What tests should be done?

Similar documents
Electrosurgery Units (ESU s) and their variants

Electrosurgery Units (ESU s) and their variants. D. J. McMahon rev cewood

Andy Falconer. Principles of surgical diathermy

Application Notes. Electrosurgical Analyzer Primer

RF leakage current in electrosurgical units: Influence of the layout in taking measurements

ESU-2000 Series Product Overview

perioperativecpd.com The basics of diathermy

Electrosurgical Generator and Smoke Evacuation System Service Manual. Utah Medical Products, Inc. Catalog Numbers ESU-110 & ESU-220

RAISING THE LEVEL OF PROTECTION IN YOUR ORS

Rigel 377. Instruction Manual 367A550 Issue 2.0. Electrosurgical Analyser. August Seaward Group Issue 2.0

White Paper: Technology and Clinical Overview of the Genii TouchSoft Coagulator monopolar probe

ALSATOM SU MPC SU 50 MPC SU 100 MPC SU 140 MPC SU 140/D MPC SU 140/BD MPC

Know and understand you and your co-workers competence level before opening a live panel to diagnose

EXCELL NHP/T. A new step AheAd to the future ELECTROSURGICAL UNITS FOR MONOPOLAR AND BIPOLAR USE ISO ALSA IT ISO

MICROWAVE DIATHERMY AND SURGICAL DIATHERMY DIATHERMICS

Rigel Medical 24 Month Warranty Statement

U.S. ARMY MEDICAL DEPARTMENT CENTER AND SCHOOL FORT SAM HOUSTON, TEXAS ELECTROSURGICAL APPARATUS SUBCOURSE MD0363 EDITION 100

WHALETEQ. ESU Neutral Electrodes Impedance Tester. Model: HFPA150. User Manual

Electrosurgery. KLS Martin ME MB 2 SAFETY BY EASY USE

PA FAN PLATE ASSEMBLY 188D6127G1 SYMBOL PART NO. DESCRIPTION. 4 SBS /10 Spring nut. 5 19A702339P510 Screw, thread forming, flat head.

The Variable Threshold Neutral Isolator (VTNI)

Instruction Manual. SSQ-2F Controller Board. For the. v1.41 For Rife Plasma Tube Systems. Manual v by Ralph Hartwell Spectrotek Services

ELECTROSURGICAL UNIT ANALYZER

ELEKTROTOM 505. Service Manual (E)

MINIMAL ENERGY. MAXIMUM PRECISION. 1,2, The newest addition to Valleylab energy made to enhance the performance of your electrosurgical handpieces1,2,

QA-ES III Electrosurgical Analyzer

ET 61 - Electrician Theory Examination Marking Schedule

Anaesthesia. Update in. Electricity and Magnetism. David Sanders, Bruce McCormick* *Correspondence

QA-ES III Electrosurgical Analyzer

ELECTROSURGICAL GENERATOR USER S GUIDE

P R O T R O N I X D i g i t a l P o w e r M e t e r M o d e l : A I n s t r u c t i o n M a n u a l

Central Electronics Model 600L Linear Amplifier

ULTRASTAB IHF Precision Current Transducer

Boulder W Class A Stereo Power Amplifier

POWER DELEGATOR SERIES 7200A POWER DISTRIBUTION UNIT WITH POWER CONDITIONING GENERAL SPECIFICATIONS

ULTRASTAB I U Precision

Low / Medium Power Electrosurgery Review 2002

DLVP A OPERATOR S MANUAL

CA60 ma Current Clamp

Thornwood Drive Operating Manual: Six-SCR General Purpose Gate Firing Board FCOG6100 Revision R

MAINTENANCE MANUAL AUDIO MATRIX BOARD P29/

SAFETY AND HEALTH STANDARD ELECTRICAL GROUNDING Effective Date: 07/17/10 Standard: Document Number: KUCSH0039 Rev: 4

BASIC ELECTRICITY/ APPLIED ELECTRICITY

BASIC ELECTRICITY/ APPLIED ELECTRICITY

AC/DC DIGITAL CLAMP METER OPERATION MANUAL

Current Probes. User Manual

Innovating Together. An introduction to Electrosurgery

Auto Diagnosis Test #2 Review

MASTR II AUXILIARY RECEIVER 19D417546G7 & G8 & ANTENNA MATCHING UNITS 19C321150G1-G2. Maintenance Manual LBI-30766L. Mobile Communications

COM-POWER OPERATION MANUAL ACS W

1 FUNCTIONAL DESCRIPTION WAY SPLITTER/INPUT BOARD FET RF AMPLIFIERS WAY POWER COMBINER VSWR CONTROL BOARD...

OPERATOR S INSTRUCTION MANUAL

Basic Circuitry and X ray Production. Lynn C. Sadler, MSRS, R.T.(R)(QM) President, WCEC, Inc.

An Alternative Technique to Perform Surgeriesin Hospital by Surgical Diathermy

User s Guide Aaron 950

Operator s Manual E L E C T R O S U R G I C A L U N I T

Licensed Electricians Practical Assessment (LEP)

Transformer Waveforms

PERMA Series Revision 2 Amplifiers

WATT TRANSDUCER. Instruction Manual

LEEP System 1000 Table of Contents

QUICK SETUP GUIDE SECULIFE ESPRO

Materials. Eight pin DIP socket 0.1 µf capacitor

Feed back element in high performance gradient amplifiers Feed back element in precision current regulated power supplies

Vocabulary. Electric Current. Electric Circuit. Open Circuit. Conductors. Insulators. Ohm s Law Current. Voltage. Resistance.

PHASE CUT SIGNAL DRIVERS

Model BSI-1A BIPHASIC STIMULUS ISOLATOR

Product overview. Features. Product specifications. Order codes. 1kΩ Resistance Output Module

OPERATING INSTRUCTIONS AND SYSTEM DESCRIPTION FOR THE. ISO-STIM 01D STIMULUS ISOLATION UNIT ±100 V / ±10 ma, bipolar output

Instruction Manual. (English-English)

F O R T H E L O V E O F M U S I C LP100 OWNER'S MANUAL AND INSTALLATION GUIDE INTRODUCTION

Troubleshooting accelerometer installations

Data Sheet. Hall effect transducers, current and voltage

Primary Metering. What is Primary Metering?

J-470 ELECTROSURGE. Radio Frequency Electrosurgery Unit. Operation Instructions

TRANSFORMER OPERATION

Data Isolation Cards. 2-Wire HDSL/56KBS Isolation Card P Wire HDSL/56KBS Isolation Card P30050

INSTRUCTION MANUAL FOR MICROELECTRODE AC AMPLIFIER MODEL 1800

USER S GUIDE. User s Guide Bovie IDS-400

CMA-100. Counter Measures Amplifier. Owner s Guide

END-OF-SUBCOURSE EXAMINATION

ARC 200 HF SURGICAL GENERATOR SIMPLY SAFE

The Discussion of this exercise covers the following points:

HO118 Binning Interface User Manual * * Version 02. User Manual. Test & Measurement

Ametek, Inc. Rotron Technical Products Division. 100 East Erie St., Suite 200 Kent, Ohio User's Guide. Number Revision F

Licensed Electricians Practical Assessment (LEP)

Instruction Manual for Digital Grounding Resistance Meter. Table of Contents

Electrosurgery. KLS Martin ME 102. The compact Electrosurgical Unit with high Capacity

MS8250A/B OPERATION MANUAL MS8250A. Hz% FUNC REL RANGE REL HOLD OFF 10A. Hz% A NCV. Hz% COM. A ma 10A FUSED 600V CAT IV.

CPCO Series DC-AC Current Probe, Clamp On, 160mm, ±1000A ±2000A ±4000A ±8000A ±12000A, ±16000A

Exercise 1: The DC Ammeter

LCM100 USER GUIDE. Line Carrier Modem INDUSTRIAL DATA COMMUNICATIONS

MS-AM402. User/Installation Manual FUSIONENTERTAINMENT.COM

***note: pallet may draw up to 5.0 amps on a 32V supply.

GROUND TESTERS For all of your Ground Testing needs...

MAINTENANCE MANUAL AUDIO BOARDS 19D902188G1, G2 & G3

PROFESSIONAL POWER AMP

Broadband Current Probe Series Operation Manual

AIM & THURLBY THANDAR INSTRUMENTS

Back to the Basics Current Transformer (CT) Testing

Transcription:

Valleylab Force FX periodic testing Electrosurgery The passage of high frequency electrical current through tissue to create a desired clinical effect. 2/7/2009 1

What tests should be done? 1. Output tests in each different mode; cut, coag and bi-polar. 2. REM high and low trip points 3. RF Leakage test 2/7/2009 2

Different loads for different modes. Usually bi-polar mode is tested at 100 ohms load, cut mode at 300 ohms and coagulation mode at 500 ohms. Measure the current, not the watts. RF power output circuitry is 1 to 6 N channel MOSFET s in parallel. With age and thermal damage, they begin to fail and the output voltage drops. Circuitry compensates by increasing the current thru the MOSFET s. Current times voltage equals wattage. Measuring wattage will not detect the drop in voltage, but measuring current will detect the voltage drop. Valleylab specifically states to test for current under specified load. 2/7/2009 3

ESU Math, not quite rocket science Watt setting of ESU / Load setting of tester = desired test result Bi-Polar 10 Watts =.1=.316 amp = 316 ma 100 Ω Coag 30 W = 0.06 =.245 A = 245 ma 500 Ω Cut 75W =.25 =.5A = 500mA 300 Ω 2/7/2009 4

REM testing Testing must be done with the middle pin of the r.e.m. connector present. Without the pin, the e.s.u. switches to non r.e.m. circuitry. If an e.s.u. is sent to mfr. for repair with a r.e.m. fault, testing must be done with the actual patient pad. Normal power output testing bypasses r.e.m. circuitry. 2/7/2009 5

ESU Vocabulary page one 1. Current The number of electrons moving past a given point per second, measured in amperes. 2. Current Density The amount of current flow per unit of surface area; current concentration directly proportional to the amount of heat generated. 3. Current Division Electrical current leaving the intended electrosurgical circuit and following an alternate path of least resistance to ground; typically the cause of alternate site burns when using a grounded generator. 4. Cut A low-voltage, continuous waveform optimized for electrosurgical cutting. 5. Cutting Use of the cut waveform to achieve an electrosurgical effect that results from high current density in the tissue causing cellular fluid to burst into steam and disrupt the structure. Voltage is low and current flow is high. 2/7/2009 6

ESU Vocabulary page two 1. Desiccation The electrosurgical effect of tissue dehydration and protein denaturation caused by direct contact between the electrosurgical electrode and tissue. Lower current density/concentration than cutting. 2. Diathermy The healing of body tissue generated by resistance to the flow of high-frequency electric current. 3. Direct Coupling The condition that occurs when one electrical conductor (the active electrode) comes into direct contact with another secondary conductor (scopes, graspers). Electrical current will flow from the first conductor into the secondary one and energize it. 4. Direct Current A flow of electrons in only one direction. 2/7/2009 7

ESU Vocabulary page three 1. Fulguration: Using electrical arcs (sparks) to coagulate tissue. The sparks jump from the electrode across an air gap to the tissue. 2. Generator: The machine that coverts low-frequency alternating current to high-frequency electrosurgical current. 3. Ground, Earth Ground: The universal conductor and common return point for electric circuits. 4. Grounded Output: The output on a electrosurgical generator referenced to ground. 2/7/2009 8

ESU Vocabulary page four Impedance: A form of electrical resistance observed in an alternating current that is analogous to the classic electrical resistance that occurs in a direct current. Insulation Failure: The condition that occurs when the insulation barrier around an electrical conductor is breached. As a result, current will travel outside the intended circuit. Isolated Output: The output of an electrosurgical generator that is not referenced to earth ground. Leakage Current: Current that flows along an undesired path, usually to ground; in isolated electrosurgery, RF current that regains its ground reference. 2/7/2009 9

ESU Vocabulary page five Monopolar Output: A grounded or isolated output on an electrosurgical generator that directs current through the patient to a patient return electrode. Patient Return Electrode (PRE): A conductive plate or pad (dispersive electrode) that recovers the therapeutic current from the patient during electrosurgery, disperses it over a wide surface area, and returns it to the electrosurgical generator. Power: The amount of heat energy produced per second, measured in watts. Power Efficiency Rating (PER) A measure of the ability of an electrosurgical generator to accurately deliver the selected power into a wide range of tissue types. 2/7/2009 10

ESU Vocabulary page six Radio Frequency (RF): An electrical current that alternates the poles in the radio frequency range (300 khz 36 Hz); the highfrequency current used in electrosurgery. Resistance The lack of conductivity or the opposition to the flow of electric current, measured in ohms. Voltage The force that pushes electric current through impedance/resistance; electromotive force or potential difference expressed in volts. Watt The unit of measurement for power. 2/7/2009 11

Test jigs used for preventative maintenance Calibrated ESU analyzer Return/dispersive/patient cable Active/output cable Variable 1 to 150 ohm resistor 2/7/2009 12

Preventative Maintenance Factory recommended schedule every 6 months refer to p 5-19, 5-24 Bipolar output test (page 5-18) 1.Set analyzer load resistance to 100 ohms. 2.Set bipolar power to 10. 3.Press footswitch and measure current in all 3 modes. (Current delivered should be 315 ma +/- 24 ma rms) 2/7/2009 13

Cut mode output test (page 5-19) 1. Set the analyzer load resistance to 300 ohms. 2. Set CUT power to 75 watts. 3. Press CUT footswitch and measure current in the Pure, Low & Blend modes. (Current delivered should be 499 ma +/- 38 ma.) 4. Repeat this test on the MONOPOLAR 2 using a hand piece. 2/7/2009 14

Coag output test ( page 5-20) 1. Set the analyzer load resistance to 500 ohms. 2. Set CUT power to 75 watts. 3. Press CUT footswitch and measure current in the Pure, Low & Blend modes. (Current delivered should be 245 ma +/- 19 ma.) 4. Repeat this test on the MONOPOLAR 2 using a hand piece.

Return Electrode Monitoring testing 1. Set var. resistor to 120 ohms and connect to REM jack. 2. Slowly increase resistance until led turns RED and alarms sounds, record trip point. (135 =/- 5 ohms) 3. Disconnect, set to 20 ohms, connect. 4. Slowly decrease resistance until red led and alarm, record trip point. (3-6 ohms) 2/7/2009 16

RF Leakage Testing refer to page 5-23 Checking monopolar High Frequency leakage current 1. Set load to 200 ohms, set cut to 300 watts & pure mode, set coag to 120 watts and spray mode. 2. Activate footswitch in 4 different settings: Open/Closed and Active/Dispersive 3. Record highest current, should not exceed 150 ma 2/7/2009 17

If High Frequency leakage is high: 1. Check REM pad socket connector 2. P.N. S202701854, $160 3. Compare your result with past tests and other similar units