MEASURING THE PERFORMANCE OF Sub-2 µm PARTICLE COLUMNS: IMPACT OF EXTRA COLUMN BAND SPREADING AND LC OPERATING PRESSURE

Similar documents
TechnicalNOTE EVALUATION OF LCT PREMIER NEGATIVE ION MODE ESI EXACT MASS MEASUREMENT PERFORMANCE. Introduction. Objective

Automated Frequency Response Measurement with AFG31000, MDO3000 and TekBench Instrument Control Software APPLICATION NOTE

Optimizing Instruments for Modern HPLC Columns

P7500 Series Probes Tip Selection, Rework and Soldering Guide

Low Capacitance Probes Minimize Impact on Circuit Operation

Verifying Power Supply Sequencing with an 8-Channel Oscilloscope APPLICATION NOTE

High-voltage Differential Probes

Simplifying DC-DC Converter Characterization using a 2600B System SourceMeter SMU Instrument and MSO/DPO5000 or DPO7000 Series Scope APPLICATION NOTE

Process Control Calibration Made Easy with Agilent U1401A

KickStart Instrument Control Software Datasheet

io n Data Sheet or The P5205 is a 100 MHz Active Differential Probe capable of measuring fast rise times of signals in floating circuits. This 1,300 V

Frame through-beam sensors

Isolation Addresses Common Sources of Differential Measurement Error

Measuring Vgs on Wide Bandgap Semiconductors APPLICATION NOTE

CISCO ONS /100-GHZ INTERLEAVER/DE-INTERLEAVER FOR THE CISCO ONS MULTISERVICE TRANSPORT PLATFORM

SOURCE MEASURE UNITS. Make Multiple Measurements Accurately Using a Single Instrument All While Saving Space, Time and Money

Power Analysis Application Module DPO4PWR MDO3PWR Datasheet

Measurement Statistics, Histograms and Trend Plot Analysis Modes

Encapsulated Transformers 115V + 115V Primary, Low Profile

PatternPro Error Detector PED3200 and PED4000 Series Datasheet

TDA7231A 1.6W AUDIO AMPLIFIER OPERATING VOLTAGE 1.8 TO 15 V LOW QUIESCENT CURRENT HIGH POWER CAPABILITY LOW CROSSOVER DISTORTION SOFT CLIPPING

Soldering a P7500 to a Nexus DDR Component Interposer

WOODWORKING TECHNOLOGY IN EUROPE: HIGHLIGHTS European Federation of Woodworking Technology Manufacturers

Simplifying FET Testing with 2600B System SourceMeter SMU Instruments APPLICATION NOTE

In-circuit Measurements of Inductors and Transformers in Switch Mode Power Supplies APPLICATION NOTE

Stress Calibration for Jitter >1UI A Practical Method

Keysight Technologies E1834E/G/J/M/Z Mounted Beam Delivery Optics. Preliminary Data Sheet

Using the Model 4225-RPM Remote Amplifier/ Switch to Automate Switching Between DC I-V, C-V, and Pulsed I-V Measurements APPLICATION NOTE

Through-beam ring sensors

A RELIABLE FOUNDATION FOR IC ANALYSIS THERMO SCIENTIFIC DIONEX AQUION IC SYSTEM

Ultraminiature Rocker&Lever Switches V ac or dc

Who Reads and Who Follows? What analytics tell us about the audience of academic blogging Chris Prosser Politics in

2018/2019 HCT Transition Period OFFICIAL COMPETITION RULES

Passive Voltage Probes

Passive High Voltage Probes P5100 P5102 P5120 P6015A

Debugging SENT Automotive Buses with an Oscilloscope APPLICATION NOTE

SUPPORT OF NETWORK FORMATS BY TRIMBLE GPSNET NETWORK RTK SOLUTION

Fundamentals of AC Power Measurements

Process Control HPP-25

30 A AC/DC Current Probe TCP0030A Datasheet

Power Measurement and Analysis Software

Keysight Technologies Using Non-Contact AFM to Image Liquid Topographies. Application Note

AC Current Probes CT1 CT2 CT6 Data Sheet

DPO7OE1 33 GHz Optical Probe

S540 Power Semiconductor Test System Datasheet

MBR10xCT Series TO-220AB

PRODUCT DATA. Sound Intensity Calibrator Type 4297

Using the Ramp Rate Method for Making Quasistatic C-V Measurements with the 4200A-SCS Parameter Analyzer APPLICATION NOTE

12.5 Gb/s PatternPro Programmable Pattern Generator PPG1251 Series Datasheet

Keysight Technologies Enhance EMC Testing with Digital IF. Application Note

Keysight TC GHz Frequency Doubler

EM Insights Series. Episode #1: QFN Package. Agilent EEsof EDA September 2008

Keysight Technologies Making Current-Voltage Measurement Using SMU

Keysight Technologies Improving Test Efficiency of MEMS Electrostatic Actuators Using the E4980A Precision LCR Meter.

S540 Power Semiconductor Test System Datasheet

20X Low Capacitance Probe P6158 Datasheet

High-voltage Differential Probes TMDP THDP THDP P5200A - P5202A - P5205A - P5210A

10 GHz Linear Amplifier PSPL5866 Datasheet

12.5 Gb/s Driver Amplifier LABware Module PSPL8001 Datasheet

Passive High Voltage Probes P5100A-TPP0850-P5122-P5150-P6015A Datasheet

Tire Pressure Monitoring Systems and Remote/Passive Keyless Entry

HRS2 Series PCB Relays

Getting Started with the LabVIEW DSP Module

Measuring Power Supply Switching Loss with an Oscilloscope

Keysight Technologies 7 Hints That Every Engineer Should Know When Making Power Measurements with Oscilloscopes. Application Note

Catalogue NE 0506 FIXTURING SYSTEM. Modular Zero-Point Fixturing for Pallets, Workpieces and Fixtures

Solutions for Solar Cell and Module Testing

Don t Let EMI/EMC Compliance Certification Slow You Down TUTORIAL

Keysight Technologies Why Magnification is Irrelevant in Modern Scanning Electron Microscopes. Application Note

Cisco ONS Metropolitan Dense Wavelength Division Multiplexing 100-GHz FlexLayer Filter Solution

Remote participation in Question sessions Audio options VoIP

AC/DC Current Probe TCP0150 Datasheet

Performing Safe Operating Area Analysis on MOSFETs and Other Switching Devices with an Oscilloscope APPLICATION NOTE

Keysight Technologies Improving the Test Efficiency of MEMS Capacitive Sensors Using the E4980A Precision LCR Meter.

Table of Contents Executive Summary 29

Characterize Phase-Locked Loop Systems Using Real Time Oscilloscopes

How big is China s Digital Economy

Welcome to the IFR Press Conference 30 August 2012, Taipei

PRODUCT DATA. PULSE Multichannel Sound Power Type 7748 Version 1.2

Trouble-shooting Radio Links in Unlicensed Frequency Bands TUTORIAL

12.5 Gb/s PatternPro Programmable Pattern Generator PPG1251 Series Datasheet

Japan s Leading Exhibition for Robotics Technologies Jan. 17[Wed]-19[Fri], 2018 Tokyo Big Sight, Japan

RECOVERED PAPER DATA

MEMS On-wafer Evaluation in Mass Production Testing At the Earliest Stage is the Key to Lowering Costs

7 Hints That Every Engineer Should Know When Making Power Measurements with Oscilloscopes.

Introduction. Part 1. Introduction...2

BenchTop Extraction Arms with unbeatable flexibility

Keysight Technologies Migrating Balanced Measurements from the

Comparing Contact Performance on PCBA using Conventional Testpads and Bead Probes

Agilent N8480 Series Thermocouple Power Sensors. Technical Overview

Mr. Wolfgang Metzen was trained as a mechanical

TDA0161. Proximity Detectors. Features. Description. Block Diagram. 10mA Output Current Oscillator Frequency 10MHz Supply Voltage +4 to +35V

Solar Array Simulation System Integration

OECD s Innovation Strategy: Key Findings and Policy Messages

Measuring CNT FETs and CNT SETs Using the Agilent B1500A

TOOLS & TECHNOLOGY. # techguide

Time and Frequency Measurements for Oscillator Manufacturers

Keysight Technologies 8490G Coaxial Attenuators. Technical Overview

Radiofrequency Ablation Needle Electrodes. Procedural Algorithms and Lesion Geometry

Tektronix Logic Analyzer Probes P6800/P6900 Series Datasheet

Transcription:

MEASURING THE PERFORMANCE OF Sub-2 µm PARTICLE COLUMNS: IMPACT OF EXTRA COLUMN BAND SPREADING AND LC OPERATING PRESSURE Kenneth J. Fountain, Eric S. Grumbach, Uwe D. Neue, Diane M. Diehl introduction Since their commercial introduction in the mid-199s, sub-2 μm particle columns have grown in popularity due to their potential for more chromatographic separating power and higher throughput. However, simply using these small particle columns in conjunction with conventional chromatographic systems does not produce the efficiencies and resolving power predicted by theory. Operating sub-2 μm particle columns at their optimal linear velocities generates pressures that are not achievable with conventional HPLC instrumentation. Therefore, sub-optimal flow rates are used, which deteriorates separations due to the influence of analyte diffusion. Furthermore, sub-2 μm particle columns are more susceptible to extra-column effects and band spreading. Recently published reports, presentations, and marketing literature have claimed that HPLC columns containing particles in the 2-3 μm range can be used with existing HPLC instrumentation to achieve the speed and efficiencies seen with UPLC technology [1-3]. These claims are made based on data comparing their performance to that of 1.7 μm particle columns on conventional HPLC systems. The disadvantages to this approach are that these systems have significant dwell volumes and band spreading, and they cannot reach the pressures required to operate 1.7 μm particle columns at their optimum linear velocity. In this work, the effects of extra-column band spreading and LC system operating pressure were investigated for sub-2 μm particle columns. The system contribution to band spreading and pressure limitations of conventional HPLC instrumentation were found to be the main reasons for the sub-optimal performance of 1.7 μm particle columns. Conventional instrumentation was found to be incapable of operating at the pressures required to use 1.7 μm particle columns at their optimal flow rates, resulting in a 55% reduction in the measured column efficiency (plate count). Experimental conditions Click on Part Numbers for more information LC Conditions UPLC System: ACQUITY UPLC System with an ACQUITY UPLC PDA detector HPLC System: Alliance 2695 Separations Module with a 2998 PDA detector Columns: ACQUITY UPLC BEH C 18, 2.1 x 5 mm, 1.7 μm XBridge C 18, 2.1 x 5 mm, 2.5 μm Part Number: 186385 Column Temp: 3 C Flow Rate:.5 1. ml/min Mobile Phase: 65/35 ACN/H 2 O (isocratic) Detection: 254 nm Sampling Rate: 4 Hz Time Constant: no filter Sample:.1 mg/ml thiourea and.2 mg/ml acenaphthene in 65/35 ACN/H 2 O Injection Volume: 1 μl (PLNO mode; 2 μl sample loop) The system volume and extra-column band spreading values were measured for each LC instrument used in this study. The measured system volume of the ACQUITY UPLC system is 15 μl, and the measured system volume of the Alliance 2695 is 876 μl. Extra-column band spreading measured at 5σ is 14 μl for the ACQUITY UPLC system and 36 μl for the Alliance 2695 system.

RESULTS AND DISCUSSION All HPLC columns have a theoretical limit for performance. In other words, there is a maximum efficiency/resolving power that each column can achieve. This limit is typically determined by the column dimensions (diameter, length), particle size, and packing process. However, experimental determination of a column s resolving power by HPLC can impact the efficiency measurement as well. Extra-column band spreading (also called band broadening or dispersion) is the unwanted widening of a chromatographic peak due to the HPLC system, which results in observable sub-optimal performance of an HPLC column. There are two sources of extracolumn band spreading. The first one is volumetric in nature and derives from the injection volume, the detector volume, and the volume of the tubing connecting the injector to the detector. The second contribution stems from time-related events such as the sampling rate or the detector time constant. This technical note will focus on the volumetric sources of extra-column band spreading. The impact of fixed extra-column band broadening on the observed column efficiency becomes more important for columns with a smaller diameter. For a column with a diameter of 4.6 mm, the negative effect of a particular band broadening value may be small. However, for a column with a diameter of 2.1 mm (~ 5-fold reduction in column volume from 4.6 mm i.d.), this same band broadening value will have a much larger effect [4]. Figure 1 shows the impact of column diameter on the observed efficiency. As column diameter decreases, the observed column efficiency decreases as well. 1 9 8 For an analyte with a retention factor of 5, the observed efficiency on a 2.1 mm i.d. column is half that of a 4.6 mm i.d. column. This effect is even more pronounced for analytes that are less retained. It is clear from Figure 1 that as the column diameter decreases, the influence of extra-column band spreading becomes larger, negatively impacting the observed column efficiency. However, column diameter is not the only parameter affecting column efficiency. Smaller particle columns are also more susceptible to band spreading than columns containing larger particles. There are two reasons for this. The first one is that the true performance of columns packed with smaller particles is higher than that of columns packed with larger particles. Thus, the peaks are narrower, and a given extra-column band broadening value will have a larger impact on the narrower peaks. Therefore, the performance is deteriorated more for columns packed with smaller particles. Second, shorter columns packed with smaller particles can be used to achieve the same column efficiency as longer columns packed with larger particles. A shorter column has a smaller column volume, and at equal efficiency, the impact of extra-column band spreading increases for the shorter column. Furthermore, the optimum linear velocity increases with decreasing particle size. Sub-2 μm particle columns generally cannot be operated at their optimum linear velocity on conventional HPLC instruments. As a compromise, sub-optimal flow rates are used, resulting in column efficiencies that are much lower than expected. The solution to achieving maximum efficiency on smaller particle (sub-2 μm) columns is to use a properly designed LC system that has minimal extra-column band spreading and the ability to operate at the pressures necessary for these columns (ACQUITY UPLC system) 7 Efficiency (N) 6 5 4 3 2 1 2.1 mm i.d. 3. mm i.d. 4.6 mm i.d. 1 2 3 4 5 6 7 8 9 1 Retention Factor (k ) Figure 1. Calculated effect of column diameter on observed efficiency using a conventional HPLC system. Calculations were made assuming a plate count of 1, on 5 mm length columns.

Figure 2 shows the difference in calculated efficiency for 1.7, 2.5, and 5 μm particle columns run on a conventional HPLC system and the ACQUITY UPLC system. Efficiency (N) 14 12 1 8 6 4 ACQUITY UPLC - 55 % System Conventional HPLC System - 44 % - 3 % HETP (µm) 6 5 4 1.7 µm (conventional HPLC system) 2.5 µm (conventional HPLC system) 3 2 1 2 4 6 8 1 12 14 16 18 Linear Velocity (mm/s) A 2 1.7 2.5 5. Particle Size (µm) Figure 2. Calculated maximum achievable efficiency of 1.7, 2.5, and 5 μm particle columns (2.1 x 5 mm) on a conventional HPLC system and an ACQUITY UPLC system. Plate counts shown are for an analyte with a k of 5. The number shown in red is the percentage of efficiency lost when using the same column on a conventional HPLC system. As particle size decreases, the difference between the efficiency measured on both LC systems increases. The result is more than a 5% decrease in efficiency for a 1.7 μm particle column used on a conventional HPLC system when compared to the ACQUITY UPLC system. It is also interesting to note that the 2.5 and 5 μm particle columns also benefit (although to a lesser extent) from being run on the ACQUITY UPLC system, indicating that this system can be used not only for UPLC applications, but can improve existing HPLC applications as well. A practical example of the theoretical concepts depicted in Figures 1 and 2 is shown by comparing the experimentally measured van Deemter curves for acenaphthene on 1.7 and 2.5 μm particle columns using UPLC and conventional HPLC systems (Fig. 3). HETP (µm) 2 18 16 14 12 1 8 6 4 2 2 4 6 8 1 12 14 16 18 Linear Velocity (mm/s) 1.7 µm (UPLC system) 2.5 µm (UPLC system) Figure 3. van Deemter curves for 1.7 μm and 2.5 μm particle columns on (A) a conventional HPLC system, and (B) an ACQUITY UPLC system. Acenaphthene was used as the test probe (k ~ 3.6). Efficiency was calculated using peak width measured at 4σ. There is no observable benefit of using the 1.7 μm particle column on a conventional HPLC system (Fig. 3A). It is only when both columns are analyzed using an ACQUITY UPLC system that there is a clear difference in performance. The maximum efficiency that was achieved for each column on the different LC systems is shown in Figure 4. The 2.5 μm particle column showed a 42% loss in plate count when used on the conventional HPLC system, and the 1.7 μm particle column showed a 57% loss in efficiency. This correlates well with the calculated trends shown in Figure 2. B

Plate Count (N) 1 9 8 7 6 5 4 3 2 1-57 % Particle Size (µm) ACQUITY UPLC System Conventional HPLC System - 42 % 1.7 2.5 Figure 4. Comparison of the maximum measured column efficiency as a function of particle size and LC system. Acenaphthene was the test probe. Plate counts were calculated using the peak width at 4σ. There are two reasons for this dramatic improvement in efficiency on the 1.7 μm particle column. First, the system contribution to band spreading is about 2.5 fold larger on the conventional HPLC system than on the ACQUITY UPLC system. This causes peaks to be much broader and can cause severe tailing (Fig. 5). AU AU.12.1.8.6.4.2.12.1.8.6.4.2 V.2.4.6.8 1 1.2 1.4 1.6 1.8 2 V.2.4.6.8 1 1.2 1.4 1.6 1.8 2 Minutes Figure 5. Isocratic separation of acenaphthene on an ACQUITY UPLC system and a conventional HPLC system. The column is an ACQUITY UPLC BEH C 18, 2.1 x 5 mm, 1.7 μm. Flow rate is.4 ml/min. acenaphthene acenaphthene ACQUITY UPLC N = 8,543 USP Tailing =.95 Conventional HPLC N = 3,49 USP Tailing = 1.41 Second, the optimum linear velocity for sub-2 μm particle columns usually generates pressures that are above the limit of conventional instrumentation. Therefore, one must operate these columns at sub-optimal flow rates in order to remain within the pressure constraints of conventional HPLC systems. This can be seen in Figure 3, where the optimum linear velocity for the 1.7 μm particle column is 3 mm/s (.2 ml/min) on the HPLC system and 1 mm/s (.6 ml/min) on the UPLC system. While it is possible for some HPLC systems to achieve these higher flow rates on sub-2 μm particle columns for a limited set of conditions, the performance of these systems still suffer from large extra-column band broadening. Since the ACQUITY UPLC system can routinely operate at pressures up to 15, psi (~1, bar) and has low system band spreading, sub-2 μm particle columns can be operated at their optimum linear velocity to yield highly efficient separations in a much shorter time. CONCLUSIONS The system contribution to band spreading and pressure limitations of conventional HPLC were found to be the main reasons for the misconception that 2-3 μm particle columns perform equally well or better than sub-2 μm particle columns. The performance of sub-2 µm particle columns cannot be accurately determined using conventional HPLC instrumentation due to pressure restrictions and extra-column effects. This is especially true for columns with diameters less than 4.6 mm. The efficiency of a 1.7 μm, 2.1 x 5 mm column was 13% greater when used on the ACQUITY UPLC system rather than a conventional HPLC system. Reducing the column packing particle size is limited by the pressure capability of the instrument, and no benefit is observed when using sub-2 μm particle columns on a conventional HPLC instrument. This study provides clear evidence that sub-2 μm columns must be used with a high pressure instrument that has minimal extra-column band spreading in order to achieve the efficiency and resolution that theory predicts for these particles.

REFERENCES 1. J.J. DeStefano, T.J. Langlois, J.J. Kirkland, J. Chrom. Sci. 46 (28) 254. 2. K. Butchart, T. Potter, A. Wright, A. Brien, International Labmate (August 27) 6. 3. J.A. Anspach, P.C. Rahn. Increasing HPLC Speed and Efficiency Without the Need for Ultra-High Pressures: New Columns Compatible with Existing and UHPLC Systems. Poster presented at HPLC 26, June 17-22, San Francisco, CA. 4. U.D. Neue. HPLC Columns: Theory, Technology, and Practice New York: Wiley-VCH, 1997. Sales Offices: Austria and European Export (Central South Eastern Europe, CIS and Middle East) 431 877 18 7 Australia 2 9933 1777 Belgium 32 2 726 1 Brazil 55 11 594 3788 Canada 8 252 4752 China 8621 6495 6999 CIS/Russia +7 495 3367 Czech Republic 42 2 617 11384 Denmark 45 46 59 88 Finland +358 9 5659 6288 France (33) 1 3 48 72 Germany 49 6196 46 Hong Kong 852 29 64 18 Hungary 36 1 35 586 India and India Subcontinent 91 8 2 837 19 Ireland 353 1 448 15 Italy 2 265 983 Japan (81) 3 3471 7191 Korea (82) 2 82 27 Mexico 5255 52 186 The Netherlands +31 ()76-5 87 2 Norway 47 63 84 6 5 Poland (48) 22 833 44 Puerto Rico 787 747 8445 Singapore 65 6273 1221 Spain 34 93 6 93 Sweden 46 8 555 115 Switzerland 41 56 676 7 Taiwan 886 2 2543 1898 United Kingdom 44 28 238 61 Waters, The Science of What s Possible, UPLC, ACQUITY UPLC, Alliance and XBridge are trademarks of Waters Corporation. All other trademarks are the property of their respective owners 28 Waters Corporation. Printed/Produced in the U.S.A. November 28 722793EN VW-PDF Waters Corporation 34 Maple Street Milford, MA 1757 U.S.A. T: 1 58 478 2 F: 1 58 872 199 www.waters.com