KODAK 9000C and KODAK 9000C 3D Extraoral Imaging System. Installation Guide

Size: px
Start display at page:

Download "KODAK 9000C and KODAK 9000C 3D Extraoral Imaging System. Installation Guide"

Transcription

1 KODAK 9000C and KODAK 9000C 3D Extraoral Imaging System Installation Guide

2 Notice Congratulations on your purchase of the KODAK 9000C and KODAK 9000C 3D Extraoral Imaging System. Thank you for your confidence in our products and we will do all in our power to ensure your complete satisfaction. The Installation Guide for the KODAK 9000C and KODAK 9000C 3D Extraoral Imaging System includes information on the cephalometric features. For the panoramic or the 3D features, see the KODAK 9000 and KODAK D Extraoral Imaging System (SM711) Installation Guide. We recommend that you thoroughly familiarize yourself with this Guide in order to make the most effective use of your system. The information contained in this Guide may be subject to modification without notice, justification or notification to the persons concerned. No part of this Guide may be reproduced without the express permission of Carestream Health, Inc. The US Federal law restricts this device to sale by or on the order of a physician. This document is originally written in English. Manual Name: KODAK 9000C and KODAK 9000C 3D Extraoral Imaging System Installation Guide Part Number: SM714 Revision Number: 01 Print Date: 06/2008 The brand names and logos reproduced in this Guide are copyright. KODAK is a trademark of KODAK used under License. KODAK 9000C and KODAK 9000C 3D Extraoral Imaging System, complies with Directive 93/42/CEE relating to medical equipment. 0086

3 Contents 1 About This Guide Conventions in this Guide Note to the User Warning and Safety Instructions Marking and Labeling Symbols Responsibility and Warranty KODAK 9000C AND KODAK 9000C 3D UNIT OVERVIEW General Overview Mobile Components General Functional Components Digital Sensor Locations Laser Locations Label Locations Control Panel Overview X-Ray Remote Control Overview KODAK 9000C AND KODAK 9000C 3D PACKAGING Standard Packaging SITE PREPARATION BEFORE INSTALLATION Standard Compliance Environmental Requirements Unit Dimensions Electrical Requirements X-Ray Room Requirement PC System Requirements INSTALLING THE CEPHALOSTAT Tool Requirements Technical Staff Requirements Opening the Box Installing the Cephalostat Unit Installing the Cephalostat Arm Installing the Cephalostat Head Wiring the Cephalostat Head Wiring the Cephalostat to the Unit Head Installing the Cephalometric Frankfort Positioning Laser (For Upgrade Only) Installing the Cephalostat Positioning Locker (For Upgrade Only) Post-Installation Control Adjusting the Cephalostat Head Adjusting the Inclination of the Cephalostat Head KODAK 9000C and KODAK 9000C 3D Extraoral Imaging System Installation Guide (SM714)_Ed 01 iii

4 Contents Adjusting the Height of the Cephalostat Head Adjusting the Ear Cones Alignment Adjusting the Nasion Support Selecting the Cephalometric Blade Collimator Parameters Calibrating the Cephalometic Blade Collimator Checking the Image Quality Fitting the Covers Fitting the KODAK 9000C and KODAK 9000C 3D Covers Fitting the Cephalostat Covers Maintenance Annual Maintenance TECHNICAL SPECIFICATIONS Compliance with International Standards Manufactured by: Compliance with International Regulations Compliance with European and International Standards Classification in Accordance with EN/IEC Classification in Accordance with EN/IEC Ambient Operating Conditions Storage Conditions Transport Conditions Cephalometric Unit Technical Specifications Unit Electronic Specifications X-Ray Generator Technical Specifications iv

5 Chapter 1 About This Guide Conventions in this Guide The following special messages emphasize information or indicate potential risk to persons or equipment: WARNING Warns you to avoid injury to yourself or others by following the safety instructions precisely. CAUTION Alerts you to a condition that might cause serious damage. IMPORTANT Alerts you to a condition that might cause problems. NOTE Emphasizes important information. TIP Provides extra information and hints. KODAK 9000C and KODAK 9000C 3D Extraoral Imaging System Installation Guide (SM714)_Ed

6 Note to the User Note to the User WARNING X-rays can be harmful and dangerous if not used properly. The instructions and warnings contained in this guide must be therefore carefully followed. As a manufacturer of radiology units that conform to stringent radiological protection standards in force throughout the world, we guarantee as low as reasonably achievable degree of protection against radiation hazards. Nonetheless, you are handling a radiology unit specially designed to emit x-ray doses in order to carry out a medical diagnosis. The room in which your radiology unit is to be installed must comply with all official regulations applicable to protection against radiation. You must install your radiology unit in a room protected against x-ray emission. This room must reduce the frequency interferences of the 30MHz to 1GHz band to at least 12db. Your local representative will assist you in the initial use of your radiology unit and will supply any relevant information you may require. To use and operate your panoramic unit and your digital imaging software you must follow the instructions contained in this guide. Warning and Safety Instructions When operating the KODAK 9000C and KODAK 9000C 3D unit, observe the following warning and safety instructions: DANGER OF ELECTRIC SHOCK This is an electrical unit. DO NOT expose it to water spray. Such action may cause an electric shock or a malfunction of the unit. LASER WARNING LASER RADIATION DO NOT STARE INTO BEAM CLASS 2 LASER PRODUCT Maximum laser output:1mw,650nm IEC :1993+A1:1997+A2:2001 For maximum safety, advise the patient not to look at the beam. Before turning on the beams, lower the Frankfurt plane beam to the lowest level. While making adjustments, ensure that the beam is not directed into the patient's eyes. 1 2 About This Guide

7 Warning and Safety Instructions WARNINGS: You are responsible for the operation and maintenance of this unit. Only legally qualified persons can operate this unit. They MUST have training to use the radiological equipment. DO NOT open the cover of the unit.when necessary, have a trained authorized service technician carry out inspection and maintenance operations. Install this unit in an x-ray room that complies with current installation standards. From this location, you must be able to maintain visual or audio communication with the patient and be able to access the Acquisition interface module during exposure. This unit must be permanently connected to the ground with a fixed power supply cable. DO NOT place the PC and the peripheral equipment connected to it in the immediate vicinity of the patient in the unit. Leave at least 1.5 m distance between the patient and the unit. The PC and the peripheral equipment must conform to the IEC60950 standard. See your computer installation guide for details of the data processing system, PC and screen. Leave a sufficient amount of clear space around the CPU to ensure that it is properly ventilated. To obtain maximum image quality and visual comfort, position the screen to avoid direct light reflections from internal or external lighting. DO NOT operate the unit if there is the threat of an earthquake. Following an earthquake, ensure that the unit is operating satisfactorily before using it again. Failure to observe this precaution may expose patients to hazards. X-ray equipment is hazardous to patients and the operator if you do not observe the exposure safety factors and operating instructions. DO NOT place objects within the field of operation of the unit. The patient should wear a protective lead-lined shoulder apron, unless other Radiation Protection Protocols apply locally. Disinfect any parts of the unit that come into contact with the patient and the operator after each patient has been exposed to x-rays. While adjusting the height of the unit, ensure that the patient is kept clear of the mechanism. When the unit is not in use, ensure that the ON/OFF switch is set to OFF (O). If the unit develops a fault, switch it to off (O), display an Unserviceable notice and contact a service technician. To dispose of the unit or its components, contact a service technician. Ask the patient to refrain from moving during the entire period of exposure. Ask the patient to remain still until the unit arm has stopped moving and the RESET movement has completed. DO NOT use this unit in conjunction with oxygen-rich environments. This unit is not intended for use with flammable anesthetics or flammable agents. DO NOT hang from the cephalostat. KODAK 9000C and KODAK 9000C 3D Extraoral Imaging System Installation Guide (SM714)_Ed

8 Marking and Labeling Symbols Marking and Labeling Symbols Type B device symbol complying with the IEC standard In the EEC, this symbol indicates: DO NOT discard this product in a trash receptacle; use an appropriate recovery and recycling facility. Contact your local sales representative for additional information on the collection and recovery programs available for this product WARNING and IONIZING RADIATION symbols warn you about radiation dangers. LASER RADIATION DO NOT STARE INTO BEAM CLASS 2 LASER PRODUCT Maximum laser output:1mw,650nm IEC :1993+A1:1997+A2:2001 N L LASER WARNING Laser radiation. DO NOT stare into the beam. Class 2 laser product. Maximum output power: 1mW, 650 nm (IEC standard) This unit emits laser radiation. Connecting point to the neutral conductor Connecting point to the live conductor Protection ground (G) Functional ground 1 4 About This Guide

9 Responsibility and Warranty Responsibility and Warranty Only damage affecting products themselves will be considered. The client can in no case claim damage compensation if it does not purely concern the product itself and provided that this exclusion of responsibility does not infringe any legal measures. This warranty does not cover damages and faults due to accident, incorrect usage, improper use, negligence, or wear and tear due to normal use. Carestream Health, Inc. can not be held liable for the consequences resulting from the non-application of the instructions contained in the installation and user manuals, namely bodily harm, profit loss, operational interruptions, data loss, loss of a financial nature, or any direct or indirect damage. Carestream Health, Inc. reserves the right to examine any presumed defect. Clients are not entitled to delay the payment of their bills or any financial deductions by way of products reputed to be faulty. Due to the continuous development of its products, Carestream Health, Inc. reserves the right to amend at any time the manual and products mentioned therein, without justification or notification to the people involved. Our products are guaranteed in their original packaging. Carestream Health, Inc. cannot be held liable for damages resulting from transportation. Consumable spares, software and accessories are excluded from the warranty. KODAK 9000C and KODAK 9000C 3D Extraoral Imaging System Installation Guide (SM714)_Ed

10 Responsibility and Warranty 1 6 About This Guide

11 Chapter 2 KODAK 9000C AND KODAK 9000C 3D UNIT OVERVIEW The KODAK 9000C and KODAK 9000C 3D digital imaging unit is compliant with the requirements of the EEC and international medical standards. The KODAK 9000C and KODAK 9000C 3D unit has been designed to carry out the following radiological examinations: Lateral Frontal (PA or AP) Oblique Submento-vertex Carpus General Overview The KODAK 9000C unit is composed of the following functional components: The rotative arm The fixed arm with a control panel The panoramic digital sensor The x-ray source assembly The x-ray remote control The chin rest base The panoramic chin rest and bite block The temple supports The hand grips The cephalostat arm The cephalostat head The head clamps and ear cones The nasion support The acquisition software (see Imaging Software Overview ) The KODAK 9000C 3D unit is composed of the following functional components: The rotative arm The fixed arm with a control panel The panoramic digital sensor The 3D digital sensor The x-ray source assembly KODAK 9000C and KODAK 9000C 3D Extraoral Imaging System Installation Guide (SM714)_Ed

12 General Overview The x-ray remote control The chin rest base The panoramic chin rest and bite block The 3D chin rest and bite block The temple supports The hand grips The cephalostat arm The cephalostat head The head clamps and ear cones The nasion support The acquisition software (see Imaging Software Overview ) The following figures illustrate the general overview of the KODAK 9000C and KODAK 9000C 3D unit. Mobile Components Figure 2-1 illustrates the up and down movement of the KODAK 9000C and KODAK 9000C 3D mobile component and the 180 rotation of the rotative arm. Figure 2 1 KODAK 9000C and KODAK 9000C 3D Unit Mobile Components 2 2 KODAK 9000C AND KODAK 9000C 3D UNIT OVERVIEW

13 General Overview General Functional Components Figure 2-2 illustrates the general functional components of the KODAK 9000C and KODAK 9000C 3D. Figure 2 2 KODAK 9000C and KODAK 9000C 3D Unit Functional Components b a RJ45/1 10 RJ45/ Unit fixed arm 16 Unit rotative arm 2 Control panel 17 Under the head front cover 3 Hand grips 18 Under the head back cover 4 Temple supports 19 Head cover 5 Chin rest base 20 Column side cover 5 a Panoramic chin rest 21 Column inside cover 5 b 3D chin rest and forehead support 22 ON/OFF button 6 Unit mains connecting terminal 23 Wall mounting brackets 7 Green warning lamp (ready state indicator) 24 Wall mounting brackets cover KODAK 9000C and KODAK 9000C 3D Extraoral Imaging System Installation Guide (SM714)_Ed

14 General Overview 8 Column connecting terminals 25 Column upper cover 9 X-Ray remote control 26 Cephalostat arm 10 Door safety switch 27 Cephalostat end of the arm cover 11 PC hosting the imaging and acquisition software 28 Cephalostat head top cover 12 Ethernet outlet RJ45/1 29 Cephalostat head side cover 13 LAN RJ45/2 30 Carbon fiber screen 14 Sensor 31 Head clamps and ear cones 15 X-ray source assembly 32 Nasion support 2 4 KODAK 9000C AND KODAK 9000C 3D UNIT OVERVIEW

15 General Overview Digital Sensor Locations Figure 2-3 illustrates the locations of the panoramic, 3D and cephalometric digital sensors of the KODAK 9000C and KODAK 9000C 3D. IMPORTANT * When the unit is ONLY panoramic, the 3D option is not available. Therefore, the location of the digital sensor indicates that it is ready for the 3D option. Figure 2 3 KODAK 9000C and KODAK 9000C 3D Unit Panoramic, 3D and Cephalometric Digital Sensor Locations Digital 3D * KODAK 9000C and KODAK 9000C 3D Extraoral Imaging System Installation Guide (SM714)_Ed

16 General Overview Laser Locations Figure 2-4 illustrates the location of the lasers of the KODAK 9000C and KODAK 9000C 3D. Figure 2 4 KODAK 9000C and KODAK 9000C 3D Unit Laser Positions B 5 A 1 B A D central positioning laser beam 2 Mid-sagittal positioning laser beam 3 Horizontal positioning laser beam 4 3D Field of View (FoV) positioning laser beam 5 Cephalometric Frankfurt positioning laser beam 2 6 KODAK 9000C AND KODAK 9000C 3D UNIT OVERVIEW

17 General Overview Label Locations Figure 2-5 illustrates the location of all the labels of the KODAK 9000C and KODAK 9000C 3D. The label locations of the KODAK 9000C are represented with *. IMPORTANT The labels presented are ONLY for the KODAK D option. The *Panoramic presents the labels and their locations for the KODAK Figure 2 5 KODAK 9000C and KODAK 9000C 3D Unit Label Locations 2008/04/23 K9000 C 3D XPNK004 * 0.90A 2008/04/23 X-Ray Source Assembly :CU ,5 mm AI 2,5 mmai CEI OPX * Type : CG932 * Panoramic 2008/04/23 K9000 C XPNK A KODAK 9000C and KODAK 9000C 3D Extraoral Imaging System Installation Guide (SM714)_Ed

18 Control Panel Overview Control Panel Overview The Control Panel is an alphanumeric digital soft touch console. It allows the operator to control certain unit functions. It also displays the operating parameters and error messages. Figure 2 6 Unit Control Panel kv ma S mem Height Adjustment button: Adjusts the height of the unit to the height of the patient. 2 Head Adjustment button: Adjusts the patient head to the x-ray beams. 3 3D Adjustment button: Adjusts the Unit arm movements to correctly position the patient for 3D acquisition. 4 Laser Beam button: Activates the beams to correctly position the patient. 5 3D Position Verification button: Positions the unit rotative arm at the selected or memorized 3D position. 6 Reset button: Resets the Unit arm to the initial position to enable the patient to enter and exit the Unit. 7 3D Memorization button: Memorizes the 3D current positioning parameter settings that override the default parameters. 8 Display Screen: Displays the current acquisition parameters and the error messages. 9 Ready Indicator LED: Green, indicates the unit is ready for acquisition. 10 X-Ray Emission LED: Yellow, indicates the x-rays are being emitted. 11 System Status LED: Red, indicates the error alerts. 2 8 KODAK 9000C AND KODAK 9000C 3D UNIT OVERVIEW

19 X-Ray Remote Control Overview X-Ray Remote Control Overview The x- ray remote control enables you to launch a radiological image acquisition via the exposure button from outside the x-ray room. You must press and hold the exposure button until the end of acquisition. Premature release of the exposure button interrupts the acquisition. Figure 2 7 X-Ray Remote Control 1 1 Exposure button: launches image acquisition. KODAK 9000C and KODAK 9000C 3D Extraoral Imaging System Installation Guide (SM714)_Ed

20 X-Ray Remote Control Overview 2 10 KODAK 9000C AND KODAK 9000C 3D UNIT OVERVIEW

21 Chapter 3 KODAK 9000C AND KODAK 9000C 3D PACKAGING Standard Packaging Table 3 1 Cephalometric Assembly Components Box Components Dimension (mm) Weight Cephalostat head Cephalostat arm Cephalostat covers Cephalometric Frankfurt positioning laser assembly* Cephalostat positioning locker* Documentation (User Guide and Installation Guide) 1200mm (L) x 800mm (D) x 1100mm (H) 78 kg * The listed components will be in your packaging only if you have to install the Cephalostat option at a later date than the KODAK 9000 or KODAK D. Table 3 2 Components Product Description Quantity Cephalostat head 1 KODAK 9000C and KODAK 9000C 3D Extraoral Imaging System Installation Guide (SM714)_Ed

22 Standard Packaging Table 3 2 Components (Continued) Product Description Quantity Cephalostat arm End of the arm cover 1 Head top cover 1 Head side cover 1 Cephalometric Frankfurt* positioning laser assembly 1 Cephalostat positioning locker* 1 * The listed components will be in your packaging only if you have to install the Cephalostat option at a later date than the KODAK 9000 or KODAK D. 3 2 KODAK 9000C AND KODAK 9000C 3D PACKAGING

23 Chapter 4 SITE PREPARATION BEFORE INSTALLATION IMPORTANT Prior to placing the order and before installation, carefully check the following requirements for the x-ray room. Standard Compliance Install the unit in an x-ray room compliant with all official regulations applicable to protection against radiation. This room must reduce the frequency interferences of the 30MHz to 1GHz band to at least 12db. Environmental Requirements Check the following ambient operating condition requirements of the x-ray room before installing the unit: Temperatures: 5 ~35 C Relative humidity: 30 ~ 85% Atmospheric pressure: 700 ~ 1060 hpa KODAK 9000C and KODAK 9000C 3D Extraoral Imaging System Installation Guide (SM714)_Ed

24 Unit Dimensions Unit Dimensions MAXI 1595 mm (62 3 ") mm (54 3 ") mm (15 63 ") mm (12 3 ") mm ( 25 ") mm (22 51 ") mm (61 27 ") 32 The unit dimensions illustrated in the above figure are as follows: Maximum height of the unit (2378mm) Minimum (1016mm) and maximum (1693mm) height of the chin rest Minimum (1030mm) and maximum (1706mm) height of the ear cones Width (2150mm) and depth (1595mm) of the unit WARNING If you need to add a base plate you must add 15mm to the height of the unit SITE PREPARATION BEFORE INSTALLATION

25 Electrical Requirements Electrical Requirements WARNING* You MUST select the operating voltage when placing an order. The operating voltage CANNOT be modified on site. The unit can operate at: 100/110/130 V 50/60 Hz 230/240 V 50/60 Hz Table 4 1 Optional Operating Voltages of the Unit Nominal voltage * (no load) Minimum Maximum Maximum line current 100 V V 90 V 143 V 20 A 230 V V 207 V 264 V 10 A CAUTION The power supply line must be equipped with a connection box that ensures a constant connection. It must not be possible to connect the unit to the power supply without using a tool. The unit must be protected against any accidental disconnection. If other units are installed on the same line, interference and voltage fluctuations can cause the radiological unit to operate abnormally. We strongly recommend that a separate electrical line be dedicated to supply power to the KODAK 9000 and KODAK D unit. This line should be protected by a circuit breaker with a maximum current of: 16 A at 230/240 V 20 A at 110/130 V A differential circuit breaker of 30 ma KODAK 9000C and KODAK 9000C 3D Extraoral Imaging System Installation Guide (SM714)_Ed

26 Electrical Requirements Figure 4 1 Electrical Diagram of the X-ray Room and Connections G L N L 1A 5 N RJ45 K9000C 8 L N bis 9 L N General mains 7bis X-ray warning lamp connecting terminal 2 Differential circuit breaker 8 Column connecting terminals 3 Red color actuator emergency stop push-button 9 X-ray remote control 4 Red color actuator emergency stop push-button 10 Door safety switch 5 Red warning lamp, power ON indicator 11 Mains outlet (for electric tools) 6 Unit mains connecting terminal 12 Ethernet outlet (RJ45/1) 7 Green warning lamp, ready state indicator 13 Contactor 4 4 SITE PREPARATION BEFORE INSTALLATION

27 Electrical Requirements A single-phase alternating current power supply is required. The electrical installation specifications should be as follows: Table 4 2 Electrical Installation Specifications Supply Voltage 230V/240V 100V/110V/ 130V Frequency 50/60 Hz 50/60 Hz Electrical supply 6 kw 6 kw Line current required Cable cross-section according to length 16A For 30 m: 2.5 mm2 20A For 10 m: 2.5 mm2 Max. line impedance Differential circuit breaker (2) Maximum current to trip circuit breaker 30 ma 30 ma 16A 20A Protect the power supply line with a differential circuit breaker that trips at maximum current. Specifications of 2 red color actuator emergency stop push-button (3) and (4) 250V 6A UL listed 250V 6A UL listed Install these stop push-buttons to simultaneously, switch off the current to the active conductors of the radiological installation and exclude any other electric equipment. Locate (3) inside the x-ray room, near the unit, for the operator to quickly cut the power supply if necessary. Locate (4) outside the x-ray room, near the x-ray remote control for the operator to quickly cut the power supply if necessary. Maintain them in OFF (open) position until a deliberate action is performed. KODAK 9000C and KODAK 9000C 3D Extraoral Imaging System Installation Guide (SM714)_Ed

28 Electrical Requirements Table 4 2 Electrical Installation Specifications (Continued) Specifications of the warning lamps (5) and (7) 60 W 60 W Locate the red warning lamp (5) outside the x-ray room to indicate the unit is active (1 lamp at each access point). Locate the green warning lamp (7) outside the x-ray room to indicate the ready state of the unit for acquisition. Contactor (13) 16A-250V UL listed 20A-250V UL listed Door Safety Switch (9) 1A / 250V 1A / 130V Optionally, connect the door safety switch (9) that deactivates the x-ray remote control if the door remains open. Use the following switches: Emergency Stop Switch: Manufacturer: Cutler-Hammer Ref: E22AT111 E22AT112 Contactor: Manufacturer: Telemecanique Ref: LC1 D4011 F6 (110/130 V - 60 Hz) P5 (230/240 V - 50 Hz) U6 (230/240 V - 60 Hz) IMPORTANT If you cannot use the above manufacturer devices, use an equivalent emergency switch and contactor in the UL list with the same specifications. For more information, see UL clause SITE PREPARATION BEFORE INSTALLATION

29 X-Ray Room Requirement X-Ray Room Requirement IMPORTANT Use an appropriate wall fixing system suitable for the type of wall construction. See the examples below. The following illustrations provide examples of wall types and fixations. 2 x 2 x 16 mm 16 mm 2 x 2 x 16 mm 16 mm Table 4 3 X-Ray Room Requirements Room Components Minimum Requirement Comments Width of the door 75 cm (30") Height of the ceiling 240 cm (95") If needed, a shorter column is possible by special order. Strength of the wall Withstanding an extraction force of 150 kpa at each point of attachment. It is the responsibility of the installer to choose an appropriate fixing system that withstands the extraction force. Load-bearing capacity of the floor 500 kg/m2 The floor must be solid and flat. Required space for the KODAK 9000C and KODAK 9000C 3D unit 1500 (L) x 2000 (D) x 2400 (H) mm IMPORTANT Install the unit where a minimum amount of space is provided to allow easy access for the patient or the maintenance technician (see Figure 4-2). KODAK 9000C and KODAK 9000C 3D Extraoral Imaging System Installation Guide (SM714)_Ed

30 X-Ray Room Requirement Figure 4 2 Minimum X-Ray Room Space Configuration for K9000C and K9000C 3D MINI 630 mm (24 103") 128 MINI 2230 mm (87 51 ") 64 RJ45/1 RJ45/ bis Red color actuator emergency stop push-button 9 X-ray remote control 4 Red color actuator emergency stop push-button 10 Door safety switch 5 Red warning lamp, power ON indicator 11 Mains outlet (for electrical tools) 6 Unit mains connecting terminal 12 Ethernet outlet (RJ45/1) 7 Green warning lamp, ready state indicator 13 Local Area Network (LAN), (RJ45/2) 7bis X-Ray warning lamp connecting terminal WARNING You MUST prepare the appropriate electrical requirements and configurations of the x-ray room before installing the unit. You MUST locate separately the following low voltage and high voltage connections: - Low Voltage: (12), (9), (10) - High Voltage: (6), (5), (7bis) 4 8 SITE PREPARATION BEFORE INSTALLATION

31 PC System Requirements PC System Requirements This section specifies the minimum PC system requirements for KODAK 9000C and KODAK 9000C 3D extraoral imaging system software. IMPORTANT It is MANDATORY to check that the PC system configuration is compatible with the PC system requirements for the KODAK 9000 and KODAK D software. If necessary you MUST update your PC system configuration. KODAK 9000C and KODAK 9000C 3D MUST be connected to the PC via a point-to-point Ethernet link and not via a LAN. DO NOT place the PC and the peripheral equipment connected to it in the immediate vicinity of the patient in the unit. Leave at least 1.5 m distance from the unit. The PC and the peripheral equipment must conform to the IEC standard. IMPORTANT For KODAK 9000C, you can use the Viewing graphic board configuration for acquisition. Table 4 4 Minimum PC System Requirements Item Viewing Acquisition * Comments CPU 2 GHz Intel Duo Core 3 GHz Intel Pentium 4 RAM 2 GB 2GB RAM has a major impact on system performance. Hard disk drive 1.2 GB for software installation 1.2 GB for software installation Graphic board 80 GB free space to use the software Nvidia / ATI based board supporting Open Glide 1.2 with 256 MB of video RAM on AGP x8 video bus (example: Nvidia GeForce 6800 GT) Nvidia board on PCI Express video bus, minimum GeForce 8800 GT 512 MB of video RAM, or Quadro FX MB of video RAM The video RAM has major impact on system performance. Monitor 1 monitor 17" or larger 1024 x 768 minimum screen resolution - 32 bits color mode Operating system Windows 2000 SP4 Windows XP Home / Pro edition SP2 Windows Vista 32 bits 1 monitor 17" 1280 x 880 minimum screen resolution Windows XP Home / Pro edition SP2 Your monitor is a vital component in displaying quality images. Low-quality screens will prevent you from proper diagnoses and treatment. Ethernet interface 1 Ethernet interface 2 Ethernet interfaces (100Mbits) CD/DVD drive A CD-ROM drive is required to install the product. A CD-ROM drive is required to install the product. KODAK 9000C and KODAK 9000C 3D Extraoral Imaging System Installation Guide (SM714)_Ed

32 PC System Requirements Table 4 4 Minimum PC System Requirements (Continued) Item Viewing Acquisition * Comments Backup Media Removable/portable, external hard disk drive Removable/portable, external hard disk drive. We strongly recommend a daily backup of x-ray images and patient records. * For the KODAK 9000 (panoramic) acquisition, you need only the Viewing graphic board configuration SITE PREPARATION BEFORE INSTALLATION

33 Chapter 5- INSTALLING THE CEPHALOSTAT Tool Requirements The installer must supply the following tools: Power drill Screws and heavy duty fixings Spirit level Measuring tape Cutter Metric Allen keys Metric spanners IMPORTANT The tool references mentioned in this manual are ISO tool references. Technical Staff Requirements The installation requires the following number of technicians: Unit Component Technicians Arm 2 Head 2 KODAK 9000C and KODAK 9000C 3D Extraoral Imaging System Installation Guide (SM714)_Ed

34 Opening the Box Opening the Box Before opening the cephalostat assembly box make sure that you have the required tool. To open the cephalostat assembly box, follow these steps: 1. Remove the upper cardboard from the packaging. DO NOT cut through the cardboard. 2. Remove the cephalostat arm. 3. Remove gently and set aside the cephalostat head cover. 4. Remove the protection elements. Lift off the packaging carton. 5. Remove gently the cephalostat head from the transport support and place it on the ground. 6. Remove the other accessories. 5 2 INSTALLING THE CEPHALOSTAT

35 Installing the Cephalostat Unit Installing the Cephalostat Unit WARNING You MUST comply with the standards currently in force while installing and using the radiological unit. Whenever necessary, you MUST use a shield or wear a lead-lined apron. ONLY an approved Trophy technician is qualified to inspect or maintain the unit while it is switched on and while the covers are removed. In this case, NO unqualified person must approach the unit. Before installing the cephalostat unit, make sure that: The unit is switched off. All the unit covers are removed. You have the required tools. IMPORTANT If you install the cephalostat at a later date than the KODAK 9000 and KODAK D, then you need to install the cephalostat Frankfort positioning laster and the cephalostat positioning locker. To install the cephalostat unit, follow these steps: Installing the Cephalostat Arm 1. With the help of 1 person, position the cephalostat arm on the rear of the mobile component of the column. To attach the arm, insert and then tighten the fixing screws (A) and (B). A B 13 KODAK 9000C and KODAK 9000C 3D Extraoral Imaging System Installation Guide (SM714)_Ed

36 Installing the Cephalostat Unit 2. Turn and lift the elbow tube (A) to the height adjustment sign (B). Tighten the fixing screws of the elbow tube (A). B A 4 mm 3. Bind together the cables (B) emerging from the elbow tube (A) with adhesive tape and slide them inside the tube to avoid any damage while installing the cephalostat head. A A A B B B 5 4 INSTALLING THE CEPHALOSTAT

37 Installing the Cephalostat Unit Installing the Cephalostat Head 4. With the help of 1 person, lift the cephalostat head and insert the elbow tube (A) inside the elbow tube clamps (B). 5. Insert the elbow tube (A) as far as the height adjustment sign (C). Screw the fixing screws but do not tighten them. 6. Correctly position the cephalostat head using a spirit level (D) placed on the side of the cephalostat head. Tighten the fixing screws. A C B D 6 mm KODAK 9000C and KODAK 9000C 3D Extraoral Imaging System Installation Guide (SM714)_Ed

38 Installing the Cephalostat Unit Wiring the Cephalostat Head 7. Pull the cables out of the elbow tube (A). Remove the adhesive tape. Connect the following wiring in the following order: Cable... On the... To... (B) 967/CJ653/J10 Power supply board CN1 (C) Green and yellow cable Next to power supply box Ground screw (D) Ethernet Pano (with ferrite) Control box Pano (E) Ethernet Workstation (with ferrite) Control box Workstation (F) 964/CJ732/J29 (with ferrite) Control box Synchro B C D A F E 2.5 mm 5 6 INSTALLING THE CEPHALOSTAT

39 Installing the Cephalostat Unit Wiring the Cephalostat to the Unit Head 8. Pull the cables out of the cephalostat arm. Pass the cables through the opening in the column and attach them with a cable tie to avoid cable wear due to usage as shown in the figure below. Connect the wiring in the following order: Cable... On the... To... (A) 967/CJ653/J10 CJ653 J10 (B) Ethernet Pano Ethernet Pano (C) Ethernet Workstation Ethernet Workstation (D) 964/CJ732/J29 CJ732 J29 B CJ699 CJ653 J4 LF440 J6 J10 A J5 CJ653 C CJ732 J15 J28 J29 D CJ732 J14 J21 J18 KODAK 9000C and KODAK 9000C 3D Extraoral Imaging System Installation Guide (SM714)_Ed

40 Installing the Cephalometric Frankfort Positioning Laser (For Upgrade Only) Installing the Cephalometric Frankfort Positioning Laser (For Upgrade Only) To install the cephalometric Frankfort positioning laser, follow these steps: 1. Unscrew and remove the generator front cover. 2. Position correctly the cephalometric Frankfurt positioning laser and screw the fixing screws. 3. Connect the cable LASER CEPH on the CJ732 board to the J28 LASER CEPH connector. Attach the cable with a cable tie. 4. Turn on the unit. On the control panel, click to turn on the positioning laser beams. Make sure that the cephalometric Frankfurt positioning laser beam is projected in the middle of the ear cone. If needed, adjust the laser using the fixing screw of the laser. x2 2.5 mm x2 5 8 INSTALLING THE CEPHALOSTAT

41 Installing the Cephalostat Positioning Locker (For Upgrade Only) Installing the Cephalostat Positioning Locker (For Upgrade Only) To install the cephalostat positioning locker, follow these steps: 1. Position correctly the cephalostat positioning locker and screw the fixing screws. 2. Insert the inductive sensor s tip of the cable 982/CJ822 in the hole (A). Connect the cable connector 982/CJ822 on J3 (B). IMPORTANT When the installation of the unit is complete and the unit is turned on, the LED on the inductive tip must lit. If the LED is not lit you must insert more the tip to create the necessary electric contact. 3. From the CJ804 collimator board, solder the power supply cable 981/CJ804/J2 to the cephalostat positioning locker, with cables: Brown to - (C) Red to + (D) 4. From the CJ804 collimator board, connect cable 980/CJ804/J1 to 980/CJ732/J23 on J23 (E). B CJ699 LF440 D CJ653 C A CJ732 E KODAK 9000C and KODAK 9000C 3D Extraoral Imaging System Installation Guide (SM714)_Ed

42 Post-Installation Control Post-Installation Control Before the post-installation control, check that: The installation of the unit is complete. The installation of the KODAK dental imaging software is complete. You have the test tool. The unit is on. WARNING You MUST correctly calibrate the rotative arm of KODAK 9000 or KODAK D. If the arm calibration is not correctly executed the cephalometric mode will not be operational. (see KODAK 9000 and KODAK D Service Guide (SM712) Ed02, Using the Technical Menu, Calibration procedures, Arm Calibration ) Adjusting the Cephalostat Head Adjusting the Inclination of the Cephalostat Head To adjust the cephalostat head inclination, follow these steps: 1. Position the head clamps in a frontal AP position. 2. Access the Cephalometric Acquisition Window. In the Program pane: Click. Click for a 30 x 30 format INSTALLING THE CEPHALOSTAT

43 Post-Installation Control 3. Launch an x-ray using the x-ray remote control. The head clamps must be correctly and vertically positioned on the acquired x-ray image. Check that the blade is parallel to the image border. 4. If needed, adjust the cephalostat head mechanically using the elbow tube for horizontal inclination of the cepahlostat head. Untighten the fixing screws (A) slightly. Use the fixing screws (B) to adjust the horizontal inclination. Tighten the fixing screws (A) when you have reached the desired adjustment. IMPORTANT DO NOT adjust using the blade collimator. If necessary, adjust using the cephalometric head. 2.5 mm 4 mm x4 A B x1 A B x1 KODAK 9000C and KODAK 9000C 3D Extraoral Imaging System Installation Guide (SM714)_Ed

44 Post-Installation Control Adjusting the Height of the Cephalostat Head To adjust the height of the cephalostat head, follow these steps: 1. Position the head clamps in a lateral position. 2. Access the Cephalometric Acquisition Window. In the Program pane: Click. Click for a 30 x 30 format. 3. Launch an x-ray using the x-ray remote control. 4. In the Imaging Window, on the drawing toolbar, click. On the acquired image, draw a line from the bottom of the image to the center of the ball (ear cone). The resulting value must be: Height: between 800 and 900 pixels (between mm and 114.5mm if calibrated at the nasion support reference plane). Length: between 800 and 900 pixels (between mm and 114.5mm if calibrated at the nasion support reference plane). 800 L 900 pixels 800 H 900 pixels 5 12 INSTALLING THE CEPHALOSTAT

45 Post-Installation Control 5. If needed, adjust the height, using the cephalostat arm screws (A). Untighten slightly the screws (A) and lift the cephalostat head to the correct height. Tighten the cephalostat arm screws (A). x4 A KODAK 9000C and KODAK 9000C 3D Extraoral Imaging System Installation Guide (SM714)_Ed

46 Post-Installation Control Adjusting the Ear Cones Alignment To adjust the ear cones alignment, follow these steps: 1. Position the head clamps in a lateral position. 2. Access the Cephalometric Acquisition Window. In the Program pane: Click. Click for a 30 x 30 format. 3. Launch an x-ray using the x-ray remote control. The ear cone ball (A) must be at the center of the head clamps circle (B). 4. In the Imaging Window, on the drawing toolbar, click. On the acquired image, draw a line from the center of the ball (A) to the center of the circle (B). The resulting value must be ±10 pixels. B +10 pixels A 5 14 INSTALLING THE CEPHALOSTAT

47 Post-Installation Control 5. If needed, adjust the cephalostat head mechanically: For a horizontal adjustment, use the cephalostat arm screws (A) and (B) For a vertical adjustment, use the elbow tube clamps screws (C). C C x4 A x1 B A B x1 2.5 mm 4 mm 4 mm Adjusting the Nasion Support To adjust the nasion support, follow these steps: 1. Lower the nasion support to a vertical position. 2. Access the Cephalometric Acquisition Window. In the Program pane: Click. Click for a 30 x 30 format. KODAK 9000C and KODAK 9000C 3D Extraoral Imaging System Installation Guide (SM714)_Ed

48 Post-Installation Control 3. Launch an x-ray using the x-ray remote control. The nasion support must be perpendicular to the ear cones. 4. In the Imaging Window, on the drawing toolbar, click. On the acquired image, draw a perpendicular line between the center of the ear cones and the nasion support holes to check that the nasion support is positioned vertically. 5. If needed, adjust the nasion support, using the nasion support fixing screws (A). A 2 mm 5 16 INSTALLING THE CEPHALOSTAT

49 Post-Installation Control Selecting the Cephalometric Blade Collimator Parameters To select the cephalometric blade collimator parameters, follow these steps 1. From the fixed arm, on the control panel, on press and hold and at the same time, until you see Collimator Calibration Parameters displayed on the Display Screen of the control panel. 2. Press to enter the sub-menus. The following sub-menus are displayed: Pano: Panoramic parameters Ceph: Cephalometric parameters 3D: 3D parameters 3. Press to select Ceph. The following parameters are displayed: ORG: ORG X and ORG Y DIM: DIM X and DIM Y Facing the digital sensor: A is the left blade B is the right balde C is the top blade D is the bottom balde + Dim Y C A B + Org Y Org X + + Dim X D KODAK 9000C and KODAK 9000C 3D Extraoral Imaging System Installation Guide (SM714)_Ed

50 Post-Installation Control 4. On : Press to decrease the parameters. Press to increase the parameters. Press to modify the parameters. 5. When the parameter selection is finished, press to save the selection. 6. Press until you see Collimator Calibration Parameters menu. 7. Press to see the Exit menu. The following sub-menus are displayed: Exit and save Exit without saving 8. Press to select Exit and save. Press to save and exit INSTALLING THE CEPHALOSTAT

51 Post-Installation Control Calibrating the Cephalometic Blade Collimator To calibrate the cephalometric blade collimator, follow these steps: 1. Access the Cephalometric Acquisition Window. In the Program pane: Click. Click for a 30 x 30 format. 2. Launch an x-ray using the x-ray remote control. A white margin must be displayed on all sides of the acquired image. The acquired image must be correctly centered. 3. If needed, reselect the cephalometric blade collimator parameters and relaunch the x-ray until you obtain the correct margins (see Selecting the Cephalometric Blade Collimator Parameters). NOTE This calibration is for all image formats. KODAK 9000C and KODAK 9000C 3D Extraoral Imaging System Installation Guide (SM714)_Ed

52 Post-Installation Control Checking the Image Quality To check the image quality, follow these steps: 1. Raise the nasion support. 2. Open the head clamps. 3. Position the image quality phantom (A) at the center of the digital sensor. A 4. Access the Cephalometric Acquisition Window. In the Program pane: Click. Click for a 30 x 30 format. 5. In the Parameter pane, select: 64 Kv 15 ma 0.4 s 6. Launch an x-ray to acquire an image INSTALLING THE CEPHALOSTAT

53 Post-Installation Control 7. In the Imaging Window, on the acquired image, zoom on the value grid (B) of the image quality phantom (A). The resulting value clearly visible must be minimum 2.5 (right, left, center and bottom). Check that there is no spliting lines of the grid of the image quality phantom. B B 8. Save the image. Use this image as reference for a later image quality verification. KODAK 9000C and KODAK 9000C 3D Extraoral Imaging System Installation Guide (SM714)_Ed

54 Fitting the Covers Fitting the Covers WARNING Before fitting the covers, you MUST switch off the unit. Fitting the KODAK 9000C and KODAK 9000C 3D Covers To fit the KODAK 9000C and KODAK 9000C 3D covers, follow these steps: 1. Fit the upper cover (A) on the column. Fit the cover on the wall-bracket (B). Fit the cover on the column connecting terminal (C) and tighten the fixing screws. L 4 x M5x mm 3 mm 1 x 1 x M5x20 M5x20 A 2 x M5x50 B 3x M4x8 J 1 x M5x20 1 x M5x20 D H 2 x M5x50 3x M4x8 E C 2 x M5x12 3 x M5x12 F K G 4 x M5x INSTALLING THE CEPHALOSTAT

55 Fitting the Covers 2. Switch ON the unit. To fit the protection plates (D) and (E), press and hold to position the rotative arm so that from above the unit head, you can see the 3 screw holes on each side of the rotation axe. 3x M4x8 3x M4x8 D E 2.5 mm IMPORTANT Make sure that you slide the protection plate side with the indention for the fixing screws towards the top. 3. Switch OFF the unit. From below the unit head, slide the protection plate (D) between the rotation axe and the unit head. From above the unit head, access and tighten the accessible fixing screws. Switch ON the unit. Press and hold to position the rotative arm so that you can access the inaccessible fixing screw. Switch OFF the unit. Access and tighten the fixing screw. KODAK 9000C and KODAK 9000C 3D Extraoral Imaging System Installation Guide (SM714)_Ed

56 Fitting the Covers 4. From below the unit head, slide the protection plate (D) between the rotation axe and the unit head. From above the unit head, access and tighten the accessible fixing screws. Switch ON the unit. Press and hold to position the rotative arm so that you can access the inaccessible fixing screw. Switch OFF the unit. Access and tighten the fixing screw. 5. Fit the 2 under the head covers (F) and (G) under the head and tighten the fixing screws. 6. Fit the right side column cover (H), position and attach the ON/OFF button to the cover securely with the fixing screws. 7. Fit the left side column cover (J) on the column and tighten the fixing screws. 8. Fit the inside column cover (K) to the column and tighten the fixing screws. 9. Fit the head cover (L) of the unit to the head chassis and tighten the fixing screws. You can switch on the unit now INSTALLING THE CEPHALOSTAT

57 Fitting the Covers Fitting the Cephalostat Covers To fit the cephalostat covers, follow these steps: 1. Fit the cephalostat end of the arm cover (A) and screw the fixing screws. 2. Fit the cephalostat head side covers (B) and screw the fixing screws. 3. Fit the cephalostat head top cover (C) and screw the fixing screws. C 4 x M5x12 A 2 x M5x16 B 5 x M5x12 KODAK 9000C and KODAK 9000C 3D Extraoral Imaging System Installation Guide (SM714)_Ed

58 Fitting the Covers 5 26 INSTALLING THE CEPHALOSTAT

59 Chapter 6 Maintenance Annual Maintenance We recommend that a general inspection of the unit should be carried out every year by an approved dental systems technician. The inspection should cover the following points: Check the attachment points to the floor and the wall. Check all the mobile components of the unit. Check the x-ray generator. Check the alignment of the x-ray. Make an image acquisition with the test tools and check the image. Check the focal trough and the symmetry. Check the degree of legibility of the labels. Check for damage to cables, covers, oil leaks, etc. If the results of any of these inspections are unsatisfactory, refer to the Service Manual in order to rectify any problems. If there is any doubt, do not operate the unit. KODAK 9000C and KODAK 9000C 3D Extraoral Imaging System Installation Guide (SM714)_Ed

KODAK 9000C AND KODAK 9000C 3D Extraoral Imaging System. User Guide

KODAK 9000C AND KODAK 9000C 3D Extraoral Imaging System. User Guide KODAK 9000C AND KODAK 9000C 3D Extraoral Imaging System User Guide Notice Congratulations on your purchase of the KODAK 9000C and KODAK 9000C 3D Extraoral Imaging System. Thank you for your confidence

More information

KODAK Dental Imaging Software. Quick Start Guide

KODAK Dental Imaging Software. Quick Start Guide KODAK Dental Imaging Software Quick Start Guide Notice Congratulations on your purchase of The KODAK Dental Imaging Software. Thank you for your confidence in our products and we will do all in our power

More information

CS 9300C. User Guide

CS 9300C. User Guide CS 9300C User Guide Notice Congratulations on your purchase of the CS 9300C. Thank you for your confidence in our products and we will do all in our power to ensure your complete satisfaction. The User

More information

KODAK 9500 Cone Beam 3D System. User Guide

KODAK 9500 Cone Beam 3D System. User Guide KODAK 9500 Cone Beam 3D System User Guide Notice Congratulations on your purchase of the KODAK 9500 Cone Beam 3D System. Thank you for your confidence in our products and we will do all in our power to

More information

CS 9300 Family User Guide Including

CS 9300 Family User Guide Including CS 9300 Family User Guide Including CS 9300 CS 9300 Select Notice Congratulations on purchasing this unit of the CS 9300 Family. Thank you for your confidence in our products and we will do all in our

More information

CS 8100 and CS 8100 Access. User Guide

CS 8100 and CS 8100 Access. User Guide CS 8100 and CS 8100 Access User Guide Notice Congratulations on your purchase of the CS 8100 and CS 8100 Access. Thank you for your confidence in our products and we will do all in our power to ensure

More information

2100 Intraoral X-Ray System. User s Guide

2100 Intraoral X-Ray System. User s Guide 2100 Intraoral X-Ray System User s Guide Notice Congratulations on your purchase of the KODAK 2100 Intraoral X-ray System. Thank you for your confidence in our products and we will do all in our power

More information

Neo Ultrasound Module Manual

Neo Ultrasound Module Manual Neo Ultrasound Module Manual Installation Instructions For complete User Operating Instructions, including Cautions, Warnings, Dangers, Indications, and Contraindications, refer to the User s Manuals.

More information

RadView Manual Collimator

RadView Manual Collimator OPERATOR MANUAL RadView Manual Collimator DEL MEDICAL IMAGING CORPORATION 11550 WEST KING STREET FRANKLIN PARK, IL. 60131 TECHNICAL SUPPORT (847) 288-7000 FAX (847) 288-7011 THIS DOCUMENT CONTAINS INSTALLATION,

More information

Installation & Operation Manual SAGA1-K Series Industrial Radio Remote Control

Installation & Operation Manual SAGA1-K Series Industrial Radio Remote Control Installation & Operation Manual SAGA1-K Series Industrial Radio Remote Control Gain Electronic Co. Ltd. Table Of Contents Safety Considerations ------------------------------------------------------------2

More information

Digital Panoramic X- Ray unit. Product Data

Digital Panoramic X- Ray unit. Product Data Digital Panoramic X- Ray unit Product Data Standard examination programs Standard panoramic: adult/child panoramic exam TMJ open/close mouth: 4 slices are taken in the same image (left/right condyle, open/close

More information

2200 Intraoral X-Ray System. User s Guide

2200 Intraoral X-Ray System. User s Guide 2200 Intraoral X-Ray System User s Guide Notice Congratulations on your purchase of the KODAK 2200 Intraoral X-ray System. Thank you for your confidence in our products and we will do all in our power

More information

A510S Operation Manual

A510S Operation Manual A510S Operation Manual REV 1.1 1 Table of Contents 1 General Information 1-1 Description 1-2 Potential Operational Hazards 1-3 Technical Specifications 1-4 Instrument Overview 1-5 Function Summary 2 How

More information

incorrect use or handling that exceeds normal operating limits.

incorrect use or handling that exceeds normal operating limits. User s Manual Wall mount unit for projectors Model name HAS-WM06 Thank you for purchasing a wall mount unit specially designed for Hitachi projectors. Be sure to read this manual and the User s Manual

More information

Operator Manual. Booklet Maker BM 60. Doc No. T10171

Operator Manual. Booklet Maker BM 60. Doc No. T10171 Booklet Maker BM 60 Operator Manual Doc No. T10171 Morgana Systems Limited United Kingdom www.morgana.co.uk Telephone: ( 01908 ) 608888 Facsimile: ( 01908 ) 692399 Part no: 60249 Doc no: T10171 Rev. date:

More information

Digital Interface Option "OPC-E1-DIO"

Digital Interface Option OPC-E1-DIO Instruction Manual Digital Interface Option "OPC-E1-DIO" Thank you for purchasing our digital interface option. Read through this instruction manual and be familiar with the digital interface option before

More information

NSPL-500. AIS/VHF antenna splitter. User Manual ENGLISH.

NSPL-500. AIS/VHF antenna splitter. User Manual ENGLISH. NSPL-500 AIS/VHF antenna splitter User Manual ENGLISH www.bandg.com www.simrad-yachting.com www.lowrance.com Preface As Navico is continuously improving this product, we retain the right to make changes

More information

Top spin Nr /

Top spin Nr / Top spin Nr. 1840 0000 / 1840 1000 Bedienungsanleitung 21-6680 28052014 / A Made in Germany Ideas for dental technology Top spin Nr. 1840 0000 / 1840 1000 Contents 1. Introduction...2 1.1 Symbols...2 2.

More information

Antenna Splitter ASA 1. Instruction manual

Antenna Splitter ASA 1. Instruction manual Antenna Splitter ASA 1 Instruction manual Contents Contents Important safety instructions... 2 The ASA 1 active antenna splitter... 4 Combination possibilities of ASA 1/ASA 1-1G8... 5 Delivery includes...

More information

CDR PanElite. j~áåíéå~ååé=fåëíêìåíáçåë==== English

CDR PanElite. j~áåíéå~ååé=fåëíêìåíáçåë==== English CDR PanElite j~áåíéå~ååé=fåëíêìåíáçåë==== English Schick Technologies, Inc. Maintenance Instructions CDR PanElite General information ATTENTION! Proper shielding of room and operator position is essential.

More information

USER MANUAL ENGLISH 1450 COIN COUNTER & SORTER

USER MANUAL ENGLISH 1450 COIN COUNTER & SORTER USER MANUAL ENGLISH 1450 COIN COUNTER & SORTER INTRODUCTION ENGLISH Thank you for purchasing the Safescan 1450 coin counter and sorter. For proper use and maintenance, we advise to read this user manual

More information

2011 / Circuit Tracer

2011 / Circuit Tracer INSTRUCTION MANUAL 2011 / 00521 Circuit Tracer Read and understand all of the instructions and safety information in this manual before operating or servicing this tool. 52044992 2008 Greenlee Textron

More information

LASER ENHANCED REVOLVER GRIP OWNER S MANUAL RED LASER GREEN LASER

LASER ENHANCED REVOLVER GRIP OWNER S MANUAL RED LASER GREEN LASER LASER ENHANCED RED LASER GREEN LASER REVOLVER GRIP OWNER S MANUAL LASER ENHANCED GRIP Installation Instructions Caution... 3 Safety Labels... 4 Installation...5-7 Programming...8-10 Batteries (Red Laser)...

More information

WARNING: We recommend that you consult the Safety, Regulatory and Technical Specification User Guide before using the CS 2200.

WARNING: We recommend that you consult the Safety, Regulatory and Technical Specification User Guide before using the CS 2200. CS 2200 User Guide Notice Congratulations on your purchase of the CS 2200. Thank you for your confidence in our products and we will do all in our power to ensure your complete satisfaction. The User Guide

More information

PRODUCT MANUAL 5-IN-1 MITER SAW STATION *SAVE THIS MANUAL FOR FUTURE REFERENCE

PRODUCT MANUAL 5-IN-1 MITER SAW STATION *SAVE THIS MANUAL FOR FUTURE REFERENCE PRODUCT MANUAL 5-IN-1 MITER SAW STATION *SAVE THIS MANUAL FOR FUTURE REFERENCE 92778 1 GENERAL SAFETY INFORMATION For your safety, please read these instructions carefully before use and keep them in the

More information

Instruction Manual. Self-Leveling Cross-Line Laser Level Model No and

Instruction Manual. Self-Leveling Cross-Line Laser Level Model No and 3728H_Manuals 9/2/10 8:02 AM Page 1 Self-Leveling Cross-Line Laser Level Model No. 40-6620 and 40-6625 Instruction Manual Congratulations on your choice of this Self-Leveling Cross-Line Laser Level. We

More information

RVG 5100 and RVG User Guide

RVG 5100 and RVG User Guide RVG 5100 and RVG 6100 User Guide Notice The User Guide for the RVG 5100 and RVG 6100 includes information on the devices as well as their usage. We recommend that you thoroughly familiarize yourself with

More information

Operating Instructions

Operating Instructions 02 ENG 28.01.2008 9:14 Uhr Seite 11 Contents 1 Declaration of Compliance 2 Proper Use 3 Overview 4 General Safety Rules 5 Specific Safety Rules 6 Operation 6.1 Switching the paint remover On/Off 6.2 Locking

More information

2-Slot Desktop Chassis (DC) Extended Temperature

2-Slot Desktop Chassis (DC) Extended Temperature APRIL 2008 LMC5202A 2-Slot Desktop Chassis (DC) Extended Temperature Copyright 2008. Black Box Corporation. All rights reserved 50 80105BB 01 A0 1000 Park Drive Lawrence, PA 35055 1018 724 746 5500 Fax

More information

С 800 CASSIDA C 800 HIGH SPEED COIN COUNTER

С 800 CASSIDA C 800 HIGH SPEED COIN COUNTER С 800 CASSIDA C 800 HIGH SPEED COIN COUNTER This manual contains important information on safety measures and operational features. Please read it carefully before operating your coin counter, and keep

More information

TUCANA-02 P PORTABLE END MILLING MACHINE USER S MANUAL

TUCANA-02 P PORTABLE END MILLING MACHINE USER S MANUAL TUCANA-02 P PORTABLE END MILLING MACHINE USER S MANUAL 1 CONTENTS Page 1. General Information 3 1.1. Introduction 3 1.2. Manufacturer 3 2. Machine s Description and Purpose of Use 3 2.1. Machine s description

More information

KODAK 2200 Intraoral X-ray System. User s Manual

KODAK 2200 Intraoral X-ray System. User s Manual KODAK 2200 Intraoral X-ray System User s Manual This document is originally written in English. Manual Name: KODAK 2200 Intraoral X-ray System, User Guide Document code: SM731 Revision Number: Rev 01 Printed

More information

MAN Revision 002. InSight Mini C-arm Imaging System Technical Reference Manual

MAN Revision 002. InSight Mini C-arm Imaging System Technical Reference Manual MAN-00668 Revision 002 InSight Mini C-arm Imaging System Technical Reference Manual April 2007 The information contained in this Manual is confidential and proprietary to Hologic, Inc. This information

More information

GENERAL OPERATIONAL PRECAUTIONS

GENERAL OPERATIONAL PRECAUTIONS GENERAL OPERATIONAL PRECAUTIONS WARNING! When using electric tools, basic safety precautions should always be followed to reduce the risk of fire, electric shock and personal injury, including the following.

More information

OBSOLETE. Progeny Vantage Panoramic X-ray System. Installation Guide For Vantage units manufactured during or after October 2012

OBSOLETE. Progeny Vantage Panoramic X-ray System. Installation Guide For Vantage units manufactured during or after October 2012 Progeny Vantage Panoramic X-ray System Installation Guide For Vantage units manufactured during or after October 2012 0120 Part Number: 00-02-1608 ECN: P3179 REV. G Contents Contents... 2 1 Regulatory

More information

VARIABLE SPEED WOOD LATHE

VARIABLE SPEED WOOD LATHE MODEL MC1100B VARIABLE SPEED WOOD LATHE INSTRUCTION MANUAL Please read and fully understand the instructions in this manual before operation. Keep this manual safe for future reference. Version: 2015.02.02

More information

Thank you for buying this AIS antenna splitter.

Thank you for buying this AIS antenna splitter. Thank you for buying this AIS antenna splitter. This product has been engineered to offer you the highest level of performance and durability and we hope that it will provide many years of reliable service.

More information

Performance Meets Simplicity

Performance Meets Simplicity Performance Meets Simplicity Cone Beam 3D Imaging Systems P Panoramic X-ray Systems Intraoral X-ray Systems Digital Intraoral Sensors Digital X-ray Phosphor Plates Intraoral Cameras Imaging Software One

More information

Film2USB Converter. Quick Start Guide & User s Manual. Model Number F2USB-05

Film2USB Converter. Quick Start Guide & User s Manual.   Model Number F2USB-05 Film2USB Converter Quick Start Guide & User s Manual Model Number F2USB-05 www.film2usb.com www.clearclicksoftware.com Table of Contents Safety Precautions... 3 Package Contents... 4 Description of Parts...

More information

KODAK RVG 6100 and 5100 Digital Radiography Systems

KODAK RVG 6100 and 5100 Digital Radiography Systems Guide_RVG6100_5100_GB 21/08/06 14:31 Page 1 Quick Start Guide KODAK RVG 6100 and 5100 Digital Radiography Systems Guide_RVG6100_5100_GB 21/08/06 14:31 Page 2 Warnings & Safety Instructions Always use the

More information

Tapping Screw (W/Flange) 46 Cord Armor 47 Tube (D) 48 Cord. 45 Cord Clip. Tapping Screw (W/Flange) 10 Gear Cover Ass'y. 12 Socket (B) Ass'y

Tapping Screw (W/Flange) 46 Cord Armor 47 Tube (D) 48 Cord. 45 Cord Clip. Tapping Screw (W/Flange) 10 Gear Cover Ass'y. 12 Socket (B) Ass'y W8VB The exploded assembly drawing should be used only for authoized service center. W8VB Item No. Part time 1 Magnetic Hex. Socket 2 Sub Stopper 3 O-Ring (S-16) 4 Locator (A) 5 Lock Sleeve (A) 6 O-Ring

More information

Atomscope PORTABLE VETERINARY X-RAY EQUIPMENT INSTRUCTION MANUAL. Version 1.0

Atomscope PORTABLE VETERINARY X-RAY EQUIPMENT INSTRUCTION MANUAL. Version 1.0 Atomscope PORTABLE VETERINARY X-RAY EQUIPMENT INSTRUCTION MANUAL Version 1.0 Thank you for purchasing our HFX90V portable veterinary x-ray unit. We are confident that you will be pleased with the radiographs

More information

PORTROD WALL MOUNTED HEIGHT ROD

PORTROD WALL MOUNTED HEIGHT ROD PORTROD WALL MOUNTED HEIGHT ROD USER INSTRUCTIONS P/N UMPORTROD Rev 1-052112 1 PORTROD Thank you for your purchase of this Health o meter Professional product. Please read this manual carefully, and keep

More information

Operating instructions Diffuse reflection sensor with background suppression O1D101 O1D / / 2018

Operating instructions Diffuse reflection sensor with background suppression O1D101 O1D / / 2018 Operating instructions Diffuse reflection sensor with background suppression O1D101 O1D104 706114 / 00 02 / 2018 Contents 1 Preliminary note...3 1.1 Symbols used...3 1.2 Warning signs used...3 2 Safety

More information

Wall Mount Bracket YM-81

Wall Mount Bracket YM-81 CN/JA/EN/DE/FR/ES/IT/SV/RU/PT/TR/NL/PL/FI/CS/NO/LT/TH/ID/MS/VI/TW/KO/AR Wall Mount Bracket YM-81 English User s Guide Deutsch Bedienungsanleitung Français Mode d emploi Español Italiano Svenska Português

More information

FORD MOUNT INSTALLATION INSTRUCTIONS

FORD MOUNT INSTALLATION INSTRUCTIONS WESTERN PRODUCTS, P.O. BOX 245038, MILWAUKEE, WI 53224-9538 Lit. No. 64289 June 1, 2003 FORD MOUNT INSTALLATION INSTRUCTIONS Bronco (4X4 only) F-150 (4X4 only), F-250/350 Super Duty 1980 1991 Model No.

More information

GroundControl. Follow instructions contained in this manual. Incorrect installation could result in serious injury or damage to property.

GroundControl. Follow instructions contained in this manual. Incorrect installation could result in serious injury or damage to property. GroundControl TM use supplied hardware Use only hardware supplied in your GroundControl kit or supplied by an authorized YAKIMA dealer. Use of unauthorized parts in the GroundControl system could result

More information

Instruction manual ADN-W AM. Antenna Module

Instruction manual ADN-W AM. Antenna Module Instruction manual ADN-W AM Antenna Module For your safety www www.sennheiser.com Manual Contents For your safety... 1 The ADN-W AM antenna module... 1 Package contents... 2 Components required for wireless

More information

Watcheye S AIS Splitter. manual

Watcheye S AIS Splitter. manual Watcheye S AIS Splitter manual Thank you for buying this AIS antenna splitter. This product has been engineered to offer you the highest level of performance and durability and we hope that it will provide

More information

HIGH PERFORMANCE MARITIME. em-trak S100 PRODUCTS. Antenna splitter. Product Manual. High Performance Maritime Products

HIGH PERFORMANCE MARITIME. em-trak S100 PRODUCTS. Antenna splitter. Product Manual.  High Performance Maritime Products em-trak S100 MARITIME PRODUCTS HIGH PERFORMANCE Antenna splitter Product Manual High Performance Maritime Products www.em-trak.com 201-0206:3 Contents 1 - Notices...1 1.1 - Safety warnings...1 1.2 - General

More information

Senior Swing. Caution

Senior Swing. Caution *740132* 740132 2800 Overhead Concealed Series 9500 Surface Applied Series Senior Swing Installation Instructions Caution LCN Senior Swing The Senior Swing Power Operator System is a low energy product

More information

S E L E C T I O N. Arm Curl. User manual

S E L E C T I O N. Arm Curl. User manual S E L E C T I O N T H E S T R E N G T H E V O L U T I O N User manual The identification plate of the and manufacturer, affixed behind the seat, gives the following details: A Name and address of the manufacturer

More information

Picasso-Trio 3-D Dental Imaging

Picasso-Trio 3-D Dental Imaging Picasso-Trio 3-D Dental Imaging Digital Panoramic & Cephalometric, CBCT X-ray Imaging System 3 in 1 System Product Data Model Picasso-Trio - NP Panoramic only [CT upgradeable] Picasso-Trio - NC Panoramic

More information

Single Sided Height Adjustable Benching

Single Sided Height Adjustable Benching Single Sided Height Adjustable Benching Bit Holder Power Drill Long # Square Drive Bit 5mm Hex Drive Bit T-30 Torx Drive Bit Hex Key -.5mm (Provided) SERIES 5 Hex Key - 3mm Socket Adaptor (or) If you have

More information

GENERAL OPERATIONAL PRECAUTIONS WARNING! When using electric tools, basic safety precautions should always be followed to reduce the risk of fire, electric shock and personal injury, including the following.

More information

Table of Contents. Makita LD050P 1

Table of Contents. Makita LD050P 1 Table of Contents EN Instrument Set-up - - - - - - - - - - - - - - - - - - - - - - - Introduction- - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - Overview - - - - - - - - - - - - - - - - -

More information

HOLE CUTTER SHARPENER ASSEMBLY & SERVICE MANUAL

HOLE CUTTER SHARPENER ASSEMBLY & SERVICE MANUAL HOLE CUTTER SHARPENER ASSEMBLY & SERVICE MANUAL WARNING You must thoroughly read and understand this manual before operating the equipment, paying particular attention to the Warning & Safety instructions.

More information

the slim retracting clothesline product manual

the slim retracting clothesline product manual &6 the slim retracting clothesline product manual for 4 and 6-line models HLS0046_4-6 line_owners Manual_v6.indd 1 2 Introduction Congratulations Congratulations on the purchase of your new Hills slim

More information

Loading Dock Safety Gate

Loading Dock Safety Gate Installation Instructions/Operation and Maintenance Manual Models LDSG-120-PCY LDSG-144-PCY Table of Contents Product Information...2 Parts List...3 Installation Instructions...5 Operation...13 Inspection

More information

OPERATION & SERVICE MANUAL FOR FC 110 AC POWER SOURCE

OPERATION & SERVICE MANUAL FOR FC 110 AC POWER SOURCE OPERATION & SERVICE MANUAL FOR FC 100 SERIES AC POWER SOURCE FC 110 AC POWER SOURCE VERSION 1.3, April 2001. copyright reserved. DWG No. FC00001 TABLE OF CONTENTS CHAPTER 1 INTRODUCTION... 1 1.1 GENERAL...

More information

D C 01/2019 3

D C 01/2019 3 D-0117968-C 01/2019 3 4 D-0117968-C 01/2019 Screw Driver Screw Driver Unplug both the Red & Blue connectors. (see above) Place a small flat head screw driver on the small orange tabs and push down while

More information

Please read BOTH these Installation Instructions and the General Instructions prior to installing or operating this equipment.

Please read BOTH these Installation Instructions and the General Instructions prior to installing or operating this equipment. Attachment Tab Height: 16-1/2 Serial Number Attachment Tab Width: 24 Please read BOTH these and the General Instructions prior to installing or operating this equipment. 1. Blue Ox towing products and

More information

USER'S GUIDE. EcoQuest International 310 T. Elmer Cox Drive Greeneville, TN (423) Setup. Operation. Specifications.

USER'S GUIDE. EcoQuest International 310 T. Elmer Cox Drive Greeneville, TN (423) Setup. Operation. Specifications. USER'S GUIDE Setup Operation Specifications Warranty (43) 638-746 CAUTION: Read manual carefully for proper procedures and operation. CONTENTS SPECIFICATIONS Specifications...............................................................3

More information

AC 3. Active Antenna Combiner. Instruction manual

AC 3. Active Antenna Combiner. Instruction manual AC 3 Active Antenna Combiner Instruction manual Contents Contents Important safety instructions... 2 The AC 3 active transmitter combiner... 4 Delivery includes... 4 Operating controls... 5 Block diagram...

More information

eflex 75x/300x Digital Microscope with Flexible Neck User Guide

eflex 75x/300x Digital Microscope with Flexible Neck User Guide eflex 75x/300x Digital Microscope with Flexible Neck User Guide Table of Contents Important information... 4 Product description... 4 Computer requirements... 5 Windows based PC... 5 Mac OS based PC...

More information

ROTARY HAMMER OWNER'S MANUAL

ROTARY HAMMER OWNER'S MANUAL ROTARY HAMMER OWNER'S MANUAL WARNING: Read carefully and understand all INSTRUCTIONS before operating. Failure to follow the safety rules and other basic safety precautions may result in serious personal

More information

Connevans.info. DeafEquipment.co.uk. This product may be purchased from Connevans Limited secure online store at

Connevans.info. DeafEquipment.co.uk. This product may be purchased from Connevans Limited secure online store at Connevans.info Solutions to improve the quality of life Offering you choice Helping you choose This product may be purchased from Connevans Limited secure online store at www.deafequipment.co.uk DeafEquipment.co.uk

More information

Maintenance and other important non-personal injury and non-material damage instructions or statements that should be observed.

Maintenance and other important non-personal injury and non-material damage instructions or statements that should be observed. VIGO INDUSTRIES INSTALLATION GUIDE FOR SHOWER ENCLOSURE ()! SAFETY PRECAUTIONS This Installation Guide uses the following symbols to indicate important information. Always observe the instructions indicated

More information

OBSOLETE. Progeny Vantage Panoramic X-ray System. Installation Guide For Vantage units manufactured during or after October 2012

OBSOLETE. Progeny Vantage Panoramic X-ray System. Installation Guide For Vantage units manufactured during or after October 2012 Progeny Vantage Panoramic X-ray System Installation Guide For Vantage units manufactured during or after October 2012 Kit #60-A2050 Part Number: 00-02-1608 REV. F ECN: P3031 REV. F Contents Contents...

More information

loqelmelp=pi=oa=l=loqelmelp=pi=oa=`ééü= loqelmelp=pi=pa=l=loqelmelp=pi=pa=`ééü båöäáëü=erpf

loqelmelp=pi=oa=l=loqelmelp=pi=oa=`ééü= loqelmelp=pi=pa=l=loqelmelp=pi=pa=`ééü båöäáëü=erpf kéï=~ë=çñw= MTKOMNR loqelmelp=pi=oa=l=loqelmelp=pi=oa=`ééü= loqelmelp=pi=pa=l=loqelmelp=pi=pa=`ééü j~áåíéå~ååé=fåëíêìåíáçåë båöäáëü=erpf Cover page = 2 D3632.103.01.01.23 07.2015 Table of contents Table

More information

LED PowerBlinder 4 ORDERCODE 41320

LED PowerBlinder 4 ORDERCODE 41320 LED PowerBlinder 4 ORDERCODE 41320 Congratulations! You have bought a great, innovative product from Showtec. The Showtec LED PowerBlinder brings excitement to any venue. Whether you want simple plug-&-play

More information

Please read and understand the OnBoard Timpani Cart Owner s Manual before using the Timpani Cart.

Please read and understand the OnBoard Timpani Cart Owner s Manual before using the Timpani Cart. Assembly and Owner s Manual OnBoard Timpani Cart Performance Position Towing Position CONTENTS Important User Information...........................2 General......................................2 Manufacturer.................................2

More information

Quolis 5000 Light INSTALLATION AND OPERATION INSTRUCTIONS. Model AL-820 IMPORTANT

Quolis 5000 Light INSTALLATION AND OPERATION INSTRUCTIONS. Model AL-820 IMPORTANT Quolis 5000 Light Model AL-80 INSTALLATION AND OPERATION INSTRUCTIONS IMPORTANT After installation is completed, check all the bolts, screws and fasteners to confirm that they are securely fastened. R

More information

Displacement Sensor CD5 Series

Displacement Sensor CD5 Series Sensor Head Instruction Manual Displacement Sensor CD5 Series Laser Type (CD5-L_25,CD5-_30,CD5-_85,CD5-_150,CD5-W350,CD5-W500,CD5-W2000) Table of Contents page SAFETY PRECAUTIONS... 3 Meanings of Safety

More information

TETRIS User's Guide. High Impedance Active Probe DO177-1

TETRIS User's Guide. High Impedance Active Probe DO177-1 TETRIS 1500 High Impedance Active Probe User's Guide DO177-1 TETRIS 1500 Copyright 2010 Ltd. All rights reserved. Information in this publication supersedes that in all previously published material. Specifications

More information

RIGOL. User s Guide. RP1000D Series High Voltage Differential Probe. Feb RIGOL Technologies, Inc

RIGOL. User s Guide. RP1000D Series High Voltage Differential Probe. Feb RIGOL Technologies, Inc User s Guide RP1000D Series High Voltage Differential Probe Feb. 2013 RIGOL Technologies, Inc Guaranty and Declaration Copyright 2012 RIGOL Technologies, Inc. All Rights Reserved. Trademark Information

More information

MRI SIMULATOR PST OPERATOR MANUAL

MRI SIMULATOR PST OPERATOR MANUAL MRI SIMULATOR PST-100355 OPERATOR MANUAL MRI Simulator Operator Manual PST-100444 Rev 7 Copyright Copyright 2012 Psychology Software Tools, Inc. All rights reserved. The information in this document is

More information

VARIABLE SPEED WOOD LATHE. Model DB900 INSTRUCTION MANUAL

VARIABLE SPEED WOOD LATHE. Model DB900 INSTRUCTION MANUAL VARIABLE SPEED WOOD LATHE Model DB900 INSTRUCTION MANUAL 1007 TABLE OF CONTENTS SECTION...PAGE Technical data.. 1 General safety rules....1-3 Specific safety rules for wood lathe.....3 Electrical information.4

More information

Knife grinding machine K3 H/K

Knife grinding machine K3 H/K Knife grinding machine K3 H/K 01.11.2010 GS-Schleiftechnik, Leyher Str. 61 a, D-90431 Nürnberg Phone 0049/9193/4404, Fax 0049/9193/4391 e-mail: info@gs-de.eu 1 Table of Contents A. SAFETY... 3 A.1 THE

More information

VueNet EasyLink. ver User`s manual External wireless video and audio transmission system for IP cameras HD and UHD in the band

VueNet EasyLink. ver User`s manual External wireless video and audio transmission system for IP cameras HD and UHD in the band VueNet EasyLink ver. 1.2 User`s manual External wireless video and audio transmission system for IP cameras HD and UHD in the band 5.1-5.8 GHz Thank you for choosing the VueNet EasyLink. We trust that

More information

SUNNY BEAM REPEATER Transmission Range Increase for Sunny Beam

SUNNY BEAM REPEATER Transmission Range Increase for Sunny Beam SUNNY BEAM REPEATER Transmission Range Increase for Sunny Beam User Manual SBeamRep-BEN091911 98-0002611 Version 1.1 EN SMA Solar Technology AG Table of Contents Table of Contents 1 Notes on this Manual..............................

More information

FS-7006PA FS-7012PA YA-7000 POWER AMPLIFIERS. AMPLIFIER AUTO SWITCHING MODULE (Optional) OPERATING INSTRUCTIONS

FS-7006PA FS-7012PA YA-7000 POWER AMPLIFIERS. AMPLIFIER AUTO SWITCHING MODULE (Optional) OPERATING INSTRUCTIONS OPERATING INSTRUCTIONS POWER AMPLIFIERS AMPLIFIER AUTO SWITCHING MODULE (Optional) FS-7006PA FS-7012PA YA-7000 Thank you for purchasing TOA's Power Amplifiers. Please carefully follow the instructions

More information

On-Line Cardio Theater Wireless Digital Transmitter Installation and Instruction Manual

On-Line Cardio Theater Wireless Digital Transmitter Installation and Instruction Manual On-Line Cardio Theater Wireless Digital Transmitter Installation and Instruction Manual Full installation instructions accompany your Cardio Theater equipment order. This On-Line version of our Installation/Instruction

More information

METAL DETECTOR INSTRUCTION GUIDE

METAL DETECTOR INSTRUCTION GUIDE METAL DETECTOR INSTRUCTION GUIDE SET UP STEP 1. STEP 2. Your NATIONAL GEOGRAPHIC detector requires no assembly or tools. Simply remove the detector from the box. Press down the red UNLOCK button on both

More information

Tube Facing Tool.

Tube Facing Tool. www.swagelok.com Tube Facing Tool This manual contains important information for the safe and effective operation of the Swagelok TF72 series tube facing tool. Users should read and understand its contents

More information

User Manual. Flooring Removal Machine (Patent Pending) Maintenance and Operating Instructions

User Manual. Flooring Removal Machine (Patent Pending) Maintenance and Operating Instructions User Manual Flooring Removal Machine (Patent Pending) Maintenance and Operating Instructions Copyright 2012 by Carpet Concepts LLC. All Rights Reserved. For technical questions or replacement parts please

More information

DE EN FR IT ES RU PL DA NL NO SV

DE EN FR IT ES RU PL DA NL NO SV DE FR IT ES RU PL DA NL NO SV Table of Contents Instrument Set-up - - - - - - - - - - - - - - - - - - - - - - - Introduction- - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - Overview - - -

More information

TH450A-T TH550A-T THP550-T/TS3000

TH450A-T TH550A-T THP550-T/TS3000 0 TH450A-T TH550A-T THP550-T/TS3000 INSTRUCTION MANUAL CEILING TYPE (OVERHEAD TRAVELING TYPE) INDUSTRIAL ROBOT SPECIFICATIONS Notice 1. Make sure that this instruction manual is delivered to the final

More information

PROFESSIONAL STEREO AMPLIFIERS USER GUIDE

PROFESSIONAL STEREO AMPLIFIERS USER GUIDE PROFESSIONAL STEREO AMPLIFIERS USER GUIDE 967/9673 - September 00 - Version.0 ENGLISH Ti SERIES - Professional stereo amplifiers Page Ti SERIES - Professional stereo amplifiers ENGLISH - Safety information

More information

RIGOL. User s Guide. RP5600 Passive Probe. July 2010 RIGOL Technologies, Inc.

RIGOL. User s Guide. RP5600 Passive Probe. July 2010 RIGOL Technologies, Inc. User s Guide RP5600 Passive Probe July 2010 RIGOL Technologies, Inc. Guaranty and Declaration Copyright 2010 RIGOL Technologies, Inc. All Rights Reserved. Trademark Information RIGOL is a registered trademark

More information

Progeny Vantage Panoramic X-ray System

Progeny Vantage Panoramic X-ray System Progeny Vantage Panoramic X-ray System Installation Guide Kit #60-A2050 Part Number: 00-02-1608 REV. A ECN: P2019 REV. B Contents 1 Regulatory Information... 3 Indications for Use... 3 Contraindications...

More information

S E L E C T I O N. Upper Back. User manual

S E L E C T I O N. Upper Back. User manual and S E L E C T I O N T H E S T R E N G T H E V O L U T I O N User manual and and The identification plate of the and manufacturer, affixed to the frame on the side opposite the padded rest, gives the

More information

Model 5100F. Advanced Test Equipment Rentals ATEC (2832) OWNER S MANUAL RF POWER AMPLIFIER

Model 5100F. Advanced Test Equipment Rentals ATEC (2832) OWNER S MANUAL RF POWER AMPLIFIER Established 1981 Advanced Test Equipment Rentals www.atecorp.com 800-404-ATEC (2832) OWNER S MANUAL Model 5100F RF POWER AMPLIFIER 0.8 2.5 GHz, 25 Watts Ophir RF 5300 Beethoven Street Los Angeles, CA 90066

More information

Media Storage Systems Fixed Media Cabinets

Media Storage Systems Fixed Media Cabinets Owner s Manual Media Storage Systems Fixed Media Cabinets Contents 1-Column Fixed Media Storage Cabinet Important User Information...........................2 Safety Precautions.................................3

More information

MaxLite Linear Strip ECO Series

MaxLite Linear Strip ECO Series General Safety Information To reduce the risk of death, personal injury or property damage from fire, electric shock, falling parts, cuts/abrasions, and other hazards read all warnings and instructions

More information

UPLIFT 2-Leg Height Adjustable Standing Desk

UPLIFT 2-Leg Height Adjustable Standing Desk UPLIFT -Leg Height Adjustable Standing Desk Also watch our assembly video http://bit.ly/9ywwh DIRECTIONS FOR ASSEMBLY AND USE TABLE OF CONTENTS PAGE Safety and Warnings Usage Parts List Assembly Instructions

More information

Random & Sinusoidal Jitter Injector. Main Unit Operation Manual

Random & Sinusoidal Jitter Injector. Main Unit Operation Manual Random & Sinusoidal Jitter Injector RJI12G Main Unit Operation Manual Rev 1.0 September 2012 Introduction... 2 Safety Instruction... 2 1. General... 4 1 1 Features... 4 1 2 Functions & Characteristics...

More information

MANUAL SEALLESS STEEL STRAPPING TOOL MODEL A332

MANUAL SEALLESS STEEL STRAPPING TOOL MODEL A332 OPERATION MANUAL / SPARE PARTS LIST MANUAL SEALLESS STEEL STRAPPING TOOL MODEL A332 13.2250.01 INDEX PAGE 1 SAFETY INSTRUCTIONS 2 2 WARRANTY CONDITIONS AND LIABILITY 3 3 APPROPRIATE USE 3 4 TECNICAL DATA

More information

FORTUS 360mc/400mc and FDM 360mc/400mc 3D Production System. Site Preparation Guide Rev E

FORTUS 360mc/400mc and FDM 360mc/400mc 3D Production System. Site Preparation Guide Rev E FORTUS 360mc/400mc and FDM 360mc/400mc 3D Production System Site Preparation Guide 107167-0005 Rev E Liability Statement The information in this document is subject to change without notice. Stratasys,

More information

Installation Guide. English. English

Installation Guide. English. English Installation Guide Safety Instructions For your safety, read all the instructions in this guide before using the setting plate. Incorrect handling that ignores instructions in this guide could damage the

More information