Successful endovascular treatment for BTK lesion using wire rendezvous technique and retrograde knuckle wire technique by collateral approach

Similar documents
How to Penetrate the Proximal and Distal Fibrous Caps in Chronic Total Occlusion?

The Current CTO Toolbox

Guidewires for the CTO Interventionalist. William L. Lombardi MD FACC FSCAI Peacehealth St. Joseph Medical Center Bellingham WA

Coronary guidewires for chronic total occlusion procedures: function and design

CROSS YOUR WAY SUPPORT CATHETERS GUIDEWIRES CROSSING CATHETERS RE-ENTRY DEVICES

ROLL CALL PURCHASE RECOMMENDATIONS

Welcome to the 7 th European Bifurcation Club October LISBON. Side Branch access session

Medtronic Company Overview

Wire Technology: How to Choose Based on Anatomy and Strategy

IVUS in CTO PCI. sidebranch technique and intralesional rewiring

Accepts Wire Guide Diameter inch. Stent Diameter mm

Medtronic Payer Solutions

A case of tortuous calcified LAD CTO treated using a "GAIA" guidewire

Unofficial Ottawa Fellowship Guide to Endovascular interventions, Anton Sharapov, MD, FRCS, Vascular Surgeon Updated October 26, 2014

Unofficial Ottawa Fellowship Guide to Endovascular interventions, Anton Sharapov, MD, FRCS, Vascular Surgeon Updated February 21, 2012

Patient-specific simulation in carotid artery stenting

Downhole Monitoring Solutions. Medical Instrumentation

MVP MICRO VASCULAR PLUG SYSTEM AN OPTION FOR PERIPHERAL EMBOLIZATION CASE REPORTS

A 360-Degree Approach to Neuroendovascuar Therapy

Precision Vascular Robotics. Corindus Vascular Robotics (CVRS) January 2017

Iterative Reconstruction

(12) Patent Application Publication (10) Pub. No.: US 2012/ A1

Simultaneous Multi-Slice (Slice Accelerated) Diffusion EPI

Driving profitable growth in Greater China. Andy Ho Chief Market Leader Greater China

X-ray detectors in healthcare and their applications

Slide 1. Slide 2. Slide 3 ACR CT Accreditation. Multi-Slice CT Artifacts and Quality Control. What are the rules or recommendations for CT QC?

Personal Information. Contact information. Education / Training. Position

2D, 3D CT Intervention, and CT Fluoroscopy

Charles Dotter and the Birth of Interventional Radiology: A Sleeping-Beauty with a Restless Sleep

Epona Medical simulation products catalog Version 1.0

Boston Scientific - Company Profile

Minnesota Historical Society

Cardiovascular Interventions

Arterial Tissue Perforation Using Ultrasonically Vibrating Wire Waveguides

ULP Wireless Technology for Biosensors and Energy Harvesting

BIOTRONIK // Celebrating 50 years of excellence BIOTRONIK. Setting the pace, pioneering the future

Alberta Health Services and Advancing Uptake of HTA & Innovation

PATIENT ADVOCACY: THE MEDTRONIC EXPERIENCE. Increase patient safety while building a more collaborative and resilient cardiac OR team.

Yinsheng Li 1, Peter Bannas 2, M.D., Perry Pickhardt M.D. 2, Meghan Lubner M.D. 2, Ke Li Ph.D. 1,2, and Guang-Hong Chen Ph.D. 1,2

The National Stock Exchange of India Ltd. Corporate Communications Department Corporate Services Department Exchange Plaza, 5th Floor,

CHINA MED DEVICE. China Medtech Will Continue Its Double Digit Growth in Years to Come

MEDICAL DEVICE INNOVATION IN GERMANY

Forward Looking Statements

Virtual Reality Gets Real in Healthcare Session 47, March 6, 2018 Rick Krohn, President, HealthSense, Inc. David Metcalf, Ph.D.

Vascular. Development of Trinias Series unity edition Angiography Systems. 1. Introduction. 3. Three Concepts of "unity"

VIRTUAL REALITY IN BIOMEDICAL

Collaborating with the Office of Technology Transfer

Participants Osama R. Mawlawi, PhD, Houston, TX (Presenter) Research Grant, General Electric Company; Research Grant, Siemens AG

Using Web-Based Computer Graphics to Teach Surgery

Disclosures of Relevant Financial Relationships

LINC MIDDLE EAST 2016

Automatic Selection of Mask and Arterial Phase Images for Temporally-Resolved MR Digital Subtraction Angiography

SIGNA Explorer Lift revives our MR

A Coupled Fluid-structure Model of a Therapeutic Ultrasound Angioplasty Wire Waveguide

2018 Mid-Atlantic Conference 8 th Annual Current Concepts in Vascular Therapies

TREND OF SURGICAL ROBOT TECHNOLOGY AND ITS INDUSTRIAL OUTLOOK

Comparison of high-resolution C-arm cone-beam CT imaging and conventional 3D rotational angiography for the better microvascular visualization

Endovascular aortic repair Fenestrated

Locating Blood Vessels in Retinal Images by Piece-wise Threshold Probing of a Matched Filter Response

Chairman Estrada asked the Secretary to call upon the registered public speakers. The Secretary called upon the following registered speaker:

STEVEN W. ORLOW, MD, MMI, FACP, FACC

Evaluation of image lag in a flat-panel, detector-equipped cardiovascular X-ray machine using a newly developed dynamic phantom

Leading the Way in Guide Wire Technology

160-slice CT SCANNER / New Standard for the Future

Anatomy of Contemporary CTO Guidewires and Selection Criteria

Asia Pacific s leading event serving the entire value chain of the medical device market. CMEF & ICMD Spring 2016 Post- Show Report

NEW DESIGN FOR S.F.A STENT : Dual Component

NOVEL THREE-DIMENSIONAL BRAIDING APPROACH

Research Support. Dual-Source CT: What is it and How Do I Test it? Cynthia H. McCollough, Ph.D.

Summer Engineering Research Internship for US Students (SERIUS) Department of Biomedical Engineering. (

HEARTS/NASCI/RSNA/SAIP/SCAI/SCCT/SCMR/SNMMI 2014 Health Policy Statement on Use of Noninvasive Cardiovascular Imaging

The Trend of Medical Image Work Station

Methods to Evaluate and Increase the RF Safety of Cardiovascular Interventional Devices During MRI. Gregory H. Griffin

How Machine Learning and AI Are Disrupting the Current Healthcare System. Session #30, March 6, 2018 Cris Ross, CIO Mayo Clinic, Jim Golden, PwC

DEMCON: High-end system supplier. Development and manufacturing of medical devices

Exploring Critical Findings and their Communication in Radiology Text Reports using Natural Language Understanding ACR 2017

JP Morgan Healthcare Conference January 9, 2007 Larry Best EVP and Chief Financial Officer

Digitalization and TITLE OF. Devices May 2018 PRESENTATION

12/21/2016. Siemens Medical Systems Research Agreement Philips Healthcare Research Agreement AAN and ASN Committees

(Biomedical Engineering Design) Spring 2009

Medical Robotics Overview of the course

Presenters. Sharon Raggio, CEO Colorado West. Monica Oss, CEO OPEN MINDS. Scott Cousino, CEO mystrength, Inc.

PREPARE. A guide to help people and their loved ones prepare for medical decision making. Name:

Integre Pro Scan combines pattern scanning and multi-color photocoagulation in our unique all-in-one laser/slit lamp design.

GMDN An Introduction. Mark Wasmuth CEO, GMDN Agency

Enhanced Functionality of High-Speed Image Processing Engine SUREengine PRO. Sharpness (spatial resolution) Graininess (noise intensity)

Teaching Digital Radiography and Fluoroscopic Radiation Protection

Medical Extrusion & Secondary Ops SPOTLIGHT Conference November 10, 2016 Arburg Technical Center Irvine, CA

Minimally Invasive Therapies Group

An Uneasy Marriage: Health Care Technology and its Relationship to Quality

EC Guideline 109: Guidance on Diagnostic Reference Levels REFERENCE LEVELS AT EUROPEAN LEVEL FOR CARDIAC INTERVENTIONAL PROCEDURES

HTA Position Paper. The International Network of Agencies for Health Technology Assessment (INAHTA) defines HTA as:

STUART B MITCHELL FOUNDER AND CEO NOVUSON SURGICAL, INC. JUNE 2018

Physician s. Zenith. pocket reference guide ENDOVASCULAR GRAFTS. Indications for use. Component ordering information. Introduction system information

SELECTIVE AND REGULATED RF HEATING OF STENT TOWARD ENDOHYPERTHERMIA TREATMENT OF IN-STENT RESTENOSIS

Software and Hardware in CCTA. Elly Castellano PhD

What to expect at your outpatient consultation. Hospitals + Health Checks + Physio + Gyms

(12) Patent Application Publication (10) Pub. No.: US 2004/ A1

April By Type of Approach- Transfemoral and Transapical. By Region- North America, Europe, APAC and RoW

Clinical Safety & Effectiveness Cohort # 11

Transcription:

Successful endovascular treatment for BTK lesion using wire rendezvous technique and retrograde knuckle wire technique by collateral approach Katsutoshi Takayama, MD, Ph.D Department of Radiology and Interventional Neuroradiology Ishinkai Yao General Hospital, Yao, Osaka, Japan

Disclosure Speaker name: Katsutoshi Takayama, M.D., ph. D I have the following potential conflicts of interest to report: Consulting Employment in industry Stockholder of a healthcare company Owner of a healthcare company Other(s) I do not have any potential conflict of interest

What is Wire rendezvous technique? Bidirectional approach using two guidewires and microcatheters to recanalize for long CTO.

What is Wire rendezvous technique?

What is Wire rendezvous technique?

What is Wire rendezvous technique?

What is Wire rendezvous technique? Antegrade approach for recanalization is not successful.

What is Wire rendezvous technique?

What is Wire rendezvous technique?

What is Wire rendezvous technique? Retrograde approach for recanalization is not successful.

What is Wire rendezvous technique?

What is Wire rendezvous technique?

What is Wire rendezvous technique? Advance antegrade wire and retrograde micro into CTO lesion.

What is Wire rendezvous technique?

What is Wire rendezvous technique? Advance antegrade guidewire into the retrograde microcatheter (Rendezvous).

What is Wire rendezvous technique?

What is Wire rendezvous technique?

What is Wire rendezvous technique? Advance antegrade micro beyond CTO segment.

What is Wire rendezvous technique? Advance antegrade micro beyond CTO segment.

What is Wire rendezvous technique? Exchange antegrade guidewire

What is Wire rendezvous technique?

What is Wire rendezvous technique?

What is Wire rendezvous Perform PTA. technique?

What is Wire rendezvous technique?

What's the key benefits? Minimize subintimal tracking Less traumatic for recanalization Goose neck wire is not necessary Much higher recanalization rate of long CTO

What is benefit of knuckle wire technique? Journal of Clinical and Diagnostic Research. 2016 Sep, Vol-10(9):

What is benefit? Knuckle wire could be advanced within the sub-intimal space without perforation.

What is benefit? The loop could be advanced within the subintimal space without causing perforation. Technical success rate 86% Clinical success rate 80% Using 0.035 inch wire J Endovasc Ther 2002;9:411 416.

Case Female / 80 year-old C/C : Ischemic rest pain (Fontaine classification: III, Rutherford category 4) P/Hx : DM, HL 7 years ago->medication Tx Percutaneous Coronary Intervention 6 years ago Laparoscopic cholecystectomy 7 years ago Stenting for bilateral SFA stenosis 1 years ago Stenting for bilateral CIA stenosis 1 years ago

CTA

CTA

CTA

long CTO of left BTK lesion 80F

long CTO of left BTK lesion 80F

My strategy in this case Firstly I try to cross ATA occlusion using microcatheter and 0.014 inch guidewire by antegrade approach. After recanalization I try to cross peroneal artery occlusion using knuckle wire technique by retrograde collateral approach. And finaaly I planed to cross peroneal artery occlusion using Rendezvous technique.

Recanalization of left ATA occlusion

POBA for left ATA Rapid Cross 2.5/3mm x 21cm (Medtronic Inc, Minneapolis, MN)

POBA for left Popliteal artery Rapid Cross 2.5/3mm x 21cm

POBA for left popliteal artery Rapid Cross 2.5/3mm x 21cm

Post POBA

Post POBA

Collateral approach Prominent Bta, GT 0.014 inch 45 angle

Collateral approach Prominent Bta GT 0.014 inch 45 angle

Rendezvous Technique Retro : prominent Bta, Ante : prominent NEO GT 0.014 inch 45 angle 135cm

Cross the lesion

POBA for peroneal artery occlusion PTA balloon 3mm x 15cm

POBA for peroneal artery occlusion PTA balloon 3mm x 15cm

Post PTA Post

Post

ABI Pre : 0.57 Post : 0.90 Post

Conclusion Wire rendezvous and retrograde knuckle wire technique by collateral approach may be safe and effective for the long CTO of BTK lesion.

Successful endovascular treatment for BTK lesion using wire rendezvous technique and retrograde knuckle wire technique by collateral approach Katsutoshi Takayama, MD, Ph.D Department of Radiology and Interventional Neuroradiology Ishinkai Yao General Hospital, Yao, Osaka, Japan