Minnesota Rules, Chapter 4732 X-ray Revision

Similar documents
Minnesota Rules, Chapter 4732 X-ray Revision

Minnesota Rules, Chapter 4732 X-ray Revision

Minnesota Rules, Chapter 4732 X-ray Revision

Minnesota Rules, Chapter 4732 X-ray Revision PROPOSED NON-MEDICAL ANALYTICAL X-RAY SYSTEMS, 1.0

Minnesota Rules, Chapter 4732 X-ray Revision PROPOSED NON-MEDICAL CABINET X-RAY SYSTEMS, 1.0

Minnesota Rules, Chapter 4732 X-ray Revision

TABLE OF CONTENTS. References

Minnesota Rules, Chapter 4732 X-ray Revision

RADIATION SAFETY REQUIREMENTS FOR ANALYTICAL X-RAY EQUIPMENT

(a) This subchapter applies to installations using analytical x-ray equipment and establishes requirements for their use.

APPENDIX K UNF RADIATION GENERATING DEVICE SAFETY PROCEDURES

PART H RADIATION SAFETY REQUIREMENTS FOR NON-HEALING ARTS RADIATION GENERATING DEVICES (RGD)

By: Georgia Institute of Technology

1. Oklahoma Administrative Code (OAC) 252:410, titled Radiation Management

Ch. 223 VETERINARY MEDICINE CHAPTER 223. VETERINARY MEDICINE GENERAL PROVISIONS X-RAYS RADIOACTIVE MATERIAL. Authority

NJDEP Medical Physicist s Radiographic QC Survey Registration Number:

RULES OF TENNESSEE DEPARTMENT OF ENVIRONMENT AND CONSERVATION DIVISION OF RADIOLOGICAL HEALTH CHAPTER USE OF X-RAY APPARATUS

SUBCHAPTER 14. THERAPEUTIC INSTALLATIONS

GE AMX 4+ Portable X-Ray

LASER RADIATION REQUIREMENTS

VACUUM INTERRUPTER APPLICATION NOTES Filename: VIAN X-Rays and Vacuum Interrupters Revision: 0 PAGE 1

MaxRay Handheld X-ray Systems Operator Training Exam

Agilent 5DX Series 2L, 3, and 5000 X-ray Safety Test Report

Agilent 5DX System. X-ray Safety Test Procedure. for use with Software Release 8.20 and later

QXR-W XRF SAFETY MANUAL

Keene State College Center for Environmental BioGeoChemistry Radiation Protection Program Revised January

05/20/14 1. Philips CM200T. Standby Condition

1-1. GENERAL 1-2. DISCOVERY OF X-RAYS

FEI Titan Image Corrected STEM

Test Equipment for Radiology and CT Quality Control Contents

Atomscope PORTABLE VETERINARY X-RAY EQUIPMENT INSTRUCTION MANUAL. Version 1.0

X-RAY SAFETY REFRESHER EXAM

SPECIFICATION. Kilovoltage X-ray calibration system for protection and diagnostic level dosimetry. Prepared by

Nuclear Associates , , CT Head and Body Dose Phantom

Model #2 UVB Midband Ultraviolet Lamp Assembly with Timer

Series 70 Servo NXT - Modulating Controller Installation, Operation and Maintenance Manual

DENTAL X-RAY OPERATOR'S INSTRUCTIONS. (for USA) Wall Mount Type...WK

ISO INTERNATIONAL STANDARD. Non-destructive testing of welds Radiographic testing of fusionwelded

RadView Manual Collimator

Joint ICTP/IAEA Advanced School on Dosimetry in Diagnostic Radiology and its Clinical Implementation May 2009

Safety Precaution Relating to Intense Radio Frequency Radiation

Nuclear Associates

Series 48 Air Cooled Laser & UC-2000 Quick Start Guide

OPERATOR'S INSTRUCTIONS

Radiological Safety Analysis Document for the CLAS12 Engineering and the first physics run of Run Group A

ANALYTICAL X-RAY EQUIPMENT RADIATION SAFETY MANUAL

Y11-DR Digital Radiography (DR) Image Quality

Series 48 Water Cooled Laser & UC-2000 Quick Start Guide

ISO INTERNATIONAL STANDARD. Non-destructive testing of welds Radiographic testing Part 1: X- and gamma-ray techniques with film

Engineering Policy & Procedure

CITY AND COUNTY OF DENVER POLICY DENVER FIRE DEPARTMENT. Emergency Responder Radio Enhancement Coverage System (RES)

Beaming Optical Length Tester (BOLT)

GT-1050A 2 GHz to 50 GHz Microwave Power Amplifier

REQUIREMENTS FOR USE OF NONMEDICAL X;RAY GENERATORS AT SLAC*

Angular Dependence Testing of the Innovative Industrial Solutions (IIS) WAM 500 D Area Monitor

IChapter Number,ChapterTiiie I0400: _OJ TJse of X-Ray Apparatus! Rule Number Rule Title

Firmware Version d & higher Installation & Operation

Safety Manual. Omni-ID UHF Tags

I-01 NAVIGATIONAL WARNING RECEIVERS

Guide to Obtaining DOT Approval of Jet Perforating Guns using AESC/IME Perforating Gun Specifications. Version 02 Effective Date: September 1, 2017

ELECTRONIC CONTROL CONCEPTS 160 Partition Street Saugerties, NY or local phone

RMV25 / RMV50 RMU25 / RMU45

9/17/90 (rev.) STATE OF MINNESOTA HINNESOTA DEPARTMENT OF HEALTH

IDEAL INDUSTRIES, INC. TECHNICAL MANUAL MODEL: MODEL: Multimeter Service Information

Product Guide Verizon Delaware LLC. Section 31 Delaware LLC Original Sheet 1. Connection With Certain Facilities of Others

X-RAYS - NO UNAUTHORISED ENTRY

Teaching Digital Radiography and Fluoroscopic Radiation Protection

Nemko-CCL, Inc West Alexander Street Salt Lake City, UT

RULES AND REGULATIONS. Title 58 - RECREATION PENNSYLVANIA GAMING CONTROL BOARD [58 PA. CODE CH. 525] Table Game Internal Controls

USER'S GUIDE. EcoQuest International 310 T. Elmer Cox Drive Greeneville, TN (423) Setup. Operation. Specifications.

Operating Manual. isolate501

I. PERFORMANCE OF X-RAY PRODUCTION COMPONENTS FLUOROSCOPIC ACCEPTANCE TESTING: TEST PROCEDURES & PERFORMANCE CRITERIA

External brake resistor AX2090-BW5x

White Space Devices (WSDs)

QL50 Luminescence Sensor

AC-TX Volt/100 Volt Transformer Panel for VR61 and VR62 Loudspeakers and the CLA37 Column Loudspeaker AC-TX128

Safety Information (Attach ed at the upper cover of t his product) Safety Warning (Attached on the output fiber isolato r & collimator) Figure 1 Laser

Ludlum Medical Physics

Application Note. ipix A Gamma imager to support various applications. Introduction. An easy to carry and deploy instrument

Signature 1100H+ Performance Verification Manual Version July 11, 2013

NDE SOLUTIONS RADIOGRAPHY COURSE OUTLINE

Notice of Rulemaking Hearing

PUBLICATION 213. Think Safety First

Basic Understanding of FCC 700 MHz Rules

SPRINGFIELD TECHNICAL COMMUNITY COLLEGE ACADEMIC AFFAIRS

AT&T INDIANA GUIDEBOOK. PART 2 - General Terms and Conditions 1st Revised Sheet 1 SECTION 9 - Connections

Rotating Anode X-Ray Tube Housing Assembly. General Data

Pipe Laser Model No Instruction Manual

Digital Imaging started in the 1972 with Digital subtraction angiography Clinical digital imaging was employed from the 1980 ~ 37 years ago Amount of

Air Monitoring Directive Chapter 9: Reporting

C-ARM FLUOROSCOPIC AND SPOT-FILM SYSTEMS

Firmware Version d & higher Installation & Operation

ProScale. Compact LCD Readout. Installation & Operation

ProScale Compact LCD Readout. Operation. Firmware version C 2.xxx and higher

STEREOTACTIC BREAST BIOPSY EQUIPMENT SURVEYS

Operating instructions Diffuse reflection sensor with background suppression O1D101 O1D / / 2018

Visibility of Detail

FM Safety Manual. Omni-ID UHF Tags

Honeywell, Automation and Control Solutions

NEW YORK CITY FIRE DEPARTMENT. Notice of Adoption of Final Rule. 3 RCNY , entitled In-Building Auxiliary Radio Communication Systems

Transcription:

Minnesota Rules, Chapter 4732 X-ray Revision DRAFT NON-MEDICAL HAND-HELD X-RAY SYSTEMS, 2.0 4732.#### NON-MEDICAL HAND-HELD X-RAY SYSTEMS. Summary of Changes MDH made a number of changes to the Non-Medical Hand-Held X-ray Systems v1.0 rule draft based on the industrial focus group s review and feedback at the February 9, 2018 meeting. The changes are described below. Subp. 3. Warning lights and devices Deleted item D The shutter open-closed status must be located near each collimator on the x-ray source housing if the useful beam is controlled with a shutter. Subp. 4. Beam ports. Deleted this subpart. Subp. 4. Labeling. [Renumbered] No changes. Subp. 5. Shutters. Deleted this subpart. Subp. 5. Safety device evaluation. [Renumbered] Revised item A. Subp. 6. Radiation emission limit. [Renumbered] No changes. Subp. 7. Area survey. [Renumbered] Form changes. Added new item C An x-ray fluorescence non-medical hand-held x-ray system is exempt from this subpart. Subp. 8. Safety procedures. [Renumbered] Item B. Revisions to subitem (3) in bold: responding to equipment malfunctions

Subp. 10. Utilization data [New] Added new subpart. Subp. 11. Storage and security; notification in event of theft or loss. [Renumbered] No changes. Subp. 12. Operator protection. [Renumbered] Added other than the operator to item D (formerly item E). Added new item F. Requirement to use a tube stand in certain situations. Revised item G (formerly item F) to include internal reference to items D and E. Subp. 13. Records. [Renumbered] 2

4732.#### NON-MEDICAL HAND-HELD X-RAY SYSTEMS. Subpart 1. Applicability. A registrant s non-medical hand-held x-ray system must comply with the requirements of this part. X-RAY SYSTEMS Subp. 2. Safety device. A registrant is responsible for the requirements of this subpart. The exposure switch of a non-medical hand-held x-ray system must: A. be able to terminate the exposure at any time during an exposure of greater than 0.5 seconds; and B. prevent an exposure when the timer is set to "zero" or "off" position if either Commented [BB(1]: Similar: NE, NC, ME, UT, SSRCR: Sec.H.8 (open beam lists as they should consider having safety device if RGD does not permit) Exempts Safety device: TX, RI, OH, Commented [BB(2]: Ohio 3701:1-68-03 Commented [JC(3]: Ohio 3701:1-68-03 Commented [JC(4]: Ohio 3701:1-68-03 position is provided. Subp. 3. Warning lights and devices. Commented [BB(5]: Similar: OH, RI, PA, UT A. A visible and discernible warning light, labeled with the x-ray words X-RAY ON, or words having similar meaning, must be: (1) located on or near the x-ray source and its controls; and (2) illuminated when the x-ray source is energized. B. The x-ray tube "on-off" status must be located near the radiation source housing. This requirement may be met if the warning lights are visible and discernible by anyone near the useful beam. Commented [TP(6]: Advisory Committee discussion: This wording is used in Industrial Radiography. Discussion on this: X-ray source vs. Switch (1) located near a switch that energizes an x-ray tube Commented [BB(7]: SSRCR H8.(b)(i) Similar: OH, TX, RI, ME, UT 3

C. Warning devices must be labeled so that the purpose is easily identified. For equipment that is installed after the effective date of this part, a warning device must have a fail-safe design. Subp. 4. Labeling. A registrant is responsible for labeling a non-medical hand-held x-ray system according to this subpart. A. A non-medical hand-held x-ray system must be labeled near any switch that Commented [JC(8]: Similar: OH, ME, UT energizes an x-ray tube with a visible and discernible sign bearing the radiation symbol and the words "CAUTION RADIATION - THIS EQUIPMENT PRODUCES IONIZING RADIATION WHEN ENERGIZED", or other words having similar meaning. B. A non-medical hand-held x-ray system must be labeled at or near the x-ray exit Commented [BB(9]: Similar: OH, ME, UT beam port to identify the location of the beam with the words CAUTION - HIGH INTENSITY X-RAY BEAM", or other words having similar meaning. Subp. 5. Safety device evaluation. A registrant is responsible for a safety device Commented [JC(10]: SSRCR; page H6, Section H.6 (j); evaluation of a non-medical hand-held x-ray system: A. upon arrival; and B. at intervals not to exceed 180 days. Commented [JC(11]: NC: 3 months C. For purposes of this subpart, a safety device evaluation includes: (1) safety device under subpart 2; (2) warning lights; (3) warning devices; and (4) shutters, if applicable. 4

D. A safety device evaluation must verify that: (1) a non-medical hand-held x-ray safety device is functioning as designed; and (2) all labels are visible and discernible. E. If a non-medical hand-held x-ray safety device is not functioning as designed, then it must be: (1) labeled immediately as defective; and (2) removed from service until the safety device is repaired. Commented [JC(12]: SSRCR; page H9, Section H.6 (j) F. A registrant must maintain a record of safety device evaluations for a nonmedical hand-held x-ray system. The record must include: (1) the dates of evaluations; (2) a list of the safety devices evaluated; (3) the results of the evaluation; (4) the name of the individual performing the evaluation; and (5) corrective actions recommended and performed for any safety device that fails the required evaluation. G. When a non-medical hand-held x-ray system is returned to service after being locked-out and tagged, it must be evaluated before use if the date of the last safety device evaluation exceeds the 180-day interval under item B. H. A non-medical hand-held x-ray system that is locked out and tagged DO NOT Commented [BB(13]: Similar: OH USE by the radiation safety officer is exempt from this subpart. Subp. 6. Radiation emission limit. A non-medical hand-held x-ray system must be Commented [BB(14]: Similar: OH, RI, ME, UT located and arranged to include sufficient shielding or access controls to prevent radiation 5

emission that may result in a dose to an individual in excess of the dose limits under part 4732.####. The dose limits must be met at the maximum operating parameters. AREA SURVEY REQUIREMENTS Subp. 7. Area survey. A. A registrant is responsible for an area survey of a non-medical hand-held x-ray system that complies with the radiation emission requirements under subpart 6. B. Qualified personnel must perform an area survey with radiation survey instruments calibrated according to part 4732.####. C. An x-ray fluorescence non-medical hand-held x-ray system is exempt from this subpart. CONDITIONS OF OPERATION Subp. 8. Safety Procedures. A registrant must develop and comply with operating and Commented [BB(15]: SSRCR; page H4, in definitions emergency procedures for a non-medical hand-held x-ray system. A. Operating and emergency procedures may be maintained in electronic or written form; B. Operating and emergency procedures must include: (1) storage and security of a non-medical hand-held x-ray system to prevent unauthorized use, removal, or accidental production of ionizing radiation when the non-medical hand-held x-ray system is not under the control and constant surveillance of an operator or the registrant; 6

(2) daily visual and operability checks of the non-medical hand-held x-ray system before use on each day to verify that a) the equipment is in good working condition; and b) required labeling under subpart 4 is present. c) If equipment problems are found during a daily check, then the equipment must be removed from service until it is repaired. (3) responding to equipment malfunctions; (4) minimizing exposure of individuals in the event of an accident; (5) notifying proper personnel in the event of an accident; and (6) locked out and tagged. C. No individual may operate a non-medical hand-held x-ray system in any manner Commented [BB(16]: SSRCR; page H8, section H.6. (h) other than that specified in the operating procedures unless the individual has obtained approval from the radiation safety officer (RSO). Approval may be authorized electronically or in written form. D. Operating and emergency procedures must be available to an operator of a Commented [BB(17]: SSRCR, Sec. I.10 (f), hand-held x-ray system. Subp.9. Temporary job site. A registrant must maintain copies of the following documents and records with a non-medical hand-held x-ray system: A. certificate of registration; and B. operating and emergency procedures. Subp. 10. Utilization data. A registrant must maintain utilization data for a non-medical hand-held x-ray system when used off-site. Utilization data include: Commented [BB(18]: Ohio 3701:1-68-03 Similar: NJ, NC 7

A. manufacturer, model number, and serial number; B. location and dates of use; and C. identity and authorization or signature of the operator. D. Utilization data for a non-medical hand-held x-ray system may be maintained in electronic or written form. Subp. 11. Storage and security; notification in event of theft or loss. A. A registrant must develop and implement written procedures for storage and security of a non-medical handheld x-ray system to prevent unauthorized use or removal when an x-ray system for non-medical handheld x-ray system is not under the control and constant surveillance of an operator or the registrant. B. A registrant must notify the commissioner of the theft or loss of a non-medical hand-held x-ray system according to part 4732.####. subpart. Subp. 12. Operator protection. A registrant is responsible for the requirements of this A. An individual or an operator must not hold a sample while the non-medical Commented [JC(19]: OH has a similar provision. hand-held x-ray system is operating; B. An operator must not aim the useful beam at any individual during operation of the non-medical hand-held x-ray system; C. The exposure switch of a non-medical hand-held x-ray system must be a "deadman" type; D. An individual, other than the operator, must maintain a distance of at least 6 8 feet (1.8 meters) from the tube housing assembly and the useful beam while

exposures are made using a non-medical hand-held x-ray system. If an individual must be within 6 feet of the tube housing assembly, then the individual must wear a 0.5 millimeter lead equivalent apron; and E. An operator is not required to wear a lead apron while using a non-medical hand-held x-ray system if it is equipped with a backscatter shield. F. If an operator is using a non-medical hand-held x-ray unit in a position that increases the operator s exposure to scatter radiation, then the operator must be protected from the scatter radiation by using a tube stand. G. An x-ray fluorescence non-medical hand-held x-ray system is exempt from items D and E. Subp. 13. Records. 9