Glass Electrode Meter

Similar documents
INSTRUCTION MANUAL FOR MICROELECTRODE AC AMPLIFIER MODEL 1800

INSTRUCTION MANUAL FOR CELL SIMULATOR MODEL 2410

Four-Channel Differential AC Amplifier

Neuroprobe Amplifier

2001A. 200KHz Function Generator Instruction Manual. 99 Washington Street Melrose, MA Phone Toll Free

Warranty Terms & Conditions

Document Version 1.2

301 & 601 Mic/Line Mixers Operation Manual

Broadband Current Probe Series Operation Manual

Line 6 L I M I T E D R E F URBISH E D (B-ST O C K) W A RR ANT Y

Model 113 Scintillation Preamplifier Operating and Service Manual

DA6002D-DA10004D. INSTALLATION / OWNER'S MANUAL Mobile Power Amplifiers

305 Stimulus Isolator. Description

Broadband Step-Up Transformer. User Manual

201AP Charge Amplifier User Manual

Broadband Current Probe Series Operation Manual

742A Series Resistance Standards

CT-2 and CT-3 Channel Taggers OPERATION MANUAL

DA560D COMPACT SERIES. INSTALLATION / OWNER'S MANUAL Mobile Power Amplifiers

SX204. Headphone Amplifier. Symetrix. Owner s Manual

XPR522 XPR540. XPR SERIES INSTALLATION / OWNER'S MANUAL Mobile Power Amplifiers

PARALLEL MULTI-AMP KIT for 7200 Series AMPLIFIERS INSTRUCTION SHEET

P5100A & P5150 High Voltage Probes Performance Verification and Adjustments

DA604D DA954D DA501D DA801D COMPACT SERIES. INSTALLATION / OWNER'S MANUAL Mobile Power Amplifiers

RFTX-1 Installation Manual

Gypsy Statement of Limited Warranty. Part 1 General Terms

80i-600A AC Current Probe

Tilting, Swiveling & Rotating Flat Panel Wall Mount

Advanced Test Equipment Rentals ATEC (2832)

An American Control Electronics Brand PCM4 SERIES USER MANUAL PCM4.

Model BSI-1A BIPHASIC STIMULUS ISOLATOR

INSTRUMENTS, INC. Model 2960AX Disciplined Quartz Frequency Standard 2960AX. Section Page Contents

AMP-13 OPERATOR S MANUAL

XPA2100 XPA4100 XPA6100. XPA SERIES INSTALLATION/OWNER S MANUAL Mobile Power Amplifiers

Vi6208, Vi6216, Vi6232 Vi6308, Vi6316, Vi6332

P5100A & P5150 High Voltage Probes Performance Verification and Adjustments

Safety. This symbol, adjacent to a terminal, indicates that, under normal use, hazardous voltages may be present.

Model 935A Current Source Operation Manual

ADA416-XLR DISTRIBUTION AMPLIFIERS OPERATING AND MAINTENANCE MANUAL

HDTV Antenna. Contemporary Indoor. User s Guide

12B/18 MultiMeter Instruction Sheet

ISO-80. Isolated Differential Amplifier. INSTRUCTION MANUAL. Serial No. World Precision Instruments

Model 9305 Fast Preamplifier Operating and Service Manual

PCO-7114 Laser Diode Driver Module Operation Manual

SI-125 Power Amplifier Manual 6205 Kestrel Road; Mississauga, Ontario; Canada; L5T 2A1 November 2016, Rev 0.5

Current Probe Fixture Instruction Manual

Operator s Manual. PP016 Passive Probe

HTA125A/250A. Power Amplifiers. Installation & Use Manual

Copyright 2014 by Minarik Drives

MICROTOOLS MICRONETBLINK KIT

Classic Series Public Address Amplifiers C10 & C20 Models

KRAMER ELECTRONICS LTD. USER MANUAL MODEL: 912 Power Amplifier. P/N: Rev 2

Instruction Manual Model M Switch, DPDT, Manual Select

41P Portable Calibrator User Manual

DL102 Counter Loop Amplifier

HHMA2 DC / TRUE RMS AC NON-CONTACT MILLIAMMETER

Wilcom MODEL T136BGMZW CIRCUIT TEST SET. Operating Instructions

Preliminary Data Sheet!

36 AC/DC True RMS. Clamp Meter. P Read First: Safety Information

DA208 & DA416 DISTRIBUTION AMPLIFIERS OPERATING AND MAINTENANCE MANUAL

Model DB Disc Caliper Brake AIR CHAMP PRODUCTS. User Manual. (i) MTY (81)

10 Amp Digital PWM Motor Speed Controller CV-2110-HD and CV-2110-HDS

15 Amp Digital High Frequency PWM Motor Speed Controller SPD-315-D and SPD-315-DS

XIA3145 INSTALLATION/OWNER S MANUAL 2/1-Channel Mobile Power Amplifier

Single Flex and Double Flex Couplings (i)

40 Amp Digital Bidirectional PWM Motor Controller with Regenerative Braking BIDIR-340-DR

BUNN AF PR-2 AF PR-3 BUNN-O-MATIC CORPORATION POST OFFICE BOX 3227 SPRINGFIELD, ILLINOIS PHONE: (217) FAX: (217)

MODEL W Power Amplifier

Radio Remote Controls Manual K Series

Model 7000 Low Noise Differential Preamplifier

TRIAXIAL FLUXGATE MAGNETOMETER OWNER'S MANUAL

LUDLUM MODEL 421 AND PMT BASE WITH PREAMPLIFIER/BIAS SUPPLY. October 2014 Serial Number and Succeeding Serial Numbers

SSR VL-12P-xC-NCS

EQ-AMP60 60W Mixer Amplifier

AMP-12 OPERATOR S MANUAL

INSTALLATION AND MAINTENANCE MANUAL FOR GROUND MONITOR GM-250 COPYRIGHT 1983 AMERICAN MINE RESEARCH, INC.

INSTRUCTION MANUAL. March 11, 2003, Revision 3

FRYCORDER Power Cord Burn-in Generator

Tilting Flat Panel Wall Mount Installation Guide

User Manual. Smart-UPS On-Line. Uninterruptible Power Supply. Isolation Transformer Models: SURT001 and SURT002

JMAA-1800HR15. Installation and Operation Instructions. 180w Performance Series Amplifier Kit For Harley RoadGlide/ Ultra

SSR V-12C. Industrial Switching Applications: Hellroaring Technologies, inc. P.O. Box 1521, Polson, Montana 59860

Radio Remote(s) (Installation Manual)

Model 935A. Dual Tone Sender INSTRUCTION MANUAL

DUAL CHANNEL BROADBAND LINEAR AMPLIFIER Model A800D

Directed Energy, Inc Oakridge Dr., Suite 100, Fort Collins, CO

Manual Carton Closing Staplers

DC200A Displacement Clipper User Manual

Wilcom MODEL T136BSBZW CIRCUIT TEST SET. Operating Instructions

Inline Antenna Signal Amplifier

MODEL 3810/2 Line Impedance Stabilization Network

Operator s Manual. PP017 and PP018 Passive Probes

34134A AC/DC DMM Current Probe. User s Guide. Publication number April 2009

VideoEase VGA 1x4 Distribution Hub (500150, ) Installation Guide

Smart-UPS On-Line Isolation and Step-Down Transformer SRT5KTF

TIA-527 Balanced Optical/Electrical Converter. Operating Instructions

Angular Rate Sensor. Owner's Manual

PHASE ROTATION METER. Operating and Instruction Manual. a n d A C C E S S O R I E S

automatic embosser & die cutter USER MANUAL

COMMERCIAL TRANSMITTER INSTRUCTIONS

Transcription:

Glass Electrode Meter

INSTRUCTION MANUAL FOR Glass Electrode R/C Meter MODEL 2700 Serial # Date PO Box 850 Carlsborg, WA 98324 U.S.A. 360-683-8300 800-426-1306 FAX: 360-683-3525 http://www.a-msystems.com Version 4.0 April, 2010

Contents General Description... 1 Instrument Features... 1 Controls and Connectors... 2 Operating Instructions... 4 Typical Set-Up Procedure... 4 Safety considerations... 7 Problem Solving... 7 Calibration Procedures... 8 Initial conditions:... 8 Power supply:... 8 Oscillator:... 8 Gain:... 9 Specifications... 10 Input... 10 Internal signal source... 10 Meter Output... 10 Output... 10 Power... 10 Physical Dimensions... 11 Model 8750 Remote power supply... 11 Warranty and Service... 12 Each R/C Electrode Meter is delivered complete with: One 3 Foot BNC Input Cable One 3 Foot Banana GND cable Rack Mount Hardware Instruction Manual Power Supply Cord Power Supply NOTE This instrument is not intended for clinical measurements using human subjects. does not assume responsibility for injury or damage due to the misuse of this instrument.

General Description Instrument Features The Model 2700 R/C Electrode Meter is designed to measure resistance and capacitance of glass microelectrodes. Resistance and capacitance are important indicators of electrode recording and stimulating properties. The Model 2700 uses a low level, 1 khz signal to measure resistance and capacitance in order not to affect the recording properties of the electrode. The resulting values are displayed on a digital meter. 1

Controls and Connectors INPUT: This BNC connector is used to connect an electrode to the meter. The driven shield of the BNC is not grounded; but, rather it is driven with a voltage equal to the measurement signal to minimize cable capacitance. The center conductor of the BNC should be connected to the electrode to be measured. WARNING: The driven shield should not be connected to anything, except the front panel BNC input connector. GND: This connector provides a ground or reference point for measuring electrode resistance. OUTPUT: This BNC connector provides an output waveform of the voltage across the electrode. The impedance of the electrode can be calculated from the peak value of this waveform by the following equation or by reading the value from the chart on page 6. Vout peak Z electrode * R 1 Vout where Range switch 0.2M 2M 20M 100M 400M 1000M peak range R range 50k 500k 5M 25M 100M 200M CAPACITY ADJ.: This knob is used to adjust an active feedback circuit to compensate and measure up to 200 pf of total capacitance from the electrode and input cable. The capacitance compensation can be accurately monitored in the experimental setup by connecting the OUTPUT connector to an oscilloscope. The CAPACITY ADJ. knob should be adjusted to obtain the sharpest corners possible on the square-wave. Clockwise rotation of this control increases the capacity compensation. If the knob is turned too far clockwise the warning "over compensated" will be displayed on the meter. 2

RANGE: This knob controls the range of resistance to be measured and the frequency at which it is measured. If the knob's position is below the electrode's resistance, the warning "Increase Range" will be displayed on the meter. If the knob's position is above the electrode's resistance the warning "Decrease Range" will be displayed on the meter. DISPLAY This switch controls both the test signal and the meter display. When the switch is in the MEASURE position, there will be a continuous application of the test signal to the electrode being measured. While the switch is in the HOLD position, the last measurement will be displayed, and the test signal will be turned off. POWER This switch turns the R/C Electrode Meter ON or OFF. Rear Panel POWER INPUT This jack connects the DC power source to the Model 2700. The DC supply contains +15V, -15V, and +5V. 3

Operating Instructions Typical Set-Up Procedure This is a generalized procedure for setting up the Model 2700 Glass Electrode R/C Meter for measuring electrode characteristics. Portions of this procedure may need to be modified for your specific application. 1. Use a beaker filled with physiological saline (or the solution in which the tissue will be bathed). The solution should have the same temperature and ionic strength as in which your experiments will be made. 2. Connect a ground wire (generally an Ag/AgCl reference electrode) from the GND connector on the front panel to the saline within the beaker. 3. Place the ground wire (Ag/AgCl reference electrode) 1-2 inches deep in the saline solution. 4. Connect the center conductor on the input cable to your electrode to be measured. WARNING: Do not connect the shield of the input cable to ground. 5. Connect the input cable's BNC connector to the instrument's INPUT connector. 6. Dip the glass microelectrode tip into the beaker of physiological saline solution. Note: immerse the micropipette to approximately the same depth as will be used during the experiment. 7. Set the instrument controls as follows: POWER DISPLAY CAPACITY ADJ. RANGE OUTPUT OFF HOLD FULLY COUNTERCLOCKWISE Set to the maximum value you think the electrode resistance might be. For example if you think the electrode will have a resistance below 5 M then set the range to 20 M. This can be connected to an oscilloscope if you want to observe the test signal. 4

8. Turn the POWER switch to ON and allow the Model 2700 to warm up for 1 minutes. 9. Now set the DISPLAY switch to MEASURE, and then read the value on the meter. 10. If the meter reads Increase Range or Decrease Range then turn the RANGE knob clockwise to increase the range and counterclockwise to decrease the range. 11. In most situations, the value displayed will be close to the actual resistance. But because the electrode and cable have capacitance, you should increase the CAPACITY ADJ. (rotate the knob clockwise), until the meter reads Over Compensated. Finally, turn the knob a small amount counterclockwise until the display no longer reads Over Compensated. The capacitance value displayed on the meter will be the actual capacitance of the electrode and cable. Note: using a resistor in place of the electrode will give the capacitance of the cable. Once the electrode has been properly adjusted for capacitance, the resistance value displayed on the meter will be the actual resistance of the electrode. It is useful to view the test signal on an oscilloscope when using the CAPACITY ADJ. Exact measurement of resistance and capacitance occurs when the square wave signal at the OUTPUT connector is squared up and does not have high frequency oscillations on top of the signal. 12. Set the DISPLAY switch to HOLD to stop the test signal. 5

Electrode Impedance 1000.00 1000M Range 100.00 400M Range 10.00 100M Range Impedance (MW) 20M Range 1.00 2M Range 0.10 0.2M Range 0.01 0 0.1 0.2 0.3 0.4 0.5 0.6 0.7 0.8 0.9 1 Output Voltage (Vpeak) 6

Safety considerations Removal of the cover will expose electronic components. Capacitors inside the instrument may still be electrically charged even when the instrument has been disconnected from all power sources. Do not ground the shield of the input cable. The shield has a voltage applied to it; and, if it is tied to ground, it will cause the Glass Electrode R/C Meter to give erroneous results. The shield of the input connector should only be connected to the INPUT BNC connector. Driven Shield and Ground Connections Since the input is enclosed by a driven shield maintained at the same potential as the input connector, there is no electric field between the input connections and shield. Consequently, there is no capacitive shunting do to the cable. This shield is connected to the external connector of the BNC. Using a driven shield for electrode measurements has the advantage of not introducing any additional shunt capacitance nor a path for the leakage current to ground. Problem Solving If the Model 2700 does not function properly, consult the following list of suggested solutions to the most common problems. If you need further assistance, please contact technical support at, Inc. Problem Values keep changing The display reads over compensated even though the CAPACITY ADJ. is off. Cause / Solution Possibly, this is caused by too much noise in your setup. Observe the output on an oscilloscope. If the waveform is not a square wave and looks more like a sine wave on top of a square wave then you are getting too much power-line interference. Try turning off unnecessary equipment. Try to shield the Model 2700, glass microelectrode, ground electrode, and input cables from external noise. This can be due to excess noise. The treatment is the same for Values keep changing SEE ABOVE. 7

Calibration Procedures Initial conditions: Controls Power off Display Hold Capacity Adj. fully counterclockwise Range 0.2 Power supply: Controls Inputs / Observations Adjust / Check Power: On Observe voltages at power Check for + 15 V ± 0.50 V Check for - 15 V ± 0.50 V Check for + 5 V ± 0.10 V Oscillator: Controls Inputs / Observations Adjust / Check Power: on Display: measure Observe voltages at TP3 (TP1 is ground) Check for 100m Vp-p Square wave of 1kHz Adjust R28 for frequency Adjust R27 for amplitude 8

Gain: Controls Inputs / Observations Adjust / Check Power: On Display: measure Range: 0.2 Place 50k resistor between input and gnd Observe voltage at output Check for clean 1.0 Vp-p Square wave of 1kHz. Adjust R27 until meter reads 50k 9

Specifications Input Impedance Capacitance Working range Maximum range 10 13 1pF Adjustable to zero ± 5.0 V ± 10 V Internal signal source Frequency Amplitude 1kHz Range < 1.99 M 100 Hz Range 2-99 M 10 Hz Range 100-1000M Variable current to assure that the potential across the electrode stays under 100mV. Meter Output Resistance 10k - 1G +/- 5% of RANGE Capacitance 0-200pF +/- 5% Output External Signal Source Voltage Output Current Output Output Resistance ± 5 V ± 5 ma 100 Power +15 Vdc 50mA - 15 Vdc 50mA + 5 Vdc 100mA DC voltages are used in the METER, so it can be used inside a Faraday cage 10

Physical Dimensions Width Height Depth Weight 8.45 inches (21.46cm) 2.55 inches (6.477 cm) 3.95 inches (10 cm) 12 lbs. Model 8750 Remote power supply Input: EING Output +15Vdc 2.0A 100-240VAC 50/60 Hz, 1.0A MAX. -15Vdc 0.5A +5Vdc 3.0A 36W Max 11

Warranty and Service LIMITED WARRANTY What does this warranty cover?, LLC (hereinafter, ) warrants to the Purchaser that the Instrument, including cables, Headstage Probes and any other accessories shipped with the Instrument,(hereafter the hardware ) is free from defects in workmanship or material under normal use and service for the period of three (3) years. This warranty commences on the date of delivery of the hardware to the Purchaser. What are the obligations of under this warranty? During the warranty period, agrees to repair or replace, at its sole option, without charge to the Purchaser, any defective component part of the hardware. To obtain warranty service, the Purchaser must return the hardware to or an authorized distributor in an adequate shipping container. Any postage, shipping and insurance charges incurred in shipping the hardware to must be prepaid by the Purchaser and all risk for the hardware shall remain with purchaser until such time as takes receipt of the hardware. Upon receipt, A-M Systems will promptly repair or replace the defective unit, and then return the hardware (or its replacement) to the Purchaser, postage, shipping, and insurance prepaid. may use reconditioned or like new parts or units at its sole option, when repairing any hardware. Repaired products shall carry the same amount of outstanding warranty as from original purchase, or ninety (90) days which ever is greater. Any claim under the warranty must include a dated proof of purchase of the hardware covered by this warranty. In any event, liability for defective hardware is limited to repairing or replacing the hardware. What is not covered by this warranty? This warranty is contingent upon proper use and maintenance of the hardware by the Purchaser and does not cover batteries. Neglect, misuse whether intentional or otherwise, tampering with or altering 12

LIMITED WARRANTY, cont What are the limits of liability for under this warranty? shall not be liable for loss of data, lost profits or savings, or any special, incidental, consequential, indirect or other similar damages, whether arising from breach of contract, negligence, or other legal action, even if the company or its agent has been advised of the possibility of such damages, or for any claim brought against you by another party. THIS EQUIPMENT IS NOT INTENDED FOR CLINICAL MEASUREMENTS USING HUMAN SUBJECTS. A-M SYSTEMS DOES NOT ASSUME RESPONSIBILITY FOR INJURY OR DAMAGE DUE TO MISUSE OF THIS EQUIPMENT. Jurisdictions vary with regard to the enforceability of provisions excluding or limiting liability for incidental or consequential damages. Check the provision of your local jurisdiction to find out whether the above exclusion applies to you. This warranty allocates risks of product failure between the Purchaser and. hardware pricing reflects this allocation of risk and the limitations of liability contained in this warranty. The agents, employees, distributors, and dealers of are not authorized to make modifications to this warranty, or additional warranties binding on the company. Accordingly, additional statements such as dealer advertising or presentations, whether oral or written, do not constitute warranties by and should not be relied upon. This warranty gives you specific legal rights. You may also have other rights which vary from one jurisdiction to another. THE WARRANTY AND REMEDY PROVIDED ABOVE IS IN LIEU OF ALL OTHER WARRANTIES AND REMEDIES, WHETHER EXPRESS OR IMPLIED. A-M SYSTEMS DISCLAIMS THE WARRANTIES OF MERCHANTIBILITY AND FITNESS FOR A PARTICULAR USE, WITHOUT LIMITATION. 13

2700 Manual DRW-5027700 rev 4 Revision History Rev Date Description 3 6/30/06 Initial Document Control release 4 4/28/10 DCR201200 Warranty and Company info