ME1000 RF Circuit Design. Lab 1. Calibration with Spectrum Analyzer

Similar documents
ME1000 RF Circuit Design. Lab 4. Filter Characterization using Vector Network Analyzer (VNA)

Antenna and Propagation

External Source Control

Agilent X-Series Signal Analyzer This manual provides documentation for the following X-Series Analyzer: CXA Signal Analyzer N9000A

Agilent PSA Series Spectrum Analyzers Noise Figure Measurements Personality

Frequency and Time Domain Representation of Sinusoidal Signals

PGT313 Digital Communication Technology. Lab 6. Spectrum Analysis of CDMA Signal

Keysight 2-Port and 4-Port PNA-X Network Analyzer

Demo / Application Guide for DSA815(-TG) / DSA1000 Series

Agilent PSA Series Spectrum Analyzers Noise Figure Measurements Personality

Agilent Technologies PSA Series Spectrum Analyzers Test and Adjustment Software

Advanced Test Equipment Rentals ATEC (2832)

Swept Return Loss & VSWR Antenna Measurements using the Eagle Technologies RF Bridge

Impedance 50 (75 connectors via adapters)

Application Note: Swept Return Loss & VSWR Antenna Measurements using the Eagle Technologies RF Bridge

AP034-OM-E Rev D ISSUED: January 2000 ²

Agilent 2-Port and 4-Port PNA-X Network Analyzer. N5241A - 10 MHz to 13.5 GHz N5242A - 10 MHz to 26.5 GHz Data Sheet and Technical Specifications

Contents. CALIBRATION PROCEDURE NI PXIe-5668R 14 GHz and 26.5 GHz Signal Analyzer

VVM measurement with E5061B for replacing 8508A vector voltmeter. May 2013 Agilent Technologies

Site Master Cable and Antenna Analyzer with Spectrum Analyzer

Xeta4. Occupied Bandwidth. Measurements

PNA Family Microwave Network Analyzers (N522x/3x/4xB) CONFIGURATION GUIDE

PXIe Contents CALIBRATION PROCEDURE

Evaluation of Filter and Amplifier Transmission Characteristics

DSA-815 Demo Guide. Solution: The DSA 800 series of spectrum analyzers are packed with features.

LB480A Pulse Profiling USB PowerSensor+ Data Sheet

S3602A/B Vector Network Analyzer Datasheet

Improving Amplitude Accuracy with Next-Generation Signal Generators

Exercise 5: Power amplifier measurement

772D coaxial dual-directional coupler 773D coaxial directional coupler. 775D coaxial dual-directional coupler 776D coaxial dual-directional coupler

PXIe Contents. Required Software CALIBRATION PROCEDURE

Agilent PSA Series Spectrum Analyzers Self-Guided Demonstration for Phase Noise Measurements

Signal Generators for Anritsu RF and Microwave Handheld Instruments

LB680A Pulse Profiling USB PowerSensor+ Data Sheet

Test & Calibration Benefits from a New Precision RF/Microwave Calibrator

Keysight X-Series Signal Analyzers

Agilent 4-Port PNA-L Network Analyzers

Signal Analysis Measurement Guide

TEST & MEASURING INSTRUMENTS. Analyzer. (4 Ports) 4 Ports

Contents. CALIBRATION PROCEDURE NI PXIe GHz and 14 GHz RF Vector Signal Analyzer

PXIe Contents. Required Software CALIBRATION PROCEDURE

Understanding RF and Microwave Analysis Basics

LB480A Pulse Profiling USB PowerSensor+ Data Sheet

P a g e 1 ST985. TDR Cable Analyzer Instruction Manual. Analog Arts Inc.

Reconfigurable 6 GHz RF Vector Signal Transceiver with 1 GHz Bandwidth

Obtaining Flat Test Port Power with the Agilent 8360 s User Flatness Correction Feature. Product Note

PXA Configuration. Frequency range

EXHIBIT 7: MEASUREMENT PROCEDURES Pursuant 47 CFR 2.947

RIGOL. User s Guide. DSA800 Options and Accessories. Dec RIGOL TECHNOLOGIES, INC.

GET10B Radar Measurement Basics- Spectrum Analysis of Pulsed Signals. Copyright 2001 Agilent Technologies, Inc.

LTE Advanced Test Guide CellAdvisor JD745B or JD785A/B Series (FW 3.085)

Power Meter. Measurement Guide. for Anritsu RF and Microwave Handheld Instruments BTS Master Site Master Spectrum Master Cell Master

Migrating 4195A to E5061B LF-RF Network Analyzer. April 2010 Agilent Technologies

PXIe Contents CALIBRATION PROCEDURE. Reconfigurable 6 GHz RF Vector Signal Transceiver with 200 MHz Bandwidth

ECE 4670 Spring 2014 Lab 1 Linear System Characteristics

FCC PART 15C TEST REPORT FOR CERTIFICATION On Behalf of. DEI Sales Inc. dba Definitive Technology. Model Number: STUDIO SLIM SUBWOOFER

DEPARTMENT OF THE ARMY TECHNICAL BULLETIN CALIBRATION PROCEDURE FOR SIGNAL GENERATOR SG-1207/U (HEWLETT-PACKARD, MODEL 8642M)

RF and Microwave Test and Design Roadshow 5 Locations across Australia and New Zealand

Circuit Characterization with the Agilent 8714 VNA

S3602C Vector Network Analyzer Datasheet

Analysis of RF transceivers used in automotive

AirScope Spectrum Analyzer User s Manual

Agilent PSA Series Spectrum Analyzers Self-Guided Demonstration for Spectrum Analysis

Agilent FieldFox RF Analyzer N9912A

Agilent 4396B Network/Spectrum/Impedance Analyzer PERFORMANCE TEST MANUAL SERIAL NUMBERS. This manual applies directly to instruments with serial

Schematic-Level Transmission Line Models for the Pyramid Probe

Keysight Technologies PNA Microwave Network Analyzers

Agilent 8902A Measuring Receiver Product Note

PN9000 PULSED CARRIER MEASUREMENTS

Configuration of PNA-X, NVNA and X parameters

Agilent CSA Spectrum Analyzer

Keysight Technologies E8257D PSG Microwave Analog Signal Generator. Data Sheet

Keysight X-Series Signal Analyzers

Contents. CALIBRATION PROCEDURE PXIe-5673 Vector Signal Generator

FCC Test Report - LTE Band 26 (part 22)

A Test Lab Techno Corp. Report Number:1410FR27

Keysight Technologies Active Differential Probes U1818A 100 khz to 7 GHz U1818B 100 khz to 12 GHz. Technical Overview

Compact Series: S5065 & S5085 Vector Network Analyzers KEY FEATURES

Keysight 2-Port and 4-Port PNA-X Network Analyzer

VR5-HD Functionality Verification Procedure. Application Note

Agilent X-Series Signal Analyzer

Handheld 3.3GHz Spectrum Analyzer

SIGNAL GENERATORS. MG3633A 10 khz to 2700 MHz SYNTHESIZED SIGNAL GENERATOR GPIB

Agilent PSA Series Spectrum Analyzers Self-Guided Demonstration for GSM and EDGE Measurements

PGT313 Digital Communication Technology. Lab 3. Quadrature Phase Shift Keying (QPSK) and 8-Phase Shift Keying (8-PSK)

PLANAR 814/1. Vector Network Analyzer

GT 9000 GT 9000S MICROWAVE

Spectrum Analyzer TEN MINUTE TUTORIAL

Keysight X-Series Signal Analyzers

SAV502 NB-IoT Modular Signal Analyzer Datasheet

Faculty of Electrical & Electronics Engineering BEE4233 Antenna and Propagation. LAB 1: Introduction to Antenna Measurement

Agilent PSA Series Spectrum Analyzers Self-Guided Demonstration for Spectrum Analysis

Keysight Technologies External Source Control X-Series Signal Analyzers Option ESC. Demo Guide

FieldFox Handheld Analyzers

Keysight Technologies E8257D PSG Microwave Analog Signal Generator

OPEN TEM CELLS FOR EMC PRE-COMPLIANCE TESTING

ECE 2111 Signals and Systems Spring 2009, UMD Experiment 3: The Spectrum Analyzer

Lab Assignment #3 Analog Modulation (An Introduction to RF Signal, Noise and Distortion Measurements in the Frequency Domain)

ENGINEERING COMMITTEE Interface Practices Subcommittee AMERICAN NATIONAL STANDARD ANSI/SCTE

SWR/Return Loss Measurements Using System IIA

Transcription:

ME1000 RF Circuit Design Lab 1 Calibration with Spectrum Analyzer This courseware product contains scholarly and technical information and is protected by copyright laws and international treaties. No part of this publication may be reproduced by any means, be it transmitted, transcribed, photocopied, stored in a retrieval system, or translated into any language in any form, without the prior written permission of Acehub Vista Sdn. Bhd. The use of the courseware product and all other products developed and/or distributed by Acehub Vista Sdn. Bhd. are subject to the applicable License Agreement. For further information, see the Courseware Product License Agreement. Objectives i) To perform scalar offset calibration with a spectrum analyzer (SA) ii) To verify the signal purity of a signal generator (SG) iii) To determine the losses contributed by the cables, connectors, and PCB traces Equipments requiredrequired i) Agilent N9320B 3 GHz RF Spectrum Analyzer ii) Agilent N9310A RF Signal Generator Accessories requiredrequired i) TRM standard calibration kit ii) 2 SMA(m)-to-SMA(m) coaxial cable ME1000 RF Circuit Design Lab 1-1/9

1. Introduction 1.1 The Need for Scalar Offset Calibrations A typical frequency domain measurement of a device-under-test (DUT) is shown in Figure 1Figure 1. When measuring the frequency response of a DUT, you will collect the power transferred from a SG to a SA via the DUT. To ensure that the data collected represents the actual response of the DUT only and not portions related to the cables, connectors, and the losses within the measurement equipment itself it is necessary to measure the total path loss of the interconnections from the equipment to the input and output terminals of the DUT. In principle, the actual effect of the DUT is the measured value minus the losses due to the interconnections on both ends of the DUT. The process of accounting for the effect of the interconnection during a measurement is called Calibration. Since we are only interested in power measurement, the calibration process described here is known as Scalar Offset Calibration, and it is usually performed at various frequency of interest. Figure 1 The Need to Calibrate Scalar Offset of Cables, Connectors, and PCB Trace Losses ME1000 RF Circuit Design Lab 1-2/9

2. SG Verification and Calibration of Loss Contributed by the Cable Important: Figure 2 SG Verification and SMA Coaxial Cable Loss Calibration Excessive input power levels can damage the SA. The threshold for damage is different for various models. The input setting can be as low as 20 dbm (0.1 W or 3.2 V [1] in a 50 Ω load). Observe the caution notice on the front panel of the equipment. [1] We are referring to the time-averaged power, V P where R = 50 Ω. 2R 2 First, we must investigate whether the SG we use is a perfectly linear device by connecting the SG directly to the SA as shown in Figure 2. 1. Use the following settings for the SG: CW frequency: Power level: 868 MHz 40 dbm (to be increased in 10 db steps until 0 dbm) N9310A setting: Preset to default settings: Frequency: Amplitude: Turn Off Mod: Turn On RF Out: [Preset] [Frequency] > [868] > {MHZ} [Amplitude] > [+/ ] > [40] > {dbm} [Mod On/Off] [RF On/Off] 2. Use the following settings for the SA: Preset the SA to its default settings. Resolution bandwidth: 100 khz Input attenuation: 10 db (or auto) N9320B setting: Preset to default settings: Attenuation: [Preset/System] > {Preset} [SPAN] > {FULL SPAN} [AMPLITUDE] > {Attenuation} > [10] > {db} [BW/Avg] > {Res BW} > {100} > {KHZ} ME1000 RF Circuit Design Lab 1-3/9

Exercises a) What are the frequencies that appear in the SA when signal power is 40 dbm? Frequency = MHz, MHz, and MHz Note: Use marker function Peak Search if necessary. b) Increase the power level by 10 db each step until 0 dbm and note the different frequencies displayed. Are there any other frequencies that appear during this power increase? YES, they are MHz and MHz or NO, no other frequencies have appeared. c) Explain why there is more than a single frequency component at the measured output. d) Explain the difference between the measured and the expected output power level when the input power level is at 0 dbm. e) Fill in the table below: Input Power from Signal Generator, Psig_gen (dbm) Fundamental Power (dbm) 2 nd Harmonic Power (dbm) 3 rd Harmonic Power (dbm) 40 30 20 10 0 We can calibrate the cable losses before measuring any DUT by using the same setup as shown in Figure 2. 3. Use the following settings to determine the loss contribute by the cable: SG settings CW frequency: Power: N9310A setting: Frequency: Amplitude: SA settings Centre frequency: Input Averaging: N9320B setting: Center frequency: Attenuation: Averaging: 868 MHz 25 dbm [Frequency] > [868] > {MHZ} [Amplitude] > [+/ ] > [25] > {dbm} 868 MHz 10 MHz (approximately 1% of centre frequency) Attenuation: 10 db (or auto) 100 khz (or auto) On [FREQUENCY] > [868] > {MHZ} [SPAN] > [10] > {MHZ} [AMPLITUDE] > {Attenuation} > [10] > {db} [BW/Avg] > {Res BW} > {100} > {KHZ} [BW/Avg] > {Average} ME1000 RF Circuit Design Lab 1-4/9

Exercises a) What is the frequency and power display on the SA when the input power level is 25 dbm? Frequency = MHz Power level = dbm Note: Use the marker function if necessary. b) Fill in the table below to determine the cable loss for various test signals. You should change the frequency and power level settings on the SG, as well as the center frequency and span settings on the SA. Use a span setting approximately 1% of the centerre frequency. SG Frequency, fsig_gen (MHz) SG Power Level, Psig_gen (dbm) 50 25 SA Measured Peak Frequency, fsa (MHz) SA Measured Output Power, Psa (dbm) Loss Contributed by Cable, Lcable = Psig_gen Psa (db) 50 20 50 15 50 10 50 5 100 25 100 10 500 15 500 10 818 25 818 20 818 15 818 10 818 5 818 0 868 25 868 20 868 18 868 15 868 13 868 10 868 8 868 5 868 0 1000 15 1736 15 1736 5 2604 15 2604 5 c) Does the cable loss remain the same for different input frequencies? d) By how much does the SA measured signal power level drop at 50 MHz and 868 MHz when the SG power level setting is reduced by 10 db? ME1000 RF Circuit Design Lab 1-5/9

3. Path Loss Calibration The total path loss in a frequency response measurement consists of two parts: the input path loss and the output path loss. Each path loss is in turn due to the coaxial cable and connector losses, and the loss from the copper trace on the printed circuit board (PCB) that leads to the DUT terminals. We can measure the path loss with the use of a calibration structure called the TRM (Through-Reflect-Match) board. Input path loss: Cable-SG loss and half of the loss from the TRM board (PCB traces and connectors). Output path loss: Cable-SA loss and half of the loss from the TRM board. Assumption: The input and output paths of the TRM board are identical. Since both input and output path loss consists of two parts, these two parts can be determined individually from the following procedures and summed up to give us the total path loss. 3.1.1. Procedures 1. Measure the cable loss from the Cable-SG, Lcable_sg and Cable-SA, Lcable_sa. 2. Measure the loss from the TRM board (through path). 3.1.2. Equipment Setup Figure 3 - SMA coaxial cable loss measurement setup We can measure the cable losses by using the setup shown in Figure 3. 1. Connect the SMA coaxial cable used for connecting the SG as shown in Figure 3. This cable will be referred to as Cable-SG. Label this cable for identification. 2. Use the following settings the for SG and SA: SG settings CW frequency: Power: Frequency of interest 25 dbm N9310A setting: Frequency: Amplitude: [Frequency] > [868] > {MHZ} [Amplitude] > [+/ ] > [25] > {dbm} ME1000 RF Circuit Design Lab 1-6/9

SA settings Centre frequency: Input attenuation: Averaging: Frequency of interest 10 MHz 10 db (or auto) 100 khz (or auto) On N9320B setting: Center frequency: Attenuation: Averaging: [FREQUENCY] > [868] > {MHZ} [SPAN] > [10] > {MHZ} [AMPLITUDE] > {Attenuation} > [10] > {db} [BW/Avg] > {Res BW} > {100} > {KHZ} [BW/Avg] > {Average} Note: 50 MHz and 868 MHz are required for the other labs to compensate for the losses. Cable loss = SA reading SG amplitude setting Since it is a loss, omit the negative sign of the final answer for the equation above. 1. Make the measurements and record the losses for Cable-SG in Table 1: Table 1 Path Loss Calibration Data Frequency (MHz) Cable-SG Loss, Lcable_sg Cable-SA Loss, Lcable_sa 50 818 868 750 800 850 900 950 1000 1050 1100 1150 1200 1250 1300 1350 1400 2. Connect the SMA coaxial cable used for connecting the SA as shown in Figure 3. This cable will be referred to as Cable-SA. Label this cable for identification. 3. Make the measurements and record the losses for Cable-SA in Table 1. ME1000 RF Circuit Design Lab 1-7/9

4. Losses Contributed by Connector and PCB Trace 5. Losses Contributed by Connector and PCB Trace Using a TRM board that represents the copper traces on the trainer board, we can measure the loss up until the DUT end. Figure 4 SMA Connector and PCB Trace Loss Calibration 1. Make the connection as shown in Figure 4. 2. The measurements from this connection are the total loss contributed by the Cable-SG, Cable-SA, and the full through path of the TRM board. a. Lfull_path = Measured signal amplitude SG amplitude setting (Lcable_sg + Lcable_sa) b. Lhalf_path = 0.5 * Lfull_path Assumption: The input and output path losses are equal. Note: The cable losses are of positive values as taken from Table 1. Omit the negative sign for the Lfull_path at the end of calculation. 3. Make the measurements and record the losses for the TRM through loss under the Measured Power at the SA column in Table 2. 4. Calculate the Lfull_path and Lhalf_path and record it in Table 2. ME1000 RF Circuit Design Lab 1-8/9

5. Calculate the total input and output path loss as shown below : a. Total input path loss, Linput-path = Lcable-sg + Lhalf-path b. Total output path loss, Loutput-path = Lcable-sa + Lhalf-path Table 2 Input and Output Path Loss Calibration Data Frequency (MHz) 50 Cable-SG Loss, Lcable_sg Cable-SA Loss, Lcable_sa Measured Power at SA TRM Through Loss, Lfull_path TRM Through Loss, Lhalf_path Total Input Path Loss, Linput _path Total Output Path Loss, Loutput _path 818 868 750 800 850 900 950 1000 1050 1100 1150 1200 1250 1300 1350 1400 References [1] Presentation slides, A Seminar on RF Measurement Spectrum Analysis Basics, Agilent Technologies, 2001 [2] Thomas H. Lee, Planar Microwave Engineering, Cambridge University Press, 2004 ME1000 RF Circuit Design Lab 1-9/9