Evaluation of Filter and Amplifier Transmission Characteristics

Similar documents
Signal Generators for Anritsu RF and Microwave Handheld Instruments

DSA-815 Demo Guide. Solution: The DSA 800 series of spectrum analyzers are packed with features.

DMR Tx Test Solution. Signal Analyzer MS2830A. Reference Specifications

TETRA Tx Test Solution

ME1000 RF Circuit Design. Lab 1. Calibration with Spectrum Analyzer

RADIO COMMUNICATIONS TEST INSTRUMENTS

Demo / Application Guide for DSA815(-TG) / DSA1000 Series

Signal Analyzer. MS2830A Microwave

Spectrum Analyzers 2680 Series Features & benefits

Impedance 50 (75 connectors via adapters)

Agilent X-Series Signal Analyzer This manual provides documentation for the following X-Series Analyzer: CXA Signal Analyzer N9000A

Swept Return Loss & VSWR Antenna Measurements using the Eagle Technologies RF Bridge

Application Note: Swept Return Loss & VSWR Antenna Measurements using the Eagle Technologies RF Bridge

Signal Analyzer MS2840A

Measuring ACLR/CACLR for Non-contiguous Carrier Aggregation

ENGINEERING COMMITTEE Interface Practices Subcommittee AMERICAN NATIONAL STANDARD ANSI/SCTE

Product Brochure. MS2690A/MS2691A/MS2692A Signal Analyzer MS2690A: 50 Hz to 6.0 GHz MS2691A: 50 Hz to 13.5 GHz MS2692A: 50 Hz to 26.

Keysight Technologies FieldFox Handheld Analyzers

Pulsed S-Parameter Measurements using the ZVA network Analyzer

External Source Control

Agilent Technologies PSA Series Spectrum Analyzers Test and Adjustment Software

EXHIBIT 7: MEASUREMENT PROCEDURES Pursuant 47 CFR 2.947

FREQUENCY SYNTHESIZERS, SIGNAL GENERATORS

Development of MT8870A Universal Wireless Test Set

PXIe Contents CALIBRATION PROCEDURE

ECE 2111 Signals and Systems Spring 2009, UMD Experiment 3: The Spectrum Analyzer

Signal Analysis Measurement Guide

Product Brochure. MS2690A/MS2691A/MS2692A Signal Analyzer MS2690A: 50 Hz to 6.0 GHz MS2691A: 50 Hz to 13.5 GHz MS2692A: 50 Hz to 26.

Transcom Instruments. Product Brochure TRANSCOM INSTRUMENTS. Product Brochure

Signal Analyzer MS2830A

LA Techniques Ltd

Signal Analyzer MS2840A. dbc/hz. dbc/hz (meas.)

Agilent PSA Series Spectrum Analyzers Self-Guided Demonstration for Spectrum Analysis

Keysight Technologies N9320B RF Spectrum Analyzer

Measurement of Digital Transmission Systems Operating under Section March 23, 2005

Agilent CSA Spectrum Analyzer

Agilent N9320B RF Spectrum Analyzer

Agilent PSA Series Spectrum Analyzers Self-Guided Demonstration for Phase Noise Measurements

Transcom Instruments. Product Brochure TRANSCOM INSTRUMENTS. Product Brochure

Digital Wireless Measurement Solution

PARAMETER CONDITIONS TYPICAL PERFORMANCE Operating Supply Voltage 3.1V to 3.5V Supply Current V CC = 3.3V, LO applied 152mA

Spectrum Analyzers U3741/3751. Compact Design with High Performance Pioneering 3 GHz/8 GHz Spectrum Analyzers are Now Available!

Product Brochure. MS2690A/MS2691A/MS2692A Signal Analyzer MS2690A: 50 Hz to 6.0 GHz MS2691A: 50 Hz to 13.5 GHz MS2692A: 50 Hz to 26.

PLANAR 814/1. Vector Network Analyzer

DMR Rx Test Solution. Signal Analyzer MS2830A. Reference Specifications

G6M VSG Module and A6M VSA Module

Keysight FieldFox Handheld Analyzers

R3131A. Advanced Test Equipment Rentals ATEC (2832) A personal spectrum analyzer for use in diverse applications R3131A

R3477. Ideal for mobile communication applications including base stations and handsets, from the development stage to production and installation

ENGINEERING COMMITTEE Interface Practices Subcommittee AMERICAN NATIONAL STANDARD ANSI/SCTE Measurement Procedure for Noise Power Ratio

Attenuators, Couplers and Filters

Agilent PSA Series Spectrum Analyzers Noise Figure Measurements Personality

Transcom Instruments. Product Brochure TRANSCOM INSTRUMENTS. Product Brochure

Agilent PN 4395/96-1 How to Measure Noise Accurately Using the Agilent Combination Analyzers

Agilent PSA Series Spectrum Analyzers Self-Guided Demonstration for Spectrum Analysis

Agilent AN How to Characterize CATV Amplifiers Effectively

Frequency and Time Domain Representation of Sinusoidal Signals

FieldFox Handheld Analyzers

Keysight Technologies FieldFox Handheld Analyzers

8 Hints for Better Spectrum Analysis. Application Note

Agilent X-Series Signal Analyzer

A Guide to Calibrating Your Spectrum Analyzer

Keysight Technologies FieldFox Handheld Analyzers

Keysight Technologies FieldFox Handheld Analyzers

SIGNAL GENERATORS. MG3633A 10 khz to 2700 MHz SYNTHESIZED SIGNAL GENERATOR GPIB

Understanding RF and Microwave Analysis Basics

Siglent Technologies SSA3021X Spectrum Analyzer and TG-SSA3000X Tracking Generator Reviewed by Phil Salas AD5X

TRANSCOM Manufacturing & Education

9 Hints for Making Better Measurements Using RF Signal Generators. Application Note 1390

Network Analyzers. R3765G/3767G Series. R3765G/67G Series Network Analyzers. 300 khz to 3.8 GHz/300 khz to 8 GHz

Migrating 4195A to E5061B LF-RF Network Analyzer. April 2010 Agilent Technologies

R3132/3162. Advanced Test Equipment Rentals ATEC (2832) R3132 : 9kHz to 3GHz R3162 : 9kHz to 8GHz

Handheld Spectrum Analyzer R&S FSH khz to 3 GHz

Configuration of PNA-X, NVNA and X parameters

SPECTRUM ANALYZERS. MS2602A 100 Hz to 8.5 GHz GPIB SPECTRUM ANALYZER

ECE 4670 Spring 2014 Lab 1 Linear System Characteristics

Noise by the Numbers

Agilent X-Series Signal Analyzer

Development of Signal Analyzer MS2840A with Built-in Low Phase-Noise Synthesizer

7. Transmitter Radiated Spurious Emissions and Conducted Spurious Emission

DSA800. No.1 RIGOL TECHNOLOGIES, INC.

Agilent PSA Series Spectrum Analyzers Noise Figure Measurements Personality

Preliminary Users Manual for the Self Contained Return Loss and Cable Fault Test Set with Amplified Wideband Noise Source Copyright 2001 Bryan K.

AMERICAN NATIONAL STANDARD

Vector Network Analyzers T - Series

SPECTRUM ANALYZERS. MS710C/D/E/F 10 khz to 23 GHz (18 to 140 GHz) GPIB SPECTRUM ANALYZER

Analysis of RF transceivers used in automotive

PXIe Contents CALIBRATION PROCEDURE. Reconfigurable 6 GHz RF Vector Signal Transceiver with 200 MHz Bandwidth

2015 Interference 101. Robin Jackman Application Engineer

Spectrum Analyzers U3771/3772. Handling frequencies of up to 43 GHz!! Our new microwave spectrum analyzer, ideal for field use, is now available.

CAA-100A Cable & Antenna Analyzer + Spectrum Analyzer

PLANAR S5048 and TR5048

Introduction to RF measurements and instrumentation. Daniel Valuch, CERN BE/RF,

Keysight X-Series Signal Analyzers

2.9GHz SPECTRUM ANALYZER

Agilent N9923A FieldFox RF Vector Network Analyzer 2 MHz to 4/6 GHz. Data Sheet

Agilent 8920A RF Communications Test Set Product Overview

Measurements on RF and AF Filters with Rohde & Schwarz Value Instruments

GET10B Radar Measurement Basics- Spectrum Analysis of Pulsed Signals. Copyright 2001 Agilent Technologies, Inc.

TECHNICAL SPECIFICATION FOR RF (TEST &MEASUREMENT) DEVICE

Transcription:

Product Introduction Evaluation of Filter and Amplifier Transmission Characteristics - Internal Signal Generator Control Function Option (Functions equivalent to tracking generator) Signal Analyzer MS2830A

- Contents - 1. Wide Applications in R&D, Manufacturing and I&M 2. Filter and Amplifier Transmission Characteristics 2-1. Introduction to Main Functions 2-2. Frequency and Output Level Ranges 3. Required Option Configuration 3-1. New Purchase 3-2. Retrofit 4. Functions and Operation Methods 4-1. Common Settings 4-2. Normalize Function 4-3. N db Bandwidth Measurement Function 4-4. Correction Value Capture Function 4-5. Trace Math Function 4-6. Reflection Characteristics Measurement 2

1. Wide Applications in R&D, Manufacturing and I&M The Signal Analyzer MS2830A integrates all the functions required for evaluating the TRx characteristics of various wireless technologies to support a wider R&D, production, and I&M application range. Spectrum Analyzer Digital Modulation Analysis Vector Signal Generator Frequency Counter Analog Radio Tester Cost Reduction All-in-One Measurement! Low Energy Small Footprint 3

1. Wide Applications in R&D, Manufacturing and I&M The spectrum analyzer configuration can be expanded to support required functions. Some measurement functions are listed bellow; refer to the product brochure for more details. Spectrum Analyzer Signal Analyzer Modulation Analysis Signal Generator MS2830A Measurement functions Standard Option Spurious Emission - Frequency Counter - Adjacent Channel Leakage Power - Internal Signal Generator Control Function - FFT(Fast Fourier Transformed) Spectrum Trace - Capture & Playback (Outputs captured RF signal from VSG) - Digital Analysis (π/4dqpsk, 4FSK, etc.,) - Analog Analysis (FM, ΦM, AM) - LTE, W-CDMA, GSM Analysis, etc., - Vector Signal Generator (Pre-installed waveform patterns for each communications systems) - Analog Signal Generator (FM, ΦM, AM) - BER - 4

- Contents - 1. Wide Applications in R&D, Manufacturing and I&M 2. Filter and Amplifier Transmission Characteristics 2-1. Introduction to Main Functions 2-2. Frequency and Output Level Ranges 3. Required Option Configuration 3-1. New Purchase 3-2. Retrofit 4. Functions and Operation Methods 4-1. Common Settings 4-2. Normalize Function 4-3. N db Bandwidth Measurement Function 4-4. Correction Value Capture Function 4-5. Trace Math Function 4-6. Reflection Characteristics Measurement 5

2. Filter and Amplifier Transmission Characteristics The Internal signal generator control function operates in conjunction with the spectrum analyzer (SPA) function and built-in signal generator (SG) option to measure the transmission characteristics of filters, amplifiers, etc. Devices Under Test Filters Attenuators Amplifiers Isolators Switches : SG (Output) SPA (Input) DUT Measure Both Passive and Active Devices The DUT input signal source has a frequency range of 100 khz to 6 GHz, an output level range of -136 to +15 dbm, a step resolution of 0.01 db, and a level accuracy of ±0.5 db to measure both passive and active devices using the built-in high-performance SG. Accurate Frequency Characteristics The SPA function displays the measured frequency characteristics results with an excellent linearity error of just ±0.07 db to display the frequency characteristics of bandpass filters, etc., accurately. 6

2-1. Introduction to Main Functions Normalize Function Calibrate frequency characteristics of cables, etc., connected at measurement. N db Bandwidth Measurement Function Measure frequency bandwidth from the peak marker point to any amplitude (N db) to measure cutoff frequency of bandpass filter. Correction Value Capture Function Capture the measured path insertion loss correction value as a.csv file, and load the.csv file into the MS2830A. 7

2-2. Frequency and Output Level Ranges The frequency and output level ranges differ according to the built-in SG type. For details, refer to the product brochure. MS2830A-052 Internal Signal Generator Control Function *1 SG Type Frequency Range Output Level Range Vector SG Analog SG Vector SG 40 40 250 khz ~ 3.6 GHz or 250 khz ~ 6 GHz 136 136 100 khz ~ 3.6 GHz 100 khz ~ 3.6 GHz + or Analog SG 100 khz ~ 6 GHz 127 127 136 136 *1 : Requires any one of MS2830A-020, or -021, or -088 options. ~ ~ or ~ ~ ~ ~ ~ ~ +2 +20 3 +15 3 +15 3 +15 dbm dbm dbm dbm dbm dbm dbm dbm ( 25 MHz) (> 25 MHz) ( 25 MHz) (> 25 MHz) ( 25 MHz) (> 25 MHz) ( 25 MHz) (> 25 MHz) 8

- Contents - 1. Wide Applications in R&D, Manufacturing and I&M 2. Filter and Amplifier Transmission Characteristics 2-1. Introduction to Main Functions 2-2. Frequency and Output Level Ranges 3. Required Option Configuration 3-1. New Purchase 3-2. Retrofit 4. Functions and Operation Methods 4-1. Common Settings 4-2. Normalize Function 4-3. N db Bandwidth Measurement Function 4-4. Correction Value Capture Function 4-5. Trace Math Function 4-6. Reflection Characteristics Measurement 9

3-1. At New Purchase This slide lists the required configuration at new purchase. There are also other various options. For details, refer to the configuration guide to select additional options in easy steps. When Built-in Vector SG : 1 2 Model Name Note MS2830A-040 MS2830A-041 MS2830A-043 MS2830A-020 MS2830A-021 3 MS2830A-022 3.6 GHz Signal Analyzer 6GHz Signal Analyzer 13.5 GHz Signal Analyzer 3.6 GHz Vector Signal Generator 6 GHz Vector Signal Generator Low Power Extension for Vector Signal Generator 4 MS2830A-052 Internal Signal Generator Control Function Choose. Choose any one. Choose any one. When either Analog SG or Both Vector SG and Analog SG : 1 2 Expands lower output level from standard 40 dbm to 136 dbm. Choose if necessary. Model Name Note MS2830A-040 MS2830A-041 MS2830A-088 MS2830A-029 3.6 GHz Signal Analyzer 6GHz Signal Analyzer 3.6 GHz Analog Signal Generator Analog Function Extension for Vector Signal Generator 3 MS2830A-052 Internal Signal Generator Control Function Choose. Choose any one. Choose any one. MS2830A-088: Requires separate MS2830A-066, A0086, and MX269018A selection. MS2830A-029: Option for adding analog SG to vector SG. Requires separate vector SG (MS2830A-020 or -021), MS2830A-022, MS2830A-066, A0086, and MX269018A 10

3-2. At Retrofit This slide explains retrofitting the internal Signal Generator Control Function option when the Signal Analyzer MS2830A has been purchased previously. Choose by confirming the actual equipment using the following steps. Step 1. Confirm option supporting retrofit : The option can be installed in the 3.6, 6, and 13.5 GHz models. Confirm the frequency printed at the top right of the front panel. Addition *1 Frequency 9kHz-3.6GHz 9kHz-6GHz 9kHz-13.5GHz - 9kHz-26.5GHz - 9kHz-43GHz *1 : The SG cannot be retrofitted if the MS2830A-066 Low Phase Noise option has been installed. Step 2. Confirm whether built SG installed or not : Check for the presence of an N-type RF connector at the bottom left of the front panel. SG not installed Go to Step 3. SG installed Go to Step 4. 11

3-2. At Retrofit Step 3. Note on Built-in SG Retrofit : The MS2830A must be returned to the factory for upgrading. When Installing Vector SG Model Name Note 1 MS2830A-120 3.6 GHz Vector Signal Generator Retrofit Choose any one. MS2830A-121 6 GHz Vector Signal Generator Retrofit 2 MS2830A-122 Low Power Extension for Vector Signal Generator Retrofit Expands lower output level from standard 40 dbm to 136 dbm. Choose if necessary. 3 Z1345A Installation kit Choose. When installing either Analog SG or Both Vector SG and Analog SG : * The analog SG cannot be installed in the 13.5 GHz model (MS2830A-043). Model Name Note 1 MS2830A-188 3.6 GHz Analog Signal Generator Retrofit Choose. Requires MS2830A-066, A0086, or MX269018A. 2 MS2830A-189 Vector Function Extension for Analog Signal Generator Retrofit Option for adding analog SG to vector SG. Choose if necessary. 3 Z1345A Installation kit Choose. Step 4. Note on Installing this Option : The license is shipped on an installation DVD. This function can be used by installing the license in the MS2830A. This upgrade does not require return to the factory. Model Name Note 1 MS2830A-352 Internal Signal Generator Control Function User-Installable Choose. 2 Z1345A Installation Kit Choose.

- Contents - 1. Wide Applications in R&D, Manufacturing and I&M 2. Filter and Amplifier Transmission Characteristics 2-1. Introduction to Main Functions 2-2. Frequency and Output Level Ranges 3. Required Option Configuration 3-1. New Purchase 3-2. Retrofit 4. Functions and Operation Methods 4-1. Common Settings 4-2. Normalize Function 4-3. N db Bandwidth Measurement Function 4-4. Correction Value Capture Function 4-5. Trace Math Function 4-6. Reflection Characteristics Measurement 13

4-1. Common Settings This slide introduces the functions and operation using measurement of the frequency characteristics of a 200-MHz bandpass filter as an example. The first step is setting the main parameters common to each function. Connecting BPF and setting each parameter BPF 200MHz BPF Switch to Spectrum Analyzer mode. > [Application Switch] > [Spectrum Analyzer] Initialize parameters. > [Preset] > [F1:Preset] Set Signal Generator Tracking Function to On. > [Measure] >[ (2of2)] > [F6:SG Control] > [F1:SG Control] = On Set Output Level ( 10 dbm as example) and set to On. > [F2:SG Output Level] = [-10] [F1:dBm] > [F3:SG Output] = On *At output of 3 dbm or more, set Start Frequency to higher than 25 MHz. Set Frequency (200 MHz as example). > [Frequency] > [F1:Center] = [200][F2:MHz] *Both Frequency Start/Stop can be set. Set Span (5 MHz as example). > [Span] > [5][F2:MHz] *Can be set using Up/Down keys. Set RBW (30 khz as example) > [BW] > [F2:RBW Value] > [30][F3:kHz] *Can be set using Up/Down keys. *The sweep speed changes according to the RBW setting. Set Detection mode (RMS as example). > [Trace] > [F8:Detection] > [F2:RMS] To next slide 14

4-1. Common Settings Adjust Reference Level ( 10 dbm as example). > [Amplitude] > [F1:Reference Level] > [-10][dBm] *Can be set using Up/Down keys. Set Trace Points (201 as example). > [Time/Sweep] >[F4:Trace Points] > (201)[F7:Set] *Can be set using Up/Down keys. *The sweep speed varies with the set number of Trace Points. Set Frequency Switching Speed (Fast Tuning as example). > [Frequency] > [F4:Switching Speed] > [F1:Fast Tuning] Fast Tuning Normal Increases measurement speed by speeding up frequency switching. Gives good SPA phase noise performance. Use when monitoring phase noise performance of narrowband filters, etc. Setting SG output to OFF when changing DUT Set SG output level to OFF. > [Measure] >[ (2of2)] > [F6:SG Control] > [F3:SG Output] = Off *Can set using SG On/Off key at bottom left of front panel 15

4-2. Normalize Function This function is for calibrating the frequency characteristics of cables, etc., connected at measurement. It is used when accurate measurement is required for assessing the impact of insertion loss of coaxial cables, etc. An example of the operation is explained below. Disconnecting DUT and connecting only cables to be calibrated as shown in Fig. 1 Fig. 1 Ref. Trace Set each parameter. > Refer to previous Common Settings. Set the Normalize function to On. > [Measure] >[ (2of2)] > [F6:SG Control] > [F6:Normalize] Perform Single sweep. > [ Single ] *Wait until sweep finishes. Save Reference Trace. > [Measure] >[ (2of2)] > [F6:SG Control] > [F6:Normalize] > [F3:Store Ref.] *Always execute [F3:Store Ref.] after the sweep has finished. If this command is executed before the sweep has finished, the previous sweep data is saved and the Normalize function does not operate correctly. Set Normalize function to On. > [F1:Nomalize] = On Perform Continuous sweep. > [Continuous ] Set Reference Trace to Off. > [F5:Ref Trace Display] > [F2:Blank] To next slide

4-2. Normalize Function Connecting BPF as shown in Fig. 2 and measuring The level at the marker point displays the relative value based on the level displayed when Normalize is executed. The displayed level value is read as the insertion loss. Fig.2 BPF *After executing the Normalize function, Normalize is initialized if parameters such as the frequency, trace points, RBW, etc., are changed. Execute Normalize again after initialization. 17

4-3. N db Bandwidth Measurement Function This function is used when measuring the bandwidth of a bandpass filter (BPF) from the selected marker peak point. The following example explains operation using a BPF with a cut-off frequency of 3.01 db (1/2 pass-through power). Connecting BPF and setting each parameter Set each parameter. > Refer to previous Common Settings. Execute Normalize. > Refer to previous Normalize Function. BPF Set N db bandwidth Measurement Function to On. > [Measure] >[ (2of2)] > [F6:SG Control] > [F7:N db Bandwidth] > [F2:N db] = [-3.01][F1:dB] > [F1:NdB Bandwidth] = On 18

4-4. Correction Value Capture Function At radio TRX tests, measurement is performed by switching the measurement path containing filters, amplifiers, attenuators, etc. This function captures the correction value for the measurement path as a.csv file. The captured.csv file can be read by the MS2830A Correction Table function to correct the frequency characteristics of the measurement path. Measurement Path ATT BPF SW SW AMP SW SW Capture each path as.csv file Frequency(Hz),Level(dB) 197500000,10.102 197525000,10.101 197550000,10.101 197575000,10.100 :.CSV 19

4-4. Correction Value Capture Function Connecting cables for calibration as shown in Fig. 1 and executing Normalize Set each parameter *1 > Refer to previous Common Settings. Execute Normalize. > Refer to previous Normalize Function. Fig. 1 *1: The maximum settable SPA Trace point number is 10001 but the Correction Table function can read up to 4096 point. Be careful when setting the Trace point number. 基準トレース Connecting path for capturing calibration value as shown in Fig. 2 Perform single sweep. > [ Single ] *Wait until sweep finishes Save calibration value. > [Save] > [F6:Save Correction CSV DATA] *Set save name Default name: Corr + yyyymmdd + sequential number > [F7:Set] = On (example) Corr20161024_003 Fig. 2 The calibration is saved as a.csv file in the following path: [MS2830A] D:/Anritsu Corporation Signal Analyzer/User Data Corrections/ 20

4-4. Correction Value Capture Function Using Correction Table function The captured.csv file can be read by the MS2830A Correction Table function to calibrate the frequency characteristics of the measurement path. Read calibration value. > [System Config] > [ (2of2)] ] > [F7:Correction] > [F1:Correction] = On > [F3:Recall Correction Table] > [F7:Recall Correction Table] *The calibration value file to be read can be selected using the cursor keys. > [F7:Recall] *The Correction Table function is part of the System Config function and the settings are saved even after executing initialization (Preset). For example, the Correction setting remains On even after executing Preset at Correction = On. *Correction On displayed at bottom of SPA screen 21

4-5. Trace Math Function Each trace has a Write function for updating the display at each measurement, and a View function for temporarily saving measured data. These functions can be used to compare each trace and calculate measurement data differences. Math type *1 Op1 Op2 (Power) Op1 + Op2 (Power) Op1 Op2 (Log) Op1 + Op2 (Log) Function Converts logarithm value (dbm) to true value (W) and performs subtraction. Result displayed as logarithmic value (dbm). Converts logarithm value (dbm) to true value (W) and performs addition. Result displayed as logarithmic value (dbm). Subtracts logarithm values (db). Adds logarithm values (db). *The Op (Operand) can be selected from any of the A, B, C, D, E, and F traces. Adding true values of trace A ( 10 dbm) and trace B ( 10 dbm) as power and displaying results at trace C ( 7 dbm) MRK1: Trace C = -7 dbm MRK2: Trace A and B = -10 dbm Save trace A measurement data (View). > [Trace] >[F1:Active Trace] = A > [F2:Trace Type] > [F2:View] Save trace B measurement data (View). > [Trace] >[F1:Active Trace] = B > [F2:Trace Type] > [F2:View] Make trace C Active. > [Trace] > [F1:Active Trace] = C Calculate trace. > [Trace] > [F7:Math Type] > [F3:Op1 + Op2 (Power)] = A 22

4-6. Reflection Characteristics Measurement Reflection characteristics can be measured by using with either a Return Loss Bridge (SWR Bridge) or Directivity Bridge Coupler, etc. The measurement error is determined by the directivity performance of the bridge being used *1. Connecting cables as shown in Fig. 1 and executing Normalize. Set the terminal connected to the DUT to either Open or Short. Take care about the connected terminals. SG SPA (Output) Fig.1 IN Return Loss Bridge OUT (Input) Set each parameter > Refer to previous Common Settings. Execute Normalize. > Refer to previous Normalize Function. DUT To next slide Open or Short CPL OUT Directivity Bridge Coupler IN Open or Short *1 For example, the measurement error is +1.7/ 1.4 db when measuring a DUT with a return loss of 20 db using a bridge with a directivity of 35 db. 23

4-6. Reflection Characteristics Measurement Connecting BPF and measuring Return Loss. Capture the measurement result as shown in the following figure. The numeric value at the marker level display is read as the Return Loss. SG (Output) SPA (Input) Fig.2 IN Return Loss Bridge OUT DUT BPF 50Ω Terminator OUT CPL Directivity Bridge Coupler IN BPF 50Ω Terminator The equation for conversion from Return Loss (RL) to VSWR is: RL VSWR = 10( 20 ) +1 10 (RL 20 ) 1 24

公知 2017-3 MG No. MS2830A-E-L-34-(2.00)