Optimization of Axial Resolution in Ultrasound Elastography

Similar documents
Introduction. Parametric Imaging. The Ultrasound Research Interface: A New Tool for Biomedical Investigations

CHAPTER 1 INTRODUCTION

Parameter Estimation Techniques for Ultrasound Phase Reconstruction. Fatemeh Vakhshiteh Sept. 16, 2010

Ultrasound Beamforming and Image Formation. Jeremy J. Dahl

COMPUTER PHANTOMS FOR SIMULATING ULTRASOUND B-MODE AND CFM IMAGES

Ihor TROTS, Andrzej NOWICKI, Marcin LEWANDOWSKI

APPLYING SYNTHETIC APERTURE, CODED EXCITATION, AND TISSUE HARMONIC IMAGING TECHNIQUES TO ALLOW ULTRASOUND IMAGING WITH A VIRTUAL SOURCE ROBYN T.

Resolution Enhancement and Frequency Compounding Techniques in Ultrasound.

Coded excitations NINE. 9.1 Temporal coding

The Simulation for Ultrasonic Testing Based on Frequency-Phase Coded Excitation

Ultrasound Bioinstrumentation. Topic 2 (lecture 3) Beamforming

Improving image contrast using coded excitation for ultrasonic imaging

Ultrasound Physics. History: Ultrasound 2/13/2019. Ultrasound

Indicator of Small Calcification Detection in Ultrasonography using Decorrelation of Forward Scattered Waves

Ultrasonic Linear Array Medical Imaging System

768 ieee transactions on ultrasonics, ferroelectrics, and frequency control, vol. 54, no. 4, april 2007

Real Time Deconvolution of In-Vivo Ultrasound Images

AN AUTOMATED ALGORITHM FOR SIMULTANEOUSLY DETERMINING ULTRASONIC VELOCITY AND ATTENUATION

AIP (2015) 34. AIP ISBN

Joshua S. Ullom, 1 Michael L. Oelze, 2 and Jose R. Sanchez Introduction

Implementation of Orthogonal Frequency Coded SAW Devices Using Apodized Reflectors

We are IntechOpen, the world s leading publisher of Open Access books Built by scientists, for scientists. International authors and editors

Pulse Compression. Since each part of the pulse has unique frequency, the returns can be completely separated.

Interaction of Sound and. logarithms. Logarithms continued. Decibels (db) Decibels (db) continued. Interaction of Sound and Media continued

Virtual ultrasound sources

Session: 1E CONTRAST AGENTS II Chair: K. Ferrara University of California-Davis. 1E-1 10:30 a.m.

A STUDY OF AM AND FM SIGNAL RECEPTION OF TIME MODULATED LINEAR ANTENNA ARRAYS

SMALL LESION DETECTION WITH RESOLUTION ENHANCEMENT COMPRESSION: A METHOD OF CODED EXCITATION/PULSE COMPRESSION PAUL MITCHELL LINDEN THESIS

Optimal Control by Transmit Frequency in Tissue Harmonic Imaging

Lesson 02: Sound Wave Production. This lesson contains 24 slides plus 11 multiple-choice questions.

STUDY ON SAW ATTENUATION OF PMMA USING LASER ULTRASONIC

12/26/2017. Alberto Ardon M.D.

Breast Ultrasound QA Phantom Recommended by Japan Association of Breast and Thyroid Sonology

Comprehensive Ultrasound Research Platform

Physics of ultrasound

The Quantitative Study of TOFD influenced by the Frequency Window of Autoregressive Spectral Extrapolation

Non-coherent pulse compression - concept and waveforms Nadav Levanon and Uri Peer Tel Aviv University

Multi-Element Synthetic Transmit Aperture Method in Medical Ultrasound Imaging Ihor Trots, Yuriy Tasinkevych, Andrzej Nowicki and Marcin Lewandowski

Table 1 The wheel-set security system of China high-speed railway

The physics of ultrasound. Dr Graeme Taylor Guy s & St Thomas NHS Trust

Exercise 2: Simulation of ultrasound field using Field II

Evaluation of in vivo liver tissue characterization with spectral RF analysis versus elasticity

Image Quality/Artifacts Frequency (MHz)

A Novel Crack Location Method Based on the Reflection Coefficients of Guided Waves

CHAPTER 6 SIGNAL PROCESSING TECHNIQUES TO IMPROVE PRECISION OF SPECTRAL FIT ALGORITHM

IMPROVING THE DETECTION OF INTERNAL RAIL CRACKS BY USING RADON TRANSFORM OF BSCAN IMAGE

Available online at ScienceDirect. Physics Procedia 70 (2015 )

Simulation of Ultrasonic Testing of Rail Wheel Face using Phased Array and DDF technique

Lesson 06: Pulse-echo Imaging and Display Modes. These lessons contain 26 slides plus 15 multiple-choice questions.

Linear arrays used in ultrasonic evaluation

Reduction of Dispersive Wave Modes in Guided Wave Testing using Split-Spectrum Processing

Implementation of Digital Signal Processing: Some Background on GFSK Modulation

A Passive Suppressing Jamming Method for FMCW SAR Based on Micromotion Modulation

Design on LVDT Displacement Sensor Based on AD598

1831. Fractional derivative method to reduce noise and improve SNR for lamb wave signals

A Reflectometer for Cable Fault Location with Multiple Pulse Reflection Method

Chapter 4. Pulse Echo Imaging. where: d = distance v = velocity t = time

A Modified Synthetic Aperture Focussing Technique Utilising the Spatial Impulse Response of the Ultrasound Transducer

An Energy-Division Multiple Access Scheme

Side-lobe Suppression Methods for Polyphase Codes

Choosing an Ultrasonic Sensor for Ultrasonography

Digital data (a sequence of binary bits) can be transmitted by various pule waveforms.

Designing Non-linear Frequency Modulated Signals For Medical Ultrasound Imaging

Transducer product selector

Phd topic: Multistatic Passive Radar: Geometry Optimization

THE INFLUENCE OF THE TRANSDUCER BANDWIDTH AND DOUBLE PULSE TRANSMISSION ON THE ENCODED IMAGING ULTRASOUND

CCD Automatic Gain Algorithm Design of Noncontact Measurement System Based on High-speed Circuit Breaker

Answer: TGC is needed to amplify echoes from deeper structures so that they appear as bright as similar structures located at more shallow depths.

Session: 2A NEW ULTRASOUND SYSTEMS Chair: H. Ermert University of Bochum 2A-1 10:30 a.m.

Low Power LFM Pulse Compression RADAR with Sidelobe suppression

SIDELOBES REDUCTION USING SIMPLE TWO AND TRI-STAGES NON LINEAR FREQUENCY MODULA- TION (NLFM)

A self-adaptive Contrast Enhancement Method Based on Gradient and Intensity Histogram for Remote Sensing Images

Development of the air-coupled ultrasonic vertical reflection method

Compound quantitative ultrasonic tomography of long bones using wavelets analysis

Lesson 06: Pulse-echo Imaging and Display Modes. This lesson contains 22 slides plus 15 multiple-choice questions.

Planar Leaky-Wave Antennas Based on Microstrip Line and Substrate Integrated Waveguide (SIW)

Design of High-Precision Infrared Multi-Touch Screen Based on the EFM32

Further development of synthetic aperture real-time 3D scanning with a rotating phased array

An Improved Adaptive Median Filter for Image Denoising

Multi Level Temperature Measurement Using a single 90 bend waveguide

Blind Single-Image Super Resolution Reconstruction with Defocus Blur

Keywords: cylindrical near-field acquisition, mechanical and electrical errors, uncertainty, directivity.

White Rose Research Online URL for this paper: Version: Accepted Version

Study on Imaging Algorithm for Stepped-frequency Chirp Train waveform Wang Liang, Shang Chaoxuan, He Qiang, Han Zhuangzhi, Ren Hongwei

Detection of Targets in Noise and Pulse Compression Techniques

Characteristics of point-focus Simultaneous Spatial and temporal Focusing (SSTF) as a two-photon excited fluorescence microscopy

A variable step-size LMS adaptive filtering algorithm for speech denoising in VoIP

Ultrasonic Testing using a unipolar pulse

MULTI-FREQUENCY ULTRASOUND IMAGING: PHANTOM STUDY

Attenuation estimation in biological tissue with ultrasound

Theoretical and Experimental Investigation of Fiber Bragg Gratings With Different Lengths for Ultrasonic Detection

The Physics of Echo. The Physics of Echo. The Physics of Echo Is there pericardial calcification? 9/30/13

Design of Ultrasonic Liquid Level Sensor Based on STM32 with MODBUS Protocol

Design & Analysis the parameters of strain based FBG sensors using Optigrating

Analyzing Pulse Position Modulation Time Hopping UWB in IEEE UWB Channel

Finite Element Analysis and Test of an Ultrasonic Compound Horn

Ultrasound physical principles in today s technology

Attenuation and velocity of ultrasound in solid state materials (transmission)

Global Journal of Engineering Science and Research Management

ACOUSTIC MICRO IMAGING ANALYSIS METHODS FOR 3D PACKAGES

Transcription:

Sensors & Transducers 24 by IFSA Publishing, S. L. http://www.sensorsportal.com Optimization of Axial Resolution in Ultrasound Elastography Zhihong Zhang, Haoling Liu, Congyao Zhang, D. C. Liu School of Computer Science, Sichuan University, Chengdu 665, China Tel.: +28-73-539997889 E-mail: z.zhihong85@gmail.com Received: 3 March 24 /Accepted: 3 June 24 /Published: 3 July 24 Abstract: To improve the axial resolution of ultrasonic elastography, by taking the advantage of code excitation and frequency compounding, multi-frequency with coded excitation for elastography (FCCE) was proposed. FCCE adopts the Chirp signal excitation scheme and strikes a balance in the selection of sub-signal bandwidth, the bandwidth overlap and the number of sub-strain image based on theoretical derivation, so as to further improve the axial resolution of elastic image. On MATLAB, multi-frequency with coded excitation for elastography was implemented and compared with short pulse. Experiments have proved that, compared with the short pulse, the elastographic signal-to-noise ratio (SNRe) and contrast-to-noise ratio (CNRe) were improved significantly. Moreover, probing depth, axial resolution and target detection were improved too. Therefore, the FCCE technology can effectively improve the elastography quality and can be applied to ultrasonic clinical trials. Copyright 24 IFSA Publishing, S. L. Keywords: Ultrasonic elastography, Axial resolution, Coded excitation, Multi-frequency compounding L.. Introduction Elastography, the imaging modality of elastic properties of biological tissue using ultrasound, was firstly presented by A. Hassen in 99 []. In general, a small external mechanical pressure is applied on soft tissue. Axial strain calculation compares the signal before the application of force, or precompression signal, with the post-compression signal. Elastogram is calculated as the gradient of the axial displacements and displayed as a gray scale image where the dark areas correspond to hard tissue while the light areas indicate soft tissue. With ultrasonic elastography, elastic properties are measured quantitatively, which offers more information than qualitative analysis of palpation. In 2, transmit-side frequency compounding elasticity imaging technique (TSFC) was proposed in B. Park [2]. Simulation data and anthropomorphic phantom data demonstrate that TSFC can effectively eliminate amplitude modulation noise and improve elastographic signal-to-noise ratio (SNRe) and elastographic contrast to noise ratio (CNRe) of ultrasound elastic strain imaging. In 25, coed excitation for ultrasound elatography was introduced for the first time in J. W. Goodman [3]. It pointed out that coded excitation can effectively suppress the noise of decorrelation in ultrasound elastography when the system SNR is low. In 22, 5 coded excitation solutions based on Chirp signals were proposed and applied in ultrasound elastography in C. B. Burckhardt [4]. Based on the research on transmit-side multi-frequency compounding (FC) and coded excitation (CE), it can be said that the combination of these two technologies and their applications in elastography are very promising. 24 http://www.sensorsportal.com/html/digest/p_226.htm

2. Method 2.. Theoretical Background From the combination of principles and processes of noise suppression, TSFC can act as the right assistive technology for CE. Based on above theoretical study, transmit-side multi-frequency coded excitation compounding technology (FCCE) is proposed. By carrying out weighted compounding on sub strain images excited by coded signals with different center frequencies and utilizing the decorrelation of strain noise, the strain noise and the noise introduced by coded excitation can be further suppressed. Finally, the image quality is improved with the advantage of increased detecting depth and enhanced strain detection capability. For TSFC technology, the greater the decorrelation of signals is, the further will noise be eliminated after weighted average while useful information is protected. Therefore, for compounding technology, the smaller the cross-correlation of two signals is, the better noise-elimination capability it has. Theoretically, the SNRe of sub strain images after compounding can be expressed as: ( ) N + ( ) ( N ) E X SNRe = = μ Var X ρ, σ () where N is the number of sub strain images. μ and σ are, respectively, the mean and standard deviation of the strain estimates in a region of uniform elasticity. ρ is the normalized cross-correlation coefficient of sub-strain images. Suppose that A and B are two sub strain images with the size of m n obtained by sending two ultrasound with different center frequencies, then ρ can be expressed as: ρ = m n ( A A)( B B) i, j i, j i= j= m n m n 2 2 ( Ai, j A) ( Bi, j B) i= j= i= j=, (2) The following conclusions can be derived from Eqs. and 2: ) When sub strain images are completely decorrelated, i.e. ρ=, the SNRe of elastic strain images after compounding will be improved by a factor of N ; 2) When sub strain images are completely correlated, i.e. ρ=, the SNRe after compounding remains unchanged; 3) When sub strain images are partially correlated, i.e. <ρ<, the SNRe after compounding will be improved by a factor of N ( + ( N ) ρ ). It can be seen that the smaller the normalized cross-correlation coefficient ρ of sub strain images is, the more improvement TSFC will bring to image quality. And if the respective ρ of sub images is of same value, then the more sub images there are, the better SNRe is. However, it can be derived from Eq., if SNRe of compounding result of K+ sub strain images is higher than that of K sub strain images, then the following equation must hold: ρ SNRe SNRe K K ρ K K ( ) ( 2 ) K+ + K+ < 2 + + K K SNReK SNReK, () where ρ N are the cross-correlation coefficients of N sub strain images, SNReN is the average SNRe of N sub strain images. From last formula, ρ K+ is affected mainly by ρ K, K and SNRe / K + SNReK. So there are many factors affecting the imaging results, and the increase of the number of sub strain images cannot guarantee that ρ remain unchanged. Therefore, the number of sub strain images is not the sole determining factor. Cramer-Rao lower bound was proposed and the theoretical minimum variance of displacement 2 distribution σ CRLB was provided [4], as shown in Eq. 4. σ 2 CRLB 2 3, 2 3 2 + 2 2π T( B + 2Bf ) ρ SNRs (4) where T is the duration time of the signal, B is the bandwidth, f is the center frequency of the probe, ρ is the cross-correlation coefficient of two frame signals before and after compression, SNRs is the signal-to-noise of the system. One of CE s characteristics is that the long coded signals with big bandwidth to excite probes can be sent. As can be seen from Eq. 4, theoretically the longer the coded signal sustains and the bigger the bandwidth is, the better the imaging effect is. Due to the beam synthesis and imaging safety, the length of the encoded signal actually requires an appropriate range, while the coded bandwidth is affected by the probe bandwidth. Therefore, the cross-correlation coefficient ρ of bandwidth B and sub-strain image becomes the most important balance point of the combination of CE and TSFC. The introduction of TSFC into CE requires coded signals with different center frequencies to excite probes. At this point, in addition to the bandwidth of the probes, the coded signal bandwidth can be limited by the bandwidth overlaps of multi frame coded signals. This is because that the higher the bandwidth overlaps are, the higher the cross-correlation coefficient ρ of sub strain images is. then TSFC cannot effectively eliminate noise through compounding. If the bandwidth of coded signal was reduced recklessly to make the cross-correlation coefficient ρ smaller, the effect of CE technology will be affected, weakening its ability of increasing detecting depth and improving target detection, and the ability to fight against attenuation. Therefore, tradeoffs need to be made when selecting the sub-signal bandwidth, 24

bandwidth overlap and the number of sub strain images. image with center frequency of f i. If using averaging method, then α i =/N (i=, 2,, N). 2.2. Technical Principles and Processes Coded signals with different center frequencies are used to excite probes in FCCE. Then the sub strain images are obtained by displacement and strain estimation. Finally, compounding strain images are obtained through weighted compounding to improve detecting depth, detecting capability, SNRe and CNRe. Every sub strain image is obtained through coded excitation and has higher SNRe. Smaller displacement estimation signal window can be used to get better resolution. Therefore, the quality starting point of compounding images is higher. Strain images excited by Chirp signal, 3-bit Barker code, 32-bit Suboptimal code and 32-bit Golay code were compared through experiments respectively in J.W. Goodman [3]. Under different esnr and attenuation coefficients, best imaging results and best robustness were obtained by Chirp code and Golay code excitations. While when adopting FCCE technology, it needs transmit multiframe signals during the movement of tissues, but the movement among tissues will weaken the sidelobe suppression capability of Golay code. In conclusion, Chirp linear frequency modulation signal is employed as the coded signal in this paper. The process of FCCE is shown in Fig.. Step : Set coded signals as excitation signal for probes, such as Chirp code and Barker code. And configure the center frequency (f, f 2,, f N ), bandwidth, pulse duration, amplitude suppressing window function of the signal sequences and excitation sequence (center frequency): f, f 2,, f N, f, f 2,, f N.; Step 2: The probe was used to compress tissues and was controlled to send 2N ultrasound signals with N different center frequencies alternatively; Step 3: Carry out pulse compression on echoes after receiving 2N frames of echo sequences; Step 4: Establish N echo buffer queues {q j ( j=,2,,n )}. q j will store the echo signals of center frequency f i ( i=,2,, N ) and echo signals will be classified according to center frequency; Step 5: Utilizing two signal frame in each echo buffer queue {q j ( j=,2,,n )}, sub strain image queue{s i (i=,2,,n)} is obtained through displacement estimation and strain estimation; Step 6: Weighted compounding: Use a weighted compounding method to combine N sub strain images into one strain image: FC( N ) = N i= αix ( fi), N α i= i (5) where N is the number of sub strain images, α i is the weight of i th sub strain image, X( f i ) is the sub strain Fig.. Flow chart of FCCE technology. 3. Experimental Step RF data of each frame were simulated using Field II [5]. The st uniform phantom consists of 3 5 scatterers (fully developed) distributed uniformly throughout mm 3 mm mm volumes with scattering strengths Gaussian distributed. The ideal strain image of the phantom is shown in Fig. 2 (a), while the modular ratio is 3.2. The second uniform phantom consists of 2 5 scatterers (fully developed) distributed uniformly throughout 5 mm 4 mm mm volumes with scattering strengths Gaussian distributed. The ideal strain image of the phantom is shown in Fig. 2 while the modular ratio is.5. Transducer parameters and processing parameters in experiments are shown in Table. 2 3 4 5 6 7 8 9 - width[mm] Fig. 2 (a). Ideal strain diagram software organization simulation motion model. No. Body model spherical hard lesion. 242

2 3 4 appear noticeably in short pulse image in the deeper area. At the depth of 7 mm, lesions are almost submerged by strain noise, and the edge is obfuscated. In FCCE image, by contrast, noise nearby the lesions is less. The lesions edge is clear and the shape is complete, although there are slight calculation errors when the depth is greater than 9 mm. 5-2 - 2 width[mm] Fig. 2 (b). Ideal strain diagram software organization simulation motion model. No. 2 Body model: 4 horizontal stripe soft lesions. Table. Default values for the simulation parameter study. Parameters Um, (V) Is, (ma) RF Sampling frequency fs MHz Pitch of transducer element ω.28 mm Height of transducer element he 5 mm kerf Ke.35 mm Number of elements N 92 Number of active elements Na 64 Transmit apodization A xmt hamming Receive apodization A rec hamming Focus Z 3 mm Assumed speed of sound C 54 m/s Elevation focus Re 2 mm Modular ratio of N. 3.2 Phantom N.2.5 Attenuation coefficient ρ.5 db/cm/mhz Echo signal-to-noise ratio esnr 5 db (a) short pulse SNRe =3.28 CNRe =3.89 (b) FCCE SNRe =6.8 CNRe =8.4 Experiment chooses the first simulation body model (containing hard spherical lesion) and test the strain signal to noise ratio SNRe and contrast to noise ratio CNRe of resulting image. Experiment 2 chooses No. 2 simulation body model for imaging and calculates the strain signal to noise ratio SNRe and CNRe contrast to noise ratio of resulting image. (c) short pulse detail (7 mm- mm) 3. Experiment Results and Discussions 3.. Experiment Fig 3 (a) and Fig. 3 (b) show the experimental resulting image of short pulse and FCCE while the attenuation coefficient is.7 db/cm/mhz and esnr is 5 db. Fig. 3 (c) and Fig. 3 (d) show the ROI (the depth of 7 mm- mm) details of Fig. 3 (a) and Fig. 3 (b). It can be seen that the calculation errors (d) FCCE detail (7 mm- mm) Fig. 3 Attenuation coefficient is.7 db/cm/mhz, experimental resulting image of short pulse and FCCE with details. 243

Fig. 4 shows the longitudinal strain values in the middle area of the short pulse and FCCE strain images in Fig. 3. In the healthy background tissue when the depth is below 65 mm, the disturbance of FCCE strain value is less severe than that of short pulse, however, the difference is insignificant. When the depth is between 65 mm and 75 mm in internal hard lesion, the disturbance of FCCE strain value is less violent than that of short pulse, which shows the disturbance of FCCE strain value is relatively even in lesion parts. When the depth is above 75 mm, the disturbance of short pulse strain value is severe, which represents that the depth is beyond the effective detection range of short pulses. Besides, although the disturbance of FCCE goes up as depth grows, the variation is relatively stable which indicates that it is still within the detection depth. strain.4.3.2. short pulse TFCCE 2 3 4 5 6 7 8 9 Fig. 4. Attenuation coefficient is.7 db/cm/mhz, longitudinal strain values in the middle location of the short pulse and FCCE. From the above analysis, under the condition of different attenuation coefficient, FCCE can improve both the image s SNRe and CNRe. Especially in high tissue attenuation, short pulse is unable to guarantee the basic quality of the image, while FCCE can help to get a strain image with medium quality. 3.2. Experiment 2 The contrast strain images in this experiment are mainly from the following two types of simulation. ) Regular short pulse (two cycles Gaussian modulation cosine signal) with center frequency of 5 MHZ. 2) The results of FCCE process. In Fig. 5 (a) Fig. 5 (d), attenuation coefficient is.4 db/cm/mhz, and esnr is 3 db. Figs.5 (a), 5 (b) and Fig. 5 (c) respectively show the strain image of short pulse process when the displacement of the estimated signal window length is 2, 3, 4 pixels. Fig. 5 (d) shows the strain image of FCCE process when the displacement of the estimated signal window length is 2 pixels. The calculated of ROI in the red box in Fig. 5 are 4.83, 5.86, 8.68 and 8.7. Fig. 5 (a) and Fig. 5 (d) show that the longitudinal resolution of strain image is close when the displacement of the estimated signal window length is 2 pixels. Fig. 5 (c) and Fig. 5(d) show that the SNRe obtained by the short pulse whose window length is 4 pixels is close to the one obtained by FCCE whose window length is 2 pixels. (a) short pulse, W=2 pt; SNRe =4.83 (b) short pulse, W=3 pt; SNRe=5.86 (c) short pulse, W=4 pt; SNRe =8.68 (d) FCCE, W=2 pt; SNRe =8.7 Fig. 5. The resulting image of short pulse and FCCE with different displacement estimation window. 244

Fig. 6 shows the longitudinal strain values in the middle location of the strain images of short pulse whose window length is 4 pixels and FCCE whose window length is 2 pixels. The horizontal arrow marks the interval width among soft white lesions. This approach can visually help us to analyze the effects of short pulse and FCCE on the longitudinal resolution. The interval widths among lesions of short pulse image are all wider than those of FCCE image, indicating that the longitudinal resolution of FCCE is better, which also can be seen in Fig. 5(a) and Fig. 5(b). n o rm a liz e d st ra in.8.6.4.2 short pulse,window length=4 TFCCE,window length=2 5 5 2 25 3 35 4 45 5 Fig. 6. Longitudinal strain values in the middle location of the strain images of short pulse with window length of 4 and FCCE with window length of 2. Based on the analysis above, we can get the following conclusions: FCCE can use a shorter displacement estimation of signal window than short pulse in the case of the same noise level, which indicates that FCCE has a potential for increasing longitudinal resolution of the strain image. 5. Conclusions The coded excitation and frequency compounding in ultrasound elastography was discussed. The academic background and the advantage of coded excitation and Transmit-side Frequency Compounding were introduced. An ultrasound elasticity imaging improvement technique based on coded excitation and Transmit-side Frequency Compounding technology naming Transmit side multi-frequency Compounding with Coded Excitation for Elastography (FCCE) was proposed. Re-SN decreases with increasing attenuation coefficient. SNRe and CNRe of FCCE increase by 85 % and 373 % than those of short pulse when the attenuation coefficient is.7 db/(cm MHz). The lesion at 7 mm is almost overwhelmed by the strain noise and its edge is not clear, while in the results of FCCE, the lesion is complete and its edge is clear although there is slight miscalculation at the depth of 9 mm. The impact of FCCE on resolution has also been analyzed. Probing depth, axial resolution and target detection were improved too. Based on experimental data analysis, we can come to the conclusion that the use of FCCE can improve the quality of ultrasound elastography. Reference []. A. Hassen, Predicting percentage of intramuscular fat using two types of real-time ultrasound equipment, Journal of Animal Science, Vol. 79, Issue, 2, pp. -8. [2]. B. Park, Measuring intramuscular fat in beef with ultrasonic frequency analysis, Journal of Animal Science, Vol. 72, Issue 3, 994, pp. 7-25. [3]. J. W. Goodman, Some fundamental properties of speckle, Journal of the Optical Society of America, Vol. 66, No. 6, 976, pp. 45-5. [4]. C. B. Burckhardt, Speckle in ultrasound B-Mode scans, IEEE Transactions on Sonics and Ultrasonics, Vol. 25, Issue 8, 978, pp. -6. 24 Copyright, International Frequency Sensor Association (IFSA) Publishing, S. L. All rights reserved. (http://www.sensorsportal.com) 245