Imaging Open-Path Fourier Transform Infrared Spectrometer For 3D Cloud Profiling

Size: px
Start display at page:

Download "Imaging Open-Path Fourier Transform Infrared Spectrometer For 3D Cloud Profiling"

Transcription

1 See discussions, stats, and author profiles for this publication at: Imaging Open-Path Fourier Transform Infrared Spectrometer For 3D Cloud Profiling Article in Proceedings of SPIE - The International Society for Optical Engineering October 2007 Impact Factor: 0.2 DOI: / CITATIONS 2 READS 24 7 authors, including: David Mansur Optra Systems 19 PUBLICATIONS 16 CITATIONS Lori A Todd University of North Carolina at Chapel Hill 39 PUBLICATIONS 457 CITATIONS SEE PROFILE SEE PROFILE Kathleen Mottus University of North Carolina at Chapel Hill 14 PUBLICATIONS 164 CITATIONS SEE PROFILE All in-text references underlined in blue are linked to publications on ResearchGate, letting you access and read them immediately. Available from: Kathleen Mottus Retrieved on: 08 May 2016

2 Imaging Open-Path Fourier Transform Infrared Spectrometer For 3D Cloud Profiling Julia Rentz Dupuis, David J. Mansur, Robert Vaillancourt, David Carlson, Thomas Evans, and Elizabeth Schundler OPTRA, Inc. 461 Boston St., Topsfield, MA phone: (978) fax: (978) Lori Todd, Kathleen Mottus University of North Carolina 121 Rosenau Hall, Chapel Hill, NC ABSTRACT OPTRA is developing an imaging open-path Fourier transform infrared (I-OP-FTIR) spectrometer for 3D profiling of chemical and biological agent simulant plumes released into test ranges and chambers. An array of I-OP-FTIR instruments positioned around the perimeter of the test site, in concert with advanced spectroscopic algorithms, enables real time tomographic reconstruction of the plume. The approach is intended as a referee measurement for test ranges and chambers. This Small Business Technology Transfer (STTR) effort combines the instrumentation and spectroscopic capabilities of OPTRA, Inc. with the computed tomographic expertise of the University of North Carolina, Chapel Hill. In this paper, we detail the design and build of a prototype I-OP-FTIR instrument. Key Words: FTIR spectrometer, cloud profiling, tomography 1 INTRODUCTION Under a U.S. Army Small Business Technology Transfer (STTR) solicitation, a need was identified for a near real-time 3D plume concentration profiler system for chemical and biological agent simulants released into test ranges and chambers. The technology developed under this opportunity will be considered as a candidate referee sensor for the test range or chamber. The requirements of the application include the ability to generate time-dependent concentration information with sensitivity better than that of the sensors under test at the range. The referee sensor must be responsive to a number of different chemical and biological agent simulants, and it must be able to operate in the zero-temperature contrast scenario between the plume and the background against which it is being viewed. In response to this opportunity, OPTRA is developing an imaging open-path Fourier transform infrared (I-OP-FTIR) spectrometer which is used with an array of plastic injection molded retroreflector arrays to simultaneously monitor an array of open paths (OPs) for chemical content and concentration by infrared spectroscopy. Via the optical design presented in this paper, from a single FTIR modulator the I-OP-FTIR emits an array of discrete OP beams to respective retroreflector arrays which return the beams to the instrument for multiplexed detection and processing. Our present design emits an array of eleven OP beams spread over a 30 fan (i.e. spaced at 3 increments). All beams operate over 7 to 14 μm (1428 to 714 cm -1 ) with a spectral resolution of 4 cm -1. Unlike previous work employing a single channel OP- FTIR with a scanner 1, the OPTRA approach simultaneously monitors an angularly resolved region thereby eliminating errors due to the movement of the plume over the duration of the single channel OP-FTIR scan. Unlike previous work using passive hyperspectral imaging 2, the OPTRA approach is quantitative without the need for background, plume, and air column radiance estimation; the OPTRA approach is also valid in the zero temperature contrast scenario where passive IR spectrometers, radiometers, and imagers cannot detect the plume. With a number of I-OP-FTIR systems strategically positioned around the test range or chamber (Figure 1) in concert with spectroscopic multicomponent algorithms, an accurate tomographic reconstruction of the plume concentration can Copyright 2009 Society of Photo-Optical Instrumentation Engineers. This paper will be published in The Proceedings of Chemical, Biological, Radiological, Nuclear, and Explosives (CBRNE) Sensing X and is made available as an electronic preprint with permission of SPIE. One print or electronic copy may be made for personal use only. Systematic or multiple reproduction, distribution to multiple locations via electronic or other means, duplication of any material in this paper for a fee or for commercial purposes, or modification of the content of the paper are prohibited.

3 be made. Under this STTR effort, OPTRA has teamed with University of North Carolina, Chapel Hill, which brings tomographic expertise for this IR spectroscopic monitoring application. Table 1 details the specifications for the I-OP-FTIR system. The initial application of this work will be a point sensor test range with the associated requirements listed in Table 1. Future work may scale the range of our instrument to on the order of 1 km test grids for standoff sensor test ranges. Figure 1: System Concept Table 1: System Performance Specifications REQUIREMENT VALUE OPTRA s I-OP-FTIR (x3) Noise equivalent concentration < 0.1 mg/m 3 Maximum concentration 20 mg/m 3 Concentration accuracy ± 10% Update rate < 5 sec Test grid size 100x100x10 m Grid spatial resolution 2.5 m Plastic Retroreflector Arrays plume OP IR beams Test range Figure 1 (left) shows the system concept. Each I-OP- FTIR emits an array of discrete IR beams which are projected through the plume to remotely located retroreflector arrays and back again through the plume to the instrument for multiplexed detection and processing. This approach allows for simultaneous interrogation of the plume unlike scanner-based systems. This approach also supports detection in the zero temperature contrast scenario unlike passive IR spectrometers. The following contains the details of our prototype I-OP-FTIR design and build. 2 DESCRIPTION OF TECHNICAL APPROACH 2.1 Multichannel Concept Our approach is similar to a traditional monostatic OP-FTIR in that we employ a single IR source from which IR radiation is coupled by a set of lenses and directed through a Michelson interferometer before being expanded and collimated by a projector optic and directed to a remotely located retroreflector array. The retroreflector array then returns the OP beam to the sensor for detection. However, in this case, we couple the radiation exiting the interferometer into an IR fiber optic assembly composed of eleven IR fibers bundled at the input end and free at the output end. The free/output ends of these fibers are then positioned along the focal plane of the projector optic. The result is eleven OP beams whose divergence is determined by a throughput match between the fiber diameter and numerical aperture (NA) with the projector aperture. The pointing of the OP beams is set by the off-axis location of each fiber relative to the optical axis of the projector lens. Each beam returned by the respective retroreflector array is then focused by the projector lens and folded onto a respective detector using an intensity beamsplitter. 2.2 Radiometric Projections The radiometric model is based on a per-channel noise equivalent spectral radiance (NESR) calculation along with projected signal to noise (SNR) and a noise equivalent concentration (NEC) for a short list of compounds of interest to the sponsor. The radiometric model takes into account the off axis performance of the projector lens and associated radiometric efficiency due to overfilling the retroreflector arrays which is a function of range. The model also takes into account the performance of the retroreflector arrays as well as atmospheric attenuation. NESR is given by: Ad W NESR = [ = ] (1) 2 1 D * η Θ Δσ Δt cm ster cm where each of these values are detailed in Table 2. Radiometric signal to noise (SNR) is given by - 2 -

4 εsource N( σ,tsource) SNR = [ = ]unitless (2) NESR where N(σ,T source ) is the Planck equation for IR source temperature, T source, and ε source is the source emissivity (unitless). Noise equivalent concentration is given by 1 1 mg NEC = ln 1 [ = ] (3) 3 α L SNR m where α and L are the line strength (m 2 /mg) and pathlength (m) respectively. Figures 2a and 2b show the projected per-channel SNR and NEC for m-cresol which is the weakest absorber on the simulant list as a function of range assuming 1 m retroreflector arrays are used. Note that the other simulants of interest to the army are sulfur hexafluoride, acetic acid, triethyl phosphate, and methyl salicylate. Values for ε source and T source are 0.9 and 1450 C (1723 K), respectively. We assume an absorption line strength of m 2 /mg for m-cresol. We assume the plume fills half of the standoff range (i.e. pathlength = range owing to double pass). Table 2: Radiometric Projection Variables VARIABLE DESCRIPTION VALUE Δσ (cm -1 ) Spectral resolution 4 d int (cm) Interferometer mirror diameter 1.27 u int (rad) * Interferometer ½ angle.016 Ω int (ster) ** Interferometer solid angle Θ (cm 2 ster) ** Etendue D* (cm Hz/W) *** Detectivity d tel (cm) Telescope aperture diameter 15 η (unitless) **** Radiometric efficiency η(r) A D (cm 2 ) *** Detector area Δt (s) Integration time 2.5 Notes for Table 2: * This angle is set by the throughput of the fibers (900 μm core, 0.25 NA) matched to that of the interferometer aperture diameter (1.27 cm) and is below the obliquity limit associated with 4 cm -1 spectral resolution and the highest optical frequency of 1428 cm -1 (7 μm). Ω int = 2π(1-cos(u int )) and Θ int = A int Ω int where A int = 1.27 cm 2. *** Eleven 1 mm uncooled lithium tantalite (LiTaO 3 ) single element detectors are used. **** Radiometric efficiency is based on a surface and vignetting loss analysis of the I-OP-FTIR as well as an analysis of the range dependent losses of the retroreflector arrays. Figure 2c shows the results of a model predicting the retroreflector performance as a function of range taking into account the following: 1. Geometric efficiency (geo) the efficiency associated with each point emitted at the sensor aperture returning on itself with a diameter equal to twice the retroreflector element diameter of 1.9 cm. 2. Edge efficiency (edge) the efficiency associated with edge and bevel losses (assumes.001 edges and bevels) 3. Angular accuracy efficiency (ang. acc.) the efficiency associated with the angular accuracy of the retroreflector elements. Note that this effect dominates diffraction (at 10 μm) for the plastic injection molded retroreflector arrays. 4. Fill efficiency the efficiency associated with the geometry of the retroreflector elements which have triangular facets and an associated 66% radiometric efficiency. 5. Atmospheric losses the efficiency associated with water vapor and carbon dioxide absorption in the atmosphere. This projection uses Hitran 3 reference spectra and integrates over the 7 to 14 μm spectral range. 6. Overfill efficiency the efficiency associated with overfilling the retroreflector array with the open path beam

5 Figure 2a: SNR vs. Range Figure 2b: NEC vs. Range for M-Cresol Figure 2c: Retroreflector Array Efficiency vs. Range According to our radiometric model and based on our pathlength assumptions, all but the two outside channels (± 15 ) meet the NEC requirement of 0.1 mg/m 3 for m-cresol (the weakest absorber) over the full range of standoffs between 0 and 100 m. The requirement is fully met for standoffs up to 80 m and for all of the other chemical agent simulants of interest over the full range of standoffs between 0 and 100 m. 2.3 Optical Design Figure 3a shows the optical layout for the prototype I-OP-FTIR. We show one arm of the interferometer where the second arm is opto-mechanically identical. Radiation emitted by the IR source is coupled into the interferometer using a doublet lens to form an image of the source on the interferometer mirrors. The radiation traverses and exits the interferometer and is focused to a field point at the common end of the fiber optic bundle (also called the fiber optic image transformer). Each fiber therefore samples radiation from every point on the interferometer mirrors over an angular subtense which is below the obliquity limit associated with the spectral resolution and highest optical frequency of this system. Each fiber then transmits its respective portion of the total radiation to a point along the curved focal plane of the projection lens where it passes through an intensity beamsplitter prior to illuminating the projector lens. The result is an array of quasi-collimated beams transmitted to the remotely located retroreflector arrays where the divergence of each beam is dictated by the throughput match between the 900 μm, f/2 fibers and the 15 cm diameter projector lens. In other words, the divergence is about d fiber u fiber /d proj = (900 μm) (0.223 rad)/(15 cm) = 1.3 mrad where rad is the half angle associated with the f/2 fibers. The pointing of the beams (i.e. the ± 15 spread) is dictated by the off-axis location of each fiber. All of the lenses shown in Figure 3a are germanium with a broadband antireflective (AR) coating for 7-14 μm. The secondary beamsplitters are also germanium with a 50% intensity beamsplitter coating for 45 incidence (with a broadband AR coating on the other side). The interferometer mirrors are diamond turned aluminum, and the interferometer beamsplitter/compensator pair are zinc selenide

6 IR Source Source Lenses Figure 3a: Optical System Layout 11x Secondary Beamsplitters Exit Lens Interferometer Mirror Beamsplitter/ Compensator Projection Lens The fiber optic image transformer (Figure 3b) is composed of eleven silver halide IR fibers which are bundled at one end and free at the other. The image shows all fiber ends covered because exposure to ultraviolet (UV) light will degrade the IR transmission. The fiber cores are 900 μm, and the NA is 0.25 (approximately f/2). The purpose of the image transformer is to evenly distribute the modulated radiation exiting the interferometer into an array of (modulated) sources spread along the curved focal plane of the projector lens. The off-axis location of each source relative to the optical axis of the projector lens determines the pointing of each OP beam within the fan. Figure 3b: Free Output End Common Input End The projector lens is a germanium asphere designed for optimal off-axis collimation (±15 ) at f/2 set by the fibers. Figure 3c shows the spot sizes at a range of 100 m for on-axis and 3, 6, 9, 12, and 15 off-axis where the scale bar is 1 m. We extracted radiometric efficiency predictions used in the radiometric projections (Section 2.2) from these spot diagrams. 2.4 Mechanical Design Figure 4a is a solid model showing the mechanical design of the I-OP-FTIR. Figures 4b and 4c show close-ups of the interferometer and mounting for the fiber optic image transformer from two views. 1 m 1 m Figure 3c: Spot Diagram at Retroreflector Arrays 15 off-axis 12 off-axis 9 off-axis 1 m 1 m 6 off-axis 3 off-axis on-axis 1 m 1 m - 5 -

7 Figure 4a: Solid Model of I-OP-FTIR Sighting Scope Cover Projector Lens Mount Image Mount DAQ Terminal Board Interferometer Heat fins for IR source Figure 4a: Solid Model of I-OP-FTIR Figure 4a is a solid model of the I-OP-FTIR mechanical system. The interferometer, fiber optic image transformer mount, projector lens, detectors with amplifiers, and terminal board for the data acquisition are all housed in the sensor head. The fiber optic bundle is not shown in this model. Electronics to control the interferometer as well as the data acquisition are located on PCI boards housed by a lunchbox PC. The lunchbox PC also contains the power supply for the IR source. The mechanical design allows for adequate (passive) heat dissipation for the IR source

8 Figure 4b: Solid Model Interferometer View Figure 4c: Solid Model Image View Fiber Bundle Launch Mount Mount 11x LiTaO3 Detectors IR Source with Heat Sink Detector Preamp Boards (x2) Moving Mirror Interferometer Encoder Reference Figure 4b: Solid Model Interferometer View Figure 4b shows a close-up of the interferometer which employs an encoder-based reference system. The encoder-based interferometer is a variant of our J-Series modulator 4 developed for the Joint Services Lightweight Standoff Chemical Agent Detector (JSLSCAD). The IR portion of the interferometer is identical to the J-Series with diamond turned aluminum mirrors and a zinc selenide beamsplitter/compensator pair. 2.5 System Build and Integration Figure 5a is a photograph of the partially-integrated I- OP-FTIR system (without detectors) showing the germanium projector lens and a laser sighting system that aids in the alignment of the retroreflector arrays. The laser sighting system rotates and snaps into eleven fixed positions every 3 (corresponding to the pointing of the eleven OP beams). The fiber optic image transformer mount is shown in the background. Figure 5b shows a side view of the I-OP-FTIR. This shows the relative position of the projector lens to the fiber optic image transformer (operating at f/2). The fiber optic image transformer is shown on the right where the common/input end is held at the field point at the exit port of the interferometer by a tip/tilt mount with x, y, and z adjustment. The free/output ends are held in their bores by set screws and ultimately a UV epoxy. The eleven secondary beamsplitters are germanium and therefore seal off the free/output ends from all visible and UV light which would otherwise damage the fibers. The eleven LiTaO 3 signal detectors Fiber Bundle Launch Mount Figure 4b: Solid Model Image View Figure 4c shows a close-up of the image transformer mount which holds the eleven free/output fiber ends, eleven secondary beamsplitters, and eleven LiTaO 3 detectors. The free fiber ends are plane ferrules which can be adjusted in z within the mount (relative to the projector lens) to set the collimation of each OP beam. Not shown is an initialization detector opposite the center LiTaO 3 detector. The initialization detector samples the interferogram from the first secondary beamsplitter reflection and does not require the retroreflector to be in place. This detector is used to both initialize the scan as well as check the alignment of the interferometer. Figure 5a: Integrated I-OP-FTIR Front View Laser Sight Germanium Projector Lens Mount - 7 -

9 are mounted above the secondary beamsplitters, and the single initialization detector (not shown) is mounted below the secondary beamsplitter opposite the center channel signal detector. This photo also shows the two detector preamplifier boards and the IR source heat sink. Figure 5b: Integrated I-OP-FTIR Side View Laser Sight Germanium Projector Lens 11x LiTaO3 Signal Detectors Mount Fiber Optic Image Detector Preamp Boards Launch Mount IR Source Heat Sink Figure 5c is a close-up of the fiber optic image transformer and mount before the detectors are mounted. This figure shows the curved profile of the mount which effectively allows us to negate the effects of field curvature of the projector lens on the collimation of the OP beams. Figure 5c: Close-up Mount Interferometer 11x LiTaO3 Signal Detectors IR Source Heat Sink Launch Mount Figure 5d is a close up of the interferometer showing the encoder and moving and stationary mirrors. The lens assemblies which image the source onto the interferometer mirrors and focus the interferometer output to a field point are enclosed in aluminum mounts. The IR source is a miniature igniter manufactured by St. Gobain

10 Figure 5d: Interferometer Close-up IR Source Source Imaging Lenses Encoder Reference Launch Mount Moving Mirror Stationary Mirror Exit Lens The fiber optic image transformer was aligned as follows. With the IR source aligned to the interferometer and the interferometer itself aligned, the common/input end of the fiber optic image transformer was aligned to the field point by serially monitoring the peak-to-peak interferogram amplitude (on an oscilloscope) of the energy exiting each fiber and adjusting the common end in x, y, z, θ x, and θ y until all were optimized. Note that the alignment of the common end is re-optimized as a final step once all detectors are in place and we have the ability to simultaneously monitor all eleven channels. The free/output fiber ends were then positioned for collimation relative to the projector lens using an IR camera focused at infinity to look back through the projector lens at each source. The fiber ends were each in turn adjusted in z to minimize the spot imaged by the IR camera. Figures 6a through 6k show these images for the eleven channels. The image of the 900 μm fiber formed by the 150 mm focal length IR camera lens should be about d image = d fiber F IR camera /F proj = (900 μm) (150 mm)/(360 mm) = 375 μm or about seven or eight, 50 μm IR camera pixels (the camera format is pixels). Following this step, a retroreflector array was then positioned in fronts of the I-OP- FTIR, and each detector was brought into focus by adjusting each in x, y, and z to optimize the peak-to-peak interferogram amplitude (Figure 7). The I-OP-FTIR was rotated to direct each beam in turn onto the retroreflector array for each detector alignment. Note the interferogram shown in Figure 7 was acquired at a much higher mirror velocity and associated modulation frequency than will be used in operation (to provide a constant signal); because of the detector bandwidth, a much cleaner interferogram will result at the operating mirror velocity. The interferogram shown in Figure 7 is also acquired with open-loop scanning. Once all eleven channels were aligned and the detectors wired to the data acquisition system, a final optimization was made by adjusting the common end of the fiber optic image transformer while simultaneously monitoring all eleven detector outputs

11 Figure 6a: -15º OP Beam Figure 6b: -12º OP Beam Figure 6c: -9º OP Beam Figure 6d: -6º OP Beam Figure 6e: -3º OP Beam Figure 6f: 0º OP Beam Figure 6g: 3º OP Beam Figure 6h: 6º OP Beam Figure 6i: 9º OP Beam Figure 6j: 12º OP Beam Figure 6k: 15º OP Beam Figure 7: Interferogram Used for Alignment 2.6 Next Steps At this point in time the I-OP-FTIR is fully integrated and is awaiting test. System characterization will include SNR measurements over a series of standoff ranges up to 100 m as well as spectral resolution using a 10.6 μm carbon dioxide laser. The interferometer itself has already been previously characterized for sampling errors, etc. Following system characterization, a library or background spectra will be acquired for the purpose of training the automated multicomponent algorithms 6 which will be applied to identify and quantify the chemical agent simulants of interest to the customer as well as a short list of compounds we will test at OPTRA. We will then perform a short series of evaluation to verify the multicomponent algorithms are working properly prior to shipping the instrument to University of North Carolina for a full tomographic demonstration

12 3 SUMMARY AND FUTURE PLANS In summary, we have presented system level, optical, and mechanical designs of the I-OP-FTIR system for 3D cloud profiling of chemical agent simulant plumes at test ranges and chambers. We have completed the build and integration of an I-OP-FTIR prototype and will begin testing shortly. Test plans include characterization of SNR at a series of standoff ranges as well as spectral resolution. We will perform a series of spectroscopic multicomponent demonstrations at OPTRA before sending the prototype to University of North Carolina for a full tomographic demonstration. University of North Carolina will demonstrate the tomography using a small array of Teflon balloons they have developed specifically for this purpose to serve as stationary plumes which can be filled to known concentrations with various chemical agent simulants. The I-OP-FTIR prototype will be serially positioned around the balloons and spectra will be collected at each position (simulating multiple instruments). The results along with sensor geometry will then be used to tomographically reconstruct the stationary plumes for each of multiple compounds. Future direction for this work may include extended standoff range operation as well as processing upgrades to handle the real time requirements which were not addressed during the Phase II. Acknowledgements This research was conducted under a Small Business Technology Transfer Phase II contract funded by the U.S. Army. The technical monitor is Dr. Alan Samuels, Edgewood Chemical Biological Center. References 1 L. A. Todd, S. K. Farhat, K. M. Mottus, and G. J. Mihlan, Experimental Evaluation of an Environmental CAT Scanning System for Mapping Chemicals in Air in Real-Time, Appl. Occupational and Environmental Hygiene, Vol. 16, No. 1, (2001). 2 B. R. Cosofret, D. Konno, A. Faghfouri, H. S. Kindle, C. M. Gittins, M. L. Finson, T. E. Janov, M. J. Levreault, R. K. Miyashiro, and W. J. Marinelli, Imaging Sensor Constellation for Tomographic Chemical Cloud Mapping, Appl. Opt. Vol. 48, No. 10, (2009). 3 Ontar Corporation, North Andover, MA. 4 J. R. Engel, J. H. Rentz, and D. L. Carlson, FTIR modulator for first responder applications, Proc. SPIE Vol. 5268, Chemical and Biological Standoff Detection, Photonics East, Providence, RI, (2003). 5 World Wide Web < 6 J. R. Dupuis, D. J. Mansur, J. R. Engel, R. Vaillancourt, L. Todd, and K. Mottus Imaging Open-Path Fourier Transform Spectrometer for 3D Cloud Profiling, Proc. SPIE Vol. 6954, Chemical, Biological, Radiological, Nuclear, and Explosives (CBRNE) Sensing IX, Defense and Security, Orlando, FL, (2008)

Observational Astronomy

Observational Astronomy Observational Astronomy Instruments The telescope- instruments combination forms a tightly coupled system: Telescope = collecting photons and forming an image Instruments = registering and analyzing the

More information

Progress in Standoff Surface Contaminant Detector Platform

Progress in Standoff Surface Contaminant Detector Platform Physical Sciences Inc. Progress in Standoff Surface Contaminant Detector Platform Julia R. Dupuis, Jay Giblin, John Dixon, Joel Hensley, David Mansur, and William J. Marinelli 20 New England Business Center,

More information

Development and Applications of a Sample Compartment FTIR Microscope

Development and Applications of a Sample Compartment FTIR Microscope Application Note Development and Applications of a Sample Since the early to mid-1940 s, scientists using infrared spectroscopy have been trying to obtain spectral data from ever smaller samples. Starting

More information

SIGIS 2. Innovation with Integrity. Long Distance Identification, Visualization and Quantification of Gases FTIR

SIGIS 2. Innovation with Integrity. Long Distance Identification, Visualization and Quantification of Gases FTIR SIGIS 2 Long Distance Identification, Visualization and Quantification of Gases Innovation with Integrity FTIR SIGIS 2 The SIGIS 2 is a scanning imaging remote Key Features sensing system based on the

More information

Spatially Resolved Backscatter Ceilometer

Spatially Resolved Backscatter Ceilometer Spatially Resolved Backscatter Ceilometer Design Team Hiba Fareed, Nicholas Paradiso, Evan Perillo, Michael Tahan Design Advisor Prof. Gregory Kowalski Sponsor, Spectral Sciences Inc. Steve Richstmeier,

More information

SIGIS 2. Innovation with Integrity. Long Distance Identification, Visualization and Quantification of Gases FT-IR

SIGIS 2. Innovation with Integrity. Long Distance Identification, Visualization and Quantification of Gases FT-IR SIGIS 2 Long Distance Identification, Visualization and Quantification of Gases Innovation with Integrity FT-IR The SIGIS 2 is a scanning imaging remote sensing system based on the combination of an infrared

More information

Physics 431 Final Exam Examples (3:00-5:00 pm 12/16/2009) TIME ALLOTTED: 120 MINUTES Name: Signature:

Physics 431 Final Exam Examples (3:00-5:00 pm 12/16/2009) TIME ALLOTTED: 120 MINUTES Name: Signature: Physics 431 Final Exam Examples (3:00-5:00 pm 12/16/2009) TIME ALLOTTED: 120 MINUTES Name: PID: Signature: CLOSED BOOK. TWO 8 1/2 X 11 SHEET OF NOTES (double sided is allowed), AND SCIENTIFIC POCKET CALCULATOR

More information

Applications of Steady-state Multichannel Spectroscopy in the Visible and NIR Spectral Region

Applications of Steady-state Multichannel Spectroscopy in the Visible and NIR Spectral Region Feature Article JY Division I nformation Optical Spectroscopy Applications of Steady-state Multichannel Spectroscopy in the Visible and NIR Spectral Region Raymond Pini, Salvatore Atzeni Abstract Multichannel

More information

Improving the Collection Efficiency of Raman Scattering

Improving the Collection Efficiency of Raman Scattering PERFORMANCE Unparalleled signal-to-noise ratio with diffraction-limited spectral and imaging resolution Deep-cooled CCD with excelon sensor technology Aberration-free optical design for uniform high resolution

More information

MR-i. Hyperspectral Imaging FT-Spectroradiometers Radiometric Accuracy for Infrared Signature Measurements

MR-i. Hyperspectral Imaging FT-Spectroradiometers Radiometric Accuracy for Infrared Signature Measurements MR-i Hyperspectral Imaging FT-Spectroradiometers Radiometric Accuracy for Infrared Signature Measurements FT-IR Spectroradiometry Applications Spectroradiometry applications From scientific research to

More information

Real-Time Scanning Goniometric Radiometer for Rapid Characterization of Laser Diodes and VCSELs

Real-Time Scanning Goniometric Radiometer for Rapid Characterization of Laser Diodes and VCSELs Real-Time Scanning Goniometric Radiometer for Rapid Characterization of Laser Diodes and VCSELs Jeffrey L. Guttman, John M. Fleischer, and Allen M. Cary Photon, Inc. 6860 Santa Teresa Blvd., San Jose,

More information

MR-i. Hyperspectral Imaging FT-Spectroradiometers Radiometric Accuracy for Infrared Signature Measurements

MR-i. Hyperspectral Imaging FT-Spectroradiometers Radiometric Accuracy for Infrared Signature Measurements MR-i Hyperspectral Imaging FT-Spectroradiometers Radiometric Accuracy for Infrared Signature Measurements FT-IR Spectroradiometry Applications Spectroradiometry applications From scientific research to

More information

EE119 Introduction to Optical Engineering Spring 2003 Final Exam. Name:

EE119 Introduction to Optical Engineering Spring 2003 Final Exam. Name: EE119 Introduction to Optical Engineering Spring 2003 Final Exam Name: SID: CLOSED BOOK. THREE 8 1/2 X 11 SHEETS OF NOTES, AND SCIENTIFIC POCKET CALCULATOR PERMITTED. TIME ALLOTTED: 180 MINUTES Fundamental

More information

A novel tunable diode laser using volume holographic gratings

A novel tunable diode laser using volume holographic gratings A novel tunable diode laser using volume holographic gratings Christophe Moser *, Lawrence Ho and Frank Havermeyer Ondax, Inc. 85 E. Duarte Road, Monrovia, CA 9116, USA ABSTRACT We have developed a self-aligned

More information

Submillimeter Pupil-Plane Wavefront Sensing

Submillimeter Pupil-Plane Wavefront Sensing Submillimeter Pupil-Plane Wavefront Sensing E. Serabyn and J.K. Wallace Jet Propulsion Laboratory, 4800 Oak Grove Drive, California Institute of Technology, Pasadena, CA, 91109, USA Copyright 2010 Society

More information

Applications of Optics

Applications of Optics Nicholas J. Giordano www.cengage.com/physics/giordano Chapter 26 Applications of Optics Marilyn Akins, PhD Broome Community College Applications of Optics Many devices are based on the principles of optics

More information

Digital Camera Technologies for Scientific Bio-Imaging. Part 2: Sampling and Signal

Digital Camera Technologies for Scientific Bio-Imaging. Part 2: Sampling and Signal Digital Camera Technologies for Scientific Bio-Imaging. Part 2: Sampling and Signal Yashvinder Sabharwal, 1 James Joubert 2 and Deepak Sharma 2 1. Solexis Advisors LLC, Austin, TX, USA 2. Photometrics

More information

Chemistry 524--"Hour Exam"--Keiderling Mar. 19, pm SES

Chemistry 524--Hour Exam--Keiderling Mar. 19, pm SES Chemistry 524--"Hour Exam"--Keiderling Mar. 19, 2013 -- 2-4 pm -- 170 SES Please answer all questions in the answer book provided. Calculators, rulers, pens and pencils permitted. No open books allowed.

More information

ECEN. Spectroscopy. Lab 8. copy. constituents HOMEWORK PR. Figure. 1. Layout of. of the

ECEN. Spectroscopy. Lab 8. copy. constituents HOMEWORK PR. Figure. 1. Layout of. of the ECEN 4606 Lab 8 Spectroscopy SUMMARY: ROBLEM 1: Pedrotti 3 12-10. In this lab, you will design, build and test an optical spectrum analyzer and use it for both absorption and emission spectroscopy. The

More information

WHITE PAPER MINIATURIZED HYPERSPECTRAL CAMERA FOR THE INFRARED MOLECULAR FINGERPRINT REGION

WHITE PAPER MINIATURIZED HYPERSPECTRAL CAMERA FOR THE INFRARED MOLECULAR FINGERPRINT REGION WHITE PAPER MINIATURIZED HYPERSPECTRAL CAMERA FOR THE INFRARED MOLECULAR FINGERPRINT REGION Denis Dufour, David Béland, Hélène Spisser, Loïc Le Noc, Francis Picard, Patrice Topart January 2018 Low-cost

More information

PHY 431 Homework Set #5 Due Nov. 20 at the start of class

PHY 431 Homework Set #5 Due Nov. 20 at the start of class PHY 431 Homework Set #5 Due Nov. 0 at the start of class 1) Newton s rings (10%) The radius of curvature of the convex surface of a plano-convex lens is 30 cm. The lens is placed with its convex side down

More information

Compact Dual Field-of-View Telescope for Small Satellite Payloads

Compact Dual Field-of-View Telescope for Small Satellite Payloads Compact Dual Field-of-View Telescope for Small Satellite Payloads James C. Peterson Space Dynamics Laboratory 1695 North Research Park Way, North Logan, UT 84341; 435-797-4624 Jim.Peterson@sdl.usu.edu

More information

Enhanced Chemical Identification Using High-Throughput Virtual-Slit Enabled Optical Spectroscopy and Hyperspectral Imaging

Enhanced Chemical Identification Using High-Throughput Virtual-Slit Enabled Optical Spectroscopy and Hyperspectral Imaging Enhanced Chemical Identification Using High-Throughput Virtual-Slit Enabled Optical Spectroscopy and Hyperspectral Imaging tornado-spectral.com INTRODUCTION There is a growing opportunity for the use of

More information

Department of Electrical Engineering and Computer Science

Department of Electrical Engineering and Computer Science MASSACHUSETTS INSTITUTE of TECHNOLOGY Department of Electrical Engineering and Computer Science 6.161/6637 Practice Quiz 2 Issued X:XXpm 4/XX/2004 Spring Term, 2004 Due X:XX+1:30pm 4/XX/2004 Please utilize

More information

Spotlight 150 and 200 FT-IR Microscopy Systems

Spotlight 150 and 200 FT-IR Microscopy Systems S P E C I F I C A T I O N S Spotlight 150 and 200 FT-IR Microscopy Systems FT-IR Microscopy Spotlight 200 with Frontier FT-IR Spectrometer Introduction PerkinElmer Spotlight FT-IR Microscopy Systems are

More information

Supplementary Materials

Supplementary Materials Supplementary Materials In the supplementary materials of this paper we discuss some practical consideration for alignment of optical components to help unexperienced users to achieve a high performance

More information

3.0 Alignment Equipment and Diagnostic Tools:

3.0 Alignment Equipment and Diagnostic Tools: 3.0 Alignment Equipment and Diagnostic Tools: Alignment equipment The alignment telescope and its use The laser autostigmatic cube (LACI) interferometer A pin -- and how to find the center of curvature

More information

MTF characteristics of a Scophony scene projector. Eric Schildwachter

MTF characteristics of a Scophony scene projector. Eric Schildwachter MTF characteristics of a Scophony scene projector. Eric Schildwachter Martin MarieUa Electronics, Information & Missiles Systems P0 Box 555837, Orlando, Florida 32855-5837 Glenn Boreman University of Central

More information

AgilOptics mirrors increase coupling efficiency into a 4 µm diameter fiber by 750%.

AgilOptics mirrors increase coupling efficiency into a 4 µm diameter fiber by 750%. Application Note AN004: Fiber Coupling Improvement Introduction AgilOptics mirrors increase coupling efficiency into a 4 µm diameter fiber by 750%. Industrial lasers used for cutting, welding, drilling,

More information

Using molded chalcogenide glass technology to reduce cost in a compact wide-angle thermal imaging lens

Using molded chalcogenide glass technology to reduce cost in a compact wide-angle thermal imaging lens Using molded chalcogenide glass technology to reduce cost in a compact wide-angle thermal imaging lens George Curatu a, Brent Binkley a, David Tinch a, and Costin Curatu b a LightPath Technologies, 2603

More information

Presented by Jerry Hubbell Lake of the Woods Observatory (MPC I24) President, Rappahannock Astronomy Club

Presented by Jerry Hubbell Lake of the Woods Observatory (MPC I24) President, Rappahannock Astronomy Club Presented by Jerry Hubbell Lake of the Woods Observatory (MPC I24) President, Rappahannock Astronomy Club ENGINEERING A FIBER-FED FED SPECTROMETER FOR ASTRONOMICAL USE Objectives Discuss the engineering

More information

Physics 3340 Spring Fourier Optics

Physics 3340 Spring Fourier Optics Physics 3340 Spring 011 Purpose Fourier Optics In this experiment we will show how the Fraunhofer diffraction pattern or spatial Fourier transform of an object can be observed within an optical system.

More information

IMAGE SENSOR SOLUTIONS. KAC-96-1/5" Lens Kit. KODAK KAC-96-1/5" Lens Kit. for use with the KODAK CMOS Image Sensors. November 2004 Revision 2

IMAGE SENSOR SOLUTIONS. KAC-96-1/5 Lens Kit. KODAK KAC-96-1/5 Lens Kit. for use with the KODAK CMOS Image Sensors. November 2004 Revision 2 KODAK for use with the KODAK CMOS Image Sensors November 2004 Revision 2 1.1 Introduction Choosing the right lens is a critical aspect of designing an imaging system. Typically the trade off between image

More information

Performance Comparison of Spectrometers Featuring On-Axis and Off-Axis Grating Rotation

Performance Comparison of Spectrometers Featuring On-Axis and Off-Axis Grating Rotation Performance Comparison of Spectrometers Featuring On-Axis and Off-Axis Rotation By: Michael Case and Roy Grayzel, Acton Research Corporation Introduction The majority of modern spectrographs and scanning

More information

OCT Spectrometer Design Understanding roll-off to achieve the clearest images

OCT Spectrometer Design Understanding roll-off to achieve the clearest images OCT Spectrometer Design Understanding roll-off to achieve the clearest images Building a high-performance spectrometer for OCT imaging requires a deep understanding of the finer points of both OCT theory

More information

Wavelength Stabilization of HPDL Array Fast-Axis Collimation Optic with integrated VHG

Wavelength Stabilization of HPDL Array Fast-Axis Collimation Optic with integrated VHG Wavelength Stabilization of HPDL Array Fast-Axis Collimation Optic with integrated VHG C. Schnitzler a, S. Hambuecker a, O. Ruebenach a, V. Sinhoff a, G. Steckman b, L. West b, C. Wessling c, D. Hoffmann

More information

3550 Aberdeen Ave SE, Kirtland AFB, NM 87117, USA ABSTRACT 1. INTRODUCTION

3550 Aberdeen Ave SE, Kirtland AFB, NM 87117, USA ABSTRACT 1. INTRODUCTION Beam Combination of Multiple Vertical External Cavity Surface Emitting Lasers via Volume Bragg Gratings Chunte A. Lu* a, William P. Roach a, Genesh Balakrishnan b, Alexander R. Albrecht b, Jerome V. Moloney

More information

HR2000+ Spectrometer. User-Configured for Flexibility. now with. Spectrometers

HR2000+ Spectrometer. User-Configured for Flexibility. now with. Spectrometers Spectrometers HR2000+ Spectrometer User-Configured for Flexibility HR2000+ One of our most popular items, the HR2000+ Spectrometer features a high-resolution optical bench, a powerful 2-MHz analog-to-digital

More information

!!! DELIVERABLE!D60.2!

!!! DELIVERABLE!D60.2! www.solarnet-east.eu This project is supported by the European Commission s FP7 Capacities Programme for the period April 2013 - March 2017 under the Grant Agreement number 312495. DELIVERABLED60.2 Image

More information

ECEN 4606, UNDERGRADUATE OPTICS LAB

ECEN 4606, UNDERGRADUATE OPTICS LAB ECEN 4606, UNDERGRADUATE OPTICS LAB Lab 2: Imaging 1 the Telescope Original Version: Prof. McLeod SUMMARY: In this lab you will become familiar with the use of one or more lenses to create images of distant

More information

OPAL Optical Profiling of the Atmospheric Limb

OPAL Optical Profiling of the Atmospheric Limb OPAL Optical Profiling of the Atmospheric Limb Alan Marchant Chad Fish Erik Stromberg Charles Swenson Jim Peterson OPAL STEADE Mission Storm Time Energy & Dynamics Explorers NASA Mission of Opportunity

More information

746A27 Remote Sensing and GIS. Multi spectral, thermal and hyper spectral sensing and usage

746A27 Remote Sensing and GIS. Multi spectral, thermal and hyper spectral sensing and usage 746A27 Remote Sensing and GIS Lecture 3 Multi spectral, thermal and hyper spectral sensing and usage Chandan Roy Guest Lecturer Department of Computer and Information Science Linköping University Multi

More information

EXPRIMENT 3 COUPLING FIBERS TO SEMICONDUCTOR SOURCES

EXPRIMENT 3 COUPLING FIBERS TO SEMICONDUCTOR SOURCES EXPRIMENT 3 COUPLING FIBERS TO SEMICONDUCTOR SOURCES OBJECTIVES In this lab, firstly you will learn to couple semiconductor sources, i.e., lightemitting diodes (LED's), to optical fibers. The coupling

More information

A fast F-number 10.6-micron interferometer arm for transmitted wavefront measurement of optical domes

A fast F-number 10.6-micron interferometer arm for transmitted wavefront measurement of optical domes A fast F-number 10.6-micron interferometer arm for transmitted wavefront measurement of optical domes Doug S. Peterson, Tom E. Fenton, Teddi A. von Der Ahe * Exotic Electro-Optics, Inc., 36570 Briggs Road,

More information

Optical Coherence: Recreation of the Experiment of Thompson and Wolf

Optical Coherence: Recreation of the Experiment of Thompson and Wolf Optical Coherence: Recreation of the Experiment of Thompson and Wolf David Collins Senior project Department of Physics, California Polytechnic State University San Luis Obispo June 2010 Abstract The purpose

More information

NASTER System Definition Proposal

NASTER System Definition Proposal Remote Sensing Team NASTER System Definition Proposal All rights reserved. - 7/14/03 Page 1 Overview Review and comment the mid-ir requirements Presentation of ABB s current platform technology Proposed

More information

Material analysis by infrared mapping: A case study using a multilayer

Material analysis by infrared mapping: A case study using a multilayer Material analysis by infrared mapping: A case study using a multilayer paint sample Application Note Author Dr. Jonah Kirkwood, Dr. John Wilson and Dr. Mustafa Kansiz Agilent Technologies, Inc. Introduction

More information

Maya2000 Pro Spectrometer

Maya2000 Pro Spectrometer now with triggering! Maya2000 Pro Our Maya2000 Pro Spectrometer offers you the perfect solution for applications that demand low light-level, UV-sensitive operation. This back-thinned, 2D FFT-CCD, uncooled

More information

Testing Aspherics Using Two-Wavelength Holography

Testing Aspherics Using Two-Wavelength Holography Reprinted from APPLIED OPTICS. Vol. 10, page 2113, September 1971 Copyright 1971 by the Optical Society of America and reprinted by permission of the copyright owner Testing Aspherics Using Two-Wavelength

More information

Test procedures Page: 1 of 5

Test procedures Page: 1 of 5 Test procedures Page: 1 of 5 1 Scope This part of document establishes uniform requirements for measuring the numerical aperture of optical fibre, thereby assisting in the inspection of fibres and cables

More information

MicroSpot FOCUSING OBJECTIVES

MicroSpot FOCUSING OBJECTIVES OFR P R E C I S I O N O P T I C A L P R O D U C T S MicroSpot FOCUSING OBJECTIVES APPLICATIONS Micromachining Microlithography Laser scribing Photoablation MAJOR FEATURES For UV excimer & high-power YAG

More information

A laser speckle reduction system

A laser speckle reduction system A laser speckle reduction system Joshua M. Cobb*, Paul Michaloski** Corning Advanced Optics, 60 O Connor Road, Fairport, NY 14450 ABSTRACT Speckle degrades the contrast of the fringe patterns in laser

More information

1.6 Beam Wander vs. Image Jitter

1.6 Beam Wander vs. Image Jitter 8 Chapter 1 1.6 Beam Wander vs. Image Jitter It is common at this point to look at beam wander and image jitter and ask what differentiates them. Consider a cooperative optical communication system that

More information

University of Wisconsin Chemistry 524 Spectroscopic Components *

University of Wisconsin Chemistry 524 Spectroscopic Components * University of Wisconsin Chemistry 524 Spectroscopic Components * In journal articles, presentations, and textbooks, chemical instruments are often represented as block diagrams. These block diagrams highlight

More information

MASSACHUSETTS INSTITUTE OF TECHNOLOGY Department of Electrical Engineering and Computer Science

MASSACHUSETTS INSTITUTE OF TECHNOLOGY Department of Electrical Engineering and Computer Science Student Name Date MASSACHUSETTS INSTITUTE OF TECHNOLOGY Department of Electrical Engineering and Computer Science 6.161 Modern Optics Project Laboratory Laboratory Exercise No. 6 Fall 2010 Solid-State

More information

White Paper: Modifying Laser Beams No Way Around It, So Here s How

White Paper: Modifying Laser Beams No Way Around It, So Here s How White Paper: Modifying Laser Beams No Way Around It, So Here s How By John McCauley, Product Specialist, Ophir Photonics There are many applications for lasers in the world today with even more on the

More information

Lithography. 3 rd. lecture: introduction. Prof. Yosi Shacham-Diamand. Fall 2004

Lithography. 3 rd. lecture: introduction. Prof. Yosi Shacham-Diamand. Fall 2004 Lithography 3 rd lecture: introduction Prof. Yosi Shacham-Diamand Fall 2004 1 List of content Fundamental principles Characteristics parameters Exposure systems 2 Fundamental principles Aerial Image Exposure

More information

Coherent Laser Measurement and Control Beam Diagnostics

Coherent Laser Measurement and Control Beam Diagnostics Coherent Laser Measurement and Control M 2 Propagation Analyzer Measurement and display of CW laser divergence, M 2 (or k) and astigmatism sizes 0.2 mm to 25 mm Wavelengths from 220 nm to 15 µm Determination

More information

EUV Plasma Source with IR Power Recycling

EUV Plasma Source with IR Power Recycling 1 EUV Plasma Source with IR Power Recycling Kenneth C. Johnson kjinnovation@earthlink.net 1/6/2016 (first revision) Abstract Laser power requirements for an EUV laser-produced plasma source can be reduced

More information

Technical Synopsis and Discussion of:

Technical Synopsis and Discussion of: OPTI-521, Fall 2008 E.D. Fasse, Page 1 Technical Synopsis and Discussion of: Optical Alignment of a Pupil Imaging Spectrometer by Stephen Horchem and Richard Kohrman Proc. of SPIE Vol. 1167, Precision

More information

EE119 Introduction to Optical Engineering Fall 2009 Final Exam. Name:

EE119 Introduction to Optical Engineering Fall 2009 Final Exam. Name: EE119 Introduction to Optical Engineering Fall 2009 Final Exam Name: SID: CLOSED BOOK. THREE 8 1/2 X 11 SHEETS OF NOTES, AND SCIENTIFIC POCKET CALCULATOR PERMITTED. TIME ALLOTTED: 180 MINUTES Fundamental

More information

NIST Agency Report May 2012 OUTLINE. The case for traceability NMI capabilities A view to the future the HIP Current/recent NIST activities

NIST Agency Report May 2012 OUTLINE. The case for traceability NMI capabilities A view to the future the HIP Current/recent NIST activities NIST Agency Report May 2012 OUTLINE The case for traceability NMI capabilities A view to the future the HIP Current/recent NIST activities The case for traceability Earth Radiation Budget: Solar irradiance

More information

Thin holographic camera with integrated reference distribution

Thin holographic camera with integrated reference distribution Thin holographic camera with integrated reference distribution Joonku Hahn, Daniel L. Marks, Kerkil Choi, Sehoon Lim, and David J. Brady* Department of Electrical and Computer Engineering and The Fitzpatrick

More information

QE65000 Spectrometer. Scientific-Grade Spectroscopy in a Small Footprint. now with. Spectrometers

QE65000 Spectrometer. Scientific-Grade Spectroscopy in a Small Footprint. now with. Spectrometers QE65000 Spectrometer Scientific-Grade Spectroscopy in a Small Footprint QE65000 The QE65000 Spectrometer is the most sensitive spectrometer we ve developed. Its Hamamatsu FFT-CCD detector provides 90%

More information

Detection of the mm-wave radiation using a low-cost LWIR microbolometer camera from a multiplied Schottky diode based source

Detection of the mm-wave radiation using a low-cost LWIR microbolometer camera from a multiplied Schottky diode based source Detection of the mm-wave radiation using a low-cost LWIR microbolometer camera from a multiplied Schottky diode based source Basak Kebapci 1, Firat Tankut 2, Hakan Altan 3, and Tayfun Akin 1,2,4 1 METU-MEMS

More information

Fiber Optic Communications

Fiber Optic Communications Fiber Optic Communications ( Chapter 2: Optics Review ) presented by Prof. Kwang-Chun Ho 1 Section 2.4: Numerical Aperture Consider an optical receiver: where the diameter of photodetector surface area

More information

Imaging Fourier transform spectrometer

Imaging Fourier transform spectrometer Rochester Institute of Technology RIT Scholar Works Theses Thesis/Dissertation Collections 2001 Imaging Fourier transform spectrometer Eric Sztanko Follow this and additional works at: http://scholarworks.rit.edu/theses

More information

An integral eld spectrograph for the 4-m European Solar Telescope

An integral eld spectrograph for the 4-m European Solar Telescope Mem. S.A.It. Vol. 84, 416 c SAIt 2013 Memorie della An integral eld spectrograph for the 4-m European Solar Telescope A. Calcines 1,2, M. Collados 1,2, and R. L. López 1 1 Instituto de Astrofísica de Canarias

More information

SUPPLEMENTARY INFORMATION

SUPPLEMENTARY INFORMATION Making methane visible SUPPLEMENTARY INFORMATION DOI: 10.1038/NCLIMATE2877 Magnus Gålfalk, Göran Olofsson, Patrick Crill, David Bastviken Table of Contents 1. Supplementary Methods... 2 2. Supplementary

More information

Introduction. Chapter 16 Diagnostic Radiology. Primary radiological image. Primary radiological image

Introduction. Chapter 16 Diagnostic Radiology. Primary radiological image. Primary radiological image Introduction Chapter 16 Diagnostic Radiology Radiation Dosimetry I Text: H.E Johns and J.R. Cunningham, The physics of radiology, 4 th ed. http://www.utoledo.edu/med/depts/radther In diagnostic radiology

More information

1. INTRODUCTION 2. LASER ABSTRACT

1. INTRODUCTION 2. LASER ABSTRACT Compact solid-state laser to generate 5 mj at 532 nm Bhabana Pati*, James Burgess, Michael Rayno and Kenneth Stebbins Q-Peak, Inc., 135 South Road, Bedford, Massachusetts 01730 ABSTRACT A compact and simple

More information

How-to guide. Working with a pre-assembled THz system

How-to guide. Working with a pre-assembled THz system How-to guide 15/06/2016 1 Table of contents 0. Preparation / Basics...3 1. Input beam adjustment...4 2. Working with free space antennas...5 3. Working with fiber-coupled antennas...6 4. Contact details...8

More information

Computer Generated Holograms for Optical Testing

Computer Generated Holograms for Optical Testing Computer Generated Holograms for Optical Testing Dr. Jim Burge Associate Professor Optical Sciences and Astronomy University of Arizona jburge@optics.arizona.edu 520-621-8182 Computer Generated Holograms

More information

Spectroscopy in the UV and Visible: Instrumentation. Spectroscopy in the UV and Visible: Instrumentation

Spectroscopy in the UV and Visible: Instrumentation. Spectroscopy in the UV and Visible: Instrumentation Spectroscopy in the UV and Visible: Instrumentation Typical UV-VIS instrument 1 Source - Disperser Sample (Blank) Detector Readout Monitor the relative response of the sample signal to the blank Transmittance

More information

Beam Shaping and Simultaneous Exposure by Diffractive Optical Element in Laser Plastic Welding

Beam Shaping and Simultaneous Exposure by Diffractive Optical Element in Laser Plastic Welding Beam Shaping and Simultaneous Exposure by Diffractive Optical Element in Laser Plastic Welding AKL`12 9th May 2012 Dr. Daniel Vogler Page 1 Motivation: Quality and flexibility diffractive spot shaping

More information

capabilities Infrared Contact us for a Stock or Custom Quote Today!

capabilities Infrared Contact us for a Stock or Custom Quote Today! Infrared capabilities o 65+ Stock Components Available for Immediate Delivery o Design Expertise in SWIR, Mid-Wave, and Long-Wave Assemblies o Flat, Spherical, and Aspherical Manufacturing Expertise Edmund

More information

Lab Report 3: Speckle Interferometry LIN PEI-YING, BAIG JOVERIA

Lab Report 3: Speckle Interferometry LIN PEI-YING, BAIG JOVERIA Lab Report 3: Speckle Interferometry LIN PEI-YING, BAIG JOVERIA Abstract: Speckle interferometry (SI) has become a complete technique over the past couple of years and is widely used in many branches of

More information

Optical Components - Scanning Lenses

Optical Components - Scanning Lenses Optical Components Scanning Lenses Scanning Lenses (Ftheta) Product Information Figure 1: Scanning Lenses A scanning (Ftheta) lens supplies an image in accordance with the socalled Ftheta condition (y

More information

The FTNIR Myths... Misinformation or Truth

The FTNIR Myths... Misinformation or Truth The FTNIR Myths... Misinformation or Truth Recently we have heard from potential customers that they have been told that FTNIR instruments are inferior to dispersive or monochromator based NIR instruments.

More information

Be aware that there is no universal notation for the various quantities.

Be aware that there is no universal notation for the various quantities. Fourier Optics v2.4 Ray tracing is limited in its ability to describe optics because it ignores the wave properties of light. Diffraction is needed to explain image spatial resolution and contrast and

More information

LTE. Tester of laser range finders. Integrator Target slider. Transmitter channel. Receiver channel. Target slider Attenuator 2

LTE. Tester of laser range finders. Integrator Target slider. Transmitter channel. Receiver channel. Target slider Attenuator 2 a) b) External Attenuators Transmitter LRF Receiver Transmitter channel Receiver channel Integrator Target slider Target slider Attenuator 2 Attenuator 1 Detector Light source Pulse gene rator Fiber attenuator

More information

Big League Cryogenics and Vacuum The LHC at CERN

Big League Cryogenics and Vacuum The LHC at CERN Big League Cryogenics and Vacuum The LHC at CERN A typical astronomical instrument must maintain about one cubic meter at a pressure of

More information

CubeSat-Scale Hyperspectral Imager for Middle Atmosphere Investigations

CubeSat-Scale Hyperspectral Imager for Middle Atmosphere Investigations CubeSat-Scale Hyperspectral Imager for Middle Atmosphere Investigations Rick Doe 1, Steve Watchorn 2, John Noto 2, Robert Kerr 2, Karl van Dyk 1, Kyle Leveque 1, and Christopher Sioris 3 1 SRI International

More information

Design of the cryo-optical test of the Planck reflectors

Design of the cryo-optical test of the Planck reflectors Design of the cryo-optical test of the Planck reflectors S. Roose, A. Cucchiaro & D. de Chambure* Centre Spatial de Liège, Avenue du Pré-Aily, B-4031 Angleur-Liège, Belgium *ESTEC, Planck project, Keplerlaan

More information

Advanced Target Projector Technologies For Characterization of Staring-Array Based EO Sensors

Advanced Target Projector Technologies For Characterization of Staring-Array Based EO Sensors Advanced Target Projector Technologies For Characterization of Staring-Array Based EO Sensors Alan Irwin, Steve McHugh, Jack Grigor, Paul Bryant Santa Barbara Infrared, 30 S. Calle Cesar Chavez, Suite

More information

OPERATING MANUAL. 100 MHz CENTER FREQUENCY OFF AXIS ACOUSTO-OPTIC BEAM DEFLECTOR MODEL NUMBER: DEG-.51 DOCUMENT NUMBER: 51A12229A

OPERATING MANUAL. 100 MHz CENTER FREQUENCY OFF AXIS ACOUSTO-OPTIC BEAM DEFLECTOR MODEL NUMBER: DEG-.51 DOCUMENT NUMBER: 51A12229A OPERATING MANUAL 100 MHz CENTER FREQUENCY OFF AXIS ACOUSTO-OPTIC BEAM DEFLECTOR MODEL NUMBER: DOCUMENT NUMBER: 51A12229A Document approved for release: W Seale Date: 8/18/06 US OFFICE: NEOS Technologies,

More information

Dense Spatial Multiplexing Enables High Brightness Multi-kW Diode Laser Systems

Dense Spatial Multiplexing Enables High Brightness Multi-kW Diode Laser Systems Invited Paper Dense Spatial Multiplexing Enables High Brightness Multi-kW Diode Laser Systems Holger Schlüter a, Christoph Tillkorn b, Ulrich Bonna a, Greg Charache a, John Hostetler a, Ting Li a, Carl

More information

MODULAR ADAPTIVE OPTICS TESTBED FOR THE NPOI

MODULAR ADAPTIVE OPTICS TESTBED FOR THE NPOI MODULAR ADAPTIVE OPTICS TESTBED FOR THE NPOI Jonathan R. Andrews, Ty Martinez, Christopher C. Wilcox, Sergio R. Restaino Naval Research Laboratory, Remote Sensing Division, Code 7216, 4555 Overlook Ave

More information

Contouring aspheric surfaces using two-wavelength phase-shifting interferometry

Contouring aspheric surfaces using two-wavelength phase-shifting interferometry OPTICA ACTA, 1985, VOL. 32, NO. 12, 1455-1464 Contouring aspheric surfaces using two-wavelength phase-shifting interferometry KATHERINE CREATH, YEOU-YEN CHENG and JAMES C. WYANT University of Arizona,

More information

Akinori Mitani and Geoff Weiner BGGN 266 Spring 2013 Non-linear optics final report. Introduction and Background

Akinori Mitani and Geoff Weiner BGGN 266 Spring 2013 Non-linear optics final report. Introduction and Background Akinori Mitani and Geoff Weiner BGGN 266 Spring 2013 Non-linear optics final report Introduction and Background Two-photon microscopy is a type of fluorescence microscopy using two-photon excitation. It

More information

7 CHAPTER 7: REFRACTIVE INDEX MEASUREMENTS WITH COMMON PATH PHASE SENSITIVE FDOCT SETUP

7 CHAPTER 7: REFRACTIVE INDEX MEASUREMENTS WITH COMMON PATH PHASE SENSITIVE FDOCT SETUP 7 CHAPTER 7: REFRACTIVE INDEX MEASUREMENTS WITH COMMON PATH PHASE SENSITIVE FDOCT SETUP Abstract: In this chapter we describe the use of a common path phase sensitive FDOCT set up. The phase measurements

More information

Development of a new multi-wavelength confocal surface profilometer for in-situ automatic optical inspection (AOI)

Development of a new multi-wavelength confocal surface profilometer for in-situ automatic optical inspection (AOI) Development of a new multi-wavelength confocal surface profilometer for in-situ automatic optical inspection (AOI) Liang-Chia Chen 1#, Chao-Nan Chen 1 and Yi-Wei Chang 1 1. Institute of Automation Technology,

More information

TIME-PRESERVING MONOCHROMATORS FOR ULTRASHORT EXTREME-ULTRAVIOLET PULSES

TIME-PRESERVING MONOCHROMATORS FOR ULTRASHORT EXTREME-ULTRAVIOLET PULSES TIME-PRESERVING MONOCHROMATORS FOR ULTRASHORT EXTREME-ULTRAVIOLET PULSES Luca Poletto CNR - Institute of Photonics and Nanotechnologies Laboratory for UV and X-Ray Optical Research Padova, Italy e-mail:

More information

Properties of Structured Light

Properties of Structured Light Properties of Structured Light Gaussian Beams Structured light sources using lasers as the illumination source are governed by theories of Gaussian beams. Unlike incoherent sources, coherent laser sources

More information

Design, calibration and assembly of an Offner imaging spectrometer

Design, calibration and assembly of an Offner imaging spectrometer Journal of Physics: Conference Series Design, calibration and assembly of an Offner imaging spectrometer To cite this article: Héctor González-Núñez et al 2011 J. Phys.: Conf. Ser. 274 012106 View the

More information

BMC s heritage deformable mirror technology that uses hysteresis free electrostatic

BMC s heritage deformable mirror technology that uses hysteresis free electrostatic Optical Modulator Technical Whitepaper MEMS Optical Modulator Technology Overview The BMC MEMS Optical Modulator, shown in Figure 1, was designed for use in free space optical communication systems. The

More information

LightPath. Infrared Optics. Leaders in aspheric optics and assemblies TECHNOLOGIES

LightPath. Infrared Optics. Leaders in aspheric optics and assemblies TECHNOLOGIES LightPath TECHNOLOGIES Infrared Optics Leaders in aspheric optics and assemblies Infrared Optics from the Experts in Molded Glass Optics Leaders in chalcogenide glass Molding Enhanced thermal performance

More information

Spectroscopic Instrumentation

Spectroscopic Instrumentation Spectroscopic Instrumentation Theodor Pribulla Astronomical Institute of the Slovak Academy of Sciences, Tatranská Lomnica, Slovakia Spectroscopic workshop, February 6-10, 2017, PřF MU, Brno Principal

More information

Dario Cabib, Amir Gil, Moshe Lavi. Edinburgh April 11, 2011

Dario Cabib, Amir Gil, Moshe Lavi. Edinburgh April 11, 2011 New LWIR Spectral Imager with uncooled array SI-LWIR LWIR-UC Dario Cabib, Amir Gil, Moshe Lavi Edinburgh April 11, 2011 Contents BACKGROUND AND HISTORY RATIONALE FOR UNCOOLED CAMERA BASED SPECTRAL IMAGER

More information

Exoplanet transit, eclipse, and phase curve observations with JWST NIRCam. Tom Greene & John Stansberry JWST NIRCam transit meeting March 12, 2014

Exoplanet transit, eclipse, and phase curve observations with JWST NIRCam. Tom Greene & John Stansberry JWST NIRCam transit meeting March 12, 2014 Exoplanet transit, eclipse, and phase curve observations with JWST NIRCam Tom Greene & John Stansberry JWST NIRCam transit meeting March 12, 2014 1 Scope of Talk NIRCam overview Suggested transit modes

More information