Perform instrument startup and verification check following the manufacturer s recommendations.

Similar documents
Perform instrument startup and verification check following the manufacturer s recommendations.

Setup Procedure for Beckman Coulter CytoFLEX Flow Cytometer

Technical Bulletin. Guide for Using BD Cytometric Bead Array (CBA) Flex Sets with the BD Accuri C6 Flow Cytometer. Introduction

BD OneFlow Setup Beads

LEGENDplex. Please read the entire manual before running the assay. BioLegend.com. Please read the entire manual before running the assay.

Enabling Legendary Discovery. LEGENDplex. LEGENDplex. Mul -Analyte Flow Assay Kit. Mul -Analyte Flow Assay Kit

LEGENDplex. Please read the entire manual before running the assay. BioLegend.com. Please read the entire manual before running the assay.

Enabling Legendary Discovery. LEGENDplex. LEGENDplex. Mul -Analyte Flow Assay Kit

LEGENDplex Mul -Analyte Flow Flow Assay Kit Kit

capabilities today. Flexibility for tomorrow.

ab Firefly Cytometer Setup Particles

LEGENDplex. Cat. No Human Proinflammatory Chemokine Panel (13-plex) Please read the entire manual before running the assay.

Enabling Legendary Discovery. LEGENDplex. LEGENDplex. Mul -Analyte Flow Assay Kit. Please read the entire manual before running the assay.

LEGENDplex. Cat. No Human Cytokine Panel 2 (13-plex) Please read the entire manual before running the assay.

LEGENDplex. Cat. No Human CD8/NK Panel (13-plex ) Please read the entire manual before running the assay.

BD FC Beads 7-Color Kit

AbC Total Antibody Compensation Bead Kit

BD Trucount Controls IVD

BD FC Beads 7-Color Kit

Flow cytometer instrument set up

Manual Rosetta Calibration v

Quick Guide. NucleoCounter NC-3000

30 Plex Human Luminex (Invitrogen Kit, Single Plate)

BD LSRFortessa X-20. Special Order Product. Designed for limited space and boundless potential

Průtokový cytometr CytoFLEX

LEGENDplex Mul -Analyte Flow Flow Assay Kit Kit

Trademarks Cytek, the Cytek logo, and all other trademarks are property of Cytek Biosciences Cytek

MAKE SURE YOUR SLIDES ARE CLEAN (TOP & BOTTOM) BEFORE LOADING DO NOT LOAD SLIDES DURING SOFTWARE INITIALIZATION

MicroLab 500-series Getting Started

Comparison of the Analysis Capabilities of Beckman Coulter MoFlo XDP and Becton Dickinson FACSAria I and II

SOP-P094. BioMek 2000 Compound Protocol Cyan/Hypercyt + Analysis

PAVING THE ROAD TO UTOPIA VIA INSTRUMENT STANDARDIZATION

How to Create and use Manual Linearization Profiles in Matan PhotoPrint Server

a Beckman Coulter Life Sciences: White Paper

BD LSRFortessa X-20. Special Order Product. Designed for limited space and boundless potential

You can easily print images using the Capture NX print function. Here we will explain the process for printing

MIF ZEISS VIOLET CONFOCAL ZEN 2009 PROTOCOL

Estimated Time Required to Complete: 45 minutes

Excel Lab 2: Plots of Data Sets

Calculate Ratiometric FRET-Images with the FRET-Image-Script

Spectradyne ncs1 TM. Operation Manual. Spectradyne LLC

ScanArray Overview. Principle of Operation. Instrument Components

Importing and processing gel images

How to Make a Run Chart in Excel

EE 210 Lab Exercise #3 Introduction to PSPICE

MATHEMATICAL FUNCTIONS AND GRAPHS

INTRODUCTION TO DATA STUDIO

Working with your TADS Information in Cornerstone

Best Practices with the Bio-Plex system. Alasdair I Reid Ph.D. Field Application Specialist Bio-Rad Laboratories

SHORT GUIDE TO LASER MICRODISSECTION USING THE PALM COMBI SYSTEM

Box Tray Geometry in MasterCAM

This tutorial will lead you through step-by-step to make the plot below using Excel.

Zeiss 780 Training Notes

Batch Counting of Foci

Cytometry tutorial: The impact of adjusting PMT voltages on spillover and compensation

DC Operating Point, I-V Curve Trace. Author: Nate Turner

Volocity Tutorial Creating a Measurement Protocol in Volocity Software

Title: Leica SP5 Confocal User Manual

facewho? Requirements Analysis

Excel Tool: Plots of Data Sets

Lanyon Smart Events Cloud. Meetings and Events. DocuSign Integration

Chapter 6 Title Blocks

USTER TESTER 5-S800 APPLICATION REPORT. Measurement of slub yarns Part 1 / Basics THE YARN INSPECTION SYSTEM. Sandra Edalat-Pour June 2007 SE 596

FlashChart. Symbols and Chart Settings. Main menu navigation. Data compression and time period of the chart. Chart types.

Creating a Watercolor Painting From a Photo. Open the photo you want to appear to be a watercolor painting. I am using the photo below:

Supplemental Reference Guide

Logging into Five9 Telephony

SCHEDULE USER GUIDE. Version Noventri Suite Schedule User Guide SF100E REV 08

Census Data Tools. Hands-on exercises July 17 & 19, LULAC National Convention

Image Capture TOTALLAB

Swept-Field User Guide

Ansoft Designer Tutorial ECE 584 October, 2004

TeleTrader FlashChart

Creating a Color Compensation file for Roche LightCycler 480 vers.i & vers. II

Axioscan - Startup. 1. Turn on the Axioscan (button to the left) and turn on the computer. 2. Log on and start the ZEN Blue software from the desktop

The operation manual of spotlight 300 IR microscope

b. Turn the power switch and key to on position for blue laser.

Section Table of Contents: Section 16.0

Working With Drawing Views-I

DOSELAB TOMOTHERAPY TG-148 QA QUICK GUIDE TG-148 RECOMMENDED TESTS 1. V.B.1.C. - Y-JAW DIVERGENCE/BEAM CENTERING

XI. Rotary Attachment Setups

SHORT GUIDE TO LASER MICRODISSECTION USING THE PALM COMBI SYSTEM

Chlorophyll Fluorescence Imaging System

Ricoh SG 3110DN - MacProfile

GXCapture 8.1 Instruction Manual

quantiflash Calibration Light Source for Cytometry

Create all plan and profile sheets in the current drawing. Create all plan and profile sheets in individual drawings.

Leica SPEII confocal microscope. Short Manual

EmbryoCellect. RHS Scanning and Analysis Instructions. for. Genepix Pro Software

The Basic RAW Image Processing Workflow In PSE 2018

BoTest Matrix E Botulinum Neurotoxin Detection Kit Protocol

IncuCyte ZOOM Scratch Wound Processing Overview

Veterinary Digital X-Ray System Quick Start Guide

Creating and Editing Plot Style Tables

LIGHT STUDIO 485. Software Instrrctions Mancal. Read trhe mancal tarefclly before operaing trhis software

MYGRAPHICSLAB: ADOBE ILLUSTRATOR CS6

Training Guide for Carl Zeiss LSM 7 MP Multiphoton Microscope

Minolta Scanner Plugin

The Zeiss AiryScan System, Confocal Four.

Celigo Cytometer User Guide

Transcription:

Setup procedure for guava easycyte with a Single Laser Important note: The guava instruments are based on microcapillary flow. This technology offers some advantages for general flow cytometry but the instruments are different from traditional flow cytometry and the performance of LEGENDplex is also different. The Forward and Side Scatter profile of the beads look different and the guava instrument may have a lower resolution in the Forward Scatter channel. While the guava easycyte may not be the best instrument for LEGENDplex in this regards, following these instructions will help in getting the best data possible. For comparison of data generated in different experiments machine settings should remain the same. Analysis of the data should also remain as similar as possible with the LEGENDplex software. 1. Start up the Instrument Perform instrument startup and verification check following the manufacturer s recommendations. 2. Select the Application from the main Menu: Several applications, which are called assays, are provided in the guavasoft package. This guide is written based in ExpressPlus and ExpressPro. The ExpressPlus Assay is for performing one-, two-, and three-color assays. ExpressPro is for performing assays with up to six colors, or where time, area, and/or width parameters are necessary. InCyte can also be used to collect six colors, time, area, and/or width parameters. When using instruments with two lasers there is no need to do compensations. Instructions for ExpressPro are also provided. 2.1 Select ExpressPlus from the dropdown menu (Figure 1) 1

Figure 1. 2.2 Create a Worklist following the guava EasyCyte User s manual. Follow the software instructions if prompted to put required tubes/solutions for instrument tray check. If you created a Worklist already, click Start Worklist. 3. Select an instrument settings option. 3.1 Click Settings (Figure 2) Note: To define, edit and allocate gates any time during the procedure please refer to the guava easycyte User s guide. 2

Figure 2. 3.2 A window will pop up (Figure 3). To adjust instrument settings, click Adjust Settings. If you performed this step in a previous LEGENDPlex assay, retrieve the settings by clicking Retrieve Settings. Select a setting file and click Open. Figure 3. 3.3 Once Adjust Settings is selected, a dialog box appears prompting you to select the sample for adjusting settings (Figure 4). Click to select the well/tube used to adjust settings, then click OK. 3

Figure 4. 3.4 The Adjust Settings screen appears (Figure 5), allowing you to adjust the threshold, the PMT voltages, and optimize the display of the data. 4. Set up PMT Voltages and Compensation The Setup Beads 3: Raw Beads are used to set up the PMT voltage of the Classification Channel RED and channel GRN. The Setup Beads 2: PE Beads are used to set up the PMT voltage of the reporter channel YLW. All three setup beads are needed for setting up Compensation. Follow the instructions below for setting up the PMT voltages and compensation: 4.1 Sonicate the Raw Beads for 1 minute in a bath sonicator to resuspend and reduce aggregates. If you do not have access to a sonicator vortex the vial for at least 2 minutes. 4.2 Transfer 400μL of the Raw Beads to a fresh FACS tube. 4.3 Make sure the events rate is within the recommended range of the machine (usually the recommended flow rate is medium; the events/μl should be less than or equal to 500). 4.4 Carefully adjust the gain and voltage for FSC and SSC so that both beads populations are visible and easy to gate. Adjust the FSC threshold to exclude unwanted events (Figure 5). 4.5 Adjust the GRN voltage so that the signal for the majority of beads is between 10 0 and 10 1 (Figure 5). 4.6 Adjust the voltage for YLW in the fluorescent dot plots so that the signal for majority of beads is between 10 0 and 10 1 (Figure 5). 4

Figure 5. 4.7 Adjust the RED settings so that the signal for all beads is between 10 1 and 5 x 10 3 (Figure 6). Figure 6. 5

4.8 Sonicate for 1 minute or vortex the vial of the Setup Beads 1: FITC Beads for 2 minutes to resuspend the beads and minimize aggregates. 4.9 Transfer 200 μl of the FITC Beads to a fresh FACS tube. Add 200 μl of Raw Beads and mix well (This will generate FITC-positive and FITC-negative populations of beads and is needed for proper compensation). 4.10 Run the mixed FITC and Raw Beads. 4.11 On the GRN vs YLW dot plot (Figure 7), the beads will display as FITC- negative and FITC-positive population (indicated by an arrow). Note: FITC beads are only of small size, falling in the Beads A gate. Figure 7. 4.12 Adjust the YLW-%GRN compensation setting so that FITC negative and FITC positive populations have similar mean YLW Fluorescence intensity (Figure 8). 6

Figure 8. 4.13 Sonicate for 1 minute or vortex the vial of the Setup Beads 2: PE Beads for 2 minutes to resuspend the beads and minimize aggregates. 4.14 Transfer 200 μl of the PE Beads to a fresh FACS tube. Add 200 μl of Raw Beads and mix well (This will generate PE-positive and PE-negative populations of beads and is needed for proper compensation. 4.15 Run the mixed PE and Raw Beads. 4.16 On the GRN vs YLW dot plot (Figure 9), the beads will display as PE- negative and PE-positive population (indicated by an arrow). Note: PE beads are only of small size, falling in the Beads A gate. 4.17 Adjust the YLW settings so that median fluorescent intensity of the PE beads falls within the lot-specific range labeled on the PE Beads vial. 7

Figure 9. 4.18 Adjust the GRN-%YLW compensation setting so that PE negative and PE positive populations have similar mean GRN fluorescence intensity. Figure 10. 8

4.19 On the RED vs YLW dot plot, the mixed beads now display as PE negative and PE positive populations (Figure 11, indicated by an arrow). Figure 11. 4.20 Adjust the RED-%YLW compensation setting so that the RED-low population and the PE positive population have similar median RED signal. Figure 12. 9

4.21 The flow cytometer is now compensated and ready for sample analysis. Additional recommendations when collecting and analyzing your samples: 1) Collect 2500 beads in either bead gate A or B, or about 300 beads per analyte. Do not collect just the 5,000 events total that is set as default in ExpressPro or ExpressPlus. Do not collect more than 10,000 beads as this could cause problems during data analysis. 2) When analyzing the samples with the LEGENDPlex Data Analysis Software do manual gating to select the beads, do not use the autogate function. Save the settings and try to use the same gating for subsequent experiments. 10