RA Duplicating Film X-OMAT 5000 RA Processor; RP X-OMAT Chemicals; 90 Second Cycle; Diffuse Visual Densitometry. Density

Similar documents
KODAK X-OMAT BT Film / 4530 / XBT

KODAK MIN-R S Film / 4906

KODAK Medical X-ray Film / General Purpose / Blue / 3528 / MXB

KODAK EKTASCAN IR Laser Imaging Film / 1356 / EIR

KODAK Simulation Film / 7157

KODAK INDUSTREX AA400 Film Versatility is the key.

EASTMAN TRI-X Reversal Film 7278

KODAK Panchromatic Separation Film 2238

KODAK PROFESSIONAL PRO IMAGE II Paper

Guide to Manual Processing of NDT Films

KODAK EKTACHROME 64T Professional Film

EASTMAN PLUS-X Reversal Film / 7276

KODAK EKTACHROME RADIANCE III Paper

Multiple Choice Identify the letter of the choice that best completes the statement or answers the question.

KODAK PROFESSIONAL ELITE Chrome 200 Film

KODAK EKTACHROME 64 Professional Film

KODAK EKTACHROME 100 Plus Professional Film

KODAK PROFESSIONAL T-MAX P3200 Black & White Negative Film

AVITONE PD 1 P-OS AND PD 3 P-OS

KODAK PROFESSIONAL T-MAX P3200 Black & White Negative Film

TECHNICAL DATA / COLOR REVERSAL FILM January 1998 E-154

KODAK PROFESSIONAL Display and Print Materials

KODAK VERSALITE TDF Film

TECHNICAL DATA / COLOR PAPER July 2003 E-19

KODAK EKTACHROME 64 Professional Film

KODAK IMAGELITE ESY Scanner Film KODAK GEN 5 GAI Recording Film KODAK IMAGELITE IRF Scanner Film KODAK GEN 5 GIR Recording Film

TECHNICAL DATA / COLOR PAPER July 2003 E-23

Computer Output Microfilm Data Sheet

EASTMAN EXR 200T Film 5287, 7287

KODAK PROFESSIONAL EKTACHROME Films E100G and E100GX Grain as Fine as Your Art.

KODAK PROFESSIONAL TRI-X 320 and 400 Films

ELITE Chrome 100 Film

5 Using KODAK EKTACOLOR RA Chemicals with Tray, Drum, and Rotary-Tube Processors

KODAK Infrared AEROGRAPHIC Film 2424

Panchromatic negative film for aerial photography

KODAK EKTACHROME 160T Professional Film / EPT

EASTMAN EXR 500T Film 5298

TECHNICAL DATA / COLOR NEGATIVE FILM February 1998 E-44

IMAGELINK Reference Archive Media

KODAK AEROCHROME III MS Film 2427

Using KODAK Control Strips for Processes AR-5 and AN-6

KODAK PROFESSIONAL ENDURA Transparency and Clear Display

Camera Negative Microfilm Data Sheet

EASTMAN EXR 200T Film / 5293, 7293

KODAK VISION Expression 500T Color Negative Film / 5284, 7284

KODALITH Ortho Films 2556,6556, Type 3

1 MPTVI DATA SHEET XXXXXXXXXXX

ELITE Chrome 400 Film

KODAK High Definition 400 Film

KODAK PROFESSIONAL KODABROME II RC Paper

KODAK High Speed Infrared Film

MINIMUM REQUIREMENTS FOR A DARKROOM.

Monitoring Process RA-4 for Silver-Halide Digital Printers

KODAK PROFESSIONAL PORTRA ENDURA Paper and. KODAK PROFESSIONAL SUPRA ENDURA Paper TECHNICAL DATA / COLOR PAPER. September 2008 E-4021

EASTMAN EKTACHROME High Speed Film (Tungsten) 7250

KODAK PROFESSIONAL PORTRA ENDURA Paper and KODAK PROFESSIONAL SUPRA ENDURA Paper

KODAK TRI-X Pan and KODAK TRI-X Pan Professional Films

KODAK ROYAL Digital Paper

Preparing Smaller-Than-Package-Size Amounts of KODAK Processing Chemicals

KODACHROME 25, 64, and 200 Professional Film

KODAK PROFESSIONAL ENDURA Transparency and Clear Display

BLACK AND WHITE, ORTHOCHROMATIC COPY FILM

KODAK PROFESSIONAL ENDURA Metallic Paper

FUJICOLOR CRYSTAL ARCHIVE PAPER SUPREME HIGH DEFINITION

Panchromatic negative film for aerial photography

KODAK PROFESSIONAL EKTACHROME Film E100VS

KODAK PROFESSIONAL AZO Paper

KODAK PROFESSIONAL T-MAX Films

TECHNICAL DATA / COLOR TRANSPARENCY FILM

400BW TECHNICAL DATA / BLACK-AND-WHITE FILM. BLACK and WHITE FILM. February 2001 F-4012

Kodak Thermal Print Microfilms

Anatomix RP X-Ray System

Diazo C Microfilm 1957, 2957, 3957 Diazo D Microfilm 1956, 2956, 3956

Processing and. Photography. Printing

AVIPHOT ASP 400 S PE1/PE0

KODAK PROFESSIONAL POLYCONTRAST III RC Paper

ISO 400/27 o BLACK AND WHITE PROFESSIONAL FILM FOR HIGH PRINT QUALITY AND FLEXIBILITY IN USE

INDUSTREX Film. For Non-Destructive Testing WHEN CLARITY IS CRITICAL

RPF & OPF Technical Information

KODAK PROFESSIONAL EKTACHROME. Films E100S and E100SW TECHNICAL DATA / COLOR REVERSAL FILM KODAK PROFESSIONAL EKTACHROME DESRIPTION FILM E100S

Compliance Guidance for RADIOGRAPHIC QUALITY CONTROL (5 th Edition)

KODAK PROFESSIONAL ELITE COLOR Films / 200 and 400

Anatomix Plus X-Ray System

KODAK PROFESSIONAL EKTACHROME Film E100VS

KODAK VERICOLOR III Professional Film

AF3-024E COLOR REVERSAL FILMS 1. FEATURES AND USES 2. SPEED

400BW TECHNICAL DATA / BLACK-AND-WHITE FILM. BLACK and WHITE FILM. July 2003 F-4012

PRODUCT INFORMATION BULLETIN

Process Monitoring: Sensitometric Parameters

8 PROCESS CONTROL. Monitoring the Process with KODACHROME 64 Control Film

Using the KODAK Negative Film Processing Cartridges FC1 and FC2

KODAK PRIMETIME 640T Teleproduction Film / 5620,7620

DEVELOPMENT. The following developers are recommended: Ilford ID-2 Developer For Fine Grain Safety Positive Film.

KODAK T-MAX 100 Direct Positive Film Developing Outfit

TECHNICAL INFORMATION

KODAK VISION3 Color Digital Intermediate Film 5254 / 2254

ISO 400/27º, BLACK AND WHITE PROFESSIONAL FILM FOR HIGH PRINT QUALITY AND FLEXIBILITY IN USE

RD-90 Developer-Replenisher F-Fix Fixer-Replenisher

FUJICOLOR CRYSTAL ARCHIVE PAPER TYPE CA

KODAK T-MAX Professional Films

Transcription:

Technical Data Sheet Issued 2014-02 RA Duplicating Film / PicturePerfect Duplicating Film RA Duplicating Film provides excellent image quality and accurate information due to its sensitometric curve shape. The direct reversal photographic emulsion used in this film is coated on one side of the approximately 0.2 mm (7- mil) polyester support. The support is blue-tinted. The finished sheet has a triple V notch and rounded corners to facilitate handling. It has an anti-curl, anti-halation, gloss-reduction coating on the side opposite the emulsion. The film is to be viewed through the anti-halation side (the anti-halation side faces the viewer). This film is designed to be processed with standard or rapid processing cycles using RP X-OMAT, X-OMAT EX II or X-0MAT MX chemicals. It may be necessary to make adjustments and/or changes to the duplicators and printers that will be using this product to optimize results. Changes include: changing the bulb(s), increasing/decreasing exposure time, and on some units, changing the slot width. These changes can be made in the darkroom or office. Please follow the manufacturer's recommendations for all changes.. Sensitometric Properties: RA Duplicating Film X-OMAT 5000 RA Processor; RP X-OMAT Chemicals; 90 Second Cycle; Diffuse Visual Densitometry Speed Measured at 1.0 OD above Gross Fog Contrast Measured as slope of the straight line portion of the sensitometric curve, and computed as the value for the rise for any three consecutive steps. Gross Fog Density of film base plus processing fog. Recommended Starter Volumes Density 4 3 2 Developer RP, EX II, X-OMAT MX Starter (Added to processor developer tank 89 ml (3 fl. Oz.) per 3.78 Litres (1 gallon) 1 0-4 -3-2 -1 0 Relative Log Exposure Notice: The data in this publication represent product tested under the conditions of exposure and processing specified. They are representative of production coatings, and therefore do not apply to a particular box or roll of photographic material. They do not represent standards or specifications that must be met by Carestream Health, Inc. The company reserves the right to change and improve product characteristics at any time. 1/5

Automatic Processing Recommendations: In general, processing is recommended in X-OMAT, RP X-OMAT and MIN-R Mammography Processors using RP X-OMAT, X-OMAT EX II or X-OMAT MX Developer and Replenisher and RP X-OMAT LO or X-OMAT MX Fixer and Replenisher. To minimize processing defects, insert the film into the processor emulsion side up (lighter colored side). Influence of developer temperature in case of automatic processing -2 ºC Ref +2 ºC 0 Base fog 0 0 Sensitivity 0 0 Contrast 0 Replenishment Rate Recommendations for X-OMAT, RP X-OMAT and MIN-R Mammography Processors (Replenishment by length) Film Size Processed 35 x 35 cm (only) Average size intermix 35 x 43 cm (only) Use Condition Average Number of Films per 8 hours Replenishment Rates (ml per 35 x 43 cm) processor operation Developer Fixer 90 sheets or more 30 90 sheets 30 sheets or less* 115 sheets or more 40 115 sheets 40 sheets or less* 75 sheets or more 25 75 sheets 25 sheets or less* 50 65 50 65 60 70 70 120 *If sensitometry does not stay within control limits, flooded replenishment may be needed. Please refer to Service Bulletin No. 30, available on the Carestream website or upon request, for additional processing recommendations. Exposure Guidelines Orientation for Exposure The emulsion side of the radiograph to be duplicated (the original) must be placed in intimate contact with the emulsion side of the duplicating film. (The emulsion side is face-up when the notch is in the upper right corner of the sheet or when the lighter colored side of the film, identified under safelight conditions, is up.) This "sandwich" must be oriented so that the original is closest to the light source and the emulsion side of the duplicating film faces the exposing light. Maintain intimate contact between the original radiograph and the duplicating film through use of a printing frame, a contact printer, or a similar device. Recommended Exposing Devices and Light Sources This film is designed for use with an ultraviolet exposure source and to function properly at an optimum exposure setting. Once this setting is found for a typical, commonly used original type, the exposing equipment can generally be left at that setting for subsequent films. The density of a duplicate can be increased (made darker) by reducing the exposure setting and can be decreased (made lighter) by increasing the exposure setting. Factors that can cause variation in the exposure of the duplicate even if the device setting is kept constant are: bulb age, amount of bulb warm-up time, bulb drift, timer drift, calibration, inconsistent slit width, etc. An optimum exposure setting is defined as one which results in the same density being produced on a duplicate at a specific point as occurs on the original at that same point. The recommendation is to pick a point around a density of 1.1 to 1.2 on the original and vary the exposure setting on the exposing device until that same density is matched on the duplicate. While this should give the most accurate duplicate, some customers may prefer darker or lighter films and the exposure setting can be adjusted according to individual customer preference. The following examples are suggested starting points for making duplicates of medical radiographs. Individual devices should be optimized and adjusted according to the guidelines given: 2/5

BLU-RAY DUPLICATOR: MarkII MarkIV MarkV Bulb Type GE BLB GE BLB GE BLB Filter Wrap None None None 1 Slot Width 0.381 cm (0.15 inches) 0.635 cm (0.25 inches) 0.381 cm (0.15 inches) Intensity/Speed 8 7 195.6 cm/min 2 (77 inches/min) Time Units --- --- 9-10 seconds 1 Filter must be removed. 2 It should take 9.4 seconds for a film sheet to travel 30.5 cm (12 inches). Since the film is slower, it is advisable to make several one time only changes to the duplicators. These include changing the bulb, exposing time, and/or slit width. We recommend following the manufacturers instructions for changing the bulb and slit width. DUPONT DUPLICATOR: Cronex Bulb Type Filter Wrap Slit Width Intensity/Speed (2)GE BLB None N/A N/A Time Units 3.6 BYERS DUPLICATOR: Lab 315A Lab 355A Bulb Type (2)GE BLB GE BLB Filter Wrap None 10 % Filter Slit Width N/A N/A Intensity/Speed N/A N/A Time Units 39 Units 37 Units 3/5

TECHNO-AIDE DUPLICATOR: 195D 360D 450D 2500D/S Bulb Type GE BLB GE BLB GE BLB (2) GE BLB Filter Wrap 0.635 cm (1/4 inch) Slot Width N/A N/A N/A N/A Intensity/Speed N/A N/A N/A N/A Time Units 3 seconds 8 seconds 3 seconds 3.2 seconds NOTE: Since following conditions and BLB light intensity vary, it may be necessary to make further minor adjustments to the exposure/intensity settings. Storage and Handling Storage - Unexposed: 10 24 C (50 75 F) Do not refrigerate or freeze as this can cause condensation to occur. 30 50 %RH Protect from heat and radioactive sources. Film is to be properly shielded from x-rays, gamma rays, or penetrating radiation. Exposed: Processed: Keep cool, dry, and properly shielded from penetrating radiation. Process as soon as possible. 16 27 C (60 F), 30 50 %RH The film should be used before the expiration date indicated on the box with the lot (emulsion) number. Handling - Hands must be clean, dry and free of lotions, etc. Film should be handled carefully by the edges to avoid physical strains such as pressure, creasing, or buckling. Luminous watches, cell phone and darkroom light leaks should be avoided. Do not re-use. Film is a single use medical device. Safelight Filter Use a safelight filter such as a GBX-2 Safelight Filter / ruby red, a 6B Safelight Filter / brown (for European use only), or a OA Safelight Filter / greenish yellow, with a 15-watt bulb located at least 1.22 metres (48 inches) from the film. 4/5

The contents of this publication are subject to change without notice. Carestream Health France 1, rue Galilée 93192 NOISY-LE- GRAND CEDEX FRANCE Carestream Health, Inc., 2014. CARESTREAM, MIN-R and X-OMAT are trademarks of Carestream Health.. Carestream Health, Inc. 150 Verona Street - Rochester, NY, USA 14608 Carestream Health, Inc. 2014, Publication No. CHSP-8606, rev. 2 5/5