Technical Information Smartec CLD18

Similar documents
Technical Information Smartec CLD18

Technical Information Smartec CLD18

Technical Information Indumax CLS54

Indumax CLS54D. Technical Information

Technical Information Memosens CLS82D

Smartec S CLD134. Technical Information

Technical Information Indumax CLS50D/CLS50

Technical Information Condumax CLS21D/CLS21

Conductivity measurement SmarTec S CLD 132

Technical Information Condumax CLS15D/CLS15

Indumax H CLS54. Technical Information

Condumax W CLS19. Technical Information. Conductivity sensors Two-electrode sensors with cell constants k = 0.01 cm 1 or k = 0.

Condumax CLS21 and CLS21D

Conductivity Transmitter

Digital flow ELEMENT transmitter for continuous flow measurement

Condumax W CLS15 and CLS15D

Condumax CLS16 and CLS16D

Electromagnetic Flow Transmitter

Indumax CLS50/CLS50D. Technical Information

Type PLC Fitting. Valve for Continuous control. Diaphragm valve. Output

Digital flow ELEMENT transmitter for continuous flow measurement

Digital inductive conductivity transmitter

Inductive Conductivity Meter ILM

The transducer component converts the measured signal and displays the actual value.

Condumax CLS15 and CLS15D

Conductivity sensor for hygienic applications

Condumax CLS15 and CLS15D

Endress Hauser. SmarTec M CLD 133 Conductivity Measurement

Condumax W CLS19. Technical Information. Conductivity sensors Two-electrode sensors with cell constants k = 0.01 cm 1 or k = 0.

Digital conductivity transmitter

Ceramax CPS341D. Technical Information

prosonic T FMU230, FMU231

Conductivity Measuring Cells for High-Temperature Applications ConduMax H CLS 16

RTD Temperature Sensor omnigrad T -TSM 480, TR 480. Hygienic RTD sensor with Pt 100, class A TSM 480 with electronics programmable via PC

ph or ORP Transmitter

JUMO Dtrans T100 Screw-in RTD Temperature Probe with/without Transmitter

Condumax CLS21 and CLS21D

omnigrad T - TSM 480, TR 480. Technical Information. RTD Temperature Sensor Hygienic design TSM 480 with electronics programmable via PC

Inductive Conductivity/Concentration and Temperature Transmitter with switch contacts. Temperature sensor. Graphics LC display.

INLINE flowmeter for continuous flow measurement

INSERTION paddle wheel flowmeter for continuous flow measurement

INLINE flowmeter for continuous flow measurement

Technical Information Orbipore CPS91D and CPS91

Pressure transmitter For sanitary applications Model SA-11

Continuous Level Sensor NSL-M-00

INLINE flowmeter for continuous flow measurement

Magnetic-Inductive flow Meter fmi

Inductive conductivity measurement mypro CLM 431 / CLD 431 inductive

Technical Information Nivector FTC968, FTC968Z

Inductive conductivity transmitter JUMO CTI-Junior

Digital flow transmitter for continuous flow measurement

OPTIFLUX 5000 Technical Datasheet

LCI. Inductive Conductivity/ Concentration and Temperature Transmitter. Compact or separate version

Condumax W CLS15 and CLS15D

SITRANS F flowmeters SITRANS F M. Transmitter Transmag 2 with sensor 911/E 4/77. Overview

Digital flowmeter for continuous flow measurement

Technical Information Orbipac CPF81D/CPF82D and CPF81/CPF82

Continuous Level Sensor NSL-M

P R o D u c T I n F o R m at I o n. TDR Level Sensor LFP. The clean solution

Pharma Temperature Sensor G3/8"

Ultrasonic level transmitter, non-contact

Digital flowmeter for continuous flow measurement

FLOWave SAW-Flowmeter

Conductive conductivity measurement mypro CLM 431 / CLD 431 conductive

Temperature Sensor with G1/2" hygienic

Pharma Temperature Sensor G3/8"

Pharma-Temperature Sensor M12 hygienic

Flush pressure transmitter For viscous and solids-containing media Model S-11

Temperature Sensor Tri-Clamp

JUMO ecotrans Lf 03 Microprocessor Transmitter / Switching Device for conductivity or resistivity and temperature

In-Line Resistance Thermometers Model TR25

Flow Measurement SITRANS F M

OPTIFLUX 5000 Technical Datasheet

Temperature Sensor with M12 hygienic

Technical Information Liquisys M CLM223/253

Temperature Sensor with M12 hygienic

Conductivity Sensor for High-Temperature Applications ConduMax W CLS 13

OEM pressure transmitter for general industrial applications Models O-10 (T), O-10 (5)

Memosens CPS96D. Technical Information

Technical Information OUSBT66

Smartec S CLD134. Operating Instructions. Conductivity Measuring System. BA401C/07/en/ Software version 1.

Turbimax CUS71D. Technical Information. Ultrasonic interface sensor Immersion sensor for interface measurement

PD30ETB20xxIS. Photoelectrics, Background Suppression reflective with IR light. Main features. Description

OPTIFLUX 1000 Technical Datasheet

High-quality pressure transmitter For general industrial applications Model S-10

BATCHFLUX 5500 C Technical Datasheet

Temperature Sensor Tri-Clamp

A Higher Level of Performance. Data Sheet. Series Pressure and Level Transmitters. For more information, please visit >

Temperature Sensor. Temperature Transmitter MPU-LCD with Display Product Information TSM and TSB FOOD SENSORS FOR FOOD AND LIFE SCIENCES.

Technical Information Memosens CCS50D

JUMO CTI-500. Inductive Conductivity/Concentration and Temperature Transmitter with switch contacts. Type Brief description.

Valve Island with Electronic I/O

Electromagnetic Flow Measuring System Two-wire, loop-powered PROline promag 23 H

Intrinsically safe pressure transmitter MBS 4201, MBS 4251, MBS 4701 and MBS 4751

Digital batch controller

PD30CNB25xxPS. Photoelectrics, Background suppression reflective - PointSpot. Main features. Main functions. Description

Technical Information Proline Promag 10P

Temperature Sensor without thread

Miniature resistance thermometer Model TR33, thread-mounted

ED701 General Industry Pressure Transmitter

Transcription:

TI01080C/07/EN/03.16 71314717 Products Solutions Services Technical Information Smartec CLD18 Compact inductive conductivity measurement for the food and beverage industry Application Compact measuring system for inductive conductivity measurement in liquids with medium to high conductivity. The structure is made from highly durable, food-safe polyetheretherketone (PEEK). The excellent chemical resistance properties of the sensor mean it can also be used in applications outside the food industry. The measuring system is perfectly suited for: Phase separation of product/water mixtures in the beverage industry Control of cleaning-in-place facilities (CIP), concentration control, separation in the cleaning return line Industrial water monitoring Rinsing processes in pickling plants Your benefits Hygienic sensor design in accordance with EHEDG and 3-A requirements, no risk of recontamination Stainless steel or plastic transmitter housing, IP 69K, can be cleaned with high-pressure steam High level of repeatability, 0.5 % of reading, always guarantees same separation or monitoring

Function and system design Measuring principle Inductive conductivity measurement An oscillator (1) generates an alternating magnetic field in the primary coil (2), which induces a current flow (3) in the medium. The strength of the current depends on the conductivity and thus on the ion concentration in the medium. The current flow in the medium, in turn, generates a magnetic field in the secondary coil (4). The resulting induced current is measured by the receiver (5) and used to determine the conductivity. 1 5 2 4 3 1 Inductive conductivity measurement 1 Oscillator 2 Primary coil 3 Current flow in the medium 4 Secondary coil 5 Receiver A0004894 Advantages of inductive conductivity measurement: No electrodes and therefore no polarization effects Accurate measurement in media with a high degree of pollution and a tendency to form buildup Complete galvanic isolation of the measurement and the medium 2 Endress+Hauser

Key properties Hygiene The PEEK sensor (polyetheretherketone) has excellent chemical, mechanical and thermal resistance properties, is entirely seamless and gap-free and is therefore safe to use in hygienic applications. The sensor has 3A approval which is required for the hygiene sector. All wetted materials are FDA-listed. The sensor has been designed according to the latest guidelines of the American Society of Mechanical Engineers - Bioprocessing Equipment (ASME BPE) and meets the requirements of EC Regulation No. 1935/2004 of the European Parliament on materials and articles intended to come into contact with food. Process connections The measuring system is available with all the process connections commonly used in the hygiene sector. The measuring system is also available with a G1½ process adapter and 2¼" coupling nut for use in non-hygienic areas. Process temperature, process pressure The sensor is suitable for continuous operating temperatures up to 110 C (230 F). It can be operated up to 130 C (266 F) for short periods (max. 60 min) for sterilization phases. The pressure resistance is 13 bar (188 psi) abs up to 50 C (122 F). If the PVC process adapter is used, the maximum pressure resistance drops to 9 bar (130 psi) abs at 25 C (77 F) and the maximum process temperature drops to 60 C (140 F). Temperature measurement A Pt 1000 temperature sensor is integrated into the sensor. This ensures efficient phase separation at different and quickly changing process temperatures. The temperature sensor is embedded in the PEEK body. Temperature compensation Linear temperature compensation with user-configurable temperature coefficients α is implemented in the measuring system. Input Measured values Conductivity Temperature Measuring range Conductivity: Recommended range: 200 μs/cm to 1000 ms/cm (uncompensated) Temperature: 10 to +130 C (+14 to +266 F) Temperature measurement Pt 1000 Binary input The binary input is used for measuring range switching. Voltage range Voltage HIGH min. Voltage LOW max. Current consumption at 24 V Undefined voltage range 0 V to 30 V 12 V 9.0 V 30 ma 9.0 to 12 V Endress+Hauser 3

Output Output signal Conductivity: 0 / 4 to 20 ma, galvanically isolated Temperature: 0 / 4 to 20 ma, galvanically isolated Load Max. 500 Ω Characteristic Linear Signal resolution Resolution: > 13 bit Accuracy: ± 20 μa Alarm output The alarm output is implemented as an "open collector". Max. current Max. voltage Error or device without supply voltage No error 200 ma 30 V DC Alarm output blocked (0 ma) Alarm output open (up to 200 ma) 4 Endress+Hauser

I O ut1+ I O ut1- I O ut2+ I O ut2- MRS+ MRS- Out+ Out- L- Smartec CLD18 Power supply Electrical connection IOut1 ma Lf MRS IOut2 ma L Alarm Out X A0019115 2 Electrical connection Terminal assignment 1 LL+ + - L 2 X A0029684 3 Terminal assignment IOut1 Current output conductivity IOut2 Current output temperature Out Binary input (measuring range switch) MRS Alarm output (open-collector) L+/L- Power supply X Grounding pin (flat male tab 4.8 mm) 1 Cover on electronics box 2 Electronics box Supply voltage 24 V DC ±20 %, protected against reverse polarity Endress+Hauser 5

Power consumption 3 W Cable specification Recommendation 0.5 mm 2 max. 1.0 mm 2 Performance characteristics Response time Conductivity: t95 < 1.5 s Temperature: t90 < 50 s Maximum measured error Conductivity: ± (2.0 % of measured value + 20 µs/cm) Temperature: ± 1.5 K Signal outputs ± 50 μa Repeatability Conductivity: max. 0.5 % of measured value ± 5 µs/cm ± 2 digits Cell constant 11.0 cm -1 Temperature compensation Range 10 to +130 C (+14 to +266 F) Types of compensation None Linear with user-configurable temperature coefficient Reference temperature 25 C (77 F) 6 Endress+Hauser

Installation Installation instructions For a 3-A compliant installation, the following must be noted: After the instrument is installed its hygienic integrity shall be maintained. The instrument shall be installed with the leakage detection at the lowest point of the assembly. Furthermore all process connections must be 3-A compliant. The sensor must be completely immersed in the medium. Avoid air bubbles in the area of the sensor. > 1 m (3.3 ft) 4 Orientation of conductivity sensors A0017691 If the flow direction changes (after pipe bends), turbulence in the medium can result. Install the sensor at a distance of at least 1 m (3.3 ft) downstream from a pipe bend. The product should flow along the hole of the sensor (see the arrows on the housing). The symmetrical measuring channel allows flow in both directions. The ionic current in the liquid is affected by the walls in confined installation conditions. This effect is compensated by what is referred to as the installation factor. The installation factor can be entered in the transmitter for the measurement or the cell constant is corrected by multiplying by the installation factor. The value of the installation factor depends on the diameter and the conductivity of the pipe nozzle as well as the distance a between the sensor and the wall. The installation factor can be disregarded (f = 1.00) if the distance to the wall is sufficient (a > 20 mm, from DN 60). If the distance to the wall is smaller, the installation factor increases for electrically insulating pipes (f > 1) and decreases for electrically conductive pipes (f < 1). It can be measured using calibration solutions, or a close approximation can be determined from the following diagram. a A0018962 5 Installation of CLD18 a Wall distance Endress+Hauser 7

f 1.05 2 1. 00 1 0.95 0 10 20 30 40 a [mm] 0.39 0.79 1.18 1.57 a [inch] 6 Relationship between installation factor f and wall distance a 1 Electrically conductive pipe wall 2 Electrically insulating pipe wall Install the measuring system in such a way that the housing is not exposed to direct sunshine. A0020517 8 Endress+Hauser

Environment Ambient temperature range Stainless steel process connection: -20 to +60 C (-4 to +140 F) PVC process connection: -10 to +60 C (14 to 140 F) Storage temperature Stainless steel process connection: -25 to +80 C (-13 to +176 F) PVC process connection: -10 to +60 C (14 to 140 F) Humidity 100 %, condensating Climate class Climate class 4K4H as per EN 60721-3-4 Degree of protection IP 69k as per EN 40050:1993 Degree of protection NEMA TYPE 6P as per NEMA 250-2008 Shock resistance Complies with IEC 61298-3, certified up to 5 g Vibration resistance Complies with IEC 61298-3, certified up to 5 g Electromagnetic compatibility Interference emission as per EN 61000-6-3:2007 + A1:2011 and EN 55011:2009 + A1:2010 Interference immunity as per EN 61326-1:2013 Process Process temperature Stainless steel process connection: -10 to +110 C (14 to 230 F) Max.130 C (266 F) up to 60 minutes PVC process connection: -10 to +60 C (14 to 140 F) Absolute process pressure Stainless steel process connection: 13 bar (188.5 psi), abs to up to 50 C (122 F) 7.75 bar (112 psi), abs at 110 C (230 F) 6.0 bar (87 psi), abs at 130 C (266 F) max. 60 minutes 1 to 6 bar (14.5 to 87 psi), abs in CRN environment tested with 50 bar (725 psi) PVC process connection: 9 bar (130.5 psi), abs to up to 50 C (122 F) 6.0 bar (87 psi), abs at 60 C (140 F) 1 to 6 bar (14.5 to 87 psi), abs in CRN environment tested with 50 bar (725 psi) Endress+Hauser 9

Pressure-temperature ratings [psi] 188.5 p (abs.) [bar] 13 1 130.5 87 9 6 2 A 14.5 1 50 60 110 130 T[ C] 122 140 230 266 T[ F] 7 Pressure-temperature ratings 1 Stainless steel process connection 2 PVC process connection A Process temperature increased briefly (max. 60 minutes) A0030822-EN Flow velocity max. 10 m/s (32.8 ft/s) for low-viscosity media in pipe DN 50 10 Endress+Hauser

Mechanical construction Design and dimensions 28 (1.10) 85 (3.35) A 78 (3.07) 23 (0.91) 127 (5.00) A 28 (1.10) 73.5 (2.89) 86 (3.39) 136.5 (5.37) 101.7 (4.00) 69 (2.72) 53.3 (2.10) 6 (0.24) 112 (4.41) 85 (3.35) B 5.2 (0.20) 123 (4.84) B 20.5 (0.81) 53.3 (2.10) 69 (2.72) 114.5 (4.51) C 12.3 (0.49) 17.1 (0.67) 32.8 (1.29) 123 (4.84) D 8 Dimensions in mm (inch) and versions (examples) A Plastic housing with thread G 1½ B Stainless steel housing with ISO 2852 clamp 2" C Stainless steel housing with Varivent DN 40 to 125 D Plastic housing with coupling nut 2¼" PVC A0018942 Endress+Hauser 11

Process connections GGE / GCP MDK MEK Ø 48.5 Ø 92 (3.62) Ø 78 (3.07) (1.90) 69 (2.72) 53.3 (2.10) 69 (2.72) 53.3 (2.10) 69 (2.72) 53.3 (2.10) MOK MQK MXK Ø 92 (3.62) Ø 78 (3.07) Ø 82 (3.23) MYK TXJ TDK Ø 94 (3.70) Ø 84 (3.31) Ø 64 (2.52) LQP 32.8 (1.29) 17.1 (0.67) TSK Ø 84 (3.31) 101.7 (4.00) 86 (3.39) 7 (0.28) 32.7 (1.29) Ø 83 (3.27) Ø 39.7 (1.56) 9 GGE GCP MDK MEK MOK MQK MXK MYK TXJ TDK TSK LQP Process connections, dimensions in mm (inch) Thread G1½ Thread G1½ PVC Aseptic DIN 11864-1-A DN 50 Aseptic DIN 11864-1-A DN 40 Dairy fitting DIN 11851 DN 50 Dairy fitting DIN 11851 DN 40 Dairy fitting DIN 11853-2 DN 40 Dairy fitting DIN 11853-2 DN 50 SMS 2" Tri-Clamp ISO 2852 2" Varivent N DN 40 to 125 Coupling nut 2¼" PVC A0018955 12 Endress+Hauser

Weight Stainless steel housing: up to 1.870 kg (4.12 lbs) Plastic housing: up to 1.070 kg (2.36 lbs) Materials In contact with medium Sensor: PEEK (polyetheretherketone) Process connection: Stainless steel 1.4435 (AISI 316 L), PVC-U Seal: EPDM Not in contact with medium Stainless steel housing: Stainless steel 1.4308 (ASTM CF-8, AISI 304) Plastic housing: PBT GF20 Seals: EPDM Window: PC Cable glands: PA, TPE Operability Display and operating elements 2 1 3 Main µs/cm _ - ESC- + E 4 10 Display and keys of the CLD18 1 Parameters 2 Measured value 3 Unit 4 Operating keys The ASTN display (Advanced Super Twisted Nematic) is split into two sections. The segment section displays the measured value. The dot-matrix section displays the parameter and unit. The operating texts are displayed in English. In the event of an error the device automatically alternates between displaying the error and the measured value. A0018963 Endress+Hauser 13

Certificates and approvals Hygiene FDA All materials in contact with the product are FDA-listed materials (apart from the PVC process connections). EHEDG Certified cleanability according to EHEDG Type EL Class I. 3-A When using the sensor in hygienic applications, please note that the cleanability of the sensor also depends on the way the sensor is installed. To install the sensor in a pipe, use the appropriate and EHEDG-certified flow vessels for the particular process connection. Certified according to 3-A Standard 74- ("3-A Sanitary Standards for Sensor and Sensor Fittings and Connections Used on Milk and Milk Products Equipment"). EC Regulation No. 1935/2004 The sensor meets the requirements of EC Regulation No. 1935/2004 on materials and articles intended to come into contact with food. Pressure approval Canadian pressure approval for pipes according to ASME B31.3 mark The product meets the requirements of the harmonized European standards. As such, it complies with the legal specifications of the EC directives. The manufacturer confirms successful testing of the product by affixing to it the mark. Ordering information Product page Product Configurator www.endress.com/cld18 On the product page there is a "Configuration" button to the right of the product image. 1. Click this button. The Configurator opens in a separate window. 2. Select all the options to configure the device in line with your requirements. In this way, you receive a valid and complete order code for the device. 3. Export the order code as a PDF or Excel file. To do so, click the appropriate button on the right above the selection window. For many products you also have the option of downloading CAD or 2D drawings of the selected product version. To do so, click the "CAD" tab and select the desired file type using dropdown lists. Scope of delivery The delivery comprises: A Smartec CLD18 measuring system in the version ordered Operating Instructions BA01149C/07/EN 14 Endress+Hauser

Accessories The following are the most important accessories available at the time this documentation was issued. For accessories not listed here, please contact your service or sales office. Standard solutions Conductivity calibration solutions CLY11 Precision solutions referenced to SRM (Standard Reference Material) by NIST for qualified calibration of conductivity measuring systems in accordance with ISO 9000 CLY11-C, 1.406 ms/cm (reference temperature 25 C (77 F)), 500 ml (16.9 fl.oz) Order No. 50081904 CLY11-D, 12.64 ms/cm (reference temperature 25 C (77 F)), 500 ml (16.9 fl.oz) Order No. 50081905 CLY11-E, 107.00 ms/cm (reference temperature 25 C (77 F)), 500 ml (16.9 fl.oz) Order No. 50081906 Technical Information TI00162C Endress+Hauser 15

www.addresses.endress.com