Design on Electrocardiosignal Detection Sensor

Similar documents
Differential Amplifier : input. resistance. Differential amplifiers are widely used in engineering instrumentation

Lecture 4 Biopotential Amplifiers

STM32 microcontroller core ECG acquisition Conditioning System. LIU Jia-ming, LI Zhi

Design of Accelerometer Pre-regulation Circuit and Performance Analysis of the Key Components

EE301 ELECTRONIC CIRCUITS

Instrumentation amplifier

Homework Assignment 03

Special-Purpose Operational Amplifier Circuits

AD8232 EVALUATION BOARD DOCUMENTATION

Chapter 4 4. Optoelectronic Acquisition System Design

ELC224 Final Review (12/10/2009) Name:

Biomedical Sensor Systems Laboratory. Institute for Neural Engineering Graz University of Technology

Dimensions in inches (mm) .268 (6.81).255 (6.48) .390 (9.91).379 (9.63) .045 (1.14).030 (.76) 4 Typ. Figure 1. Typical application circuit.

Single Supply, Rail to Rail Low Power FET-Input Op Amp AD820

Operational Amplifiers

Low Cost, General Purpose High Speed JFET Amplifier AD825

Lesson number one. Operational Amplifier Basics

BIOMEDICAL INSTRUMENTATION PROBLEM SHEET 1

Dual operational amplifier

Active Filter Design Techniques

Analog Filters D R. T A R E K T U T U N J I P H I L A D E L P H I A U N I V E R S I T Y, J O R D A N

Linear IC s and applications

Lecture Notes Unit-III

High Common-Mode Rejection. Differential Line Receiver SSM2141 REV. B FUNCTIONAL BLOCK DIAGRAM FEATURES. High Common-Mode Rejection

GOVERNMENT OF KARNATAKA KARNATAKA STATE PRE-UNIVERSITY EDUCATION EXAMINATION BOARD II YEAR PUC EXAMINATION JULY-2012 SCHEME OF VALUATION

Circuit Design and Implementation of Micro-Displacement Measurement System of Laser Self-Mixing Interference

Micropower, Single-Supply, Rail-to-Rail, Precision Instrumentation Amplifiers MAX4194 MAX4197

CHAPTER 3. Instrumentation Amplifier (IA) Background. 3.1 Introduction. 3.2 Instrumentation Amplifier Architecture and Configurations

CA-550 Series / CA-650 Series

HUMAN DETECTION AND RESCUE USING BIO POTENTIAL SIGNALS

A 1MHz-64MHz Active RC TI-LPF with Variable Gain for SDR Receiver in 65-nm CMOS

Dimensions in inches (mm) .021 (0.527).035 (0.889) .016 (.406).020 (.508 ) .280 (7.112).330 (8.382) Figure 1. Typical application circuit.

Active Filters - Revisited

TRANSDUCER INTERFACE APPLICATIONS

Constant Current Control for DC-DC Converters

Preliminary simulation study of the front-end electronics for the central detector PMTs

Precision INSTRUMENTATION AMPLIFIER

OPERATIONAL AMPLIFIER PREPARED BY, PROF. CHIRAG H. RAVAL ASSISTANT PROFESSOR NIRMA UNIVRSITY

Design on LVDT Displacement Sensor Based on AD598

Audio Applications for Op-Amps, Part III By Bruce Carter Advanced Analog Products, Op Amp Applications Texas Instruments Incorporated

APPLICATION NOTE 6206 SIMPLE, EFFECTIVE METHOD AND CIRCUIT TO MEASURE VERY-LOW 1/F VOLTAGE REFERENCE NOISE (< 1ΜV P-P, 0.

Experiment 1: Amplifier Characterization Spring 2019

Operational Amplifier BME 360 Lecture Notes Ying Sun

IFB270 Advanced Electronic Circuits

Low Pass Filter Introduction

Single Supply, Rail to Rail Low Power FET-Input Op Amp AD820

Introduction (cont )

Kanchan S. Shrikhande. Department of Instrumentation Engineering, Vivekanand Education Society s Institute of.

Chapter 15: Active Filters

LM4562 Dual High Performance, High Fidelity Audio Operational Amplifier

LM4562 Dual High Performance, High Fidelity Audio Operational Amplifier

Florida Atlantic University Biomedical Signal Processing Lab Experiment 2 Signal Transduction: Building an analog Electrocardiogram (ECG)

Research and design of PFC control based on DSP

DESIGN OF OTA-C FILTER FOR BIOMEDICAL APPLICATIONS

Model 176 and 178 DC Amplifiers

Analog Electronics. Lecture. Op-amp Circuits and Active Filters. Muhammad Amir Yousaf

ECE-342 Test 1: Sep 27, :00-8:00, Closed Book. Name : SOLUTION

Shock sensor PKGS series Application manual of peripheral circuit Feb. 3, 2003 Example circuit for charge sensitivity type shock sensor.

LME49710 High Performance, High Fidelity Audio Operational Amplifier

Intruder Alarm Name Mohamed Alsubaie MMU ID Supervisor Pr. Nicholas Bowring Subject Electronic Engineering Unit code 64ET3516

Precision, Low-Power and Low-Noise Op Amp with RRIO

Models 900CT & 900BT. Tunable Active Single Channel Certified Filter Instrument

SALLEN-KEY LOW-PASS FILTER DESIGN PROGRAM

Introduction to Analog Interfacing. ECE/CS 5780/6780: Embedded System Design. Various Op Amps. Ideal Op Amps

Electrocardiogram (ECG)

High Accuracy INSTRUMENTATION AMPLIFIER

Research and Design of Envelope Tracking Amplifier for WLAN g

Simulation of small signal resonant amplifier based on Multisim Dan Ren

16 V, 1 MHz, CMOS Rail-to-Rail Input/Output Operational Amplifier ADA4665-2

GOVERNMENT OF KARNATAKA KARNATAKA STATE PRE-UNIVERSITY EDUCATION EXAMINATION BOARD II YEAR PUC EXAMINATION MARCH-2012 SCHEME OF VALUATION

OPERATING INSTRUCTIONS AND SYSTEM DESCRIPTION FOR THE EXT-02 B EXTRACELLULAR AMPLIFIER

A.C. FILTER NETWORKS. Learning Objectives

ECEN Network Analysis Section 3. Laboratory Manual

Bioelectric Signal Analog Front-End Module Electrocardiograph

Chapter 2. The Fundamentals of Electronics: A Review

LINEAR IC APPLICATIONS

Module 2. Measurement Systems. Version 2 EE IIT, Kharagpur 1

High Speed FET-Input INSTRUMENTATION AMPLIFIER

Dual, Ultralow Distortion, Ultralow Noise Op Amp AD8599

Internally Trimmed Integrated Circuit Multiplier AD532

LOW NOISE AMPLIFIER SA SERIES

Assist Lecturer: Marwa Maki. Active Filters

EKT 314 ELECTRONIC INSTRUMENTATION

SENSOR AND MEASUREMENT EXPERIMENTS

Description. Output Stage. 5k (10k) - + 5k (10k)

Implementation of wireless ECG measurement system in ubiquitous health-care environment

Low Cost Instrumentation Amplifier AD622

Vocal fader IC BA3837 / BA3837F / BA3838F. Multimedia ICs

MCP6031/2/3/ µa, High Precision Op Amps. Features. Description. Applications. Design Aids. Package Types. Typical Application

High Common-Mode Voltage Difference Amplifier AD629

About the Tutorial. Audience. Prerequisites. Copyright & Disclaimer. Linear Integrated Circuits Applications

Four-Channel Sample-and-Hold Amplifier AD684

Pankaj Naik Electronic and Instrumentation Deptt. SGSITS, Indore, India. Priyanka Sharma Electronic and. SGSITS, Indore, India

FET-Input, Low Power INSTRUMENTATION AMPLIFIER

High Speed, Precision Sample-and-Hold Amplifier AD585

FYS3240 PC-based instrumentation and microcontrollers. Signal sampling. Spring 2015 Lecture #5

Laboratory 6. Lab 6. Operational Amplifier Circuits. Required Components: op amp 2 1k resistor 4 10k resistors 1 100k resistor 1 0.

EXPERIMENT 1: Characteristics of Passive and Active Filters

Fast-Settling FET-Input INSTRUMENTATION AMPLIFIER

150MHz phase-locked loop

Transcription:

Sensors & Transducers 203 by IFSA http://www.sensorsportal.com Design on Electrocardiosignal Detection Sensor Hao ZHANG School of Mathematics and Computer Science, Tongling University, 24406, China E-mail: ahzhanghao@yeah.net eceived: 8 August 203 /Accepted: 8 September 203 /Published: 3 October 203 Abstract: The key of using a sensor to detect electrocardiosignal (ECG) is on extracting a weak signal from strong noise background and performing the high-gain amplification. The ECG characteristics and circuit model were analyzed in this study. The design included the detection sensor of four parts-preamplifier, high-pass and low-pass filter circuits, 50 Hz notch filter and main amplifier, and the sensor has a high common-mode rejection ratio (CM) and uses a special chip UAF42 to filter power frequency interference. It has been successfully applied to the ECG monitoring system to achieve good results. Copyright 203 IFSA. Keywords: ECG, Driving circuit of right leg, Filter.. Introduction Tissues and body fluids around heart can conduct electricity. The sum of action potential changes of numerous myocardial cells can be conducted and reflected to the body surface, and the bioelectrical signal generated by myocardial activation during this heart activity is ECG. A large number of domestic and foreign experimental data and literatures suggest that the ECG amplitude of healthy human is about 0~4000 µv, and the frequency is 0.05~00 Hz. The 90 % of spectrum energy is concentrated on 0.25~35 Hz []. ECG amplitude is small, and the signal to noise ratio (SN) is low; it is vulnerable to electromagnetic radiation interference in the surrounding environment and internal electronic noise interference of instrumentation, so how to extract the ECG from strong noise background and perform high-gain amplification is a difficult problem to be solved. A sensor with a high CM was designed in this study, and the detection and acquisition of ECG are effectively achieved. 2. ECG Signal Characteristics and Circuit Model The measuring position on the skin surface determines that the ECG has the following characteristics when a disposable ECG electrode with JK- type of electrode model was used for an acquisition: Metal electrode is in contact with the electrolyte of skin, and a potential difference will be produced between electrode and skin. It can be overcome by the common-mode rejection characteristic of preamplifier if the polarization potentials of two electrodes are equal. Otherwise, the direct-current (DC) differential-mode signal is formed and enters the preamplifier, resulting in a deviation of static work point, or even making the amplifier enter saturation. ECG resistance includes contact impedance between electrode and skin and the body resistance of human soft tissue. The specific contact impedance value is in connection with body parts, contact 26 Article number P_38

tightness, skin cleanliness, humidity, seasonal changes and many other factors. The 50 Hz power frequency interference and induced potential existing in the body surface and space are necessarily reflected to the preamplifier through electrode to form interference on ECG. In summary, the ECG source can be simplified to the equivalent circuit, as shown in Fig.. equation (2), which is a digital having the same magnitude as the measured signal. It is impossible to improve the SN to no matter how high the CM of differential amplifier may be at this time. It is essential to improve the contact between electrode and skin to make s s2 as small as possible and meanwhile increase the input impedance of the amplifier in order to reduce the differential mode u u2 generated by the common mode signals at the input end of prestage. The above analysis shows that ECG detection sensor should have the following basic properties: CM 00 db and input impedance sr 00MΩ and needs a 50 Hz narrow band-stop filter, which should have small whole circuit noise and be suitable for miniaturization and integration. Here a practical ECG detection sensor was designed on the basis of the above analysis. Fig.. Simplified equivalent circuit of ECG source. Where e, e2: ECG potential to be tested; s, Cs: contact impedance; e2: 50 Hz power frequency signal; sr : input impedance; E, E 2 : polarization potential of electrode; U sc : output voltage. The use of a high CM of differential amplifier to inhibit power frequency interference is reasonable because 50 Hz common-mode signal is much stronger than the potential of ECG. The common mode rejection ratio is represented by CM if the differential-mode amplification of preamplifier is K in ideal conditions without considering the internal resistance of the signal source. The output Usc can be represented by Formula (). K U = K sc ( e e ) + e CM 2 2. () Formula () indicates that the SN can be increased by improving the CM. SN min=5 when CM = 00 db if the minimum usable signal is selected to be 50 µv, which is acceptable. The situation is more complicated with the source resistance being considered. The equivalent circuit with common mode input was further simplified for ease of discussion, as shown in Fig. 2. The contact resistance s s 2 can generate differential mode between potentials u and u2 of two input ends if the earth input impedances of two input ends are sr and sr2 in differential circuit, and sr = sr2 = sr is set. u u = e (, ) 设 (2) s s2 2 2 sr s s2 sr Supposing. = 0 MΩ, = KΩ, e = V sr s s2 2 u u = e /0000 = 00 µ V can be derived from 2 2 Fig. 2. A simplified equivalent circuit with only common-mode input being considered. 3. Design on Acquisition Circuit Firstly, the high fidelity amplification of weak ECG signal was performed through anti-high frequency interference and preamplifier section after the ECG was acquired by electrodes. The gain of the preamplifier circuit cannot be designed to be too large since DC signal a few times bigger than useful ECG exists at the input end. Then, it was amplified by the main amplifier for the second time after the interference was filtered through low-pass filter, high pass filter and 50 Hz trap wave. Finally, the A/D conversion of ECG was performed [2]. 3.. Anti-High Frequency Interference and Preamplifier Circuit The frequency range of effective ECG is 0.05~00 Hz. ECG is added to the front of the amp from the body via electrodes, and an anti-highfrequency circuit was specially designed in order to ensure the proper display of ECG. The capacitors C and C2 are anti-high-frequency interference capacitors in Fig. 3. The capacitive reactance of the capacitor C is as following when the center frequency of ECG is 0 Hz: 27

X c = = = 72MΩ 6 2π fc 2π 0 220 0 (3) It has greater input impedance than the monitor, so the circuit is equivalent to an open circuit. X c is only 720 Ω which is much lower than the input impedance of the monitor when the frequency is high (such as MHz), so the circuit is equivalent to the short circuit of high-frequency interference signals to ground. and C, 2 and C2 constitute passive low-pass filters, respectively, whose cutoff frequencies are: fh = = = 33 khz 3 6 2πC 2π 0 22 0 220 0 (4) The ECG entered the preamplifier after the highfrequency interference above 33 khz was filtered. ECG belongs to differential mode signal, so instrumentation amplifier was used as a preamplifier, which has many characteristics: high input impedance (generally up to above 0 9 Ω), low bias current, high CM and single-ended output. This design selected a low-power, high-precision and low noise instrumentation amplifier AD620. The maximum of input imbalance 50 mv, Input imbalance drift 0.6 mv/ C, CM 20 db (G=0). The calculation formula of gain is: 49.4kΩ G = + G, (5) where G is the resistance between pin and pin 8, and G is the gain value. The gain value should be large enough in order to avoid interference from other signals since ECG signal is weak, however, on the other hand, saturation may appear on the first stage if the gain is too large, and so the appropriate value of G should be selected to obtain the proper gain. The testing shows that G should be set as 8.25 kom in the design, and G=7 is obtained through the calculation according to the formula (4). A common-mode feedback circuit was constituted by the AD705J, C5, 4, 5, 6 and 7 in order to further improve the CM. The common mode noise that is entered by the body and mixed in the original ECG was detected by two equal-value resistances: 3 and 5. Its first level phase reversal was amplified by AD705J, and then it returned to the right leg and was superimposed on the common mode voltage of body, thus reducing the absolute value of human common mode interference and improving SN. Such a system is conventionally called the driving system of right foot. AD620 is used in the previa differential circuit so that the amplifier has many characteristics: high input impedance, high CM and low noise, thus realizing the ECG differential amplification and effectively inhibiting the interference of 50 Hz common mode signal [3]. 3 22k 2 22k C 220pF 24.9K 5 24.9K 4 8.25K 8 2 6 IAOUT 3 Preamplifier output C2 AD620A 220pF 6 M 7 0K C3 0.uF AD705J 8 ES Floating Tracking Fig. 3. Anti-high frequency interference and preamplifier circuit. 3.2. High Pass Filter and Low-Pass Filter Amplifier temperature drift, changes in skin resistance, breathing and body muscle contraction can cause a baseline drift phenomenon of ECG, which means that the output ECG slowly moves up and down some horizontal line. These effects can be attributed to the low-frequency noise interference from the perspective of spectrum. High-pass filter can remove this effect. C4, C5, 9, 0, U constitute a second-order active high-pass filter, as shown in the left half of Fig. 4. Butterworth filter was selected in order to make the value of element simple and response characteristic flat; cutoff frequency is designed to be f = 0.05 Hz, and C = C = 2 µ F 4 5 hp and 9 = 20 are taken in order to minimize the loss of ECG low-frequency compositions and meanwhile effectively filter the interference of low frequency compositions. = 2.25 M Ω and 9 =.254 M Ω can be calculated by the formula 0 f hp =, which can approximatively 2π CC 4 5 9 0 adopt = 2.26 MΩ and =.3 M Ω. 9 0 A fourth-order Butterworth low-pass filter (right half of Fig. 4) was used in order to eliminate the high-frequency interference signals of more than 00 HZ. Component value calculations are shown in [4]. 28

C4 IAOUT 2uF 0.3M C5 2uF U 9 2.26M High-pass filter 5k 2 5k 3 0k C7 0.uF A C8 0.0047uF U2 4 20k Low-pass filter 5 20k 6 56k C9 0.uF C0 0.022uF U3 LP F_OUT Fig. 4. High pass and low pass filters. The simulation analyses of amplitude-frequency characteristics of high-pass and low-pass filters were performed in MULTISI M0 software, and Fig. 5 shows the amplitude-frequency characteristics of high-pass and low-pass filters. The amplitude attenuation is -3.7 db at 49.98 mhz and -3.02 db at 98.48 Hz. Fig. 5 shows that out-band attenuation is relatively rapid (-80 db/0 octave) because a fourth-order Butterworth low-pass filter was used. The signals sent by the electrocardio electrodes were limited to 0.05~00 Hz after the high-pass filtering and low-pass filtering. low-pass and band-pass filter designs. A typical analog architecture of state variable is used, and an inverting amplifier and two integrators are integrated inside. The integrators include 000 pf (±0.5 %) capacitance and thus solve a problem of getting low loss capacitor in the active filter design well. Meanwhile, a dedicated CAD software FILTE42 is used for the design, and the software can calculate the corresponding element values as long as the parameter is input into FILTE42 according to the requirements of designed circuit, very convenient, and has advantages of a high-precision frequency and high Q values. The designed filter circuit is shown in Fig. 6, and Fig. 7 shows response curves obtained by a simulation software Tina-TI. 3 2k 4 2k 5 5 3.2M 2 3.2M 5 20k C 0n 6 2.8k In HP FAdj BP FAdj2 LP A4 - Out Out Out Vout + In2 In3 Gnd A4 Out A4 + C2 0n Fig. 5. Amplitude-frequency characteristics of filter. Vin m U UAF42E 3.3. 50 Hz Notch Filter Fig. 6. 50 Hz notch filter. The preamplifier circuit has a strong inhibitory effect on common mode interference, but some power frequency interference enters a circuit in a differential mode signal way, and the frequency is within the frequency band of ECG. The output remains to have strong power frequency interference after the front preamplifier, high-pass filter and lowpass filter because of the electrode and input loop instability and other factors besides these, so it must be specially filtered out. The double-t trap circuit is often used against 50 Hz frequency interference. Such a circuit requires a high symmetry, and the precision demanding of components is strict, or the notch frequency and quality factor (Q value) are directly affected, so this brings difficulty to production process or device screening, and it is quite difficult to adjust. The UAF42 Universal Active Filter of BB (Burr-Brown) company has overcome this drawback, and it can be widely used in high-pass, T Gain (db) 0.00-0.00-20.00-30.00 0 25 50 75 00 Frequency (Hz) Fig. 7. The amplitude-frequency response curve of 50 Hz notch filter (Q=0). Trap wave point is f NOTCH =50 Hz, and the frequency value may be changed by and 2 values. Quality factor (Q value) is determined by 6, and their relationship is: 29

6 = 25k Q (6) The Q value should be 0 if 6 =2.8 k is set. Q-value must be consistent with the ratios of 5 / 3 or 5 /. Thereby, the ratio of 4 feedback resistance 5 to input resistance 3 or 4 is also 0. The stopband bandwidth meets BW -3dB =f NOTCH /Q=5 Hz. The high Q is corresponding to a narrower stopband for the notch filter, and low Q is corresponding to a wider stopband. Stopband bandwidth and attenuation multiple is contradictory: if bandwidth is small, then attenuation is small, or if bandwidth is big, then attenuation is big. It is essential to find a balance between stopband bandwidth and attenuation multiple when ECG is processed. The stopband bandwidth is 5 Hz in Fig. 7, and the gain is about -28 db to ensure the integrity of the ECG and meanwhile effectively filter out the 50 Hz power frequency interference. 7 0k 9 00k U4 8 0k 20 ES Fig. 8. The main amplifier. Fig. 9. Electrocardiogram. 22 200k U5 2 0k A/D 3.4. The Main Amplifier 5. Conclusions ECG amplitude is about 0~4000 µv which needs to be amplified for 000 times to a volt level before entering the A/D. Preamplifier magnification is 7 times, and finally it just needs to be amplified for 43 times through the main amplifier. Post amplification circuit is the two-stage reversed-phase amplifier, as shown in Fig. 8. The first stage of amplifier gain is 9, and the second stage 7 = 0 22 of amplifier gain is. is the adjustable 22 20 resistance of 200 kω which can be used to adjust the overall gain size of post amplification circuit to suit the requirements of different subsequent circuits. 4. The Applications on ECG Detection Sensor The testing shows the performance indicators of ECG detection sensor as follows: CM: 08 db; Pass band: 0.05~00 Hz; Input impedance: greater than 00 MΩ; Differential mode voltage gai: 000 times; The efficiency attenuation of power frequency filter: about 30 db. Fig. 9 shows a section of electrocardiogram obtained after the sensor was applied to some portable ECG monitor. The ECG is clear and stable, so the sensor is fully able to meet ECG monitoring requirements. The desired results are also got. ECG is a complex physiological signal, and there exist many factors to affect the accuracy and reliability of measurement in ECG measurement. ECG detection sensor was designed on the basis of the characteristics of ECG, which can effectively inhibit common mode signal, amplify weak ECG and has an ideal CM. The 50 Hz notch filter designed using special UAF42 filter chip especially has the advantages of a high precision frequency, narrow bandwidth and high attenuation and solves a difficult problem: obtaining the capacitor with a stringent tolerance, low loss and high degree of matching in the design of filter. eferences 203 Copyright, International Frequency Sensor Association (IFSA). All rights reserved. (http://www.sensorsportal.com) []. arnon Pallas-Areny, Interference rejection characteristics of Biopotential amplifiers, a comparative analysis, IEEE Transaction on Biomedical Engineering, Vol. 35, Issue, 998, pp. 20-24. [2]. Zhang Wei, Zhang Shi, Bao Xirong, et al., Design of preamplifier and anti-interference for the portable ECG monitor, Information of Medical Equipment, Vol. 20, Issue 0, 2005, pp. 7-9, 34. [3]. Huang Jinwen, esearch the effect of grounding in medical signal acquiring, Journal of Yunnan University (Natural Sciences Edition), Vol. 29, Issue S, 2007, pp. 9-23. [4]. The precise design of active power filter, Publishing House of Electronic Industry, Beijing, 984. 30