High specification CCD-based spectrometry for metals analysis

Similar documents
Thermo Scientific icap 7000 Plus Series ICP-OES: Innovative ICP-OES optical design

Performance Comparison of Spectrometers Featuring On-Axis and Off-Axis Grating Rotation

Experimental Analysis of Luminescence in Printed Materials

Superior ICP-OES optical design for unmatched speed and performance

For Client Review Only. All Rights Reserved. Advanstar Communications Inc. 2005

CONFIGURING. Your Spectroscopy System For PEAK PERFORMANCE. A guide to selecting the best Spectrometers, Sources, and Detectors for your application

True simultaneous ICP-OES for unmatched speed and performance

QE65000 Spectrometer. Scientific-Grade Spectroscopy in a Small Footprint. now with. Spectrometers

ULS24 Frequently Asked Questions

University of Wisconsin Chemistry 524 Spectroscopic Components *

metcon meteorologieconsultgmbh, Instruments for Atmospheric Research W1aa_Feb_2017_1.doc 1 -

Abstract No. 32. Arne Bengtson and Tania Irebo. Swerea KIMAB AB, Isafjordsgatan 28A, SE Kista, Sweden

Improving the Collection Efficiency of Raman Scattering

Components of Optical Instruments

Maya2000 Pro Spectrometer

BaySpec SuperGamut OEM

SCCH 4: 211: 2015 SCCH

DETECTORS Important characteristics: 1) Wavelength response 2) Quantum response how light is detected 3) Sensitivity 4) Frequency of response

Improved Spectra with a Schmidt-Czerny-Turner Spectrograph

Spectroscopy in the UV and Visible: Instrumentation. Spectroscopy in the UV and Visible: Instrumentation

Chemistry Instrumental Analysis Lecture 7. Chem 4631

Chemistry 524--"Hour Exam"--Keiderling Mar. 19, pm SES

TriVista. Universal Raman Solution

Applications Information

Applications of Steady-state Multichannel Spectroscopy in the Visible and NIR Spectral Region

CHAPTER 7. Components of Optical Instruments

Properties of a Detector

Introduction to the operating principles of the HyperFine spectrometer

Observational Astronomy

Better Imaging with a Schmidt-Czerny-Turner Spectrograph

Aqualog. CDOM Measurements Made Easy PARTICLE CHARACTERIZATION ELEMENTAL ANALYSIS FLUORESCENCE GRATINGS & OEM SPECTROMETERS OPTICAL COMPONENTS RAMAN

Oriel MS260i TM 1/4 m Imaging Spectrograph

2013 LMIC Imaging Workshop. Sidney L. Shaw Technical Director. - Light and the Image - Detectors - Signal and Noise

Spectral Analysis of the LUND/DMI Earthshine Telescope and Filters

Content Spectrophotometers

8.2 Common Forms of Noise

Spark Spectral Sensor Offers Advantages

CCDS. Lesson I. Wednesday, August 29, 12

MS260i 1/4 M IMAGING SPECTROGRAPHS

Specifications Summary. Register well depth (typical) Standard mode High Capacity mode High Sensitivity mode. Maximum spectra per sec 1,612

Dual-FL. World's Fastest Fluorometer. Measure absorbance spectra and fluorescence simultaneously FLUORESCENCE

DESIGN AND CHARACTERIZATION OF A HYPERSPECTRAL CAMERA FOR LOW LIGHT IMAGING WITH EXAMPLE RESULTS FROM FIELD AND LABORATORY APPLICATIONS

ECEN. Spectroscopy. Lab 8. copy. constituents HOMEWORK PR. Figure. 1. Layout of. of the

Aqualog. Water Quality Measurements Made Easy PARTICLE CHARACTERIZATION ELEMENTAL ANALYSIS FLUORESCENCE

Photons and solid state detection

MTF and PSF measurements of the CCD detector for the Euclid visible channel

X-Ray Spectroscopy with a CCD Detector. Application Note

Gamma Spectrometer Initial Project Proposal

the need for an intensifier

PCS-150 / PCI-200 High Speed Boxcar Modules

Company synopsis. MSU series

Historical. McPherson 15 Mount

ROTATING SHADOWBAND SPECTRORADIOMETER MODEL RSS-1024/UVRSS-1024 BULLETIN RSS/UVRSS-1024

Image Capture TOTALLAB

PUV3402 LED multiwave photometer A new approach to online process photometry

HR2000+ Spectrometer. User-Configured for Flexibility. now with. Spectrometers

The only simultaneous absorbance and f uorescence system for water quality analysis! Aqualog

Noise Analysis of AHR Spectrometer Author: Andrew Xiang

WIDE ANGLE GEOMETRY EDXRF SPECTROMETERS WITH SECONDARY TARGET AND DIRECT EXCITATION MODES

Camera Test Protocol. Introduction TABLE OF CONTENTS. Camera Test Protocol Technical Note Technical Note

Aqualog. Water Quality Measurements Made Easy FLUORESCENCE

SPECTRAL SCANNER. Recycling

An Introduction to the Silicon Photomultiplier

Prodigy DC Arc. The Ultimate Solution for Elemental Analysis of Solid Samples

Radiation transducer. ** Modern electronic detectors: Taking the dark current into account, S = kp + bkgnd over the dynamic range.

event physics experiments

OPTI510R: Photonics. Khanh Kieu College of Optical Sciences, University of Arizona Meinel building R.626

Miniature Spectrometer Technical specifications

contraa Hit the Mark! High-Resolution Continuum Source AAS

3 General Principles of Operation of the S7500 Laser

Upgrade of the ultra-small-angle scattering (USAXS) beamline BW4

Control of Noise and Background in Scientific CMOS Technology

LYNXEYE XE. Innovation with Integrity. High-Resolution Energy-Dispersive Detector for 0D, 1D, and 2D Diffraction XRD

CHAPTER 9 POSITION SENSITIVE PHOTOMULTIPLIER TUBES

Basic Components of Spectroscopic. Instrumentation

Spectrophotometer. An instrument used to make absorbance, transmittance or emission measurements is known as a spectrophotometer :

INTERNATIONAL STANDARD

The HERA-B Ring Imaging Cerenkov ˇ Detector

Welcome to: LMBR Imaging Workshop. Imaging Fundamentals Mike Meade, Photometrics

A Measurement of the Photon Detection Efficiency of Silicon Photomultipliers

IBIL setup operation manual for SynerJY software version

Available online at ScienceDirect. Procedia Technology 17 (2014 )

Minimizes reflection losses from UV-IR; Optional AR coatings & wedge windows are available.

Isolator-Free 840-nm Broadband SLEDs for High-Resolution OCT

Optical In-line Control of Web Coating Processes

Performance of High Pixel Density Multi-anode Microchannel Plate Photomultiplier tubes

SpectraPro 2150 Monochromators and Spectrographs

Multi-channel imaging cytometry with a single detector

AIXUV's Tools for EUV-Reflectometry Rainer Lebert, Christian Wies AIXUV GmbH, Steinbachstrasse 15, D Aachen, Germany

Cerro Tololo Inter-American Observatory. CHIRON manual. A. Tokovinin Version 2. May 25, 2011 (manual.pdf)

Total Absorption Dual Readout Calorimetry R&D

Where detectors are used in science & technology

Ultrafast instrumentation (No Alignment!)

Lecture 30: Image Sensors (Cont) Computer Graphics and Imaging UC Berkeley CS184/284A

Report on BLP Spectroscopy Experiments Conducted on October 6, 2017: M. Nansteel

GenePix Application Note

SPECTROGRAPHS FOR ANALYZING NANOMATERIALS

Parameter Selection and Spectral Optimization Using the RamanStation 400

Development of a spectrometry system Using lock-in amplification technique

F/48 Slit Spectroscopy

Transcription:

High specification CCD-based spectrometry for metals analysis New developments in hardware and spectrum processing enable the ARL QUANTRIS CCD-based spectrometer to achieve the performance of photo-multiplier tube instruments for the determination of low levels of carbon, nitrogen, sulphur and phosphorus with excellent precision, coupled with an equivalent or superior precision for the measurement of key alloying elements. Emile Muller, Jean-Marc Böhlen and Edmund Halász Thermo Electron The analytical demands of the metal industry are constantly increasing in terms of detection limits, precision, accuracy, speed of analysis and flexibility. More and more metal processors are looking for reliable instruments that can analyse a greater variety of material grades, yet which are flexible and need no hardware modifications for further extensions. Optical Emission Spectrometry (OES) for elemental analysis of metals offers a high level of performance at extremely high speed, coupled with simple sample preparation and low cost of ownership. Spectrometer manufacturers rapidly adopted solid-state detectors because they allow the design of smaller optics and hence small instruments, like portables and bench tops. Among the solid-state detectors available, charge coupled devices (CCDs) already becoming an industry standard in digital photography, scientific and medical imaging were adopted in OES. H o w e v e r, until recently, it was thought that CCDbased instruments, despite their high versatility, had lower performance in comparison with photomultiplier tube (PMT) instruments, in particular having higher detection limits and lower precision. PMTbased instruments, however, especially for multimatrix applications, do have geometric limitations. This article introduces the ARL QUANTRIS (Figure 1) produced by the Thermo Electron Corporation a second-generation optical emission spectrometer based on CCD technology. New developments in Figure 1 The ARL QUANTRIS CCD-based metals a n a l y s e r hardware and spectrum processing enable the instrument to achieve the performance of PMT instruments with, for instance, the determination of carbon, nitrogen, sulphur and phosphorus in steels at low levels, and with excellent precision. Instrument architecture The instrument has an innovative architecture, built on proven modern technologies and consists of three main modules: the spark stand with the CCS, the optics with the CCD detectors, and the control and readout electronics. Sample stand and CCS The analysis stand is water-cooled for better heat evacuation, and has a very low stand-by flow and a fast flush and dust blow-out system. The sample excitation source is the patented Thermo CCS. This source was introduced in 1994 on the high-specification ARL 4460 340

Flat field Wavelength Grating Average Average band s p e c t r o g r a p h range (nm) (groove/mm at d i s p e r s i o n pass par pixel grating centre) ( n m / m m ) ( p m / p i x e l ) VUV spectrograph 130-200 3240 1. 2 8 Basic spectrograph 200-410 1, 105 3. 5 24 Optional alkaline s p e c t r o g r a p h 410-780 590 6. 7 43 Table 1 ARL QUANTRIS optics characteristics Figure 2 Schematics of the optics of the ARL QUA N T R I S Figure 3 (a) Example of spectrum pixe l i s a t i o n, (b) The effect of 0.25 pixel drift to the right on the above spectrum PMT spectrometer and this is the first time that such a high-performance source has become available on a C C D-based instrument, offering significant advantages in comparison to any other spark generator currently used for OES. It is the only servo-controlled d i g i t a l source on the market and its high degree of flexibility enables all analytical figures to be optimised for each type of metal and alloy. Flat field optics with CCD detectors The CCD of the ARL QUANTRIS is specifically designed for high-specification industrial, scientific or military applications and has three parallel lines of 8,640 detectors of 7 by 9.8 micrometers. The optics are based on the concept of the flat field spectrograph and consist of a primary slit and a flat field grating that focuses the light onto a plane surface, where a CCD detector is placed. There are three flat- f i e l d spectrographs (see Figure 2), covering the spectral ranges: 130 200, 200 410 and 410 780nm. Each of them has its own focusing lens and holographic a b e r r a t i o n -corrected concave grating, enabling optimised light collection on the high-performance linear CCD. The first spectrograph CCD is covered with a fluorescent coating for increased sensitivity and the detectors are also cooled (with Peltier coolers) and t h e r m o-controlled within ±0.06 C at a temperature close to 0 C. The middle spectrograph is oriented vertically and the two others horizontally. The letters indicate the position of the light source (A), the lens (B), the primary slit (C), the grating (D), and the CCD (E) of the third spectrograph. The CCD replaces the photomultiplier detectors used in traditional spectrometers and gives the instrument immense flexibility. The spectrometer body is made of cast iron, operates under vacuum, and ensures excellent short and long-term temperature and ambient pressure s t a b i l i t y, together with minimal absorption of light in the ultra-violet (UV) domain. The characteristics of the optics are shown in Table 1. The light is directed to the optics in directreading mode. This is a key feature because it eliminates the need for fibre optics that suffer from effects that degrade light transmission such as aging and solarisation. Control and readout electronics All instrument functions are monitored and controlled by an internal controller. 341

The spectra from each CCD are stored in a file to allow graphical display and are processed to calculate peak intensities or element concentrations. The pixel scale is converted into wavelength range through a calibration function. For each measuring program analytical wavelengths are defined and spectral information is converted into concentrations. Apart from the new architecture of the hardware, real progress in the instrument s performance has been made thanks to mathematical treatment of the spectra. Several problems had to be overcome by mathematical treatment and measurement strategy: Figure 4 Comparison between a raw spectrum and a geometrically interpolated spectrum with enhanced digital resolution The CCD readings are processed by dedicated powerful 32 bit RISC ARM signal processors and digitised by triple 16 bit A/D converters. The communication with the external PC is done via Fa s t Ethernet (100Mbit/s) and TCP/IP protocol. Novel spectrum processing The pixels of the CCD have a photosensitive layer in which incident photons are detected, and are connected with a potential well where the resulting photoelectrons are collected. At the end of the acquisition period the content of the well is gated out, processed and digitally converted. Elementary acquisition parameters, including the elementary acquisition period, are defined for each CCD and for every matrix, in order to optimise the intensity of lines of interest, while avoiding saturation. Spectrum pixelisation Because the pixel acts as a wavelength integrator over the pixel band pass, the spectrum obtained has a histogram shape that may hide details. As a consequence, integration windows for analytical wavelength are difficult to define with high accuracy. Figure 3a is an example of a peak with Full Width at Half Maximum (FWHM) of 2.5 pixels. D r i f t Drift is unavoidable. With a pixel width of only 7 micrometers, it is sensitive to the drift of the spectrum. Figure 3b represents a simulation of drift of the peak displayed in Figure 3a of 2 microns, which is equivalent to an 8pm displacement of the spectrum on the second spectrograph. S e n s i t i v i t y Ty p i c a l l y, the sensitivity of CCD pixels is between one and three orders of magnitude lower than the sensitivity of a PMT. A way to compensate for lower sensitivity is to use the information from more pixels instead of the peak pixel by increasing the active band pass for the analytical window and to reduce the spectral noise. Spectral noise can be defined as pixel flickering as the result of summation of all sources of noise applied to a pixel. Width of the Peak FWHM Relative intensity Relative intensity integration VUV spectrograph variation for variation for w i n d ow (pixels) Peak FWHM truncation error truncation error S t a n d a r d of 1 pixel (%) of 1/ 4 pixel (%) s p e c t r o g r a p h 7 4 9. 5 2. 0 2. 5 2. 2 0. 2 5 4 17. 2 3. 8 2. 5 10. 3 1. 5 3 4 32. 7 7. 7 2. 5 30. 7 5. 9 Table 2 Simulation of integrated peak intensity jumps caused by truncation error during drift correction E l e m e n t SD no SD with Improvement f i l t e r i n g f i l t e r i n g f a c t o r A l 1. 1 0. 5 2. 2 C 3. 0 1. 6 1. 9 C r 5 1. 5 3. 3 C u 2 0. 5 4 M n 40 6 6. 7 M o 6 3. 2 1. 9 N 31 1. 6 19. 4 N i 13. 5 5 2. 5 P 20 2. 3 8. 7 S 6 1. 8 3. 3 S i 15 4. 8 3. 1 Table 3 Typical improvements by applying digital filtering on pure iron sample 5 RE12 342

N o i s e The CCD detectors and the readout electronics are subject to different type of noise that can be referred to as thermal noise related to thermally generated charges in the CCD structure, photonic or shot noise related to statistical fluctuations of the flux of incident photons, and noise related to the amplifier and signal processor. The thermal noise and the shot noise follow the Poisson law, being expressed as the square root of the number of electrons or photons. Part of the thermal noise is suppressed by cooling the CCD for, as with a silicon structure, the thermal noise doubles every 16ºC in the range 0 70ºC. Spectral interferences OES spectra are very complex and, depending on E L E M E N T A l B B i C C a C e C o C r C u DL [ppm] 6. 3 1. 2 9 6. 2 0. 8 28 15 6 4. 5 Precision 0. 001 0. 00002 0. 00002 0. 0001 0. 005 0. 00015 0. 00006 0. 0002 0. 0002 0. 0002 0. 0003 0. 00025 0. 00015 0. 01 0. 00015 0. 0001 0. 00025 0. 0003 0. 0004 0. 0007 0. 0003 0. 0003 0. 00015 0. 05 0. 0003 0. 0005 0. 00065 0. 0005 0. 001 0. 0005 0. 00035 0. 0003 0. 1 0. 0004 0. 001 0. 0011 0. 0007 0. 0015 0. 0008 0. 0005 0. 00045 0. 5 0. 0015 0. 005 0. 002 0. 003 0. 0015 0. 002 1 0. 003 0. 01 0. 004 0. 005 0. 003 0. 0035 5 0. 017 0. 032 0. 009 0. 016 10 0. 085 0. 015 20 0. 03 40 0. 06 Table 4 A R L Q UA N T R I S detection limits (3 sigma) and typical precision values (1 sigma) for iron base E L E M E N T L a M g M n M o N N b N i P P b DL [ppm] 8. 5 0. 7 25 30 16 10 20 20 13 Precision 0. 001 0. 00002 0. 005 0. 00025 0. 0001 0. 0008 0. 0005 0. 0004 0. 0005 0. 0004 0. 0004 0. 01 0. 0004 0. 0002 0. 0008 0. 0008 0. 00055 0. 0004 0. 0005 0. 0005 0. 00075 0. 05 0. 0016 0. 0015 0. 0009 0. 001 0. 00075 0. 0007 0. 0007 0. 0008 0. 003 0. 1 0. 0030 0. 0035 0. 0011 0. 0015 0. 001 0. 001 0. 0009 0. 0015 0. 5 0. 0025 0. 004 0. 005 0. 004 0. 002 0. 007 1 0. 0045 0. 007 0. 02 0. 009 0. 004 0. 02 5 0. 020 0. 04 0. 09 0. 014 10 0. 045 0. 10 0. 025 20 0. 10 0. 05 40 0. 09 E l e m e n t S S b S i S n Ti V W Z n Z r DL [ppm] 13 5 18 10 6 15 150 6 11 P r e c i s i o n 0. 001 0. 005 0. 0003 0. 00015 0. 0004 0. 00025 0. 0002 0. 0005 0. 0002 0. 0003 0. 01 0. 0004 0. 0003 0. 0004 0. 00025 0. 0003 0. 0005 0. 00025 0. 0004 0. 05 0. 001 0. 0007 0. 0005 0. 0005 0. 00065 0. 0007 0. 0035 0. 0008 0. 001 0. 1 0. 002 0. 001 0. 0007 0. 001 0. 001 0. 001 0. 004 0. 002 0. 5 0. 002 0. 005 0. 003 0. 006 0. 01 1 0. 0035 0. 01 0. 006 0. 01 5 0. 017 0. 03 0. 03 10 0. 06 0. 06 20 0. 13 40 343

matrix and sample excitation parameters, between 20,000 and 30,000 lines are visible in the range of 130 780nm. Original and robust treatments have been successfully applied to the ARL QUA N T R I S spectra, leading to top analytical performance. Each of these treatments solves one of the above problems, but the final effects are interconnected. Enhancement of numerical resolution An original mathematical treatment for numerical resolution enhancement was developed for ARL Q UA N T R I S. The treatment corresponding to a geometrical interpolation enhances the numerical resolution by reducing the step dimension of the spectrum. Figure 4 shows a comparison between a raw spectrum and a geometrical interpolated spectrum with a numerical resolution enhancement factor of 4 (ie, numerical resolution is 0.25 pixel after treatment). The numerical resolution enhancement improves accuracy in peak intensity, peak position and spectral resolution. The letters indicate the accuracy improvement for the peak intensity (a), peak position (b) and resolution (c). Drift correction The peak positions of well-defined and resolved lines, computed by special peak-fitting algorithms, are compared to reference positions on the CCDs. The drift correction algorithm moves and deforms the spectrum in order to find the smallest differences with the reference lines. By introducing numerical resolution enhancement, the accuracy of drift correction increases and consequently the intensity jumps as the problem of rounding to the nearest pixel diminishes. Table 2 shows the results of simulations of the intensity jumps due to truncation, for various symmetrical integration windows on two peaks (the integration windows are symmetrical and centred on the peak pixel): one situated on the VUV spectrograph, with a FWHM of 4 pixels, the second situated on the standard spectrograph with FWHM of 2.5 pixels. Spectrum noise reduction Several low-pass digital filters were adopted on the ARL QUA N T R I S. The filters are matched to the characteristics of the spectral region of interest through resolution and integration window width. Digital filtering improves reproducibility and lowers the detection limits. Table 3 shows the typical improvements for a pure iron sample 5RE12, expressed as standard deviations (SD) obtained on 10 measurements, without and with filtering. Flexible intensity computation Calculating the line intensity by summation over an integration window increases CCD sensitivity so window widths are rigorously selected in order to minimise spectral overlaps. Off peak and on peak background subtractions are implemented for selected lines in order to lower detection limits and to compensate for interferences. One of the biggest advantages of having the whole spectrum available is the capability to select for each analytical line, the b e s t-suited internal standard line that behaves similarly and compensates for the instrumental changes. By selecting homologous-type internal standard lines, reproducibility is increased by a factor of 2 to 4. This is clearly a strong advantage over PMT instruments, where the number of available lines is limited. For the iron base, 14 carefully tested internal standard lines are selected, providing excellent performance. Detection limits and precision Table 4 shows the actual performances obtainable for analysis needed in the steel industry. The detection limits (DL) and the typical precision levels are similar to those attainable with PMT instruments. The instrument can be factory calibrated for numerous variants of the iron matrix, such as low alloy steels, Cr-Ni steels, Cr steels, cast and nodular iron, high speed and tool steels. C o n c l u s i o n s Thanks to the new hardware concept and the novel spectrum treatment, the ARL QUANTRIS bridges the performance gap between CCD and PMT i n s t r u m e n t s. Even if the resolution is limited and the sensitivity of the CCD detector is lower in comparison to PMT instruments, the full spectrum coverage and the adopted numerical methods generate very good detection limits that satisfy most of the analytical demands in the steel industry, and offer an equivalent or superior precision for the measurement of key alloying elements. Moreover, it works as a multi-matrix instrument, with unlimited applications, for production and d e v e l o p m e n t. Emile Muller is Technical Director OES, Software & Automation, Jean-Marc Böhlen is Manager OES Applications and Edmund Halász is Application Specialist OES at Thermo Electron in Ecublens, S w i t z e r l a n d. 344