FLUORESCENCE MICROSCOPY. Matyas Molnar and Dirk Pacholsky

Similar documents
FLUORESCENCE MICROSCOPY

Boulevard du Temple Daguerrotype (Paris,1838) a busy street? Nyquist sampling for movement

Imaging Introduction. September 24, 2010

BioVis Core Facility. Fluorescence. Quantitative Microscopy Course CBA Dirk Pacholsky

Education in Microscopy and Digital Imaging

Microscopy: Fundamental Principles and Practical Approaches

microscopy A great online resource Molecular Expressions, a Microscope Primer Partha Roy

Figure 3.4 Approximate size of various types of cells. ~10 um. Red Blood Cells = mm 1500 um. Width of penny Pearson Education, Inc.

Point Spread Function. Confocal Laser Scanning Microscopy. Confocal Aperture. Optical aberrations. Alternative Scanning Microscopy

BASICS IN BIOIMAGING AND OPTICS PLATFORM EPFL SV PTBIOP LIGHT MICROSCOPY

Light Microscopy. Upon completion of this lecture, the student should be able to:

Microscopy Training & Overview

Lecture 23 MNS 102: Techniques for Materials and Nano Sciences

CCAM Microscope Objectives

VISUAL PHYSICS ONLINE DEPTH STUDY: ELECTRON MICROSCOPES

Tissue Preparation ORGANISM IMAGE TISSUE PREPARATION. 1) Fixation: halts cell metabolism, preserves cell/tissue structure

Observing Microorganisms through a Microscope

Microscopy. Matti Hotokka Department of Physical Chemistry Åbo Akademi University

Very short introduction to light microscopy and digital imaging

Chapter 2 The Study of Microbial Structure: Microscopy and Specimen Preparation

Microscope anatomy, image formation and resolution

MULTIPHOTON MICROSCOPY. Matyas Molnar Dirk Pacholsky

ANSWER KEY Lab 2 (IGB): Bright Field and Fluorescence Optical Microscopy and Sectioning

Observing Microorganisms through a Microscope LIGHT MICROSCOPY: This type of microscope uses visible light to observe specimens. Compound Light Micros

Nikon. King s College London. Imaging Centre. N-SIM guide NIKON IMAGING KING S COLLEGE LONDON

Microscopy Techniques that make it easy to see things this small.

Why and How? Daniel Gitler Dept. of Physiology Ben-Gurion University of the Negev. Microscopy course, Michmoret Dec 2005

1 Co Localization and Working flow with the lsm700

Introduction to Light Microscopy. (Image: T. Wittman, Scripps)

Invitation for a walk through microscopy. Sebastian Schuchmann Jörg Rösner


BASICS IN LIGHT MICROSCOPY

Biology 29 Cell Structure and Function Spring, 2009 Springer LABORATORY 1: THE LIGHT MICROSCOPE

Bi/BE 227 Winter Assignment #3. Adding the third dimension: 3D Confocal Imaging

MULTIPHOTON MICROSCOPY

Final Exam, 150 points PMB 185: Techniques in Light Microscopy

Bio 407. Applied microscopy. Introduction into light microscopy. José María Mateos. Center for Microscopy and Image Analysis

Compound Light Microscopy. Microscopy. Things to remember... 1/22/2017. This is what we use in the laboratory

3D light microscopy techniques

Reflection! Reflection and Virtual Image!

INTRODUCTION TO OPTICAL MICROSCOPY

Motorized Axio Observer Start-up instructions

Applications of Optics

Supplemental Figure 1: Histogram of 63x Objective Lens z axis Calculated Resolutions. Results from the MetroloJ z axis fits for 5 beads from each

Basics of confocal imaging (part I)

2/4/15. Brightfield Microscopy! It s all about Magnification..! or is it?!

Rates of excitation, emission, ISC

Things to check before start-up.

Introduction to light microscopy

Practical work no. 3: Confocal Live Cell Microscopy

Introduction to light microscopy

Katarina Logg, Kristofer Bodvard, Mikael Käll. Dept. of Applied Physics. 12 September Optical Microscopy. Supervisor s signature:...

Resolution. Diffraction from apertures limits resolution. Rayleigh criterion θ Rayleigh = 1.22 λ/d 1 peak at 2 nd minimum. θ f D

Chapter 3 Op,cal Instrumenta,on

3. are adherent cells (ie. cells in suspension are too far away from the coverslip)

Chapter 3 Op+cal Instrumenta+on

SETTING UP OF A TOTAL INTERNAL REFLECTION FLUORESCENT MICROSCOPE (TIRFM) SYSTEM: A DETAILED OVERVIEW

Confocal and 2-photon Imaging. October 15, 2010

INTRODUCTION THIN LENSES. Introduction. given by the paraxial refraction equation derived last lecture: Thin lenses (19.1) = 1. Double-lens systems

Supplementary Information for. Surface Waves. Angelo Angelini, Elsie Barakat, Peter Munzert, Luca Boarino, Natascia De Leo,

LSM 710 Confocal Microscope Standard Operation Protocol

CCAM s Selection of. Zeiss Microscope Objectives

MICROSCOPE LAB. Resolving Power How well specimen detail is preserved during the magnifying process.

Microscopy, Staining, and Classification

BASICS OF CONFOCAL IMAGING (PART I)

START-UP PROCEDURE 1 THE MICROSCOPE STAND 3 OBJECTIVES 5 STARTING WITH LAS (SOFTWARE) AND SETTING UP THE MICROSCOPE STAND 7

Light microscopy BMB 173, Lecture 14, Feb. 21, 2018

PHYSICS. Chapter 35 Lecture FOR SCIENTISTS AND ENGINEERS A STRATEGIC APPROACH 4/E RANDALL D. KNIGHT

Systems Biology. Optical Train, Köhler Illumination

Quick Guide for Zeiss 710 Laser Scanning Confocal MGH Cancer Center

Burton's Microbiology for the Health Sciences

High resolution extended depth of field microscopy using wavefront coding

BIOIMAGING AND OPTICS PLATFORM EPFL SV PTBIOP BASICS IN LIGHT MICROSCOPY

a) How big will that physical image of the cells be your camera sensor?

TRAINING MANUAL. Multiphoton Microscopy LSM 510 META-NLO

Microscopic Structures

Lecture 16. OMX - Structured Illumination Microscopy Ian Dobbie x Microscopy Course Lecture 16 1

LSM 510 META in Chang Gung University

Leica SP8 TCS Users Manual

Nature Methods: doi: /nmeth Supplementary Figure 1. Schematic of 2P-ISIM AO optical setup.

Nikon Instruments Europe

Olympus Fluoview 1000S Spectral Confocal Microscope Introduction to the NRI-MCDB Microscopy Facility Spectral Confocal Microscope

Chapter 17: Wave Optics. What is Light? The Models of Light 1/11/13

EE119 Introduction to Optical Engineering Spring 2003 Final Exam. Name:

Low Voltage Electron Microscope

Leica SP8 TCS Users Manual

The Nature of Light. Light and Energy

Nature Protocols: doi: /nprot Supplementary Figure 1. Schematic diagram of Kőhler illumination.

Leica TCS SP8 Quick Start Guide

Lecture M2 - Bespoke Microscopes. Ian Dobbie

Instruction Manual T Binocular Acromat Research Scope T Trinocular Acromat Research Scope

Examination, TEN1, in courses SK2500/SK2501, Physics of Biomedical Microscopy,

LSM 780 Confocal Microscope Standard Operation Protocol

Zeiss 880 Training Notes Zen 2.3

Leica TCS SP8 Quick Start Guide

A BRIEF INTRODUCTION TO MICROSCOPY The two key properties of a microscope that allow you to see microbes are resolution and magnification.

Zeiss LSM 510 Confocor III Training Notes. Center for Cell Analysis & Modeling

Digital Camera Technologies for Scientific Bio-Imaging. Part 2: Sampling and Signal

2017 MICROSCOPE REVIEW by Karen L. Lancour RELATIVE SIZE OF MICROBES

Development of a High-speed Super-resolution Confocal Scanner

Transcription:

FLUORESCENCE MICROSCOPY Matyas Molnar and Dirk Pacholsky 1

The human eye perceives app. 400-700 nm; best at around 500 nm (green) Has a general resolution down to150-300 μm (human hair: 40-250 μm) We need a tool to see smaller things or more of the spectral range Microscope (Objective/Filter) Camera/Film 2

WHY FLUORESCENCE MICROSCOPY? 3

Microscopical techniques Brightfield - low contrast for thin or transparent specimen - staining to enhance contrast needed (histochemical staining) Phase contrast - contrast via optical element (Phase ring) - intracellular structures can be seen - good for cell culture applications - negative: halos around cell bodies - combining with other techniques is generally poor (e.g. overlay with fluorescent image) 4

Microscopical techniques DIC/Nomarski - contrast via polarized light & optical element (Wollaston prism) - gives a (fake) topographical view - excellent for combination with fluorescence and histochemical staining Fluorescence - contrast via fluorescent staining - You see only what you stain (...in the perfect world) - special optical elements are needed (filter cubes) high resolution, high contrast, good for quantification (area + intensity) - Staining is sensitive - it can fade. 5

THE MICROSCOPE 6

Optical pathway of a microscope Objective projects image of specimen via Tube lens to Primary image plane Eyepiece magnifies this image. * Objectives are not interchangeable 7

Optical pathway of a microscope 8

The objective or lens The heart of a microscope, may contain up to e.g. 12 lenses Specification and Identification - Magnification (enlargement) - Numerical aperture (resolution) - Immersion medium (should fit to embedding medium) - corrections (spherical; chromatical) - working distance - tube length (infinity or 160 mm) - coverslip thickness 9

Objective magnification and resolution A microscope magnifies a specimen with a certain resolution. 10x / high NA 10x / low NA objectives 20x N.A. 0.75 0.37 µm resolution 40x N.A. 0.75 0.37 µm resolution 60x N.A. 0.75 0.37 µm resolution Magnification without resolution is useless : empty magnification! 10

Illumination of the specimen Lens Light to/ from point source focal point in XZ Sample Z view Focal point Focal plane resolution in XY approx 2 x better than in Z The better the objective the smaller the focal point better means Numerical Aperture See next slide Field of view 11

Illumination of the specimen - resolution d = λ 2nsin θ d: point resolution, the shorter the lenght is better for us [nm] λ: wavelength of light used ([nm], visible light 400-700 nm) n: refractive index (air: 1, water: 1.3, oil: 1.4-1.5) ѳ: the maximum cone of light that can enter or exit the lens nsin(2 ѳ) : numerical aperture (NA) 12

Resolution, Airy disk, NA & WD WD 13

Objective Resolution and Airy disk A point of light will not be a point of light Light originally coming from a point and passing through lenses etc. will not be a point again in the image, but rather a dot (1st maxima, AiryDisk) with several side maxima separated by mininima (interference pattern). he yellow dots shall indicate infinite points, where light originally came from. z Airy disk in XZ x The Spreading from Point light source to a Airy disk Image is called Point Spread Function (PSF). PSF gets bigger with mismatch of embedding medium and objective

Match embedding medium to objective. Light coming from high to low density medium (glas vs air) gets refracted away from the vertical of the incident angle, eventually misses the lense and is lost for imaging. Appliance of immersion (oil, glycerin, water) between Coverslip and Lense with similar refractive index as Glas will reduce refraction and enhance light yield which in turn gives better Airy pattern (resolution) Light coming from one point source will get scattered and refracted into different angles the point gets spreaded. By applying High Numerical Aperture this effects are kept to a minimum. PSF gets bigger with mismatch of embedding medium and objective

Practical tips Bad approach Match objective with correct - Immersion media - Coverslip thickness - Embedding media - Imaging setup - Sample preparation good approach - Objectives indicate for which immersion media they are made for - Objectives indicate for which coverslip thickness they are made for - Embedding medium has optimally same RI like immersion media - Place the sample as close as possible towards the objective lens - RI of sub-cellular components considerably lower than that of immersion media, and in many cases these RI are uncertain and vary throughout the specimen. Different fixations might destroy antigen to be targeted or might quench fluorescence (of e.g. GFP)

THE FLUORESCENCE MICROSCOPE 17

Fluorescence Examples of fluorescent probes Principle of fluorescence Principle of fluorescent microscope Excitation-Emission filter cube

Combining fluorescent dyes - crosscheck To avoid false positive images in Fluorescence microscopy check for Seeing is Believing BUT Is it true? What s to be seen in pos/neg control stained unstained Crossreact AbX with AbY? AbX AbY(1) AbX AbY(2) X Y X Y Unspecific binding by Ab? cell with - without target X X Appropr. fixation? Fixation A Fixation B X Crosstalk/ Bleeding through? ex Use quality objectives, correct filter, embedding medium to avoid abberations 19

Widefield vs Optical section Kidney sample 10µm thick, 63x/NA 1.43, Widefield image and optical section using Apotome technique. 20

Bleaching Bleaching before and after 100x imaging same area with Widefield microscopy. Test sample is a strong stain and so bleaching might be subtle and only clearly be see in LUT (look-up-tables) Intensities of emission are shown in LUT Black to white LUT= Blue, over green, yellow, red You might not see the subtle changes But would like to compare Intensities Between image 1 and 2? Be aware... 21

IMAGING 22

Imaging digital camera - pixel 0-255 (8bit) All pixels of the camera will be exposed to light at once; image is processed all pixels at once Black&White cameras pixel does not care about color. For Fluorescence microscopy use B&W cameras (with appropriate filtercubes) 23

Imaging Features of a digital camera Spatial Resolution: ability to capture fine specimen details without pixels being visible in image (1308x1040 pixel, 6.45x6.45µm pixel on 2/3 chip) Light-Intensity Resolution: dynamic range or number of gray levels that are distinguishable in image. (12 bit or 16 bit) Time Resolution: frame rate - the ability to follow movement or rapid kinetic processes (38 fps) Signal-to-Noise Ratio: visibility and clarity of specimen signals relative to the image background Spectral Sensitivity: range of wavelength on which camera reacts (350-1000 nm) (data from Zeiss Axiocam MRm) 24

Image quality dynamic range Acquire your image with appropriate grey level (8bit at least) to represent different intensity levels. Incoming photons 12 or 16 (4096 65536 grey levels )bit images would also allow you e.g. to enhance features lying in the dark to stretch them into the light... 25

Imaging color vs black&white In Color Cameras each pixel is overlaid by color filter lense pattern The Bayer mosaic. Reduction of sensitivity and actual resolution + +/- +/- +/- + - + - Color pixel red only lets pass light in red range (+) signal. Rest of pixels are calculated in respect to surrounding pixels (+/-). i.e. 66% (2 of 3 colors /px). More green in the Bayer mosaic, therefore human eye is more sensitive to green. Problem: Actual resolution is 2x2 pixel i.e. 4x less Solution: camera with moveable chip are used each pixel will sample light from (9) different positions. High resolution Brightfield Black&White cameras pixel does not care about color. For Fluorescence microscopy use B&W cameras (with appropriate filtercubes) 26

Publishing photos Use the highest bit depth if needed, and take care about over/underexposure Use the highest quality/resolution as possible Use TIF files Crop image if neccessary Use measure bar to show scale Do not use total magnification, e. g. objective magnification - 60X (which is of course not 60X magnification but 600X at least) Magnification 1000x VS Measure bar (20.000 light years) http://apod.nasa.gov/apod/ap090917.html 27

THANKS FOR YOUR ATTENTION! 28