PoS(PhotoDet 2012)058

Similar documents
Characterization of a prototype matrix of Silicon PhotoMultipliers (SiPM s)

Nuclear Instruments and Methods in Physics Research A

Characteristics of a prototype matrix of Silicon PhotoMultipliers (SiPM)

Tutors Dominik Dannheim, Thibault Frisson (CERN, Geneva, Switzerland)

A Measurement of the Photon Detection Efficiency of Silicon Photomultipliers

Characterisation of SiPM Index :

PoS(PhotoDet 2012)061

Silicon Photomultiplier

Direct Measurement of Optical Cross-talk in Silicon Photomultipliers Using Light Emission Microscopy

Characterizing a single photon detector

PoS(PhotoDet 2012)051

Silicon Carbide Solid-State Photomultiplier for UV Light Detection

Low Dark Count UV-SiPM: Development and Performance Measurements P. Bérard, M. Couture, P. Deschamps, F. Laforce H. Dautet and A.

Direct Measurement of Optical Cross-talk in Silicon Photomultipliers Using Light Emission Microscopy

Silicon Photo Multiplier SiPM. Lecture 13

PoS(PhotoDet 2012)022

Andrea WILMS GSI, Helmholtzzentrum für Schwerionenforschung, Darmstadt, Germany

Silicon Photodiodes - SXUV Series with Platinum Silicide Front Entrance Windows

InGaAs SPAD BIOMEDICAL APPLICATION INDUSTRIAL APPLICATION ASTRONOMY APPLICATION QUANTUM APPLICATION

Recent Development and Study of Silicon Solid State Photomultiplier (MRS Avalanche Photodetector)

NON-AMPLIFIED PHOTODETECTOR USER S GUIDE

NON-AMPLIFIED HIGH SPEED PHOTODETECTOR USER S GUIDE

Department of Electrical Engineering and Computer Science, Massachusetts Institute of Technology, 77. Table of Contents 1

INGAAS FAST PIN (RF) AMPLIFIED PHOTODETECTORS

SiPMs for solar neutrino detector? J. Kaspar, 6/10/14

InGaAs SPAD freerunning

Silicon Photomultiplier Evaluation Kit. Quick Start Guide. Eval Kit SiPM. KETEK GmbH. Hofer Str Munich Germany.

High collection efficiency MCPs for photon counting detectors

Détecteurs SiPM. Nicoleta Dinu. Laboratory of Linear Accelerator, Orsay

PRELIMINARY RESULTS OF PLASTIC SCINTILLATORS DETECTOR READOUT WITH SILICON PHOTOMULTIPLIERS FOR COSMIC RAYS STUDIES *

Detectors for microscopy - CCDs, APDs and PMTs. Antonia Göhler. Nov 2014

Measure the roll-off frequency of an acousto-optic modulator

Scintillator/WLS Fiber Readout with Geiger-mode APD Arrays

Total Absorption Dual Readout Calorimetry R&D

HIGH SPEED FIBER PHOTODETECTOR USER S GUIDE

J-Series High PDE and Timing Resolution, TSV Package

SPMMicro. SPMMicro. Low Cost High Gain APD. Low Cost High Gain APD. Page 1

event physics experiments

PMT tests at UMD. Vlasios Vasileiou Version st May 2006

PRELIMINARY. Specifications are at array temperature of -30 C and package ambient temperature of 23 C All values are typical

Study of Silicon Photomultipliers for Positron Emission Tomography (PET) Application

Recent Developments in Ultra-High Speed and Large Area Photomultiplier Tubes

Quantum Efficiency Measurement System with Internal Quantum Efficiency Upgrade

PCS-150 / PCI-200 High Speed Boxcar Modules

arxiv: v1 [astro-ph.im] 19 Nov 2014

Silicon Avalanche Photodiode SAR-/SARP-Series

CHAPTER 9 POSITION SENSITIVE PHOTOMULTIPLIER TUBES

Redefining Measurement ID101 OEM Visible Photon Counter

PoS(PhotoDet 2012)016

RAPSODI RAdiation Protection with Silicon Optoelectronic Devices and Instruments

Red, Green, Blue (RGB) SiPMs

InGaAs Avalanche Photodiode. IAG-Series

Timing and cross-talk properties of BURLE multi-channel MCP PMTs

Near Ultraviolet (NUV) SiPMs

Performance of the MCP-PMTs of the TOP counter in the first beam operation of the Belle II experiment

SILICON PHOTOMULTIPLIERS: FROM 0 TO IN 1 NANOSECOND. Giovanni Ludovico Montagnani polimi.it

In the name of God, the most merciful Electromagnetic Radiation Measurement

A new single channel readout for a hadronic calorimeter for ILC

SensL B-Series Silicon Photomultipliers for TOF- PET. NDIP2014 Kevin O Neill 4 th July, 2014

AND9770/D. Introduction to the Silicon Photomultiplier (SiPM) APPLICATION NOTE

An Introduction to the Silicon Photomultiplier

SiPM Module PRELIMINARY

Photon Count. for Brainies.

HF Upgrade Studies: Characterization of Photo-Multiplier Tubes

Week 9: Chap.13 Other Semiconductor Material

A Study of Silicon Photomultiplier Sensor Prototypes for Readout of a Scintillating Fiber / Lead Sheet Barrel Calorimeter

ECEN 4606, UNDERGRADUATE OPTICS LAB

A-CUBE-Series High Sensitivity APD Detector Modules

NMI3 Meeting JRA8 MUON-S WP1: Fast Timing Detectors High Magnetic Field µsr Spectrometer Project at PSI Status Report

arxiv: v1 [physics.ins-det] 16 Jun 2010

Ground-based optical auroral measurements

An electrical photon source

High granularity scintillating fiber trackers based on Silicon Photomultiplier

Page 1. Ground-based optical auroral measurements. Background. CCD All-sky Camera with filterwheel. Image intensifier

Introduction to silicon photomultipliers (SiPMs) White paper

Timing Resolution Performance Comparison for Fast and Standard Outputs of SensL SiPM

The Benefits of Photon Counting... Page -1- Pitfalls... Page -2- APD detectors... Page -2- Hybrid detectors... Page -4- Pitfall table...

Review of Solidstate Photomultiplier. Developments by CPTA & Photonique SA

TCSPC at Wavelengths from 900 nm to 1700 nm

arxiv: v2 [physics.ins-det] 10 Jan 2014

Development of the first prototypes of Silicon PhotoMultiplier (SiPM) at ITC-irst

Engineering Medical Optics BME136/251 Winter 2018

Real-time Laser Beam Position Detector. XY-4QD User Manual

Exp 3 COLCULATE THE RESPONSE TIME FOR THE SILICON DETECTOR

Homework Set 3.5 Sensitive optoelectronic detectors: seeing single photons

Amplified Photodetectors

arxiv: v2 [physics.ins-det] 14 Jan 2009

Characterization of SiPMs for Large Scale Applications

Non-amplified Photodetectors

Application Notes: Discrete Amplification Photon Detector 5x5 Array Including Pre- Amplifiers Board

Instruction manual and data sheet ipca h

Highlights of Poster Session I: SiPMs

CHAPTER 11 HPD (Hybrid Photo-Detector)

Single-Photon Avalanche Diodes (SPAD) in CMOS 0.35 µm technology

Overview Full Featured Silicon Photomultiplier Module for OEM and Research Applications The is a solid state alternative to the Photomultiplier Tube (

OFCS OPTICAL DETECTORS 11/9/2014 LECTURES 1

KLauS4: A Multi-Channel SiPM Charge Readout ASIC in 0.18 µm UMC CMOS Technology

SUPPLEMENTARY INFORMATION

SINPHOS SINGLE PHOTON SPECTROMETER FOR BIOMEDICAL APPLICATION

PMF the front end electronic for the ALFA detector

Transcription:

Absolute Photo Detection Efficiency measurement of Silicon PhotoMultipliers Vincent CHAUMAT 1, Cyril Bazin, Nicoleta Dinu, Véronique PUILL 1, Jean-François Vagnucci Laboratoire de l accélérateur Linéaire, Orsay, France E-mail: chaumat@lal.in3p3.fr The absolute measurement of the Photo Detection Efficiency (PDE) of a silicon photomultiplier over the whole visible range using continuous light is not a straightforward task as it requires the accurate determination of the detector gain. We have developed at LAL a procedure to achieve an absolute PDE measurement with an accuracy of ± 6 %. 1.Introduction Silicon photomultipliers (SiPM) are recent detectors studied by many groups. Their proprieties (compact form, insensitivity to magnetic fields, low voltage power supplies...) are promising for various scientific fields, from high energy physics calorimetry to medical imaging application. An accurate measurement of the SiPM parameters is needed to integrate these devices in experimentation. Absolute Photo Detection Efficiency over the visible range is a decisive parameter for most applications. To measure the absolute SiPM PDE, we have carried out two methods at LAL: the Counting method using short laser pulses at fixed wavelengths and the Current method, using continuous monochromated light for a continous wavelength scan from 400 to 700nm. After describing the experimental set-up and explaining in detail the two methods, we will show the absolute PDE measurement results for three SiPMs with different geometry. International Workshop on New Photon-detectors, LAL Orsay, France June 13-15, 2012 1 Speaker: V.Chaumat, V.Puill Copyright owned by the author(s) under the terms of the Creative Commons Attribution-NonCommercial-ShareAlike Licence. http://pos.sissa.it

2. Experimental 2.1 Description of the tested devices and experimental set-up The SiPMs studied are MPPCs (S10362-11) from HAMAMATSU Photonics with 1 mm² of active area. Measurements were performed in order to determine the working range of each SiPM: its breakdown voltage (V BD ), single pixel avalanche gain, dark count rate (DCR) and recovery time were measured. For more details about the employed set-up and the theory regarding measurements, refer to [1]. Breakdown voltage at 20 C for MPPCs of 25 µm, 50 µm and 100 µm of pixel size were measured at 69.9 V, 68.3 V and 68.7 V respectively. Fig.1 presents the experimental set-up used for the PDE measurements. It is composed of a dark box of 1 m 3 that assures a high temperature stability (20 0.1 C). The test bench was equipped with several light sources: a continuous source: a halogen lamp (100 W) followed by an ORIEL grating monochromator (350-800nm) with a wavelength accuracy of 2 nm. 3 pulsed Pilas laser diodes (405, 467 and 635 nm) driven at a repetition rate of 500 khz with 1 Chaumat,Puill.spectral width below 3 nm and pulse timing width below 50 ps. The incident flux is measured with a reference (see 2.2) for continuous and pulsed light configurations. The light intensity is set by neutral density filters inside a motorized wheel. The readout electronics consist of a 500 MHz MITEQ voltage amplifier (gain = 350, 50). A SMA cable sends the signal to a Wavepro 750ZI LeCroy digital oscilloscope (40 GSamples/s, 4GHz of analog bandwidth). Polarization and current measurement of the SiPM are done by the Keithley source meter (2612). 2.2 Test-bench calibration Fig 1: experimental setup for the SiPM PDE measurement We monitored the temperature with six Pt100 probes installed inside the dark box and readout by a Keithley acquisition unit and a thermo regulator (water chiller). A 20 0.1 C regulation is achieved. The light sources are placed at a stabilized temperature of 22 C 1 C. Three reference photodetectors have been used to calibrate the light incident flux: 2 PIN Photodiodes (HAMAMATSU-S3590-18 and a Gamma Scientific-UDT 221) and a photomultiplier () (HAMAMATSU-R7400U-01). They are fixed on a rigid support on 3 axes translation stages, allowing movement in front of the light spot with an accuracy of 0.1 mm. The relative position between all the detectors is checked with a focusing CCD camera with an accuracy of 0.2 mm. The beam light is spreaded with lenses to achieve a 10 mm Copyright owned by the author(s) under the terms of the Creative Commons Attribution-NonCommercial-ShareAlike Licence. http://pos.sissa.it

diameter surface whose homegeneity has been controlled with the (with a 1mm² pin hole) to 2%.These precisions allow placement errors to be negleted. Although the was supplied with a quantum efficiency curve, we had to correct these data to obtain a good agreement (5%) with the 2 PIN photodiodes. Once this correction has been made, the is used as a unique reference detector. The incident flux on the detector is defined as: I qe Flux incident( ph / mm² / s) (Eq1) h c Rs G S where q e is the electron charge (C), I the photo generated current (A), Rs the anode sensibility (A/W), G its gain and S its illuminated surface (mm²). The main errors on the incident flux determination come from the errors on the illuminated surface (3% over 38.3mm²) and on the product of the gain by the sensitivity of the (5%). After a one hour warming, the current is measured with a Keithley source meter (2612) with around 1% accuracy (few na over hundreds na). The incident flux on the SiPM is then known with an accuracy of 6 %. 2.3 PDE measurement methods The SiPM PDE [Eq2] is defined as the ratio of the number of converted photons (N photoe- ) to the incident number of photons (N photon ), it takes into account the quantum efficiency and the fill factor of the SiPM: Flux incident is evaluated by the calibrated whereas N photoe- is evaluated either by the Counting Method, or by the Current Method. 2.3.1 The Counting Method N PDE(%) 100 Flux The pulse coming from the Pilas laser diodes is short enough (50 ps) to be considered as a Dirac pulse. For low Flux incident, below 5 photons per pulse, the obtained photoelectron distribution follows the Poisson probability (Fig. 2). Therefore the number of photoelectrons produced in the SiPM is defined as: photoe N photoe ln((p0)) where P(0) is the probability of non converted events that are counted over 40000 events. Incident (Eq3) (Eq2) Fig 2 : oscilloscope view of the SiPM answer to short light pulses with few photons The major error of the PDE measurement with this method comes from the measurement of the incident number of photons which had been estimated at 6%. Copyright owned by the author(s) under the terms of the Creative Commons Attribution-NonCommercial-ShareAlike Licence. http://pos.sissa.it

Absolute PDE measurement Vincent Chaumat 2.3.2 The Current Method The number of photons coming from the monochromator that had been converted into photoelectrons by the SiPM (N photoe- ) is given by: I pg N 100 (Eq4) photoe q G where I pg is the photogenerated current (A) and G SiPM the SiPM gain measured in dark conditions. The G SiPM value directly affects N photoe- and therefore the PDE. The determination of the gain of a SiPM has to take into account all the secondary effects (cross-talk, after-pulses) that affect it. To perform this measurement, we consider the SiPM signals in the dark (Fig 3): the oscilloscope is triggered on the rising edge of a primary avalanche signal. The signal is integrated (charge C) over a time window whose width represents around 5 times the recovery time of the device to take account of all its secondary effects. The pedestal charge (charge P), corresponding to the non-correlated signals, is determined by integrating the charge of these signals over a time window of the same length. Fig 3: SiPM dark signal and its correlated secondary effects integrated over the green window and the non correlated pedestal signals integrated over the maroon window The SiPM gain is then given by: G SiPM where Z oscillo is the impedance of the oscilloscope (50 ). The gain of the MPPCs measured as a function of the bias voltage is shown in Fig. 4. e SiPM ( C P) q Z e Oscillo (Eq5) Fig 4: gain of the S10362-11-25, 50 and 100 measured in dark at 20 C as a function of the bias voltage 4

Absolute PDE measurement Vincent Chaumat 3 MPPC PDE measurements results 3.1 Comparison of the Counting Method and the Current Method results Figure 5 shows the ratio of the counting PDE to the current PDE at 3 wavelengths (405, 467 and 635 nm) as a function of the bias voltage. We observe that the PDE measured by counting is larger than the one determined by current measurement for the MPPC with 25 and 50 µm of pixel size whereas this is the contrary for the 100 µm. For the 3 devices, the ratio increases with the bias voltage and reaches around 1.1 at high overvoltages. For a same MPPC, the ratio does not depend on the wavelength. The difference between these two results can come from a bad determination either of the incident flux on the SiPM or of the number of converted photons. With the rigorous optical calibration of the test bench, the incident flux is known with 6 % accuracy. The most likely hypothesis is an incorrect SiPM gain measurement that distorted the number of converted photons determined with the current method. To check this hypothesis, we determine the PDE using the current method with the laser and the monochromator. With the ratio of two measurements, shown in Figure 6, the gain of the MPPC is no longer a variable of the equation. We observe that the PDE ratio is close to 1 and constant with the bias voltage. Therefore, we can conclude that the inaccurate measurement of the SiPM gain explains what we observed on Figure 5. Fig 5: counting PDE to current PDE ratio of the 3 MPPC S10362-11 as a function of the bias voltage and for 3 wavelength Fig 6: ratio of the SiPM PDE measured when illuminated by the laser over the one when the light comes from the monochromator To determine the absolute PDE of a SiPM over the whole visible range of wavelengths with a high precision, the measurements with the current method have to be corrected. There are two ways to make this correction: either we normalize the current PDE curve with the counting PDE measurements at 3 wavelengths [3], or another method to determine the gain of the SiPM with the desire accuracy is developed. The latter method is explained afterwards. 3.2 The SiPM gain measurement using the Counting Method As the determination of the gain of a SiPM is difficult to do very accurately with the method explained in 2.3, we calculate it from the PDE results obtained with the counting method. The resultant gain is given by: ( Ilight I dark) Gaincalculated ( Eq6) Flux PDE q incident 5 count e

Absolute PDE measurement Vincent Chaumat where I light is the SiPM photo generated current (A), I dark the SiPM dark current (A), Flux incident the number of photons at the SiPM surface (ph/mm²/s) and PDE count the PDE measured with the counting method. Fig 7 shows that gain measured from the counting method is independent from the wavelength. 3.3 The absolute SiPM PDE over all the visible light range To determine the absolute PDE of the MPPC S10362-11 at all the wavelengths from 400 to 700 nm, we have performed the measurements with the current method using the gain calculated from measurements of PDE at 3 wavelengths with the counting method. The results for the 3 devices are shown on Fig 8. The agreement between the two PDE values is within 5 %. Fig 7: S10362-11-25, 50 and 100 gain measurement from pulsed light at 20 C 4 Conclusion and further work Fig 8: S10362-11-25, 50 and 100 absolute PDE as a function of the wavelength, at 20 C After a long calibration process, we manage to obtain a 6% absolute PDE accuracy measurement over a wide spectral range from 400 to 700nm and at various polarization voltages. We develop a method which fits equally well for continuous light or pulsed light. We highlight the difficulty to measure the real SiPM gain with the dark signal. Despite our efforts, we did not achieve a direct gain measurement precise enough, only 10 to 20% accuracy, to realize a user friendly PDE measurement process. Owing to a very precise bench calibration (6%), we extract the SiPM gain from a calculation based on counting PDE measurement. Detailed analysis of secondary effects (after-pulses, crosstalk) and dark noise distribution is needed (ongoing study in collaboration with Fermilab) to perform a better gain measurement from the dark signal. References [1] Dinu, N. et al, Electro-optical characterization of SiPM: a comparative study,nim A610, 423-426 (2009). [2] P. Eckert et al,,characterisation studies of silicon photomultiplier, NIM. A620,217-226 (2010). [3] Vacheret.A et al, Characterisation and simulationof the response of Multi Pixel Photon counters to lox light level. arxiv:1101.1996v1 [physics in-det] (2011) 6