R I T. Title: Wyko RST Plus. Semiconductor & Microsystems Fabrication Laboratory Revision: A Rev Date: 05/23/06 1 SCOPE 2 REFERENCE DOCUMENTS

Similar documents
Bruker Optical Profilometer SOP Revision 2 01/04/16 Page 1 of 13. Bruker Optical Profilometer SOP

Title: Amray 1830 SEM#2 Semiconductor & Microsystems Fabrication Laboratory Revision: D Rev Date: 03/18/2016

Advanced Technology and Manufacturing Institute. Zygo ZeScope

Standard Operating Procedure

KEYENCE VKX LASER-SCANNING CONFOCAL MICROSCOPE Standard Operating Procedures (updated Oct 2017)

FRAUNHOFER AND FRESNEL DIFFRACTION IN ONE DIMENSION

PANalytical X pert Pro Gazing Incidence X-ray Reflectivity User Manual (Version: )

GlassSpection User Guide

Zygo Optical Interferometer SOP

Adobe Photoshop. Levels

Figure 1 The Raith 150 TWO

Stitching MetroPro Application

MEASUREMENT CAMERA USER GUIDE

JEOL 6500 User Manual

Technical Note How to Compensate Lateral Chromatic Aberration

Digital Portable Overhead Document Camera LV-1010

Difrotec Product & Services. Ultra high accuracy interferometry & custom optical solutions

RENISHAW INVIA RAMAN SPECTROMETER

Things to check before start-up.

Olympus LEXT OLS 4000 Confocal Laser Microscope

Supplemental Figure 1: Histogram of 63x Objective Lens z axis Calculated Resolutions. Results from the MetroloJ z axis fits for 5 beads from each

Instruction Manual for HyperScan Spectrometer

Nikon Capture NX "How To..." Series

Leica DMi8A Quick Guide

Guidance on Using Scanning Software: Part 5. Epson Scan

Characterization Microscope Nikon LV150

How to combine images in Photoshop

Renishaw InVia Raman microscope

Zeiss AxioImager.Z2 Brightfield Protocol

ISCapture User Guide. advanced CCD imaging. Opticstar

You can easily print images using the Capture NX print function. Here we will explain the process for printing

Zeiss LSM 780 Protocol

of surface microstructure

Kigamo Scanback which fits in your view camera in place of conventional film.

Using the Hitachi 3400-N VP-SEM

ImagesPlus Basic Interface Operation

FEI Tecnai G 2 F20 Operating Procedures

Contents STARTUP MICROSCOPE CONTROLS CAMERA CONTROLS SOFTWARE CONTROLS EXPOSURE AND CONTRAST MONOCHROME IMAGE HANDLING

EVOS M5000 Imaging System

DAVID SLS-1. Getting Started Guide

Minolta Scanner Plugin

The operation manual of spotlight 300 IR microscope

Check that the pneumatic hose is disconnected!!!! (unless your using the BSE detector, of course)

Bruker Dimension Icon AFM Quick User s Guide

Axioscan - Startup. 1. Turn on the Axioscan (button to the left) and turn on the computer. 2. Log on and start the ZEN Blue software from the desktop

KNOW YOUR CAMERA LEARNING ACTIVITY - WEEK 9

SUPRA Optix 3D Optical Profiler

Application Note #548 AcuityXR Technology Significantly Enhances Lateral Resolution of White-Light Optical Profilers

Scanning Electron Microscope FEI INSPECT F50. Step by step operation manual

Color and More. Color basics

CAPTURING IMAGES ON THE HIGH-MAGNIFICATION MICROSCOPE

Horiba LabRAM ARAMIS Raman Spectrometer Revision /28/2016 Page 1 of 11. Horiba Jobin-Yvon LabRAM Aramis - Raman Spectrometer

Training Guide for Leica SP8 Confocal/Multiphoton Microscope

MIF ZEISS VIOLET CONFOCAL ZEN 2009 PROTOCOL

Bruker Dimension Icon AFM Quick User s Guide

ISIS A beginner s guide

Installation & User Manual Micro-Image Capture 7

MIF ZEISS LSM510 CONFOCAL USER PROTOCOL

Operating Checklist for using the Laser Scanning Confocal Microscope. Leica TCS SP5.

Before you start, make sure that you have a properly calibrated system to obtain high-quality images.

Large Field of View, High Spatial Resolution, Surface Measurements

3D Scanning Guide. 0. Login. I. Startup

Information & Instructions

Micro-Image Capture 8 Installation Instructions & User Guide

Digital Director Troubleshooting

Be aware that there is no universal notation for the various quantities.

Standard Operating Procedure for the Amray 1810 Scanning Electron Microscope Version: 29 NOVEMBER 2014

Zeiss LSM 880 Protocol

Table of Contents 1. Image processing Measurements System Tools...10

University of MN, Minnesota Nano Center Standard Operating Procedure

Lab Report 3: Speckle Interferometry LIN PEI-YING, BAIG JOVERIA

Using the Advanced Sharpen Transformation

Camera Club of Hendersonville

1.1. In regular TEM imaging mode, find a region of interest and set it at eucentric height.

Brightfield Microscopy and Image Acquisition on Spotcam1. by Ryan Taylor/Nancy Kleene Last modified 10/02/05 by Birgit Ehmer

Workflow for Betterlight Imaging

Operating Rausch ScanCam within POSM.

Olympus Digital Microscope Camera (DP70) checklist

1. What is SENSE Batch

User Guide for TWAIN / DirectX interface for GRYPHAX USB 3.0 cameras

User Manual for HoloStudio M4 2.5 with HoloMonitor M4. Phase Holographic Imaging

GXCapture 8.1 Instruction Manual

MAKE SURE YOUR SLIDES ARE CLEAN (TOP & BOTTOM) BEFORE LOADING DO NOT LOAD SLIDES DURING SOFTWARE INITIALIZATION

Procedures for Performing Cryoelectron Microscopy on the FEI Sphera Microscope

WITec Alpha 300R Quick Operation Summary October 2018

MC3 Motion Control System Shutter Stream Quickstart

OCULUS VR, LLC. Oculus User Guide Runtime Version Rev. 1

Everest System / Slidebook Operating Procedures

INTERFEROMETER VI-direct

OPERATION OF THE HITACHI S-450 SCANNING ELECTRON MICROSCOPE. by Doug Bray Department of Biological Sciences University of Lethbridge

DOCUMENT SCANNER INSTRUCTIONS. Space. Backup. Count Only. New File. Scanner. Feeding Option Manual Auto Semi-Auto

The ideal K-12 science microscope solution. User Guide. for use with the Nova5000

WORKFLOW GUIDE. Trimble TX8 3D Laser Scanner Camera and Nodal Ninja R1w/RD5 Bracket Kit

1 Set up the confocal light path for imaging a green dye (Alexa488-EGFP). For example, the

This document assumes the user is already familiar with basic operation of the instrument in TEM mode and use of the Microscope Control interface.

Therefore, all descriptions and illustrations in this instruction manual, including all specifications are subject to change without notice.

CHAPTER1: QUICK START...3 CAMERA INSTALLATION... 3 SOFTWARE AND DRIVER INSTALLATION... 3 START TCAPTURE...4 TCAPTURE PARAMETER SETTINGS... 5 CHAPTER2:

PANalytical X pert Pro High Resolution Specular and Rocking Curve Scans User Manual (Version: )

Contents Technical background II. RUMBA technical specifications III. Hardware connection IV. Set-up of the instrument Laboratory set-up

Standard Operating Procedure of Atomic Force Microscope (Anasys afm+)

Transcription:

Approved by: Process Engineer / / / / Equipment Engineer 1 SCOPE The purpose of this document is to detail the use of the Wyko RST Plus. All users are expected to have read and understood this document. It is not a substitute for in-person training on the system and is not sufficient to qualify a user on the system. Failure to follow guidelines in this document may result in loss of privileges. 2 REFERENCE DOCUMENTS o RST Plus Operator s Guide o Wyko Surface Profilers Technical Reference Manual 3 DEFINITIONS n/a 4 TOOLS AND MATERIALS 4.1 General Description The Wyko is a surface profiler that uses optical interference techniques to perform a non-contact analysis of a sample. The software is version 1.8. 4.2 Capabilities 4.2.1 PSI Mode The PSI mode uses phase shifting interferometry, has a maximum height resolution of 1600Å between adjacent pixels and is intended for smooth surfaces. In this mode the optical path length of the test and reference beams is modified causing a shift in the fringes. The shift in fringes in monitored by a detector. This mode uses a red filter. 4.2.2 VSI Mode The VSI mode uses vertical scanning interferometry, has a maximum height resolution of 500 µm and can be used to measure surfaces that are rough or that have a lot of surface topography. In this mode the interference fringes using white light are monitored as the focus of the system is varied. This mode is good for most measurements. RIT SMFL Page 1 of 10

5 SAFETY PRECAUTIONS 5.1 Hazards to the Tool 5.1.1 The Wyko is a delicate optical instrument that should be operated carefully. 5.1.2 The Wyko is very sensitive to vibration and must be still during a measurement. 5.1.3 Always make sure that there is enough space between the lens and the sample to prevent damage. 6 INSTRUCTIONS 6.1 Initial State Check 6.1.1 The system is controlled by a computer which is normally left on. If the computer is off it may be restarted under the table. 6.1.2 Turn on the camera monitor. 6.1.3 Verify the air table is working properly. If the air table is not floating, have a staff member adjust it. 6.2 Resetting the System 6.2.1 Occasionally the computer will lock up and need to be restarted. 6.2.2 If you ever need to stop a measurement, press the Escape key. 6.3 Sample Setup 6.3.1 Start the Vision for RST Plus software for the system. 6.3.2 Place the sample on the stage and select the proper magnification lens using care not to run into the sample. 6.3.3 There is a selector ring on the front of the Wyko that may be used to further vary the magnification. 6.3.4 Make sure the filter is in the correct position. VSI mode uses white light and PSI mode uses red light. RIT SMFL Page 2 of 10

6.3.5 The sample illumination is controlled through the software. Select Hardware and then Intensity. Adjust the illumination manually or use the Auto Intensity button. The image on the screen should appear bright but not red. If you see red areas, then back off the intensity some. RIT SMFL Page 3 of 10

6.3.6 Adjust the focus using care not to hit the sample which could damage the lens. The sample is in focus when the interference fringes are visible. This can be viewed on the video monitor or the intensity screen on the computer. It may be easier to find the fringes using the red filter. For course focus you may close the field and aperture stops and focus on the resulting polygon. Make sure to open them before taking a measurement. 6.3.7 Adjust the tilt using the tilt knobs on the front and left side of the stage. If there is too much tilt there will be many small fringes. As the tilt is reduced, the fringes will broaden and rotate. The fringes should be as broad as possible with the darkest fringe centered. It may also be necessary to adjust the fine focus while you are adjusting the tilt. If the intensity still looks good, you can close the window by selecting OK. Make sure that the fringes are on the upper surface of the steps. RIT SMFL Page 4 of 10

6.4 Measurement setup for PSI mode 6.4.1 Verify that the red filter is in place. 6.4.2 Select Hardware and then Measurement Options. 6.4.3 Select PSI Mode. 6.4.4 Select Options. 6.4.5 Enter a Modulation Threshold. This is used to eliminate data that is outside an acceptable signal to noise ratio. PSI is typically 7-10% 6.4.6 Select a Resolution in pixels. Higher resolutions will increase the measurement time. 6.4.7 Select a data average number to calculate results. The default is 2. 6.4.8 Select an A/F Scanback and a 2 nd Scanback. 6.4.9 To make a measurement select File and then New Measurement. RIT SMFL Page 5 of 10

6.5 Measurement setup for VSI mode 6.5.1 Verify that the filter is positioned to provide white light. 6.5.2 Select Hardware and then Measurement Options. 6.5.3 Select VSI Mode. 6.5.4 Select Options. 6.5.5 Enter a Scan Depth. Use an approximate peak to valley range. 6.5.6 Enter a Modulation Threshold. This is used to eliminate data that is outside an acceptable signal to noise ratio. VSI is typically 2%. 6.5.7 Enter a Scan Back to move the objective up before starting a measurement. 6.5.8 Select a Resolution in pixels. Higher resolutions will increase the measurement time. 6.5.9 Choose a scan speed. High speed is the default. 6.5.10 Select a data average number to calculate results. The default is 2. 6.5.11 To make a measurement select File and then New Measurement. RIT SMFL Page 6 of 10

6.6 Data Analysis 6.6.1 After completion of a measurement the following plot will be displayed. RIT SMFL Page 7 of 10

6.6.2 If a 3D plot is needed, go to Analysis, Processed and select 3D. Right click on the resulting plot to change the view or to zoom in. RIT SMFL Page 8 of 10

6.6.3 If a 2D profile is needed, go to Analysis and select 2D Profile. Use the mouse to move the vertical and horizontal lines to obtain different profiles. Right click for additional options. 6.6.4 Some other options that are available under Analysis include Filtered Histogram, Multiple Region, Step Measurement and Fourier. 6.7 Data Export to another Computer 6.7.1 Press the Print Screen button 6.7.2 From the Start Menu select Programs, Accessories and Paint. 6.7.3 From Paint select Edit and then Paste. 6.7.4 Save your file to 3 ½ Floppy (A:) 6.7.5 Alternately you may right click on a plot and choose to save just the plot to a tiff file. This may result in a higher resolution image. 6.8 System Shutdown 6.8.1 When finished using the system turn down the intensity. Select Hardware, Intensity and then minimize the intensity using the slider bar. 6.4.2 Turn off the video monitor. RIT SMFL Page 9 of 10

6.9 Errors during Run 6.9.1 A measurement may be stopped at any time using the Escape key. 6.6.2 If the system locks up try restarting the computer. 7 APPROPRIATE USES OF THE TOOL 8 ATTACHMENTS REVISION RECORD Original Issue Summary of Changes Originator Rev/Date Sean O Brien A-05/23/06 RIT SMFL Page 10 of 10