Retinal stray light originating from intraocular lenses and its effect on visual performance van der Mooren, Marie Huibert

Similar documents
Retinal stray light originating from intraocular lenses and its effect on visual performance van der Mooren, Marie Huibert

Retinal stray light originating from intraocular lenses and its effect on visual performance van der Mooren, Marie Huibert

The design is distinctive. The outcomes are clear. Defocus tolerance 1 Glistening-free performance 1,2 Predictable outcomes 1

COMPARISON OF THE MEDICONTUR 860FAB

Improving Lifestyle Vision. with Small Aperture Optics

Choices and Vision. Jeffrey Koziol M.D. Thursday, December 6, 12

NOW. Approved for NTIOL classification from CMS Available in Quar ter Diopter Powers. Accommodating. Aberration Free. Aspheric.

SEE BEYOND WITH FULLRANGE OPTICS. Developed by Hanita Lenses

Choices and Vision. Jeffrey Koziol M.D. Friday, December 7, 12

Raise your expectations. Deliver theirs.

Maximum Light Transmission. Pupil-independent Light Distribution. 3.75D Near Addition Improved Intermediate Vision

Glistening-Free Hydrophobic Acrylic IOL. Glistening-Free Hydrophobic Acrylic IOL

Long-term quality of vision is what every patient expects

*Simulated vision. **Individual results may vary and are not guaranteed. Visual Performance When It s Needed Most

Treatment of Presbyopia during Crystalline Lens Surgery A Review

NIH Public Access Author Manuscript J Refract Surg. Author manuscript; available in PMC 2007 January 8.

Crystalens AO: Accommodating, Aberration-Free, Aspheric Y. Ralph Chu, MD Chu Vision Institute Bloomington, MN

FOR PRECISE ASTIGMATISM CORRECTION.

Multifocal Intraocular Lenses for the Treatment of Presbyopia: Benefits and Side-effects

University of Groningen. Young eyes for elderly people van Gaalen, Kim

Ocular Scatter. Rayleigh Scattering

Evolution of Diffractive Multifocal Intraocular Lenses

Wide Diopter Range and Different Haptic Platforms

Retinal stray light originating from intraocular lenses and its effect on visual performance van der Mooren, Marie Huibert

Retinal stray light originating from intraocular lenses and its effect on visual performance van der Mooren, Marie Huibert

The Appearance of Images Through a Multifocal IOL ABSTRACT. through a monofocal IOL to the view through a multifocal lens implanted in the other eye

Energy balance in apodized diffractive multifocal intraocular lenses

Unique Aberration-Free IOL: A Vision that Patients

Dr. Magda Rau Eye Clinic Cham, Germany

Sulcoflex. For when perfection is the only option! Pseudophakic Sulcus Fixated Secondary IOLs. Sulcoflex Aspheric. Sulcoflex Toric

The Aberration-Free IOL:

Product Portfolio. Sulcoflex Pseudophakic Supplementary IOLs. Your skill. Our vision.

IOL Types. Hazem Elbedewy. M.D., FRCS (Glasg.) Lecturer of Ophthalmology Tanta university

Patient information. Your options for cataract treatment Enjoy clear vision at all distances with multifocal IOLs

Clinical Evaluation 3-month Follow-up Report

PROGRESSIVE VISION WITHIN FULL ACCOMMODATIVE RANGE

Clinical Update for Presbyopic Lens Options

Design of a Test Bench for Intraocular Lens Optical Characterization

Product Portfolio. Sulcoflex Pseudophakic Supplementary IOLs

HOYA aspherical IOL with ABC (Aspheric Balanced Curve) Design

day night convinced supreme contrast sensitivity THE IOL FOR DAY & NIGHT

EDoF IOL. ZEISS AT LARA 829MP Next generation Extended Depth of Focus Intraocular Lens. NEW EDoF IOL from ZEISS

Customized intraocular lenses

Visual Outcomes of Two Aspheric PCIOLs: Tecnis Z9000 versus Akreos AO

Quality of Vision With Multifocal Progressive Diffractive Lens: Two-Year Follow-up

RayOne Hydrophobic IOL. New design. New standard MADE IN UK

Correlation of pupil size with visual acuity and contrast sensitivity after implantation of an apodized diffractive intraocular lens

Wavefront Aberrations in Eyes With Acrysof Monofocal Intraocular Lenses

REVISITING POTENTIAL HAZARD OF LED SOURCES TO CAUSE BLH IN SPECIFIC POPULATION

ROTATIONAL STABILITY MAKES THE DIFFERENCE

THE XTRAFOCUS IS AN ELEGANT SOLUTION TO COMPLEX CASES.

Quality Testing of Intraocular Lenses. OptiSpheric IOL Family and WaveMaster IOL 2

Citation for published version (APA): Franssen, L., & Coppens, J. E. (2007). Straylight at the retina : scattered papers

Customized Correction of Wavefront Aberrations in Abnormal Human Eyes by Using a Phase Plate and a Customized Contact Lens

FOR EXCELLENT VISION QUALITY TO SUPPORT AN ACTIVE LIFESTYLE

AT LISA tri 839MP and AT LISA tri toric 939MP from ZEISS The innovative trifocal IOL concept providing True Living Vision to more patients

Special Publication: Ophthalmochirurgie Supplement 2/2009 (Original printed issue available in the German language)

Chapter 25. Optical Instruments

WHY EDOF INTRAOCULAR LENSES? FOR EXCELLENT VISION QUALITY TO SUPPORT AN ACTIVE LIFESTYLE PATIENT INFORMATION. Cataract treatment

Comparison of higher order aberrations with spherical and aspheric IOLs compared to normal phakic eyes

Comparison of contrast sensitivity and color discrimination after clear and yellow intraocular lens implantation

Quality Testing of Intraocular Lenses. OptiSpheric IOL Family and WaveMaster IOL 2

NEW THE WORLD S FIRST AND ONLY SINUSOIDAL TRIFOCAL IOL

NEW. AT LISA tri 839MP and AT LISA tri toric 939MP from ZEISS The innovative trifocal IOL concept providing True Living Vision to more patients

Multifocal and Accommodative

ASPIRA PUBLICATION LIST

Accommodating IOL s History and Clinical Management

Prospective sual evaluation of apodized diffractive intraocular lenses

The Blueprint for Improved Image Quality

Maximising Treatment Outcomes with Premium IOL Technology. Saturday 13 September 2014 XXXII Congress of the ESCRS London, UK.

Effects of intraocular lenses with different diopters on chromatic aberrations in human eye models

CLINICAL SCIENCES. Corneal Optical Aberrations and Retinal Image Quality in Patients in Whom Monofocal Intraocular Lenses Were Implanted

Contrast sensitivity in the presence of a glare light. Theoretical concepts and preliminary clinical studies. L.-E. Paulsson and J.

Impact of scattering and spherical aberration in contrast sensitivity

7 DO IT. A SAFER WAY TO TRIFOCALITY * : ELEVATED PHASE SHIFT (EPS) ** 7 DIFFRACTIVE RINGS FOR OPTIMAL LIGHT DISTRIBUTION AND LESS DISTURBANCE

In vitro comparative optical bench analysis of a spherical and aspheric optic design of the same IOL model

Design and qualification of a diffractive trifocal optical profile for intraocular lenses

Repeatability of measurements with a double-pass system

Aberrations Before and After Implantation of an Aspheric IOL

INTRODUCTION THIN LENSES. Introduction. given by the paraxial refraction equation derived last lecture: Thin lenses (19.1) = 1. Double-lens systems

OptiSpheric IOL. Integrated Optical Testing of Intraocular Lenses

PHYSICS. Chapter 35 Lecture FOR SCIENTISTS AND ENGINEERS A STRATEGIC APPROACH 4/E RANDALL D. KNIGHT

Multifocal Progressive Diffractive Lens to Improve Light Distribuition and Avoid Light Loss: Two Years Clinical Results

Vision for What They Love.

Assessing Visual Quality With the Point Spread Function Using the NIDEK OPD-Scan II

Refractive Surgery: Vance Thompson, MD, FACS Refractive Surgeon. Oculeve Wavetec Zeiss Mynosys LRG Equinox Precision Lens ORA Amaken EXCELLens

In Vitro Strehl Ratios with Spherical, Aberration-Free, and customized spherical aberration-correcting METHODS

American National Standard for Ophthalmics. Extended Depth of Focus Intraocular Lenses

WaveMaster IOL. Fast and accurate intraocular lens tester

Physics 208 Spring 2008 Lab 2: Lenses and the eye

Not everyone can do this. Introducing RayOne with patented Lock & Roll TM technology for the smallest fully preloaded IOL incision

Section 2 concludes that a glare meter based on a digital camera is probably too expensive to develop and produce, and may not be simple in use.

Clinical Trials of Exterior Non Implanted Interference-Based Extended Depth of Focus Intra Ocular Lens Design

Abetter understanding of the distribution of aberrations in

Postoperative Wavefront Analysis and Contrast Sensitivity of a Multifocal Apodized Diffractive IOL (ReSTOR) and Three Monofocal IOLs

Applications of Optics

Removal of the crystalline lens followed by implantation of a

Forget Most Everything! The Surgical Management of Presbyopia 2/23/2016. Refraction vs. Diffraction. Presbyopic IOL s Patient Expectations

Optics of the crystalline lens and accommodative response

4th International Congress of Wavefront Sensing and Aberration-free Refractive Correction ADAPTIVE OPTICS FOR VISION: THE EYE S ADAPTATION TO ITS

Transcription:

University of Groningen Retinal stray light originating from intraocular lenses and its effect on visual performance van der Mooren, Marie Huibert IMPORTANT NOTE: You are advised to consult the publisher's version (publisher's PDF) if you wish to cite from it. Please check the document version below. Document Version Publisher's PDF, also known as Version of record Publication date: 2016 Link to publication in University of Groningen/UMCG research database Citation for published version (APA): van der Mooren, M. H. (2016). Retinal stray light originating from intraocular lenses and its effect on visual performance [Groningen]: Rijksuniversiteit Groningen Copyright Other than for strictly personal use, it is not permitted to download or to forward/distribute the text or part of it without the consent of the author(s) and/or copyright holder(s), unless the work is under an open content license (like Creative Commons). Take-down policy If you believe that this document breaches copyright please contact us providing details, and we will remove access to the work immediately and investigate your claim. Downloaded from the University of Groningen/UMCG research database (Pure): http://www.rug.nl/research/portal. For technical reasons the number of authors shown on this cover page is limited to 10 maximum. Download date: 16-10-2018

Impact of intraocular lens material and design on light scatter: In vitro study Reprinted from Journal of Cataract and Refractive Surgery, Vol. 40, Langeslag MJM, van der Mooren M, Beiko GHH, Piers PA Explanted multifocal intraocular lenses, Pages 2120 2127, Copyright 2014, with permission from ASCRS and ESCRS (Elsevier) AMO Groningen BV, Netherlands (Langeslag, van der Mooren, Piers) St. Catharines, Canada (Beiko) http://dx.doi.org/10.1016/j.jcrs.2014.10.017

PURPOSE: To determine the typical in vitro straylight levels for intraocular lenses (IOLs) of different materials and designs. SETTING: Abbott Medical Optics, Inc., Groningen, the Netherlands. DESIGN: Experimental study. METHODS: Two optical bench setups were used to determine baseline straylight levels of IOLs placed in a saline-filled cuvette: one for forward scatter positions between 0.6 and 3.0 degrees and one for positions up to 22.0 degrees. Line-spread functions were measured using the small angle setup, and scattered light intensity was measured using the wide-angle setup. From these measurements, the angular dependent straylight parameter was calculated. Ten IOLs of different materials (hydrophobic and hydrophilic) and designs (monofocal or diffractive multifocal and spheric or aspheric) were studied, and their measured straylight levels were compared with the levels in a 20-year-old and a 70-year-old healthy noncataractous human crystalline lens. RESULTS: Irrespective of the material or design, monofocal IOLs had straylight levels below or close to those of a 20-year-old human crystalline lens. Diffractive multifocal IOLs had straylight levels higher than those of monofocal IOLs but less than those of a 70-year-old human crystalline lens. With increasing angle, hydrophobic IOLs showed a gradual decrease in straylight level. After an initial decrease, hydrophilic IOLs showed an increase in straylight level for larger angles. CONCLUSIONS: The baseline straylight levels of IOLs were design and material dependent (hydrophobic < hydrophilic; monofocal < diffractive multifocal). Most monofocal IOLs had straylight levels below the levels in a 20-year-old human crystalline lens. 82

Impact of intraocular lens material and design on light scatter: In vitro study Imperfections in ocular media cause light entering the eye to be scattered. Natural imperfections can be irregularities that occur at the interface of ocular media (eg, cornea to aqueous) or particles, such as proteins, that are present in the medium. This causes part of the light, initially directed toward the focus point on the retina, to disperse onto the surrounding retina. Consequently, less light will reach the focal point on the retina. The scattered light, or straylight, thus diminishes the contrast of the optical image formed. 1,2 The natural crystalline lens also contributes to the scattering of light. In young crystalline lenses, the amount of scatter is minimal, but as the natural lens ages, the amount of straylight originating from the crystalline lens increases because of the natural aging process of the lens. 3 6 The increased straylight causes visual symptoms such as disability glare during night driving and hindrance from low sun during daytime. 2 Cloudy lenses or cataracts are the extreme of the aging process; light passing through a cataractous lens is more scattered, resulting in a significant loss of contrast. 7 Even though straylight may cause as much functional vision loss as visual acuity and the impact of light scattering from cataracts is well understood, current clinical assessment for cataract surgery is primarily based on visual acuity measurements and slitlamp examination of the density of the crystalline lens. Retinal straylight from light scattering due to cataracts may, however, case as much functional vision loss as poor visual acuity. 5,8 11 Additionally, retinal straylight and visual acuity behave independently. 1,5 7,9,12 Therefore, several clinical investigators have advocated the use of an increase in straylight as an indication for cataract surgery. 2,5,8 11 Previous in vivo measurements using the C-Quant straylight meter (Oculus Optikgeräte GmbH) show that pseudophakic eyes have straylight levels comparable to or below that of healthy eyes. 5 This finding demonstrates that the cataract is the largest source of intraocular light scatter and that its extraction and replacement with an intraocular lens (IOL) lowers the amount of straylight and the related visual effects. Commercially available IOLs differ in their material, design, and manufacturing method, all of which are potential sources of light scatter 5,13 ; however, the extent to which they introduce straylight and how that relates to typical straylight levels in healthy young and old eyes have not been well defined. Increased knowledge of the straylight introduced by different types of IOLs would contribute to a broader and better understanding of the visual consequences of IOL designs and materials. This knowledge would also help in managing patient expectations, which is especially important given the increased attention to light scatter and its role in functional vision loss. In this study, in vitro measurements were performed to determine the baseline straylight levels for IOLs differing in design, material, and/or manufacturing method. The obtained results were compared between individual types of IOLs as well as with the straylight levels published for 20-year-old and 70-year-old healthy crystalline lenses. 13 83

MATERIALS AND METHODS The straylight levels of 10 types of IOLs were studied. Only sterile IOLs were used. The IOLs were made from 4 acrylic materials labeled as hydrophobic and 2 acrylic materials labeled as hydrophilic. They had 4 optical designs: spheric or aspheric and monofocal or diffractive multifocal. Table 1 summarizes the details of the IOLs and the nomenclature used to address them in this article (acrylic A to F). 14,15,A The IOLs' straylight was measured in 2 laboratory-based setups, together covering an angular domain from the optical axis up to 22 degrees. A short overview of both setups is given below. A more detailed description of the measurements and their background has been given by van der Mooren et al. 13 Straylight The straylight parameter s(q) introduced by van der Berg 4 allows quantification of the amount of straylight, both in vivo and in vitro. By definition, the straylight parameter is the equivalent luminance divided by the illuminance of the glare source falling into the eye x angle square. 16 The straylight corresponds to the skirts of the point-spread function (PSF) as presented in equation 1. 16 s(θ) = PSF(θ)xθ 2 [degrees 2 / sr]. (1) Setup 1: Small Angle (up to 3.0 Degrees) Figure 1. Small-angle measurement setup. The first setup, to measure scatter for small angles, is shown in Figure 1. 13 After being hydrated in saline for at least 24 hours, the IOLs were placed in a saline-filled cuvette and 84

Impact of intraocular lens material and design on light scatter: In vitro study placed on an optical bench in an eye model that reproduces the spherical (c [4.0] = 0.27 mm) and chromatic (0.4 mm between hydrogen F and C lines) aberration of an average pseudophakic eye 17,B using white light and a 4.0 mm aperture. A 3.0 mm slit, corresponding to 0.5 degree, was used to collect information on the outer skirts of the PSF. 13 The light intensity distribution of each IOL was obtained using a stepwise lateral displacement of the charge coupled device (CCD) camera, which captured the linespread function. Using 5 steps, an intensity range of 6 orders of magnitude over an angle of ±3 degrees with respect to the optical axis was obtained (Figure 2). For each position, a different shutter time was applied to cover the maximum light intensity range for that particular position. The PSF was obtained by integrating the line-spread function along the image slit. The angular position was calculated from the pixel position, the magnification, and the nodal point. From the PSF and the angular positions, the angular dependent straylight parameters, s(θ), for forward scatter positions between 0.6 and 3.0 degrees were calculated using equation 1. 13,16 Figure 2. Five line-spread function measurements represented by 5 colors were stitched together to obtain the line-spread function over the entire angular range. The lateral displacement of the CCD with respect to the central optical axis is denoted in the legend for each line-spread function recording. 85

Setup 2: Wide Angle (up to 22.0 Degrees) Figure 3. Wide-angle measurement setup. The wide-angle setup is shown in Figure 3. 13 After the IOLs were hydrated in saline for at least 24 hours, the straylight was measured by placing the IOLs in a saline-filled cuvette. A light source was defined and directed using several lenses and filters. The intensity of the light transmitted by the IOL was measured using a CCD camera at different angles relative to the incident beam. The CCD camera and the cuvette were shielded by a black curtain. Similar to the small-angle setup, the camera's shutter times were increased with increasing measurement angle to detect the decreasing light intensity. 13 All analyses were performed using a 4.0 mm aperture. The PSF was calculated from the measured light intensities, light power P(θ), using equation 2, in which Ω represents the solid angle and P 0 is the total light transmitted by the IOL. 13 PSF(θ) = P(θ) / (Ωx P 0 ) (4) Subsequently, the straylight parameter s(θ) was calculated from the PSF and the forward scatter angle using equation 1. 13 Data Analysis The straylight levels for the IOL designs and materials were compared with each other and with the published levels for healthy noncataractous 20- and 70-year-old human crystalline lenses. 13 By definition, the setup measures the tail of the line-spread function; however, the straylight parameter s(θ) is calculated from the PSF. Because in the smallangle method the linespread function approximates the PSF for angles greater than 0.5 degree, results are presented for 0.6 degree outward. For angles smaller than 0.6 degree, no valid PSF values could be measured because of the use of a 0.5 degree slit target. For 86

Impact of intraocular lens material and design on light scatter: In vitro study both measurement setups, the measurements were considered to be symmetrical around the optical axis. Therefore, the results from the negative and positive angles were averaged. The results from the small-angle and wide-angle measurements were combined for each IOL. One IOL was measured for each model listed in Table 1. For 4 IOL models, straylight from at least 2 different IOLs was measured. For each of the IOL models, the maximum standard deviation was 5% of the mean straylight values for the complete angular range. Based on this outcome, it was considered reasonable to use the measurements on 1 IOL for each model as a representable baseline. Table 1. Characteristics of the IOLs in the study. Description Labeled Material Refractive index* Water content IOL Model IOL Design Manufacturer Acrylic A Hydrophobic 1.47 (at 35 C) ~2% 14,15 AAB00 Monofocal/spherical Abbott Medical ZCB00 Monofocal/aspheric Optics, Inc. ZMB00 Multifocal/aspheric Acrylic B Hydrophobic 1.55 ~1% 14,15 SN60AT Monofocal/spherical Alcon, Inc. SN60WF Monofocal/aspheric SN60D3 Multifocal/spherical Acrylic C Hydrophobic 1.540 (at 35 C) ~5% 14 MX60 Monofocal/aspheric Bausch & Lomb, Inc. Acrylic D Hydrophobic 1.520 (at 23 C) ~1% 15 FY-60AD Monofocal/aspheric Hoya Corp. Acrylic E Hydrophilic 1.459 (hydrated, at 20 C) 26% A ADAPT-AO Monofocal/aspheric Bausch & Lomb, Inc. Acrylic F Hydrophilic 1.460 26%* Softec-HD Monofocal/bi-aspheric Lenstec, Inc. *Source: Directions for use RESULTS Material Comparison The straylight levels for the 4 monofocal aspheric IOLs made from materials labeled as hydrophobic acrylic were well below those of the straylight levels of a 20-year-old human crystalline lens. For small forward scatter angles in the range of 0.6 to 3.0 degrees, straylight levels of the IOLs were similar to each other and decreased with increasing angle (Figure 4). For larger angles, the curve representing the straylight levels for acrylics A and B tended to flatten out; however,the curves for acrylics C and D showed a small increase in straylight levels as the forward scatter angle increased from 4 to 20 degrees. In Figure 5, the straylight levels for IOLs made from materials labeled as hydrophilic acrylic are presented. For small forward light scatter angles (up to 3.0 degrees), straylight levels were well below the levels of a 20-year-old human crystalline lens. For angles larger than 3.0 degrees, the straylight levels increased, reaching that of a 20-year-old human crystalline lens at forward scatter angles of approximately 20.0 degrees. 87

Figure 4. Material comparison between monofocal aspheric IOLs made from 4 different hydrophobic acrylic materials (acrylica to D). Figure 5. Material comparison between monofocal aspheric IOLs made from 2 different hydrophilic materials (acrylics E and F). 88

Impact of intraocular lens material and design on light scatter: In vitro study Design Comparison To determine whether design-related differences influenced straylight, 1 spheric and 1 aspheric monofocal IOL made from hydrophobic acrylics A and B were compared (Figure 6). All the spheric and aspheric monofocal IOLs showed decreasing straylight levels for the small forward scatter angles, with the overall levels remaining well below those of a 20- year-old human crystalline lens. In general, the aspheric monofocal IOLs had slightly lower straylight levels than the spheric monofocal IOLs. Figure 7 compares the straylight levels for monofocal and diffractive multifocal IOLs made from hydrophobic acrylics. Both diffractive multifocal IOLs had a 4.0 diopter addition power. Straylight levels for diffractive multifocal IOLs were higher than those for monofocal IOLs but remained below those of a 70-year-old human crystalline lens. Additionally, for scatter angles greater than 2.0 degrees (aspheric design) or 7.5 degrees (spheric design), straylight levels of diffractive multifocal IOLs decreased to levels below those of a 20-year-old human crystalline lens and flattened out with larger angles. Figure 6. Design comparison of 1 spheric and 1 aspheric monofocal IOL made from acrylics A and B. 89

Figure 7. Design comparison of IOLs made from (a) acrylic A having a monofocal aspheric and a diffractive multifocal aspheric design (top) and (b) acrylic B having a monofocal spheric and a diffractive multifocal spheric design (bottom). 90

Impact of intraocular lens material and design on light scatter: In vitro study DISCUSSION Intraocular lenses were measured in an ideal in vitro situation. The absolute and relative straylight levels measured should therefore be considered as a baseline situation. In vivo, the amount of straylight will be defined not only by the type of IOL, but also by the presence of other light-scattering interfaces such as the capsular bag, striae, and the presence of posterior and/or anterior capsule opacification. These biological aspects as well as material inhomogeneity may vary from person to person and may change over time. Monofocal IOLs had straylight levels below or close to the levels of a 20-year-old human crystalline lens, irrespective of the material or design (spheric or aspheric). Diffractive multifocal IOLs had relatively higher straylight levels, but the levels did not exceed those of a 70-year-old human crystalline lens, and with increasing angle, the levels dropped to less than those of a 20-year-old human crystalline lens. These larger forward scatter angles correspond to the region that is considered to be outside the region affected by the halo that results from a diffractive multifocal design. 18 The differences in straylight levels between monofocal and diffractive multifocal IOL designs can be assigned to the straylight originating from the higher-order foci of the diffractive surface. Removing a cataractous lens and replacing it with an IOL is expected to lower the lens' contribution to straylight in the eye. This is consistent with the observations by van den Berg et al., 5 who compared in vivo straylight levels before and after cataract surgery. Analysis of straylight levels at different forward scatter angles revealed that most of the hydrophobic acrylic IOLs showed a decrease in straylight levels with increasing forward scatter angles until they reached a plateau for the large angles (Figure 4). In contrast, for hydrophilic IOLs, the initial decrease in straylight levels for smaller angles was followed by a similar gradual increase for larger angles (Figure 5). A veil is apparent on the hydrophobic IOLs at large scatter angles; for example Figure 8 shows the photographs taken during the measurements for a hydrophobic and a hydrophilic acrylic IOL from the same manufacturer at a small (1.5 degrees) and a large (22.0 degrees) forward scatter angle. A veil is clearly seen on the hydrophilic IOL at 22.0 degrees forward scatter angle. Water content may also vary among the hydrophobic acrylic IOLs (Table 1). A review of Figures 4 and 5 shows a trend toward increasing straylight parameter values at scatter angles of 3.0 degrees onward for acrylic material C that is similar to that for hydrophilic IOLs. The absolute increase in the straylight parameter value for hydrophobic material C is not as large as that for the hydrophilic IOLs. Material C has the highest water content of all hydrophobic materials tested (Table 1) but not as high as the hydrophilic materials; this may point toward a relationship between the IOL materials' water content and the straylight parameter behavior for large forward scatter angles. Future work may focus on investigating possible mechanisms explaining this observation. In addition to IOL material and design, surface and bulk irregularities arising from the manufacturing process may influence the amount of straylight measured. In this study, IOLs from different 91

manufacturers were used and therefore the manufacturing process may have influenced the straylight levels observed. The eye model used for the small-angle measurements was designed to reproduce the average corneal spherical aberration. Straylight levels of spheric and aspheric IOLs measured in this study were low and their relative differences may not be attributed to spherical aberration because this factor is expected to play a role for angles that are much smaller than 0.6 degrees. The higher straylight values obtained from the small-angle setup measurements in the overlapping angular range for all IOLs except the multifocal spheric lens B (Figure 7) may be explained by a difference in the base straylight level of the measurement setups used in this study. Straylight measurements require the accurate determination of light intensities over a large range of values. In the low intensity regions, accurate straylight measurements become increasingly challenging. This may lead to variation in absolute straylight parameter values when the same IOL is measured under different conditions (eg, on another day) or when the same IOL model from a different manufacturing batch is measured. This is especially true for monofocal IOLs as they showed straylight levels close to the straylight levels of the setup without an IOL. Intraocular lenses were only measured when the straylight value for the setup alone was at least 0.5 log unit below the level of a healthy noncataractous 20-year-old human crystalline lens, as described by the in-depth explanation of this method by van der Mooren et al. 13 Additionally, a repeatability study has been performed on a monofocal and a multifocal IOL with 3 repetitions on each IOL. The maximumstandard deviation was 13% of the mean straylight values for the complete angular range. The results presented in this study can help predict the consequences of a change in IOL design, material, or manufacturing process on the postoperative straylight level in pseudophakic eyes. The baseline levels that we measured can be used to inform patients of the expected straylight level and possible improvement in visual effects. Future work may include in vivo measurements of aged IOLs and/or the comparison of the in vitro straylight levels presented in this paper with the straylight levels from in vivo measurements in pseudophakic eyes. In conclusion, this study found that most monofocal IOLs have straylight levels below the levels published for a healthy noncataractous 20-year-old human crystalline lens. Light scattering is influenced by IOL material and design; hydrophobic IOLs tend to have lower straylight levels than hydrophilic IOLs for scatter angles larger than 3.0 degrees and IOLs with a monofocal IOL design have lower straylight levels than IOLs having a diffractive multifocal design. 92

Impact of intraocular lens material and design on light scatter: In vitro study Figure 8. Acrylic C (left) and acrylic E (right) monofocal IOLs taken in the wide-angle setup at a forward scatter angle of 1.5 degrees (top) and 22.0 degrees (bottom) (white light, in saline). A veil is clearly visible on the hydrophilic IOL at large angles (bottom right). WHAT WAS KNOWN Intraocular light scatter due to cataracts is a known source of functional vision loss. Extraction of the cataractous lens and replacement with an IOL lowers the amount of straylight. WHAT THIS PAPER ADDS In vitro straylight measurements of different IOLs show that baseline straylight levels depend on their design and material (monofocal <diffractive multifocal; hydrophobic < hydrophilic). Irrespective of the material, all monofocal IOLs show in vitro baseline straylight levels below the levels reported for a 20-year-old healthy human crystalline lens. 93

REFERENCES 1. Zuckerman JL, Miller D, Dyes W, Keller M. Degradation of vision through a simulated cataract. Invest Ophthalmol 1973; 12:213 224. Available at: http://www.iovs.org/content/12/3/213.full.pdf.accessed September 1, 2014 2. vandenbergtjtp,franssenl,coppens JE.Ocularmedia clarity and straylight. In: Dartt DA, ed, Encyclopedia of the Eye. Oxford,UK, Academic Press, 2010; 3, 173 183. Available at: http://www.nin.knaw.nl/portals/0/department/berg/documents/vdberga2010encycleye00 230.pdf. Accessed September 1, 2014 3. IJspeert JK, de Waard PWT, van den Berg TJTP, de Jong PTVM. The intraocular straylight function in 129 healthy volunteers; dependence on angle, age and pigmentation. Vision Res 1990; 30:699 707 4. van den Berg TJTP. Analysis of intraocular straylight, especially in relation to age. Optom Vis Sci 1995; 72:52 59 5. van den Berg TJTP, van Rijn LJ, Michael R, Heine C, Coeckelbergh T, Nischler C, Wilhelm H, Grabner G, Emesz M, Barraquer RI, Coppens JE, Franssen L. Straylight effects with aging and lens extraction. Am J Ophthalmol 2007;144:358 363 6. de Waard PWT, IJspeert JK, van den Berg TJTP, de Jong PTVM. Intraocular light scattering in age-related cataracts. Invest Ophthalmol Vis Sci 1992; 33:618 625. Available at: http://www.iovs.org/cgi/reprint/33/3/618.pdf. Accessed September 1,2014 7. van den Berg TJTP. Importance of pathological intraocular light scatter for visual disability. Doc Ophthalmol 1986; 61:327 333 8. van der Meulen IJE, Gjertsen J, Kruijt B, Witmer JP, Rulo A,Schlingemann RO, van den Berg TJTP. Straylight measurements as an indication for cataract surgery. J Cataract Refract Surg 2012; 38:840 848 9. Elliott DB, Hurst MA, Weatherill J. Comparing clinical tests of visual loss in cataract patients using a quantification of forward light scatter. Eye 1991; 5:601 606. Available at: http://www.nature.com/eye/journal/v5/n5/pdf/eye1991104a.pdf. Accessed September 1, 2014 10. Paulsson L-E, Sj ostrand J.Contrast sensitivity in the presence of a glare light; theoretical concepts and preliminary clinical studies. Invest Ophthalmol Vis Sci 1980; 19:401 406. Available at: http://www.iovs.org/cgi/reprint/19/4/401. Accessed September 1, 2014 11. Elliott DB, Bullimore MA. Assessing the reliability, discriminative ability, and validity of disability glare tests. Invest Ophthalmol Vis Sci 1993; 34:108 119. Available at: http://www.iovs.org/cgi/reprint/34/1/108.pdf. Accessed September 1, 2014 12. Hess R, Woo G. Vision through cataracts. Invest Ophthalmol Vis Sci 1978; 17:428 435. Available at: http://www.iovs.org/content/17/5/428.full.pdf. Accessed September 1, 2014 13. van der Mooren M, van den Berg T, Coppens J, Piers P.Combining in vitro test methods for measuring light scatter in intraocular lenses. Biomed Opt Express 2011; 2:505 510. Available at: http://www.ncbi.nlm.nih.gov/pmc/articles/ PMC3047356/pdf/505.pdf. Accessed September 1, 2014 14. Bozukova D, Pagnoulle C, Jérôme C. Biomechanical and optical properties of 2 new hydrophobic platforms for intraocular lenses. J Cataract Refract Surg 2013; 39:1404 1414 94

Impact of intraocular lens material and design on light scatter: In vitro study 15. Werner L. Glistenings and surface light scattering in intraocular lenses. J Cataract Refract Surg 2010; 36:1398 1420 16. Vos JJ, van den Berg TJTP. Report on disability glare. CIE Collection 1999. Vision and Colour; Physical Measurement of Light and Radiation. CIE 1999; 135:1 9. Available at: http://www.nin.knaw.nl/portals/0/department/berg/documents/vos_vdb1999.pdf. Accessed September 1, 2014 17. Norrby S, Piers P, Campbell C, van der Mooren M. Model eyes for evaluation of intraocular lenses. Appl Opt 2007; 46:6595 6605 18. Pieh S, Lackner B, Hanselmayer G, Z ohrer R, Sticker M,Weghaupt H, Fercher A, Skorpik C. Halo size under distance and near conditions in refractive multifocal intraocular lenses. Br J Ophthalmol 2001; 85:816 821. Available at: http://www.ncbi. nlm.nih.gov/pmc/articles/pmc1724058/pdf/v085p00816.pdf. Accessed September 1, 2014 OTHER CITED MATERIAL A. Behndig A, Bellucci R, Pynson J, Brooks L. The Akreos IOLs; clinically proven to provide quality of vision.uspublication,april 2009. Available at: http://www.bausch.com/portals/109/- /m/bl/united%20states/files/downloads/ecp/surgical/the-akreos-iols.pdf. Accessed September 1, 2014 B. van dermoorenm,weeberh, Piers P.Verification of the average cornea eye ACE model. IOVS 2006; 47:ARVO E-abstract 309. Available at: http://abstracts.iovs.org//cgi/content/abstract/47/5/309?sidz7a0d8347-7f43-4724-97bcb0d09298fbca. Accessed September 1, 2014 95