Performance of Image Intensifiers in Radiographic Systems

Similar documents
Measurements of MeV Photon Flashes in Petawatt Laser Experiments

SHADOWGRAPH ILLUMINIATION TECHNIQUES FOR FRAMING CAMERAS

Digital Detector Array Image Quality for Various GOS Scintillators

CORRECTED RMS ERROR AND EFFECTIVE NUMBER OF BITS FOR SINEWAVE ADC TESTS

Introduction. Chapter 16 Diagnostic Radiology. Primary radiological image. Primary radiological image

U.S. Air Force Phillips hboratoq, Kirtland AFB, NM 87117, 505/ , FAX:

High Explosive Radio Telemetry System. Federal Manufacturing & Technologies. R. Johnson, FM&T; B. Mclaughlin, FM&T;

Up-conversion Time Microscope Demonstrates 103x Magnification of an Ultrafast Waveforms with 300 fs Resolution. C. V. Bennett B. H.

Sandia National Laboratories MS 1153, PO 5800, Albuquerque, NM Phone: , Fax: ,

Amorphous Selenium Direct Radiography for Industrial Imaging

Detector And Front-End Electronics Of A Fissile Mass Flow Monitoring System

Nanosecond, pulsed, frequency-modulated optical parametric oscillator

Investigation of Effective DQE (edqe) parameters for a flat panel detector

GYROTRON-BASED MILLIMETER-WAVE: BEAMS FOR MATERIAL PROCESSING. Thomas Hardek Wayne Cooke. William P e r r y D a n i e l Rees

Introduction to Radar Systems. Radar Antennas. MIT Lincoln Laboratory. Radar Antennas - 1 PRH 6/18/02

IMAGE SENSOR SOLUTIONS. KAC-96-1/5" Lens Kit. KODAK KAC-96-1/5" Lens Kit. for use with the KODAK CMOS Image Sensors. November 2004 Revision 2

Performance of Keck Adaptive Optics with Sodium Laser Guide Stars

Detection of Targets in Noise and Pulse Compression Techniques

Examination, TEN1, in courses SK2500/SK2501, Physics of Biomedical Microscopy,

AN IN-LINE POWER MONITOR FOR HE11 LOW LOSS TRANSMISSION LINES

Digital Images & Image Quality

Stimulated Emission from Semiconductor Microcavities

Five-beam Fabry-Perot velocimeter

BIG PIXELS VS. SMALL PIXELS THE OPTICAL BOTTLENECK. Gregory Hollows Edmund Optics

Defense Technical Information Center Compilation Part Notice

High collection efficiency MCPs for photon counting detectors

PHOTOMULTIPLIER TUBE SATURATION INDICATOR. Jeffery F. Ruch P.O. Box 79 West Mifflin, PA. David J. Urban. West Mifflin, PA DISCLAIMER

Detectors for microscopy - CCDs, APDs and PMTs. Antonia Göhler. Nov 2014

Radiology Physics Lectures: Digital Radiography. Digital Radiography. D. J. Hall, Ph.D. x20893

INFRARED MEASUREMENTS OF THE SYNTHETIC DIAMOND WINDOW OF A 110 GHz HIGH POWER GYROTRON

Miniature Spectrometer Technical specifications

REAL-TIME X-RAY IMAGE PROCESSING; TECHNIQUES FOR SENSITIVITY

A new Photon Counting Detector: Intensified CMOS- APS

Optical design of a high resolution vision lens

Recent advances in ALAMO

Observational Astronomy

Modulation Transfer Function

Photons and solid state detection

X-ray light valve (XLV): a novel detectors technology for digital mammography*

Performance of Smoothing by Spectral Dispersion (SSD) with Frequency Conversion on the Beamlet Laser for the National Ignition Facility

A new Photon Counting Detector: Intensified CMOS- APS

Optical System Case Studies for Speckle Imaging

AIGaAs/InGaAIP Tunnel Junctions for Multifunction Solar Cells. Sharps, N. Y. Li, J. S. Hills, and H. Hou EMCORE Photovoltaics

Image Enhancement by Edge-Preserving Filtering

COMMERCIAL IN CONFIDENCE. SUBJECT : Low Light Level Solid State TV Imaging. AUTHOR : S. H. Spencer & N. J. Catlett

CHAPTER 9 POSITION SENSITIVE PHOTOMULTIPLIER TUBES

EXPERIENCE WITH AND STUDIES OF THE SNS* TARGET IMAGING SYSTEM

IPD3. Imaging Photon Detector APPLICATIONS KEY ATTRIBUTES

Introduction to Radar Systems. The Radar Equation. MIT Lincoln Laboratory _P_1Y.ppt ODonnell

Lawrence Berkeley National Laboratory Recent Work

Criteria for Optical Systems: Optical Path Difference How do we determine the quality of a lens system? Several criteria used in optical design

10/3/2012. Study Harder

High Performance. Image Intensifiers

10/26/2015. Study Harder

The Open University s repository of research publications and other research outputs

BASICS OF FLUOROSCOPY

Comparisons Between Digital Gamma-Ray Spectrometer (DSPec) and Standard Nuclear Instrumentation Methods (NIM) Systems

On-line spectrometer for FEL radiation at

Radiographic sensitivity improved by optimized high resolution X -ray detector design.

On spatial resolution

Determining MTF with a Slant Edge Target ABSTRACT AND INTRODUCTION

Imaging Optics Fundamentals

Hardware-in-the-Loop Testing of Wireless Systems in Realistic Environments

U.S. DEPARTMENT OF ENERGY. YlAMT-485 Y-I 2. Project Accomplishment Summary for Project Number 92-Y12P-013-B2 HYDROFORMING DESIGN AND PROCESS ADVISOR

The ACT External HEPA Push-Through Filter Assembly. A. A. Frigo, S. G. Wiedmeyer, D. E. Preuss, E. F. Bielick, and R. F. Malecha

The Development of an Enhanced Strain Measurement Device to Support Testing of Radioactive Material Packages*

DISCLAIMER. Portions of this document may be illegible in electronic image products. Images are produced from the best available original document.

&wf-9+/ob/--21*~~ II. Ron Harper and Robert A. Hike

MINIATURE X-RAY SOURCES AND THE EFFECTS OF SPOT SIZE ON SYSTEM PERFORMANCE

ENGINEERING CHANGE ORDER ECO No. COS-057 Center for Astrophysics & Space Astronomy Date 13 February 2001 University of Colorado, Boulder Sheet 1 of 6

MTF and NPS of single-shot dual-energy sandwich detectors

MAPPING INDUCED POLARIZATION WITH NATURAL ELECTROMAGNETIC FIELDS FOR EXPLORATION AND RESOURCES CHARACTERIZATION BY THE MINING INDUSTRY

Modulation Transfer Function

Understanding Infrared Camera Thermal Image Quality

TSBB09 Image Sensors 2018-HT2. Image Formation Part 1

High-]FrequencyElectric Field Measurement Using a Toroidal Antenna

Y f OAK RIDGE Y4 2 PLANT. Lionel Levinson General Electric Company. November 24, Approved for Public Release; distribution is unlimited.

Preliminary Modulation Transfer Function Study on Amorphous Silicon Flat Panel System for Industrial Digital Radiography

Single- Crystal Sapphire Optical Fiber Sensor Instrumentation

BASLER A601f / A602f

ECEN. Spectroscopy. Lab 8. copy. constituents HOMEWORK PR. Figure. 1. Layout of. of the

Charged Coupled Device (CCD) S.Vidhya

High resolution images obtained with uncooled microbolometer J. Sadi 1, A. Crastes 2

THE MEASURED PERFORMANCE OF A 170 GHz REMOTE STEERING LAUNCHER

ISO INTERNATIONAL STANDARD. Photography Electronic still-picture cameras Resolution measurements

Development of dual MCP x-ray imager for 40 ~ 200 kev region

Image acquisition. In both cases, the digital sensing element is one of the following: Line array Area array. Single sensor

Spectral Analysis of the LUND/DMI Earthshine Telescope and Filters

Production of HPDs for the LHCb RICH Detectors

Cascaded Wavelength Division Multiplexing for Byte-Wide Optical Interconnects

Unit thickness. Unit area. σ = NΔX = ΔI / I 0

Some of the important topics needed to be addressed in a successful lens design project (R.R. Shannon: The Art and Science of Optical Design)

Optical basics for machine vision systems. Lars Fermum Chief instructor STEMMER IMAGING GmbH

Performance Comparison of Spectrometers Featuring On-Axis and Off-Axis Grating Rotation

Marconi Applied Technologies CCD30-11 Inverted Mode Sensor High Performance CCD Sensor

Sub-nanometer Interferometry Aspheric Mirror Fabrication

DESIGN NOTE: DIFFRACTION EFFECTS

Ultra-high resolution 14,400 pixel trilinear color image sensor

Basler aca640-90gm. Camera Specification. Measurement protocol using the EMVA Standard 1288 Document Number: BD Version: 02

High Performance. Image Intensifiers

Transcription:

DOE/NV/11718--396 LA-UR-00-211 Performance of Image Intensifiers in Radiographic Systems Stuart A. Baker* a, Nicholas S. P. King b, Wilfred Lewis a, Stephen S. Lutz c, Dane V. Morgan a, Tim Schaefer a, Mark D. Wilke b a Bechtel Nevada, Los Alamos Operations, P.O. Box 809, Los Alamos, NM 87544 b Los Alamos National Laboratory, Los Alamos, NM 87545 c Bechtel Nevada, Special Technologies Laboratory, Santa Barbara, CA 93101 ABSTRACT Electronic charge-coupled device (CCD) cameras equipped with image intensifiers are increasingly being used for radiographic applications. These systems may be used to replace film recording for static imaging, or at other times CCDs coupled with electro-optical shutters may be used for static or dynamic (explosive) radiography. Image intensifiers provide precise shuttering and signal gain. We have developed a set of performance measures to calibrate systems, compare one system to another, and to predict experimental performance. The performance measures discussed in this paper are concerned with image quality parameters that relate to resolution and signal-to-noise ratio. Keywords: Image intensifier, radiography, CCD, camera 1. INTRODUCTION There are many different types of image intensified cameras in use today. Imaging objectives are used to define experimental requirements and desired system performance. Recording specifications may include field of view, resolution, dynamic range, sensitivity, shutter times, and number of images. To aid in the process of identifying and refining requisite system performance characteristics, we have established a set of performance measures that can be used to provide consistent system comparisons along with modeling information. Optimal system performance is dependent on compatible tube selection. This paper investigates the use of various image tubes used in radiographic imaging systems and discusses the relative merits of commercially available planar diode and micro-channel plate (MCP) image intensifiers. Image intensifiers are sometimes used in conjunction with other image tubes such as framing cameras or electrostatic minifiers, and some of these combinations are discussed. 2. INTENSIFIER PARAMETERS Image intensifier selection parameters include image tube type, photocathode and phosphor material, input and output window material, resolution, extinction ratio, dynamic range and sensitivity. 2.1 Intensifier Type MCP image intensifiers (MCPII) are used extensively for night vision equipment, which typically employs small (18 mm) diameter tubes. MCPIIs are also used in camera systems, sometimes utilizing large diameter MCPs, up to 75 mm in diameter. Diameters of 40 mm and 25 mm are more common. The MCPII is designed for high gain or fast gating applications with moderate to high resolution. Planar diode image intensifiers are designed to provide low noise, high resolution, gated images. Image intensifiers used in * Correspondence: E-mail: bakersa@nv.doe.gov, Telephone: 505-663-2039, Fax: 505-663-2003. 1

front of a framing camera require a fast phosphor output screen so that the temporal information in the radiographed scene is not lost due to smearing from long phosphor decay times. Phosphor selection alsoaffects system gain due to the spectral matching of the phosphor output to the second photocathode. Photocathode selection can also be impacted by its spectral sensitivity to the radiation-to-light converter emission, which is often blue for fast scintillators. Input and output windows are selected for either fiber optic or lens coupling. 2.2 System Performance Parameters Imaging systems are often calibrated for specific experimental conditions. Typical calibration measurements include limiting resolution, magnification, and a transfer curve. Limiting resolution is defined as the maximum detectable spatial frequency, in line pairs per millimeter (lp/mm), referenced to either the image or object plane. System magnification is the ratio of the CCD pixel size to the corresponding pixel size at the radiographic object (M = image/object). A transfer curve maps the system signal level versus input energy. The most useful units provide average pixel counts versus input energy density. A transfer curve describes the dynamic range of the system indicating the noise floor, minimum detectable signal, and saturation level and identifies the linear operating region for the system. Detective quantum efficiency (DQE) is a measure of system efficiency and noise characteristics. Other parameters are used to measure system performance in the frequency domain. Modulation transfer function (MTF), noise power spectra, and DQE are frequency domain parameters 1 used in system modeling. MTF is used to describe system fidelity as a function of input sine wave spatial frequency. Noise power spectrum is used to model the noise contributions over a range of signal levels. Other useful parameters include signal gain and scene contrast. Signal gain is helpful when comparing the relative signal from one system to another for a constant input. Scene contrast can be a very critical parameter for radiographic imaging systems. For good scene contrast, we look for high resolving power near a bright scene. Scene contrast is degraded when a system has poor low frequency response, indicated by tailing or blurring in a sharp edge response. When edge response is not sharp, any scene structure near a bright area may be obscured by roll off in the edge response. Scene contrast is established by calculating the modulation of a resolution pattern near a bright edge. For camera comparison, a standard measure is needed. We have chosen a relatively coarse resolution pattern (10 lp/mm) for each camera. The optical target used has a clear background with dark bar patterns to allow modulation detail near a bright background to be examined. The square bar pattern modulation is the contrast transfer function (CTF). Table 1, below, lists some measured values for different types of image tubes. We are interested in generalizing image tube response, by type, in order to help identify typical tube performance. Examples of measurement techniques are discussed later. Limiting Resolution (lp/mm) MTF = 50%, f c (lp/mm) Scene Contrast CTF @ 10 lp/mm DQE @ 430 nm QE @ 430 nm CCD @ -27 o C 22 17.8 0.71 0.116 ---- CCD + Diode 20 8.1 0.57 0.079 0.15 CCD + 40 mm MCP CCD + Hi Blue MCP Framing Camera 18 9 0.41 0.053 0.14 20 7 0.45 0.114 0.19 14 4.8 0.26 0.004 0.06 Table 1. A comparison of intensifier performance characteristics. 2

All of the image tubes in Table 1 incorporate a p-43 phosphor screen and all have fiber optic input and output windows, with the exception of the Hi-Blue MCP which has a quartz input window. The diode intensifier used is a planar-focused 40-mm diode intensifier. The Hi-Blue MCP is a 25-mm image tube with photocathode sensitivity optimized for blue response. It has a high-resolution, small pore-sized MCP. The framing camera uses a grid-gated electro-static image tube with fiber optic input and output windows. 3. MEASUREMENT TECHNIQUES The measurements discussed in this paper are accomplished using a ring strobe light source in an integrating sphere that illuminates an optical target imaged onto the detector system. The target is imaged through a spectral filter to reduce chromatic aberrations for resolution measurements and to perform sensitivity measurements in the spectral region best matching the intended region of radiographic use. 3.1 Resolution Measurements Limiting resolution is measured by imaging a resolution pattern onto the detector to determine the highest resolvable spatial frequency. For our purposes, we have defined resolvable frequency to be any frequency with better than 5% modulation. From an extracted profile, modulation is defined as: Modulation = m = (S max S min ) / (S max + S min ), where, S max and S min are the maximum and minimum signal levels above background, respectively. 3 Sine wave and square wave modulation correspond to MTF and CTF, respectively. 2 The MTF is the output modulation divided by the input modulation as a function of frequency: MTF(f) = m output (f) / m input (f), m input (f) = 1 for resolution targets. Modulation transfer curves are used to view modulation over a spatial frequency range and to predict a system response in modeling for cascading component MTFs. An MTF curve can be generated by analyzing numerous line pair patterns or by analyzing the system edge response. We use edge response to calculate MTF curves. A profile is extracted from the image to give the edgespread function (ESF). The ESF is then differentiated to produce a line-spread function (LSF). The Fourier transform of the LSF gives the MTF curve. 1 MTF = F (d(esf)/dx) = F (LSF) Scene contrast CTF is calculated as shown below with values taken from the modulation profile in Figure 1. S min = 12.75 10.94 = 1.81, S max = 12.75 6.06 = 6.69 CTF = (S max S min ) / (S max + S min ) = (6.69 1.81) / (6.69 + 1.81) = 0.57 3

14 10 3 Scene Contrast 12.75 12 10 3 S min 10.94 Signal (Counts) 10 10 3 8 10 3 S max 6 10 3 6.06 4 10 3 0 0.2 0.4 0.6 0.8 1 1.2 Image Plane (millimeter) Figure 1. A typical scene contrast plot, a 10 lp/mm bar target is resolved with 57% modulation against a bright background. 3.2 Detective Quantum Efficiency (DQE) DQE is used to measure system sensitivity and noise characteristics. DQE is the square of the output signal-to-noise ratio divided by the square of the input signal-to-noise ratio. 1 The ideal detector has a DQE equal to one. We are measuring DQE of the detector over a large pixel area, approximately 10,000 pixels. This technique calculates DQE in a low spatial frequency response range, which facilitates the removal of fixed pattern noise. DQE = (S/N) out 2 /(S/N) in 2. (S/N) in 2 = (q/q 1/2 ) 2 = q, quanta input. The quanta input is measured with an energy meter in joules (J). The input energy is then converted to number of photons per resolution element as follows: 4 q p = Q e / [(A D )(ev)(hc/e)/λ ], where: q p = photons per pixel, Q e = irradiant energy (Joules) A D = Area of detector illumination (pixels) ev = 1.6022 x 10 19 J (electron energy) hc/e = 1.23985 x 10-6 V m ~= 1240 ev nm (volt-wavelength conversion) λ = input wavelength (nm) q p = Q e / [(A D )( 1.6022 x 10 19 )(1240/λ)], 4

q res.el. = q p x A res.el. = photons per resolution element. A res.el. = area of a resolution element in pixels. The area of a resolution element is determined experimentally using a point source illumination. A single mode fiber, 6-micron core, is used as a point source illumination on the detector. The point source image is background subtracted, then normalized to one. The area of illumination is then integrated to determine the number of total pixels in the spatial impulse response. Ideal resolution element size is one pixel (a lot of spread in the impulse response results in many pixels comprising a resolution element). Signal-to-noise out is defined as the mean signal divided by the standard deviation in the signal after fixed pattern noise has been removed. Fixed pattern noise can be removed by dividing two flat field images or by averaging multiple flat field images. Using two images is less labor-intensive but loses the ability to track the mean signal value. Two image method: Signal noise is determined by subtracting two flat field images to remove the fixed pattern noise. Signal noise = standard deviation (flat1-flat2) / 2 1/2. Ten image method: Ten flat field images are summed then normalized to one. The corrected flat field is divided into a signal image to remove fixed pattern noise. Signal noise = standard deviation (corrected signal). This technique has the advantage of retaining the mean signal values. For MCP image intensifiers, DQE measurements are found to vary with MCP gain voltage as shown in Figure 2. 0.12 Intensified Camera DQE 0.10 0.08 DQE 0.06 0.04 0.02 0.00 400 500 600 700 800 900 1000 MCP Voltage Figure 2. Intensified camera DQE as a function of MCP voltage. 5

3.3 Dynamic Range The dynamic range of a system is measured with a transfer curve, which maps the output signal as a function of input energy density. Figure 3 shows a transfer curve of a MCPII at two different gain settings. Both gain settings allow CCD saturation before MCP saturation is reached. The higher gain setting has both higher sensitivity and higher noise. The dynamic range of the system is the ratio of the saturation threshold input energy to the minimum detectable signal level. The minimum detectable signal is identified as the input energy needed to produce a signal-to-noise ratio of 2 to 1. The minimum detectable signal level is approximately two times the noise equivalent or noise equivalent quanta (NEQ) level. For the transfer curves in Figure 3, the output is approximately linear over the entire dynamic range. A linearity measurement would calculate the camera response deviations from a perfectly linear response over the dynamic range of the camera. For gain 10x, dynamic range = 1 x 10-6 / 4.64 x 10-11 = 2.16 x 10 4. For gain 100x, dynamic range = 1 x 10-8 / 3.08 x 10-12 = 3.25 x 10 3. 10X Gain 1000X Gain 10 5 Transfer Curve CCD Saturation = 65 x 10 4 CCD Readout Signal (Counts) 10 4 10 3 10 2 10 1 s:n = 2:1 @ 20 s:n = 2:1 @ 2 10 0 10-13 10-12 10-11 10-10 10-9 10-8 10-7 10-6 10-5 Input Energy (uj/cm 2 ) Figure 3. MCP intensified CCD transfer curve at two different gain settings. 3.4 Phosphor Coupling Image intensifiers are sometimes positioned in front of a second image tube, such as a framing tube, when additional gain or wavelength shifting will improve system response. The phosphors used on intensifiers employed in this fashion must have decay times that are as short, or shorter, than the desired framing time. The gain study below (Figure 4) indicates the relative gains of three different systems that incorporate planar diodes with different phosphor screens. Some of these results were presented in Large Format Radiographic Imaging, 5 which compared the relative gains of various components viewing a blue light 6

source and considered spectral shifting issues of the phosphor matching to the second photocathode or CCD sensitivity. Figure 4. Relative gain of phosphor screens. CONCLUSION This work is part of an ongoing effort to improve image quality in electronic recording of radiographic images. We continue to investigate the potential benefits of various electro-optical devices. Further investigation is needed into tradeoffs between using lenses for magnification and electro-static minifiers, and in phosphor selection in image intensifiers for gain with consideration given to wavelength shifting, timing, and resolution issues associated with intensifier use in front of framing tubes. We are in the process of developing a software package to be used with a standard set of camera system evaluation tests. This package will enable the user to acquire a complete data set and process the data with a stand-alone package. The package will simplify the processing now associated with analyzing this type of data. The camera evaluation package will provide a consistent method of obtaining system performance and modeling information. ACKNOWLEDGEMENTS This work was supported by the U.S. Department of Energy, Nevada Operations Office, under Contract No. DE-AC08-96NV11718. 7

REFERENCES 1. Dainty, J. C. & Shaw, R. Image Science. Academic Press. 1974. 2. Csorba, I. P. Image Tubes. Howard W. Sams & Co., Inc. Publishing. 1985. 3. Holst, G. C. CCD Arrays, Cameras, and Displays. SPIE Optical Engineering Press. 1995. 4. RCA. Electro-Optics Handbook. 1974 5. Baker, S. A., King, N. S. P., Lewis, W., Rohrer, J. S., Stewart, L., Wilke, M. D. Large Format Radiographic Imaging. in SPIE Vol. 3768, Hard X-Ray, Gamma Ray, and Neutron Detector Physics. 1999. 6. Boreman. G. D. Handbook of Optics, 2 nd Ed. "MTF." McGraw-Hill. 1995 DISCLAIMER This report was prepared as an account of work sponsored by an agency of the United States Government. Neither the United States Government nor any agency thereof, nor any of their employees, nor any of their contractors, subcontractors or their employees, makes any warranty, express or implied, or assumes any legal liability or responsibility for the accuracy, completeness, or any third party's use or the results of such use of any information, apparatus, product, or process disclosed, or represents that its use would not infringe privately owned rights. Reference herein to any specific commercial product, process, or service by trade name, trademark, manufacturer, or otherwise, does not necessarily constitute or imply its endorsement, recommendation, or favoring by the United States Government or any agency thereof or its contractors or subcontractors. The views and opinions of authors expressed herein do not necessarily state or reflect those of the United States Government or any agency thereof. DISTRIBUTION LIST DOE Nevada Operations Office U.S. Department of Energy DOE/NV Public Technical Information Resource Center Office of Scientific and Technical Reading Facility P.O. Box 98518 Information P.O. Box 98521 89193-8518 P.O. Box 62 Las Vegas, NV Las Vegas, NV 89193-8518 Oak Ridge, TN 37831 8