WaveMaster IOL. Fast and accurate intraocular lens tester

Similar documents
WaveMaster IOL. Fast and Accurate Intraocular Lens Tester

OptiSpheric IOL. Integrated Optical Testing of Intraocular Lenses

Quality Testing of Intraocular Lenses. OptiSpheric IOL Family and WaveMaster IOL 2

Quality Testing of Intraocular Lenses. OptiSpheric IOL Family and WaveMaster IOL 2

OPTINO. SpotOptics VERSATILE WAVEFRONT SENSOR O P T I N O

SpotOptics. The software people for optics L E N T I N O LENTINO

Ron Liu OPTI521-Introductory Optomechanical Engineering December 7, 2009

OPAL. SpotOptics. AUTOMATED WAVEFRONT SENSOR Single and double pass O P A L

"SIMPLE MEASUREMENT, ADVANCED RESULTS"

Industrial quality control HASO for ensuring the quality of NIR optical components

Explanation of Aberration and Wavefront

3.0 Alignment Equipment and Diagnostic Tools:

Puntino. Shack-Hartmann wavefront sensor for optimizing telescopes. The software people for optics

OMI-SWIR. SpotOptics FAST & ACCURATE WAVEFRONT SENSOR S W I R

Reference and User Manual May, 2015 revision - 3

PROCEEDINGS OF SPIE. Automated asphere centration testing with AspheroCheck UP

Ocular Shack-Hartmann sensor resolution. Dan Neal Dan Topa James Copland

1.1 Singlet. Solution. a) Starting setup: The two radii and the image distance is chosen as variable.

SpotOptics. The software people for optics OPAL O P A L

Adaptive Optics for LIGO

Breadboard adaptive optical system based on 109-channel PDM: technical passport


Paper Synopsis. Xiaoyin Zhu Nov 5, 2012 OPTI 521

GEOMETRICAL OPTICS AND OPTICAL DESIGN

Exercise 1 - Lens bending

Customized Correction of Wavefront Aberrations in Abnormal Human Eyes by Using a Phase Plate and a Customized Contact Lens

Vision Research at. Validation of a Novel Hartmann-Moiré Wavefront Sensor with Large Dynamic Range. Wavefront Science Congress, Feb.

Lecture 2: Geometrical Optics. Geometrical Approximation. Lenses. Mirrors. Optical Systems. Images and Pupils. Aberrations.

Infra Red Interferometers

Aberrations and adaptive optics for biomedical microscopes

CHARA AO Calibration Process

Advanced Lens Design

ImageMaster. Worldwide Leader in MTF Testing

Lecture 2: Geometrical Optics. Geometrical Approximation. Lenses. Mirrors. Optical Systems. Images and Pupils. Aberrations.

Optical Design with Zemax

AST Lab exercise: aberrations

Lens Design I. Lecture 5: Advanced handling I Herbert Gross. Summer term

Tolerancing in Zemax. Lecture 4

Proposed Adaptive Optics system for Vainu Bappu Telescope

Optical System Design

Difrotec Product & Services. Ultra high accuracy interferometry & custom optical solutions

Optical design of a high resolution vision lens

ΘΘIntegrating closedloop adaptive optics into a femtosecond laser chain

Compact and Modular Interferometers

ECEN 4606, UNDERGRADUATE OPTICS LAB

10/25/2017. Financial Disclosures. Do your patients complain of? Are you frustrated by remake after remake? What is wavefront error (WFE)?

Sequential Ray Tracing. Lecture 2

PROCEEDINGS OF SPIE. Measurement of low-order aberrations with an autostigmatic microscope

OPTICAL IMAGING AND ABERRATIONS

Tutorial Zemax 8: Correction II

phone extn.3662, fax: , nitt.edu ABSTRACT

What is Wavefront Aberration? Custom Contact Lenses For Vision Improvement Are They Feasible In A Disposable World?

Design of a Test Bench for Intraocular Lens Optical Characterization

MASSACHUSETTS INSTITUTE OF TECHNOLOGY Mechanical Engineering Department. 2.71/2.710 Final Exam. May 21, Duration: 3 hours (9 am-12 noon)

Lecture 4: Geometrical Optics 2. Optical Systems. Images and Pupils. Rays. Wavefronts. Aberrations. Outline

Wavefront Sensing In Other Disciplines. 15 February 2003 Jerry Nelson, UCSC Wavefront Congress

Shaping light in microscopy:

Laboratory experiment aberrations

High contrast imaging lab

IAC-08-C1.8.5 OPTICAL BEAM CONTROL FOR IMAGING SPACECRAFT WITH LARGE APERTURES

Nature Methods: doi: /nmeth Supplementary Figure 1. Schematic of 2P-ISIM AO optical setup.

On the study of wavefront aberrations combining a point-diffraction interferometer and a Shack-Hartmann sensor

Beam expansion standard concepts re-interpreted

NIMO TR1504 CONTACT US. Lambda-X s.a. Av. Robert Schuman 102 B-1400 NIVELLES Belgium

Digital Wavefront Sensors Measure Aberrations in Eyes

INTRODUCTION TO ABERRATIONS IN OPTICAL IMAGING SYSTEMS

ADALAM Sensor based adaptive laser micromachining using ultrashort pulse lasers for zero-failure manufacturing

Exercises Advanced Optical Design Part 5 Solutions

OPTOCRAFT Unleashed wavefront technology

Hartmann Wavefront Analyzer

Metrology and Sensing

Geometrical Optics for AO Claire Max UC Santa Cruz CfAO 2009 Summer School

StockOptics. CATALOG 2018 Europe

ATS 200 & ATS 400 Alignment Turning Stations with Ful y Integrated Measurement Technology

Multi aperture coherent imaging IMAGE testbed

J. C. Wyant Fall, 2012 Optics Optical Testing and Testing Instrumentation

Lens Design I Seminar 5

DETERMINING CALIBRATION PARAMETERS FOR A HARTMANN- SHACK WAVEFRONT SENSOR

Using Stock Optics. ECE 5616 Curtis

Testing Aspherics Using Two-Wavelength Holography

Wavefront sensing by an aperiodic diffractive microlens array

Chapter 3. Introduction to Zemax. 3.1 Introduction. 3.2 Zemax

Shack-Hartmann wavefront sensor: technical passport

12.4 Alignment and Manufacturing Tolerances for Segmented Telescopes

Measurement of the Modulation Transfer Function (MTF) of a camera lens. Laboratoire d Enseignement Expérimental (LEnsE)

Be aware that there is no universal notation for the various quantities.

INTERFEROMETER VI-direct

Handbook of Optical Systems

Optical Engineering 421/521 Sample Questions for Midterm 1

Performance Factors. Technical Assistance. Fundamental Optics

AY122A - Adaptive Optics Lab

VATT Optical Performance During 98 Oct as Measured with an Interferometric Hartmann Wavefront Sensor

AgilEye Manual Version 2.0 February 28, 2007

Calibration of AO Systems

Cardinal Points of an Optical System--and Other Basic Facts

Vision. The eye. Image formation. Eye defects & corrective lenses. Visual acuity. Colour vision. Lecture 3.5

Tutorial Zemax 3 Aberrations

DIMENSIONAL MEASUREMENT OF MICRO LENS ARRAY WITH 3D PROFILOMETRY

The Aberration Structure of the Keratoconic Eye

Lens Design I. Lecture 3: Properties of optical systems II Herbert Gross. Summer term

Transcription:

WaveMaster IOL Fast and accurate intraocular lens tester

INTRAOCULAR LENS TESTER WaveMaster IOL Fast and accurate intraocular lens tester WaveMaster IOL is a new instrument providing real time analysis of monofocal and multifocal intraocular lenses (IOL) by measuring and analyzing the wavefront with high speed and accuracy. The ultra-accurate wavefront mapping enables analysis of most complicated lenses with spherical, aspheric or toric shape. Among many other parameters WaveMaster IOL provides results for Diopter power as well as high resolution power mapping of the complete lens aperture Lower and higher orders of lens aberrations Modulation Transfer Function (MTF) and Point Spread Function (PSF) Effective focal length (EFL) Various wavefront sensors with different spatial resolutions and available accessories allow for using WaveMaster IOL in different environments like production sites or research and development laboratories. In combination with a temperature stabilized Model Eye all types of IOL can be characterized in situ according to EN/ISO 11979. It is also possible to simply measure the lenses in air. The software provides various ways of displaying the results. It includes extended analysis features and data saving options. In addition theoretical data can be loaded and compared in real time with measurements. For ultra-fast measurement of IOL in high volume production, an automated tray system with multiple lens seats and attachable model eye is available. Setup WaveMaster IOL is based on a Shack- Hartmann wavefront sensor measurement system in infinite conjugate set up. Wavefront sensors measure aberrations which are introduced to a wavefront while passing through any kind of optical material, lens or objective. In the infinite conjugate set up a point light source is placed in the focal plane of the sample under test. A telescope images the exit pupil of the sample under test onto the wavefront sensor. At the same time the image is magnified in order to use the full area of the wavefront sensor to guaranty highest spatial resolution. Wavefront sensor CCD chip Microlens array Imaging telescope Sample under test Laser point light source Infinite conjugate set up of the wavefront measurement system 2

SETUP System components Wavefront sensor The standard design of a Shack-Hartmann sensor mainly consists of a CCD camera which is placed in the focal plane of a microlens array. An incoming wavefront is sampled by the lenses of the microlens array and the foci form a spot pattern on the camera which would be evenly spaced in case of a plane wavefront. Any aberration introduced by the sample lens leads to a curvature of the wavefront thus resulting in small local wavefront tilts. These induce a measurable shift of each focus spot position. An integration of the obtained slope information allows for reconstruction of the wavefront profile with high accuracy. Sample holder The sample holder can be adapted to the requirements of the individual IOL measurement task. In-air measurement Single IOL sample jigs which allow for a measurement of the lenses in air can be used with WaveMaster IOL. The lenses are placed in the jig which is then inserted into the holder base of the instrument. Because of the high precision holder base no extended sample alignment is necessary. Due to the easy handling, lenses can be exchanged very quickly resulting in a high throughput. Using state of the art computers this wavefront reconstruction can be done within the CCD camera frame rate i.e. within fractions of a second even if microlens arrays with a large number of lenses are used to obtain high spatial resolution. The WaveMaster IOL with sample holder for in-air measurements Measurement principle of the Shack-Hartmann wavefront sensor 3

SETUP In-situ measurement When in-situ measurements are required a model eye can be used. It fits into the high precision holder base of the instrument and is easily exchangeable with the single lens jigs. The Model Eye has been designed to simulate the effect of the real human eye during measurements of IOL and is defined in the ISO 11979 standard. It is composed of two parallel plates delimitating an area filled with saline solution. The IOL is placed at a specific position between these two plates on a sample holder including the aperture stop. An achro- 3mm Aperture 1 2 3 Saline Water n =1.336 4 5 T = 35 C 7 8 IOL under Test 9 Model Cornea (Achromatic Doublet) parallel plates (BK7) Image Plane Design of the Model Eye according to the ISO 11979 standard The WaveMaster IOL with Model Eye as sample holder for in-situ matic doublet simulates the eye cornea. The complete Model Eye is held at a constant temperature of 35 C. The aperture size can be changed due to measurement requirements. By default it is Ø = 3mm. The model cornea can be removed easily to allow for measurement of optical power of the IOL. High volume production testing In case of testing in a high volume production environment the system can be factory configured with a tray system which allows for fully automatic measurement of a multitude of IOL. This tray system which contains a fully motorized high precision XY linear stage can be proposed for in-air as well as in-situ measurements including the Model Eye. 4 WaveMaster IOL Model Eye with accessories

OPERATING PRINCIPLE During the measurement an automatic classification of all IOL in different quality classes is done according to user defined pass fail criteria. A classification map is displayed at the end of each tray measurement allowing for an easy sorting of the lenses. Light source The point light source consists of a stabilized monochromatic laser light source with an absolute wavefront error < λ/15. It is available with different wavelengths, numerical apertures and working distances. A kinematic mount allows for easy change. The point light source is placed on an automatic high precision linear focusing stage. Operating Principle Optimized for simplicity in use and high accuracy WaveMaster IOL provides a fast and accurate tool for testing intraocular lenses in a research and development as well as in a production environment. The wavefront is measured and analyzed with frame rates of up to 16 Hz and provides real time power map, -PSF and -MTF as well as analysis of the lower and higher order lens aberrations. In addition WaveMaster IOL also determines EFL and average Diopter power. Already during the measurement a real time comparison with design or reference data can be performed. Since simultaneous measurements of different aperture diameters are possible the IOL can be analyzed at e.g. Ø = 3mm and Ø= 6mm at the same time without changing the instrument setup. All measurements on IOL can be done in-air or in-situ. Hardware as well as software support an easy and fast change between the different measurement conditions. On the software side this is enabled by the possibility to save and load individual measurement settings files, on the hardware side by using the high precision sample holder. In addition alignment error compensation and automatic focusing allow for an easy and fast sample setup. When measuring large numbers of the same type of IOL, statistic functions in the software are available. Diopter and power map The Diopter power of an IOL is defined as the reciprocal of the reduced paraxial focal length in aqueous humor. From the measured effective focal length of the IOL, the power can be determined directly when the measurement is made in situ. When measuring in air the WaveMaster IOL software converts the Diopter power of the IOL for in situ conditions taking the design conversion factor of the lens into account. Measuring in air leads to easier handling of the lenses and fast throughput in production conditions for example when using a tray with several tens of IOL. In the power map the local effective focal length for each point of the aperture of the IOL is displayed. It is directly deduced from the measured wavefront. By measuring the wavefront defocus coefficients, EFL and Diopter power of the IOL can be easily determined 5

OPERATING PRINCIPLE Software module for measuring MTF and PSF PSF und MTF The Modulation Transfer Function (MTF) is a parameter describing objectively the performance of optical imaging systems by testing the ability of an optical system to transfer the details of an object to the image in terms of contrast. It has the value 1 for a perfect contrast reproduction and the value 0 for a system being unable to produce any image contrast. Lens aberrations The measured wavefront can be decomposed into a linear combination of polynomials of the Zernike- or the Seidel series which describe typical optical properties and errors of a lens as e.g. coma and astigmatism or spherical aberrations. This allows for real time measurement and comparison of lower and higher order lens aberrations of the IOL. PSF as well as MTF can be calculated directly from the measured wavefront and are displayed in real time. 6

SOFTWARE Software module IOL measurement Software The advanced software is designed to work under Windows operating systems and fulfils the need of the optical test for easy, intuitive operation. It offers a high level of speed and accuracy and provides consistent, reliable results. All aspects of the data acquisition starting with the powerful autofocus system, data calculation, calibration and the display of the data are under software control and fully automated. The software package provides menu-driven operator guidance and advanced data management. First 6 orders of the Zernike coefficients 7

TECHNICAL DATA Wavefront acquisition and display Absolute or relative measurement Subtraction of background illumination Real time display of 2D- and 3D-wavefront Peak-to-Valley and Root-Meansquare values Intensity Slope data Raw camera image Real time correction of sample misalignment (tilt and defocus) Available units: µm or λ Frame rate up to 16 Hz Intraocular lens measurement Diopter power and power map Real time calculation and 2D- and 3D display Customized zonal display of powermap EFL measurement and display MTF and PSF Real time calculation and 2D- and 3D display Real time numerical and graphical (2D and 3D) display Lower and higher lens aberration (Zernike and Seidel) Real time aberration analysis Real time numerical and graphical (2D and 3D) display of aberration coefficients Data saving and documentation Measurement certificate showing Graphical and numerical display of all measurement results Measurement conditions and sample related information Results can be saved in various formats to allow for further analysis with external software Detailed measurement settings can be saved into separate files and reloaded Technical data Aperture diameter: 0.5 to 15 x 15 mm Number of lenslets: up to 140 x 140 Power Range: -100 to +100 D Power absolute accuracy: 0.1 to 0.3 % Power resolution: 0.01 D Power map lateral resolution: 20 µm for a 3 mm aperture MTF absolute accuracy: 2 % MTF Wavefront measurement absolute (relative) accuracy rms: <λ/20 (λ /50) Measurement time for 1 IOL including Power, MTF and aberrations: 5 sec Wavelength: 546 nm ±10 Light Source numerical aperture: up to 0.95 Focus position accuracy: 1.5 µm Pass/Fail classification according to user defined criteria Real time comparison of measurement results with design data TRIOPTICS GmbH. Optische Instrumente Hafenstr. 35-39. D-22880 Wedel / Germany Phone: +49-4103 - 18006-0. Fax: +49-4103 - 18006-20 E-mail: info@trioptics.com http://www.trioptics.com 2010 TRIOPTICS GmbH. All rights reserved