Quick Start Guide LIQ-QSG-226, Rev F June Rosemount 226. Toroidal Conductivity Sensors

Similar documents
Rosemount 225. Toroidal Conductivity Sensors. Product Data Sheet July Toroidal Conductivity Sensor LIQ-PDS-225

Submersion/Insertion Toroidal Sensor

Submersion/Insertion Toroidal Sensor

Rosemount TM 498CL ph-independent Free Chlorine Sensor

General Purpose Toroidal Conductivity Sensors

General Purpose Toroidal Conductivity Sensors

Rosemount 5408 and 5408:SIS Level Transmitters

TUpH 396RVP and 398RVP Retractable ph/orp Sensors

Model Q40CT. System Features. Toroidal Conductivity Monitor...

Model 50 Ground Loop ph Sensor Model 55 Solution Ground ph Sensor

TB404 Conductivity sensors. Perfect for use in standard to highly corrosive solutions

Guidelines for Choosing and Installing Radar in Stilling Wells and Bypass Chambers

Signet ph/orp Electrodes General Purpose

Signet ph/orp Electrodes

Type PLC Fitting. Valve for Continuous control. Diaphragm valve. Output

LEVEL TECHNOLOGIES TO POWER THE FUTURE

Best Practices for Power and Transient Protection on Rosemount Radar Transmitters

Magnetic Flowmeter Simulator

The transducer component converts the measured signal and displays the actual value.

HBPH-MK2 Brine/NH3 Leakage Sensor

Mobrey Ultrasonic. MSP422, MSP400RH, and MSP900GH Level Transmitters. Product Data Sheet February 2015 IP2045, Rev EA

Rosemount 333 HART Tri-Loop

SITRANS F flowmeters. SITRANS F System information MAGFLO electromagnetic flowmeters 4/9

Reliance SG777 Series Steel Water Gage Valves

Digital inductive conductivity transmitter

Model 144H NEW. PC-Programmable Temperature Transmitter. Product Data Sheet

B Operating Instructions V1.00/EN/

Signet 2750 DryLoc ph/orp Sensor Electronics

Signet 2751 DryLoc ph/orp Smart Sensor Electronics

Fisher 1052 Size 20 Diaphragm Rotary Actuator with H Mounting Adaptation

Rosemount 3101, 3102, and 3105

Model 8800C and Model 8800A Smart Vortex Flowmeter English Rev. JA

Measuring and Controlling Unit Multronic OC

MJK Expert 700 / 800 / 900 / 1100 / 2100 MJK Expert 1400 / 3400

Flat Sample Holder Part Number

DEUTSCH Instructional Manual for HDP-400 Power Crimper (TE CONNECTIVITY PN )

Rosemount 248 Wireless Temperature Transmitter

Indumax CLS50/CLS50D. Technical Information

Installation, Maintenance and Operating Instructions

Emission-Pak Assemblies

SPES Underwater Eddy Current Proximity Sensor User s Manual P/N M2A-107

Rosemount 5300 Series

Reliance SG800 Series Steel Water Gage Valves

Technical Information Smartec CLD18

Digital conductivity transmitter

Magnetic Inductive Flow Sensor induq

Customer Drawings Product Selection

Instruction MI December Magnetic Flow System

CX105 Conductivity/Resistivity Transmitter

DissolvedOxygenMonitor

Roline L1 Series. Humidity - Temperature Transmitters INSTRUCTION MANUAL

ph or ORP Transmitter

SPES Underwater Eddy Current Proximity Sensor User s Manual P/N M3A-103

[ Rosemount 648 Wireless Temperature Transmitter. Rosemount 648 Wireless. Quick Installation Guide , Rev CA August 2011

Remote Controlled FOUNDATION Fieldbus Two-Wire ph/orp Transmitter

Rosemount TGU 65 Tank Radar Gauge, Still Pipe Antenna

Endress Hauser. SmarTec M CLD 133 Conductivity Measurement

S8000 Series ph/orp Electrode Platform. Self Cleaning Flat Surface ph/orp Electrodes LIQUID ANALYTICAL SOLUTIONS THE NEXT GENERATION

I-VICFLEX.VS1. Victaulic VicFlex Style VS1 Dry Sprinkler WARNING WARNING TECHNICAL DATA BEND CHARACTERISTICS INSTALLATION INSTRUCTIONS SIDEWALL

Technical Information Condumax CLS21D/CLS21

Ensure reliable and repeatable multi-component gas analysis.

TYPE 3740/3740D WET/DRY RUNNING CARTRIDGE SPLIT SEAL

Rosemount 3051 Pressure Transmitter and Rosemount 3051CF DP Flowmeters

KPS Turbidity Meter User s Manual

4973 Series Conductivity Cells Installation and Maintenance Manual Rev 1 6/99

Electromagnetic Flow Transmitter

M1 Series. Humidity - Temperature Transmitter INSTRUCTION MANUAL

2-wire level transmitter

Flow Technologies Overview. For internal use only / Siemens AG All Rights Reserved.

Ricochet Alphasonic Level Transmitter Model LA15/20 Owner s Manual

LF Conductivity Measuring Probe

Condumax CLS21 and CLS21D

1/2 2 (DN 15 50) Standard Bore Series /2 1-1/2 (DN 15 40) Full Bore Series 9000 Flanged Ball Valves with ISO Bonnet

TYPE 3740XL EXTRA LARGE WET RUNNING CARTRIDGE SPLIT SEAL

USER MANUAL. Maxwell Technologies BOOSTCAP Ultracapacitor Energy Storage Modules for Low Duty Cycle Applications

Rosemount Wireless Dam Monitor (WDM)

Rosemount 753R Remote Web Based Monitoring Indicator

ED701 General Industry Pressure Transmitter

Process ph/orp Analysis Systems

11A Series Absolute Pressure Transmitters

Assmann Corporation of America TANK INSTALLATION AND USE GUIDELINES FOR BULK STORAGE TANKS

4.00 ULTRASONIC ANALYZER

INSTALLATION INSTRUCTIONS POLE/PITCH- ADJUSTABLE MOUNT Model: TPM-2000 SERIES

Technical Information Smartec CLD18

SMARTPAT PH 1590 Technical Datasheet

SITRANS F flowmeters. SITRANS F M System information MAGFLO electromagnetic flowmeters. 4/18 Siemens FI

SITRANS F flowmeters SITRANS F M. Flow sensor MAG /48 Siemens FI Overview

LCI. Inductive Conductivity/ Concentration and Temperature Transmitter. Compact or separate version

600T Series Transmitters. Specification Sheet. Model 653T Electronic temperature transmitter (PC programmable)

Flow Measurement SITRANS F M. Flow sensor MAG /49. Overview

Electromagnetic flowmeters and switches DWM 1000/2000

Sight glass window Sight flow indicator

Rosemount 648 Wireless Temperature Transmitter

Broadband Step-Up Transformer. User Manual

Non-Submersible Industrial Pressure Transducer

Troubleshooting RTDs and Thermocouples

Operating instructions OI/FPD350-EN Rev. H. FPD350 Torbar Averaging pitot tubes

FMG70B Series Magnetic Inductive Flow Sensor

Technical Information Smartec CLD18

B Operating Instructions /

Transcription:

Quick Start Guide LIQ-QSG-226, Rev F June 2017 Rosemount 226 Toroidal Conductivity Sensors

Safety Information WARNING! HIGH PRESSURE AND TEMPERATURE HAZARD Before removing the senosr, reduce the process pressure to 0 psig and cool down the process temperature. Failure to reduce the pressure and temperature may cause serious injury to personnel. CAUTION! EQUIPMENT DAMAGE The wetted sensor materials may not be compatible with process composition and operating conditions. Application compatibility is entirely your responsibility.

Contents Contents Chapter 1 Description and Specifications...1 1.1 Overview...1 1.2 Specifications...1 1.3 Unpacking and inspection... 1 1.4 Ordering Information... 2 Chapter 2 Install...3 2.1 Installing the Sensor... 3 2.1.1 Submersion Mounting... 5 2.1.2 Insertion Mounting... 5 2.1.3 Sensor Cable...5 2.2 Wiring the Sensor...6 Chapter 3 Calibration... 11 3.1 Sensor Calibration... 11 3.2 Calibrating against a Standard Solution... 11 3.3 Calibrating against a Referee In Process...13 3.4 Calibrating against a Referee - Grab Sample... 15 Chapter 4 Troubleshooting... 17 4.1 Maintaining the sensor...17 4.2 Troubleshooting... 18 Chapter 5 Accessories... 21 Rosemount 226 Sensors i

Contents ii Quick Start Guide

Description and Specifications 1 Description and Specifications 1.1 Overview The Rosemount 226 sensor is a toroidal (inductive) conductivity sensor. These sensors work well for measuring in highly conductive liquids up to 2 S/cm (2,000,000 μs/cm). Unlike metal electrode based conductivity sensors, toroidal conductivity sensors, like the Rosemount 226, are resistant to fouling, coating, and chemical attack. Sensors are molded with highly corrosion-resistant glass-filled PEEK (polyetheretherketone). The sensors include an integral Pt-100 RTD for temperature compensation. With a large bore hole opening, the Rosemount 226 greatly resists plugging when used in liquids containing high amounts of suspended solids. PEEK is not recommended for greater than 50% concentrations (at 77 F (25 C) ) of H2SO4, HNO3, and H3PO4. PEEK is not recommended for use with HF at all. 1.2 Specifications Table 1-1: Rosemount 226 Toroidal Conductivity Sensor Specifications Description Conductivity Range Wetted Materials Operating Temperature Maximum Pressure Material and Units Refer to transmitter product data sheet Glass-filled PEEK, EPDM Gasket 32 to 248 F (0 to 120 C) 295 psig (2135 kpa [abs]) Cable Length 20 ft. (6.1 m) Maximum Cable Length Up to 200 ft. maximum Process Connections 7/8 in. 9 UNC threads for flange mounting and 1 in. MNPT (with -80 option); see dimensional drawings for more details. Weight/Shipping Weight 2 lbs/3 lbs (1.0 kg/1.5 kg) 1.3 Unpacking and inspection Complete the following steps when you unpack your instrument. 1. Inspect the shipping container. If there is damage, contact the shipper immediately for instructions. 2. If there is no apparent damage, remove the sensor. 3. Ensure that all items shown on the packing list are present. If items are missing, contact your local Customer Care representative. Rosemount 226 Sensors 1

Description and Specifications 4. Save the shipping container and packaging. They can be reused to return the sensor to the factory in case of damage. 1.4 Ordering Information Table 1-2: Model Rosemount 226 Toroidal Conductivity Sensor ordering information Sensor type 226 Toroidal Conductivity Sensor Materials of Construction 02 Glass-filled PEEK (1) Transmitter Compatibility (2) 54 Standard Cable (3) 56 EMI/RFI Shielded Cable (4) Mounting Kit 80 Submersion (5) 81 Insertion Through User-Supplied Flange (6) 82 No Kit Required (7) Typical Model Number: 226-02-56-80 (1) The sensor is supplied with an EPDM gasket (a Viton gasket PN 33151-01 is also available; see accessories). (2) Cables may be extended using the remote junction box PN 23550-00 (sold separately) and extension cables (see accessories). (3) Recommended for use with Rosemount legacy transmitter models 1054 and 2054. (4) Recommended for use with Rosemount transmitter models 54C, 54eC, 81T, 2081T, 3081T, 4081T, XMT, 56, 1056, 1066, 5081, and 56. (5) Includes a 1-in. MNPT PEEK adapter (6) Includes spacer and nut (7) This option does not include any mounting kit and is for replacement sensors only. 2 Quick Start Guide

Install 2 Install 2.1 Installing the Sensor To ensure accurate readings, it is recommended the sensor be installed so that there is at least 2.4 inches of clearance between the sensor and tank or pipe walls. If installed too closely to the walls, an error in readings will be induced by wall effects. Wall effects arise from the interaction between the current induced in the sample by the sensor and nearby pipe or vessel walls. As Figure 2-1 shows, the measured conductivity can either increase or decrease depending on the wall material. This effect can be seen by watching the conductivity readings change as the sensor is moved closer to the sides of the pipe, tank, or beaker. Ensure that the sensor is completely submerged in the process liquid. Mounting the sensor in a vertical pipe run with flow running from bottom to top is recommended. If the sensor must be installed in a horizontal pipe run, mount the sensor in the 3 o clock or 9 o clock position. Figure 2-1: Measured conductivity as a function of clearance between sensor and walls Rosemount 226 Sensors 3

Install Figure 2-2: Rosemount 226 with 1 in. MNPT process connection mounting adapter (-80 option) dimensional drawing Figure 2-3: Rosemount 226 with 7/8 in. 9 UNC thread and insertion through flange mounting adapter (-81 option) dimensional drawing 4 Quick Start Guide

Install 2.1.1 Submersion Mounting The sensor must be mounted in conduit or stand pipe to protect the back end from process leakage. Use Teflon tape for a good seal. 2.1.2 Insertion Mounting The sensor is designed to be mounted through any user-supplied flange. The user is responsible for cutting a hole through the flange to fit the sensor. The flange may be drilled and tapped for the sensor s 7/8 in. 9 UNC thread. Alternatively, a simple 15/16 in. (2.4 cm) drilled hole will accommodate the 7/8-in. 9 UNC thread. 2.1.3 Sensor Cable CAUTION! ELECTRICAL HAZARD Do not run sensor cable in same conduit as the AC power wiring or near heavy electrical equipment. Cables run in the same conduit with power wiring or near heavy electrical equipment may cause measurement errors and damage the sensor. CAUTION! MOISTURE DAMAGE Sensor cables routed in conduit must be sealed or plugged with sealing compound. Failure to properly seal the conduit may allow accumulated moisture in the transmitter housing and damage the sensor and the transmitter. Rosemount 226 Sensors 5

Install 2.2 Wiring the Sensor For additional wiring information on this product, including sensor combinations not shown here, please refer to the Transmitter Wiring Diagrams. Figure 2-4: Wire functions 6 Quick Start Guide

Install Figure 2-5: Wiring for Rosemount 226-54 and Rosemount 226-56 sensors to 1056 and 56 transmitters Rosemount 226 Sensors 7

Install Figure 2-6: Wiring for Rosemount 226-54 and Rosemount 226-56 sensors to 1066 transmitter Figure 2-7: Wiring for Rosemount 226-54 and Rosemount 226-56 sensors to 5081-T transmitter 8 Quick Start Guide

Install Figure 2-8: Wiring sensors through a remote junction box Rosemount 226 Sensors 9

Install 10 Quick Start Guide

Calibration 3 Calibration 3.1 Sensor Calibration The nominal cell constant of the Rosemount 226 sensor is 1.2/cm. The error in cell constant is about ± 10%, so conductivity readings made using the nominal cell constant will have an error of at least ± 10%. Wall effects as shown in Figure 2-1 will likely make the error greater. There are two basic ways to calibrate a toroidal sensor: against a standard solution or against a referee meter and sensor. A referee meter and sensor is an instrument that has been previously calibrated and is known to be accurate and reliable. The referee instrument can be used to perform either an in-process or a grab sample calibration. Regardless of the calibration method used, the connected transmitter automatically calculates the cell constant once the known conductivity is entered. For more detailed information on calibration methods, please refer to Application Data Sheet (ADS 43-025). 3.2 Calibrating against a Standard Solution Calibration against a standard solution requires removing the sensor from process piping. This calibration method is practical only if wall effects are absent or if the sensor can be calibrated in a container identical to the process piping. Ideally, the conductivity of the standard used should be close to the middle of the range that the sensor will be used in. Generally, toroidal conductivity sensors have good linearity and so standards greater than 5000 μs/cm at 25 C may also be used. Prerequisites CAUTION! Before removing the sensor, be absolutely certain that the process pressure is reduced to 0 psig and the process temperature is lowered to a safe level! Immerse the rinsed sensor in the standard solution and adjust the transmitter reading to match the conductivity of the standard. For an accurate calibration several precautions are necessary: Procedure 1. If wall effects are absent in the process installation, use a sufficiently large container for calibration to ensure that wall effects are absent. 2. To check for wall effects, fill the container with solution and place the sensor in the center submerged at least ¾ of the way up the stem. Rosemount 226 Sensors 11

Calibration 3. Note the reading. Then move the sensor small distances from the center and note the reading in each position. The readings should not change. 4. If wall effects are present, be sure the vessel used for calibration has exactly the same dimensions as the process piping. 5. Also, ensure that the orientation of the sensor with respect to the piping is exactly the same in the process and calibration vessels. Figure 3-1: Calibration installation orientation 6. Turn off automatic temperature compensation in the transmitter. This eliminates error in the cell constant. 7. Use a good quality calibrated thermometer to measure the temperature of the standard solution. The thermometer error should be less than ±0.1 C. 8. Allow adequate time for the solution and sensor to reach thermal equilibrium. If the sensor is being calibrated in an open beaker, keep the thermometer far enough away from the sensor so it does not introduce wall effects. 9. If the sensor is being calibrated in a pipe tee or similar vessel, it will probably be impractical to place the thermometer in the standard solution. 10. Instead, put the thermometer in a beaker of water placed next to the calibration vessel. 11. Let both come to thermal equilibrium with the ambient air before continuing calibration. 12 Quick Start Guide

Calibration Figure 3-2: Measuring standard temperature 12. Make sure that the air bubbles are not adhering to the sensor. An air bubble trapped in the toroid opening has a particularly severe effect on the reading. 3.3 Calibrating against a Referee In Process This method involves connecting the process and referee sensors in series and allowing the process liquid to flow through both sensors. The process sensor is calibrated by adjusting the process analyzer reading to match the conductivity measured by the referee instrument. Prerequisites For a successful calibration, several precautions are necessary: Procedure 1. If possible, adjust the conductivity of the process liquid so that it is near the midpoint of the operating range. If this is not possible, adjust the conductivity so that it is at least 5000 μs/cm. 2. Orient the referee sensor so that air bubbles always have an easy escape path and cannot get trapped. Rosemount 226 Sensors 13

Calibration Figure 3-3: Calibration with a referee instrument example 3. Tap and hold the flow cell in different positions to allow bubbles to escape. 4. Turn off automatic temperature compensation in the transmitter. This eliminates error in the cell constant. 5. Keep tubing runs between the sensors short and adjust the sample flow to as high a rate as possible. Short tubing runs and high flow ensure that the temperature of the liquid does not change as it flows from one sensor to another. 6. Wait for readings to stabilize before starting the calibration. 14 Quick Start Guide

Calibration 3.4 Calibrating against a Referee - Grab Sample This method is useful when calibration against a standard is impractical or when in-process calibration is not feasible because the sample is hot, corrosive, or dirty, making handling the waste stream from the referee sensor difficult. Prerequisites The method involves taking a sample of the process liquid, measuring its conductivity using a referee instrument, and adjusting the reading from the process analyzer to match the measured conductivity. Prerequisites For a successful calibration, several precautions are necessary: Procedure 1. If possible, adjust the conductivity of the process liquid so that it is near the midpoint of the operating range. If this is not possible, adjust the conductivity so that it is at least 5000 μs/cm. 2. Take the sample from a point as close to the process sensor as possible. Be sure the sample is representative of what the sensor is measuring. 3. Keep temperature compensation with the transmitter turned on. 4. Confirm that the temperature measurements in both the process and referee instruments are accurate, ideally to within ±0.5 C. 5. Wait until readings are stable before starting the calibration. Rosemount 226 Sensors 15

Calibration 16 Quick Start Guide

Troubleshooting 4 Troubleshooting 4.1 Maintaining the sensor Generally, the only maintenance required is to keep the opening of the sensor clear of deposits. Cleaning frequency is best determined by experience. CAUTION! Make sure that the sensor is cleaned of process liquid before handling. Rosemount 226 Sensors 17

Troubleshooting 4.2 Troubleshooting Table 4-1: Troubleshooting Problem Probable cause Solution Off-scale reading Noisy reading Reading seems wrong (lower or higher than expected) Sluggish response Wiring is wrong. RTD is open or shorted. Sensor is not in process stream. Sensor is damaged. Sensor is improperly installed in the process stream. Sensor cable is run near high voltage process stream. Sensor cable is moving. Bubbles are trapped in the sensor, particularly in the toroid opening. Sensor is not completely submerged in the process stream. Cell constant is wrong. Wall effects are present. Wrong temperature correction algorithm is being used. Verify and correct wiring. Check the RTD for open or short circuits. See Figure 4-1. Confirm that the sensor is fully submerged in the process stream. See Section 2.1. Perform isolation checks. See Figure 4-1. Confirm that the sensor is fully submerged in the process stream. See Section 2.1. Move cable away from high voltage conductors. Keep sensor cable stationary. Install the sensor in a vertical pipe run with the flow against the toroidal opening. Incresase flow if possible. Confirm that the sensor is fully submerged in the process stream. See Section 2.1. Calibrate the sensor in place in the process piping. See Section 3.1. Check that the temperature correction is appropriate for the sample. See transmitter manual for more information. Temperature reading is inaccurate. Disconnect the RTD leads (Figure 4-1 ) and measure the resistance between the in and common leads. Resistance should be close to the value in Table 4-2. Slow temperature response to sudden changes in temperature. Sensor is in a dead area in the piping. Slow temperature response to sudden changes in temperature. Use an RTD in a metal thermowell for temperature compensation. Move sensor to a location more representative of the process liquid. Use an RTD in a metal thermowell for temperature compensation. 18 Quick Start Guide

Troubleshooting Table 4-2: Resistance vs. Temperature for Temperature Compensation (PT-100 RTD) Temperature Resistance 10 C (50 F) 103.9 Ω 20 C (68 F) 107.8 Ω 25 C (77 F) 109.7 Ω 30 C (86 F) 111.7 Ω 40 C (104 F) 115.5 Ω 50 C (122 F) 119.4 Ω Figure 4-1: Sensor resistance check Rosemount 226 Sensors 19

Troubleshooting 20 Quick Start Guide

Accessories 5 Accessories Part number Description 2001492 Stainless steel tag (must specify marking) 23550-00 Remote junction box without preamplifier 23294-00 Interconnecting extension cable, unshielded (for use with remote junction box) 23294-05 Interconnecting extension cable, shielded (for use with remote junction box) 33151-00 EPDM gasket 33151-01 Viton gasket 33185-01 Submersion mounting adapter, 1 in. MNPT, 3 in. length, PEEK (spare for -80 option sensors) 33185-02 Flange insertion mounting adapter, 1 in. length, PEEK (spare for -81 option sensors) 33219-00 7/8 in. 9 UNC X 1 in. MNPT for conduit connection (spare for -81 option sensors) Rosemount 226 Sensors 21

Accessories 22 Quick Start Guide

Accessories Rosemount 226 Sensors 23

LIQ-QSG-226 Rev F 2017 www.emerson.com/rosemountliquidanalysis Emerson Automation Solutions 8200 Market Blvd Chanhassen, MN 55317 Toll Free +1 800 999 9307 F +1 952 949 7001 liquid.csc@emerson.com www.emerson.com/rosemountliquidanalysis EUROPE Emerson Automation Solutions Neuhofstrasse 19a P.O. Box 1046 CH-6340 Baar Switzerland T + 41 (0) 41 768 6111 F + 41 (0) 41 768 6300 liquid.csc@emerson.com www.emerson.com/rosemountliquidanalysis MIDDLE EAST AND AFRICA Emerson Automation Solutions Emerson FZE Jebel Ali Free Zone Dubai, United Arab Emirates, P.O. Box 17033 T +971 4 811 8100 F +971 4 886 5465 liquid.csc@emerson.com www.emerson.com/rosemountliquidanalysis ASIA-PACIFIC Emerson Automation Solutions 1 Pandan Crescent Singapore 128461 Singapore T +65 777 8211 F +65 777 0947 liquid.csc@emerson.com www.emerson.com/rosemountliquidanalysis 2016 Rosemount. All rights reserved. The Emerson logo is a trademark and service mark of Emerson Electric Co. Rosemount is a mark of one of the Emerson Process Management family of companies. All other marks are the property of their respective owners. The contents of this publication are presented for information purposes only, and, while effort has been made to ensure their accuracy, they are not to be construed as warranties or guarantees, express or implied, regarding the products or services described herein or their use or applicability. All sales are governed by our terms and conditions, which are available on request. We reserve the right to modify or improve the designs or specifications of our products at any time without notice.