ENGINEERING TEST REPORT #: A LSR JOB #: C-2411

Similar documents
ENGINEERING TEST REPORT # C LSR Job #: C-2411 Compliance Testing of: RM186-SM

L.S. Compliance, Inc. W66 N220 Commerce Court Cedarburg, WI

Laird Attn: Bill Steinike W66 N220 Commerce Ct. Cedarburg, WI Report Constructed by: Zach Wilson, EMC Technician Signature: Date: June 21, 2017

TEST REPORT... 1 CONTENT...

TEST REPORT # TCB LSR Job #:C-626

Advanced Compliance Solutions, Inc FAU Blvd, Suite 310 Boca Raton, Florida (561)

EN 55015: 2013 Clause Pass. EN 55015: 2013 Clause Pass. EN 55015: 2013 Clause Pass

EMC Test Report. Report Number: M030826

Harmonic Current emission EN :2014 Class A Pass. Voltage Fluctuation and Flicker EN :2013 Clause 5 Pass

EMC TEST REPORT. Report No.: TS EME Model No.: 33XR-A Issued Date: Jan. 08, 2009

Discontinuous Disturbance (Click) EN :2006+A1:2009+A2:2011 Clause N/A** Radiated Emission, 30MHz to 1000MHz

EMC TEST REPORT for LEDELS LIGHTING CO., LTD. LED module Model No. : LL-F12T4815X6B

EMC TEST REPORT For MPP SOLAR INC Inverter/ Charger Model Number : PIP 4048HS

Test Certificate A sample of the following product received on May 13, 2011 and tested on May 13, 18 and 19, 2011 complied with the requirements of,

2310 to 2390 MHz, 3m distance MCS8 (MIMO) to 2500 MHz Restricted band MCS8 (MIMO)

EMC standards. Presented by: Karim Loukil & Kaïs Siala

EMC REPORT. ShenZhen KY Technology Co.,Ltd. No.369, BaoTian 1st RD, TieGang Industrial Park, Xixiang Town, Baoan District, ShenZhen, PRC.

EMC TEST REPORT for : DONGGUAN EVER DEVELOPMENT ELECTRONIC CO., Electronic calculator Model No.: KF15758

2620 Modular Measurement and Control System

Report for Excelsys EMC Measurements for 6Xgen Purchase Order: by Project Number EMT08J027

EMC REPORT DONGGUAN FIT-WATCH CO., LTD. 18#,Hedong No.1 road,jinsha village,changan town, Dongguan City, Guangdong Province.

Sunlight Supply, Inc.

TRANSMITTER MODEL: KAS-2030M

ETSI EN V2.2.0 ( ) ETSI EN V2.1.1 ( ) TEST REPORT. United GULF GATE Co.

CENTRE OF TESTING SERVICE INTERNATIONAL

EN 55022: 2010+AC:2011 Clause 6.1 Pass. Harmonic Current EN :2006+A1:2009+A2:2009 Class A N/A

Report for Excelsys EMC Measurements for 4Xgen Purchase Order: Project Number EMT07J026 Rev. B

BS EN TESTS ON THE IT TOXIC GAS DETECTOR MODULE

FCC CFR47 PART 15 SUBPART C INDUSTRY CANADA RSS-GEN AND RSS-210 CERTIFICATION TEST REPORT FOR BROADCOM BLUETOOTH MODULE MODEL NUMBER: BCM92046MD

Test Report. Ningbo Goldmore Industrial Co.,Ltd.

EN61326 EMC COMPLIANCE REPORT on the LP Series Ultrasonic Transmitter Remote Amplifier and Transducer for Hawk Measurement Systems Pty Ltd

EMC VERIFICATION SUMMARY Report No.: SZHH

Test Report GUANGDONG EAST POWER CO., LTD

TEST REPORT FROM RFI GLOBAL SERVICES LTD

Certificate of Test AND KEEPS ALL REQUIREMENTS ACCORDING THE FOLLOWING REGULATIONS IEC :2001 IEC :2007

EMC TEST REPORT. For. Boardcon Technology Limited. MINI9G45 Computer on Module. Model No. : MINI9G45

1 SUMMARY OF STANDARDS AND RESULTS

EMC TEST REPORT. NORTE SIRIUS ENTERPRISE CO., LTD , Shin-Sheng St., Chung-Ho Dist, New Taipei City, Taiwan

FCC 47 CFR PART 15 SUBPART C INDUSTRY CANADA RSS-210 ISSUE 8 BLUETOOTH LOW ENERGY CERTIFICATION TEST REPORT FOR. 2.4GHz LE MODULE MODEL NUMBER: RN4020

Test Report. Guangdong East Power Co., Ltd. Fully Automatic AC Voltage Regulator. Brand Name:

Nemko Canada Inc., 303 River Road, R.R. 5, Ottawa, Ontario, Canada, K1V 1H2

TABLE OF CONTENTS. Report No.: SH E03 1. TEST RESULT CERTIFICATION...3

BS EN TESTS ON THE RS-485 FFP2 THE BRAIN REVISION C

ACCORDING TO: FCC part 15 subpart C, and subpart B FOR:

C-Tick EMC TEST REPORT. For. AC/DC Switching Power Supply

EMC TEST REPORT For ShenZhen ZhangQing Electronic LTD poe tester Model No.: WS POE Tester/Detector Additional Model No: Please Refer to Page 8

EMC TEST REPORT - Addendum

Prepared for Address. : KST DIGITAL TECHNOLOGY LIMITED : No.226, Pangu Street, Meixian, Meizhou, Guangdong. Prepared by

ELECTRICAL TESTING ACCORDING TO: FCC 47CFR part 15 subpart C (DTS) and subpart B FOR:

FCC & RSS-216 (Class II Permissive Change) Wireless Power Transfer Report. for A Acer Incorporated

Page 1 of 51 Report No.: T TEST REPORT FCC ID: 2AGJ5WAP-30. In Accordance with: FCC PART 15, SUBPART C : 2015 (Section 15.

EN v3.1.1 TEST REPORT FOR. Bluetooth Serial Port Module MODEL NUMBER: LMX9838 REPORT NUMBER: 14U17620-E1V3 ISSUE DATE: AUGUST 31, 2017

EN :2007+A1:2011 Electromagnetic compatibility Emission standard for residential, commercial and light-industrial environments

ITUNER NETWORKS CORPORATION EMC REPORT Fremont Blvd. Fremont, CA

FCC PART TEST REPORT. Weccan Industrial Limited

TEST SUMMARY. Prüfbericht - Nr.: Test Report No.: Seite 2 von 25. Page 2 of 25

FCC Verification TEST REPORT

Page: 1 of 20 EMC TEST REPORT EN55024:1998+A2:2003

FCC 15B Test Report. : Bluetooth 4.2 Dual Mode USB HCI Module (Refer to item for more details) (Refer to item for more details)

Medtronic MiniMed TEST REPORT FOR. GST3 Glucose Sensor Transmitter, MMT-7763A. Tested To The Following Standards:

Chapter I - Federal Communications Commission Subchapter A - General Part 15 - Radio Frequency Devices Subpart C - Intentional Radiators

TEST REPORT. For RFID READER/WRITER. In conformity with. FCC CFR 47 Part15 Subpart C

Itron, Inc. TEST REPORT FOR. RF Telemetry Device Model: CCU100. Tested To The Following Standards:

To «Test_Standards» Test of: Radwin Ltd. Outdoor Subscriber Radio Unit. To: FCC CFR 47 Part 15B; ICES-003 Issue 6: 2016

BUREAU VERITAS. itaste svd

Overview of EMC Regulations and Testing. Prof. Tzong-Lin Wu Department of Electrical Engineering National Taiwan University

Measurement of RF Emissions from a Caterpillar Inc. MSS3s RF ID Key Fob

PARTIAL RADIATED EMISSIONS AND RADIATED IMMUNITY TEST REPORT

FCC Radio Test Report. Orpak Systems Ltd.

FCC CERTIFICATION TEST REPORT

EMC Test report for LED Panel Light Models , , , , ,

Test Report. Applicant ASUSTeK COMPUTER INC. 4F, No. 150, Li-Te Rd., Peitou, Taipei, Taiwan. Date of Receipt Jan. 28, Issued Date Mar.

EMC TEST REPORT for SHENZHEN MUST ENERGY TECHNOLOGY CO., LTD

EMC TEST REPORT CE TEST REPORT. for Audio / Video receivers and associated equipment

TEST REPORT FROM RFI GLOBAL SERVICES LTD

Measurement of RF Interference from a Canopy 900MHz Access Point and Subscriber Module Using A Yagi Antenna

EN v1.2.1 ( ) TEST REPORT FOR n 1X2 PCIe MINICARD TRANSCEIVER MODEL NUMBER: AR5B91 REPORT NUMBER: 08U

Table of Contents 1. GENERAL INFORMATION SYSTEM TEST CONFIGURATION CONDUCTED EMISSIONS TEST RADIATED EMISSION TEST...

TEST REPORT FROM RFI GLOBAL SERVICES LTD

EMI T E S T R E P O R T

CTS (NINGBO) TESTING SERVICE TECHNOLOGY INTERNATIONAL OPERATE ACCORDING TO ISO/IEC EMC TEST REPORT

EMC TEST REPORT RADISYS CORPORATION. Tel: Fax:

APPLICATION FOR CERTIFICATION On Behalf of Futaba Corporation Radio Control Model No.:T10CG-2.4G FCC ID:AZPT10CG-24G Brand : Futaba

FCC 47 CFR PART 15 SUBPART B

EMC TEST REPORT. for. Coliy Technology Co.,Ltd. Fluxgate Gaussmeter

EMC Test Report. Product. Model Number. Prepared for : Shenzhen top technology co., LTD Address : No 2.First Rd Of Lucky,Longgang Ping Shenzhen China

FCC 47 CFR PART 15 SUBPART C CERTIFICATION TEST REPORT FOR. Bluetooth Remote Control for Video Set Top Box MODEL NUMBER: IPRC1000 FCC ID: 2ABTE-L3YJC9

FCC CFR47 PART 15 SUBPART C INDUSTRY CANADA RSS-247 ISSUE 1 BLUETOOTH LOW ENERGY CERTIFICATION TEST REPORT FOR

Model: M /800 MHz Mobile Radio

Certificates and reports shall not be reproduced except in full, without the written permission of MET Laboratories, Inc.

CENTRE OF TESTING SERVICE INTERNATIONAL

8370 Court Avenue, Suite B-1 Ellicott City, Maryland (410) FCC CERTIFICATION

RF Emissions Test Report To Determine Compliance With: FCC, Part 15 Rules and Regulations

Kutta Radio & charger Report Number: TRE0026-5/17 Page 1 of 27 EMC TEST REPORT. TEST SAMPLE (EUT) : Kutta Medium frequency radio & Charger

RADIO TEST REPORT. For MODEL NO FCC ID: C3K1703 IC ID: 3048A Test Report No. R-TR190-FCCIC-UNII-1 Issue Date: 14 September 2015

TEST REPORT. draft draft draft. TRF Originator...: Shenzhen CTL Testing Technology Co., Ltd. Master TRF...: Dated

EMC TEST REPORT. Report No.: CE10-LIE040101E

Title: Test on 5.8 GHz Band Outdoor WiFi (802.11b/g) Wireless Base Station

ECHOSTAR 54.0 BRISBANE VOICE REMOTE 2017 MODEL: URC-2027BC0-R

By order of ZHONGSHAN LIANGYI LIGHTING CO., LTD. at Zhongshan, China

Transcription:

W66 N220 Commerce Court Cedarburg, WI 53012 Phone: 262.375.4400 Fax: 262.375.4248 www.lsr.com Compliance Testing of: RM186-SM Test Date(s): 4-12-16 4-25-16 Prepared For: Attention: N. Zach Hogya Laird Technologies W66 N220 Commerce Court Cedarburg, Wisconsin 53012 ENGINEERING TEST REPORT #: 316062A LSR JOB #: C-2411 This Test Report is issued under the Authority of: Shane Dock, EMC Engineer Signature: Date: 5/6/16 Quality Assurance by: Michael Hintzke, EMC Engineer Report By: Shane Dock, EMC Engineer Signature: Date: 5/6/16 Signature: Date: 5/6/16 This Test Report may not be reproduced, except in full, without written approval of LS Research, LLC.

TABLE OF CONTENTS LS Research, LLC In Review... 3 Product and General Information... 4 Introduction... 4 Product Description... 4 Test Requirements... 5 Summary of Test Report... 6 Definition of Performance Criteria... 7 Radiated Emissions Test... 8 AC Mains Conducted Emissions Test... 15 Electrostatic Discharge (ESD) Test... 20 Radiated RF Immunity Test... 23 Appendix A: Test Equipment List... 28 Appendix B: Applicable Referenced Standards.... 29 Appendix C: Uncertainty Statement... 30 LS Research, LLC Page 2 of 30

LS Research, LLC In Review As an EMC Testing Laboratory, our Accreditation and Assessments are recognized through the following: A2LA American Association for Laboratory Accreditation Accreditation based on ISO/IEC 17025: 2005 with Electrical (EMC) Scope of Accreditation A2LA Certificate Number: 1255.01 Federal Communications Commission (FCC) USA Listing of 3 Meter Semi-Anechoic Chamber based on Title 47 CFR Part 2.948 FCC Registration Number: 90756 Industry Canada On file, 3 Meter Semi-Anechoic Chamber based on RSS-GEN Issue 4 File Number: IC 3088-2 On file, 3 Meter Semi-Anechoic Chamber based on RSS-GEN Issue 4 File Number: IC 3088-3 LS Research, LLC Page 3 of 30

Product and General Information Manufacturer: Laird Technologies Date(s) of Test: 4/12/16 4/25/16 Tested By: Shane Dock, Zach Wilson Voltage: 3.6 1.8VDC Environmental Conditions in the Test Lab: Temperature: 20-25 C Atmospheric Pressure: 86 kpa - 106 kpa Humidity: 30-60% Introduction Between 4/12/16 and 4/25/16, a series of emissions and immunity tests were performed on the RM186-SM henceforth referred to as the Equipment Under Test or EUT. The emissions and immunity tests performed on the EUT are in accordance with ETSI EN 301 489-3 and ETSI EN 301 489-1 standards. The tests were performed to allow verification of the product to meet the EMC Directive 2014/30/EU. Product Description The RM186-SM module is designed to enable OEMs to add a long range LoRa radio link as well as central role Bluetooth Low Energy (BLE) to small, portable, power-conscious devices. The RM186-SM module is enabled with Laird s smart BASIC, an event-driven programming language that enables OEMs to make their product development quicker and simpler, significantly reducing time to market. smartbasic enables customers to develop a complete embedded application inside the compact RM191 hardware, connecting to a wide array of external sensors via its I2C, SPI, UART, ADC or GPIO interfaces. The module is based on the world-leading Nordic Semiconductor nrf51822 (BLE) and Semtech Sx1272 (LoRa) chipsets, the RM186-SM module provides ultra-low power consumption with outstanding wireless range using the LoRa radio link and local BLE connections. LS Research, LLC Page 4 of 30

Test Requirements Product/Product Family EN 301 489-17/EN 301 489-1 Application EMC for Radio Equipment, Specific conditions for Broadband DTS/EMC standard for radio equipment and services Basic Standards (Emissions) Port Tested Test Descriptions Results EN 55022 Enclosure Radiated Emissions Pass AC Power Ports Conducted Emissions Pass IEC 61000-3-2 EN 61000-3-2 AC Power Ports Harmonic Current Emissions (through Amendment 14 of IEC 61000-4-7) Note 2 IEC 61000-3-3 EN 61000-3-3 AC Power Ports Voltage Fluctuation and Flicker Note 2 Basic Standards IEC EN Port Tested Test Specifications Results IEC 61000-4-2 EN 61000-4-2 Enclosure ESD, 4kV Contact, 8kV Air Discharge Pass IEC 61000-4-3 EN 61000-4-3 Enclosure Radiated RF Immunity, 3 V/m, 80% AM @ 1kHz, 1% step of the previous frequency 80-1000, 1400-2700 MHz Pass IEC 61000-4-4 EN 61000-4-4 AC Power Ports EFT Burst: Note 0.5 kv - 2kV 3,4 Signal, Control & IEC 61000-4-4 EN 61000-4-4 Telecomm and DC power EFT Burst 0.5kV Note 3 ports IEC 61000-4-5 EN 61000-4-5 AC Input Ports Surge 1kV Differential Mode (line to line) 2kV Common Mode (line to ground) Note 3 IEC 61000-4-5 EN 61000-4-5 Telecommunication Ports Surge ± 0.5kV Common Mode Note 3 IEC 61000-4-6 EN 61000-4-6 Conducted RF Immunity: 3 V 80% AM modulation @ 1kHz Note 4 IEC 61000-4-6 EN 61000-4-6 IEC 61000-4-8 EN 61000-4-8 AC Power Ports Signal, Control, Telecommunication and DC Power Ports (>3m) Display & Magnetic Sensors IEC 61000-4-11 EN 61000-4-11 AC input ports Conducted RF Immunity: 3 V, 80% AM modulation @ 1kHz Notes: 1. See Appendix B for current test standard publication dates. EUT was tested per requirements of current standard publications. 2. Not applicable based on table 1 of ETSI EN 301 489-1 v1.9.2. The EUT is a portable equipment. 3. Not applicable based on table 2 of ETSI EN 301 489-1 v1.9.2 The EUT is a portable equipment. 4. Not Applicable. EUT is not AC powered. Note 4 Magnetic Field Immunity 3 A/m Note 3 Voltage Dips and Short Interruptions at test Voltage level: 70%, 40% and 5% nominal for 10ms, 100ms and 5 sec (50Hz) Note 3,4 LS Research, LLC Page 5 of 30

Summary of Test Report DECLARATION OF COMPLIANCE The EUT was found to MEET the requirements as described within the specifications of ETSI EN 301 489-3 V1.6.1 and ETSI EN 301 489-1 V1.9.2 for emissions and immunity tests. In addition, based on the review of the data and measurement method, the EUT was found to MEET the requirements as described within the specifications ETSI EN 301 489-1 V1.9.2. The enclosed test results pertain to the sample(s) of the test item listed, and only for the tests performed on the data sheets. Any subsequent modification or changes to the test items could invalidate the data contained herein, and could therefore invalidate the findings of this report. If some emissions are seen to be within 3dB of their respective limits, these levels are within the tolerances of the test equipment and site employed. There is a possibility that this unit, or a similar unit selected out of production may not meet the required limit specification if tested by another agency. LS Research, LLC Page 6 of 30

Definition of Performance Criteria EUT setup during immunity and susceptibility testing: The EUT was connected via BLE to a BLE sensor and via LoRa to a LoRa gateway, respectively. The percentage of successful packets sent to each versus the total number sent was monitored to determine the criterion met, along with what mode the unit was in. Performance Criterion A: The unit must stay in the test mode and must maintain a packet integrity rate of 98% or better. Performance Criterion B: The unit must stay in the UART mode and must be able to start up in test mode successfully. Performance Criterion C: The unit does not stay in the UART mode and must be reset to that mode. LS Research, LLC Page 7 of 30

Radiated Emissions Test Test Setup The test setup was assembled in accordance with EN 55022 per requirements of ETSI EN 301 489-3. Test Procedure The frequency range from 30 MHz to 6000 MHz was investigated, and levels were manually noted at the various fixed degree settings of azimuth on the turntable and antenna height. The EUT was placed in a 3 Meter Semi-Anechoic Chamber, with the antenna mast placed such that the antenna was 3 meters from the test object. A Biconical Antenna was used to measure emissions from 30 MHz to 200 MHz, a Log Periodic Antenna was used to measure emissions from 200 MHz to 1000 MHz. Emissions between 1000 to 6000 MHz were measured using a double ridged horn antenna. The procedure for finding the maximum radiated RF emission for each scan was as follows: locating the antenna elevation (between 1 and 4 meters) that produced a peak emission. Next, the EUT was rotated on the turntable, obtaining the azimuth of the maximum signal level. These two steps were repeated until the maximum emission stayed at the same level. Individual scans were run using horizontal and vertical antenna polarities in the different frequency ranges. - Test Equipment Utilized A complete list of test equipment can be found in Appendix A. Correction factors and cable loss factors were entered into the EMI Receiver database. As a result, the data taken from the EMI Receiver accounts for the antenna correction factor as well as cable loss, and can therefore be entered into the database as a corrected meter reading. The EMI Receiver was operated with a resolution bandwidth of 120 khz for measurements below 1 GHz and 1MHz for measurements above 1GHz. Test Results The EUT was found to MEET the Radiated Emissions requirements of the ETSI EN 301 489-3 standard. The frequencies with significant signals were recorded and plotted as shown in the Data Charts and Graphs. LS Research, LLC Page 8 of 30

Calculation of Radiated Emissions Limits The following table presents the EN 55022 limits. Frequency (MHz) 10 m Limit db V/m 3 m Limit db V/m 30-230 30 40.5 230-1000 37 47.5 Frequency (MHz) Peak Limit (3m) db V/m Average Limit (3m) db V/m 1000 3000 70 50 3000-6000 74 54 Sample conversion of limits between 10 meters and 3 meters: 3m limit (db V/m = 10m limit (db V/m) +) 20 log (10m/3m) db or 3m limit (db V/m) = 10m limit (db V/m) + 10.5 db from 30-230 MHz for example: 3m limit (db V/m) = 30.0 db V/m + 10.5 db 40.0 db V/m = 30.0 db V/m + 10.5 db LS Research, LLC Page 9 of 30

Radiated Emissions Data Chart Manufacturer: Laird Technologies Date(s) of Test: 4/12/16 4/14/16 Tested By: Shane Dock Voltage: 3.6 VDC (Worst-Case) Distance: X 3 Meters 10 Meters EUT Power: Single phase 120 VAC, 20 VDC Lithium Ion Battery X 60Hz EUT Placement: X 80cm non-conductive table 10cm Spacers EUT Test 3 Meter Semi-Anechoic FCC 3/10m OATS X Location: Listed Chamber Measurements: Pre-Compliance Preliminary X Final Detectors Used: X Peak X Quasi-Peak X Average Environmental Conditions in the Lab: Temperature: 20 25 C Relative Humidity: 30 60 % LS Research, LLC Page 10 of 30

The table depicts the level of significant radiated emissions found: Below 1000 MHz Frequency Height Azimuth Quasi-Peak QP Limit (MHz) (cm) (Degrees) (dbm) (dbm) Polarization Orientation Note 193.88 100 0 18.9 40.5 H V Noise Floor 196.88 100 0 19.9 40.5 V V Noise Floor 922.058 100 0 27.0 47.5 H V Noise Floor 987.28 100 0 26.9 47.5 V V Noise Floor Above 1000 MHz Frequency Height Azimuth Peak Average Plimit Alimit (MHz) (cm) (Degrees) (dbm) (dbm) (dbm) (dbm) Polarization Orientation Note 5920.81 150 0 40.17 28.05 74 54 H V Noise Floor 5941.66 150 0 40.43 28.02 74 54 V V Noise Floor LS Research, LLC Page 11 of 30

Photo(s) Taken During Testing Setup for Radiated Emissions Test LS Research, LLC Page 12 of 30

Graphs Note: All screen captures below feature worst-case emissions. Only noise floor measurements were taken, and emissions seen in the plots are not a function of the EUT. Signature Scan of Peak Radiated Emissions 30-200 MHz, at 3m Horizontal Vertical Emissions above are not a function of the EUT. 200-1000 MHz, at 3m Horizontal Vertical LS Research, LLC Page 13 of 30

1000 MHz 6000 MHz at 3m Horizontal Vertical LS Research, LLC Page 14 of 30

AC Mains Conducted Emissions Test Test Setup The test setup was assembled in accordance with EN 55022 per requirements of ETSI EN 301 489-3. A generic DC supply was used to power the EUT during testing (The charger is provided separately from the unit.) Test Procedure The EUT was investigated in continuous modulated transmit mode for this portion of the testing. The appropriate frequency range and bandwidths were selected on the EMI Receiver, and measurements were made. The bandwidth used for these measurements is 9 khz, as specified in CISPR 16-1, Section 1, Table 1, for Quasi-Peak and Average detectors in the frequency range of 150 khz to 30 MHz. Final readings were then taken and recorded. Test Equipment Utilized A list of the test equipment and accessories utilized for the Conducted Emissions test is provided in Appendix A. This list includes calibration information and equipment descriptions. All equipment is calibrated and used according to the operation manuals supplied by the manufacturers. Calibrations of the LISN and Limiter were performed at an IEC/ISO 17025 accredited calibration laboratory, traceable to the SI standard. All cables are calibrated and checked periodically for conformance. The emissions are measured on the EMI System, which has automatic correction for all factors stored in memory and allows direct readings to be taken. Test Results The EUT was found to MEET the Radiated Emissions requirements of the ETSI EN 301 489-3 standard. The frequencies with significant signals were recorded and plotted as shown in the Data Charts and Graphs. LS Research, LLC Page 15 of 30

CISPR 22 Class B Limits of Conducted Emissions at the AC Mains Ports Frequency Range Class B Limits (dbµv) Measuring (MHz) Quasi-Peak Average Bandwidth 0.150-0.50 * 66-56 56-46 RBW = 9 khz 0.5 5.0 56 46 VBW 9 khz for QP 5.0 30 60 50 VBW = 1 Hz for Notes: Lower limit applies at transition frequencies. Average * The limit decreases linearly with the logarithm of the frequency in this range. LS Research, LLC Page 16 of 30

CONDUCTED EMISSIONS TEST DATA CHART Manufacturer: Laird Technologies Date(s) of Test: 4/20/16 4/21/16 Project Engineer: Shane Dock Test Engineer: Shane Dock Voltage: 3.6 VDC Operation Mode: Sleep/Standby Mode Environmental Conditions in the Temperature: 20-25 C Relative Humidity: 40% Lab: Test Location: X AC Mains Test area Chamber X 40cm from Vertical Ground Plane 10cm Spacers EUT Placed On: X 80cm above Ground Plane Other: Measurements: Pre-Compliance Preliminary X Final Detectors Used: Peak X Quasi-Peak X Average Line Frequency (MHz) Q-Peak Reading (dbμv) Q-Peak Limit (dbμv) Quasi- Peak Margin (db) Average Reading (dbμv) Average Limit (dbμv) Average Margin (db) 1 0.181 41.0 64.4 23.4 31.4 54.4 23.0 1 0.226 39.4 62.6 23.2 29.5 52.6 23.1 1 0.604 29.6 56.0 26.4 20.9 46.0 25.1 2 0.150 36.7 66.0 29.3 25.5 56.0 30.5 2 0.235 27.9 62.3 34.4 20.0 52.3 32.3 2 0.622 31.6 56.0 24.4 23.2 46.0 22.8 LS Research, LLC Page 17 of 30

Test Setup Photo(s) Conducted Emissions Test LS Research, LLC Page 18 of 30

Screen Captures Conducted Emissions Test The screen captures represent Peak Emissions. For conducted emission measurements, both a Quasi-Peak detector function and an Average detector function are utilized. Line 1 Line 2 LS Research, LLC Page 19 of 30

Electrostatic Discharge (ESD) Test Test Setup The ESD Immunity Test was setup in accordance with the IEC 61000-4-2 standard per the requirements of ETSI EN 301 489-3. The tests were performed with a Horizontal Coupling Plane (HCP) 1.6m x 0.8m placed on top of a table which is 80 cm height above the ground plane. A 0.5 mm thin sheet of non-conductive material was placed underneath the EUT for isolation of the Coupling Plane. The tests were performed using a Schaffner NSG438 Gun, which is compliant with the ESD standard IEC 61000-4-2, and was grounded to the ground reference plane through a 2-meter cable. The unit was placed in a production running mode. The unit was set up to communicate with both a BLE sensor and a LoRa gateway, alternating between transmitting data to these units and receiving data from them. Test Procedure The EUT was tested up to a level of 4 kv Contact discharge, and a level of 8 kv Air discharge, as required by ETSI EN 301 489-3 standard. Both Vertical Coupling Plane (VCP) and Horizontal Coupling Plane (HCP) discharges were applied at locations around the EUT. Contact Discharge The electrostatic discharges were applied to points of the EUT that were accessible during normal usage. The pointed tip of the ESD Gun was applied in contact mode to several key points on the EUT, and other accessible areas. The test voltage was increased from a minimum level of 2kV to a level of 4kV, in steps, to determine the EUT s susceptibility threshold. At least 10 single discharges were applied to selected points in both positive and negative polarities. The ESD Gun was held perpendicular to the surface of the discharge point to ensure repeatability of test results. During the ESD testing, the EUT continued to operate while it was monitored for any distortion caused by the ESD. The test sample showed no apparent failure during contact discharge from 2 kv to 4 kv. ESD contact discharges were also applied to a VCP spaced 10cm from the EUT, and a HCP around the EUT. Air Discharge The charged round tip of the ESD Gun was moved around various key points on the EUT to determine locations that would take a discharge. The tip was moved away from the EUT s surface after each discharge. This process was repeated until the discharges were completed from levels of 2 kv to 8 kv. Test Equipment Utilized A complete list of test equipment can be found in Appendix A. Test Results The EUT was found to MEET the ESD requirements of ETSI EN 301 489-3. Details of the test results can be found in the Data Charts. LS Research, LLC Page 20 of 30

Electrostatic Discharge Data Chart Manufacturer: Laird Technologies Date(s) of Test: 4/22/16 4/25/16 Test Engineer: Zach Wilson Voltage: 3.6 VDC A (X) indicates that a test specification has been observed. Discharge Voltage Level: X Level 1 X Level 2 X Level 3 Other Discharge Impedance: 150 Ohm/150pF X 330 Ohm/150pF Discharge Factor: X > 1 Second Other 5 Seconds Number of Discharges: X > 10 Hits at all locations Other Hits at all locations Performance Criterion: X Criterion A X Criterion B Criterion C Test Results: X Passed Failed Test Voltages Discharge Type Test Points +2kV -2kV +4kV -4kV +8kV -8kV Air Contact Criteria Horizontal CP Pass Pass Pass Pass N/A N/A X B Vertical CP Pass Pass Pass Pass N/A N/A X B Base of Anten. Pass Pass Pass Pass Pass Pass X A LS Research, LLC Page 21 of 30

Photo(s) Taken During Testing Setup for Electrostatic Discharge (ESD) Test LS Research, LLC Page 22 of 30

Radiated RF Immunity Test Test Setup The EUT was operated within the 3 Meter Semi-Anechoic Chamber during the Radiated RF Immunity Test. The test was performed in accordance with IEC 61000-4-3 per the requirements of ETSI EN 301 489-3. A level calibrated field wa-s first determined by using the following steps. A transmitting antenna was placed 3 meters from the EUT s position, and 1.5 meters above the Ground Reference Plane. A swept carrier signal was then transmitted via the antenna, over a frequency range of 80 MHz to 1000 MHz. The resulting field intensity was monitored over a planar area of 1.5 m x 1.5 m, starting at 0.8 meters above the floor at the EUT position. A grid of 16 points within the 1.5 m x 1.5 m aperture area was measured using a Field Intensity Probe. The forward power required to create this field strength at each frequency step was measured with a Directional Coupler and stored in a calibration file. The Power Meter, Signal Generator, and Field Intensity Probe were all controlled and monitored by a laptop computer running the Teseq software program. This calibration was performed in both horizontal and vertical polarities. A similar setup and process was used to calibrate a 9 pt. grid for 1400 MHz to 2700 MHz. A Double Ridge Wave Guide Horn Antenna was used for the transmitting antenna. The calibration was conducted in both the vertical and horizontal antenna polarities. The unit was placed in a production running mode. The unit was set up to communicate with both a BLE sensor and a LoRa gateway, alternating between transmitting data to these units and receiving data from them. Test Procedure The EUT was placed on an 80 cm non-conductive table and the Antenna Mast was placed such that the antenna was 3 meters from the EUT. A Bilog Antenna was used from 80 MHz to 1000 MHz and a Double-Ridged Waveguide Horn Antenna was used from 1400MHz to 2700 MHz to transmit RF power to the EUT for both vertical and horizontal polarities. The transmitting antenna was placed at a fixed position above the Ground Reference Plane at a height of 1.5 meters for the lower frequency range and 1.3m for the higher frequency range. The field strength was monitored using a Field Sensor Probe placed on the test table near the EUT as a verification check. The intensity of the established field strength was checked prior to the actual testing. The Teseq software program was set to dwell for 3 seconds at each frequency and control the Signal Generator modulation of 1 khz, 80% A.M. Each frequency was increased by 1% of the previous frequency in a logarithmic fashion. The test was repeated for each test configuration (front and back of the EUT) using both vertical and horizontal antenna polarities. The parameters of the RF Immunity testing for each sweep were set and automatically controlled and recorded in the Data Charts. Test Equipment Utilized A complete list of test equipment can be found in Appendix A. Test Results The EUT was found to MEET the radiated immunity requirements of ETSI EN 301 489-3 The EUT was monitored remotely from the Control Room during the test. LS Research, LLC Page 23 of 30

RF IMMUNITY TEST CONFIGURATION LS Research, LLC Page 24 of 30

Radiated RF Immunity Test Data Chart Manufacturer: Laird Technologies Date(s) of Test: 4/18/16 4/20/16 Test Engineer: Shane Dock Voltage: 3.6 VDC Areas Tested: X Front X Back Left Right Environmental Conditions in the Lab: Temperature: 20 25 C Atmospheric Pressure: 86 kpa 106 kpa Relative Humidity: 30 60% A (X) indicates that a test specification has been observed. Frequency Range: X 1.4 2.7 GHz X 80 MHz 1000 MHz 26 MHz 80 MHz 900 ± 5 MHz 27 MHz 500 MHz Other Field Strength: Level 1: 1 V/m Level 3: 10 V/m X Level 2: 3 V/m Other Antenna Range-EUT X 2 m X 3 m Antenna Polarity: X Horizontal X Vertical Modulation: X AM: 80%, 1 khz Pulse 50% duty, 200 Hz AM: 80%, 0.5 Hz Other: Pulse 50% duty, 1 Hz Step: X 1% of Fund. Freq., 3 Sec. Dwell 1 MHz Step, Sec. Dwell Performance Criteri X Criterion A Criterion B Criterion C Test Results: X Passed Failed LS Research, LLC Page 25 of 30

Test Config 80-1000MHz 1% Frequency Increments 3 V/m Stress Level 3 sec. Dwell Time 80%-1kHz Sine Modulation EUT Side Polarity Results Criteria Front Horizontal Pass A Front Vertical Pass A Back Horizontal Pass A Back Vertical Pass A Test Config 1400-2700MHz 1% Frequency Increments 3 V/m Stress Level 3 sec. Dwell Time 80%-1kHz Sine Modulation Side Polarity Results Criteria Front Horizontal Pass A Front Vertical Pass A Back Vertical Pass A Back Horizontal Pass A LS Research, LLC Page 26 of 30

Photo(s) Taken During Testing LS Research, LLC Page 27 of 30

Appendix A: Test Equipment List LS Research, LLC Page 28 of 30

/ Appendix B: Applicable Referenced Standards. Standard Date Am. 1 Am. 2 EN 55022 2006 2007 ETSI EN 301 489-3 V1.6.1 2013 ETSI EN 301 489-1 V1.9.2 2011 EN 61000-4-2 2009 IEC 61000-4-2 2008 EN 61000-4-3 2006 2008 2010 IEC 61000-4-3 2006 2007 2010 LS Research, LLC Page 29 of 30

Appendix C: Uncertainty Statement Table of Expanded Uncertainty Values, (K=2) for Specified Measurements Measurement Type Particular Configuration Uncertainty Values Radiated Emissions 3 Meter chamber, Biconical Antenna 4.82 db Radiated Emissions 3-Meter Chamber, Log Periodic 4.88 db Antenna Radiated Emissions 3-Meter Chamber, Horn Antenna 4.85 db Radiated Emissions 10-Meter OATS, Biconical Antenna 4.32 db Radiated Emissions 10-Meter OATS, Log Periodic Antenna 3.63 db Absolute Conducted Emissions Agilent PSA/ESA Series 1.38 db AC Line Conducted Emissions Shielded Room/EMCO LISN 3.20 db Radiated Immunity 3 Volts/Meter in 3-Meter Chamber 2.05 Volts/Meter Conducted Immunity 3 Volts level 2.33 V EFT Burst, Surge, VDI 230 VAC 54.4 V ESD Immunity Discharge at 15kV 3200 V Temperature/Humidity Thermo-hygrometer 0.64 / 2.88 %RH LS Research, LLC Page 30 of 30