Technical Information Indumax CLS54

Similar documents
Indumax CLS54D. Technical Information

Technical Information Smartec CLD18

Technical Information Smartec CLD18

Indumax H CLS54. Technical Information

Technical Information Smartec CLD18

Technical Information Memosens CLS82D

Smartec S CLD134. Technical Information

Technical Information Indumax CLS50D/CLS50

Technical Information Condumax CLS21D/CLS21

Technical Information Condumax CLS15D/CLS15

Condumax CLS16 and CLS16D

Indumax CLS50/CLS50D. Technical Information

Condumax CLS21 and CLS21D

Condumax CLS15 and CLS15D

Condumax W CLS19. Technical Information. Conductivity sensors Two-electrode sensors with cell constants k = 0.01 cm 1 or k = 0.

Condumax CLS15 and CLS15D

Condumax W CLS15 and CLS15D

Conductivity Measuring Cells for High-Temperature Applications ConduMax H CLS 16

Condumax CLS21 and CLS21D

Condumax W CLS19. Technical Information. Conductivity sensors Two-electrode sensors with cell constants k = 0.01 cm 1 or k = 0.

Condumax W CLS15 and CLS15D

Conductivity measurement SmarTec S CLD 132

Conductivity Sensor for High-Temperature Applications ConduMax W CLS 13

Technical Information Orbipore CPS91D and CPS91

Conductivity sensor for hygienic applications

Technical Information Liquisys M CLM223/253

Ceramax CPS341D. Technical Information

Inductive Conductivity Meter ILM

Inductive conductivity measurement mypro CLM 431 / CLD 431 inductive

The transducer component converts the measured signal and displays the actual value.

Technical Information Orbipac CPF81D/CPF82D and CPF81/CPF82

Smartec S CLD134. Operating Instructions. Conductivity Measuring System. BA401C/07/en/ Software version 1.

In-Line Resistance Thermometers Model TR25

Liquisys M CLM223/253

OPTIFLUX 5000 Technical Datasheet

Technical Information Memosens CPS96D

Technical Information OUSBT66

Inductive conductivity transmitter JUMO CTI-Junior

In-line resistance thermometer Model TR25

Pressure transmitter For sanitary applications Model SA-11

OPTIFLUX 5000 Technical Datasheet

Magnetic-Inductive flow Meter fmi

Digital conductivity transmitter

Electromagnetic Flow Transmitter

Inductive Conductivity/Concentration and Temperature Transmitter with switch contacts. Temperature sensor. Graphics LC display.

LCI. Inductive Conductivity/ Concentration and Temperature Transmitter. Compact or separate version

Level sensor Magnetostrictive, high-resolution measuring principle For sanitary applications, model FLM-H

Memosens CPS96D. Technical Information

Rosemount 225. Toroidal Conductivity Sensors. Product Data Sheet July Toroidal Conductivity Sensor LIQ-PDS-225

Tophit CPS441 and CPS441D

Continuous Level Sensor NSL-M-00

Conductive conductivity measurement mypro CLM 431 / CLD 431 conductive

Type PLC Fitting. Valve for Continuous control. Diaphragm valve. Output

Endress Hauser. SmarTec M CLD 133 Conductivity Measurement

Technical Information Memosens CCS50D

Orbipac CPF81/CPF82 and CPF81D/CPF82D

FLOWave SAW-Flowmeter

Flow Measurement SITRANS F M

Assembly 249 (9.80) 259 (10.2) Example: level probe with Pt100 and HUT temperature head transmitter ( ma) ma. Core colors: RD BK YE BU

Process transmitter Model UPT-20, with pressure port Model UPT-21, with flush diaphragm

INLINE flowmeter for continuous flow measurement

A Higher Level of Performance. Data Sheet. Series Pressure and Level Transmitters. For more information, please visit >

omnigrad M TR 46 Technical Information Hygienic in-line RTD thermometer PCP ( ma), HART or PROFIBUS-PA electronic

Memosens New technology for contactless, inductive, digital transmission of ph signals

Digital inductive conductivity transmitter

Submersible pressure sensor For oils and fuels Model LF-1

BATCHFLUX 5500 C Technical Datasheet

Submersible pressure sensor For superior applications Model LF-1

RTD Temperature Sensor omnigrad T -TSM 480, TR 480. Hygienic RTD sensor with Pt 100, class A TSM 480 with electronics programmable via PC

SITRANS F flowmeters SITRANS F M. Flow sensor MAG /48 Siemens FI Overview

LF Conductivity Measuring Probe

Flow Measurement SITRANS F M. Flow sensor MAG /49. Overview

OPTIFLUX 1000 Technical Datasheet

Turbimax CUS71D. Technical Information. Ultrasonic interface sensor Immersion sensor for interface measurement

P R o D u c T I n F o R m at I o n. TDR Level Sensor LFP. The clean solution

Submersible pressure sensor For water and wastewater Model LW-1

ED701 General Industry Pressure Transmitter

Conductivity-Converter

Intrinsically safe pressure transmitter MBS 4201, MBS 4251, MBS 4701 and MBS 4751

Type F-20-3A Sanitary Pressure Transmitter NEMA 4X with Integral Junction Box

Pharma Temperature Sensor TFP-EX

Temperature Sensor with G1/2" hygienic

Conductivity measurement Mycom S CLM 153

prosonic T FMU230, FMU231

Electromagnetic Flow Measuring System Two-wire, loop-powered PROline promag 23 H

In-Line Resistance Thermometers Model TR472

Capacitive Limit Switch Pharma NCS

Capacitive Limit Switch Pharma NCS

Mass Flowmeter CoriolisMaster FCM2000

JUMO Dtrans T100 Screw-in RTD Temperature Probe with/without Transmitter

Sterilizable IsFET sensor for ph Measurement TopHit CPS 441

TIDALFLUX 2300 F Technical Datasheet

omnigrad T - TSM 480, TR 480. Technical Information. RTD Temperature Sensor Hygienic design TSM 480 with electronics programmable via PC

Temperature Sensor Tri-Clamp

Temperature Sensor. Temperature Transmitter MPU-LCD with Display Product Information TSM and TSB FOOD SENSORS FOR FOOD AND LIFE SCIENCES.

Page Type Resolution (mm) min. max. Output

Content. Flow meter F2-UN. Description Technical Data Dimension drawings Operation... 8 Mounting instructions...

SMARTPAT COND 1200 Technical Datasheet

Capacitive Limit Switch Pharma NCS

Continuous Level Sensor NSL-M

Transcription:

TI00400C/07/EN/13.16 71313416 Products Solutions Services Technical Information Indumax CLS54 Hygienic inductive conductivity sensor for applications in the food, beverage, pharmaceutical and biotech industries Application The CLS54 conductivity sensor is specifically designed for use in hygienic applications in the food, beverage and pharmaceutical industries and in biotechnology. Thanks to its hygiene certificates and its food-safe, virgin PEEK design without any joints or crevices, it meets the strict demands of these industries. The CLS54 is ideal for: Phase separation of product/water and product/product mixtures in pipe systems Control of cleaning in place (CIP) processes in the return line Concentration control in the remaking of CIP cleaning agents Product monitoring in pipes, bottling plants and quality assurance Leakage monitoring in the following industries: Dairies Breweries Beverages (water, juices, soft drinks) Pharmaceuticals and biotechnology Use with the Liquiline CM42 and Liquisys CLM223/253 transmitters; integral part of the Smartec CLD134 measuring system. Your benefits Unique hygienic design, therefore no risk of recontamination Has all the hygiene certificates required in the hygiene sector Has all the process connections commonly used in the hygiene sector Durable thanks to fully encapsulated, sealless design Fast measurement with temperature response time t 90 under 26 s, ensuring safe and efficient phase separation Insensitive to electrode contamination and polarization thanks to inductive measuring principle Explosion protection

Function and system design Measuring principle Inductive conductivity measurement An oscillator (1) generates an alternating magnetic field in the primary coil (5), which induces a current flow (4) in the medium. The strength of the current depends on the conductivity and thus on the ion concentration in the medium. The current flow in the medium, in turn, generates a magnetic field in the secondary coil (3). The resulting induced current is measured by the receiver (2) and used to determine the conductivity. 1 2 5 3 1 2 3 4 5 Oscillator Receiver Secondary coil Current flow in the medium Primary coil Advantages of inductive conductivity measurement: No electrodes and therefore no polarization effects Accurate measurement in media with a high degree of pollution and a tendency to form buildup Complete galvanic isolation of the measurement and the medium 4 A0024926 Important features of Indumax CLS54 Hygiene The injection-molded PEEK sensor offers outstanding chemical, mechanical and thermal resistance and is hygienically safe thanks to its seamless, crevice-free design. As only virgin PEEK is in contact with the medium, highest biological safety for the food, beverage and pharmaceutical sectors is guaranteed. The sensor has been designed according to the latest guidelines of the American Society of Mechanical Engineers - Bioprocessing Equipment (ASME BPE) and meets the requirements of EC Regulation No. 1935/2004 of the European Parliament on materials and articles intended to come into contact with food. Certified hygiene The sensor has all the necessary listings and approvals for the hygiene sector, such as exclusive use of FDA-listed materials for surfaces in contact with medium and certification according to EHEDG (European Hygienic Engineering & Design Group) and 3-A standards. A certificate for biological reactivity testing according to USP part <97> and <88> class VI including lot number tracking back to compounding is also available as an option. Process connections The sensor is available with all the process connections commonly used in hygienic applications. The most common connections can be supplied as standard. Other process connections can be ordered as a custom-made product. Process temperature, process pressure The use of special materials means that the sensor is suitable for exposure to continuous temperatures of up to 125 C (257 F). It may be exposed to temperatures of up to 150 C (300 F) for a short time (max. 60 min)). The sensor's pressure resistance is 12 bar (174 psi) at temperatures up to 90 C (194 F) and at higher temperatures is always above the steam pressure. The sensor is suitable for underpressure applications. Temperature measurement The sensor has an integrated temperature sensor with a temperature response time t 90 of less than 26 s, making efficient phase separations possible at varying and quickly changing process temperatures. The temperature sensor is embedded into the PEEK body without any seals, thereby ensuring a long sensor service life. 2 Endress+Hauser

Measuring system A complete measuring system consists of the following components at least: The CLS54 inductive conductivity sensor A transmitter, e.g. Liquiline CM42 1 2 3 A0005431 1 Example of a measuring system 1 Indumax CLS54 2 Liquiline CM42 transmitter 3 Measuring cable Input Measured values Conductivity Temperature Measuring ranges Conductivity Recommended range: 100 μs/cm to 2000 ms/cm (uncompensated) Temperature 10 to +150 C (+14 to +302 F) Cell constant k = 6.3 cm 1 Temperature measurement Pt 1000 (Class A according to IEC 60751) Endress+Hauser 3

} } } Indumax CLS54 Power supply Electrical connection The sensor is supplied with a fixed cable. The wiring diagram is provided in the Operating Instructions of the transmitter used. Connection via a VBM junction box is necessary for a cable connection. The extension to the transmitter is via the CLK6 cable. Cable WH RD Coax WH Coax RD Ferrule RD Ferrule BU Ferrule RD Ferrule BU GN WH YE BN n.c. Screen Inner connector Screen Inner connector Pt 1000 2 Fixed cable/special measuring cable CLK6 A0005433-EN Cable length: Max. total length of 55 m (180 ft) Performance characteristics Temperature response time Maximum measured error t 90 26 s ± (0.5 % of reading + 10 μs/cm) after calibration (plus uncertainty of the conductivity of the calibration solution) 4 Endress+Hauser

Installation Orientation The sensor must be completely immersed in the medium. Avoid air bubbles in the area of the sensor. > 1 m (3.3 ft) 3 Installation positions of the conductivity sensor A0017691 If the flow direction changes (after pipe bends), turbulence in the medium can result. Install the sensor at a distance of at least 1 m (3.3 ft) downstream from a pipe bend. The medium should flow parallel to the sensor bore (see arrows on the sensor body). The symmetrical measuring channel allows flow in both directions. For a 3-A compliant installation, the following must be noted: After the instrument is installed its hygienic integrity shall be maintained. All process connections must be 3-A compliant. Endress+Hauser 5

Installation factor The ionic current in the liquid is affected by the walls in confined installation conditions. This effect is compensated by what is referred to as the installation factor. The installation factor can be entered in the transmitter for the measurement or the cell constant is corrected by multiplying by the installation factor. The value of the installation factor depends on the diameter and the conductivity of the pipe nozzle as well as the distance a between the sensor and the wall. The installation factor f (f = 1.00) can be disregarded if the distance to the wall is sufficient (a > 15 mm, from DN 65). If the distance to the wall is smaller, the installation factor increases for electrically insulating pipes (f > 1), and decreases for electrically conductive pipes (f < 1). It can be measured using calibration solutions, or a close approximation can be determined from the diagram below. f 4 CLS54 installation a Wall distance a A0005453 1.40 2 1.20 1.00 0.80 1 0 5 10 15 20 25 a [mm] 0.20 0.39 0.59 0.79 0.98 a [inch] 5 Relationship between installation factor f and wall distance a 1 Electrically conductive pipe wall 2 Electrically insulating pipe wall A0005441 Air set To compensate residual coupling in the cable and between the two sensor coils, zero adjustment in air ("air set") must be performed before installing the sensor. Follow the instructions provided in the Operating Instructions of the transmitter used. Environment Ambient temperature range Storage temperature -20 to +60 C (-4 to 140 F) -25 to +80 C (-13 to +176 F) Humidity 5 to 95 % Degree of protection IP 68 / NEMA type 6 (1 m (3.3 ft) water column, 50 C (122 F), 168 h) 6 Endress+Hauser

Process Process temperature Sterilization Process pressure (absolute) Temperature/pressure ratings -10 to +125 C (+14 to +257 F) 150 C (302 F) / 5 bar (72.5 psi) (max. 60 min.) 13 bar (188.5 psi) up to 90 C (194 F) 9 bar (130.5 psi) at 125 C (257 F) 1 to 6 bar (14.5 to 87 psi) in CRN environment tested with 50 bar (725 psi) Underpressure down to 0.1 bar (1.45 psi) p (abs.) [psi] [bar] 188.5 130.5 13 9 B 87 6 A 14.5 1 10 0 10 30 50 70 90 110 125 150 [ C] 32 50 86 122 158 194 230 257 302 [ F] 6 A B Pressure/temperature ratings Temporarily for sterilization (max. 60 min.) MAWP (maximum allowable working pressure) according to ASME-BPVC Sec. VIII, Div 1 UG101 for CRN registration A0008379 Endress+Hauser 7

Mechanical construction Design 2 3 4 1 7 Indumax CLS54 1 Temperature sensor 2 Housing 3 Process connection 4 Flow opening A0005430 Dimensions 126(4.96) Ø 64(2.52) 54.5(2.15) 28.5(1.12) 78.5(3.09) 58(2.28) Ø 9 (0.35) 33(1.30) Ø 6(0.24) 11.5(0.45) 10.2(0.40) 8 Dimensions in mm(inch) (long version) A0005429 Weight 0.3 to 0.5 kg (0.66 to 1.1 lb.) depending on version plus cable 8 Endress+Hauser

Materials In contact with medium Virgin PEEK Not in contact with medium PPS-GF40 Stainless steel 1.4404 (AISI 316L) Screws: 1.4301 (AISI 304) Cable gland: PVDF Seals: FKM, EPDM Cable: TPE Surface roughness Ra 0.8 μm (smooth, injection-molded PEEK surface) at surfaces in contact with medium Endress+Hauser 9

Process connections A B 126(4.96) 126(4.96) Ø 89.5 (3.52) 17(0.67) Ø 84 (3.31) 12.5(0.49) 0 16(0.63) 0 16(0.63) 36.5(1.44) Ø 6(0.24) 36.5(1.44) Ø 6(0.24) A0004949 9 Process connections for CLS54 (short version), dimensions in mm (inch) A NEUMO BioControl D50 for pipe connection: DN 40 (DIN 11866 series A, DIN 11850); DN 42.4 (DIN 11866 series B, DIN EN ISO 1127); 2" (DIN 11866 series C, ASME-BPE) B Varivent N DN 40 to 125 A B 126(4.96) 126(4.96) 11(0.43) 0 62.6(2.46) 4(0.16) 0 Ø 57.7(2.27) Ø 64.7(2.55) 58(2.28) 78.5(3.09) 58(2.28) 78.5(3.09) C D 126(4.96) 126(4.96) 2.85(0.11) 0 54.5(2.15) 64(2.52) 0 66.85(2.63) 58(2.28) 78.5(3.09) 58(2.28) 78.5(3.09) A0005436 10 Process connections for CLS54 (long version), dimensions in mm (inch) A Sanitary connection DIN 11851, DN 50 B SMS coupling 2" C Clamp ISO 2852, 2" D Aseptic coupling DIN 11864-1 form A, for pipe according to DIN 11850, DN 50 10 Endress+Hauser

Chemical resistance Medium Concentration PEEK Caustic soda NaOH 0 to 15 % 20 to 90 C (68 to 194 F) Nitric acid HNO 3 0 to 10 % 20 to 90 C (68 to 194 F) Phosphoric acid H 3 PO 4 0 to 15 % 20 to 80 C (68 to 176 F) Sulfuric acid H 2 SO 4 0 to 30 % 20 C (68 F) Peracetic acid H 3 C-CO-OOH 0.2 % 20 C (68 F) Errors and omissions excepted Endress+Hauser 11

Certificates and approvals mark Hygiene Declaration of Conformity The product meets the requirements of the harmonized European standards. As such, it complies with the legal specifications of the EC directives. The manufacturer confirms successful testing of the product by affixing to it the mark. FDA All materials in contact with the product are listed by the FDA. EHEDG Certified cleanability according to EHEDG TYPE EL-class I. 3-A When using the sensor in hygienic applications, please note that the cleanability of the sensor also depends on the way the sensor is installed. To install the sensor in a pipe, use the appropriate and EHEDG-certified flow vessels for the particular process connection. Certified according to 3-A Standard 74- ("3-A Sanitary Standards for Sensor and Sensor Fittings and Connections Used on Milk and Milk Products Equipment"). Biological reactivity (USP class VI) (option) Biological reactivity test certificate (Certificate of Compliance) according to USP (United States Pharmacopoeia) part<87> and part <88> class VI with lot number traceability of materials in contact with the medium. Ex approvals ATEX II 1G Ex ia IIC T3/T4/T6 CSA IS/NI Cl. I Div. 1 & 2 GP A - D in conjunction with the Liquiline M CM42 transmitter Pressure approval Canadian pressure approval for pipes according to ASME B31.3 Regulation (EC) No. 1935/2004 Meets the requirements of Regulation (EC) No. 1935/2004 12 Endress+Hauser

Ordering information Product page Product Configurator Scope of delivery www.endress.com/cls54 The navigation area is located on the right of the product page. 1. Under "Device support" click "Configure your selected product". The Configurator opens in a separate window. 2. Select all the options to configure the device in line with your requirements. In this way, you receive a valid and complete order code for the device. 3. Export the order code as a PDF or Excel file. To do so, click the appropriate button at the top of the screen. The scope of delivery includes: Sensor in the version ordered Operating Instructions Accessories The following are the most important accessories available at the time this documentation was issued. For accessories not listed here, please contact your service or sales office. Cable extension Measuring cable Measuring cable CLK6 Extension cable for inductive conductivity sensors, for extension via VBM junction box Sold by the meter, order number: 71183688 Junction box VBM Junction box for cable extension 10 terminal strips Cable entries: 2 x Pg 13.5 or 2 x NPT ½" Material: aluminum Degree of protection: IP 65 Order numbers Cable entries Pg 13.5 : 50003987 Cable entries NPT ½": 51500177 Desiccant pouch Desiccant pouch with color indicator for VBM junction box Order No. 50000671 Calibration solutions Conductivity calibration solutions CLY11 Precision solutions referenced to SRM (Standard Reference Material) by NIST for qualified calibration of conductivity measuring systems in accordance with ISO 9000 CLY11-B, 149.6 μs/cm (reference temperature 25 C (77 F)), 500 ml (16.9 fl.oz) Order No. 50081903 CLY11-C, 1.406 ms/cm (reference temperature 25 C (77 F)), 500 ml (16.9 fl.oz) Order No. 50081904 CLY11-D, 12.64 ms/cm (reference temperature 25 C (77 F)), 500 ml (16.9 fl.oz) Order No. 50081905 CLY11-E, 107.00 ms/cm (reference temperature 25 C (77 F)), 500 ml (16.9 fl.oz) Order No. 50081906 Technical Information TI00162C Endress+Hauser 13

www.addresses.endress.com