Calibrating the NI 5653 requires you to install one of the following packages on the calibration system. NI-RFSA 2.4 or later NI-RFSG 1.

Similar documents
PXIe Contents CALIBRATION PROCEDURE. 10 GHz or 20 GHz RF Analog Signal Generator

Contents. CALIBRATION PROCEDURE NI PXIe-5698

Required Software. Related Documentation. Password. Calibration Interval

Contents. CALIBRATION PROCEDURE PXIe-5673 Vector Signal Generator

NI PXIe Contents CALIBRATION PROCEDURE. Four Channel Precision Source-Measure Unit (SMU)

PXIe Contents. Required Software CALIBRATION PROCEDURE

NI PXIe Contents. Required Software CALIBRATION PROCEDURE. Dual-Output Programmable DC Power Supply

Contents. Software Requirements. CALIBRATION PROCEDURE NI PXIe-5663E

NI PXIe Contents CALIBRATION PROCEDURE. Single-Channel Precision Source-Measure Unit (SMU)

Contents. Software Requirements CALIBRATION PROCEDURE NI PXI-5663

NI PXIe-5171R. Contents. Required Software CALIBRATION PROCEDURE

Contents. Software. Documentation. CALIBRATION PROCEDURE NI PXIe-6674T

PXIe, 7½-Digit, ±1,000 V, Onboard 1.8 MS/s Isolated Digitizer, PXI Digital Multimeter

Contents. CALIBRATION PROCEDURE NI PXIe GHz and 14 GHz RF Vector Signal Analyzer

CALIBRATION PROCEDURE PXIe-4302/4303 and TB-4302C 32 Ch, 24-bit, 5 ks/s or 51.2 ks/s Simultaneous Filtered Data Acquisition Module. ni.

Contents. CALIBRATION PROCEDURE NI PXIe-5668R 14 GHz and 26.5 GHz Signal Analyzer

Contents. CALIBRATION PROCEDURE NI PXIe-4463 DSA Analog Output

Reconfigurable 6 GHz RF Vector Signal Transceiver with 1 GHz Bandwidth

Contents. Software Requirements

Power Amplifier. Input Characteristics. SPECIFICATIONS NI PXIe-4610 Power Amplifier. Signal Range

PXIe Contents CALIBRATION PROCEDURE. Reconfigurable 6 GHz RF Vector Signal Transceiver with 200 MHz Bandwidth

CALIBRATION PROCEDURE NI PXIe-4330/4331. Contents

GETTING STARTED GUIDE NI khz to 100 MHz, 50 Ω, AC Coupled RF Receiver Module

NI PXIe Contents CALIBRATION PROCEDURE

PXIe Contents. Required Software CALIBRATION PROCEDURE

Contents. Software Requirements. Documentation Requirements. Calibration Interval CALIBRATION PROCEDURE NI TB-4300B

NI PXIe-5667 (3.6 GHz)

NI 6624 Calibration Procedure

GETTING STARTED GUIDE NI AI, ±5 V, 24 Bit, 51.2 ks/s/ch Simultaneous, AC/DC Coupling, IEPE AC Coupling

DEVICE SPECIFICATIONS Phase Matrix PXI GHz PXI Vector Signal Analyzer

PXIe Contents CALIBRATION PROCEDURE

GETTING STARTED GUIDE NI AI, ±10 V, 24 Bit, 50 ks/s/ch Simultaneous

GETTING STARTED GUIDE NI AI, ±60 V, 24 Bit, 1 ks/s/ch Simultaneous

NI PXIe-5601 Specifications

NI PXIe-5630 Specifications

Input Characteristics. Measurement Range. SPECIFICATIONS NI PXIe-4357

NI PXIe-5663 Specifications

PXIe Contents SPECIFICATIONS. RF Amplifier

PXIe Contents SPECIFICATIONS. 14 GHz and 26.5 GHz Vector Signal Analyzer

These specifications apply to the PXIe-5113 with 64 MB and 512 MB of memory.

PXIe Contents SPECIFICATIONS. 250 khz to 20 GHz RF Analog Signal Generator

DEVICE SPECIFICATIONS NI PXIe Ch, 24-bit, 25.6 ks/s Simultaneous AC LVDT Input Module

Analog Characteristics

PXIe Contents SPECIFICATIONS. RF Preselector Module

PXIe Contents SPECIFICATIONS. 3.3 GHz RF Analog Signal Generator

Note Using the PXIe-5785 in a manner not described in this document might impair the protection the PXIe-5785 provides.

NI PXI Contents DEVICE SPECIFICATIONS. RF Attenuator

GETTING STARTED GUIDE NI AI Differential/32 AI Single-Ended, ±200 mv to ±10 V, 16 Bit, 250 ks/s Aggregate, Fuel-Cell Measurements

Contents. CALIBRATION PROCEDURE NI 5421/ MS/s Arbitrary Waveform Generator

if it is used in a manner not described in the user documentation.

PXI Contents SPECIFICATIONS. 6.6 GHz RF Analog Signal Generator

PXIe Contents. Definitions SPECIFICATIONS. 8.5 GHz Vector Network Analyzer

USRP-2940/2942/2943/2944/2945

Contents CALIBRATION PROCEDURE NI PXI-5422

PXIe, 24-Channel, ±24 V, 50 ma Precision PXI Source Measure Unit

PXI Contents SPECIFICATIONS. 1.3 GHz RF Analog Signal Generator

NI Contents CALIBRATION PROCEDURE

PXIe Contents. Definitions SPECIFICATIONS. 8.5 GHz Vector Network Analyzer

NI PXI-2797 Specifications 40 GHz 6 1 Terminated 50 Ω Multiplexer

Contents CALIBRATION PROCEDURE NI 5412

100 MHz Bandwidth, 12-Bit PXI Waveform Generator

Terminology. Output Characteristics. DEVICE SPECIFICATIONS NI PXIe-4463 DSA Analog Output. ni.com/manuals

NI sbrio-9632/9642 Verification Procedure

NI PXI/PXIe-2540 Specifications

SPECIFICATIONS PXIe Ch (8 ADC), 2 MS/s, bit, Flexible Resolution PXI Analog Input Module

DEVICE SPECIFICATIONS NI PXIe-4304/ Ch, 24-bit, ±42 V, 5 ks/s or 51.2 ks/s Simultaneous Filtered Data Acquisition Module. ni.

DEVICE SPECIFICATIONS NI PXIe-4302/ Ch, 24-bit, 5 ks/s or 51.2 ks/s Simultaneous Filtered Data Acquisition Module

USRP Contents. Definitions SPECIFICATIONS. 10 MHz to 6 GHz Tunable RF Transceiver

is used in a manner not described in the user documentation.

NI PXI/PXIe-2541 Specifications 300 MHz Ω Matrix

NI PXI-2521 Specifications 40-Channel DPST Relay Module

Reconfigurable 6 GHz Vector Signal Transceiver with I/Q Interface

NI PXI/PCI-5412 Specifications 14-Bit 100 MS/s Arbitrary Waveform Generator

SCXI MHz Quad 8x1 50 Ω Multiplexer

SPECIFICATIONS PXIe channel, 16-bit, 400 ks/s/ch, Ch-Ch Isolated Analog Input Module. ni.com/manuals

NI PXI-5422 Specifications 16-Bit 200 MS/s Arbitrary Waveform Generator

Contents CALIBRATION PROCEDURE NI PXI-5404

SCXI Terminal Block

NI PXI-2520 Specifications 80-Channel SPST Relay Module

NI Contents CALIBRATION PROCEDURE

Current ON state current ma minimum, 6 ma typical, 10 ma maximum OFF state current ma maximum

OPERATING INSTRUCTIONS AND SPECIFICATIONS. NI 9211E 4-Channel Thermocouple Input Module

NI PXI/PCI-5402/5406 Specifications 14/16-Bit, 20/40 MHz Arbitrary Function Generator

mmwave Transceiver System

NI PXIe Contents. About These Specifications DEVICE SPECIFICATIONS. 3 GHz Ω Multiplexer (SP16T)

Contents. USER MANUAL NI ISM-7400 Integrated Stepper

80 MHz Bandwidth, 16-Bit PXI Waveform Generator. These specifications apply to the one-channel and two-channel PXIe-5433.

NI PXI/PXIe-2529 Specifications

Contents. CALIBRATION PROCEDURE NI PXIe-6555/6556. ni.com/manuals

PXIe Contents SPECIFICATIONS. 100 MVector/s PXI Digital Pattern Instrument

USRP Contents. Definitions SPECIFICATIONS. Software Defined Radio Reconfigurable Device

USRP Contents. Definitions. Conditions SPECIFICATIONS. Software Defined Radio Reconfigurable Device

PXI MICROWAVE LOCAL OSCILLATOR MODULE

PXIe Contents SPECIFICATIONS. 32-Channel Digital Pattern Instrument

NI 2865A 0.3 A Matrix Cards for NI SwitchBlock

USRP Contents. Definitions. Conditions SPECIFICATIONS. Software Defined Radio Device

USRP Contents. Definitions SPECIFICATIONS. 50 MHz to 2.2 GHz Tunable RF Transceiver

USRP Contents. Definitions. Conditions SPECIFICATIONS. Software Defined Radio Device

This document lists the specifications for the NI PXI The PXI-5900 is a differential amplifier.

PXIe, 40 MHz Bandwidth, 16-Bit PXI Waveform Generator. These specifications apply to the one-channel and two-channel PXIe-5423.

Transcription:

CALIBRATION PROCEDURE NI PXIe-5653 This document contains the verification and adjustment procedures for the National Instruments PXIe-5653 RF synthesizer (NI 5653). Refer to ni.com/calibration for more information about calibration solutions. Contents Software... 2 Documentation... 2 Test Equipment... 2 Test Conditions... 5 Initial Setup... 5 Self-Calibrating the NI 5653... 5 As-Found and As-Left Limits... 6 Verification... 6 Verifying 10 MHz Reference Frequency Accuracy... 6 Verifying 10 MHz Reference Amplitude Accuracy... 7 Verifying 100 MHz Reference Frequency Accuracy... 7 Verifying 100 MHz Reference Amplitude Accuracy... 8 Verifying LO1 Frequency Accuracy... 9 Verifying LO1 Amplitude Accuracy... 10 Verifying LO2 Amplitude Accuracy... 10 Verifying LO3 Amplitude Accuracy... 11 Verifying LO1 Phase Noise... 11 Verifying LO2 Phase Noise... 12 Verifying LO3 Phase Noise... 14 Adjustment... 15 Adjusting Reference Accuracy... 15 Adjusting YIG Frequency... 16 Adjusting LO1, LO2, and LO3 Output Power... 17 EEPROM Update... 17 Reverification... 17 Reverifying 10 MHz Reference Amplitude Accuracy... 18 Reverifying 100 MHz Reference Amplitude Accuracy... 18 Reverifying LO1 Amplitude Accuracy... 18 Reverifying LO2 Amplitude Accuracy... 19 Reverifying LO3 Amplitude Accuracy... 19 Reverifying Other Device Performance Specifications... 20 Worldwide Support and Services... 20

Software Calibrating the NI 5653 requires you to install one of the following packages on the calibration system. NI-RFSA 2.4 or later NI-RFSG 1.7 or later Note NI-RFSG automatically installs when you install NI-RFSA. You can download all required software from ni.com/downloads. The software supports programming the calibration procedures in the LabVIEW, C, and LabWindows /CVI application development environments (ADE). When you install the software, you need to install support only for the ADE that you intend to use. Documentation You might find the following documents helpful as you perform the calibration procedure: NI 5665 RF Vector Signal Analyzer Getting Started Guide NI RF Signal Generators Help NI PXIe-5665 Specifications The latest versions of these documents are available on ni.com/manuals. Test Equipment Table 1 lists the equipment NI recommends for the performance verification and adjustment procedures. If the recommended equipment is not available, select a substitute using the minimum requirements listed in the table. 2 ni.com NI PXIe-5653 Calibration Procedure

Table 1. Required Equipment Specifications for NI 5653 Calibration Equipment Recommended Model Where Used Minimum Requirements NI PXIe-5653 Calibration Procedure National Instruments 3 Signal source analyzer Frequency reference Rohde & Schwarz (R&S) FSUP Signal Source Analyzer Symmetricon 8040C Rubidium Frequency Standard Verifying 10 MHz Reference Frequency Accuracy Verifying 100 MHz Reference Frequency Accuracy Verifying LO1 Frequency Accuracy Verifying LO1 Phase Noise Verifying LO2 Phase Noise Verifying LO3 Phase Noise Adjusting Reference Accuracy Adjusting YIG Frequency Accuracy Verifying 10 MHz Reference Frequency Accuracy Verifying 100 MHz Reference Frequency Accuracy Verifying LO1 Frequency Accuracy Adjusting Reference Accuracy Adjusting YIG Frequency Accuracy Frequency range: 9 MHz to 10 GHz Frequency accuracy: 1 ppb over the frequency range, 900 MHz to 9 GHz recommended Noise floor: <-152 dbm/hz Phase noise measurement using cross-correlation Frequency counter marker feature Spectrum analysis capabilities Frequency: 10 MHz Frequency accuracy: 1 * 10-9 (typically ±5E-11)

4 ni.com NI PXIe-5653 Calibration Procedure Power meter SMA (m)-to-sma (m) Cable Table 1. Required Equipment Specifications for NI 5653 Calibration (Continued) Equipment Recommended Model Where Used Minimum Requirements Anritsu ML2438A with a MA247xD Series Diode Sensor Verifying 10 MHz Reference Amplitude Accuracy Verifying 100 MHz Reference Amplitude Accuracy Verifying LO1 Amplitude Accuracy Verifying LO2 Amplitude Accuracy Verifying LO3 Amplitude Accuracy Adjusting LO1, LO2, and LO3 Output Power Verifying 10 MHz Reference Frequency Accuracy Verifying 100 MHz Reference Frequency Accuracy Verifying LO1 Frequency Accuracy Verifying LO1 Phase Noise Verifying LO2 Phase Noise Verifying LO3 Phase Noise Adjusting Reference Accuracy Adjusting YIG Frequency Accuracy Adjusting LO1, LO2, and LO3 Output Power Range: 0 dbm to +20 dbm Frequency range: 9 MHz to 8.3 GHz Accuracy: 0.5% Linearity: Up to +15 dbm (MA2475 only) Length: 36 in.

Test Conditions Follow these guidelines to optimize the connections and the environment during calibration: Keep cabling as short as possible. Long cables act as antennas, picking up extra noise that can affect measurements. Verify that all connections, including front panel connections and screws, are secure. Maintain an ambient temperature of 23 C ±5 C. Keep relative humidity between 10% and 90%, noncondensing. Allow a warm-up time of at least 30 minutes after the chassis is powered on. The warm-up time ensures that the NI 5653 is at a stable operating temperature. In each verification procedure, insert a delay between configuring all devices and acquiring the measurement. This delay may need to be adjusted depending on the instruments used but should always be at least 1,000 ms for the first iteration, 1,000 ms when the power level changes, and 100 ms for each other iteration. Perform self-calibration on the NI 5653. Plug the PXI/PXI Express chassis and the calibrator into the same power strip to avoid ground loops. Use a torque wrench appropriate for the type of RF connector that you are using. NI recommends a 0.565 N m (5 lb in.) wrench for SMA connectors and an 0.90 N m (8 lb in.) wrench for 3.5 mm connectors. Connect the frequency reference source to the REF IN connector on the back of the PXI Express chassis with a standard BNC (m)-to-bnc (m) cable. This connection replaces the connection from the NI 5653 REF OUT (10 MHz) connector to the PXI Express chassis REF IN connector, if present. Ensure that the PXI/PXI Express chassis fan speed is set to HIGH, that the fan filters, if present, are clean, and that the empty slots contain filler panels. For more information, refer to the Maintain Forced-Air Cooling Note to Users document available at ni.com/ manuals. Initial Setup Refer to the NI 5665 RF Vector Signal Analyzer Getting Started Guide for information about how to install the software and hardware and how to configure the device in Measurement and Automation Explorer (MAX). Self-Calibrating the NI 5653 Note No signal connections are needed for self-calibration. You can self-calibrate the NI 5653 using the self-calibration features provided with NI-RFSG. NI PXIe-5653 Calibration Procedure National Instruments 5

As-Found and As-Left Limits The as-found limits are the published specifications for the NI 5653, found in the NI PXIe-5665 Specifications document. NI uses these limits to determine whether the NI 5653 meets the device specifications when it is received for calibration. The as-left limits are equal to the published NI specifications for the NI 5653, less guard bands for measurement uncertainty, temperature drift, and drift over time. NI uses these limits to determine whether the NI 5653 meets the device specifications over its calibration interval. Verification The performance verification procedures assume that adequate traceable uncertainties are available for the calibration references. Verifying 10 MHz Reference Frequency Accuracy This verification ensures that the internal oven-controlled crystal oscillator (OCXO) is adjusted for correct frequency accuracy. Complete the following procedure to determine the as-found status of the NI 5653. 1. Connect the NI 5653 REF OUT 10 MHz front panel connector to the signal source analyzer RF IN front panel connector. 2. Connect the signal source analyzer REF IN rear panel connector to the rubidium frequency reference output connector. 3. Create a new device session for the NI 5653. 4. Set the RF frequency to 4 GHz and commit the settings to hardware. 5. Check the signal generation status and verify that there are no reported errors or warnings. 6. Configure the signal source analyzer as follows: Center frequency: 110 MHz Reference level: -50 dbm Frequency span: 1 khz Reference clock source: External Frequency counter mode: 0.1 Hz resolution 7. Use the signal source analyzer frequency counter to measure the frequency of the peak at approximately 110 MHz. 6 ni.com NI PXIe-5653 Calibration Procedure

8. Calculate the deviation using the following formula: Δf = f measured ( MHz) 110MHz ------------------------------------------------------------------- 110MHz 9. Ensure that the result in step 8 is less than the result of the following formula: Initial Accuracy + Aging + Temperature Stability where Initial Accuracy = 50 10 9 100 10 9 Aging = ------------------------- NumberofYearsSinceLastAdjustment Year Temperature Stability = 10 10 9 10. Close the device session. Verifying 10 MHz Reference Amplitude Accuracy This verification ensures that the internal 10 MHz reference circuitry is adjusted for correct amplitude accuracy. Complete the following procedure to determine the as-found status of the NI 5653. 1. Connect the NI 5653 REF OUT 10 MHz front panel connector to the power meter. 2. Connect the signal source analyzer REF IN rear panel connector to any rubidium frequency reference output connector. 3. Create a new device session for the NI 5653. 4. Set the RF frequency to 4 GHz and commit the settings to hardware. 5. Check the signal generation status and verify that there are no reported errors or warnings. 6. After correcting for the measurement frequency, use the power meter to measure the NI 5653 REF OUT 10 MHz front panel connector power. Ensure that the power is between 1 dbm and 10 dbm. 7. Close the device session. Verifying 100 MHz Reference Frequency Accuracy This verification ensures that the internal 100 MHz reference circuitry is adjusted for correct frequency accuracy. Complete the following procedure to determine the as-found status of the NI 5653. 1. Connect the NI 5653 REF OUT 100 MHz front panel connector to the signal source analyzer RF IN front panel connector. 2. Connect the signal source analyzer REF IN rear panel connector to any rubidium frequency reference output connector. 3. Create a new device session for the NI 5653. 4. Set the RF frequency to 4 GHz and commit the settings to hardware. 5. Check the signal generation status and verify that there are no reported errors or warnings. NI PXIe-5653 Calibration Procedure National Instruments 7

6. Configure the signal source analyzer as follows: Center frequency: 100 MHz Reference level: 10 dbm Frequency span: 250 khz Reference clock source: External Frequency counter mode: 0.1 Hz resolution 7. Use the signal source analyzer frequency counter to measure the frequency of the peak at approximately 110 MHz. 8. Calculate the deviation using the following formula: Δf = f measured ( MHz) 100MHz ------------------------------------------------------------------- 100MHz 9. Ensure that the result in step 8 is less than the result of the following formula: Where: Initial Accuracy + Aging + Temperature Stability Initial Accuracy = 50 10 9 100 10 9 Aging = ------------------------- NumberofYearsSinceLastAdjustment Year Temperature Stability = 10 10 9 10. Close the device session. Verifying 100 MHz Reference Amplitude Accuracy This verification ensures that the 100 MHz reference circuitry is adjusted for correct amplitude accuracy. Complete the following procedure to determine the as-found status of the NI 5653. 1. Connect the NI 5653 REF OUT 100 MHz front panel connector to the power meter. 3. Set the RF frequency to 4 GHz and commit the settings to hardware. 4. Check the signal generation status and verify that there are no reported errors or warnings. 5. After correcting for the measurement frequency, use the power meter to measure the NI 5653 REF OUT 100 MHz front panel connector power. Ensure that the power is between 1 dbm and 10 dbm. 6. Close the device session. 8 ni.com NI PXIe-5653 Calibration Procedure

Verifying LO1 Frequency Accuracy This verification ensures that the internal local oscillator (LO) circuitry is adjusted for correct frequency accuracy. Complete the following procedure to determine the as-found status of the NI 5653. 1. Connect the NI 5653 LO1 front panel connector to the signal source analyzer RF IN front panel connector. 2. Connect the signal source analyzer REF IN rear panel connector to the rubidium frequency reference output connector. 3. Create a new device session for the NI 5653. 4. Create a list of test frequencies from 3.2 GHz to 8.3 GHz in 100 MHz increments including endpoints. 5. Set the RF frequency to the first value of the frequency array you created in step 4 and commit the settings to hardware. 6. Check the signal generation status and verify that there are no reported errors or warnings. 7. Configure the signal source analyzer as follows: Center frequency: Frequency from the list in step 4 Reference level: 20 dbm Frequency span: 1 khz Reference clock source: External 8. Measure the frequency of the peak that is returned by the signal source analyzer at approximately the corresponding point in the frequency array you created in step 4. 9. Use the following formula to calculate the deviation: Δf = f measured ( MHz) fexpected( MHz) ----------------------------------------------------------------------------------- f ( MHz) expected 10. Repeat steps 5 through 9 for each subsequent point in the frequency array created in step 4. 11. Ensure that the results in step 8 are less than the result of the following formula: Where Initial Accuracy + Aging + Temperature Stability Initial Accuracy = 50 10 9 100 10 9 Aging = ------------------------- NumberofYearsSinceLastAdjustment Year Temperature Stability = 10 10 9 12. Close the device session. NI PXIe-5653 Calibration Procedure National Instruments 9

Verifying LO1 Amplitude Accuracy This verification ensures that the internal LO1 circuitry is adjusted for correct amplitude accuracy. Complete the following procedure to determine the as-found status of the NI 5653. 1. Connect the NI 5653 LO1 front panel connector to the power meter. 3. Create a list of test frequencies from 3.2 GHz to 8.3 GHz in 100 MHz increments, including endpoints. 4. Calculate the expected output power, P Expected, for each frequency using the following formulas. For all frequencies from 3.2 GHz to 8.2 GHz: frequency GHz P Expected 10.5 3 ( ) 3.2GHz = -------------------------------------------------------------------- ( dbm) 5.0GHz For the frequency 8.3 GHz: P Expected = 6.5 dbm 5. Set the RF frequency to the first value of the frequency array you created in step 3 and commit the settings to hardware. 6. Check the signal generation status and verify that there are no reported errors or warnings. 7. After correcting for the measurement frequency, use the power meter to measure the NI 5653 LO1 front panel connector power. 8. Calculate the power deviation, P Deviation, using the following formula: P Deviation ( db) = P Measured ( dbm) P Expected ( dbm) 9. Repeat steps 5 through 8 for each point in the frequency array you created in step 3. 10. Verify that the power deviation results calculated in step 9 for each point are within ±2.5 db for LO1. 11. Close the device session. Verifying LO2 Amplitude Accuracy This verification ensures that the internal LO2 circuitry is adjusted for correct amplitude accuracy. Complete the following procedure to determine the as-found status of the NI 5653. 1. Connect the NI 5653 LO2 front panel connector to the power meter. 3. Set the RF frequency to 4 GHz and commit the settings to hardware. 4. Check the signal generation status and verify that there are no reported errors or warnings. 5. After correcting for the measurement frequency, use the power meter to measure the LO2 output power. 6. Ensure that the output power is between 6.5 dbm and 13 dbm. 7. Close the device session. 10 ni.com NI PXIe-5653 Calibration Procedure

Verifying LO3 Amplitude Accuracy This verification ensures that the internal LO3 circuitry is adjusted for correct amplitude accuracy. Complete the following procedure to determine the as-found status of the NI 5653. 1. Connect the NI 5653 LO3 front panel connector to the power meter. 3. Set the RF frequency to 4 GHz and commit the settings to hardware. 4. Check the signal generation status and verify that there are no reported errors or warnings. 5. After correcting for the measurement frequency, use the power meter to measure the LO3 output power. 6. Ensure that the output power is between 7 dbm and 13 dbm. 7. Close the device session. Verifying LO1 Phase Noise This verification ensures that the internal LO1 circuitry is adjusted for correct phase noise accuracy. Complete the following procedure to determine the as-found status of the NI 5653. 1. Connect the NI 5653 LO1 front panel connector to the signal source analyzer RF IN front panel connector. 3. Configure the signal source analyzer as follows: Phase noise measurement using cross-correlation PLLs and internal generators Automatic DUT detection Spur suppression enabled Fast Fourier transform (FFT) mode with Blackman-Harris window Measurement span from 100 Hz to 10 MHz (offset frequencies) 4. Use the signal source analyzer to perform a manual frequency sweep at the settings shown in Table 2. Table 2. LO1 Manual Sweep Settings Carrier Frequency Offset Resolution Bandwidth Cross Correlation 100 Hz to 300 Hz 10 Hz 100 300 Hz to 1 khz 30 Hz 1,000 1 khz to 3 khz 100 Hz 10,000 3 khz to 10 khz 300 Hz 10,000 10 khz to 30 khz 1 khz 10,000 30 khz to 100 khz 1 khz 5,000 100 khz to 300 khz 3 khz 1,000 NI PXIe-5653 Calibration Procedure National Instruments 11

Table 2. LO1 Manual Sweep Settings (Continued) Carrier Frequency Offset Resolution Bandwidth Cross Correlation 300 khz to 1 MHz 10 khz 1,000 1 MHz to 3 MHz 30 khz 1,000 3 MHz to 10 MHz 100 khz 1,000 5. Set the RF frequency to 5.4125 GHz and commit the settings to hardware. 6. If your signal source analyzer does not automatically center for phase noise measurements, manually set the signal source analyzer center frequency to 5.4125 GHz. 7. Check the signal generation status and verify that there are no reported errors or warnings. 8. Use the signal source analyzer to measure the phase noise. Record the measurements. 9. Set the RF frequency to 7.8125 GHz and commit the settings to hardware. 10. If your signal source analyzer does not automatically center for phase noise measurements, manually set the signal source analyzer center frequency to 7.8125 GHz. 11. Repeat steps 7 through 8 using the new frequency and record the measurements. 12. Ensure that the recorded measurements are within the limits set in Table 3. Verifying LO2 Phase Noise Table 3. LO1 Phase Noise Density (dbc/hz) Offset LO1 at 5.4125 GHz LO1 at 7.8125 GHz 100 Hz <-89 <-86 1 khz <-118 <-115 10 khz <-128 <-127 100 khz <-125 <-125 1 MHz <-141 <-141 5 MHz <-155 <-155 This verification ensures that the internal LO2 circuitry is adjusted for correct phase noise accuracy. Complete the following procedure to determine the as-found status of the NI 5653. 1. Connect the NI 5653 LO2 front panel connector to the signal source analyzer RF IN front panel connector. 3. Configure the signal source analyzer as follows: Phase noise measurement using cross-correlation PLLs and internal generators Automatic DUT detection Spur suppression enabled 12 ni.com NI PXIe-5653 Calibration Procedure

Fast fourier transform (FFT) mode with Blackman-Harris window Measurement span from 100 Hz to 10 MHz (offset frequencies) 4. Use the signal source analyzer to perform a manual frequency sweep at the settings in Table 4. Table 4. LO2 Manual Sweep Settings Carrier Frequency Offset Resolution Bandwidth Cross Correlation 100 Hz to 300 Hz 10 Hz 100 300 Hz to 1 khz 30 Hz 1,000 1 khz to 3 khz 100 Hz 10,000 3 khz to 10 khz 300 Hz 10,000 10 khz to 30 khz 1 khz 10,000 30 khz to 100 khz 1 khz 5,000 100 khz to 300 khz 3 khz 1,000 300 khz to 1 MHz 10 khz 1,000 1 MHz to 3 MHz 30 khz 1,000 3 MHz to 10 MHz 100 khz 1,000 5. Set the RF frequency to 4 GHz and commit the settings to hardware. 6. If your signal source analyzer does not automatically center for phase noise measurements, manually set the signal source analyzer center frequency to 4 GHz. 7. Check the signal generation status and verify that there are no reported errors or warnings. 8. Use the signal source analyzer to measure the phase noise. Record the results. 9. Close the device session. 10. Ensure that the recorded measurements are within the limits set in Table 5. Table 5. LO2 Phase Noise Density (dbc/hz) Offset LO2 100 Hz <-92 1 khz <-121 10 khz <-134 100 khz <-134 1 MHz <-143 5 MHz <-155 NI PXIe-5653 Calibration Procedure National Instruments 13

Verifying LO3 Phase Noise This verification ensures that the internal LO3 circuitry is adjusted for correct phase noise accuracy. Complete the following procedure to determine the as-found status of the NI 5653. 1. Connect the NI 5653 LO2 front panel connector to the signal source analyzer RF IN front panel connector. 3. Configure the signal source analyzer as follows: Phase noise measurement using cross-correlation PLLs and internal generators Automatic DUT detection Spur suppression enabled Fast Fourier transform (FFT) mode with Blackman-Harris window Measurement span from 100 Hz to 10 MHz (offset frequencies) 4. Use the signal source analyzer to perform a manual frequency sweep at the settings listed in Table 6. Table 6. LO3 Phase Noise Frequency Sweep Signal Source Analyzer Settings Carrier Frequency Offset Resolution Bandwidth Cross Correlation 100 Hz to 300 Hz 10 Hz 100 300 Hz to 1 khz 30 Hz 1,000 1 khz to 3 khz 100 Hz 10,000 3 khz to 10 khz 300 Hz 10,000 10 khz to 30 khz 1 khz 10,000 30 khz to 100 khz 1 khz 5,000 100 khz to 300 khz 3 khz 1,000 300 khz to 1 MHz 10 khz 1,000 1 MHz to 3 MHz 30 khz 1,000 3 MHz to 10 MHz 100 khz 1,000 5. Set the RF frequency to 4 GHz and commit the settings to hardware. 6. If your signal source analyzer does not automatically center for phase noise measurements, manually set the signal source analyzer center frequency to 800 MHz. 7. Check the signal generation status and verify that there are no reported errors or warnings. 8. Use the signal source analyzer to measure the phase noise. Record the results. 9. Close the device session. 14 ni.com NI PXIe-5653 Calibration Procedure

10. Ensure that the recorded measurements are within the limits set in Table 7. Table 7. LO3 Phase Noise Density (dbc/hz) Offset LO3 Adjustment 100 Hz <-104 1 khz <-135 10 khz <-148 100 khz <-149 1 MHz <-158 5 MHz <-160 Completing the adjustment procedure automatically updates the calibration date and temperature stored in the NI 5653 EEPROM. The calibration date also updates when you call the nirfsg Initialize External Calibration and the nirfsg Close External Calibration VIs. Adjusting Reference Accuracy Complete the following procedure to adjust the NI 5653 frequency accuracy performance using a signal source analyzer. This adjustment yields a more accurately tuned VCO frequency. 1. Connect the NI 5653 REF OUT 10 MHz front panel connector to the signal source analyzer RF IN front panel connector. 2. Connect the signal source analyzer REF IN rear panel connector to the rubidium frequency reference output connector. 3. Configure the signal source analyzer as follows: Center frequency: 10 MHz Reference level: 10 dbm Frequency span: 600 Hz Reference clock source: External 4. Create a new external calibration session using the appropriate password. Note The default password for password-protected operations is NI. 5. Create a new Reference Clock calibration session. NI PXIe-5653 Calibration Procedure National Instruments 15

6. Use the signal source analyzer to measure the peak power and frequency at the peak with the following parameter settings: Set frequency reference level: Spectrum Analyzer Ref Level value returned by NI-RFSG Set center frequency: Spectrum Analyzer Center Frequency value returned by NI-RFSG Frequency counter mode: 0.1 Hz resolution Measure power level: maximum peak power as measured 7. Adjust the NI 5653 Reference Clock using the measurements taken in step 6 for the Measured Ref Out Frequency Peak and Measured Ref Out Power parameters. 8. Repeat steps 6 and 7 until the Reference Clock Calibration Complete parameter returns a value of TRUE. 9. Close the external calibration session. Store the results to the EEPROM on the device. Adjusting YIG Frequency Complete the following procedure to adjust the NI 5653 YIG coarse frequency performance using a signal source analyzer. 1. Connect the NI 5653 LO1 front panel connector to the signal source analyzer RF IN front panel connector. 2. Connect the signal source analyzer REF IN rear panel connector to the rubidium frequency reference output connector. 3. Configure the signal source analyzer as follows: Center frequency: 3 GHz Reference level: 20 dbm Frequency span: 1 GHz Resolution bandwidth: 50 khz Reference clock source: External 4. Create a new external calibration session using the appropriate password. 5. Create a new YIG frequency calibration session. 6. Use the signal source analyzer to measure the peak power and frequency at the peak with the following parameter settings: Set frequency span: Spectrum Analyzer Frequency Span value returned by NI-RFSG Set center frequency: Spectrum Analyzer Center Frequency value returned by NI-RFSG Measure frequency: Frequency at peak power as measured Measure power level: Maximum peak power as measured 7. Adjust the YIG frequency using the measurements taken in step 6 for the Measured LO1 Frequency and Measured LO1 Power parameters. 16 ni.com NI PXIe-5653 Calibration Procedure

8. Repeat steps 6 through 7 until the YIG Frequency Calibration Complete parameter returns a value of TRUE. 9. Close the external calibration session. Store the results to the EEPROM on the device. Adjusting LO1, LO2, and LO3 Output Power Complete the following procedure to adjust the NI 5653 LO power performance using a power meter. This adjustment yields a more accurate output power for the NI 5653 LO front panel connectors. 1. Connect the NI 5653 LO1 front panel connector to the power meter. 2. Create a new external calibration session using the appropriate password. 3. Create a new LO gain calibration session for LO1. 4. Use the power meter to measure the LO1 output power at the frequency specified by the Calibration Frequency parameter. 5. Adjust the LO Gain using the power measured in step 4 for the Measured LO Power parameter. 6. Repeat steps 4 and 5 until the LO Gain Calibration Complete parameter returns a value of TRUE. 7. Close the external calibration session. Store the results to the EEPROM on the device. 8. Disconnect the NI 5653 LO1 front panel connector from the power meter, and connect the NI 5653 LO2 front panel connector to the power meter. 9. Repeat steps 2 through 7 for LO2. 10. Disconnect the NI 5653 LO2 front panel connector from the power meter, and connect the NI 5653 LO3 front panel connector to the power meter. 11. Repeat steps 2 through 7 for LO3. EEPROM Update When an adjustment procedure is completed, the NI 5653 internal calibration memory, stored in the EEPROM, is immediately updated. If you do not want to perform an adjustment, you can update the calibration date and onboard calibration temperature without making any adjustments by performing an external calibration in software. Reverification Complete the following sections to determine the as-left status of the device. Note If any test fails reverification after performing an adjustment, verify that you have met the Test Conditions before returning your device to NI. Refer to Worldwide Support and Services for information about support resources or service requests. NI PXIe-5653 Calibration Procedure National Instruments 17

Reverifying 10 MHz Reference Amplitude Accuracy This reverification ensures that the internal 10 MHz reference circuitry is adjusted for correct amplitude accuracy following adjustment. Complete the following procedure to determine the as-left status of the NI 5653. 1. Connect the NI 5653 REF OUT 10 MHz front panel connector to the power meter. 2. Connect the signal source analyzer REF IN rear panel connector to any rubidium frequency reference output connector. 3. Create a new device session for the NI 5653. 4. Set the RF frequency to 4 GHz and commit the settings to hardware. 5. Check the signal generation status and verify that there are no reported errors or warnings. 6. After correcting for the measurement frequency, use the power meter to measure the NI 5653 REF OUT 10 MHz front panel connector power. Ensure that the power is between 4 dbm and 7.5 dbm. 7. Close the device session. Reverifying 100 MHz Reference Amplitude Accuracy This reverification ensures that the 100 MHz reference circuitry is adjusted for correct amplitude accuracy following adjustment. Complete the following procedure to determine the as-left status of the NI 5653. 1. Connect the NI 5653 REF OUT 100 MHz front panel connector to the power meter. 3. Set the RF frequency to 4 GHz and commit the settings to hardware. 4. Check the signal generation status and verify that there are no reported errors or warnings. 5. After correcting for the measurement frequency, use the power meter to measure the NI 5653 REF OUT 100 MHz front panel connector power. Ensure that the power is between 4 dbm and 7.5 dbm. 6. Close the device session. Reverifying LO1 Amplitude Accuracy This reverification ensures that the internal LO1 circuitry is adjusted for correct amplitude accuracy following adjustment. Complete the following procedure to determine the as-left status of the NI 5653. 1. Connect the NI 5653 LO1 front panel connector to the power meter. 3. Create a list of test frequencies from 3.2 GHz to 8.3 GHz in 100 MHz increments, including endpoints. 18 ni.com NI PXIe-5653 Calibration Procedure

4. Calculate the expected output power, P Expected, for each frequency using the following formulas. For all frequencies from 3.2 GHz to 8.2 GHz: frequency GHz P Expected 10.5 3 ( ) 3.2GHz = -------------------------------------------------------------------- ( dbm) 5.0GHz For the frequency 8.3 GHz: P Expected = 6.5 dbm 5. Set the RF frequency to the first value of the frequency array you created in step 3 and commit the settings to hardware. 6. Check the signal generation status and verify that there are no reported errors or warnings. 7. After correcting for the measurement frequency, use the power meter to measure the NI 5653 LO1 front panel connector power. 8. Calculate the power deviation, P Deviation, using the following formula: P Deviation ( db) = P Measured ( dbm) P Expected ( dbm) 9. Repeat steps 5 through 8 for each point in the frequency array you created in step 3. 10. Verify that the power deviation results calculated in step 9 for each point are within ±1.0 db for LO1. 11. Close the device session. Reverifying LO2 Amplitude Accuracy This reverification ensures that the internal LO2 circuitry is adjusted for correct amplitude accuracy following adjustment. Complete the following procedure to determine the as-left status of the NI 5653. 1. Connect the NI 5653 LO2 front panel connector to the power meter. 3. Set the RF frequency to 4 GHz and commit the settings to hardware. 4. Check the signal generation status and verify that there are no reported errors or warnings. 5. After correcting for the measurement frequency, use the power meter to measure the LO2 output power. 6. Ensure that the output power is between 8.25 dbm and 9.75 dbm. 7. Close the device session. Reverifying LO3 Amplitude Accuracy This reverification ensures that the internal LO3 circuitry is adjusted for correct amplitude accuracy following adjustment. Complete the following procedure to determine the as-left status of the NI 5653. 1. Connect the NI 5653 LO3 front panel connector to the power meter. NI PXIe-5653 Calibration Procedure National Instruments 19

3. Set the RF frequency to 4 GHz and commit the settings to hardware. 4. Check the signal generation status and verify that there are no reported errors or warnings. 5. After correcting for the measurement frequency, use the power meter to measure the LO3 output power. 6. Ensure that the output power is between 9 dbm and 11 dbm. 7. Close the device session. Reverifying Other Device Performance Specifications For all other device performance specifications, repeat the Verification section to determine the as-left status of the device. Worldwide Support and Services The National Instruments website is your complete resource for technical support. At ni.com/ support you have access to everything from troubleshooting and application development self-help resources to email and phone assistance from NI Application Engineers. Visit ni.com/services for NI Factory Installation Services, repairs, extended warranty, and other services. Visit ni.com/register to register your National Instruments product. Product registration facilitates technical support and ensures that you receive important information updates from NI. A Declaration of Conformity (DoC) is our claim of compliance with the Council of the European Communities using the manufacturer s declaration of conformity. This system affords the user protection for electromagnetic compatibility (EMC) and product safety. You can obtain the DoC for your product by visiting ni.com/certification. If your product supports calibration, you can obtain the calibration certificate for your product at ni.com/calibration. National Instruments corporate headquarters is located at 11500 North Mopac Expressway, Austin, Texas, 78759-3504. National Instruments also has offices located around the world. For telephone support in the United States, create your service request at ni.com/support or dial 512 795 8248. For telephone support outside the United States, visit the Worldwide Offices section of ni.com/niglobal to access the branch office websites, which provide up-to-date contact information, support phone numbers, email addresses, and current events. Refer to the NI Trademarks and Logo Guidelines at ni.com/trademarks for more information on National Instruments trademarks. Other product and company names mentioned herein are trademarks or trade names of their respective companies. For patents covering National Instruments products/technology, refer to the appropriate location: Help»Patents in your software, the patents.txt file on your media, or the National Instruments Patents Notice at ni.com/patents. You can find information about end-user license agreements (EULAs) and third-party legal notices in the readme file for your NI product. Refer to the Export Compliance Information at ni.com/legal/export-compliance for the National Instruments global trade compliance policy and how to obtain relevant HTS codes, ECCNs, and other import/export data. 2011 2013 National Instruments. All rights reserved. 373369B-01 Dec13