EMC TEST REPORT. Report No.: TS EME Model No.: 33XR-A Issued Date: Jan. 08, 2009

Similar documents
EMC TEST REPORT. NORTE SIRIUS ENTERPRISE CO., LTD , Shin-Sheng St., Chung-Ho Dist, New Taipei City, Taiwan

TEST REPORT... 1 CONTENT...

EMC Test Report. Report Number: M030826

EMC Testing Report. Dual-120CS. Yuan Hsun Electric Co., Ltd.

EMC TEST REPORT. Report No. : EM/2004/10096 Page : 1 of 19

BUREAU VERITAS. itaste svd

EMC REPORT. ShenZhen KY Technology Co.,Ltd. No.369, BaoTian 1st RD, TieGang Industrial Park, Xixiang Town, Baoan District, ShenZhen, PRC.

EMC TEST REPORT for : DONGGUAN EVER DEVELOPMENT ELECTRONIC CO., Electronic calculator Model No.: KF15758

Report for Excelsys EMC Measurements for 4Xgen Purchase Order: Project Number EMT07J026 Rev. B

EMC TEST REPORT. For. Boardcon Technology Limited. MINI9G45 Computer on Module. Model No. : MINI9G45

Certificate of Test AND KEEPS ALL REQUIREMENTS ACCORDING THE FOLLOWING REGULATIONS IEC :2001 IEC :2007

EMC test report AU01+E04

ITUNER NETWORKS CORPORATION EMC REPORT Fremont Blvd. Fremont, CA

EN 55015: 2013 Clause Pass. EN 55015: 2013 Clause Pass. EN 55015: 2013 Clause Pass

EN :2007+A1:2011 Electromagnetic compatibility Emission standard for residential, commercial and light-industrial environments

CTS (NINGBO) TESTING SERVICE TECHNOLOGY INTERNATIONAL OPERATE ACCORDING TO ISO/IEC EMC TEST REPORT

EMC REPORT DONGGUAN FIT-WATCH CO., LTD. 18#,Hedong No.1 road,jinsha village,changan town, Dongguan City, Guangdong Province.

EMC VERIFICATION SUMMARY Report No.: SZHH

EMC Test Report FATEK AUTOMATION CORP. FATEK. FBs-TBOX N/A

EMC TEST REPORT. according to

ELECTRONICS TESTING CENTER(ETC), TAIWAN

Test Report. Ningbo Goldmore Industrial Co.,Ltd.

TEST SUMMARY. Prüfbericht - Nr.: Test Report No.: Seite 2 von 25. Page 2 of 25

EN61326 EMC COMPLIANCE REPORT on the LP Series Ultrasonic Transmitter Remote Amplifier and Transducer for Hawk Measurement Systems Pty Ltd

CE EMC Test Certification

Report for Excelsys EMC Measurements for 6Xgen Purchase Order: by Project Number EMT08J027

Harmonic Current emission EN :2014 Class A Pass. Voltage Fluctuation and Flicker EN :2013 Clause 5 Pass

EMC Test Report For Changzhou Airwheel Technology Co.,Ltd. Airwheel Model: A6S,A6,A6P,A6PS,A6T,A6TS

Shenzhen Toby Technology Co., Ltd. EMC Test Report. Report No.: TB-EMC Page: 1 of 20

Overview of EMC Regulations and Testing. Prof. Tzong-Lin Wu Department of Electrical Engineering National Taiwan University

EMC TEST REPORT for LEDELS LIGHTING CO., LTD. LED module Model No. : LL-F12T4815X6B

Discontinuous Disturbance (Click) EN :2006+A1:2009+A2:2011 Clause N/A** Radiated Emission, 30MHz to 1000MHz

EMC TEST REPORT. ADDRESS : #2 Creation Rd. 4, Science-Based Ind. Park Hsinchu Taiwan, R.O.C.

EMC TEST REPORT For ShenZhen ZhangQing Electronic LTD poe tester Model No.: WS POE Tester/Detector Additional Model No: Please Refer to Page 8

EMC TEST REPORT For MPP SOLAR INC Inverter/ Charger Model Number : PIP 4048HS

TEST SUMMARY Seite 2 von 27. Prüfbericht - Nr.: Test Report No HARMONICS ON AC MAINS RESULT: Passed

ETSI EN V2.2.0 ( ) ETSI EN V2.1.1 ( ) TEST REPORT. United GULF GATE Co.

Page: 1 of 20 EMC TEST REPORT EN55024:1998+A2:2003

EMC TEST REPORT for SHENZHEN MUST ENERGY TECHNOLOGY CO., LTD

TEST REPORT. Report No. : EM/2008/80003 Page : 1 of 54

Prepared for Address. : KST DIGITAL TECHNOLOGY LIMITED : No.226, Pangu Street, Meixian, Meizhou, Guangdong. Prepared by

EMC COMPLIANCE TEST REPORT

1 SUMMARY OF STANDARDS AND RESULTS

9. Electrostatic Discharge Immunity Test (ESD)

EMC TEST REPORT For INNOKIN TECHNOLOGY CO., LIMITED. ITaste vv. Model No.: ITaste vv

EMC TEST REPORT. For MICRO-STAR INT'L CO.,LTD. USB Card Reader

APPENDIX A. GENERAL DESCRIPTION OF EQUIPMENT UNDER TEST

CE TEST CERTIFICATION

CE EMC Test Certification

Discontinuous Disturbance (Click) EN : 2011 Clause Pass Radiated Emission, 30MHz to 1000MHz

BS EN TESTS ON THE IT TOXIC GAS DETECTOR MODULE

Test Report. Guangdong East Power Co., Ltd. Fully Automatic AC Voltage Regulator. Brand Name:

CONFORMANCE TEST REPORT FOR EN /-6

Report No.: 11A080304E-E Page 1 of 35. CE Test Report. (Declaration of Conformity) for. Electromagnetic Compatibility.

APPLICATION FOR EMC DIRECTIVE. On Behalf of. Shenzhen Qinhan Lighting Co., Ltd. led flood light. Trade Name:

TEST REPORT. draft draft draft. TRF Originator...: Shenzhen CTL Testing Technology Co., Ltd. Master TRF...: Dated

C E R T I F I C A T I O N

EN v3.1.1 TEST REPORT FOR. Bluetooth Serial Port Module MODEL NUMBER: LMX9838 REPORT NUMBER: 14U17620-E1V3 ISSUE DATE: AUGUST 31, 2017

TEST SUMMARY. Prüfbericht - Nr.: Test Report No.: Seite 2 von 27. Page 2 of 27

TEST REPORT FROM RFI GLOBAL SERVICES LTD

Test Certificate A sample of the following product received on May 13, 2011 and tested on May 13, 18 and 19, 2011 complied with the requirements of,

EMC TEST REPORT. for. Coliy Technology Co.,Ltd. Fluxgate Gaussmeter

EN 55022: 2010+AC:2011 Clause 6.1 Pass. Harmonic Current EN :2006+A1:2009+A2:2009 Class A N/A

CE TEST CERTIFICATION

TABLE OF CONTENTS 1. GENERAL DESCRIPTION OF EUT 3 2. GENERAL INFORMATION OF TEST CONDUCTED EMISSION TEST RADIATED EMISSION TEST 11

EMC Test Report. Product. Model Number. Prepared for : Shenzhen top technology co., LTD Address : No 2.First Rd Of Lucky,Longgang Ping Shenzhen China

PT. SARANA KARYA SOLUSINDO.

TEST REPORT. Table of Contents

C E R T I F I C A T I O N

CTS (NINGBO) TESTING SERVICE TECHNOLOGY INTERNATIONAL OPERATE ACCORDING TO ISO/IEC EMC TEST REPORT

Test Verification of Conformity

CE EMC CERTIFICATION

ENGINEERING TEST REPORT #: A LSR JOB #: C-2411

Test Report Report No.: G150424B02-03EN Date: May. 07, 2015 Page 1 of 49

Test Report GUANGDONG EAST POWER CO., LTD

Guangdong East Power Co., Ltd. Uninterruptible Power Systems

Kutta Radio & charger Report Number: TRE0026-5/17 Page 1 of 27 EMC TEST REPORT. TEST SAMPLE (EUT) : Kutta Medium frequency radio & Charger

EMC TEST REPORT For. Shenzhen Ruideou Keji Youxian Gongsi. Shenzhen Ruideou Keji Youxian Gongsi

EMC standards. Presented by: Karim Loukil & Kaïs Siala

CE EMC TEST REPORT. According to

TEST REPORT. Report Number: DAL-001 Project Number: G Report Issue Date: February 6, 2013

EMC TEST REPORT. According to

EMISSION TEST REPORT

TEST REPORT. : The submitted samples complied with the above EMC standards. TRF No.: AS NZS (2012)-b

CE/EMC TEST REPORT. For HK QNOVATOR TECHNOLOGY CO.,LIMITED GS02-M X1, BL02, GS01, GS01-M, GS02, GS03, GS04, GS05, GS06, GS07, GS08, GS09, GS10

Report No.: 063L107-RF-CE-P14V02-2. Test Report. MICRO-STAR INTL Co., LTD. No. 69, Li-De St., Jung-He City, Taipei Hsien, Taiwan, R.O.C.

EN v1.2.1 ( ) TEST REPORT FOR n 1X2 PCIe MINICARD TRANSCEIVER MODEL NUMBER: AR5B91 REPORT NUMBER: 08U

2620 Modular Measurement and Control System

ELECTRONICS TESTING CENTER, TAIWAN

EMC TEST REPORT For CE

EMC TEST REPORT For. Shenzhen Micsig Instruments Co., Ltd. Handheld Multifunctional Oscilloscope. Model No.: MS520S

A) Documentation Page(s) Test report 1-17 Directory 2 Test Regulations 3 General Remarks and Summary 17 Test-setups (Photos) 18-23

Report No.: 9A042401C2 Page 1 of 36. Test Report EN : Prepared for

STC Test Report. Date : Page 1 of 13 No. : HM161169

EMC Test Report. Report No.: AGC EE01 : VR 6600PRO, VR 6610, VR 6700, VRP 6000, VRP 6800 : VERO GLOBAL COMMUNICATION CO.

TABLE OF CONTENTS. 6. PHOTOGRAPH Photo of Conducted Emission Measurement Photo of Radiation Emission Measurement...

TABLE OF CONTENTS. 1. General Description of EUT General Information of Test Conducted Emission Tests 7-12

Ave output power ANT 1(dBm) Ave output power ANT 2 (dbm)

CTS (NINGBO) TESTING SERVICE TECHNOLOGY INTERNATIONAL OPERATE ACCORDING TO ISO/IEC EMC TEST REPORT

TEST REPORT. Power Spout PLT V. tested to the specification

Transcription:

Page 1 of 18 EMC TEST REPORT Report No.: TS08100063-EME Model No.: 33XR-A Issued Date: Jan. 08, 2009 Applicant: Test Method/ Standard: Test By: FLUKE CORP. 6920 Seaway Blvd, M/S 266D Everett, WA 98203 U.S.A. EN 61326-1: 2006 EN 61000-3-2: 2006 EN 61000-3-3: 1995+A1: 2001+A2: 2005 Intertek Testing Services Taiwan Ltd. No. 11, Lane 275, Ko-Nan 1 Street, Chia-Tung Li, Shiang-Shan District, Hsinchu City, Taiwan This test report consists of 18 pages in total. It may be duplicated completely for legal use with the allowance of the applicant. It shall not be reproduced except in full, without the written approval of Intertek Laboratory. The test result in this report only applies to the tested sample. The test report was prepared by: These measurements were taken by: Sign on File Sammi Liu/ Assistant Sign on File Bell Huang/ Engineer The test report was reviewed by: Name Rico Deng Title Senior Engineer

Page 2 of 18 Table of Contents 1. General Information... 3 1.1 Identification of The EUT... 3 1.2 Additional Information About The EUT... 3 1.3 Peripherals Equipment... 3 2. Test Summary... 4 3. Test Specifications... 5 3.1 Standards... 5 3.2 Mode of operation during the test... 5 3.3 Performance criteria... 6 4. CISPR 11 Radiated Emission Test... 7 4.1 Operating Environment... 7 4.2 Test Procedure... 7 4.3 Test Equipment... 8 4.4 Radiated Emission Limit... 8 4.5 Uncertainty of Radiated Emission... 8 4.6 Radiated Emission Test Data... 9 5. IEC 61000-4-2 Electrostatic Discharge Immunity Test... 11 5.1 Operating Environment... 11 5.2 Purpose... 11 5.3 Test Set-Up... 11 5.4 Test Conditions... 11 5.5 Test Equipment... 12 5.6 Test Result... 12 6. IEC 61000-4-3 Radiated, Radio-Frequency, Electromagnetic Field Immunity... 13 Test... 13 6.1 Operating Environment... 13 6.2 Purpose... 13 6.3 Test Set-Up... 13 6.4 Test Conditions... 13 6.5 Test Equipment... 14 6.6 Generation of The Electromagnetic Field... 14 6.7 Test Results... 15 Appendix A1: External photo of EUT... 16 Appendix B1: Radiated Emission Test Set-up... 18

Page 3 of 18 1. General Information 1.1 Identification of The EUT Product: Digital Multimeter Model No.: 33XR-A Rated Power: DC 9 V from battery Power Cord: N/A Data Cable: N/A Sample receiving date: Sep. 23, 2008 Testing date: Oct. 15, 2008 ~ Jan. 06, 2009 Note 1: This report is for the exclusive use of Intertek's Client and is provided pursuant to the agreement between Intertek and its Client. Intertek's responsibility and liability are limited to the terms and conditions of the agreement. Intertek assumes no liability to any party, other than to the Client in accordance with the agreement, for any loss, expense or damage occasioned by the use of this report. Only the Client is authorized to permit copying or distribution of this report and then only in its entirety. Any use of the Intertek name or one of its marks for the sale or advertisement of the tested material, product or service must first be approved in writing by Intertek. The observations and test results in this report are relevant only to the sample tested. This report by itself does not imply that the material, product, or service is or has ever been under an Intertek certification program. Note 2: The test report only allows to be revised within three years from its original issued date unless further standard or the requirement was noticed. Note 3: When determining the test conclusion, the Measurement Uncertainty of test has been considered. 1.2 Additional Information About The EUT The EUT is a Digital Multimeter. It has to fulfill the requirements of CISPR 11 emission relevant standards. For more detail features, please refer to user s Manual. 1.3 Peripherals Equipment Resistance 1

Page 4 of 18 2. Test Summary Emission (EN 61326-1: 2006) Standard Test Type Result Remarks CISPR 11: 2003 Class B Radiated Test PASS Pass by -8.90 db at 782.700 MHz with antenna polarization Vertical Immunity (EN 61326-1: 2006) Standard Test Type Result Performance Criteria Test Judgment ESD test PASS Criterion B Meets the requirements of Performance Criterion A IEC 61000-4-2: 1995 +A1: 1998 +A2: 2000 IEC 61000-4-3: 2006 +A1: 2007 RS test PASS Criterion A Meets the requirements of Performance Criterion A Remark: The EUT has been tested/evaluated and pass the above standards without modification.

Page 5 of 18 3. Test Specifications 3.1 Standards EN 61326-1: 2006 Electrical equipment for measurement, control and laboratory use EMC requirements Part 1: General requirements. EN 61000-3-2:2006 Electromagnetic compatibility- Part 3. Limits Section 2. Limits for harmonic current emissions (equipment input current 16 A per phase) EN 61000-3-3: 1995+A1: 2001+A2: 2005 Electromagnetic compatibility - Part 3. Limits Section 3. Limitation of voltage fluctuations and flicker in low-voltage supply systems for equipment with rated current 16 A CISPR 11: 2003 Limits and methods of measurement of radio disturbance characteristics of industrial, scientific and medical (ISM) radio-frequency equipment Since the EUT is not connected to AC source, directly the tests are not applicable for it. The following test items were exempted based on engineering judgment. EN 61000-3-2 Harmonic test EN 61000-3-3 Flicker test IEC 61000-4-4 EFT test IEC 61000-4-5 Surge test IEC 61000-4-6 CS test IEC 61000-4-11 Dip test 3.2 Mode of operation during the test The EUT was supplied with DC 9 V from battery and tested in resistance function mode.

Page 6 of 18 3.3 Performance criteria The performance criterias listed below are based on those regulated in the standard, specified by the manufacturer or derived from the product specification. Criteria A: During testing, normal performance within the specification limits. Criteria B: During testing, temporary degradation, or loss of function or performance which is self-recovering. Criteria C: During testing, temporary degradation, or loss of function or performance which requires operator intervention or system reset occurs.

Page 7 of 18 4. CISPR 11 Radiated Emission Test 4.1 Operating Environment Temperature: 25 Atmospheric Pressure: 1023 hpa Relative Humidity: 50 % Test Voltage: DC 9 V from battery 4.2 Test Procedure Antenna Tower EUT & Peripherals 10 meter 1.0~4.0 meter Receiver Antenna 0.8 meter Ground Plane RF Test Receive Radiated testing was performed at an 10 meter open area test site. The equipment under test were placed on a turntable top 0.8meter above ground. The table was 360 degrees to determine the position of the highest radiation. EUT is set 10 meters from the EMI receiving antenna, which is mounted on a variable height mast. The antenna height is varied between one meter and four meters above ground to find the maximum value of the field strength. Both horizontal polarization and vertical polarization of the antenna was set to conduct the measurement. The bandwidth was set on the EMI meter 120 khz. The levels are quasi peak value readings. The frequency spectrum from 30MHz to 1000MHz was investigated.

Page 8 of 18 4.3 Test Equipment Equipment Brand Model No. EMI Receiver Rohde & schwarz ESCS30 Bilog Antenna Schaffner CBL6112B Note: The above equipments are within the valid calibration period. 4.4 Radiated Emission Limit Frequency (MHz) Distance(m) Class A Class B 30~230 10 40 30 230~1000 10 47 37 Note: 1. The tighter limit shall apply at the edge between two frequency bands. 2. Distance refers to the distance in meters between the measuring instrument antenna and the closet point of EUT. 4.5 Uncertainty of Radiated Emission Expanded uncertainty (k=2) of radiated emission measurement is ±4.68 db.

Page 9 of 18 4.6 Radiated Emission Test Data Polarity: Model No.: Test Condition: Vertical 33XR-A Resistance function mode Remark: 1. Level (db V/m)= Factor (db/m)+ Read Level (db V) 2. Factor = Antenna Factor (db/m) + Cable Loss (db) 3. Over Limit (Margin) (db) = Level (db V/m) Limit Line(dB V/m)

Page 10 of 18 Polarity: Model No.: Test Condition: Horizontal 33XR-A Resistance function mode Remark: 1. Level (db V/m)= Factor (db/m)+ Read Level (db V) 2. Factor = Antenna Factor (db/m) + Cable Loss (db) 3. Over Limit (Margin) (db) = Level (db V/m) Limit Line(dB V/m)

Page 11 of 18 5. IEC 61000-4-2 Electrostatic Discharge Immunity Test 5.1 Operating Environment Temperature: 25 Atmospheric Pressure: 1023 hpa Relative Humidity: 55 % Test Voltage: DC 9 V from battery 5.2 Purpose The object of the test is to evaluate the ESD immunity performance of EUT. 5.3 Test Set-Up A horizontal coupling plane (HCP) was placed on a non-metallic table 0.8 m above a reference ground plane (RGP) and connected to it with a cable with two 470 k resistors. The EUT was placed on an insulation sheet on the HCP and was operated according to the specified operating mode. A vertical coupling plane (VCP) was connected to the RGP with a cable with two 470 k resistors. 5.4 Test Conditions Test level: Air discharge ------- +/- 8 kv Contact discharge ------- +/- 4 kv Single discharge at 1 second interval positive discharge and negative discharge

Page 12 of 18 5.5 Test Equipment Equipment Manufacturer Model No. Electrostatic Discharge System NoiseKen ESS-2002 Note: The above equipments are within the valid calibration period. 5.6 Test Result Point of Discharge Contact Test Point Applied Voltage (kv) Total No. of Discharge (Each Point) Result Criteria Level Remark ±2 20 P A - ±4 20 P A - ±2 20 P A - Air Test Point ±4 20 P A - ±8 20 P A - VCP ±2 20 P A - (4 sides) ±4 20 P A - Description HCP ±2 20 P A - (4 sides) ±4 20 P A - Contact Discharge: 6 test points Metallic Screws Metallic Case Metallic Connect ports Metallic Junctions Others: Note: 1. P means the EUT pass the test. Note: 2. - means not applicable Air Discharge Plastic Screws Plastic Case (gap) Plastic Connect ports Plastic Junctions LED indicator Panel Board Others:

Page 13 of 18 6. IEC 61000-4-3 Radiated, Radio-Frequency, Electromagnetic Field Immunity Test 6.1 Operating Environment Temperature: 23 Atmospheric Pressure: 1023 hpa Relative Humidity: 53 % Test Voltage: DC 9 V from battery 6.2 Purpose This test method subjects the EUT to a power source of disturbance comprising electric and magnetic field, simulating those coming from intentional RF transmitters. 6.3 Test Set-Up The EUT was placed on a non-metallic table 0.8 meter above the reference ground plane (RGP) and was operated according to its specified operating mode. Ferrite tiles/absorbers were placed on the RGP between the EUT and the antenna to reduce the reflections from the RGP. The EUT and its cables were exposed for the electromagnetic field for 1.5m vertically and 1.5m horizontally. The distance between antenna and EUT is 3 meter. 6.4 Test Conditions Test level Test field strength V/m Modulation 1 1 1 khz 80% AM 2 3 1 khz 80% AM 3 10 1 khz 80% AM X Special 1 khz 80% AM The frequency steps Dwell time Frequency range Test ports Test field strength : 1, Log sweep : 3 sec : 80 MHz~1GHz, 1.4 GHz~2 GHz, 2 GHz~2.7 GHz : Enclosure port : 1V/m & 3V/m

Page 14 of 18 6.5 Test Equipment Equipment Manufacture Model No. An-echoic chamber 7m 3m 3m Comtest Instrumentation 9708093 RF signal Generator Marconi 2023B Dual Band RF Power Amplifier High Power Microwave Amplifier Series Kalmus MILMEGA 757LCB AS0102-30 Bi-log Antenna EMCO 3141 RF Power Meter Boonton 4230 Power Sensor Boonton 51011-EMC Power Sensor Boonton 51011-EMC Field Probe Holaday HI-4422 Note: The above equipments are within the valid calibration period. 6.6 Generation of The Electromagnetic Field The electromagnetic field is generated from a computer controlled signal generator. The output power is amplified and then radiated from broadband log periodic antennas. For each sweep a pre-recorded empty chamber calibration file is used to establish the required field strength. When using these files the field strength inside an area of 1.5/1.0 m x 1.5m is in accordance with the standard.

Page 15 of 18 6.7 Test Results Exposed Side: Front Left Rear Right H: Horizontal V: Vertical Frequency (MHz) Horizontal/ Vertical Result Criteria Level Remark 80 MHz to 1 GHz H P A 3 V/m 80 MHz to 1 GHz V P A 3 V/m 1.4 GHz to 2 GHz H P A 3 V/m 1.4 GHz to 2 GHz V P A 3 V/m 2 GHz to 2.7 GHz H P A 1 V/m 2 GHz to 2.7 GHz V P A 1 V/m Note: 1. P means the EUT pass the test. Note: 2. - means not applicable

Appendix A1: External photo of EUT Report No.: TS08100063-EME Page 16 of 18

Page 17 of 18

Appendix B1: Radiated Emission Test Set-up Report No.: TS08100063-EME Page 18 of 18