Level instruments. Continuous level measurement - Radar transmitters SITRANS LR300. 5/180 Siemens FI Overview

Similar documents
SITRANS L Level instruments

Level Measurement. Continuous level measurement - Radar transmitters SITRANS LR400 5/229. Overview

Level Measurement. Continuous level measurement - Radar transmitters SITRANS LR400 5/229. Overview

Level instruments. Continuous level measurement - Radar transmitters SITRANS LR200. 5/200 Siemens FI Configuration.

Level Measurement. Continuous level measurement Radar transmitters SITRANS LR460. 4/258 Siemens FI

Level Measurement Continuous level measurement Radar transmitters

Level Measurement. Continuous level measurement - Radar transmitters SITRANS LR200. 5/200 Siemens FI Overview.

Level Measurement Continuous level measurement Radar transmitters

Level Measurement Continuous level measurement - Radar transmitters

Level Measurement Continuous level measurement Radar transmitters SITRANS LR200. Overview. Configuration

Continuous level measurement Radar transmitters

Continuous level measurement Radar transmitters

INDUSTRIAL CONTROL INSTRUMENTS FOR PROCESS AUTOMATION. Configuration. Installation

SITRANS LR 300: High Performance Radar for Level Measurement

Level Measurement Continuous level measurement - Radar transmitters

Level instruments. Continuous level measurement - Radar transmitters SITRANS LR250. 5/210 Siemens FI Overview.

Level Measurement Continuous level measurement - Radar transmitters

Level Measurement. Continuous level measurement - Capacitance transmitters SITRANS LC300 5/271. Overview. Configuration

Level Measurement. Continuous level measurement Capacitance transmitters SITRANS LC300. icenta Controls Ltd

SITRANS T measuring instruments for temperature

SITRANS L Level instruments

SITRANS F flowmeters SITRANS F US. Transmitter FUS060. Overview

Siemens AG Level Measurement

Flow Measurement SITRANS F M. Flow sensor MAG /49. Overview

Level Measurement. 4/2 Overview

SITRANS F flowmeters SITRANS F M. Flow sensor MAG /48 Siemens FI Overview

Level Measurement. 4/2 Product overview

Temperature Measurement

Level Measurement. 4/2 Overview

SITRANS F flowmeters SITRANS F M. Transmitter Transmag 2 with sensor 911/E 4/77. Overview

Level measurement Radar

RHE12. Hazardous Area Coriolis Mass Flow Transmitter. Features. Applications. Benefits

TEK-F LEX 4100A. Guided Wave Radar Level Transmitter. LEVEL. Technology Solutions

FLOWMETER. Siemens MAG 5000/6000. Overview

The Radar Level Meters can be installed both in metallic or non metallic tanks; their use is not harmful to humans or environment.

SITRANS L Level instruments

Instruction Manual September sitrans LR 300

Instruction Manual September sitrans LR 300 (HART)

DREXELBROOK. IMPULSE Series. Guided Wave Radar (TDR) Easy to install, Easy to use. Dependable performance. The New Impulse Guided Wave Radar

Flow Measurement SITRANS F M

OEM radar transmitter, for aggressive media level measurement

Instruction Manual May iq radar

SITRANS F C. Flowmeter SITRANS FC430 3/151. Technical specifications. Overview

DREXELBROOK. IMPULSE Series Guided Wave Radar (TDR) for Total Level and Volumetric Measurements. Easy to install, Easy to use. Dependable performance

IQ RADAR 300. Instruction Manual March 2001

RADAR TRANSMITTERS BLISSFLEX 1100 SELECTION GUIDE. The characters of the order code have been described below

2-Wire. Level Gauge BM 702

SITRANS F flowmeters SITRANS F US. SONOFLO SONO 3000 transmitter. 4/152 Siemens FI Overview

Instruction Manual May sitrans LR300

Radar Level Measurement micropilot FMR 130 for stilling wells

DREXELBROOK DREXELBROOK. Impulse. Series IMPULSE. IMPULSE Series. Guided Wave Radar (TDR) for Total Level and Volumetric Measurements DATA SHEET

ABB MEASUREMENT & ANALYTICS DATA SHEET. Ultrasonic level LST400 Open channel flow transmitter

LST400 Ultrasonic level / open channel flow transmitter

SITRANS T measuring instruments for temperature

Rosemount TGU 65 Tank Radar Gauge, Still Pipe Antenna

Flow Measurement SITRANS F M

3DLevelScanner. Technical Data Sheet. Level Sensor. Non-Contact, Multiple-Point Measurement

Coriolis Massflowmeter

TEK-TEMP 2100A. Explosion-Proof Temperature Transmitter. TEMPERATURE. Technology Solutions

IQ RADAR 160. Instruction Manual PL-550 January Rev. 3.2

Coriolis Massflowmeter

SITRANS F flowmeters SITRANS F US. Transmitter FUS080. 4/194 Siemens FI

Coriolis Mass Flow Meter Model TM

LEVEL WITH US. to Effectively Minimize Process Upsets. 25 GHz pulse radar for liquids and slurries in general service and process applications

Temperature Measurement Transmitters for mounting in sensor head

Echotel Model 910 Ultrasonic Level Switch

Flow Measurement SITRANS F M. Transmitter TRANSMAG 2 with sensor 911/E 4/101. Overview

Switching. Radar. Overview 114 VEGAMIP series Mounting adapter 120

Servo Gauge 854 XTG. Global Experience. Locally Applied. Cut costs, not repeatability.

RADIATED MICROWAVES. Application. Working frequency. 6 GHz Band C 6 GHz Band C. Working temperature C C.

Design and applications

3DLevelScanner. Technical Data Sheet. Level Sensor. Non-Contact, Multiple-Point Measurement

Magnetostrictive Level Transmitter

Technical Information Condumax CLS21D/CLS21

Process display RIA 261

Temperature Measurement

Rosemount Specifications. Product Data Sheet , Rev BA February 2009 PERFORMANCE SPECIFICATIONS

OPTIFLUX 5000 Technical Datasheet

TMU. Coriolis Mass Flow Meter. for liquids and gas

Main features. Operation. LEVEL Transmitters

Instruction Manual August sitrans LR 200

ED701 General Industry Pressure Transmitter

SITRANS F flowmeters SITRANS F US SONOCAL /184 Siemens FI

The transducer component converts the measured signal and displays the actual value.

Temperature Measurement

Temperature Measurement Transmitters for mounting in sensor head

Target disk flowmeter for liquids and gases

APT3200L SMART PRESSURE TRANSMITTER. Application Areas: WITH DIAPHRAGM SEAL

Level gauges Series LT Level indicator, switch and transmitter for liquids

MJK Expert 700 / 800 / 900 / 1100 / 2100 MJK Expert 1400 / 3400

ENRAF SERVO GAUGE 854 ATG Setting the standard in gauging.

Positioner with HART Communication Type 3780

Ultrasonic level transmitter, non-contact

STT 3000 Series STT170 SMART TEMPERATURE TRANSMITTER Models STT171, STT173, STT17H, STT17F, STT17C

PiloTREK SZINTTÁVADÓK LEVEL TRANSMITTERS PU L S E BU RS T R A DA R LE V E L TR A NSMIT TERS K- BA N D R A DA R FO R LI Q U I DS

Mobrey Ultrasonic. MSP422, MSP400RH, and MSP900GH Level Transmitters. Product Data Sheet February 2015 IP2045, Rev EA

Accuracy: ± 0.1 % of reading ±zero point stability (from liquids up to TMU-x040) Material: (316 L) / (316 Ti)/ Hastelloy C-22

PiloTREK SZINTTÁVADÓK. Pulse burst radar level transmitters 5 YEARS WARRANTY

Echotel Model 910 Ultrasonic Level Switch

Rosemount 3101, 3102, and 3105

Transcription:

Overview is a 4-wire, 6 GHz pulse radar level transmitter for continuous and process vessels including high temperature and pressure, to a range of 20 m (66 ft). Benefits Auto False-Echo Suppression Infrared Intrinsically Safe handheld or remote programming.8 GHz (U.S.A. 6.3 GHz) Built-in diagnostics Various flanges, horn and waveguide antenna options available Extremely high signal-to-noise ratio Application The is available both for standard applications and for applications that require explosion proof protection. The features a compact design and robust construction and is available with an epoxy-coated aluminum or stainless steel enclosure. Operating at low frequency and high signal transmission speed, it is virtually unaffected by atmospheric or temperature conditions. It provides reliable measurement in environments with harsh chemicals, steam, dust, encrustation, turbulence and agitation. The high resistance PTFE rod antenna is chemically immune and resistant to material buildup. The can communicate using Modbus or HART protocols. When using the flange or threaded antenna options, the SI- TRANS LR300 is easily installed by positioning the unit on a standpipe, bolting or threading it in place and connecting it to the power supply. There s no need to fill or empty the vessel for calibration or commissioning. Key Applications: liquid bulk storage tanks, agitated process vessels /180 Siemens FI 01 2009

Level instruments Configuration installation Siemens FI 01 2009 /181

Technical specifications Mode of operation Measuring principle Radar level measurement Frequency.8 GHz (U.S.A. 6.3 GHz) Input Measuring range 0.4 to 20 m (1.3 to 6 ft) Output Output signal Analog output Optically isolated, 4 to 20 ma - Load Max. 40 Ω - Accuracy 0.02 ma Communications Modbus/RS-48 connection or HART Performance (reference conditions) Error in measurement at +20 C ±1 mm from 0.4 to 10 m (+68 F) ±0.1% from 10 to 20 m - Temperature drift < ±0.2% of range, -40 to +60 C (-40 to +140 F) - Repeatability ±2 mm, up to 3 m ±3 mm, from 3 m to m ± mm, from m to 10 m ±10 mm, from 10 m to 20 m - Fail-safe ma signal programmable as high, low or hold (Loss of Echo) Rated operating conditions Installation conditions Location Indoor/outdoor Ambient conditions (enclosure) Ambient temperature -40 to +60 C (-40 to +140 F) Installation category II Pollution degree 4 Medium conditions Dielectric constant ε r ε r > 1.6 (For εr < 3, use waveguide antenna or stillpipe) Temperature -40 to +200 C (-40 to +392 F) Pressure (vessel) Dependent on process connection and temperature (refer to Pressure/Temperature curves) Design Enclosure - Material Aluminium, epoxy coated or optional stainless steel - Cable inlet 2 x ½" NPT or M20x1. Degree of protection Type 4X/NEMA 4X, Type 6/NEMA 6, IP67, IP68 Weight 6. kg (14.3 lbs) with 2"/10 psi flange; weight varies depending on flange size and pressure rating Dielectric rod antennas - Material PTFE - Dimensions 41 cm (16.3") long including integral gasket (other antenna types available) Process connections Flanged Other connections Flat faced flanges: 316L stainless steel, 0, 80, 100, 10, 200 mm (2, 3, 4, 6, 8"), bolt hole pattern to ASME, DIN and JIS sizes Available (please see order information) Power supply Universal power supply unit 48 to 230 V AC ± 1%, 40 to 70 Hz, 28 VA (11 W) 24 to 230 V DC ± 1%, 9 W Certificates and approvals Safety CSA NRTL/C, CE, FM, C-TICK Marine Lloyd s Register of Shipping ABS Type Approval Radio Europe, Industry Canada, FCC, C-TICK Explosion Proof ATEX II 1/2 G EEx de IIC T6 ATEX II 1/2 G EEx de [ia] IIC T6 CSA Class I, Div 1, Groups A, B, C, D, E, F, G FM Class I, Div.1, Groups A, B, C, D, E, F and G Sanitary 3A Sanitary (Sanitary rod antenna only) Communication Intrinsically Safe Siemens handheld Infrared receiver programmer - Approvals for handheld programmer Programmer (remote keypad) PC Display (local) IS model with ATEX EEx ia IIC T4, CSA/FM Class I, Div. 1, Groups A, B, C, D Modbus or HART; upgradable Flash via RS-48 SIMATIC PDM Alphanumeric and multi-graphic liquid crystal for readout and entry /182 Siemens FI 01 2009

Level instruments, horn antenna version C) 7 M L 4 1 1 - Antenna version Stainless steel 316L with PTFE emitter 0 Stainless steel 316L with PTFE emitter and purging kit 1) 1 Sliding waveguide system 1000 mm (40") long 1) 2) 2 Flange design (316L Stainless Steel) DN 0 PN 16, Type A, flat faced 1) AA DN 80 PN 16, Type A, flat faced BA DN 100 PN 16, Type A, flat faced CA DN 10 PN 16, Type A, flat faced DA DN 200 PN 16, Type A, flat faced EA 2" ASME, 10 lb 1) FB 3" ASME, 10 lb, flat faced GB 4" ASME, 10 lb, flat faced HB 6" ASME, 10 lb, flat faced JB 8" ASME, 10 lb, flat faced 1) KB DN 0 PN 40, flat faced 1) AC DN 80 PN 40, flat faced BC DN 100 PN 40, flat faced CC DN 10 PN 40, flat faced DC DN 200 PN 40, flat faced EC 2" ASME, 300 lb, flat faced 1) FD 3" ASME, 300 lb, flat faced GD 4" ASME, 300 lb, flat faced HD 6" ASME, 300 lb, flat faced JD 8" ASME, 300 lb, flat faced KD JIS DN 0 10K 1) AE JIS DN 80 10K BE JIS DN 100 10K CE JIS DN 10 10K DE JIS DN 200 10K EE (Note: Flange bolting patterns and facings dimensionally correspond to the applicable ASME B16., or EN 1092-1, or JIS B 2238 standard.) Communication/output 4 to 20 ma, HART, Modbus 0 Process seal/gasket FKM 0 Nitrile, only for sliding waveguide systems 1 FFKM [-3 to +200 C (-31 to +392 F)] 2 Enclosure/Cable inlet Aluminum, Epoxy painted 2 x ½" NPT 0 2 x M20x1. 1 Stainless steel 316L 2 x ½" NPT 2 2 x M20x1. 3 Horn size/waveguide options 80 mm (3") horn 3) B 100 mm (4") horn 3) C 10 mm (6") horn D) D 200 mm (8") horn E 100 mm (4") horn with 100 mm (4") waveguide D) F 3) 100 mm (4") horn with 10 mm (6") waveguide D) G 3) 100 mm (4") horn with 200 mm (8") waveguide H 3) 100 mm (4") horn with 20 mm (10") waveguide 3) J, horn antenna version C) 7 M L 4 1 1-10 mm (6") horn with 100 mm (4") waveguide K 10 mm (6") horn with 10 mm (6") waveguide L 10 mm (6") horn with 200 mm (8") waveguide M 10 mm (6") horn with 20 mm (10") waveguide N 200 mm horn with 100 mm (4") waveguide P 200 mm horn with 10 mm (6") waveguide Q 200 mm (8") horn with 200 mm (8") waveguide R 200 mm (8") horn with 20 mm (10") waveguide S Waveguide only - Waveguide length 00 mm to T 3000 mm (in 1 mm increments) (Add order code Y01 and plain text: waveguide length...mm") Horn with custom waveguide lengths 101 to U 2000 mm (Add order code Y01 and plain text: waveguide length...mm", and Add order code Y03 and plain text: horn size...mm") 4) Approvals ) General Purpose, CE, CSA US/C A CSA Class I, Div 1, Groups A, B, C, D, E, F, G, CE ) D ATEX II 1/2 G EEx de IIC T6, CE ) E FM Class I, Div 1, Groups A, B, C, D, E, F, G, FCC, F 6.3 GHz, for U.S.A. only General Purpose, FM, FCC, 6.3 GHz, for U.S.A. G only ATEX II 1/2 G EEx de [ia] IIC T6 J Pressure rating Rating per Pressure/Temperature curves in manual 0 0. bar g (7.2 psi g maximum) 1 Further designs Order code Please add -Z" to and specify Order code(s). Acceptance test certificate: Manufacturer's test C11 certificate M to DIN 30, Part 18 and to ISO 9000 Inspection Certificate Type 3.1 per EN 10204 C12 Stainless steel tag [69 mm x 0 mm (2.71 x 1.97")]: Y1 Measuring-point number/identification (max. 16 characters) specify in plain text Horn with waveguide custom lengths: Y01 Enter the total length of the waveguide in plain text description (1 mm increments) Enter the Horn size in plain text description Y03 [100 mm (3.94"), 10 mm (.91"), and 200 mm (7.87") only] Siemens FI 01 2009 /183

, horn antenna version C) 7 M L 4 1 1 - Instruction manual English C) 7ML1998-CL0 French C) 7ML1998-CL14 German C) 7ML1998-CL34 Note: Instruction manual should be ordered as a separate line item on the order. Multi-language Quick Start manual C) 7ML1998-QA8 This device is shipped with the Siemens Milltronics manual CD containing the complete ATEX Quick Start and instruction manual library. Accessories Handheld programmer for, C) 7ML830-2AH Intrinsically Safe, EEx ia Enclosure wrench 7ML1830-1HB Tag, stainless steel with hole, 7ML1930-1BJ 12 x 4 mm (0.47 x 1.77"), for fastening on mounting-eye, one text line, (e.g. for Sensors) HART modem/rs-232 (for use with a PC and D) 7MF4997-1DA HART modem/usb (for use with a PC and D) 7MF4997-1DB RS 48 to RS 232 converter, not isolated, C) 7ML1830-1HA port powered One metallic cable gland M20x1., rated -40 to 7ML1930-1AP +80 C (-40 to +176 F) for General Purpose or ATEX EEx e installations (available for HART only) SITRANS RD100 Remote display - see RD100 on page /263 SITRANS RD200 Remote display - see RD200 on page /26 1) Available with pressure rating option 1 only 2) Available with horn size option C to E only 3) For stilling well applications only 4) Order standard waveguide lengths 100, 10, 200, 20 mm (3.93,.91, 7.87, 9.84") by choosing options from Horn size/waveguide options ) Includes European Radio and Industry Canada approvals,.8 GHz, C-TICK C) Subject to export regulations AL: N, ECCN: EAR99 D) Subject to export regulations AL: N, ECCN: EAR99H /184 Siemens FI 01 2009

Level instruments, sanitary version C) 7 M L 4 1 2-7 7 777-7777 Antenna version PTFE, one piece rod antenna 0 UHMW-PE, one piece rod antenna 1 Process connection Sanitary fitting clamp A Configuration/connection size Only for rod antenna 2" (0.8 mm) connection A 3" (76.2 mm) connection B 4" (101.6 mm) connection C Antenna Without antenna 0 Mounting Clamp No mounting clamp 0 Mounting clamp included 1) 1 Enclosure/cable inlet Aluminum, epoxy coated 2 x ½" NPT 0 2 x M20x1. 1 Stainless steel 316L 2 x ½" NPT 2 2 x M20x1. 3 Communication/output 4 to 20 ma, HART, Modbus A Approvals General Purpose, CE, CSA 2) US/C A CSA Class I, Div I, Groups A, B, C, D, E, F, G, CE 2) D) D ATEX II 1/2 G EEx de IIC T6, CE 2) E FM Class I, Div I, Groups A, B, C, D, E, F, G, FCC, D) F 6.3 GHz, for U.S.A. only General Purpose, FM, FCC, 6.3 GHz, for U.S.A. D) G only ATEX II 1/2 G EEx de [ia] IIC T6 J Pressure rating Rating per Pressure/Temperature curves in manual 0 0. bar g (7.2 psi g maximum) 1 Further designs Order code Please add -Z" to and specify Order code(s). Acceptance test certificate: Manufacturer's test C11 certificate M to DIN 30, Part 18 and to ISO 9000 Inspection Certificate Type 3.1 per EN 10204 C12 Stainless steel tag [69 mm x 0 mm (2.71 x 1.97")]: Y1 Measuring-point number/identification (max. 16 characters) specify in plain text, sanitary version C) 7 M L 4 1 2-7 7 777-7777 Instruction manual English C) 7ML1998-CL0 French C) 7ML1998-CL14 German C) 7ML1998-CL34 Note: The instruction manual should be ordered as a separate line item on the order. Multi-language Quick Start manual C) 7ML1998-QA8 This device is shipped with the Siemens Milltronics manual CD containing the complete ATEX Quick Start and instruction manual library. Accessories Handheld programmer for, Intrinsically Safe, EEx ia C) 7ML830-2AH Tag, stainless steel with hole, 12 x 4 mm (0.47 x 7ML1930-1BJ 1.77"), for fastening on mounting-eye, one text line, (e.g. for sensors) Sanitary fitting clamp, stainless steel: 2" (0.8 mm) 7ML1830-1HD 3" (76.2 mm) 7ML1830-1HE 4" (101.6 mm) 7ML1830-1HF Enclosure wrench 7ML1830-1HB HART modem/rs-232 (for use with a PC and D) 7MF4997-1DA HART modem/usb (for use with a PC and D) 7MF4997-1DB RS 48 to RS 232 converter, not isolated, port C) 7ML1830-1HA powered One metallic cable gland M20x1., rated -40 to 7ML1930-1AP +80 C (-40 to +176 F) for General Purpose or ATEX EEx e installations (available for HART only) SITRANS RD100 Remote display - see RD100 on page /263 SITRANS RD200 Remote display - see RD200 on page /26 1) Available with pressure rating option 1 only 2) Includes European Radio and Industry Canada approvals,.8 GHz, C-TICK C) Subject to export regulations AL: N, ECCN: EAR99 D) Subject to export regulations AL: N, ECCN: EAR99H Siemens FI 01 2009 /18

, rod antenna version C) 7 M L 4 1 3 - Antenna version PTFE 0 Process connection Flange version (316L Stainless Steel) DN 0 PN 16, Type A, flat faced AA DN 80 PN 16, Type A, flat faced BA DN 100 PN 16, Type A, flat faced CA DN 10 PN 16, Type A, flat faced DA 2" ASME, 10 lb, flat faced FB 3" ASME, 10 lb, flat faced GB 4" ASME, 10 lb, flat faced HB 6" ASME, 10 lb, flat faced JB DN 0 PN 40, flat faced AC DN 80 PN 40, flat faced BC DN 100 PN 40, flat faced CC DN 10 PN 40, flat faced DC 2" ASME 300 lb, flat faced 1) FD 3" ASME, 300 lb, flat faced GD 4" ASME, 300 lb,flat faced HD 6" ASME, 300 lb, flat faced JD JIS DN 0 10K AE JIS DN 80 10K BE JIS DN 100 10K CE JIS DN 10 10K DE (Note: Flange bolting patterns and facings dimensionally correspond to the applicable ASME B16., or EN 1092-1, or JIS B 2238 standard.) Threaded version (316L Stainless Steel) 1½" NPT [(Taper), ANSI/ASME B1.20.1] LA 2" NPT [(Taper), ANSI/ASME B1.20.1] MA R 1½" [(BSPT), EN 10226] LC R 2" [(BSPT), EN 10226] MC G 1½" [(BSPP), EN ISO 228-1] LE G 2" [(BSPP), EN ISO 228-1] ME Antenna, or inactive shield length Without antenna 0 PTFE 0 mm (1.97") 1 100 mm (3.94") 2 Stainless steel 316L 100 mm (3.94") 2) 3 10 mm (.91") 2) 4 200 mm (7.87") 2) 20 mm (9.84") 2) 6 Custom inactive shield length 101 mm to 1000 mm 7 (in 1 mm increments) Add order code Y01 and plain text: Inactive shield length...mm" 2) Process Seal/Gasket Integral Gasket 3) 4) 0 FKM O-ring ) 1 Enclosure/Cable inlet Aluminum, Epoxy painted 2 x ½" NPT 0 2 x M20x1. 1 316L Stainless steel 2 x ½" NPT 2 2 x M20x1. 3 Communication/output 4 to 20 ma, HART, Modbus A, rod antenna version C) 7 M L 4 1 3 - Approvals 6) General Purpose, CE, CSA US/C A CSA Class I, Div I, Groups A, B, C, D, E, F, G, CE 6) D) D ATEX II 1/2 G EEx de IIC T6, CE E FM Class I, Div I, Groups A, B, C, D, E, F, G, FCC, D) F 6.3 GHz, for U.S.A. only General Purpose, FM, FCC, 6.3 GHz, for U.S.A. D) G only ATEX II 1/2 G EEx de [ia] IIC T6 J Pressure rating Rating per Pressure/Temperature curves in manual 0 0. bar g (7.2 psi g maximum) 1 Further designs Order code Please add -Z" to and specify Order code(s). Acceptance test certificate: Manufacturer's test certificate M to DIN 30, Part 18 and to ISO 9000 C11 Inspection Certificate Type 3.1 per EN 10204 C12 Inactive custom shield lengths: Enter the total Y01 length of the inactive shield in plain text description (in 1 mm increments). Stainless steel tag [69 mm x 0 mm (2.71 x 1.97")]: Y1 Measuring-point number/identification (max. 16 characters) specify in plain text Instruction manual English C) 7ML1998-CL0 French C) 7ML1998-CL14 German C) 7ML1998-CL34 Note: Instruction manual should be ordered as a separate line item on the order. Multi-language Quick Start manual C) 7ML1998-QA8 This device is shipped with the Siemens Milltronics manual CD containing the complete ATEX Quick Start and instruction manual library. Accessories Handheld programmer for, Intrinsically C) 7ML830-2AH Safe, EEx ia Antenna rod, PTFE 7ML1830-1HC Antenna, 0 mm (2"), PTFE 7ML1830-1CH Antenna, 100 mm (4"), PTFE 7ML1830-1CG Tag, stainless steel with hole, 7ML1930-1BJ 12 x 4 mm (0.47 x 1.77"), for fastening on mounting-eye, one text line, (e.g. for sensors) Enclosure wrench 7ML1830-1HB HART modem/rs-232 (for use with a PC and D) 7MF4997-1DA HART modem/usb (for use with a PC and D) 7MF4997-1DB RS 48 to RS 232 converter, not isolated, port C) 7ML1830-1HA powered One metallic cable gland M20x1., rated -40 to 7ML1930-1AP +80 C (-40 to +176 F) for General Purpose or ATEX EEx e installations (available for HART only) SITRANS RD100 Remote display - see RD100 on page /263 SITRANS RD200 Remote display - see RD200 on page /26 1) For use with pressure rating option 1 only 2) For use with all process connection options except AA, FB, AC, FD, AE, LA, LC and LE 3) Available with flat faced flange process connections only 4) Available with antenna options 0 to 2 only ) Not available for combination of flat faced flanges with antenna options 0 to 2 6) Includes European Radio and Industry Canada approvals,.8 GHz, C-TICK C) Subject to export regulations AL: N, ECCN: EAR99 D) Subject to export regulations AL: N, ECCN: EAR99H /186 Siemens FI 01 2009

Level instruments Characteristic curves Process Pressure/Temperature derating curves Schematics connections Siemens FI 01 2009 /187

Dimensional drawings dimensions /188 Siemens FI 01 2009