Preliminary studies of a new monitor ionization chamber

Similar documents
DETERMINATION OF INHERENT AND ADDITIONAL FILTRATION IN ORDER TO ESTABLISH RADIATION QUALITIES ACCORDING TO IEC 61267

Response characteristics of a tandem ionization chamber in standard X-ray beams

Determination of the tube voltage from clinic mammographic system using two types of detectors

Cylindrical Ion Chambers

Cylindrical Ion Chambers Victoreen Model 550 Series

Calibration of KAP meters

DETECTORS. Including Codes of Practice IONIZING RADIATION

SPECIFICATION. Kilovoltage X-ray calibration system for protection and diagnostic level dosimetry. Prepared by

IONIZING RADIATION 2018 / 2019 DETECTORS. Including Codes of Practice

MEASURING CONDITIONS AND UNCERTAINTIES FOR THE COMPARISON AND CALIBRATION OF NATIONAL DOSIMETRIC STANDARDS AT THE BIPM *

Development of the Mini-Ionization Chamber for High-Dose Real-Time Monitoring Inside a Gamma Irradiation Facility

Joint ICTP/IAEA Advanced School on Dosimetry in Diagnostic Radiology and its Clinical Implementation May 2009

I. Introduction.

RADIATION THERAPY DETECTORS. For Small Field Dosimetry

Electronic Brachytherapy Sources. Thomas W. Rusch

DOSE EVALUATION IN UROGRAPHY RADIOLOGICAL PROCEDURES

Technical Protocol. SIM Comparison of Calibration Coefficients at Radiotherapy Level for Orthovoltage X-ray Beams. Pilot laboratory: NIST

Estimation of signal transfer property for wireless digital detector in different measurement schemes

Dose Reduction and Image Preservation After the Introduction of a 0.1 mm Cu Filter into the LODOX Statscan unit above 110 kvp

DETECTORS UNCOMPROMISING QUALITY. The standard in dosimetry measurements for over 40 years. EXRADIN DETECTORS

Introduction of a Single Chip TLD System for Patient Dosimetry

When small things matter. Small Field Dosimetry Application Guide

Derivation of total filtration thickness for diagnostic x-ray source assembly

Predicted image quality of a CMOS APS X-ray detector across a range of mammographic beam qualities

PIXE TECHNIQUE APPLIED TO ALMEIDA JÚNIOR MATERIALS

Half value layer and AEC receptor dose compliance survey in Estonia

Commissioning and Calibrating a Linear Accelerator State-of-the-Art in 2010

Monitoring the eye lens

Received 3 December 2016; revised 17 May 2017; editorial decision 21 June 2017; accepted 4 July 2017

Test Equipment for Radiology and CT Quality Control Contents

Effect of pressure, temperature and humidity in air on photon fluence and air kerma values at low photon energies

Progress Report from the HDR Brachytherapy Working Party. Margaret Bidmead

IV. 4. An Optical Common-mode Rejection for Improving the Sensitivity Limit of a Radiochromic Imaging Film

THE SLAC PULSED X-RAY FACILITY* ABSTRACT

熊本大学学術リポジトリ. Kumamoto University Repositor

Title: A COMPARISON OF Cs-137 AND X-RAY SOURCES AS CALIBRATION REFERENCES FOR THERMOLUMINESCENT DOSIMETER CHIPS

COMPUTED RADIOGRAPHY SIMULATION USING THE MONTE CARLO CODE MCNPX

Exposure Indices and Target Values in Radiography: What Are They and How Can You Use Them?

Development of Personal Dosimeter Using Electronic Dose Conversion Method

Detectors for small field dosimetry

Development of Solid-State Detector for X-ray Computed Tomography

Update on NIST X-Ray Air-Kerma Standards and Calibrations

THE USE OF CdTe DETECTORS FOR DENTAL X-RAY SPECTROMETRY

ABSORBED DOSE DISTRIBUTIONS USING THE ISODENSITOMETRIC METHOD FOR EXPOSURES WITH FILTER EMPLOYED FOR MAMMOGRAPHIES

PEDIATRIC AND STAFF DOSE EVALUATION IN FLUOROSCOPY UPPER GASTROINTESTINAL SERIES

Chromatic X-Ray imaging with a fine pitch CdTe sensor coupled to a large area photon counting pixel ASIC

Energy Monitoring Device for Electron Beam Facilities

X-ray tube with needle-like anode

The on-line detectors of the beam delivery system for the Centro Nazionale di Adroterapia Oncologica(CNAO)

Effective energy measurement using radiochromic film: application of a mobile scanner

PATIENT EFFECTIVE DOSES IN DIAGNOSTIC RADIOLOGY, NA

2012 :15th SESSION of ESMP

A Multichannel Dosimeter Based on Scintillating Fibers for Medical Applications

ELECTRONIC CONTROL CONCEPTS 160 Partition Street Saugerties, NY or local phone

NOT FOR DISTRIBUTION JINST_083P_0914 v1

ISO INTERNATIONAL STANDARD. Non-destructive testing of welds Radiographic testing Part 1: X- and gamma-ray techniques with film

IAEA Coordinated Research Project on Development of Harmonized QA/QC Procedures for Maintenance and Repair of Nuclear Instruments

OPERATION MANUAL. Model LET-SW5. April Far West Technology, Inc. 330 South Kellogg Ave, Suite D Goleta, CA 93117

The evaluation of minimum detectable phantom thickness change using a scanning liquid filled ion chamber EPID dose response

This document is a preview generated by EVS

Nuclear Associates

ADVANCED THERAPY DOSIMETER MODEL 35040

Ultrastable Low-Noise Current Amplifiers With Extended Range and Improved Accuracy

Electron density profile reconstruction on the TCABR sweeping reflectometer

Post-irradiation colouration of Gafchromic EBT radiochromic film

Dose reduction using Cu-filter for full-spine radiografic examination of patients with adolescent idiopathic scoliosis

computed tomography, computed tomography dose index, dosimetry

Comparison of peripheral dose measurements using Ionization chamber and MOSFET detector

DEVELOPMENT OF A CHARGE-SENSITIVE PREAMPLIFIER USING COMMERCIALLY AVAILABLE COMPONENTES

High-sensitivity compact dosimeter using two silicon X-ray diodes

I. PERFORMANCE OF X-RAY PRODUCTION COMPONENTS FLUOROSCOPIC ACCEPTANCE TESTING: TEST PROCEDURES & PERFORMANCE CRITERIA

Barcelona, January 2011 Workshop on optimization of Radiation Protection of Medical Staff

Key words: fluoroscopy, dose-area-product, kerma-area-product, calibration of KAP meters, patient exposure

Abstract. This investigation was supported by PHS grant CA10953 awarded by the NCI, DHHS. Introduction

New Detectors for X-Ray Metal Thickness Measuring

The Absolute Detection Efficiency of Vacuum Electron Multipliers, to kev Protons and Ar + Ions.

A comparison of two methods for the determination of freein-air geometric efficiency in MDCT

RaySafe X2. Effortless measurements of X-ray

Improved Imaging Plate Dosimetry for X-rays in Interventional Radiology

Development of the Use of Amorphous Silicon (ASi) Electronic Portal Imaging Devices as a Physics Tool for Routine Linear Accelerator QA

Pulse-Energy-Enhanced, Strongly Modulated Er:YLF Laser for Medical Applications

Ideas for type tests of electronic dosemeters in pulsed fields

DIAGNOSTIC OF CORROSION DEFECTS IN STEAM GENERATOR TUBES USING ADVANCED SIGNAL PROCESSING FROM EDDY CURRENT TESTING

X-RAY IMAGING EE 472 F2017. Prof. Yasser Mostafa Kadah

Characterization of an in vivo diode dosimetry system for clinical use

236 IEEE TRANSACTIONS ON NUCLEAR SCIENCE, VOL. 59, NO. 1, FEBRUARY 2012

The importance of radiation quality for optimisation in radiology

A new approach to film dosimetry for high energy photon beams: Lateral scatter filtering

Imaging Technique Optimization of Tungsten Anode FFDM System

Dosimetry and Position Sensing Ionization Chamber for Ion Beam Tracking

Technical aspects on DAP calibration and CT calibration

Introduction of Computed Radiography in Two Mammography Services: Image Quality and Dose Analysis

QUALITY CONTROL TESTS IN SOME DIAGNOSTIC X-RAY UNITS IN BANGLADESH

Radiographic sensitivity improved by optimized high resolution X -ray detector design.

Amorphous Selenium Direct Radiography for Industrial Imaging

membrane sample EUV characterization

Collimation Assessment Using GAFCHROMIC XR-M2

Diagnostic x-ray equipment compliance and facility survey

SUBCHAPTER 14. THERAPEUTIC INSTALLATIONS

Analysis of the kinetics of phase and amplitude gratings recorded in azopolymer films

Transcription:

1 Preliminary studies of a new monitor ionization chamber Maíra T. Yoshizumi, Vitor Vivolo and Linda V. E. Caldas Instituto de Pesquisas Energéticas e Nucleares (IPEN-CNEN) Comissão Nacional de Energia Nuclear Av. Prof. Lineu Prestes, 2242, 05508-000, São Paulo, SP, Brazil Tel/Fax: +55 11 31339715 E-mail: mairaty@ipen.br E-mail: vivolo@ipen.br E-mail: lcaldas@ipen.br Abstract: A new monitor ionization chamber was developed at Instituto de Pesquisas Energéticas e Nucleares (IPEN) in order to monitor X-ray beams. The main difference of this monitor ionization chamber in relation to other monitor chambers is its geometry which consists of a ring shaped sensitive volume. Because of this geometry, the monitor chamber has a central 5 hole through which the direct radiation beam passes. The operational characteristics of the monitor chamber were evaluated: saturation, ion collection efficiency and polarity effect. Besides these tests, the short- and medium-term stabilities of its response were also evaluated. During the tests the leakage current was always negligible. All results showed values within those recommended internationally (IEC, 1997). Corresponding author

2 Keywords: monitor chamber, X-radiation, pre-operational characteristics. INTRODUCTION Ionization chambers are the most used dosimeters for measurements of exposure or absorbed dose. These chambers are available in different designs and they are constituted by materials depending on their application (Boag, 1987). The ionization chambers are mostly cylindrical, spherical and with parallel-plates. A transmission ionization chamber is a special type of ionization chamber, used to monitor radiation beams, needed always when the radiation generator output may vary due to fluctuations of the power-line, for example. A simple parallel-plate chamber was suggested by Attix (Attix, 1986) to function as a monitor chamber for X-ray beam applications. In this work, he suggested thick Lucite plates coated with graphite to be used as the collecting electrode, guard ring and window entrance, all of them electrically insulated. For electron beams and soft X-ray beams the window entrance can be substituted by a stretched foil. A thimble chamber (Farmer type chamber) was another suggestion for this purpose (Attix, 1986). Recently, the International Atomic Energy Agency published a report with recommendations to be used in the dosimetry of diagnostic radiology beams (IAEA, 2007). In this document there are many recommendations on the use of monitor chambers, such as their positioning in the radiation field, performance specifications, and also a calibration method of other chambers, by the substitution method, using the monitor chamber.

3 According to the same publication (IAEA, 2007), the monitor chamber should add as little filtration to the beam as possible and the performance specifications of the monitor chamber and the associated measuring assembly should be similar to that of the standard instrument. Several ionization chambers have been developed at the Calibration Laboratory of IPEN for use in the diagnostic radiology and radiotherapy energy ranges (Albuquerque and Caldas, 1989; Dias and Caldas, 1998; Costa and Caldas, 2003; Maia and Caldas, 2005; Oliveira and Caldas, 2007). In this work a new monitor ionization chamber was developed to be used for monitoring of X- ray radiation beams, diagnostic radiology level. This ionization chamber was designed to be simple and cheap. The main difference between this chamber and commercial monitor transmission chambers is its ring shaped sensitive volume. There is a central hole in the ionization chamber through which the radiation beam passes. The aim of this central hole is to avoid the interference of the ionization chamber with the direct radiation beams, maintaining their original spectra. MATERIALS AND METHODS The new monitor ionization chamber was connected to an electrometer (Model UNIDOS 10474, Physikalisch-Technische Werkstätten (PTW), Freiburg, Germany). All measurements were corrected to the environmental reference conditions (20 C and 101.3 kpa). For the pre-operational tests, an industrial X-ray equipment Pantak/Seifert, model ISOVOLT 160HS, that operates from 5 to 160 kv, was utilized. These tests were performed using the

4 diagnostic quality beam RQR 5 defined by the International Electrotechnical Commission, IEC 1267 (1994). The reference system for the established diagnostic radiology qualities in this equipment is a parallel-plate ionization chamber PTW, model 77334, with a 1 cm 3 sensitive volume, traceable to the German primary standard laboratory Physikalisch-Technische Bundesanstalt (PTB). The stability tests were performed using a 90 Sr+ 90 Y check source device, PTW, model 8921, with nominal activity of 33 MBq, 1994. RESULTS AND DISCUSSION The new ionization chamber is showed in Figure 1. This is a parallel-plate ionization chamber with a central hole. The chamber body is made of PMMA that guarantees electrical insulation and mechanical resistance. The collecting electrode is made of a ring shaped aluminum plate and the outer electrode (entrance window) is an aluminized polyester foil. The collecting electrode is surrounded by PMMA discs coated with graphite spray, which operate as guard rings (one surrounding the total sensitive volume and the other surrounding the central hole). The guard rings are necessary to keep the electric field constant on both sides of the collecting electrode. The outer diameter of the ionization chamber is 190 mm, the central hole has 60 mm of diameter and the total sensitive volume is about 200 cm 3. The final ionization chamber is showed in Figure 2. The pre-operational tests of the monitor chamber studied were: saturation curve, ion collection efficiency and polarity effect. The short- and medium-term stability measurements of the chamber response were taken during one month. During this period, the leakage current was always negligible.

5 Saturation, ion collection efficiency and polarity effect The saturation test was performed to determine the optimal applied voltage for the ionization chamber. A saturation curve was obtained from measurements of the chamber response by varying the X-ray tube voltage from -400 V to +400 V, in steps of ±50 V (Figure 3). The ionization chamber was positioned at 30 cm from the focal spot; at this distance the air kerma rate was 544.2 mgy/min. As showed in Figure 3, the ionization chamber achieved the saturation of its response. The chosen applied voltage for the monitor chamber was -400 V. The ion collection efficiency was obtained by the two-voltage method, given by the equation (IAEA, 2001): K S 2 ( V1 V2 ) 1 2 ( V V ) ( M M ) =, 1 2 1 2 where M i is the collected charge at the polarizing voltages V i (operational polarizing voltage) and V 1 / V 2 = 2. The ion collection efficiency was better than 99.9% for both polarities. This result is in accordance with the value of 5% of ionic recombination losses, recommended by IEC (1997). The polarity effect was determined by the ratio of the collected charges for similar voltages of opposite signals. In Table 1 the results for all pair of voltages are presented. Leakage current The leakage current was determined by measuring the chamber response during 20 minutes, before and after irradiation. For both cases, the leakage current represented only 0.25% of the

ionization current produced by the minimum air kerma rate used in this work. This value is within the limit recommended internationally (IEC, 1997). 6 Short- and medium-term stability For the short-term stability test, the monitor chamber was exposed to the check source, and 10 consecutive measurements were taken, under reproducible conditions. The maximum variation of the monitor chamber response was less than 0.4%, and therefore it is within the recommended limit of 3% (IEC, 1997). The medium-term stability test was obtained by taking the medium value of the 10 measurements of the short-term stability during a period of one month (Figure 4). The maximum variation obtained was 0.7%, thus within the 2% value recommended limit (IEC, 1997). CONCLUSIONS The new monitor ionization chamber developed at IPEN and presented in this work achieved the expected results in the case of all pre-operational tests realized: saturation, ion collection efficiency and polarity effect. During these tests the chamber response stability showed an adequate performance, satisfying the international recommendations. This monitor chamber is a good alternative to any laboratory or hospital, because of its simple design and low-cost materials.

7 ACKNOWLEDGMENTS The authors are thankful to Mr. M. Xavier for technical assistance, and to Conselho Nacional de Desenvolvimento Científico e Tecnológico (CNPq), Coordenação de Aperfeiçoamento de Pessoal de Nível Superior (CAPES) and Fundação de Amparo à Pesquisa do Estado de São Paulo (FAPESP), Brazil, for partial financial support of this work.

8 REFERENCES Albuquerque, M.P.P. and Caldas, L.V.E., 1989. New ionization chambers for beta and X- radiation. Nucl. Instrum. Meth. Phys. Res. A 280, 310-313. Attix, F.H., 1986. Introduction on radiological physics and radiological dosimetry. 2 nd ed., John Wiley & Sons, New York. Boag, J.W., 1987. Ionization chambers. In: Kase, K.R., Bjärngard, B.E. and Attix, F.H. (Eds.), The Dosimetry of Ionizing Radiation, Vol 2. Academic, New York, 169-243. Costa, A.M. and Caldas, L.V.E., 2003. A special ionization chamber for quality control of diagnostic and mammography X ray equipment. Radiat. Prot. Dosim. 104, 41-45. Dias, S.K. and Caldas, L.V.E., 1998. Development of an extrapolation chamber for the calibration of beta-ray applicators. IEEE Trans. Nucl. Sci. 45, 1666-1669. IAEA 2001. Absorbed dose determination in external beam radiotherapy: an international code of practice for dosimetry based on standards of absorbed dose to water. IAEA TRS 398. Vienna: International Atomic Energy Agency. IAEA 2007. Dosimetry in diagnostic radiology: an international code of practice. IAEA TRS 457. Vienna: International Atomic Energy Agency. IEC 1997. Medical electrical equipment - Dosimeters with ionization chambers and/or semiconductor detector as used in X-ray diagnostic imaging. IEC 61674. Genève: International Electrotechnical Commission. IEC 1994. Medical diagnostic X-ray equipment Radiation conditions for use in the determination of characteristics. IEC 1267. Genève: International Electrotechnical Commission. Maia, A.F. and Caldas, L.V.E., 2005. A new extended-length parallel-plate ionization chamber. Phys. Med. Biol. 50, 3837-3847.

Oliveira, M.L. and Caldas, L.V.E., 2007. Performance of a prototype of na extrapolation minichamber in various radiation beams. Appl. Radiat. Isot. 65, 975-979. 9

10 Figure captions: Figure 1 Diagram of the monitor chamber developed at IPEN. Figure 2 Picture of the new monitor chamber. Figure 3 Saturation curve of the monitor chamber using a diagnostic radiology quality beam (RQR 5). Figure 4 Medium-term stability of the monitor ionization chamber.

Fig. 1 11

Fig. 2 12

13 300 Ionization current (pa) 200 100 0-100 -200-300 -400-200 0 200 400 Voltage (V) Fig. 3

14 1.03 1.02 Relative response 1.01 1.00 0.99 0.98 0.97 0 5 10 15 Measurement Fig. 4.

15 Table 1 Polarity effect of the new monitor ionization chamber. Voltage Collected charge (V) (nc) Ratio +50 / -50 2.77 / 2.69 1.03 +100 / -100 2.75 / 2.72 1.01 +150 / -150 2.77 / 2.76 1.00 +200 / -200 2.80 / 2.78 1.01 +250 / -250 2.81 / 2.79 1.01 +300 / -300 2.82 / 2.82 1.00 +350 / -350 2.82 / 2.80 1.01 +400 / -400 2.84 / 2.82 1.01