Acoustical Testing 1

Similar documents
Non-Anechoic Testing Environment

Acoustical Testing I Loudspeaker Study

Reverberation time and structure loss factor

The Speaker Study. By: Jay Bliefnick. Acoustical Testing 1. Attn: Dr. Dominique Chéenne, Dr. Lauren Ronsse. Group Members:

Case study for voice amplification in a highly absorptive conference room using negative absorption tuning by the YAMAHA Active Field Control system

SIA Software Company, Inc.

TBM - Tone Burst Measurement (CEA 2010)

ODEON APPLICATION NOTE ISO Open plan offices Part 2 Measurements

ONLINE TUTORIALS. Log on using your username & password. (same as your ) Choose a category from menu. (ie: audio)

Please refer to the figure on the following page which shows the relationship between sound fields.

THE ANV MEASUREMENT SYSTEMS SOUND INSULATION TESTING SYSTEM INSTRUCTION MANUAL FOR FIELD TESTING OF WALLS, FLOORS & STAIRS

Supplementary User Manual for BSWA Impedance Tube Measurement Systems

Introduction to Equalization

CHAPTER 3 THE DESIGN OF TRANSMISSION LOSS SUITE AND EXPERIMENTAL DETAILS

Improving room acoustics at low frequencies with multiple loudspeakers and time based room correction

Monitor Setup Guide The right monitors. The correct setup. Proper sound.

Experimental Investigation on the Effect of Origami Geometry on the Acoustic Characteristics

Revision of ISO Standards on field sound insulation testing. Carl Hopkins

FREQUENCY RESPONSE AND LATENCY OF MEMS MICROPHONES: THEORY AND PRACTICE

Measuring procedures for the environmental parameters: Acoustic comfort

Airborne Sound Insulation

Small Room and Loudspeaker Interaction

Mei Wu Acoustics. By Mei Wu and James Black

Acoustics. Randy Zimmerman Chief Engineer

ROOM SHAPE AND SIZE ESTIMATION USING DIRECTIONAL IMPULSE RESPONSE MEASUREMENTS

Pre- and Post Ringing Of Impulse Response

In-Ear Monitors using VRT Transducers A market survey

ODEON APPLICATION NOTE Calculation of Speech Transmission Index in rooms

Orion E-STA Acoustic Test: Evaluating Predictions Against Data

ECMA-108. Measurement of Highfrequency. emitted by Information Technology and Telecommunications Equipment. 5 th Edition / December 2010

MEASURING SOUND INSULATION OF BUILDING FAÇADES: INTERFERENCE EFFECTS, AND REPRODUCIBILITY

Application Note 7. Digital Audio FIR Crossover. Highlights Importing Transducer Response Data FIR Window Functions FIR Approximation Methods

Room Acoustics. March 27th 2015

ALTERNATING CURRENT (AC)

Loudspeaker Array Case Study

A mobile reverberation cabin for acoustic measurements in an existing anechoic room

EQ s & Frequency Processing

SOUND FIELD MEASUREMENTS INSIDE A REVERBERANT ROOM BY MEANS OF A NEW 3D METHOD AND COMPARISON WITH FEM MODEL

Holographic Measurement of the Acoustical 3D Output by Near Field Scanning by Dave Logan, Wolfgang Klippel, Christian Bellmann, Daniel Knobloch

The vibration transmission loss at junctions including a column

ABSORPTION MEASUREMENTS FOR TWO TABLE SCREENS FROM GÖTESSONS INDUSTRI AB

Aalborg Universitet. Published in: Acustica United with Acta Acustica. Publication date: Document Version Early version, also known as pre-print

ECMA-108. Measurement of Highfrequency. emitted by Information Technology and Telecommunications Equipment. 4 th Edition / December 2008

UBL S119 LOUDSPEAKER SYSTEM

An Investigation on Factors That Cause Error in Reverberation Time Measurement (ISO 3382) in UTHM Lecturer Room

Direct Field Acoustic Test (DFAT)

RD75, RD50, RD40, RD28.1 Planar magnetic transducers with true line source characteristics

Acoustic Calibration Service in Automobile Field at NIM, China

Active Control of Energy Density in a Mock Cabin

Technical Data Measurement Microphones

Response spectrum Time history Power Spectral Density, PSD

What you Need: Exel Acoustic Set with XL2 Analyzer M4260 Measurement Microphone Minirator MR-PRO

(i) Sine sweep (ii) Sine beat (iii) Time history (iv) Continuous sine

Proceedings of Meetings on Acoustics

Transfer Function (TRF)

Low frequency sound reproduction in irregular rooms using CABS (Control Acoustic Bass System) Celestinos, Adrian; Nielsen, Sofus Birkedal

APPLICATION NOTE. Making Accurate Voltage Noise and Current Noise Measurements on Operational Amplifiers Down to 0.1Hz. Abstract

Understanding Sound System Design and Feedback Using (Ugh!) Math by Rick Frank

Microphone a transducer that converts one type of energy (sound waves) into another corresponding form of energy (electric signal).

A White Paper on Danley Sound Labs Tapped Horn and Synergy Horn Technologies

Measuring Instrument Combinations

Rub & Buzz Detection with Golden Unit AN 23

Production Noise Immunity

How To... Commission an Installed Sound Environment

12. PRELAB FOR INTERFERENCE LAB

ANGKASA Reverberation Acoustic Chamber Characterization

PART I: The questions in Part I refer to the aliasing portion of the procedure as outlined in the lab manual.

Validation of lateral fraction results in room acoustic measurements

Errata to Procedural Standard for Sound & Vibration Measurement 2015 Third Edition

THE SPEAKER. The decibel scale is related to the physical sound intensity measured in watts/cm 2 by the following equation:

8th AIAA/CEAS Aeroacoustics Conference June 16 18, 2002/Breckenridge, CO

New ON/OFF Absorption Technology That Includes Low Frequencies

Measurement of Weighted Harmonic Distortion HI-2

MICROPHONE TECHNIQUES

Holographic Measurement of the 3D Sound Field using Near-Field Scanning by Dave Logan, Wolfgang Klippel, Christian Bellmann, Daniel Knobloch

Lab 18 Delay Lines. m208w2014. Setup. Delay Lines

Measurement of weighted harmonic distortion HI-2

STEREO IMPACT Solar Energetic Particles Package (SEP) Dynamic Test Plan

EAW KF740 Technology Brief

3D Intermodulation Distortion Measurement AN 8

DESIGN OF VOICE ALARM SYSTEMS FOR TRAFFIC TUNNELS: OPTIMISATION OF SPEECH INTELLIGIBILITY

Measuring Center for Android User manual Software for analyzing sound signals

Informations and comments on ECE-TRANS-WP.29-GRB e

What applications is a cardioid subwoofer configuration appropriate for?

29th TONMEISTERTAGUNG VDT INTERNATIONAL CONVENTION, November 2016

INTERNATIONAL BACCALAUREATE PHYSICS EXTENDED ESSAY

sound reinforcement & microphones

DISTANCE CODING AND PERFORMANCE OF THE MARK 5 AND ST350 SOUNDFIELD MICROPHONES AND THEIR SUITABILITY FOR AMBISONIC REPRODUCTION

Problems with the INM: Part 2 Atmospheric Attenuation

SigCal32 User s Guide Version 3.0

Acoustical Improvements in Kulturhuset Oceanen Using Slitted Low Frequency Absorbers. Mattias Davidsson, Master's Thesis 2006:18

App Note Highlights Importing Transducer Response Data Generic Transfer Function Modeling Circuit Optimization Digital IIR Transform IIR Z Root Editor

Façade insulation at low frequencies influence of room acoustic properties

WITHIN GENERATOR APPLICATIONS

Envelopment and Small Room Acoustics

Musical Acoustics, C. Bertulani. Musical Acoustics. Lecture 14 Timbre / Tone quality II

Filters And Waveform Shaping

Statistical properties of urban noise results of a long term monitoring program

SPEAKER SC-COBRA-PEEK-WP

XLA 3200 Line Array Loudspeakers

Transcription:

Material Study By: IRINEO JAIMES TEAM ANDREW MILLER SAM SHROYER NATHAN NEGRU ERICH PFISTER Acoustical Testing 1 Dr. Lauren Ronsse, Dr. Dominique Chéenne 11/05/2014

Table of Contents Abstract. 3 Introduction....3 Test Method and Modifications.. 4 Measurement set up and Equipment......6 Room Modal Frequency Analysis..10 Results.. 11 Equations..13 Conclusion 14 Appendix...15

ABSTRACT The purpose of this experiment was to find absorption coefficient of Sonofiber Auralex panels. Parameters presented in The Standard Method for Sound Absorption and Sound Absorption Coefficients by the Reverberation Room Method (Designation: c423-00) paper were followed but limitations due to a pre-existing pillar and unavailable equipment required modifications to the parameters. INTRODUCTION The main focus of this experiment was to determine the absorption coefficient of Sonofiber Auralex absorption panels from 100Hz to 5000Hz using parameters set forth by The Standard Method for Sound Absorption and Sound Absorption Coefficients by the Reverberation Room Method (Designation: c420-00) paper. The team was able to calculate the absorption coefficients of the panels with some modifications to the parameters to best fit the property of the CCC chamber and to accommodate for a pre-existing pillar in the room. Columbia College Chicago s reverb chamber is located in the basement of 33 East Congress Parkway. The 72ft^2 (6.69m) area of the room dimensions, Sonofiber panels, and microphone positions are depicted in Figure 1.

TESTING METHOD AND MODIFICATIONS Figure 1: plan view of CCC reverb chamber Table 1 has been created in order to compare the standard vs. implemented parameters. In the right column (Test Standard) are the parameters, on the right column (CCC reverb Chamber) is the application of the standard and adjustments made. TEST STANDARDS 1) Volume - Room should be no less than 125 m^3 or 4414.33 ft^3 2) Diffuse Sound Field - Satisfactory approximation can be achieved with sound reflectors hung or distributed with random orientations about the volume of room. - Experience has shown that a satisfactory approximation can be achieved with a number of sound-reflective panels hung or distributed with random orientation. CCC REVERB CHAMBER Volume CCC reverb chamber is 157.69 m^3 or 5568.77 ft^3 Reflection panels were not available, concluded that the present diffuse sound field in the CCC reverb chamber was close to desired diffuse sound field. *Pre -existing pillar prevent the microphones from receiving equal amounts of energy from all directions by causing unwanted early reflections

3) Background Noise Level - Level shall be 15dB below lowest level used to calculate decay rate. 4) A-Mounting for Specimen - The specimen can be placed on floor but should avoid placing in center of the room or avoid specimen sides being parallel to any other walls. 5) Distance From Surfaces - No part of the specimen shall be closer than 2.5ft (.75m) to a reflective surface. 6) Mic Measurements - Fixed microphone needs 5 measurements at 35 different positions which are at least 5ft (1.524m) apart. 7) Mic Distance From Specimen - Fixed microphone shall be at least 1.5ft (.0457m) from surface of the test specimen. 8) Relative Humidity - Relative humidity in the room shall be greater than 40% during the test. 9) Room Construction - The room is best constructed of massive masonry or concrete materials, but other materials, such as well-damped steel, may be used. Lighter construction may be excessively absorptive at 200Hz. 10) Microphone - The microphone or microphones used to measure decay rate shall be omnidirectional with a flat ( + 1 db within any one-third octave band) 11) Measurement of sound absorption coefficient - The average absorption coefficient of the room surfaces at each frequency, determined by Absorption of the empty room area of the room surfaces Highest level was 115dB *Lowest background noise level was at 33dB at 100Hz. Figure 1, shows the placement of the specimen. The Sonofiber Auralex absorption panels were not placed in the center and avoided and avoided the specimen s sides from being parallel to any wall. The closest distance the specimen was to a reflective surface was approximately 4ft well above 2.5ft (.75m). Figure 1 shows microphone positions, and their relative position to one another. All microphones were positioned 5ft (1.524m) above specimen. Relative humidity was at 45% *Temperature was at 73% Previously the CCC reverb chamber was a bank vault and later converted. Walls made up of composite steel with an air gap behind it followed by a concrete wall. ECM800 ( specifications of microphone located in appendix) Absorption Absorption - Empty Room coefficient 100 12.24 0.059 125 12.8 0.061 160 9.45 0.045 200 10.15 0.048 250 8.26 0.039

Shall be less than or equal to.05 12) Test Signal - The test signal shall be band of random noise with continuous spectrum covering the range over which measurements are made. The frequency range of the measurements shall include the onethird octave bands with mid-band frequencies, as defined in ANSI S1.0 from 100Hz to 5000Hz 315 8.01 0.038 400 7.88 0.038 500 7.97 0.038 630 7.58 0.036 800 8.37 0.040 1000 8.51 0.041 1250 8.34 0.040 1600 7.5 0.036 2000 8 0.038 2500 9.02 0.043 3150 9.96 0.048 4000 11.79 0.056 5000 14.52 0.070 Using pink noise generated by SpectraPlus, measurements from 40Hz to 20000Hz were gathered. For test purposes, absorption coefficients from 100Hz to 5000Hz were the focus. Table 1: Conditions to qualify a room as a diffuse field compared with the conditions present/followed in the CCC reverb chamber tests Measurement set up and Equipment For test microphone an ECM8000 was used as stated in Table 1. A Motu Interface (828MK3) was used to provided a high quality signal. Figure 2: Motu interface converts analog to digital, providing good quality single

Four test speakers were used to produce sound in the reverberation chamber. The speakers where placed facing the corners of the reverberation chamber to create diffusion. Figure 3: Mackie speaker facing room corners The test specimen was Sonofiber Auralex Panels. The configuration of the specimen is shown in Figure 4. Figure 4: Sonofiber (72ft^2) panel, made up of 18 (2ft x 2ft) panels Aviom An-16/I were used to connect the inputs and outputs from the modeling lab to the reverberation chamber. - Figure 5: visual representation of Aviom An-16/i

SpectraPlus tone generator was used to excite the room with pink noise. The test microphone would then pick up the excitation then redirected back into SpectraPlus where its algorithms calculated decay time for specific frequencies. Figure 6: Computer connected through patchbay in modeling lab to patchbay in reverb chamber. Figure 7: Signal chain flow diagram of test

db Figure 7 is the signal chain diagram. A single tone was generated. SpectraPlus in this case generated pink noise. The tone went through signal processors, then played through the speakers in the reverberation chamber. The microphone in the reverberation chamber picked up the frequency energy content and then went back into signal processors and into SpectraPlus, where SpectraPlus calculated T-60 Values. Prior to testing, background noise levels was measured to satisfy condition 3 from Table 1. LEQ levels were measured in the CCC reverb chamber for 2 minutes. Longer LEQ could have been achieved, however, the CCC reverb chamber is located in the basement of Columbia College Chicago. The only outside interference that could have created drastic changes in background noise level would have been the CTA train, located in close proximity to the building. The low frequency boost would have occurred below 100Hz and would not be a factor in the final test result if the train happened to run by while testing. 40 35 30 25 20 15 10 5 SPL (db re 20uPA) 0 0 1000 2000 3000 4000 5000 6000 Freq (Hz) Figure 8: LEQ levels for the given frequencies [numerical values are located in appendix]

The highest LEQ value was measured at 100 Hz at approximately 33 db. With the standard only requiring the lowest level used to calculate decay rate to be 15 db below the highest value of LEQ, the test signals generated in the chamber was at 115 db. With decay rate levels reaching 95dB at the lowest, well above the noise floor present. Satisfying the background noise level condition, microphone positions were chosen based on condition 7 & 10 from Table 1. For each microphone position, (Mic-[B thru F], 5 measurements were calculated by SpectraPlus. T-60 values for both CCC reverb chamber empty and CCC reverb chamber with specimen+, with a total of fifty T-60 values. Taking the average of each microphone s T-60 values, decay rate, absorption of room, absorption of material, and absorption coefficient of the specimen can be derived. Calculations located in Table 2.1 in appendix. Refer to the section Equations in report to review individual formulas. ROOM MODEL FREQUENCY ANALYSIS To get a better understanding of possible sound pressure level differences, the lowest modal frequencies were calculated. Mode # (Length, Width, Height) Frequency 1, 0, 0 17.49 0, 1, 0 26.92 0, 0, 1 68.94 1, 1, 0 32.11 1, 0, 1 71.12 0, 1, 1 74.01 2, 0,0 34.99 2, 0, 1 77.31 2, 1, 0 44.15 2, 0, 2 142.25 Table 2: under the equations section, Equation 1 was used

Seconds (s) When audibly testing mode (0, 1, 0), by walking throughout the room, a sound pressure level different did occur mid-way in the width (x) dimension, as predicted by the calculation. However, throughout the room unpredictable sound pressure level drops were audibly heard. Reviewing T-60 values for the same frequency at different microphone positions yields decay rate discrepancies from one microphone to the next. RESULTS The unexpected sound pressure level drops can be better understood by graphing T-60 values vs frequencies Figure 12. 4.5 Mic-B Mic-C Mic-D Mic-E Mic-F 4 3.5 3 2.5 2 1.5 1 0.5 0 0 5000 10000 15000 20000 25000 Freq (Hz) Figure 9: T-60 over each frequency tested in the CCC reverb chamber The chart, in a perfectly diffuse sound field with no interference from axial modes would have all 5 of the microphone curves layered on top of one another and theoretically just be one plotted curve. Figure 9 shows just how much T-60 differed throughout the room. The outlined

section covering roughly 1 2 khz, has peaks that varied when they should have been similar. The differences in T-60 values at different parts of the room were taken into account when reviewing final results. 1.2 Abs. Coefficient 1 0.8 0.6 0.4 0.2 0 0 1000 2000 3000 4000 5000 6000 Figure 10: absorption coefficients of Sonofiber panels from 100Hz to 5000Hz, numerical values in Table 1.1 in Appendix Reviewing the Absorption Coefficients from Figure 10, the Sonofiber ability to reduce T-60 between 1 khz and 4 khz is high, with a roll-off happening out of this range. The expectation was below 500Hz the effectiveness of the material to reduce T-60 values would decrease due to low frequency properties. With these findings it was concluded the Sonofiber panels are great to place within an environment requiring T-60 values to be as low as possible around 1 khz to 4 KHz, in many cases would be a vocal performance space. Surprisingly, with varying T-60 values throughout the room the results did not produce skewed final absorption coefficient results.

EQUATIONS Equation 1 calculates axial modes, given the dimensions of the room. Equation 2 calculates the speed of sound when given temperature in Fahrenheit. Equation 3 calculates the decay rate, time it takes a signal to decay 60 db. Equation 4 calculates total absorption of the room when decay rate, speed of sound, and volume are known. Equation 5 calculates the total absorption due to the specimen. Equation 6 calculates the absorption coefficient of the specimen. Equation 1: Modal Frequencies = C ( l 2 Lx )2 + ( m Ly )2 + ( n Lz )2 l = Length integer w = width integer h = height integer [Hz] C= Speed of Sound Lx = length of room Ly = width of room Lz = height of room Equation 2: Speed of Sound = 49.022 459.67 + Temperature(Fahrenheit) [ft/s] Equation 3: Equation 4: Decay Rate = 60dB T60 [db/seconds] Total Absorption for both empty & with specimen = C = Speed of sound.921 (V)(Decay Rate) C V= Volume [Sabine] Equation 5: Specimen Total Absorption = [Total Abs. w/specimen - Total Abs. empty] [Sabine]

Equation 6: Absorption Coefficient of Specimen = Specimen Total Absorption Area of Specimen [Sabine/ft^2] *Area of specimen (Sonofiber Auralex panels) = 72 ft^2 or 6.69 m^2 *Volume of CCC reverb chamber = 157.69 m^3 or 5568.77 ft^3 *Temperature = 73 degrees Fahrenheit CONCLUSION The focus of the paper was to find the absorption coefficient of the Sonofiber Auralex panels by averaging calculated T-60 values throughout the reverb chamber. With a less than ideal diffuse sound field modifications to the The Standard Method for Sound Absorption and Sound Absorption Coefficients by the Reverberation Room Method (Designation: c423-00) still yielded reliable applicable absorption coefficients. Finally Sonofiber Auralex absorption panels are great for vocal performance spaces. For future studies an addition of reflection panels to the reverb chamber capable of eliminating axial modes would yield more accurate results. Expanding the test method to include a frequency sweep and an impulse response can offer more useful data to analyze. Gathering data about the modal properties of the CCC reverb chamber can prevent poor microphone placement.

Appendix * LEQ Values (numerical values for figure 8) Freq. (Hz) SPL (db re 20uPA) Freq. (Hz) SPL(dB re 20uPA) 40 36.25 1000 26.60 50 40.98 1250 26.31 63 41.22 1600 27.00 80 30.80 2000 26.43 100 32.94 2500 27.57 125 30.30 3150 28.27 160 33.40 4000 26.59 200 29.41 5000 26.28 250 29.60 6300 25.48 315 27.62 8000 24.99 400 29.57 10000 25.28 500 28.98 12500 25.17 630 28.85 16000 25.56 800 27.36 20000 30.05

Table 1.1 Table 2.1

Frequency response and directionality of microphone EMC-8000 Behringer