Physical and dosimetric aspects of a multileaf collimation system used in the dynamic mode for implementing intensity modulated radiotherapy

Similar documents
IMRT Delivery System QA. IMRT Dose Delivery. Acceptance testing. Why: specific tests for IMRT? Accuracy of leaf positioning (gaps) MLC Alignment

Monte Carlo study on a new concept of a scanning photon beam system for IMRT

The Current State of EPID-Based Linear Accelerator Quality Assurance. Disclosures. Purpose of this First Talk 8/3/2017

Evaluation of a diode array for QA measurements on a helical tomotherapy unit

A Generalized Strategy for 3D Dose Verification of IMRT/VMAT Using EPID-measured Transit Images

Commissioning. Basic machine performance MLC Dose rate control Gantry speed control End-to-end tests

Using the frame averaging of as500 EPID for IMRT verification

A 2-D diode array and analysis software for verification of intensity modulated radiation therapy delivery

Isocenter and Field of View Accuracy Measurement Software for Linear Accelerator

IQM Detector Characteristics: Signal reproducibility

Development of the Use of Amorphous Silicon (ASi) Electronic Portal Imaging Devices as a Physics Tool for Routine Linear Accelerator QA

SUN NUCLEAR. EPIDose : An Overview of EPIDose and the EPIDose Process and Algorithm. corporation. Your Most Valuable QA and Dosimetry Tools

CHAPTER 2 COMMISSIONING OF KILO-VOLTAGE CONE BEAM COMPUTED TOMOGRAPHY FOR IMAGE-GUIDED RADIOTHERAPY

Confirmation, refinement, and extension of a study in intrafraction motion interplay with sliding jaw motion

Accuracy of rapid radiographic film calibration for intensity-modulated radiation therapy verification

A feasibility study of using conventional jaws to deliver IMRT plans in the treatment of prostate cancer *

PHYSICS QUESTIONNAIRE FORM

A positioning QA procedure for 2D/2D (kv/mv) and 3D/3D (CT/CBCT) image matching for radiotherapy patient setup

7/23/2014. Acknowledgements. Implementing a new digital medical accelerator. New Generation of Medical Accelerators

8/3/2017. Use of EPIDs for Non-Routine Linac QA. Disclosure. Learning Objectives. Parts of this project received support from Varian Medical System.

SCINTILLATING FIBER DOSIMETER ARRAY

Evaluation of dosimetry parameters of photons and electron beams using a linear ionization chamber array

A new approach to film dosimetry for high energy photon beams: Lateral scatter filtering

Analysis of Post-exposure Density Growth in Radiochromic Film with Respect to the Radiation Dose

Comparison of peripheral dose measurements using Ionization chamber and MOSFET detector

QUALITY CONTROL PHANTOMS FOR RADIOTHERAPY AND MEDICAL IMAGING

Dosimetric IMRT verification with a flat-panel EPID

Accurate two-dimensional IMRT verification using a back-projection EPID dosimetry method

Volumetric Modulated Arc Therapy. David Shepard Swedish Cancer Institute Seattle, WA

The evaluation of minimum detectable phantom thickness change using a scanning liquid filled ion chamber EPID dose response

Effect of slit scan imaging techniques on image quality on radiotherapy electronic portal imaging

A new approach to measure dwell position inaccuracy in HDR ring applicators quantification and corrective QA

CyberKnife Iris Beam QA using Fluence Divergence

Characterization of an in vivo diode dosimetry system for clinical use

Initial setup and subsequent temporal position monitoring using implanted RF transponders

Clinical experience with EPID dosimetry for prostate IMRT pre-treatment dose verification

Comparison of measured Varian Clinac 21EX and TrueBeam accelerator electron field characteristics

3D Diode Array Commissioning: Building Confidence in 3D QA Technology

GafChromic EBT2 and EBT3 Films for Ball Cube II Phantom

GafChromic QuiCk Phantom with EBT3P/3+P Film and FilmQA Pro for Radiation Therapy Dosimetry Applications

A Guide to Radiochromic Film Dosimetry with EBT2 and EBT3

Clinical helical tomotherapy commissioning dosimetry

QC Testing for Computed Tomography (CT) Scanner

State of the Art Film Dosimetry

SUBCHAPTER 14. THERAPEUTIC INSTALLATIONS

DOSELAB TOMOTHERAPY TG-148 QA QUICK GUIDE TG-148 RECOMMENDED TESTS 1. V.B.1.C. - Y-JAW DIVERGENCE/BEAM CENTERING

Stability of the Helical TomoTherapy Hi Art II detector for treatment beam irradiations

An Introduction to TG-142 Imaging QA Using Standard Imaging Products. Mark Wiesmeyer, PhD, DABR Technical Product Manager Standard Imaging, Inc.

International Journal of Advance Engineering and Research Development

Beam Delivery Techniques: Passive Scattering Proton Beams. Zuofeng Li and Roelf Slopsema University of Florida Proton Therapy Institute

ABSORBED DOSE DISTRIBUTIONS USING THE ISODENSITOMETRIC METHOD FOR EXPOSURES WITH FILTER EMPLOYED FOR MAMMOGRAPHIES

Characterization, Commissioning and Evaluation of Delta 4 IMRT QA System. Ram Sadagopan 1 UTMD Anderson Cancer Center Houston, TX.

A diagnostic tool for basic daily quality assurance of a tomotherapy Hi Art machine

Monica Kishore. Medical Physics Graduate Program Duke University. Approved: Jennifer O Daniel, Co-Supervisor. Fang-Fang Yin, Co-Supervisor

Historical perspective on IMRT AAPM Summer School: June My view is not the only one. What is history? William Valentine Mayneord.

Conflict Disclosure. Rotational IMRT. Arc therapy. Dynamic Arc therapy. Intensity Modulated Arc Therapy Principles and Perspectives

Aperture Based Inverse Planning AAPM Summer School 2003

8/3/2016. The EPID Strikes Back. Novel Applications for Current EPID Technology. Joerg Rottmann, PhD. Disclosures and acknowledgements

Impact of energy variation on Cone Ratio, PDD10, TMR20 10 and IMRT doses for flattening filter free (FFF) beam of TomoTherapy Hi-Art TM machines

Performance evaluation of the RITG148 + set of TomoTherapy quality assurance tools using RTQA 2 radiochromic film

ArcCHECK, ein neuartiger QS-Ansatz bei der Rotationsbestrahlung

COMPREHENSIVE TG-142 IMAGING AND MACHINE QA

C a t p h a n. T h e P h a n t o m L a b o r a t o r y. Ordering Information

ArcCHECKTM. The Ultimate 4D QA Solution. Your Most Valuable QA and Dosimetry Tools. VMAT RapidArc TomoTherapy Pinnacle 3 SmartArc Conventional IMRT

) was derived from 50% of the central axis dose created by nominal light field using geometry and mathematical methods. Leaf position (X mlc.

IMRT verification with a camera-based electronic portal imaging system

BRANDON RICE UNIVERSITY OF FLORIDA

ArcCHECK. The Ultimate 4D QA Solution. Your Most Valuable QA and Dosimetry Tools

Sensitivity study of an automated system for daily patient QA using EPID exit dose images

Be aware that there is no universal notation for the various quantities.

ISPFILMQATM STATE-OF-THE-ART RADIOTHERAPY VERIFICATION SOFTWARE. Supports all major radiotherapy technologies! FilmQA TM

Chapter 18 Optical Elements

Introduction. Chapter 16 Diagnostic Radiology. Primary radiological image. Primary radiological image

Table 1: Available X-ray Beam Energy Combinations (MV) 6 10/ Yes Yes 6 16/15 6 Yes No 6 23/18 6 Yes No 6 25/20 6 Yes No

Commissioning an Elekta Versa HD linear accelerator

Introduction of a Single Chip TLD System for Patient Dosimetry

I. Introduction.

specifications TrueBeam System

Application Note (A11)

Characterization and evaluation of an integrated quality monitoring system for online quality assurance of external beam radiation therapy

Post-irradiation colouration of Gafchromic EBT radiochromic film

Commissioning and Calibrating a Linear Accelerator State-of-the-Art in 2010

specifications TrueBeam STx System

Automated dose control in multi-slice CT. Nicholas Keat Formerly ImPACT, St George's Hospital, London

When small things matter. Small Field Dosimetry Application Guide

Mobius3D. Software based IMRT QA

Dose-response characteristics of an amorphous silicon EPID

Gantry angle determination during arc IMRT: evaluation of a simple EPID-based technique and two commercial inclinometers

QA Considerations. QA for LGK Perfexion : : Follow NRC licensing guidelines (10( CFR ) Leksell Gamma Knife Perfexion

A Fast Monolithic System for Proton Imaging. Fritz DeJongh ProtonVDA Inc October 2017

Far field intensity distributions of an OMEGA laser beam were measured with

CHAPTER 6 QUALITY ASSURANCE OF VARIAN ON-BOARD IMAGER

Rotational total skin electron irradiation with a linear accelerator

Exposure Indices and Target Values in Radiography: What Are They and How Can You Use Them?

Department of Physics, State University of New York at Buffalo, Buffalo NY, USA

- Water resistant. - Large size.

A proposed method for linear accelerator photon beam steering using EPID

Ludlum Medical Physics

Devices & Services Company

Quality control of Gamma Camera. By Dr/ Ibrahim Elsayed Saad 242 NMT

Transcription:

Physical and dosimetric aspects of a multileaf collimation system used in the dynamic mode for implementing intensity modulated radiotherapy Thomas LoSasso, a) Chen-Shou Chui, and C. Clifton Ling Department of Medical Physics, Memorial Sloan Kettering Cancer Center, 1275 York Avenue, New York, New York 10021 Received 19 May 1997; accepted for publication 3 August 1998 The use of a multileaf collimator in the dynamic mode to perform intensity modulated radiotherapy became a reality at our institution in 1995. Unlike treatment with static fields using a multileaf collimator, there are significant dosimetric issues which must be assessed before dynamic therapy can be implemented. We have performed a series of calculations and measurements to quantify head scatter for small fields, collimator transmission, and the transmission through rounded leaf ends. If not accounted for, these factors affect the delivered dose to the prostate by 5% 20% for a typical plan. Data obtained with ion chambers and radiographic film are presented for both 6 and 15 MV x-ray beams. The impact on the delivered dose of the mechanical accuracy of the multileaf collimator, achieved during leaf position calibration and maintained during dose delivery, is also discussed. 1998 American Association of Physicists in Medicine. S0094-2405 98 02510-3 Key words: multileaf collimator, intensity modulation, dosimetry, quality assurance I. INTRODUCTION Recent advances in computer application and linac technology have provided new tools for creating and delivering optimized radiation treatments. Several groups have designed computer algorithms, based on the inverse planning method, which automatically generate optimized dose distributions by modulating the photon fluence in the radiation fields. 1,2 In parallel, software for operating multileaf collimators MLCs in the dynamic mode i.e., the leaves are in motion during radiation delivery has been developed. Varian Associates, Palo Alto, CA. Converting the radiation fluence profiles to the leaf motion patterns is a third software module. 3 These capabilities, in combination, have been implemented in our center for the delivery of intensity modulated radiotherapy IMRT. MLCs are available from several manufacturers at this time. Their designs differ in the way they couple with conventional collimators and in their physical and dosimetric characteristics. MLC hardware is already familiar in many clinics, as MLCs have been used for static field therapy for several years. Several papers have discussed the dosimetry of these devices as a replacement for metal alloy blocks used for static fields. 4 7 Varian MLCs Varian Associates, Palo Alto, CA have been used clinically at our institution to deliver static field treatments since 1992 and to deliver intensity modulated field treatments with dynamic multileaf collimation DMLC since 1995. 8 In general, the dosimetry of radiation fields is similar whether shaped by the Varian MLC or metal alloy blocks. Specifically, factors such as geometrical accuracy, depth dose, transmission, and output are comparable. 4 7 In addition, the dose distributions are similar for a multifield treatment at the boundary of the planning target volume, when electron transport and patient setup uncertainty are properly considered. 7 However, based upon our experience, these studies are inadequate to predict the dosimetry of DMLC used for intensity modulated treatments. The application of MLCs to intensity modulation engenders a unique set of problems, relating to accurate and reproducible dose delivery, which need to be resolved before DMLC therapy can be successfully implemented in the clinic. Previously, we have reported the results of tests designed to examine certain mechanical aspects of DMLC and their implications on dose delivery. 9 Tests examined the stability of leaf speed, the effect of acceleration and deceleration of leaf motion, the positional accuracy of the leaves, and the effect of lateral disequilibrium on dose profiles due to the varying intensity levels between adjacent leaves. Based upon the results, the mechanical aspects of the Varian MLC were judged to be accurate and reliable. Dosimetric verification of IMRT prostate fields in phantom have shown the dose accuracy achievable with this technique. 10 It was also clear that further studies were needed in order to characterize fully the dosimetric aspects of DMLC. In this study we carefully examined the accuracy and reproducibility of the leaf gaps, keeping in mind that the accuracy of the delivered dose with DMLC is directly dependent on the precision of the gap between opposed leaves. For the MLC used in our study, this depends upon the gap calibration methods of the manufacturer. Another factor which affects the gap precision is the ability of the leaves of the MLC to move to their prescribed coordinates at the prescribed times. For this reason, a proprietary software system interactively monitors the leaf motions and interrupts radiation delivery if the leaves are not at the specified positions to within a certain tolerance. Thus, we evaluate the impact of these factors on the dose delivery. We also assessed how the MLC design influences the photon fluence produced by DMLC with the sliding window technique. 3 In this method, because of the relatively small gaps between opposed leaves and because most re- 1919 Med. Phys. 25 10, October 1998 0094-2405/98/25 10 /1919/9/$10.00 1998 Am. Assoc. Phys. Med. 1919

1920 LoSasso, Chui, and Ling: Multileaf collimation system 1920 gions are shielded by leaves much of the time, the delivered photon fluence is quite sensitive to the transmissions through the leaves and the rounded leaf ends, the leakage between the leaves, and the head scatter, factors which are of lesser significance in the use of static fields. If not properly accounted for, these influences could result in substantial errors in dose delivered. Whereas this study involves only MLC of a specific design, the concepts and findings have broad implications for intensity modulated therapy with DMLC and may be applicable to MLC of other designs as well. II. MATERIALS AND METHODS A. MLC description The Varian MLC Mark II is mounted below the conventional block collimators on a Varian C-Series accelerator. It consists of 26 pairs of tungsten alloy leaves, approximately 6 cm thick, each projecting a 1.0 cm leaf width at isocenter, with the midleaf position at about 51 cm from the source. The leaves, mounted within carriages, move in a rectilinear fashion, in the same direction as the lower block collimators. The range of travel of individual leaves is approximately 15 cm at isocenter relative to the carriage, and 20 to 16 cm about isocenter when combined with carriage motion. Maximum leaf speed is 3 cm per second at isocenter. For DMLC operation, only the leaf motion is permitted with the current version of the software ; the carriages remain fixed during dose delivery. The upper collimators define the superior and inferior borders, and the lower collimators are positioned 0.5 cm distal to the most retracted leaf position at each side of the field, similar to the procedure for static fields. In order to ensure a relatively constant penumbra at different positions in the beam, the ends of the leaves are rounded as shown in Fig. 1 a. The central 3 cm portion of each leaf end is circular with a radius of curvature of 8.0 cm. Beyond this, the leaf end is straight, and at an angle of 11.3 relative to the vertical axis. Leakage between adjacent leaves is minimized by an interlocking tongue and groove arrangement Fig. 1 b. The rectilinear leaf motion and the rounded leaf edge, in combination, introduce a nonlinear dependence of field size on leaf position. As a leaf is retracted away from the central axis, the field-defining point on the rounded end is no longer the center of the leaf, but moves away from the source. As a leaf is extended across the central axis, the field-defining point moves toward the source. In either case, the resulting field size is slightly smaller than that predicted by the linear change in leaf position. Varian provides a generic correction file to compensate for the nonlinear dependence of field size with leaf position. In this way, the digital readout is linked to the light and radiation fields, and not the physical displacements of the leaves. Since the MLC leaves are precisely machined, the parameters in this file are common to all Varian MLCs and should not require adjustment. A second file, the Positioning Calibration File, allows the user to compensate for misalignment between the MLC and the central axis. This is performed by adjusting four parameters, the skewness of the left and right leaf banks individually, the gap between opposed leaf banks, and the centering of the leaf FIG. 1. Views of the leaves a from the side and b from the front. The leaf face is rounded near the center with a radius of curvature of 8 cm. The outer portions of the face are straight, 11.3 off the vertical. banks on the central axis. It is notable that the gap adjustment, which is most critical for dynamic dose delivery, can be accurately performed using a feeler gauge for a small field size e.g., 0.1 cm. Other fields sizes are set relative to this reference field relying upon the precision of the individual leaf motor encoders. After these correction files are activated, the digital readout should agree with the projected light field. The planning and execution of intensity modulation treatment at Memorial Sloan Kettering Cancer Center MSKCC is based upon an inverse plan optimization which strives to achieve a uniform dose within the target while keeping the normal tissue doses below their specified tolerances. 10 Dose calculations are performed using the convolution of an optimized intensity distribution with a pencil beam kernel. 11 The sliding window technique, where all leaves start at one end of the field and move unidirectionally, with different speeds, to the other end, is used to deliver the intensity modulated profiles. 3 B. Mechanical accuracy Given its importance in IMRT with the sliding window technique, the accuracy in leaf position and the related gap width were studied and monitored by several techniques. First, the reproducibility of the leaf position readout is routinely checked by static light field projection at isocenter, with a spatial uncertainty of about 0.25 mm. Second, a test has been devised to check the accuracy and reproducibility of leaf positions and gap widths in DMLC mode. 9 Briefly, an intensity-modulated radiation field consisting of narrow bands 1 mm wide of exposure spaced at 2 cm intervals is

1921 LoSasso, Chui, and Ling: Multileaf collimation system 1921 produced on film. By this method, it is possible to visually detect errors greater than 0.2 mm in an individual leaf position or gap width. A third test involves periodically measuring with an ion chamber the dose for a uniform field delivered dynamically with a small, 0.5-cm-wide, sweeping gap, and normalizing it with the dose from a static 10 10 cm 2 field defined by the jaws. Since the dose can be measured precisely and since it is very sensitive to any width change of this small gap, this test is capable of detecting less than 0.1 mm deviation. The above test, performed with a cylindrical ion chamber in air at the isocenter, was also repeated for different gantry angles 0, 90, 180, and 270 to assess the effect of gravity on the performance of the MLC in the dynamic mode. Other factors which could affect the delivered dose, such as chamber position and beam output variation with gantry angle, were accounted for by repeating the measurements with the fixed reference field defined by the jaws. The accuracy of dose delivered during DMLC operation is dependent, in part, on a proprietary software module which controls the movement of the individual leaves during radiation delivery, according to the data in a DMLC file which specifies the leaf positions as a function of monitor units MU. As described in our previous publication, the DMLC file is generated, based on the intensity modulated beam profiles produced by an inverse planning algorithm, by a software module designed by Spirou and Chui at MSKCC. 3 Every 55 ms during DMLC delivery, the control software monitors the leaf positions and compares them to their prescribed positions. The beam is interrupted momentarily if any leaf position is outside a user-defined tolerance, selectable from 0.1 to 5.0 mm. In theory, this should never happen because in designing the MSKCC software which creates the DMLC file, the capability of the MLC in the dynamic mode i.e., the maximum leaf speed, was explicitly considered. In actuality, there are occasional out-of-tolerance interrupts for reasons which are not fully understood, but which could be due to MLC motor fatigue and effects of leaf acceleration and deceleration. To assess the influence of leaf position tolerance on DMLC operation, relative dose measurements were made for a range of tolerance settings at the central axis and at off-axis points using an ion chamber and film in a polystyrene phantom for several typical DMLC fields. C. MLC transmission Average transmission through the MLC was determined with ion chamber measurements in a polystyrene phantom. A parallel plate chamber 2.5-cm-diam collection volume was placed at 100 cm source to chamber distance 46 cm from the bottom of the leaves on the central axis and at 7 cm off axis. Since the chamber s diameter spanned several leaves, the measurement averages over midleaf and interleaf transmissions. Measurements were made for fields defined by the upper and lower jaws, with the MLC in the open and closed positions. The ratio of these measurements is the transmission. For the measurements on the axis, for closed MLC, in order to avoid leakage between the rounded ends, the junction between the opposed leaves was placed off axis and under the jaw. For measurements at 7 cm off axis, the lower jaws were positioned asymmetrically. Measurements on the axis were made for jaw settings of 6 6, 10 10, and 14 14 cm 2, and at depths of d max through 30 cm, for 6 and 15 MV x rays; off-axis measurements were made for 10 10 cm 2 fields only. Additional measurements were made with radiographic film to separately quantify midleaf and interleaf transmissions for both 6 and 15 MV x rays. For jaw settings of 10 10 cm 2, Kodak XV2 film was placed at 15 cm depth in a polystyrene phantom perpendicular to the central axis and irradiated to the same average optical density, 1.0, for both the open and closed MLC fields by adjusting the MU. The film was scanned with a scanning densitometer with a 1 mm aperture. Doses were derived from densities using film calibrated at 15 cm depth for a 10 10 cm 2 field. The transmission profile for the leaves was obtained by dividing the transmitted dose profile by the open field dose profile after correcting for the relative MUs delivered. D. Head scatter An intensity modulated IM field generated by the sliding-window technique is essentially a composite of many relatively small fields. Accurate dose calculations for an IM field, given the rapid change in output factor with field size for small fields, requires detailed knowledge of the extended source distribution. Thus, we measured in-air photon output and extracted the source distribution from the derivative of the measured output as a function of field size. The derivative is approximated with primary and scatter components. In our current model, the primary component is assumed to be a delta function, and the scatter is described by two conic distributions. The heights and base radii of the two cones were chosen so that the calculated and measured output factors agree to within 1% for field sizes up to 15 15 cm 2. The primary fluence at any point can then be computed by integrating the source function over the part of the source seen by the point through the aperture. For these measurements the upper and lower jaws were set to 15 cm at the isocenter. The MLC was used to define fields from 10 10 cm 2 to less than 1 1 cm 2 with square and rectangular shapes. Two methods have been used to measure output in air. In one method, a small cylindrical ion chamber was used with a 3-mm-thick brass buildup cap with a combined outer diameter and axial length of 1.2 and 1.5 cm, respectively. For field sizes of 3 cm and less in either dimension, a film technique was used to supplement the ion chamber measurements. Kodak XV2 film in ready pack was suspended in air perpendicular to the beam central axis between a pair of 0.5-mm-thick polystyrene plates. To achieve sufficient buildup for 15 MV x rays, lead disks, 6 mm in diameter and 3 mm in thickness, were placed at the field center in contact with the film jacket, upstream and downstream from the film, with slight pressure asserted by the plates. In this configuration, the density at the center of the field was converted to dose, extending the output in-air mea-

1922 LoSasso, Chui, and Ling: Multileaf collimation system 1922 surements to field sizes less than 1 cm 2. Relative outputs were measured for 6 and 15 MV x rays. E. Rounded leaf end transmission MLCs with rounded leaf ends are offered by several manufacturers. Two approaches were used to assess the contribution of the radiation transmitted through the rounded leaf ends. First, photon attenuation by the leaf end was calculated based on an accurate description of the MLC leaf geometry Fig. 1 a and the measured linear attenuation coefficient. The attenuation coefficient was derived from the measured average MLC transmission and the known leaf thickness for the manufacturer s design. Whereas an individual leaf is nominally 6 cm thick, we calculated the average thickness to be only 5.65 cm, due to guides at the top and the bottom, and the tongue and groove at the sides Fig. 1 b. Profiles were calculated for leaf ends positioned on the central axis and at 10 cm off axis. A second approach involves measuring the in-phantom dose profiles of static MLC fields with gap widths of 0 10 cm, and then integrating over the measured profile. The integration should yield the sum of a the fluence transmitted through the MLC leaves, b the fluence through the gap opening, and c the fluence transmitted through the rounded leaf ends. The sum of b and c should be proportional to the effective gap opening, i.e., the nominal gap plus an offset. The assumption that the offset is a constant, largely independent of leaf position, is a good approximation within 5 cm about the central axis, as our calculated results will show. For both 6 and 15 MV x rays, Kodak XV2 film was placed at 100 cm from the source at a depth of 5 cm in a polystyrene phantom. The upper and lower jaws were set to 5.0 and 14.0 cm, respectively, in symmetric mode. The MLC gaps, symmetric about the central axis, varied from 0 cm with opposing leaf ends in contact to 10 cm to give a 5 10 cm 2 field. A WP102 scanning film densitometer Wellhofer, West Germany was used to scan each film parallel to the leaf motion. Each film was scanned ten times, with each scan separated by 1 mm in the direction perpendicular to the leaf motion covering the 1 cm region around the central axis, to include midleaf and interleaf contributions. After conversion to dose, all ten scans for each field were integrated over the 12 cm region, symmetric about the central axis. Another measurement was made to obtain the MLC transmission for this experimental geometry. The MLC was set to zero field size and its center positioned 7.5 cm off the central axis and 0.5 cm into the shadow of the lower jaw. The exposed film was scanned and integrated over the 12 cm extent about the central axis. This transmission, calculated per unit length of leaf, is then used to subtract the contribution of MLC transmission a from the integrated dose for each of the open fields described above. A plot of the net integral dose b plus c versus gap width, should yield the offset to the nominal gap width, when extrapolated to zero dose. FIG. 2. Effect of the leaf tolerance setting on a delivered dose and b treatment time for two of the five fields, the left anterior oblique LAO and the posterior POST, used to deliver 180 cgy in our standard IMRT prostate plan. III. RESULTS A. Mechanical accuracy A number of characteristics are important in the use of a MLC in the dynamic mode: mechanical precision, stability, and electromechanical reliability. Routine quality assurance has been performed with the Varian MLCs at our institution for several years before the commencement of DMLC operation. These tests have indicated that the calibration of the MLC is very stable. Based upon visual examination of the light fields, the reproducibility of leaf positioning is, generally, 0.25 mm. Furthermore, the operation of the MLCs have been very reliable, i.e., essentially trouble free operation with an average use factor for the MLCs on three machines of approximately 70%. During the development of the DMLC program at MSKCC, a number of acceptance and QA tests were devised using the dynamic functions. 9 One QA test which is routinely used consists of narrow bands of radiation produced on film by a start and stop leaf pattern in dynamic mode. The test results confirmed that the precision of leaf position is

1923 LoSasso, Chui, and Ling: Multileaf collimation system 1923 FIG. 4. The midleaf and interleaf transmissions measured at isocenter in phantom with radiographic film for 6 and 15 MV x rays. The transmission is the ratio of the MLC blocked field profile to the open field profile for a 10 10 cm 2 field defined by the jaws and a depth of 15 cm. mm. Further testing revealed that this variation is not due to error in gap width, but to beam instability caused by the increase in beam hold off incidence at the tighter tolerance setting. FIG. 3. The average of midleaf and interleaf transmissions measured at isocenter in phantom with a parallel plate ion chamber, 2.5 cm in diameter for a 6 MV x rays and b 15 MV x rays. The transmission is the ratio of the central axis dose for the MLC blocked field to the open field for the collimator settings and depths shown. 0.25 mm or better for both Varian MLCs used for DMLC. For one of these MLCs, more sensitive measurements of the gap width, based upon the output measured with an ion chamber for a small scanned field, were made at different gantry angles. The results showed that the precision with which the gap of the Varian MLC can be set and maintained has been found to be better than 0.1 mm, independent of gantry angle. Measurements using the same field at off-axis points are consistent with the reduced off-axis leaf end transmission presented in Sec. III E. In the test of the effect of the tolerance setting, as the tolerance was decreased from 5.0 down to 0.1 mm, the beam hold off invoked by the software becomes more frequent, indicating that the leaves are at times unable to reach their prescribed positions, and that such incidences are more likely at tighter tolerance settings. The effect of tolerance on delivered dose is shown in Fig. 2 a for two of five fields used to deliver 1.8 Gy in our standard five field intensity modulated prostate plan. 12 The maximum dose variation is less than 1% for each field, occurring for tolerances between 0.1 and 2.0 B. Treatment time As shown in Fig. 2 b, the treatment time is essentially constant for tolerances greater than 2 mm. As the tolerances are reduced, the beam was increasingly held off, thereby lengthening the treatment time. For the two examples shown the treatment time more than doubles, as the tolerance is reduced from 2.0 to 0.1 mm. C. MLC transmission The average of midleaf transmission and interleaf leakage is shown in Figs. 3 a and 3 b for 6 and 15 MV x rays, respectively, normalized to the output of the open field. The transmission, over a range of clinically useful field sizes 6 6 to 15 15 cm 2 and depths d max to 20 cm, is about 2.0%. The transmission increases with jaw opening for both energies; the exact relationship is not known, although increased scatter from the MLC is likely. The transmission for 6 MV x rays also increases with depth of measurement, while for 15 MV x rays it is almost constant. This may be due to beam hardening for 6 MV x rays within the MLC resulting in a more penetrating transmitted beam than the open beam. The increase in pair production within the MLC for the higher energies in the 15 MV beam may have offset the beam hardening effect, as there is little change with depth. The transmission at 7 cm off axis is essentially the same as it is on the central axis for each energy. Transmission profiles obtained for a 10 10 cm 2 field at 15 cm depth in polystyrene using film are displayed in Fig. 4. The average transmissions were 2.0% and 2.1% for 6 and 15 MV x rays in good agreement with the values found with the ion chamber. The range of transmission over the central por-

1924 LoSasso, Chui, and Ling: Multileaf collimation system 1924 FIG. 6. Calculated primary beam transmission through the rounded leaf end. The central axis and off-axis transmissions are based upon the geometry of the leaf and the measured effective attenuation coefficient for the MLC. The area under the central axis transmission curve less the 2% corresponding to the transmission for full leaf thickness is equivalent to the rectangular area from 0 to 1 mm, indicating that the transmission due to the rounded leaf edges may be approximated in a dynamic treatment as an offset of the leaves by 1 mm. Varian MLC is located below the lower jaws, and therefore, the field size dependence is less sensitive to the distributed source and head scatter. FIG. 5. Output measurements in air at the isocenter with an ion chamber and with radiographic film for a 6 MV x rays and b 15 MV x rays. The width W and length L are dimensions defined by the MLC. For all measurements, the jaws are set to 15 15 cm 2. tion of the MLC was from 1.7% at midleaf to 2.7% between leaves. The variation in transmission through different leaves is less than 0.1%, while interleaf values vary by as much as 0.5%. For purposes of treatment planning calculations, 2% is used for static transmission to points outside the field. However, for DMLC fields using the sliding window technique, each point in the field is shielded for a longer time by the leaves relative to the time it spends under the window, resulting in an effective transmission of 4% 6% of the total fluence. D. Head scatter Relative output measurements were made in air using an ion chamber, and radiographic film to extend the measurements to fields less than 1 cm 2. The results are shown in Figs. 5 a and 5 b for fields of various sizes and shapes. Relative to that of a 10 10 cm 2 field, the output for a 3 3 cm 2 field is reduced by 2%, and for a 1 0.6 cm 2 field by 5% 6%. This reduction is small, about half that reported by others when fields were defined by the jaws for a similar machine. 13 This difference is probably due to the fact that the E. Leaf end transmission Based upon the measured average transmission through the MLC of 2.0% and the calculated effective leaf thickness of 5.65 cm, the linear attenuation coefficient is taken to be 0.692 cm 1. Using this value and assuming a point source, the photon transmission profiles are calculated for the leaf edge positioned at the central axis and at 10 cm off axis Fig. 6. The primary transmission under the leaf falls off to 50% at 0.3 mm projected distance at isocenter, to 20% at 1.5 mm, and to 4% at 6 mm. Beyond 8 mm the transmission falls to 2%. The falloff is slightly faster when the leaf edge is off axis. The effect of the transmission through the rounded leaf on the dose delivered by DMLC can be approximated by a 1 mm offset applied to the leaf position. The offset is represented in Fig. 6 by the rectangle, which is equal to the area under the transmission curve for the leaf end, except for the 2% MLC transmission. Thus, the added transmission due to the rounded leaf ends is equivalent to enlarging the field size of a focused collimator by 2.0 mm 1.0 mm for each opposed leaf. As an example of the significance of this, the transmission through the leaf ends will contribute 10% of the total dose delivered by a nominal 2.0 cm gap moving at a constant speed across the field, since the effective gap is 2.2 cm. Although the offset varies slightly with leaf position at 10 cm off axis, it is 0.8 mm, a single value is a good approximation. The results of film dosimetry, i.e., integrating cross-field dose profiles of static MLC fields of various widths, are

1925 LoSasso, Chui, and Ling: Multileaf collimation system 1925 FIG. 7. Integrated dose vs MLC gap width measured in phantom with radiographic film for 6 and 15 MV x rays. The lines are linear fits to the data using least-squares regression. Extrapolation to zero integral dose determines the effective gap offset. The uncertainties are the standard errors of the data. given in Fig. 7, with the relative integral dose plotted against the nominal field width. Regression analysis of the data yielded offsets of 2.0 and 1.7 mm for 15 and 6 MV, respectively, in excellent agreement with that derived from calculated transmission curves of the MLC leaf end. IV. DISCUSSION In the sliding window technique of DMLC application, the delivered dose is directly related to the gap between opposed leaves as they sweep across the field. To appreciate the importance of gap width accuracy in DMLC, we related error in delivered dose to error in gap width for different nominal window widths. As shown in Fig. 8, the dose error is large for small gap width and large gap error. However, as presented earlier, the MLC position precision is better than 0.1 mm. In addition, for clinical intensity modulated treatments at this time, relatively little dose is delivered with gaps FIG. 8. Calculated results relating the error in the dose delivered to the error in the gap for a range of gap widths. FIG. 9. Comparative influences of MLC transmission, rounded edge transmission, and head scatter on the target dose. 1 cm. Therefore, the dose error due to gap error is not an issue. It is noteworthy that subtle changes in individual leaf positions may be caused by gravity, drag due to friction, or simply misalignment with the central axis; however, each of these factors tends to shift both leaves in the same direction, with the result that the gap width is unchanged. Figure 2 b indicated that a leaf position tolerance setting of 0.1 mm could increase the treatment time of a typical prostate field from about 30 to 80 s. On the other hand, accuracy in delivered dose either on the central axis Fig. 2 a or at off-axis points data not shown is hardly affected by the choice of tolerance setting. Therefore, for reasons of treatment time efficiency, dose accuracy, and potential hardware problems with the leaves during DMLC operation, 2 mm was chosen as the leaf position tolerance. Thus, under normal operating conditions the leaf position tolerance is nearly passive; however, it acts as a safety interlock should a leaf behave abnormally, e.g., becoming stuck during irradiation. The implications of MLC transmission and transmission through the rounded leaf end are more serious for DMLC than for static MLC treatments. The transmissions through the leaves and the leaf ends contribute to the dose throughout the target, not just near or outside the field boundary as for static MLC fields. Furthermore, the relatively small fields associated with the sliding window technique of DMLC treatments tend to amplify the importance of the transmitted photons. Typically 1- to 4-cm-wide gaps are used for the treatment of the prostate, which are uncommon for static field treatments. The idealized curves in Fig. 9 indicate the magnitudes of these effects for a typical prostate treatment with 15 MV. The curves are based upon the assumption that a gap of fixed width moves with constant speed across the target. The fractional contribution of the rounded leaf end is approximately the ratio of the effective gap offset, 2 mm, to the gap width. The dose transmitted through the MLC, will typically vary from 3% 6% of the total target dose and is related to the target width 7 cm for this example divided by the gap width. For DMLC in the treatment of prostate can-

1926 LoSasso, Chui, and Ling: Multileaf collimation system 1926 FIG. 10. Comparison of calculated and measured dose distributions with corrections for MLC transmissions and head scatter. The distributions are generated using the five intensity modulated fields for a particular prostate patient, superimposed upon a homogeneous cylindrical phantom. cer, these combined transmissions will add 5% 20% of the target dose for 1- to 4-cm-wide gaps. For static fields, on the other hand, the transmissions produce only slight broadening of the penumbra and 2% primary fluence outside the field, both of which can be managed as for cerrobend blocking. For a dynamic field, the influence of the MLC on output is much less than the curves in Fig. 5 suggest. As the leaf window moves across the target, each point in the target sees overlapping portions of the extended source at different times. As a result, the source distribution associated with a dynamic field as a whole is nearly the same as that for a static field with the same outer shape. It follows then, as for static fields for this MLC design, that overall head scatter for a dynamic field is essentially independent of the MLC as shown in Fig. 9; head scatter is defined mainly by the jaw setting. Actually, due to the nonuniform intensity patterns within a dynamic field, the contribution of head scatter will vary slightly within the field. In our scheme, the head scatter and leaf transmissions are accounted for during the conversion from optimized intensity profiles to leaf motion patterns. The inclusion of MLC transmission is explained in detail in a previous publication. 3 The effects of head scatter and leaf end transmission are handled in the following manner. First, without considering these effects, the leaf motions are generated from the intended intensity distribution. Then the head scatter contribution at each point, calculated by backprojecting the field opening, defined by MLC and jaws, onto the source plane, and the leaf end transmission profile are convolved with the leaf motion patterns to produce a delivered intensity distribution. Leaf motions are then generated from a third working intensity distribution, which is modified until the delivered and intended distributions agree to within 1% 3 iterations. That the corrections are properly applied is evidenced by routine comparisons of measured and calculated dose. In one verification study, shown in Fig. 10, the five intensity modulated fields for a particular prostate patient were used to expose radiographic film parallel to the central axis in a cylindrical polystyrene phantom, 25 cm in diameter. The film densities were converted to dose by calibrating the film at 12.5 cm depth. Dose calculations were then performed for the five fields in a similar phantom geometry using the threedimensional treatment planning system. The agreement within and around the target region is very good. Finally, something should be said about tongue and groove effects. These effects have been evaluated for the clinical fields used for the intensity modulated prostate treatments of our patients. They result in a reduction in the dose between leaves, which tends to counteract the interleaf leakage. For all the DMLC patients treated to date, film dosimetry has been performed in a flat homogeneous phantom for each field and the measured dose distributions are compared with similarly calculated dose distributions. For the majority of the fields, the measured dose variation at the interleaf regions is not detectable. The maximum variation observed was a 5% dose reduction along a line 3 cm in length with a full width at half-maximum of 2 mm for one of five prostate fields for an intensity modulated plan. Obviously, the severity is lessened when the composite dose for all five fields is considered. In addition, a new optimization strategy has been introduced which should alleviate tongue and groove effects. 14 This will be tested in conjunction with our inverse planning optimization algorithm to determine its benefits. V. SUMMARY Commissioning a MLC for DMLC application requires considerable effort. This paper addresses the important issues confronting the Varian MLC user. The results presented above can be divided into two categories. Those which are related to the alignments and calibration of the MLC need to be monitored carefully by the user to establish their accuracy and precision, since the DMLC application is much less tolerant of misalignment and poor calibration than conventional static field treatments. The second category comprises unavoidable factors which affect the output and which must be accurately quantified and accounted for during the plan optimization or dose delivery process. These include head scatter and transmissions associated with midleaf, interleaf, and leaf end components of the MLC. ACKNOWLEDGMENTS The authors wish to thank Dr. Wendell Lutz, Dr. Radhe Mohan, Dr. Spiridon Spirou, and Dr. Jorg Stein for helpful discussions, and Dr. Chandra Burman for generating the calculated dose distribution in Fig. 10. This work was supported in part by Grant No. CA 59017 from the National Cancer Institute, Department of Health and Human Resources, Bethesda, MD.

1927 LoSasso, Chui, and Ling: Multileaf collimation system 1927 a Electronic mail: losassot@mskcc.org 1 A. Brahme, Optimization of stationary and moving beam radiation therapy techniques, Radiother. Oncol. 12, 129 140 1988. 2 T. Bortfeld, J. Bukelbach, R. Boesecke, and W. Schlegel, Methods of image reconstruction from projections applied to conformation radiotherapy, Phys. Med. Biol. 35, 1423 1434 1990. 3 S. Spirou and C. S. Chui, Generation of arbitrary intensity profiles by dynamic jaws or multileaf collimators, Med. Phys. 21, 1031 1042 1994. 4 A. L. Boyer, T. G. Ochran, C. E. Nyerick, T. J. Waldron, and C. J. Huntzinger, Clinical dosimetry for implementation of a multileaf collimator, Med. Phys. 19, 1255 1261 1992. 5 J. M. Galvin, A. R. Smith, and B. Lally, Characterization of a multileaf collimator system, Int. J. Radiat. Oncol., Biol., Phys. 25, 181 192 1993. 6 T. LoSasso, and G. J. Kutcher, Multileaf collimation versus alloy blocks: Analysis of geometric accuracy, Int. J. Radiat. Oncol., Biol., Phys. 32, 499 506 1995. 7 T. LoSasso, C. S. Chui, G. J. Kutcher, S. A. Leibel, Z. Fuks, and C. C. Ling, The use of a multi-leaf collimator for conformal radiotherapy of carcinomas of the prostate and nasopharynx, Int. J. Radiat. Oncol., Biol., Phys. 25, 161 170 1993. 8 C. C. Ling et al., Conformal radiation treatment of prostate cancer using inversely-planned intensity-modulated photon beams produced with dynamic multileaf collimation, Int. J. Radiat. Oncol., Biol., Phys. 35, 721 730 1996. 9 C. S. Chui, S. Spirou, and T. LoSasso, Testing of dynamic multileaf collimation, Med. Phys. 23, 635 641 1996. 10 X. Wang, S. Spirou, T. LoSasso, J. Stein, C. S. Chui, and R. Mohan, Dosimetric verification of intensity-modulated fields, Med. Phys. 23, 317 327 1996. 11 C. S. Chui, T. LoSasso, and S. Spirou, Dose calculation for photon beams with intensity modulation generated by dynamic jaw or multileaf collimations, Med. Phys. 21, 1237 1244 1994. 12 C. Burman et al., Planning, delivery and quality assurance of intensity modulated radiotherapy using a dynamic multileaf collimator: A strategy for large-scale implementation for the treatment of carcinoma of the prostate, Int. J. Radiat. Oncol., Biol., Phys. in press. 13 T. C. Zhu and B. E. Bjarngard, The head-scatter factor for small field sizes, Med. Phys. 21, 65 68 1994. 14 J. P. C Santvoort and B. J. M. Heijmen, Dynamic multileaf collimation without tongue and groove underdosage effects, Phys. Med. Biol. 41, 2091 2105 1996.