Electrochemical Quartz Crystal Microbalance (eqcm)

Similar documents
NOVA. EQCM Tutorial. Version

173 Electrochemical Impedance Spectroscopy Goals Experimental Apparatus Background Electrochemical impedance spectroscopy

LECETURE 4. Piezoelectric sensor. Part 1

Recent Researches in Circuits, Systems and Signal Processing

QCM200 Quartz Crystal Microbalance Digital Controller. QCM25 5 MHz Crystal Oscillator

A potentiostat is an electronic instrument that controls the voltage between two electrodes

Universal Dummy Cell 3. Operator's Manual

Metrohm Autolab Instruments for Electrochemistry

AN INTEGRATED ULTRASOUND TRANSDUCER DRIVER FOR HIFU APPLICATIONS. Wai Wong, Carlos Christoffersen, Samuel Pichardo, Laura Curiel

Simple Quartz Crystal Models: A Review

An Oscillator Scheme for Quartz Crystal Characterization.

being developed. Most up and coming drugs are extremely expensive and limited in

Quick Check of EIS System Performance

Universal Dummy Cell 2. Operator's Manual

Characteristics of Crystal. Piezoelectric effect of Quartz Crystal

Introduction. sig. ref. sig

Design and Characterization of a Rotating Electrochemical Quartz-Crystal-Microbalance Electrode

EasyChair Preprint. Real-Time Monitoring System for Polystyrene Coating Material Deposition onto QCM Sensor using Ultrasonic Atomizer Spray

INDIAN INSTITUTE OF SCIENCE EDUCATION AND RESEARCH PUNE

Compliance Voltage How Much is Enough?

ELECTROCHEMICAL QUARTZ CRYSTAL NANOBALANCE

Introduction and QCM-D Theory. Q-Sense Basic Training, April 4-5, 2006

The effect of phase difference between powered electrodes on RF plasmas

A complete solution for your Electrochemistry research initiative

EXPERIMENTAL SETUP AIMED TO STUDY THE ELECTRICAL IMPEDANCE VARIATIONS OF A PLASMA COLUMN IN A WIDE FREQUENCY RANGE

THE ELECTROMETRIC AC-DC TRANSFER STANDARD AS PRIMARY STANDARD AT IEN FOR AC VOLTAGES FROM 300 V TO 1000 V

Crystal Characterisation with the TrewMac TE3001 Analyser

Lab 4. Crystal Oscillator

Multiply Resonant EOM for the LIGO 40-meter Interferometer

STUDENT LABORATORY WORKSHEET EXPERIMENT B: NANOSCALE THIN FILMS

Introduction to Microeletromechanical Systems (MEMS) Lecture 12 Topics. MEMS Overview

Supplementary Information

Clocking the Data ABSTRACT INTRODUCTION KEY WORDS

Measuring Batteries using the Right Setup: Dual-cell CR2032 and Battery Holder

GENERAL OPERATING INFORMATION

MEMS Real-Time Clocks: small footprint timekeeping. Paolo Frigerio November 15 th, 2018

Testing Electrochemical Capacitors Part 3: Electrochemical Impedance Spectroscopy

Setting up a Multi sine impedance measurement

Optimization of a Love Wave Surface Acoustic Device for Biosensing Application

Advanced Fuel Cell Diagnostic Techniques for Measuring MEA Resistance

SAMPLE: EXPERIMENT 2 Series RLC Circuit / Bode Plot

Design note for YIC Quartz Crystal Unit

AN5E Application Note

Introduction to LIVM Accelerometers

Potentiostat/Galvanostat/Zero Resistance Ammeter

Measurement of the equivalent circuit of quartz crystals

An Instrumentation System

sensors ISSN by MDPI

Lab 4. Crystal Oscillator

Cell-chip coupling for bioelectronic devices. Lab manual

Substrate as Efficient Counter Electrode for Dye- Sensitized Solar Cells

Resonance Tube Lab 9

Kistler portable triaxial Force Plate

Using ir compensation

Communication Circuit Lab Manual

TA Instruments RSA-G2 Dynamic Mechanical Analyzer

Crystal Resonator Terminology

Reference 3000 Potentiostat/Galvanostat/ZRA

List of Crystal XXXXXXXX Unit Model Names kHz

Investigation of An Acoustic Temperature Transducer and its Application for Heater Temperature Measurement

Nickel Thin Film Resonantly Generated at a Rate of Megahertz

Thermal Johnson Noise Generated by a Resistor

Resonance Tube. 1 Purpose. 2 Theory. 2.1 Air As A Spring. 2.2 Traveling Sound Waves in Air

Enable Highly-Stable Plasma Operations at High Pressures with the Right RPS Solution

POT/GAL 15 V 10 A and POT/GAL 30 V 2 A. Electrochemical Impedance Potentiostat Galvanostat Test Interface for Alpha-A Analyzer

PvdF Piezoelectric Film Based Force Measuring System

Passive external radio frequency filter for Langmuir probes

An Experimental Study on Fabricating an Inverted Mesa-Type Quartz Crystal Resonator Using a Cheap Wet Etching Process

Introduction to Measurement Systems

Resonant Frequency of the LRC Circuit (Power Output, Voltage Sensor)

LABORATORY #3 QUARTZ CRYSTAL OSCILLATOR DESIGN

Instruction manual and data sheet ipca h

9. How is an electric field is measured?

ECE 4670 Spring 2014 Lab 1 Linear System Characteristics

Technical Information

Chapter 2. The Fundamentals of Electronics: A Review

STRAIN, FORCE, PRESSURE, AND FLOW MEASUREMENTS

Study of Inductive and Capacitive Reactance and RLC Resonance

multi-channel Potentiostat / Galvanostat / Impedance Analyzer Rev

Laboratory Exercise 6 THE OSCILLOSCOPE

Potentiostat / Galvanostat / Impedance Analyzer

4. Digital Measurement of Electrical Quantities

A HIGH PRECISION QUARTZ OSCILLATOR WITH PERFORMANCE COMPARABLE TO RUBIDIUM OSCILLATORS IN MANY RESPECTS

Experiment: P34 Resonance Modes 1 Resonance Modes of a Stretched String (Power Amplifier, Voltage Sensor)

TUNED AMPLIFIERS 5.1 Introduction: Coil Losses:

Exercise 9: inductor-resistor-capacitor (LRC) circuits

HIGH LOW Astable multivibrators HIGH LOW 1:1

High-speed wavefront control using MEMS micromirrors T. G. Bifano and J. B. Stewart, Boston University [ ] Introduction

EECS40 RLC Lab guide

Lab E5: Filters and Complex Impedance

Technology. Construction of magnetic buzzer. Construction of magnetic buzzer. Operation principles and construction

High-End Sensors & Sensor System: How to Achieve High Metrological Performances?

Multi-spectral acoustical imaging

BANNARI AMMAN INSTITUTE OF TECHNOLOGY SATHYAMANGALAM DEPARTMENT OF PHYSICAL SCIENCES Unit III ULTRASONICS

THERMAL NOISE. Advanced Laboratory, Physics 407, University of Wisconsin. Madison, Wisconsin 53706

Lab 2: Linear and Nonlinear Circuit Elements and Networks

A NEW HIGH-FREQUENCY TORSIONAL RHEOMETER FOR BITUMINOUS BINDERS

PHYS102 Previous Exam Problems. Sound Waves. If the speed of sound in air is not given in the problem, take it as 343 m/s.

INTRODUCTION TO AC FILTERS AND RESONANCE

VersaSTAT Series. potentiostat/galvanostat

Transcription:

Electrochemical Quartz Crystal Microbalance (eqcm) 1. Fundamentals The quartz crystal microbalance (QCM) is based on the piezoelectric effect. This is where a crystal produces a potential across a material in response to a mechanical strain. The converse is also true, a mechanical strain can be produced by application of a potential across a material. Quartz is by far the most commonly used due to its availability and many different ways to cut the crystal. Application of an alternating potential (sine wave) to the crystal faces cases the crystal to oscillate. A standing wave within the crystal can be established and (Figure 1) the frequency of this oscillation is called the resonate frequency. This fundamental frequency can be calculated based on the properties of the crystal. The quartz crystal oscillations can be modeled in a similar fashion to a mechanical Figure 1. Scheme of QCM Shear deformation spring-mass-damper simple harmonic oscillator or an electrical RLC circuit. In the case of an RLC circuit (Figure 2a), R represents the energy dissipated during the oscillation, C represents the energy stored and L represents the inertial component related to the displaced mass. Therefore, the impedance spectrum of the quartz crystal can be used to extract the values of R, L and C. An example Impedance spectrum for a RLC circuit is shown in Figure 2b. Here, a sharp spike is observed at the resonate frequency ( f0 ) and therefore a change in mass results in a change in frequency. Figure 2. (a) Scheme of an RLC Circuit for modeling the Impedance response of a quartz crystal (b) Impedance spectrum for a series RCL circuit

In practice, the act of adding electrical contacts to the crystal (represented as gold contacts in Figure 1 ) adds an additional capacitance. This capacitance is added into the RCL model and is denoted the Butterworth-Van Dyke (BVD) Model (Figure 3a). This circuit model results in two resonant frequencies fs and fp,(figure 3b) which represent the series resonant frequency (same as the original RLC circuit in Figure 2a) and parallel resonant frequency. In the example of the Stanford Research Systems (SRS) QCM, the parallel capacitance is manually canceled out on the front of the instrument and the series resonant frequency is measured. Figure 3. (a) Equivalent circuit for the Butterworth-Van Dyke Model (b) Impedance spectrum for the BVD model As previously mentioned, the resonant frequency is dependent on the inductor element, which itself is a function of the mass displacement, therefore, the mass change of the quartz crystal can be correlated with a change in the resonant frequency. This dependance is represented as the Sauerbry equation (Equation 1) and can be further simplified to Equation 2 where Cf is the calibration constant. This constant is usually 56.6 Hz cm 2 µg -1 for a 5 MHz At-cut quartz crystal in air, however it is recommended to calibrate each crystal individually. Additionally this analysis assumes a thin rigid film and the decrease in frequency is small in comparison to the resonance frequency (5 MHz). f0 - Fundamental frequency of crystal (5 MHz); µq - Shear modulus of crysta ρq - Density of the crystal m - Mass added n - Harmonic number (n=1, 5 MHz driven at 5 MHz) (1) (2)

2. Setting up the Experiment Gamry Ref 600 *Orientation Matters* Figure 4. Instrument setup for eqcm To PC *Notice that the SRS QCM output is +/- 10 V and the Gamry Ref 600 is +/-3 V. Therefore you Must change the scale factor (Hz/V) such that the output is between +/- 3V*

The above setup is for eqcm and the SRS QCM doesn't need to be connected to the potentiostat if the mass change is not an electrochemical process. In addition to setup, the glass cell containing the solution must be sealed to the face of the eqcm. It is recommend that the cell have a water jacket in order to stabilize the temperature of the crystal environment. Additional details on how to assemble the electrode and electrode connections is provided in the SRS QCM manual and a video is available online at http://www.thinksrs.com/products/qcm200.htm 3. Experiment : Calibrating the Quartz crystal microbalance by Cu 2+ electrodeposition 1. Prepare an aqueous electroplating solution consisting of 10-100 mm CuSO4 and 0.5 M H2SO4. In this example the copper sulfate is 50 mm 2. Estimate the change frequency ( Δf ) In this experiment, a constant current plating and stripping process will be used to calibrate the sensitivity factor (Cf). The first step will be Cu plating for 50 seconds at a rate of -1.0 ma. This results in 0.05 C of charge being passed. The next step will be Cu stripping for 50 seconds at a rate of + 1.0 ma. The expected frequency change can be calculated from the area (A,1.38 cm 2 ), molecular weight (Mw. Cu = 63.54 g/mol), Faradays constant (F), number of electrons transferred (n, for Cu 2+ n = 2) and the estimated sensitivity factor (Cf, 56.6 Hz cm 2 µg -1 for a 5 MHz At-cut quartz crystal) using the following equation. In this example the expected frequency change is ~ 675 Hz. (3) IMPORTANT: The analogue frequency output of the SRS QCM will output a voltage proportional to Δf, the proportionally factor can be set on the instrument display. The scale factors can be set to the following (200 Hz/V, 500 Hz/V, 1 khz/ V, 2 khz/v, 5 khz/v, 10 khz/v, 20 khz/v.) Remember the output could be +/-10 V, However the voltage MUST be between +/- 3V for the Gamry Ref. 600 to read. Therefore, in this example the scale factor will be set to 500 Hz/V in order to ensure that the frequency change ( Δf ) is within +/- 3 V.

3. Fill Electrochemical cell and immerse the reference electrode (Ag/AgCl Sat. KCl ), counter electrode (graphite rod or platinum mesh) and working QCM electrode as shown in Figure 4. 4. Once immersed in the solution, flip the Co compensation switch to [ADJUST], the lights will turn on, then adjust the dial up/down until only the illuminated lights are reading [NULL]. This is the manual capacitance compensation as previously mentioned. After this process, flip the Co compensation switch back to [HOLD]. This compensation process should be repeated every time the environment of the crystal/electrode changes. 5. The [PARAMETER] button on the front of the display is used to cycle through all the available options (Gate interval (0.1 s), Scale Factor, Frequency, Resistance). The [VALUE] button is used to cycle through the frequency and resistance display modes which include the following (F: Absolute Frequency, f: Relative Frequency, m: mass displacement [ng] = relative frequency/ 0.0566, R: Absolute Resistance, r: Relative Resistance). 6. Set the display to relative frequency (f), then hold down the [VALUE] button in order to zero the relative frequency. This is the value that will be output to the Gamry REF 600 using the following equation. Analogue Frequency out [V] = Relative Frequency/-Scale Factor. The Scale factor is set using [PARAMETER] and [VALUE] buttons to find and change the current scale factor (set 500 Hz/V). *The positive voltage direction indicates an increase in mass and the more negative, a mass decrease, Hence the negative sign in the equation* 7. Open SRSQCM on the Desktop or by going to the folder C:/srsqcm200/ srsqcm200.exe. Also open up Gamry Framework. 8. Allow the QCM to stabilize after coming into contact with the liquid. Purge the solution with argon. Turn off Ar purge while experiment is running in order to minimize interference with any changes in pressure. Also eliminate any external vibrations. 9. Start collecting data by clicking the red [(click to run)] and [LOG DATA] button, then select where to save the data. zero the relative frequency by holding the [VALUE] key on the front of the SRSQCM200.

10. Configure the Gamry chronopotentiometry experiment in the following way Pre Step Current: 0 A Pre Step Delay: 10 s Step 1 Current: - 0.001 A Step 1 Time : 50 s Step 1 Current: + 0.001 A Step 1 Time : 50 s Sample Period: 0.1 s Upper Voltage: 1.0 V Lower Voltage : - 1.0 V Advanced Pstat Setup : Another window will open, select the following options CA Speed : Norm Ich: 5Hz Vch: 5Hz Ach : BNC 11. Start Experiment and collect data. the frequency should begin to decrease as Cu is deposited. This can be seen in the srsqcm200 program. 4.Data analysis and Work-up 1. Open the data in EChem Analyst and add the data from the auxiliary channel (Ach) to the dataset. This data is the voltage from the QCM which can be used to calculate the change in frequency at each point in time. The current data is used to calculate the theoretical mass change based on the coulombs passed. Again notice that a more positive voltage means an increase in mass, which is accompanied by a the passage of a negative current (cathodic). This negative current causes Cu deposition on the surface of the electrode.

4.Example Data 0.8 ma 0.5 ma 1.0 ma 1.0 ma 0.8 ma 0.5 ma

Data taken from the deposition at 1 ma Electrode Area - 1.38 cm 2 References 1) Hillman, A. R. The EQCM: electrogravimetry with a light touch, J. Solid State Electrochem., 15, 2011, 1647-1660. 2) Buttry, D. A. The Quartz Cystal Microbalance as an In Situ Tool for Electrochemistry, in Electrochemical Interfaces. Modern Techniques for In-Situ Interface Characterization, H. D. Abruña, Ed., VCH Publishers, Inc., New York, 1991, 531-66. Hillman, A. R. The Electrochemical Quartz Crystal Microbalance. In Instrumentation and Electroanalytical Chemistry; Bard, A. J., Stratmann, M., Unwin, P. R., Eds.; Encyclopedia of Electrochemistry; Wiley: New York, 2003; Vol. 3, 230-289. 3) Glassford, A. P. M. J. Vac. Sci. Technol. 1978, 15, 1836. Kanazawa, K. K.; Gordon II, J. Anal. Chem. 1985, 57, 1770. Kanazawa, K. K.; Gordon II, J. Analytica Chimica Acta 1985, 175, 99-105. 4) Sauerbrey, G. Z. Phys. 1959, 155, 206. 5) Muramatsu, H.; Tamiya, E.; Karube, I. Anal. Chem. 1988, 60, 2142-2146. Hillman, A. R.; Jackson, A.; Martin, S. J. Anal. Chem. 2001, 73, 540-549. Martin; Hillman; Etchnique 6) Gamry application note https://www.gamry.com/application-notes/physechem/basics-of-a-quartz-crystal-microbalance/