EE 230 Experiment 10 ECG Measurements Spring 2010

Similar documents
Lab E5: Filters and Complex Impedance

Lab: Using filters to build an electrocardiograph (ECG or EKG)

EXPERIMENT 5 Bioelectric Measurements

UNIVERSITY OF UTAH ELECTRICAL ENGINEERING DEPARTMENT

Massachusetts Institute of Technology MIT

BIO-ELECTRIC MEASUREMENTS

Florida Atlantic University Biomedical Signal Processing Lab Experiment 2 Signal Transduction: Building an analog Electrocardiogram (ECG)

Lab 6: Instrumentation Amplifier

The above figure represents a two stage circuit. Recall, the transfer function relates. Vout

UNIVERSITY OF UTAH ELECTRICAL AND COMPUTER ENGINEERING DEPARTMENT ELECTROMYOGRAM (EMG) DETECTOR WITH AUDIOVISUAL OUTPUT

EECE Circuits and Signals: Biomedical Applications. Lab ECG I The Instrumentation Amplifier

DEPARTMENT OF ELECTRICAL ENGINEERING AND COMPUTER SCIENCE MASSACHUSETTS INSTITUTE OF TECHNOLOGY CAMBRIDGE, MASSACHUSETTS 02139

EE4902 C Lab 7

EXPERIMENT 8 Bio-Electric Measurements

Lab 4: Analysis of the Stereo Amplifier

Bio-Potential Amplifiers

Electronics I. laboratory measurement guide Andras Meszaros, Mark Horvath

Development of Electrocardiograph Monitoring System

EL4089 and EL4390 DC Restored Video Amplifier

ECE ECE285. Electric Circuit Analysis I. Spring Nathalia Peixoto. Rev.2.0: Rev Electric Circuits I

6.101 Introductory Analog Electronics Laboratory

ECE3204 D2015 Lab 1. See suggested breadboard configuration on following page!

Experiment 1: Amplifier Characterization Spring 2019

Single Supply, Rail to Rail Low Power FET-Input Op Amp AD820

AD8232 EVALUATION BOARD DOCUMENTATION

DEPARTMENT OF ELECTRICAL ENGINEERING AND COMPUTER SCIENCE MASSACHUSETTS INSTITUTE OF TECHNOLOGY CAMBRIDGE, MASSACHUSETTS 02139

Kanchan S. Shrikhande. Department of Instrumentation Engineering, Vivekanand Education Society s Institute of.

Real Analog - Circuits 1 Chapter 11: Lab Projects

University of Jordan School of Engineering Electrical Engineering Department. EE 204 Electrical Engineering Lab

Electronic Instrumentation ENGR-4300 Fall 2002 Project 2: Optical Communications Link

Analog Circuits and Systems

EXPERIMENT 7 The Amplifier

BME 405 BIOMEDICAL ENGINEERING SENIOR DESIGN 1 Fall 2005 BME Design Mini-Project Project Title

ECE4902 C Lab 7

Equipment: You will use the bench power supply, function generator and oscilloscope.

Group: Names: Resistor Band Colors Measured Value ( ) R 1 : 1k R 2 : 1k R 3 : 2k R 4 : 1M R 5 : 1M

Experiment 1: Instrument Familiarization (8/28/06)

Homework Assignment 01

Ques on (2): [18 Marks] a) Draw the atrial synchronous Pacemaker block diagram and explain its operation. Benha University June 2013

Lab 2: Discrete BJT Op-Amps (Part I)

A Design Of Simple And Low Cost Heart Rate Monitor

Figure 1(a) shows a complicated circuit with five batteries and ten resistors all in a box. The

Experiment 8 Frequency Response

Biomedical Instrumentation (BME420 ) Chapter 6: Biopotential Amplifiers John G. Webster 4 th Edition

EE 233 Circuit Theory Lab 4: Second-Order Filters

EE100B Experiment 6. The Design of Waveform Generators. College of Engineering University of California, Riverside. Objective

*Notebook is excluded

THE AMPLIFIER. A-B = C subtractor. INPUTS Figure 1

EE 3101 ELECTRONICS I LABORATORY EXPERIMENT 9 LAB MANUAL APPLICATIONS OF IC BUILDING BLOCKS

Lecture 4 Biopotential Amplifiers

Homework Assignment 01

Single Supply, Rail to Rail Low Power FET-Input Op Amp AD820

Special-Purpose Operational Amplifier Circuits

UNIVERSITY OF PENNSYLVANIA EE 206

This is an outline of the subjects that will be touched upon during this presentation.

Laboratory 2. Lab 2. Instrument Familiarization and Basic Electrical Relations. Required Components: 2 1k resistors 2 1M resistors 1 2k resistor

EE 3305 Lab I Revised July 18, 2003

NJM4151 V-F / F-V CONVERTOR

Circuit 4 Schmitt Trigger

instead we hook it up to a potential difference of 60 V? instead we hook it up to a potential difference of 240 V?

Homework Assignment True or false. For both the inverting and noninverting op-amp configurations, V OS results in

EE 368 Electronics Lab. Experiment 10 Operational Amplifier Applications (2)

ELEC 351L Electronics II Laboratory Spring 2014

* Notebook is excluded. Features KL-720 contains nine modules, including Electrocardiogram Measurement, E lectromyogram Measurement,

Experiment 6: Biasing Circuitry

Electrocardiogram (ECG)

Experiment 1: Instrument Familiarization

Laboratory Activities Handbook

Name Kyla Jackson, Todd Germeroth, Jake Spooler Date May 5, 2010 Lab 3E Group 3 Experiment Title Project Deliverable 3

EMG Electrodes. Fig. 1. System for measuring an electromyogram.

Chapter 9: Operational Amplifiers

Integrators, differentiators, and simple filters

Used to overcome ventricular fibrillation may be due to coronary occlusion, shock, or abnormalities in blood chemistry

STM32 microcontroller core ECG acquisition Conditioning System. LIU Jia-ming, LI Zhi

Homework Assignment 03

LABORATORY 7 v2 BOOST CONVERTER

Step Response of RC Circuits

Physiological Signal Processing Primer

DiMarzio Section Only: Prelab: 3 items in yellow. Reflection: Summary of what you learned, and answers to two questions in green.

NON-AMPLIFIED HIGH SPEED PHOTODETECTOR USER S GUIDE

BIOMEDICAL INSTRUMENTATION PROBLEM SHEET 1

Concepts to be Reviewed

ME 365 EXPERIMENT 7 SIGNAL CONDITIONING AND LOADING

Lab 1: Basic Lab Equipment and Measurements

CHAPTER 3. Instrumentation Amplifier (IA) Background. 3.1 Introduction. 3.2 Instrumentation Amplifier Architecture and Configurations

Experiment 6: Biasing Circuitry

Lab 2: Linear and Nonlinear Circuit Elements and Networks

EE 233 Circuit Theory Lab 3: First-Order Filters

Sirindhorn International Institute of Technology Thammasat University at Rangsit

Amplified High Speed Photodetectors

Week 8 AM Modulation and the AM Receiver

Lab 2: Common Base Common Collector Design Exercise

Introduction to basic laboratory instruments

Laboratory Project 1: Design of a Myogram Circuit

Common-source Amplifiers

NON-AMPLIFIED PHOTODETECTOR USER S GUIDE

I1 19u 5V R11 1MEG IDC Q7 Q2N3904 Q2N3904. Figure 3.1 A scaled down 741 op amp used in this lab

Transient Response of Low-Power ECG Recoding Amplifiers for Use with Un-gelled Electrodes

Lecture #4 Special-purpose Op-amp Circuits

LF442 Dual Low Power JFET Input Operational Amplifier

Transcription:

EE 230 Experiment 10 ECG Measurements Spring 2010 Note: If for any reason the students are uncomfortable with doing this experiment, please talk to the instructor for the course and an alternative experiment will be assigned. Although we believe this experiment to be safe, electrodes will be placed on the body to take measurements and if large electrical signals are injected into these electrodes, the subject could be seriously injured or killed. We will exercise a reasonable level of precaution in this experiment but can not guarantee that the experiment is completely safe. CAUTION: MUST READ the FOLLOWING TWO PARAGRAPHS BEFORE PROCEEDING When conducting this experiment, the following procedures MUST be followed. First, the only piece of electrical equipment that should be attached to the circuit board when electrodes are attached to the body is the oscilloscope. Make sure that the ground terminal of the oscilloscope is connected to the ground in the circuit. Second, when the electrodes are attached to the body, the biasing of all circuitry should be with two 9V dc batteries. Third, students should only connect electrodes to a student when another individual is present and a TA or the course instructor is in the room. Finally, if for any reason any electrical shocks are experienced, immediately remove the electrodes from the subject. In preparation of the circuits, a Body Emulator will be built. The body emulator is a simple circuit that will provide signals of about the same amplitude as the ECG signals. The Body Emulator will be used during the development stages of this experiment. In this experiment, only two dc power supplies should be used, one should be set at +9V and the other at -9V. When working with the Body Emulator, normal laboratory procedures should be followed. Before the Body Emulator is replaced with the actual electrodes attached to the subject, disconnect the +9V and -9V dc power supplies and disconnect all test equipment except for the oscilloscope. Replace the +9V and -9V dc power supplies with 9V batteries. Verify that the circuit is still working using only the oscilloscope to make measurements with the 9V batteries. After the circuit performance is verified with the Body Emulator and the two 9V batteries, connect the electrodes on the subject in place of the two voltages on the Body Emulator. Any trouble shooting that is undertaken when the electrodes are placed on the subject should use only the oscilloscope for measurements. An optical isolator is often used when making measurements on the body and provides additional isolation between the test subject and the external instrumentation. To keep the scope of this measurement at a reasonable level, an optical isolator will not be used here. This experiment will focus on measuring electrocardiogram (ECG) signals that are present on the surface of the skin. Some refer to these as ECG signals and others use the term EKG signals to donate the same signals. The ECG signals are generated by the heart and are used by cardiologists and physicians as a part of an evaluation process to determine if the heart is operating properly. ECG signals are also monitored in Page 1 of 6

pacemakers and defribulators to assess how the heart is performing. Abnormalities or changes in the ECG trigger action by the pacemaker or defribulator. If two electrodes are placed at suitable locations on the surface of the skin, a waveform similar to that shown in Fig. 1 will be obtained. Fig. 1 Artists Depiction of Electrocardiogram (from www.mathworks.com/.../fx_files/10858/1/ecg.jpg) There will be one major peak every time the heart beats. The magnitude of this peak will depend upon where the electrodes are placed and will vary somewhat from patient to patient. The peak amplitude is in the 1mV range (400uV to 2mV are typical). As the heart muscle contracts, the shape of the ECG signal will change and there is considerable information not only in when the peaks occur and in the shape of the waveform in the vicinity of the peaks but also in the smaller peaks that appear between the major peaks. Unfortunately there is a lot of noise present on the surface of the skin and also in the amplifiers that are used to amplify these small signals so the real signal that will be obtained is not nearly as ideal as that shown in Fig. 1. It will be corrupted by a rather high level of noise that may also be in the mv range or sometimes even larger. Medical equipment that is used to obtain ECG signals includes signal processing algorithms that remove as much of the noise as possible without excessively distorting the signal information that is present in the ECG signals. In this experiment, emphasis will be placed on obtaining the ECG signal and on removing enough of the noise to make the major peaks separate themselves from the noise that occurs between. Most of the signal information in the ECG signal is at low frequencies, likely between 0.5Hz and 50Hz. Consequently a highpass filter is often used reject dc components that may be present in either the incoming signals or in the amplifiers themselves. Likewise a lowpass filter is often used to reject higher frequency components. Finally, large 60Hz signals are often present. If the lowpass filter has steep Page 2 of 6

band edge somewhere between the signal band and 60 Hz, the 60Hz interfering signal can be reduced. Sometimes a narrow-band notch filter is also included to specifically reject the 60Hz signal. Since the pole frequencies of these filters is quite low, the required capacitor values tend to get rather large making the use of one or more electrolytic capacitors is particularly attractive for building these filters. Electrolytic capacitors can be used but the normal operation of the circuit must not attempt to reverse bias the electrolytic capacitors. This issue will be discussed later in this experiment. Body Emulator This circuit will be used to generate signals that are comparable in amplitude to the ECG signals that will be obtained at the surface of the skin. An electrical model of the signals that will be picked up at the surface of the skin is shown in Fig. 2. The Fig. 2 ECG Signals on the skin surface ECG signal of interest is V ECG = V SIG1 -V SIG2. Note that the desired signal is the difference of two single-ended signals. Thus, three electrodes will ultimately be required. The placement of these electrodes on the body will be discussed later. There is a thevenin equivalent impedance in series with the two signals so the signals that will actually appear at the input to an amplifier are the signals on nodes A and B. The actual voltage that will be measured is the difference voltage V AB. Of course, the instrumentation should have a high input impedance so that there is little voltage drop across R TH1 and R TH2. 10K 10K 10K R TH2 V A V FG 10 V AB R TH1 V B Fig. 3 ECG Body Emulator Page 3 of 6

An ECG Body Emulator is shown in Fig. 3. V FG represents the output of the function generator. If the Function Generator is programmed to output a 500uV pulse of duration 10mS once every second, this will be a crude approximation of the ECG signal. The 10K to 10Ω attenuator will reduce the amplitude of this pulse to about 500uV. The output signals from the two operational amplifiers should be complimentary 500uV pulses. The resistors R TH1 and R TH2 represent the thevenin impedances of Fig. 2. Thevenin Impedances The thevenin impedance is difficult to obtain since it is comprised of the impedance of the skin contact in series with the internal impedance to the source of the voltage. Although there is not a point voltage source internal to the body, the model of Fig. 2 should suffice for this experiment. If we assume that the internal impedance is small compared to the skin contact impedance, this impedance can be rather easily measured. You might first want to measure the impedance between two points on the body by placing two test leads on the skin and then measuring the impedance with the multimeter. Then, place two electrodes with conductive grease provided by the TA on the surface of the skin and repeat the measurement with the multimeter. MAKE sure you are measuring the resistance with the multimeter. The two electrodes can be placed on ONE arm in the vicinity of the wrist about 4 inches apart. Assuming the two impedances are about the same, the thevenin impedances, R TH1 and R TH2 should be about half of the measured impedance. How much difference do you see in the impedance by placing two test leads on the skin and by placing two electrodes on the skin with conducting grease? ECG Amplifier and Filters Many different circuits can be used to build the difference amplifier and the filters. Fig. 4 shows a basic instrumentation amplifier that is widely used for measuring ECG signals. Note the input impedance is ideally infinite for both inputs. The output of this amplifier is single-ended making it possible to follow this with single-ended filters. V A R 3A R 4A R 1 R 2A V OUT R 2B R 3B R 4B V B Fig. 4 Basic Instrumentation Amplifier Ideally the A and B values of the corresponding resistors are identical. This shows everything being direct coupled and is adversely affected by offset voltages of the amplifiers if the gains of the amplifiers are large. Methods for reducing the effects of Page 4 of 6

offset voltage were discussed earlier in the course. Remember when designing the differential amplifier and any filters to use +/- 9V power supplies. Using Electrolytic Capacitors to Build Filters When it is necessary to have filters with low frequency poles, the size of the required capacitors often becomes quite large making it particularly attractive to use electrolytic capacitors. But, if the first-order lowpass filter shown in Fig. 5a were used with an electrolytic capacitor, the capacitor would become reverse biased when the output voltage goes negative for an extended time and this would be unacceptable. Fig. 5 Using electrolytic capacitors in filter circuits By simply connecting the - terminal to the most negative supply voltage in the circuit, this capacitor will never become reverse biased. It can be shown that the circuit of Fig. 5b has the same transfer function as that of Fig. 5a but in the latter case the electrolytic capacitor can be used since it never becomes reverse biased. Although this example focused on the lowpass filter, similar methods can be used when it is desired to utilize electrolytic capacitors in other filter circuits. Driving LEDs on Pulse Edges Circuits that provide a short-duration pulse on a rising or falling edge of a signal are termed monostable multivibrators. The operation of monostable multivibrators is similar to that of the triangle and square wave generators that were discussed in class. Monostable multivibrators are discussed in the text for this course. Laboratory Procedure Part 1. Design and test a Body Emulator that can be used for helping in the design of the ECG amplifier Part 2. Design and test an amplifier circuit that will amplify the ECG signals so that the peak amplitude of the ECG signal is between 5V and 8V. Your laboratory report should include several periods of the ECG signal you obtained. Be sure that you also include an analysis of any amplifier circuits and filter circuits you use along with the measured Page 5 of 6

performance with the Body Emulator in your report. These circuits should all be tested with the Body Emulator and other appropriate test equipment before the test electrodes are attached to the body. (When attaching the electrodes, attach the ground electrode to the right leg near the ankle. Attach the two signal electrodes to the left and right wrists. Use alcohol and a cotton ball to remove any surface oils prior to attaching the electrodes). Be sure to have your TA check the circuit prior to attaching the electrodes. Remember that the circuit should be powered by 9V batteries any time the electrodes are attached to the patient. Part 3. Add a circuit to the amplifier that will turn on a RED LED for 0.1sec every time the heart beats. Parts 1, 2, and 3 must be completed. As discussed above, differing levels of filtering will affect the quality of the ECG signal. The grading of this experiment will depend, in part, on how clean of a signal is obtained in Part 2. Page 6 of 6