the NMIs of the world: the Brazilian experience

The cost and benefits of the CIPM MRA to the NMIs of the world: the Brazilian experience The cost and benefits of the CIPM MRA to the NMIs of the world: the Brazilian experience João Alziro Herz da Jornada President of Inmetro- Brazil Costs and Benefits of the CIPM MRA, as perceived by Inmetro-

More information

Validation of primary hydrophone calibrations by inter-laboratory comparisons and by independent calibration methods

Validation of primary hydrophone calibrations by inter-laboratory comparisons and by independent calibration methods Validation of primary hydrophone calibrations by inter-laboratory comparisons and by independent calibration methods S. P Robinson and P. D Theobald National Physical Laboratory, ampton Road, TW11 OLW

More information

Status report of CENAM, Mexico to the 10 th Meeting of the CCAUV 2015

Status report of CENAM, Mexico to the 10 th Meeting of the CCAUV 2015 Status report of CENAM, Mexico to the 10 th Meeting of the CCAUV 2015 In the period 2013-2015, the Acoustics, Ultrasound and Vibrations Area of CENAM has been developing measurement systems, standards

More information

Table of Contents. Compendium SPEKTRA. Calibration Systems CS18. Vibration and Shock Exciters. Vibration Control Systems. Services

Table of Contents. Compendium SPEKTRA. Calibration Systems CS18. Vibration and Shock Exciters. Vibration Control Systems. Services Products & Services Table of Contents Compendium SPEKTRA Calibration Systems CS18 CS18 - Applications CS18 - Acceleration CS18 - Shock CS18 - Acoustics References, Options and Accessories Vibration and

More information

Calibration Certificate

Calibration Certificate ÌÇa8SÈ-EÇÂ!ÈfeFÎ Calibration Certificate Date of issue 2015-12-08 Calibrated by Gert Schnipper Approved by Ole F. Carlsen 1 (5) CUSTOMER: GN RESOUND A/S GTO LAUTRUPBJERG 7-9 2750 BALLERUP DEVICE UNDER

More information

MOY/SCMI/60 SPECIFICATION OF ACCURACY FOR VEE PIECES FOR EXTERNAL SCREW THREAD MEASUREMENT

MOY/SCMI/60 SPECIFICATION OF ACCURACY FOR VEE PIECES FOR EXTERNAL SCREW THREAD MEASUREMENT Centre for Basic, Thermal and Length Metrology National Physical Laboratory MOY/SCMI/60 SPECIFICATION OF ACCURACY FOR VEE PIECES FOR EXTERNAL SCREW THREAD MEASUREMENT Hardened steel vee pieces, for use

More information

Technical Documentation

Technical Documentation Technical Documentation Microphone Type 4964 for Hand-held Analyzer Types 2250, 2250-L and 2270 Supplement to Instruction Manual BE 1712 English BE 1864 11 Microphone Type 4964 for Hand-held Analyzer

More information

OIML R 130 RECOMMENDATION. Edition 2001 (E) ORGANISATION INTERNATIONALE INTERNATIONAL ORGANIZATION. Octave-band and one-third-octave-band filters

OIML R 130 RECOMMENDATION. Edition 2001 (E) ORGANISATION INTERNATIONALE INTERNATIONAL ORGANIZATION. Octave-band and one-third-octave-band filters INTERNATIONAL RECOMMENDATION OIML R 130 Edition 2001 (E) Octave-band and one-third-octave-band filters Filtres à bande d octave et de tiers d octave OIML R 130 Edition 2001 (E) ORGANISATION INTERNATIONALE

More information

Keysight Technologies Optical Power Meter Head Special Calibrations. Brochure

Keysight Technologies Optical Power Meter Head Special Calibrations. Brochure Keysight Technologies Optical Power Meter Head Special Calibrations Brochure Introduction The test and measurement equipment you select and maintain in your production and qualification setups is one of

More information

Comparison of the Josephson Voltage Standards of the LNE and the BIPM

Comparison of the Josephson Voltage Standards of the LNE and the BIPM Comparison of the Josephson Voltage Standards of the LNE and the BIPM (part of the ongoing BIPM key comparison BIPM.EM-K10.b) S. Solve and R. Chayramy Bureau International des Poids et Mesures F- 92312

More information

Certificate of Calibration

Certificate of Calibration Certificate of Calibration ISO/IEC 17025:2005 and ANSI/NCSL Z540.1-1994 Model Number Manufacturer Description Serial Number Customer Asset No. 11667A Keysight Technologies Inc DC-18 GHz power splitter,

More information

Publishable JRP Summary for Project T4 J03 JOSY. Next generation of quantum voltage systems for wide range applications

Publishable JRP Summary for Project T4 J03 JOSY. Next generation of quantum voltage systems for wide range applications Publishable JRP Summary for Project T4 J3 JOSY Next generation of quantum voltage systems for wide range applications The main objective of this project is to introduce quantum-based measurement systems

More information

CALIBRATION OF THE BEV GPS RECEIVER BY USING TWSTFT

CALIBRATION OF THE BEV GPS RECEIVER BY USING TWSTFT CALIBRATION OF THE BEV GPS RECEIVER BY USING TWSTFT A. Niessner 1, W. Mache 1, B. Blanzano, O. Koudelka, J. Becker 3, D. Piester 3, Z. Jiang 4, and F. Arias 4 1 Bundesamt für Eich- und Vermessungswesen,

More information

Report on Research Activities of KRISS CCAUV

Report on Research Activities of KRISS CCAUV Report on Research Activities of KRISS CCAUV Division of Physical Metrology Center for Fluid Flow & Korea Research Institute of Standards and Science KRISS - CCAUV Dr. H. S. Kwon hyusang@kriss.re.kr (Air,

More information

Comparison of the Josephson Voltage Standards of the CENAM and the BIPM (part of the ongoing BIPM key comparison BIPM.EM-K10.b)

Comparison of the Josephson Voltage Standards of the CENAM and the BIPM (part of the ongoing BIPM key comparison BIPM.EM-K10.b) Comparison of the Josephson Voltage Standards of the CENAM and the BIPM (part of the ongoing BIPM key comparison BIPM.EM-K10.b) S. Solve, R. Chayramy, and M. Stock Bureau International des Poids et Mesures

More information

FINAL REPORT. EUROMET project No Bilateral comparison of DC and AC voltages BEV - NCM. W. Waldmann (BEV, pilot laboratory) P.

FINAL REPORT. EUROMET project No Bilateral comparison of DC and AC voltages BEV - NCM. W. Waldmann (BEV, pilot laboratory) P. FINAL REPORT EUROMET project No. 690 Bilateral comparison of DC and AC voltages BEV - NCM W. Waldmann (BEV, pilot laboratory) P. Aladzhem (NCM) April 006 Euromet690_Final_Report.doc page 1 of 1 1. Introduction

More information

Errata to Procedural Standard for Sound & Vibration Measurement 2015 Third Edition

Errata to Procedural Standard for Sound & Vibration Measurement 2015 Third Edition Errata to Procedural Standard for Sound & Vibration Measurement 2015 Third Edition Correction Sheet #1 Issued 01 July 2018 Copyright 2018 by NEBB All rights reserved. Published 2018. Printed in the United

More information

TIME SIGNALS. Their maximum departure from the Universal Time UT1 is thus 0.9 seconds.

TIME SIGNALS. Their maximum departure from the Universal Time UT1 is thus 0.9 seconds. 112 TIME SIGNALS The time signal emissions reported here follow the UTC system, in accordance with the Recommendation 460-4 of the Radiocommunication Bureau (RB) of the International Telecommunication

More information

Organisation Internationale de Métrologie Légale

Organisation Internationale de Métrologie Légale Organisation Internationale de Métrologie Légale INTERNATIONAL RECOMMENDATION Integrating-averaging sound level meters Sonomètres intégrateurs-moyenneurs OIML R 88 Edition 1998 (E) CONTENTS Foreword...

More information

DEC Country Report. National Institute of Metrology (Thailand) By Mr.Pairoj Rattanangkul

DEC Country Report. National Institute of Metrology (Thailand) By Mr.Pairoj Rattanangkul National Institute of Metrology (Thailand) DEC Country Report By Mr.Pairoj Rattanangkul email : pairoj@nimt.or.th National Institute of Metrology (Thailand) 1 2 Table of Content Brief history of NIMT Where

More information

CCEM KEY COMPARISON CCEM.RF-K18.CL (GT-RF/00-1) Final Report. C. Eiø, D. Adamson, J. Randa, D. Allal and R. Uzdin

CCEM KEY COMPARISON CCEM.RF-K18.CL (GT-RF/00-1) Final Report. C. Eiø, D. Adamson, J. Randa, D. Allal and R. Uzdin CCEM KEY COMPARISON CCEM.RF-K18.CL (GT-RF/00-1) Noise in 50 Ω coaxial line at frequencies up to 1 GHz Final Report C. Eiø, D. Adamson, J. Randa, D. Allal and R. Uzdin Christopher Eiø National Physical

More information

PRODUCT DATA. Sound Intensity Probe Kit Type 3599 Sound Intensity Microphone Pair Type 4197 Dual Preamplifier Type 2683

PRODUCT DATA. Sound Intensity Probe Kit Type 3599 Sound Intensity Microphone Pair Type 4197 Dual Preamplifier Type 2683 PRODUCT DATA Probe Kit Type 3599 Microphone Pair Type 4197 Dual Preamplifier Type 2683 Type 3599 is a two-microphone probe kit for measuring sound intensity. Specially designed for use with Brüel & Kjær

More information

A study of Savitzky-Golay filters for derivatives in primary shock calibration

A study of Savitzky-Golay filters for derivatives in primary shock calibration ACTA IMEKO December 2013, Volume 2, Number 2, 41 47 www.imeko.org A study of Savitzky-Golay filters for derivatives in primary shock calibration Hideaki Nozato 1, Thomas Bruns 2, Henrik Volkers 2, Akihiro

More information

CIPM and CCPR What are these organizations and how do they affect my testing results. Maria Nadal Photometry, Surface Color and Appearance NIST

CIPM and CCPR What are these organizations and how do they affect my testing results. Maria Nadal Photometry, Surface Color and Appearance NIST CIPM and CCPR What are these organizations and how do they affect my testing results Maria Nadal Photometry, Surface Color and Appearance NIST CIE USA Annual Meeting October 6-7, 2014 Calibration Laboratory

More information

CALIBRATION OF THE BEV GPS RECEIVER BY USING TWSTFT

CALIBRATION OF THE BEV GPS RECEIVER BY USING TWSTFT CALIBRATION OF THE BEV GPS RECEIVER BY USING TWSTFT A. Niessner 1, W. Mache 1, B. Blanzano, O. Koudelka, J. Becker 3, D. Piester 3, Z. Jiang 4, and F. Arias 4 1 Bundesamt für Eich- und Vermessungswesen,

More information

PRODUCT DATA. Sound Intensity Probe Kit for 2260 Type 3595 Sound Intensity Microphone Pair Type 4197 Dual Preamplifier Type 2683

PRODUCT DATA. Sound Intensity Probe Kit for 2260 Type 3595 Sound Intensity Microphone Pair Type 4197 Dual Preamplifier Type 2683 PRODUCT DATA Sound Intensity Probe Kit for 2260 Type 3595 Sound Intensity Microphone Pair Type 4197 Dual Preamplifier Type 2683 Type 3595 is a two-microphone probe kit for measuring sound intensity. Specially

More information

19th Meeting of the Directors of NMIs and Member State Representatives with BIPM

19th Meeting of the Directors of NMIs and Member State Representatives with BIPM 19th Meeting of the Directors of NMIs and Member State Representatives with BIPM Session on Advanced Manufacturing, Digitization and Internet of Things Thursday, 19 th October 2017 11:10 to 13:00 Advanced

More information

Public Address Systems

Public Address Systems ISBN 978 0 11792 743 8 Specification No. 15 United Kingdom Civil Aviation Authority Issue: 2 Date: 13 September 2012 This Specification is only directly applicable to those aircraft where Issue 1 of the

More information

DC Voltage Linearity Measurements and DVM Calibration with Conventional and Programmable Josephson Voltage Standards

DC Voltage Linearity Measurements and DVM Calibration with Conventional and Programmable Josephson Voltage Standards 20th IMEKO TC4 International Symposium and 18th International Workshop on ADC Modelling and Testing Research on Electric and Electronic Measurement for the Economic Upturn Benevento, Italy, September 15-17,

More information

Designing practical on-site. on-site calibration protocols for acoustic systems: key elements and pitfalls.

Designing practical on-site. on-site calibration protocols for acoustic systems: key elements and pitfalls. Loughborough University Institutional Repository Designing practical on-site calibration protocols for acoustic systems: key elements and pitfalls This item was submitted to Loughborough University's Institutional

More information

PRODUCT DATA. Applications. Uses

PRODUCT DATA. Applications. Uses PRODUCT DATA Impedance Tube Kit (50 Hz 6.4 khz) Type 4206 Impedance Tube Kit (100 Hz 3.2 khz) Type 4206-A Transmission Loss Tube Kit (50 Hz 6.4 khz) Type 4206-T Brüel & Kjær offers a complete range of

More information

Co-opted Members: Megumi Akoshima (AIST/NMIJ), JooHyun Lee (KRISS), Lindsay Chapman (NPL), Christian Monte (PTB).

Co-opted Members: Megumi Akoshima (AIST/NMIJ), JooHyun Lee (KRISS), Lindsay Chapman (NPL), Christian Monte (PTB). CCT Task Group on Thermophysical Quantities Activity Report, May 2014 May 2017 Report prepared by: Jean-Remy Filtz, LNE Date of Report: 23 May 2017 This report summarized the activities undertaken by CCT-TG-ThQ

More information

PRODUCT DATA. Sound Intensity Calibrator Type Uses and Features

PRODUCT DATA. Sound Intensity Calibrator Type Uses and Features PRODUCT DATA Sound Intensity Calibrator Type 4297 Sound Intensity Calibrator Type 4297 is used for on site sound pressure calibration and pressure residual intensity index verification. Its most important

More information

Supplementary Comparison EUROMET.EM.RF-S25 EURAMET Project 819. Comparison of Electrical Field Strength Measurements above 1 GHz.

Supplementary Comparison EUROMET.EM.RF-S25 EURAMET Project 819. Comparison of Electrical Field Strength Measurements above 1 GHz. EURAMET.EM.RF-S25 Final Report page 1 of 120 Supplementary Comparison EUROMET.EM.RF-S25 EURAMET Project 819 Comparison of Electrical Field Strength Measurements above 1 GHz Final Report Karel Dražil Czech

More information

PRODUCT DATA. Falcon Range 1/4-inch Microphone Preamplifier Type 2670

PRODUCT DATA. Falcon Range 1/4-inch Microphone Preamplifier Type 2670 PRODUCT DATA Falcon Range 1/4-inch Microphone Preamplifier Type 2670 1/4-inch Microphone Preamplifier Type 2670 is a Falcon Range product for precision acoustic measurements with Brüel & Kjær s wide rangf

More information

Calibration of piezoelectric accelerometers at INTI

Calibration of piezoelectric accelerometers at INTI PROCEEDINGS of the 22 nd International Congress on Acoustics Acoustical Measurements and Instrumentation: Paper ICA2016-784 Calibration of piezoelectric accelerometers at INTI Alexis Zapata (a), Ramiro

More information

International Committee. for Weights and Measures. Proceedings of Session I. of the 104th meeting. (9-10 March 2015)

International Committee. for Weights and Measures. Proceedings of Session I. of the 104th meeting. (9-10 March 2015) International Committee for Weights and Measures Proceedings of Session I of the 104th meeting (9-10 March 2015) Executive Summary Session I of the 104th meeting of the CIPM (9-10 March 2015) CIPM bureau

More information

An investigation on diffuse-field calibration of measurement microphones by the reciprocity technique

An investigation on diffuse-field calibration of measurement microphones by the reciprocity technique PROCEEDINGS of the 22 nd International Congress on Acoustics Electroacoustics and Audio Engineering: Paper ICA2016-147 An investigation on diffuse-field calibration of measurement microphones by the reciprocity

More information

CALIBRATION OF LASER VIBROMETER STANDARDS ACCORDING TO ISO

CALIBRATION OF LASER VIBROMETER STANDARDS ACCORDING TO ISO XVIII IMEKO WORLD CONGRESS Metrology for a Sustainable Development September, 17 22, 2006, Rio de Janeiro, Brazil CALIBRATION OF LASER VIBROMETER STANDARDS ACCORDING TO ISO 16063-41 Dr.-Ing. Uwe Buehn

More information

Annex to the Accreditation Certificate D-K according to DIN EN ISO/IEC 17025:2005

Annex to the Accreditation Certificate D-K according to DIN EN ISO/IEC 17025:2005 Deutsche Akkreditierungsstelle GmbH German Accreditation Body Annex the Accreditation Certificate D-K-15183-01-00 according DIN EN ISOIEC 17025:2005 Period of validity: 19.12.2017 18.12.2022 Holder of

More information

SEM Magnification Calibration & Verification: Building Confidence in Your Scale Bar

SEM Magnification Calibration & Verification: Building Confidence in Your Scale Bar SEM Magnification Calibration & Verification: Building Confidence in Your Scale Bar Mark A. Koten, Ph.D. Senior Research Scientist Electron Optics Group McCrone Associates Why check your SEM image calibration?

More information

Keysight Technologies USB Preamplifiers

Keysight Technologies USB Preamplifiers Keysight Technologies USB Preamplifiers U77/A 1 MHz to 4 GHz U77/C 1 MHz to 6. GHz U77/F to GHz Technical Overview Keysight USB Preamplifiers U77A/C/F - Technical Overview Key Features and Benefits Automatic

More information

PRODUCT DATA. PULSE Multichannel Sound Power Type 7748 Version 1.2

PRODUCT DATA. PULSE Multichannel Sound Power Type 7748 Version 1.2 PRODUCT DATA PULSE Multichannel Sound Power Type 7748 Version 1.2 Multichannel Sound Power offers the easiest way yet to measure sound power and other important sound characteristics of products according

More information

NPL REPORT DEPC Engineering Measurement DEPC-EM 002. An investigation into the effect of marking mass standards of OIML class E1

NPL REPORT DEPC Engineering Measurement DEPC-EM 002. An investigation into the effect of marking mass standards of OIML class E1 NPL REPORT DEPC Engineering Measurement DEPC-EM 002 An investigation into the effect of marking mass standards of OIML class E1 J Berry and S P Downes NOT RESTRICTED APRIL 2006 National Physical Laboratory

More information

ISO INTERNATIONAL STANDARD

ISO INTERNATIONAL STANDARD INTERNATIONAL STANDARD ISO 1996-2 Second edition 2007-03-15 Acoustics Description, measurement and assessment of environmental noise Part 2: Determination of environmental noise levels Acoustique Description,

More information

Division of Physical Metrology Center for Fluid Flow & Acoustics Korea Research Institute of Science & Standards Hyu-Sang Kwon & Wan-Ho Cho

Division of Physical Metrology Center for Fluid Flow & Acoustics Korea Research Institute of Science & Standards Hyu-Sang Kwon & Wan-Ho Cho Report on the Bilateral Comparison between KRISS & NIM Division of Physical Metrology Center for Fluid Flow & Acoustics Korea Research Institute of Science & Standards Hyu-Sang Kwon & Wan-Ho Cho Introduction

More information

Certificate of Accreditation

Certificate of Accreditation PERRY JOHNSON LABORATORY ACCREDITATION, INC. Certificate of Accreditation Perry Johnson Laboratory Accreditation, Inc. has assessed the Laboratory of: Doko Bldg., 3-3-4,Minamisuna, Koto-ku, Tokyo 136-8580

More information

ISO INTERNATIONAL STANDARD

ISO INTERNATIONAL STANDARD INTERNATIONAL STANDARD ISO 13472-2 First edition 2010-05-15 Acoustics Measurement of sound absorption properties of road surfaces in situ Part 2: Spot method for reflective surfaces Acoustique Mesurage

More information

BIPM Report to the AFRIMETS General Assembly. Andy Henson Director of International Liaison and Communications

BIPM Report to the AFRIMETS General Assembly. Andy Henson Director of International Liaison and Communications BIPM Report to the AFRIMETS General Assembly Andy Henson Director of International Liaison and Communications Towards a new SI New data published in the latest edition of Metrologia. CCM At least 3 experiments,

More information

2-3 Calibration of Standard Voltage and Current Generator

2-3 Calibration of Standard Voltage and Current Generator 2-3 Calibration of Standard Voltage and Current Generator Katsumi FUJII, Kojiro SAKAI, Tsutomu SUGIYAMA, Kouichi SEBATA, and Iwao NISHIYAMA This paper describes the calibration method of standard voltage

More information

Informations and comments on ECE-TRANS-WP.29-GRB e

Informations and comments on ECE-TRANS-WP.29-GRB e Transmitted by the expert from France Informal document GRB-63-16 (63rd GRB, 16-18 February 2016, agenda item 2) Informations and comments on ECE-TRANS-WP.29-GRB-2016-02e Louis-Ferdinand PARDO (France)

More information

AFRIMETS.EM.RF-S1. Attenuation and reflection measurements for coaxials at 100 MHz, 1 GHz and 10 GHz Type N Connector

AFRIMETS.EM.RF-S1. Attenuation and reflection measurements for coaxials at 100 MHz, 1 GHz and 10 GHz Type N Connector AFRIMETS.EM.RF-S1 Attenuation and reflection measurements for coaxials at 100 MHz, 1 GHz and 10 GHz Type N Connector Main author Linoh Magagula 1 Co-authors Abdelrahman Sallam 3, Abdelkarim MALLAT 2, Nadia

More information

High-Precision Resistance Decade and Calibrator. High-Precision Resistance Decade and Calibrator. Model 1405, 1406, Decades

High-Precision Resistance Decade and Calibrator. High-Precision Resistance Decade and Calibrator. Model 1405, 1406, Decades High-Precision Resistance Decade and Calibrator High-Precision Resistance Decade and Calibrator 2008 burster präzisionsmeßtechnik gmbh & co kg All rights reserved Edition 14.04.2008 Manufacturer: burster

More information

Investigation of Two Different Techniques for Accurate Measurements of Sinusoidal Signals

Investigation of Two Different Techniques for Accurate Measurements of Sinusoidal Signals Investigation of Two Different Techniques for Accurate Measurements of Sinusoidal Signals Shereen M. El-Metwally 1, Mamdouh Halawa 2 1 Department of Systems and Biomedical Engineering, Cairo University,

More information

APMP.AUV.V-K1 Regional Key Comparison of Standard Accelerometer. Final Report. Project Number: APMP-IC Shing Chen and Hans-Jürgen von Martens

APMP.AUV.V-K1 Regional Key Comparison of Standard Accelerometer. Final Report. Project Number: APMP-IC Shing Chen and Hans-Jürgen von Martens Regional Key Comparison of Standard Accelerometer Final Report Project Number: APMP-IC-4-95 Shing Chen and Hans-Jürgen von Martens January 2001 Abstract In project APMP-IC-4-95 (BIPM comparison identifier:

More information

Intercomparison of radiation temperature measurements over the temperature range from 1600 K to 3300 K

Intercomparison of radiation temperature measurements over the temperature range from 1600 K to 3300 K INSTITUTE OF PHYSICS PUBLISHING Metrologia 40 (2003) S39 S44 METROLOGIA PII: S0026-1394(03)56056-X Intercomparison of radiation temperature measurements over the temperature range from 1600 K to 3300 K

More information

Supplementary comparison: APMP.EM.RF-S21.F. Loop antennas (9 khz to 30 MHz) Final Report

Supplementary comparison: APMP.EM.RF-S21.F. Loop antennas (9 khz to 30 MHz) Final Report Supplementary comparison: APMP.EM.RF-S21.F Loop antennas (9 khz to 30 MHz) Final Report Author Masanori Ishii 1 Co-authors Jeong Hwan Kim 2, Yu Ji 3 (formerly), Chi Hyun Cho 2, and Tim Zhang 3 1 National

More information