A COMPARATIVE STUDY ON THE PERFORMANCE OF DIGITAL DETECTOR SYSTEMS FOR HIGH ENERGY APPLICATIONS

Similar documents
Film Replacement in Radiographic Weld Inspection The New ISO Standard

Digital Radiology with Photon Counting Detectors

Performance evaluation of a photon counting detector for high energy NDT applications

Minimum Requirements for Digital Radiography Equipment and Measurement Procedures by Different Industries and Standard Organizations

Computed Radiography

ISO INTERNATIONAL STANDARD. Non-destructive testing of welds Radiographic testing Part 2: X- and gamma-ray techniques with digital detectors

APPLICATION OF THE DIGITAL RADIOGRAPHY IN WELD INSPECTION OF GAS AND OIL PIPELINES

» There are many possible techniques for making good radiographs. » Following a defined procedure will simplify technique

Moving from film to digital: A study of digital x-ray benefits, challenges and best practices

Essential Parameters and Conditions for Optimum Image Quality in Digital Radiology

Essential Parameters for the Visibility of IQIs and Small Indications in Digital Radiography

Photon Counting and Energy Discriminating X-Ray Detectors - Benefits and Applications

Radiographic testing: Increased detection sensitivity using optimum source to object distance

A Modified Detectability Criterion for Conventional Radiography Simulation

PRACTICAL CONSIDERATIONS AND EFFECTS OF METALLIC SCREEN FLUORESCENCE AND BACKSCATTER CONTROL IN GAMMA COMPUTED RADIOGRAPHY

Minimum Requirements for Digital Radiography Equipment and Measurement Procedures by Different Industries and Standard Organizations

The HOIS recommended practice for in-service computed radiography of pipes

ISO INTERNATIONAL STANDARD. Non-destructive testing of welds Radiographic testing Part 1: X- and gamma-ray techniques with film

Digital Radiography : Flat Panel

Strategies for Film Replacement in Radiography - a comparative study -

DOSE REDUCTION BY USE OF DIGITAL X-RAY DETECTORS. Bernhard REDMER

PERFORMANCE CHARACTERIZATION OF AMORPHOUS SILICON DIGITAL DETECTOR ARRAYS FOR GAMMA RADIOGRAPHY

The HOIS recommended practice for in-service computed radiography of pipes. Dr Stephen F Burch, WCNDT 2012, Durban, SA. April 2012

ISO INTERNATIONAL STANDARD

ON THE WAY TO DIGITAL RADIOGRAPHY

Fig.2: Scanner VistaScan for image plates

Migrating from traditional to Digital Radiography in Aerospace

Advanced Digital Radiography for Field NDT

NON-DESTRUCTIVE EVALUATION UTILIZING IMAGING PLATES FOR FIELD RADIOGRAPHY APPLICATIONS

Equivalent Penetrameter Sensitivity (EPS) for Performance Evaluation of Computed Radiography Systems Muzibur Khan * and Mike Brothers

Digital Detector Array Image Quality for Various GOS Scintillators

Tomographic 3D-Radiometry for the Visualisation and Measurement of the Defects of Girth Seams

Digital Radiography for the Inspection of Small Defects

Digital Radiographic Inspection replacing traditional RT and 3D RT Development

Gamex Lite 3.0 Program description and operating manual

Pre Conference Workshop on Advanced Digital imaging

Dürr NDT GmbH & Co. KG Höpfigheimer Straße 22 D Bietigheim-Bissingen Germany. Contract No. BAM ZBA Dürr

NDT Supply.com 7952 Nieman Road Lenexa, KS USA

Preliminary Modulation Transfer Function Study on Amorphous Silicon Flat Panel System for Industrial Digital Radiography

Dürr NDT GmbH & Co. KG Höpfigheimer Straße 22 D Bietigheim-Bissingen Germany. Contract No. BAM ZBA Dürr

This document is a preview generated by EVS

Research Article Digital Radiography Using Digital Detector Arrays Fulfills Critical Applications for Offshore Pipelines

ScienceDirect. Evaluation of Pipe Wall Thickness Based on Contrast Measurement using Computed Radiography (CR)

THE CR SPECIALISTS HD-CR CR 35 NDT

Experiences of users in Digital Radiography

Real Time Linear Array Imaging. Brian Caccamise

Determination of Optimum X-Ray Tube Output Parameters kv and ma for Digital Radiography Testing of Welded Tubes

Advancements In Digital Radiography: CR, DR, and DICONDE

17th World Conference on Nondestructive Testing, Oct 2008, Shanghai, China

The field deployable digital X-Ray solution for weld inspection in O&G

Part 3: Image quality classes

X-ray backscattering: Variable irradiation geometry facilitates new insights

SMALL-SIZE BETATRON FOR THE ENERGY 2.5 MEV FOR FIELD INSPECTION

IQI-Sensitivity and Applications of Flat Panel Detectors and X-Ray Image Intensifiers A Comparison

ISO INTERNATIONAL STANDARD. Non-destructive testing of welds Radiographic testing of fusionwelded

Flash-Radiography Instead of Traditional Radiography with Intermediate Carriers of Information

High Energy Digital Radiography & 3D-CT for Industrial Systems

Mobile digital radiography system for nondestructive testing of large diameter pipelines

Amorphous Selenium Direct Radiography for Industrial Imaging

In-service inspections for primary coolant circuit components of light water reactors Part 6: Radiographic testing

Hardware for High Energy Applications 30 October 2009

First Validation of CIVA RT Module with a Linear Accelerator in a Nuclear Context

Gamex CR 2.0 Program description and operating manual

Modelling Computed Radiography Detectors with a Cascaded Linear System Model

5th Pan American Conference for NDT 2-6 October 2011, Cancun, Mexico

FOCUS. Quick Industrial X-ray Testing Without Intermediate Data Carriers of Information. by V.A. Troitskiy

I. PERFORMANCE OF X-RAY PRODUCTION COMPONENTS FLUOROSCOPIC ACCEPTANCE TESTING: TEST PROCEDURES & PERFORMANCE CRITERIA

10/3/2012. Study Harder

10/26/2015. Study Harder

NEW POSSIBILITIES OF RADIATION CONTROL OF QUALITY OF WELDED JOINTS

Radiation Protection and On-Site Inspection of the Elevated Subway Construction with an Ir-192 Gamma-Source

Digital Profile Radiography Practical uses and limitations

A True Innovation in Non-Destructive Testing System FUJIFILM COMPUTED RADIOGRAPHY. Series 4 CR

A TUTORIAL FOR PERFORMING A RADIOGRAPHIC EXAMINATION

Examination of Pipe Welds by Image Plate Based Computed Radiography System

HPX1-Plus. For Non-Destructive Testing. THE BENCHMARK IN COMPUTED RADIOGRAPHY.

COMPUTED RADIOGRAPHY THE STATE OF THE ART

UNDERSTANDING THE CURRENT CAPABILITIES AND LIMITATIONS OF DIGITAL INDUSTRIAL RADIOGRAPHY

digital radiography r High-energy DT Applications lior Pick, Ron Pincu and Ofra Kleinberger over film radiog- p P This advanced

Recommended Training Curriculum For Digital Radiography Personnel (Level II)

Dose Reduction and Image Preservation After the Introduction of a 0.1 mm Cu Filter into the LODOX Statscan unit above 110 kvp

Ray Detection Digital Image Quality and Influential Factors

MC SIMULATION OF SCATTER INTENSITIES IN A CONE-BEAM CT SYSTEM EMPLOYING A 450 kv X-RAY TUBE

Digital Radiography Systems Techniques and Performance Evaluation for Space Applications

Digital Profile Radiography Practical uses and limitations

Ensuring Shielding adequacy in Lead shielded spent fuel transportation casks using gamma scanning

Radiographic sensitivity improved by optimized high resolution X -ray detector design.

COMPUTED RADIOGRAPHY CHAPTER 4 EFFECTIVE USE OF CR

X-ray Imaging. PHYS Lecture. Carlos Vinhais. Departamento de Física Instituto Superior de Engenharia do Porto

SINCE2011 Singapore International NDT Conference & Exhibition, 3-4 November 2011

Modeling computed radiography with imaging plates

HDRTR (High Definition Real Time Radiography)

international Standard 1106/l

HIGH RESOLUTION COMPUTERIZED TOMOGRAPHY SYSTEM USING AN IMAGING PLATE

Dual and Multi-energy Radiography for CFRP Composites Inspection

This document is a preview generated by EVS

DIGITAL RADIOGRAPHY. Digital radiography is a film-less technology used to record radiographic images.

Investigation of the film Kodak MX 125 in Agfa NDT S eco processing

Photomultiplier Tube

Introduction To NDT. BY: Omid HEIDARY

Transcription:

11th European Conference on Non-Destructive Testing (ECNDT 2014), October 6-10, 2014, Prague, Czech Republic More Info at Open Access Database www.ndt.net/?id=16394 A COMPARATIVE STUDY ON THE PERFORMANCE OF DIGITAL DETECTOR SYSTEMS FOR HIGH ENERGY APPLICATIONS A. Schumm 1, E. Martin 2, M. Beaumont 2, U. Ewert 3, S. Kolkoori 3, U. Zscherpel 3, N. Wrobel 3, B. Redmer 3 1 EDF R&D, 1 avenue du général de Gaulle, F-92141 Clamart Phone: +33 1 43073953, Fax: +33 1 47652713; e-mail: andreas.schumm@edf.fr 2 EDF CEIDRE, 2 rue ampère, F-93206 Saint Dénis 3 BAM Berlin, Unter den Eichen 87, D-12205 Berlin Abstract For cast components reaching or exceeding total material thicknesses of 150mm, high energy sources such as linear accelerators or Betatrons are required in order to obtain reasonable exposure times. In this study, the performance of digital detector systems, involving imaging plates (IP) and digital detector arrays (DDA), was evaluated with respect to the testing class B requirements as formulated in the standard EN ISO 17636-2. As a reference, traditional radiographic film and a Cobalt-60 source was used. With film exposures, testing class B was achieved with Co-60 and Betatron (7.5 MV) at longer exposure times. The preliminary results show that the testing class B was not obtained with the examined digital detector arrays (DDA) and the high resolution imaging plates (IP), even at 40, 60 and 80 minutes exposure time with a 7.5 MV Betatron. Class A was achieved using high resolution imaging plates with optimized metal filters between object and IPs and a high resolution DDA with intermediate Cu filters. Class A was also achieved applying a DDA with lower basic spatial resolution than required by Table B.13 of EN ISO 17636-2, but using the compensation principle as described in this standard. The next generation of digital detector arrays might potentially be able to obtain class B performance with the expected spatial resolution and sensitivity improvements. Keywords: Radiography, Betatron, digital detector array (DDA), casting; image quality, standards 1. Introduction Radiographic inspection of critical components with high material thickness (150 mm and more) requires high energy sources and rather long exposure times. Digital radiography with digital detector arrays and imaging plates hold the promise of a notable decrease in exposure time and enhanced image quality because of their increased sensitivity with respect to traditional industrial X-ray films. At the same time, the significantly larger dynamic range should improve robustness in terms of the useable incident radiation dose range. The objective of this study was to investigate the performance of digital detectors representative of the state-of-the-art in 2013 with realistic double-wall inspection configurations using high energy radiation sources. The chosen configuration is representative of the inspection of an elbow pipe with an outer diameter of 907 mm. The elbow pipe is represented by two planar mock-ups with 70 mm and 81 mm wall thickness. The digital detectors used in this study are an imaging plate (IP) with a basic spatial resolution of 40 µm, scanned with 50 µm pixel size, and two digital detector arrays with respective pixel sizes of 200 µm and 143 µm. The X-ray source used was a Betatron with a maximum energy of 7.5 MeV and a dose rate of 5.3 R/min (53 mgy/min) at a distance of 1meter. The reference image was obtained on a traditional C3 X-ray film system, in conjunction with a Co 60 source. Figure 1 shows the installation during a Betatron exposure with a digital detector array (DDA). 2. EN ISO 17636-2 standard requirements

For this application, the requirements were to achieve the image quality according to testing class B as described in the new EN ISO 17636-2 standard for weld testing using digital detectors [1]. According to this standard, which replaced the older EN1435 standard in May 2013 for digital radiography applications, a wire-type image quality indicator (IQI) with element number W8 (0.63 mm wire diameter) shall be visible (IQI positioned on the detector side) for the given penetrated thickness. The associated basic spatial resolution shall be at least 0.13mm (for a DDA or IP), which requires duplex wire IQI D9. The EN ISO 17636-2 standard describes the compensation principle (II), which allows to compensate the underperformance on basic spatial resolution (missing duplex wire pairs) by an increase of the contrast sensitivity (better single wire IQI detection), but only up to a certain point. Densimet shielding X-ray source Betatron 7.5 MeV Matrix detector (area: 40.96 x 40.96 cm 2 ) Figure 1. Preparation of experimental setup with DDA and Betatron 3. Experimental Results and Discussion The Table 1 summarizes the experimental conditions of the various exposures and the obtained results using industrial X-ray films and imaging plates. The column wire IQI indicates the just visual wire number value in the IQI set W6 FE-EN (EN ISO 19232-1). The main parameters in Table 1 are the exposure time (t exp ), source-to-detector distance (SDD), geometrical unshaprness Ug, optical density o.d. of NDT films or signal-to-noise ratio (SNR) for digital detector images, as well as the IQI value seen. For an image quality according to the standard EN ISO 17636 requires a wire IQI visibility of W6 for testing class A and W8 for testing class B. The results show that all exposures with NDT films satisfy the requirements of class B, as expected using longer exposure times. Film packages of two films were exposed synchronously to an optical density o.d. > 1.3. The films were evaluated under double film viewing condition. The usage of single films and single film viewing would require the double exposure time as given in Table 1. For exposures with imaging plates, the testing class A was achieved only with optimum detector front screen packages and back-scatter filters. For a full appreciation of the image quality, the normalised signal-to-noise ratio (SNR N ) is necessary, for which the basic spatial resolution must be known [2]. It is typically obtained using a duplex wire IQI.

Unfortunately the duplex wire IQI of EN ISO 19232-5 was not visible in IP radiographs of the high thickness materials. Therefore, the unormalized SNR values are given in Table 1. The Co-60 source used for these experiments had an activity of 30 Ci, with a focal spot size of 4 x 4.2 mm 2. The dose rate measured at a distance of 1 meter was about 393 mgy/h. The Betatron was used with its maximum energy of 7.5 MeV, with a focal spot measured to be 1.5 x 3mm 2 (own measurements, according to the data sheet, the focal spot should be smaller). The dose rate observed at a distance of 1 meter was 53 mgy/min (3180 mgy/h). The exposure time with the Betatron is limited to 40 minutes, since the device requires a cool-down period of 15 minutes after 40 minutes of use. Table 1. Results with X-ray films and imaging plates Source Detector t exp SDD Co60 Beta / U g Film C3 2 9h 1000 mm Film C3 2 / 0.4 mm 45min 1300 mm Beta Film C3 2 30min 1300 mm Beta IP HR 60min 1300 mm Beta IP HR 60min 1300mm o.d. 1 / SNR Wire IQI 3.25 W9 0.5 mm 3.20 W8 0.63mm 2.2 W7 0.80 mm 350 W5 1.25 mm 375 W6 1 optical density (o.d.) 2 double film exposure and double film viewing 1 mm Comments Vacupac 2 x 0.027 µm Pb, 0.5 mm Fe filter intermediate Lower unsharpness than Betatron radiographs Vacupac 2 x 0.027 µm Pb, 0.5 mm Fe filter intermediate 12 times shorter exposure time than Co-60 exposure W8 barely visible IP HR 50µm 35x43cm², grey level: 8500, filter: 2 mm Pb + 1mm Sn + 1mm Cu IP HR 50µm 35x43cm², grey level: 8500, filter: 1 mm Pb, + 1mm Cu + 1mm Sn The IP UR1 detector is an IP with low intrinsic unsharpness and high basic spatial resolution at low energies (down to 40 µm), but low sensitivity. The experiments clearly show that the best results were obtained with C3 film and a Co-60 source using 9 h exposure time with double film exposure. Replacing the Co-60 source with a Betatron for the film experiment allows to reduce the exposure time by a factor of approx. 12 (normalized to the SDD: factor 14), which is larger than expected from the exposure dose comparison (factor 8). On the other hand, the combination Betatron and HR imaging plates did not allow a significant decrease of exposure time with respect to the NDT film systems. The two experiments with imaging plates further show that the right choice of the filter allows to gain one IQI wire value. For the experiments with imaging plates, interpretation of the results was done with and without image processing (high pass filtering), but without a notable difference.

The experimental setup for inspecting thick-walled cast component using the highenergy Betatron and digital detector arrays is shown in Figure 2. The geometric set up (SDD and SOD) was the same as for film and IP inspection as shown in Figure 1 and Table 1. The signal-to-noise ratio in the image has been enhanced by performing optimum detector calibration (i.e. effective bad pixel correction and compensation of structure noise of the detector elements). Additionally, an optimum shielding was employed in order to reduce the influence of scattered radiation from surroundings on image quality (see Fig. 2). The results achieved using two different digital detector arrays at various exposure times are summarized in Table 2. For the experiments with the DDAs it is necessary to verify if the spatial resolution is compatible with the requirements of the EN ISO 17636-2 standard since the detector pixel size is significantly larger than the basic spatial resolution of film and IP radiographs. A series of dedicated experiments were conducted with duplex wire IQIs to obtain this quantitative information. In order to evaluate the basic spatial resolution of digital images of the thick-walled component, a CERL C duplex plate IQI was used as well as a duplex wire IQI according to EN ISO 19232-5. Detector A has a pixel size of 200 µm and a thick Gadox-Scintillator PI-100 and allowed to distinguish duplex wire D4 only. Applying the compensation principle, as permitted by the standard, inspection class A was achieved with an exposure time of 40 minutes, but the testing class B was not achievable with this detector due to its insufficient basic spatial resolution. It should be mentioned, that the required wire resolution was proven at 40 minutes exposure time corresponding to Table B.9. The image quality for class B testing was confirmed in relation to EN 1435, but not corresponding to EN ISO 17636-2. Detector B (pixel size 143 µm, basic spatial resolution 160 µm (Duplex wire D7 achieved) and thin Gadox-Szintillator DRZ Standard) satisfies the requirements of testing class A without resorting to the compensation principle at 20 minutes exposure time, and almost achieved class B (still 2 more wires to be detected) after increasing the exposure time by a factor of 4. A 4 mm copper detector front filter was used to increase the contrast-to-noise ratio (CNR) in the digital image (result of the reduction of scattered radiation detected by DDA). (a) Detector X-ray Betatron 7.5 MeV Backwall (81 mm thick) Frontwall (70 mm thick) (b) (c)

Figure 2. (a) Experimental setup showing front wall and back wall of the cast component inspected using 7.5 MeV X-ray Betatron and a digital detector array (pixel size 200µm). Measured digital X-ray image with (b) 20 min, (c) 40 min exposure times. Table 2. Results with high-resolution digital detector arrays (DDAs) Matrix Detector Type (exposure time) A (20 min) SR b (mm) SNR N Wire Type IQI 0.4 120 W7 0.8 mm CERL C Duplex Plate IQI D4 ISO 17636-2 Required class A Wire /duplex IQI A (40 min) 0.4 145 W9 0.5 mm D4 B (20 min) 0.2 144 W6 1 mm D7 B (40 min) 0.2 175 W6 1 mm D7 B (80 min) 0.2 275 W7 0.8 mm D7 SR b : Basic spatial resolution of image 4. Conclusions The results certainly show that industrial NDT films remain ahead of digital detectors in terms of its image quality for inspections at high penetrated steel thicknesses. The exposure times of NDT films in double film exposure and double film viewing technique was comparable to the exposure times used for the digital detectors. Inspection class A was obtained with imaging plates and digital detector arrays using a Betatron as high energy X-ray source (7.5 MeV). For these applications, a serious decrease in exposure time was observed in comparison to Co-60 exposures. Testing class A was achieved with imaging plates only if suitable packages of metal filters were used in front of the IP cassette and with good back scatter filtering.

The advantage of NDT film image quality as compared to imaging plates and DDAs may be explained by the lower sensitivity for scattered radiation of the radiographic film. This improves the contrast-to-noise ratio (CNR) of the film images in comparison with the CNR of the radiographs taken with imaging plates or DDAs. Especially for DDAs the filters in front and behind the detection layer need to be improved, whereas for imaging plates suitable filter combinations are given in EN ISO 17636-2. It should be possible to obtain testing class B with a DDA with improved basic spatial resolution and specialized internal screens as e.g. thick crystalline CsI screens. Currently, direct converting detectors with 100 µm pixel size and 750 2000 µm detection layer of CdTe are available, which are up to now used with success for testing class B up to 300 kev. It will require more internal shielding of the electronics to apply them at 7.5 MeV to improve the results shown here. References 1. Non destructive testing of welds radiographic testing part 2: X- and gamma-ray techniques with digital detector, EN ISO 17636-2, CEN/TC121/SC5, 2013 2. U Ewert, U Zscherpel, K Heyne, M Jechow, K Bavendik, New compensation principles for enhanced image quality in industrial radiology with digital detector arrays, Materials Evaluation, Vol 68, pp 163-168, 2010.