Application Note Oxygen Sensor

Similar documents
O2 SS OXYGEN SOLID ELECTROCHEMICAL SENSOR

NITROGEN DIOXIDE NO2 SS SOLID ELECTROCHEMICAL SENSOR

ALPHASENSE USER MANUAL. Toxic Sensor Evaluation Board Issue 4

Signal conditioning for electrochemical sensors. Pierre SENNEQUIER / AAS

Chapter 3 Electronic Circuit for MWCNT Ethylene Sensor

Electrochemical Sensors Application Note 2 Design of Electronics for Electrochemical Gas Sensors

Electrochemical Sensors Application Note 2 Design of Electronics for Electrochemical Gas Sensors

AN4348 Application note

Electrochemical Sensors Application Note 2 Design of Electronics for Electrochemical Gas Sensors

R S / R 100ppb 0.1. Fig. 1: R S /R 0 as a function of gas concentration at 50% RH and 25 C.

ULPSM-NO August 2017

Ultra-Low Power Analog Sensor Module for Sulfur Dioxide

ULPSM-Ethanol

ULPSM-RESPIRR

Photops. Photodiode-Amplifier Hybrids

V CC OUT MAX9945 IN+ V EE

Constant Voltage and Constant Current Controller. Features

Voltage-Current Regulator TLE 4305

New Op Amps TSU111, TSZ182 & P-NUCLEO-IKA02A1

V CC OUT MAX9945 IN+ V EE

OPERATING INSTRUCTIONS AND SYSTEM DESCRIPTION FOR THE TEC-B-01M VOLTAGE CLAMP MODULE FOR EPMS SYSTEMS. VERSION 1.2 npi 2014

Technical Data Sheet. Sensoric 4-20 ma Transmitter Board

CS6A4689 Stereo Headphone Amplifier Evaluation Board (Rev 1)

AP4310A. Pin Assignments. Description. Features. Applications. Typical Applications Circuit. A Product Line of. Diodes Incorporated

Lead Free. (Note 2) Note: 1. RoHS revision Glass and High Temperature Solder Exemptions Applied, see EU Directive Annex Notes 5 and 7.

INSTRUCTION MANUAL FOR CELL SIMULATOR MODEL 2410

X-FPM(4L)/X-FPM(4L)-AR

High-Side Measurement CURRENT SHUNT MONITOR

MAX44248 Evaluation Kit. Evaluates: MAX General Description. Features. Component List. Component Supplier

LM392/LM2924 Low Power Operational Amplifier/Voltage Comparator

Recommended Land Pattern: [mm]

ULPSM-CO Ultra-Low Power Sensor Module - Carbon Monoxide with Printed Sensor

FPM(L)-NIR(1100) Content PRODUCT SPECIFICATION

IS31AP4066D DUAL 1.3W STEREO AUDIO AMPLIFIER. January 2014 KEY SPECIFICATIONS

Low Drop Voltage Regulator TLE

Technical data sheet TDS0074

O 2 SENSORS Zirconium Dioxide (ZrO 2 ) Software & Hardware Design Guide

Analog Technologies. Dual Mode Laser Driver LDA1-CP1-D

2352 Walsh Ave. Santa Clara, CA U. S. A. Tel.: (408) , Fax: (408)

L08P IPV/W/IPVW SERIES

nanopower Op Amp in a Tiny 6-Bump WLP

Maxim Integrated Products 1

Compliance Voltage How Much is Enough?

NE/SA5234 Matched quad high-performance low-voltage operational amplifier

Pin # Pin Name Pin Type Description. 4 GND Signal ground Signal ground pin. Connect ADC and DAC grounds to here.

Current Transducer CTSR 1-P = 1A

DUAL OP AMP AND VOLTAGE REFERENCE General Description. Features

TL494M PULSE-WIDTH-MODULATION CONTROL CIRCUIT


Building a Microcontroller based potentiostat: A Inexpensive and. versatile platform for teaching electrochemistry and instrumentation.

Zahner 3/2011. femto-farad Probe

VM2000. Low-Noise Bottom Port Piezoelectric MEMS Microphone Data Sheet Vesper Technologies Inc. Differential Analog Output

AD596/AD597 SPECIFICATIONS +60 C and V S = 10 V, Type J (AD596), Type K (AD597) Thermocouple,

LP2902/LP324 Micropower Quad Operational Amplifier

Precision, Low Power, Micropower Dual Operational Amplifier OP290

Analytical Chemistry II

Using ir compensation

RV4141A Low-Power, Ground-Fault Interrupter

LD /03/2017. Constant Voltage and Constant Current Controller. Features. General Description. Applications. Typical Application REV.

LF442 Dual Low Power JFET Input Operational Amplifier

Low Drop Voltage Regulator TLE 4274

CGY2107HV CGY2107HV PRODUCT DATASHEET. Dual High Gain Low Noise High IP3 Amplifier. Rev 0.2 FEATURES APPLICATIONS DESCRIPTION

MODEL: 20VS2-02. Hybrid IC Isolation Amplifiers 20 Series

LDS8710. High Efficiency 10 LED Driver With No External Schottky FEATURES APPLICATION DESCRIPTION TYPICAL APPLICATION CIRCUIT

Obsolete Product(s) - Obsolete Product(s)

LM321. Single Channel Operational Amplifier

DUAL OP AMP AND VOLTAGE REFERENCE General Description. Features

LB4310. Dual Op Amp And Voltage Reference. General Description. Features. Applications. LB4310 Rev of /02/03.

Biomedical Electrodes, Sensors, and Transducers. Definition of Biomedical Electrodes, Sensors, and Transducers. Electrode: Sensor: Transducer:

NE/SA/SE532 LM258/358/A/2904 Low power dual operational amplifiers

Thermocouple Conditioner and Setpoint Controller AD596*/AD597*

Qualified for Automotive Applications. Product Validation according to AEC-Q100/101

IC Preamplifier Challenges Choppers on Drift

PolarSpeed -M(L)/PolarSpeed -M(L)-AR


Features. Applications SOT-23-5

TLE4916-1K. Datasheet. Sense & Control. Low Power Automotive Hall Switch. Rev.1.0,

TSL253R LIGHT-TO-VOLTAGE OPTICAL SENSOR

Dual-Axis, High-g, imems Accelerometers ADXL278

LM124/LM224/LM324/LM2902 Low Power Quad Operational Amplifiers

AOZ A Ultra Low Dropout Linear Regulator. Features. General Description. Applications. Typical Application AOZ2023

Introduction to Operational Amplifiers

Product Datasheet P MHz RF Powerharvester Receiver

oxy.iq Panametrics Oxygen Transmitter Applications Features bhge.com Two-wire, loop-powered 4 to 20 ma transmitter An oxygen transmitter for use in:

SKY : Direct Quadrature Demodulator GHz Featuring No-Pull LO Architecture

MAX9633 Evaluation Kit Evaluates: MAX9633

Ultra-Small, Ultra-Thin, 4-Bump Op Amp

RV4145A Low-Power Ground Fault Interrupter

ZXCT1107/1109/1110 LOW POWER HIGH-SIDE CURRENT MONITORS

Precision, High-Bandwidth Op Amp

high performance needs great design. Coverpage: AS89000 Datasheet

IXYS IXI848A. High-Side Current Monitor. General Description. Features: Applications: Ordering Information. General Application Circuit

INTEGRATED CIRCUITS. SA571 Compandor. Product specification 1997 Aug 14 IC17 Data Handbook

Use of ph or glucose probes to diagnose compartment syndrome

= +25 C, Vcc1 = Vcc2 = Vcc3 = +5V

QPB3311SR. 15 db Return Path Amplifier MHz. Product Overview. Key Features. Functional Block Diagram. Applications. Ordering Information

TA75W01FU TA75W01FU. Dual Operational Amplifier. Features Pin Connection (Top View)

AS339/339A. Description. Pin Assignments. Features. Applications LOW POWER LOW OFFSET VOLTAGE QUAD COMPARATORS AS339/339A

Ultraviolet selective thin film sensor TW30DY NEW: Read important application notes on page 4 ff.

Single Supply, Rail to Rail Low Power FET-Input Op Amp AD820

Transcription:

MEM2 Application Note Oxygen Sensor Contents 1)Sensor principle...1 Electrochemical Gas Sensors in General...1 Working Principle of the Membrapor Oxygen-Sensor...1 2)Characteristics of Membrapor Oxygen-Sensor...2 3)Designing a Potentiostatic Circuit for the Oxygen-Sensor...3 4)Connecting the Sensor to the Circuit...4 5)Usage of the Sensor...4 6)Intrinsic Safety Considerations...4 7)Product Safety...4 8)Storage...4 1) Sensor principle Electrochemical Gas Sensors in General Electrochemical sensors operate by reacting with the analyte and producing an electrical signal proportional to the analyte concentration. Electrochemical gas sensors are generally amperometric sensors, generating a current that is linearly proportional to the gas concentration. The principle behind amperometric sensors is the measurement of the current-potential relationship in an electrochemical cell where equilibrium is not established. The current is quantitatively related to the rate of the electrolytic process at the sensing electrode (also known as working electrode) where the reference electrode's potential is constant. At the counter electrode occurs a counter-reaction which balances the reaction of the sensing electrode, thus completing the electrical circuit. The ionic current between the counter and sensing electrode is transported by the electrolyte inside the sensor body, whereas the current path between is provided by wires which are terminated with pin connectors. Working Principle of the Membrapor Oxygen-Sensor Oxygen diffuses through a capillary to reach the surface of the sensing electrode. With this approach the amount of gas entering the sensor is controlled by diffusion. MEM2 Appl. Note Rev. 8/2011 Page 1 of 5

Figure 1 Working principle of the oxygen sensor The oxygen reaching the sensing electrode is then reduced and therefore consuming electrons, which leads to the electric current. The counter electrode balances the reaction of the sensing electrode by oxidizing water. Thereby protons are produced at the counter electrode. These migrate through the electrolyte towards the sensing electrode. A potentiostatic circuit maintains the potential of the sensing electrode at -600 mv with respect to the reference electrode potential. This negative bias potential is necessary to operate this amperometric fuel cell. In opposite to other biased sensors, the Membrapor oxygen sensor is stabilized within a few minutes when the circuitry is switched on. In opposite to galvanic O2-sensors this type has no consumable parts, nor does change anything inside with time. Surrounding oxygen has no impact on storage life and the sensor is RoHS compliant in opposite to lead-based sensors. 2) Characteristics of Membrapor Oxygen-Sensor Lifetime not limited In opposite to galvanic O2-sensors this type has no consumable parts, nor does change anything inside with time Easy storage Storage is unproblematic in opposite to sensors with sacrificial anode RoHS compliant This sensor does not contain any toxic metals in opposite to lead-based sensors Reliable This sensor is based on the same technology as Membrapor`s sensors for toxic gases: High quality gas sensors based on many years of experience. Performance details are given in the data sheet. MEM2 Appl. Note Rev. 8/2011 Page 2 of 5

3) Designing a Potentiostatic Circuit for the Oxygen-Sensor To operate an electrochemical sensor a control circuitry is required, referred to as the potentiostatic circuit. For a 3-electrode sensor the main purpose is to maintain a voltage between the reference electrode (Ref) and the sensing electrode (Sens) to control the electrochemical reaction and to deliver an output signal proportional to the current produced by the sensor. The potential on the counter electrode (Cnt) is not important, so long as the circuit can provide sufficient voltage and current to maintain the correct potential of the sensing electrode. Figure 2 Schematic diagram of the electronic circuit for the O2-sensor The resistors R11, R12, R13 have to be chosen in the way that a VSET of +600 mv is supplied at the positive input of the operational amplifier IC2. The potential of the sensing electrode is held at 0 V by the biasing effect of the output circuit and so the result is a -600 mv bias between Sens and Ref. The reference voltage of the virtual earth should be chosen appropriate to allow enough voltage swing of the counter electrode. The maximum voltage between counter and sensing is 1.3 V. If the supply voltage is 2.5 V for example, then the reference voltage must be below 1.2 V for the virtual earth. The measuring circuit for the electrochemical sensor is a single stage op amp IC1 in a transimpedance configuration. The sensor current is reflected across R GAIN, generating an output voltage relative to the virtual earth GND. C2 reduces high frequency noise. The recommended value of the load resistor (R load =10 Ohm) is a compromise between fastest response time and best signal-to-noise ratio. The input offset voltage of the op amp IC1 will add to the sensor bias voltage (as the sensing electrode will be offset from 0V) so the input offset should be kept low. An op amp should be chosen with low input offset voltage temperature drift, to not affect the bias voltage when temperature changes. The control op amp IC2 provides the current to the counter electrode to balance the current required by the sensing electrode. Because the counter is oxidizing, it sinks a current into IC2. Therefore, IC2 needs to have an adequate current sinking capability. The inverting input into IC2 is connected to the reference electrode and must not draw any significant current from the reference electrode. An op amp with an input bias current of less than 5nA is recommended. MEM2 Appl. Note Rev. 8/2011 Page 3 of 5

4) Connecting the Sensor to the Circuit ELECTROCHEMICAL GAS SENSORS Never solder connectors directly on to pin of the sensor. Connection should be made via a PCB mounting socket. WARNING: SOLDERING TO PINS WILL RENDER YOUR WARRANTY VOID. 5) Usage of the Sensor Sensor must not be exposed to temperature, humidity and pressure outside the range quoted in the sensor data sheet. Sensor should not be exposed to organic vapour, which may influence the baseline or even cause physical damage to the body of the sensor. Electrochemical gas sensors are classed as non-dangerous and may be transported without special packaging or labelling. Although, you are advised to check any local regulations. 6) Intrinsic Safety Considerations The O2/M-100 is a three electrode cell which produces small currents and voltages and is not able to store large quantities of energy. The current of the sensor increases linearly over the recommended operating range of the oxygen concentration and is measured as the sensor output: sensor sensitivity [μa/%] x O 2 concentration [%] = output signal [ μa] Maximum current in normal operation (0-30% O2): < 4 ma Maximum voltage in normal operation (0-30% O2): < 1.4 V Note: IEC 60079 11, 5.6 Simple apparatus: The following apparatus shall be considered to be simple apparatus: Sources of generated energy, which do not generate more than 1.5V, 100 ma and 25mW. 7) Product Safety The Membrapor Oxygen Sensor is not considered a chemical hazard in normal use. Should the housing be damaged, the electrolyte inside the sensor may leak out. Exposure to the sensor electrolyte, which is diluted sulfuric acid, is the only component that may potentially prove hazardous to health. In the event of skin contact, rinse with plenty of water and seek medical advice. MEM2 Appl. Note Rev. 8/2011 Page 4 of 5

8) Storage ELECTROCHEMICAL GAS SENSORS Sensors should be stored in their original packaging below 30 C and 90% RH. Surrounding oxygen has no impact on storage life in opposite to sensors with a sacrificial anode. Do not store sensors together with organic solvents or flammable liquids. MEMBRAPOR AG is continuously improving its products and documentations. This document is subject to change without prior notice. Membrapor AG accepts no liability for any consequential losses, injury or damage resulting from the use of this document or the information contained within it. The data is given for guidance only. MEM2 Appl. Note Rev. 8/2011 Page 5 of 5