TRANSMITTER CALIBRATION

Similar documents
TI25 - Pre-Instructional Survey

Department of Mechanical and Aerospace Engineering. MAE334 - Introduction to Instrumentation and Computers. Final Examination.

BASIC TECHNIQUES FOR ACCURATE RESISTANCE MEASUREMENT WHITE PAPER. Reduce Measurement Errors in your Application

WATER METERS M SERIES WATER METERS INSTALLATION, OPERATION AND MAINTENANCE GUIDE METAL M SERIES WATER METER PLASTIC M SERIES WATER METER

HUMITRAN-RHT & RH RELATIVE HUMIDITY/TEMPERATURE TRANSMITTER RELATIVE HUMIDITY TRANSMITTER

EE 241 Experiment #4: USE OF BASIC ELECTRONIC MEASURING INSTRUMENTS, Part III 1

Exercise 2: Temperature Measurement

EE283 Laboratory Exercise 1-Page 1

Introduction to Electronic Equipment

Group: Names: Resistor Band Colors Measured Value ( ) R 1 : 1k R 2 : 1k R 3 : 2k R 4 : 1M R 5 : 1M

Differential Amplifier : input. resistance. Differential amplifiers are widely used in engineering instrumentation

MARMARA UNIVERSITY CSE315 DIGITAL DESIGN LABORATORY MANUAL. EXPERIMENT 7: Analog-to-Digital Conversion. Research Assistant Müzeyyen KARAMANOĞLU

Exercise 3: Voltage in a Series Resistive Circuit

To Calibrate or Not to Calibrate a fieldbus transmitter? Dale Perry Pressure Marketing Manager Rosemount

EE 448 Fall Lab Experiment No. 3 04/04/2008. Transformer Experiment

Laboratory Exercise - Seven

Pre-Lab for Batteries and Bulbs

Ohm s Law and Electrical Circuits

Configurations of Resistors

The Art of Electrical Measurements

Electric Transformer. Specifically, for each coil: Since the rate of change in flux through single loop of each coil are approximately the same,

Process Calibrator. TechChek 820

PAD - Heavy Duty Differential Pressure Trasmitter

DSTS-3B DEPTHSOUNDER TEST SET OPERATOR S MANUAL

MFJ ENTERPRISES, INC.

Ohm s and Kirchhoff s Circuit Laws. Abstract. Introduction and Theory. EE 101 Spring 2006 Date: Lab Section #: Lab #2

April 1994 UCM-420A. Setpoint Controller. Operating and Installation Instructions

Resistance and Ohm s law

Series APD4059. For Load Cells/Pressure Transducer Transmitters. Field Rangeable, with Calibration Resistor. Standard Features.

KING-GAGE KING. ES2 Liquid Level Transmitter. Installation and Operation Instructions ma Liquid Level Transmitter EX-1834

User manual Firmware V1.0-V1.2 HTB230. In-head transmitter

5. Transducers Definition and General Concept of Transducer Classification of Transducers

Wiring and Measuring 4-20mA Instruments

Exercise 2: Ohm s Law Circuit Current

Process Multimeter DPM 79

Using Circuits, Signals and Instruments

Economical Digital Indicators For Process and Electrical Measurement

Isolated, Linearized RTD Input 7B34 FEATURES APPLICATIONS PRODUCT OVERVIEW FUNCTIONAL BLOCK DIAGRAM

W100/W600/W900 Controller ph Electrode Troubleshooting Guide

INSTALLATION & OPERATION MANUAL

PHYS 1112L - Introductory Physics Laboratory II

DRG-SC Series Signal Conditioners

User s Guide FLSC-45/45B. Paddlewheel Signal Conditioner. Shop online at omega.com

ECE 317 Laboratory #1 Force Sensitive Resistors

Instrument Usage in Circuits Lab

Basic Electronics. Chapter 2 Basic Electrical Principles and the Functions of Components. PHYS 401 Physics of Ham Radio

CA330 RTD Calibrator: High-speed Response and High-resolution Resistance Simulator

ANALOG SCALING MODULE

Operating Manual. Differential Pressure Transmitter DPS 100

SHRI ANGALAMMAN COLLEGE OF ENGINEERING & TECHNOLOGY (An ISO 9001:2008 Certified Institution) SIRUGANOOR,TRICHY

PIE 525. Diagnostic Thermocouple RTD & Milliamp Calibrator Operating Instructions. Practical Instrument Electronics

MEASUREMENTS & INSTRUMENTATION ANALOG AND DIGITAL METERS

SIGNAL INTERFACE MODULE

Princo MagneLevel Model LML

MFJ ENTERPRISES, INC.

User manual V1.0 RTB231 RTB232. RTD transmitter

Laboratory 2 (drawn from lab text by Alciatore)

SENSOR AND MEASUREMENT EXPERIMENTS

CEC VIBRATION TRANSMITTER

Temperature Transmitters - Head Mounted

Current, resistance, and Ohm s law

Lab 11: Circuits. Figure 1: A hydroelectric dam system.

AMERITRON SDC-102 Screwdriver Antenna Controller

KING-GAGE. D/P Transmitter 4-20 ma Pressure Transmitter ma Pressure Transmitter. Operation/Calibration Manual

CX105 Conductivity/Resistivity Transmitter

INSTALLATION & OPERATION MANUAL

MEASUREMENT AND INSTRUMENTATION QUESTION BANK UNIT I INTRODUCTION. Part A

Motomatic Servo Control

50 Ω Dummy Load for Transmitter testing and experiments

Installation and Operating Manual

Basic Electronics. Chapter 2, 3A (test T5, T6) Basic Electrical Principles and the Functions of Components. PHYS 401 Physics of Ham Radio

EKT 314/4 LABORATORIES SHEET

December 2014 Ed 21. Testing and Recalibration

CPE 100L DIGITAL LOGIC DESIGN I DESIGN LABORATORY LABORATORY 1 LAB SAFETY QUIZ & LAB EQUIPMENT USE TUTORIAL UNIVERSITY OF NEVADA, LAS VEGAS GOALS:

NOVATO (MAXREFDES16#): 4-20MA LOOP-POWERED TEMPERATURE SENSOR WITH HART

LABORATORY MODULE. Analog Electronics. Semester 2 (2005/2006)

RTD and thermocouple circuits, with millivolt calculations

STATION NUMBER: LAB SECTION: RC Oscillators. LAB 5: RC Oscillators ELECTRICAL ENGINEERING 43/100. University Of California, Berkeley

Lab Exercise # 9 Operational Amplifier Circuits

Technician License Course Chapter 3. Lesson Plan Module 4 Electricity

// Parts of a Multimeter

Interfacing the 1250B, 1250LTC and 1511-LTC Monitors to the Beckwith Tapchanger Controller

EELE 354 Lab Assignment 2: Electric Heater - Power Measurements and Considerations

KING-GAGE. D/P Transmitter 4 20 ma Pressure Transmitter ma Pressure Transmitter. Operation/Calibration Manual

MFJ-219/219N 440 MHz UHF SWR Analyzer TABLE OF CONTENTS

PHYS 1112L - Introductory Physics Laboratory II

Curriculum. Technology Education ELECTRONICS

Lab 3: Digital Multimeter and Voltage Generator

Module 1, Lesson 2 Introduction to electricity. Student. 45 minutes

ELE.B: Original Assignment Resistors in Series Classwork Homework

DC Electric Circuits: Resistance and Ohm s Law

= 7 volts Copyright , R. Eckweiler & OCARC, Inc. Page 1 of 5

KING-GAGE ES2. Installation and Operation Instructions Manual. Electronic Liquid Level Transmitter. ESR Serial No. Versions and Slimline Versions

Experiment 1: Instrument Familiarization (8/28/06)

Loop calibration and maintenance

OPTITEMP TT 20 Technical Datasheet

Ultra Precise Immersion RTD Sensors


Series and Parallel Resistors

WATT TRANSDUCER. Instruction Manual

Transcription:

TRANSMITTER CALIBRATION In this laboratory study, you will be asked to calibrate two of four different devices. To do so, you will need to write the equation defining the performance of the unit. Use this equation to predict the output of the unit with a known value of input. The transmitters under study are: TCS Model TX1504 temperature transmitter calibrated for an RTD input. The input range is either 50150 o F or 40240 o F. The output range is 420 ma. This is a loop powered current device requiring an input voltage of 835 VDC. Contractor Instruments Model P855 static pressure transmitter. Input range is either 0 to 3 in w.g. or 0.25 to 1.75 in w.g. The output range is 4 to 20 ma. This is a loop powered current device requiring an input voltage of 840 VDC. Omega Model FLSC18B flow transmitter designed for use with any number of their flow sensors. This particular transducer is a loop powered current device with a 4 to 20 ma output. The input signal is the frequency of a magnetic pulse as output from the flow sensor. PROCEDURE In all cases, you must first be able to determine the output of the device for a given input. If the manufacturer does not provide a table, you can develop the equation for the device. The equation should be written so as to predict the transmitter output with a known input of simulated temperature, pressure or flow. TCS Model TX1504 The input to the transmitter will be a resistance. The attached table correlates resistance to temperature for a 1000 Ω sensor. Select two appropriate resistors. Do not rely on the stated rating of the resistor. Use a multimeter to measure its resistance. Once the actual resistance of each is measured, determine what temperature this represents. For example, assume you selected a resistor with a resistance of 1130.5 Ω. Referring to the attached resistance table, note this falls between 92 o F and 93 o F. Interpolating, we find: 1130.5 1129.662 * ( 93 92) 92 = 92.39 1131.812 1129.662 Using the equation you developed, you can now predict the output of the transmitter. o F If a desired resistance cannot be found, you can parallel two resistors to obtain that desired resistance. To determine the resistance of two resistors in parallel, remember that: 1 1 1 = R R R 1 2 where: R=desired resistance R 1 = resistance of first resistor R 2 = resistance of second resistor Connect the device as shown below. Ensure the device is properly configured for a 1000 Ω sensor and a 200 o F temperature span as shown in the attached specification sheet. Your instructor will inspect the wiring prior to switching on the power supply.

Power Supply Milliamp Meter R W B SENSOR Signal is B to W Terminal R is compensating Note: Replace sensor with a resistor during calibration. With the power supply off, connect the resistor of the lowest value to the input terminals (B to W) or (B to R) of the transducer. Before switching on the power supply, ensure the proper voltage range is selected and the voltage output is turned down. Turn the voltage up on the power supply while monitoring it with the builtin voltmeter or a handheld multimeter until you read 24 VDC. Current output will read on the analog milliamp meter. Adjust the zero pot on the transducer until you read your desired milliamp output. Now switch the power off, replace the resistor with a higher value and turn the power back on. Adjust the span pot to read the desired value of current output. Repeat the procedure by switching resistors and adjusting zero and span until they settle out to the desired values. The transducer is now calibrated. Check your calibration by connecting any value of resistance that falls within the specified input range and observe the output. Is this what you expect? CAUTION We have found the TCS transducers often calibrate just the opposite of what is normally expected. In other words, you may need to adjust the zero pot with the high resistance connected and the span pot with the low resistance connected. CONTRACTOR INSTRUMENTS Model P855 The Contractor Instruments static pressure transmitter accepts wither a 03 in w.g. or 0.251.75 in w.g. static pressure as stated on the nameplate. The output is 420 ma. Connect the transmitter as shown in the diagram below. Apply Pressure Here Power Supply LO HI Manometer Milliamp Meter Pressure input will be measured with an Airflow Manometer. Connect a tube to the high port of the transmitter and the manometer so they both read static pressure. Be extremely cautious of attaching and removing tubes from the sensor. The barbs are plastic and break easily. By blowing into the tube, apply a pressure to the manometer/transmitter close to the upper end of its range and pinch it off with a pair of vice grips. Read the pressure on the manometer and the current from the milliamp meter. Based

on the equation you developed for this unit, determine what current you should be reading. Adjust the span pot as necessary to read the proper voltage. Now release the pressure. This should be zero pressure. Read the current. If it is not what you expect, adjust the zero pot. Repeat the procedure by alternating between zero pressure and any other pressure (close to the upper end of the sensor span) until the readings settle out to what you ve calculated. The transmitter is now calibrated. Check your calibration by applying any pressure within the range of the transmitter and observing the current. Is the current what you expect? OMEGA Model FLSC18B The input to the transducer will be a pulse at a certain frequency. The attached table correlates pulse frequency with flow in gpm. Select two appropriate values of input frequency. Determine the appropriate gpm flow these values represent and the current output of the transducer. Connect the device as shown below. The 180 Ω resistor simulates the receiver display or the controller. 1 2 Leads FUNCTION GENERATOR Power Supply 180 Ω 3 4 5 FREQUENCY COUNTER Milliamp Meter Select a square wave output on the function generator and adjust the output to your desired frequency. Use the frequency counter to measure the frequency output of the function generator. Set the frequency counter for DC Coupling, 20X attenuation, and LP (Low Pass) filter on. These settings will filter out the electrical noise within the lab as a result of the fluorescent lighting and the effects of the local radio stations (RF Interference). With the output of the function generator on low frequency, measure the output of the transmitter on the milliamp meter. Adjust the zero pot if necessary. Change the output of the function generator to your desired high frequency. Measure the output of the transmitter. Adjust the span as necessary to obtain the desired output. Repeat the procedure by alternating between high and low frequencies until the readings settle out to what is calculated. The transmitter is now calibrated. Check your calibration by applying any other frequency within the range of the transmitter and observing the output. Is the output what you expect?