Single camera spectral domain polarizationsensitive optical coherence tomography using offset B-scan modulation

Similar documents
Polarization-sensitive spectral-domain optical coherence tomography using a single line scan camera

60 MHz A-line rate ultra-high speed Fourier-domain optical coherence tomography

Simultaneous acquisition of the real and imaginary components in Fourier domain optical coherence tomography using harmonic detection

Spectral domain optical coherence tomography with balanced detection using single line-scan camera and optical delay line

Full-range k -domain linearization in spectral-domain optical coherence tomography

Temporal coherence characteristics of a superluminescent diode system with an optical feedback mechanism

Talbot bands in the theory and practice of optical coherence tomography

PHASE DETECTION WITH SUB-NANOMETER SEN- SITIVITY USING POLARIZATION QUADRATURE EN- CODING METHOD IN OPTICAL COHERENCE TOMOG- RAPHY

Optical coherence tomography

Instructions for the Experiment

Heterodyne swept-source optical coherence tomography for complete complex conjugate ambiguity removal

7 CHAPTER 7: REFRACTIVE INDEX MEASUREMENTS WITH COMMON PATH PHASE SENSITIVE FDOCT SETUP

High-speed spectral-domain optical coherence tomography at 1.3 µm wavelength

Isolator-Free 840-nm Broadband SLEDs for High-Resolution OCT

FIRST REPORTED in the field of fiber optics [1], [2],

Single-shot two-dimensional full-range optical coherence tomography achieved by dispersion control

Volumetric microvascular imaging of human retina using optical coherence tomography with a novel motion contrast technique

Parallel Digital Holography Three-Dimensional Image Measurement Technique for Moving Cells

Imaging the Subcellular Structure of Human Coronary Atherosclerosis Using 1-µm Resolution

doi: /OE

High-speed imaging of human retina in vivo with swept-source optical coherence tomography

Improved lateral resolution in optical coherence tomography by digital focusing using twodimensional numerical diffraction method

Transmission- and side-detection configurations in ultrasound-modulated optical tomography of thick biological tissues

All fiber optics circular-state swept source polarization-sensitive optical coherence tomography

(51) Int Cl.: G01B 9/02 ( ) G01B 11/24 ( ) G01N 21/47 ( )

OCT mini-symposium. Presenters. Donald Miller, Indiana Univ. Joseph Izatt, Duke Univ. Thomas Milner, Univ. of Texas at Austin Jay Wei, Zeiss Meditec

Ultrahigh speed volumetric ophthalmic OCT imaging at 850nm and 1050nm

CHAPTER 5 FINE-TUNING OF AN ECDL WITH AN INTRACAVITY LIQUID CRYSTAL ELEMENT

Gabor fusion technique in a Talbot bands optical coherence tomography system

Imaging obscured subsurface inhomogeneity using laser speckle

White-light interferometry, Hilbert transform, and noise

Large-field high-speed polarization sensitive spectral domain OCT and its applications in ophthalmology

Phase-sensitive high-speed THz imaging

Off-axis full-field swept-source optical coherence tomography using holographic refocusing

Lab Report 3: Speckle Interferometry LIN PEI-YING, BAIG JOVERIA

Kit for building your own THz Time-Domain Spectrometer

Phase-resolved optical frequency domain imaging

Supplementary Materials

Polarization Sagnac interferometer with a common-path local oscillator for heterodyne detection

some aspects of Optical Coherence Tomography

Visualization of human retinal micro-capillaries with phase contrast high-speed optical coherence tomography

Study of self-interference incoherent digital holography for the application of retinal imaging

Characteristics of point-focus Simultaneous Spatial and temporal Focusing (SSTF) as a two-photon excited fluorescence microscopy

A novel tunable diode laser using volume holographic gratings

Observational Astronomy

Real-time optical spectrum analysis of a light source using a polarimeter

Monte Carlo simulation of an optical coherence tomography signal in homogeneous turbid media

Radial Polarization Converter With LC Driver USER MANUAL

Ultrahigh-resolution high-speed retinal imaging using spectral-domain optical coherence tomography

Improving the output beam quality of multimode laser resonators

In-line digital holographic interferometry

Multi-frequency and multiple phase-shift sinusoidal fringe projection for 3D profilometry

Optimization for Axial Resolution, Depth Range, and Sensitivity of Spectral Domain Optical Coherence Tomography at 1.3 µm

Lab 5 - Electro-Optic Modulation

Removing the depth-degeneracy in optical frequency domain imaging with frequency shifting

Full-range polarization-sensitive swept-source optical coherence tomography by simultaneous transversal and spectral modulation

Holography as a tool for advanced learning of optics and photonics

Modifications of the coherence radar for in vivo profilometry in dermatology

Jones matrix analysis of high-precision displacement measuring interferometers

EE119 Introduction to Optical Engineering Fall 2009 Final Exam. Name:

LOPUT Laser: A novel concept to realize single longitudinal mode laser

Polarization Experiments Using Jones Calculus

SUPPLEMENTARY INFORMATION

Fourier Domain (Spectral) OCT OCT: HISTORY. Could OCT be a Game Maker OCT in Optometric Practice: A THE TECHNOLOGY BEHIND OCT

Section 2 ADVANCED TECHNOLOGY DEVELOPMENTS

SUPPLEMENTARY INFORMATION DOI: /NPHOTON

Laser Telemetric System (Metrology)

Use of Computer Generated Holograms for Testing Aspheric Optics

Single-shot depth-section imaging through chromatic slit-scan confocal microscopy

Pulsed illumination spectral-domain optical coherence tomography for human retinal imaging

Be aware that there is no universal notation for the various quantities.

Pulsed-source spectral-domain optical coherence tomography with reduced motion artifacts

Design of a digital holographic interferometer for the. ZaP Flow Z-Pinch

Direct observation of beamed Raman scattering

Opto-VLSI-based reconfigurable photonic RF filter

Citation for published version (APA): Nguyen, D. V. (2013). Integrated-optics-based optical coherence tomography

Programmable polarization-independent spectral phase compensation and pulse shaping by use of a single-layer liquid-crystal modulator

Outline. Motivation Experimental Set-Up Theory behind the set-up Results Acknowledgements

Optical frequency domain imaging with a rapidly swept laser in the nm range

GRENOUILLE.

High-speed optical frequency-domain imaging

Fang-Wen Sheu *, Yi-Syuan Lu Department of Electrophysics, National Chiayi University, Chiayi 60004, Taiwan ABSTRACT

Module 5: Experimental Modal Analysis for SHM Lecture 36: Laser doppler vibrometry. The Lecture Contains: Laser Doppler Vibrometry

Optical Coherence Tomography Systems and signal processing in SD-OCT

Laser Sources for Frequency-Domain Optical Coherence Tomography FD-OCT

Full Field Swept Source Optical Coherence Tomography

Methods for parallel-detection-based ultrasound-modulated optical tomography

7. Michelson Interferometer

Supplementary Figure 1. Effect of the spacer thickness on the resonance properties of the gold and silver metasurface layers.

Thin holographic camera with integrated reference distribution

Compact OAM Microscope for Edge Enhancement of Biomedical and Object Samples

PHY 431 Homework Set #5 Due Nov. 20 at the start of class

A pulsed THz Imaging System with a line focus and a balanced 1-D detection scheme with two industrial CCD line-scan cameras

Two-step-only phase-shifting interferometry with optimized detector bandwidth for microscopy of live cells

Simultaneous multiple-depths en-face optical coherence tomography using multiple signal excitation of acousto-optic deflectors

Spectrally resolved frequency comb interferometry for long distance measurement

In vivo three-dimensional microelectromechanical endoscopic swept source optical coherence tomography

Optical sectioning using a digital Fresnel incoherent-holography-based confocal imaging system

Improved phase sensitivity in spectral domain phase microscopy using line-field illumination and self phase-referencing

Photon Diagnostics. FLASH User Workshop 08.

Transcription:

Single camera spectral domain polarizationsensitive optical coherence tomography using offset B-scan modulation Chuanmao Fan 1,2 and Gang Yao 1,3 1 Department of Biological Engineering, University of Missouri, Columbia, MO 65211 USA 2 cf7b6@mail.missouri.edu 3 gyao@missouri.edu Abstract: We report a simple implementation to acquire spectral domain polarization-sensitive optical coherence tomography (PSOCT) using a single camera. By combining a dual-delay assembly in the reference arm and offset B-scan in the sample arm, the orthogonal vertical- and horizontalpolarized images were acquired in parallel and spatially separated by a fixed distance in the full range image space. The two orthogonal polarization images were recombined to calculate the intensity, retardance and fast-axis images. This system was easy to implement and capable of acquiring highspeed in vivo 3D polarization-sensitive OCT images. 2010 Optical Society of America OCIS codes: (260.1440) Polarization sensitive device; (110.4500) Optical coherence tomography. References and links 1. D. uang, E. A. Swanson, C. P. Lin, J. S. Schuman, W. G. Stinson, W. Chang, M. R. ee, T. Flotte, K. Gregory, C. A. Puliafito, and J. Fujimoto, Optical coherence tomography, Science 254(5035), 1178 1181 (1991). 2. A. F. Fercher, W. Drexler, C. K. itzenberger, and T. Lasser, Optical Coherence Tomography Principles and Applications, Rep. Prog. Phys. 66(2), 239 303 (2003). 3. M. R. ee, D. uang, E. A. Swanson, and J. G. Fujimoto, Polarization sensitive low coherence reflectometer for birefringence characterization and ranging, J. Opt. Soc. Am. B 9(6), 903 908 (1992). 4. J. F. de Boer, T. E. Milner, M. J. C. van Gemert, and J. S. Nelson, Two-dimensional birefringence imaging in biological tissue by polarization-sensitive optical coherence tomography, Opt. Lett. 22(12), 934 936 (1997). 5. S. L. Jiao, and L.. Wang, Jones-matrix imaging of biological tissues with quadruple-channel optical coherence tomography, J. Biomed. Opt. 7(3), 350 358 (2002). 6. B.. Park, C. Saxer, S. M. Srinivas, J. S. Nelson, and J. F. de Boer, In vivo burn depth determination by highspeed fiber-based polarization sensitive optical coherence tomography, J. Biomed. Opt. 6(4), 474 479 (2001). 7. S. M. Srinivas, J. F. de Boer,. Park, K. Keikhanzadeh,. E. uang, J. Zhang, W. Q. Jung, Z. Chen, and J. S. Nelson, Determination of burn depth by polarization-sensitive optical coherence tomography, J. Biomed. Opt. 9(1), 207 212 (2004). 8. B. Cense, T. C. Chen, B.. Park, M. C. Pierce, J. F. de Boer, and J. F. de Boer, Thickness and birefringence of healthy retinal nerve fiber layer tissue measured with polarization-sensitive optical coherence tomography, Invest. Ophthalmol. is. Sci. 45(8), 2606 2612 (2004). 9. C. itzenberger, E. Goetzinger, M. Sticker, M. Pircher, and A. Fercher, Measurement and imaging of birefringence and optic axis orientation by phase resolved polarization sensitive optical coherence tomography, Opt. Express 9(13), 780 790 (2001). 10. E. Götzinger, M. Pircher, B. Baumann, C. Ahlers, W. Geitzenauer, U. Schmidt-Erfurth, and C. K. itzenberger, Three-dimensional polarization sensitive OCT imaging and interactive display of the human retina, Opt. Express 17(5), 4151 4165 (2009). 11. M. Zhao, and J. A. Izatt, Single-camera sequential-scan-based polarization-sensitive SDOCT for retinal imaging, Opt. Lett. 34(2), 205 207 (2009). 12. E. Götzinger, M. Pircher, and C. K. itzenberger, igh speed spectral domain polarization sensitive optical coherence tomography of the human retina, Opt. Express 13(25), 10217 10229 (2005). 13. B. Baumann, E. Götzinger, M. Pircher, and C. K. itzenberger, Single camera based spectral domain polarization sensitive optical coherence tomography, Opt. Express 15(3), 1054 1063 (2007). 14. B. Cense, M. Mujat, T. C. Chen, B.. Park, and J. F. de Boer, Polarization-sensitive spectral-domain optical coherence tomography using a single line scan camera, Opt. Express 15(5), 2421 2431 (2007). (C) 2010 OSA 29 March 2010 / ol. 18, No. 7 / OPTICS EXPRESS 7281

15. C. Fan, Y. Wang, and R. K. Wang, Spectral domain polarization sensitive optical coherence tomography achieved by single camera detection, Opt. Express 15(13), 7950 7961 (2007). 16. B. Baumann, M. Pircher, E. Götzinger, and C. K. itzenberger, Full range complex spectral domain optical coherence tomography without additional phase shifters, Opt. Express 15(20), 13375 13387 (2007). 17. L. An, and R. K. Wang, Use of a scanner to modulate spatial interferograms for in vivo full-range Fourierdomain optical coherence tomography, Opt. Lett. 32(23), 3423 3425 (2007). 18. R. A. Leitgeb, R. Michaely, T. Lasser, and S. C. Sekhar, Complex ambiguity-free Fourier domain optical coherence tomography through transverse scanning, Opt. Lett. 32(23), 3453 3455 (2007). 1. Introduction Optical coherence tomography (OCT) [1,2] is a powerful non-invasive optical imaging technique that can acquire high resolution depth-resolved images in highly scattering tissue specimen. Polarization-sensitive optical coherence tomography (PSOCT) [3 5] enables polarization-dependent detection in OCT and can provide additional imaging contrast such as retardance and fast-axis orientation in birefringence samples. PSOCT has been successfully applied in a variety of biomedical applications such as burn depth estimation in skin [6,7] and ophthalmology [8 10]. The recent development of PSOCT has been shifted to spectral domain implementation [11 14,11,15] due to its superior speed and sensitivity that are critical for in vivo three dimensional applications. In order to extract polarization properties in a birefringence sample, at least two sample images need to be acquired with two orthogonal detection polarization states [9]. Zhao and Izatt [11] implemented a single-camera sequential-scan-based PSOCT system by using an electro-optic modulator to change the detection polarization between sequential scans. To further improve imaging speed, several single-shot PSOCT implementations have been developed. Baumann et al [13] used a polarization-sensitive beam splitter to direct the horizontal- and vertical-polarized components to two adjacent halves of a single spectrometer camera. Cense et al [14] reported a similar system where a Wollaston prism in the detection arm was used to direct the two polarization components to a single camera. Fan et al [15] later reported a different configuration that separated the two orthogonal polarization components in the reference arm instead of the detection arm. The two orthogonally polarized images were mapped into the two opposite sides in the full range image space relative to the zero delay line. This implementation used a common interference path in the detection arm and thus avoided the strict spatial alignment required when separating two polarization images in the detection arm. owever, it required an extra phase modulation in the reference arm [15]. In this paper, we presented an improved implementation over Fan et al s [15] original approach by eliminating the extra phase modulation components in the reference arm. Instead, we utilized the phase modulation provided by a galvo scanner during offset B-scans that was previously used [16 18] to achieve full range spectral domain OCT. This new implementation provided a simple solution to acquire high-speed polarization sensitive OCT images. 2. Methods 2.1 system setup A schematic diagram of the proposed PSOCT system is shown in Fig. 1. An 844nm SLD with λ = 46.8nm (SLD-371-P1, Superlum, Russia) was used as the light source. The incident laser power on the sample surface was 1.3mW. The incident light was collimated and vertically polarized () by a polarizer after passing through an optical isolator. At the reference arm, the light was split into two parts by a non-polarized beam splitter (BS). At one arm, the vertically polarized () light was directly reflected by a mirror (M); while at the other arm, a quarter-wave plate (QWP) was placed at 45 to convert the back reflected light into a horizontally polarized () light. The - and -polarized reference light was recombined by the same beam splitter (BS). In the sample arm, another QWP placed at 45 converted the -polarized incident light into circularly polarized light. The incident light was reflected at a point slightly deviated from the pivot axis at the x-y galvanometer scanner and (C) 2010 OSA 29 March 2010 / ol. 18, No. 7 / OPTICS EXPRESS 7282

redirected onto an objective lens (OL). The backscattered light from the sample interfered with the horizontal- and vertical-polarized reference light separately, and were coupled into a custom designed spectrometer via a single mode fiber. The coherence spectra were captured by a line scan CCD camera with 1024 pixels (AIIA SM2, e2v, France) and acquired through a frame grabber (PCIe-1427, National Instruments, USA). The image acquisition speed was 50k A-line/s. Fig. 1. A schematic diagram of the PSOCT system. SLD: superluminescent diode, C: collimator, P: polarizer to generate vertically polarized light. BS: non polarization beam splitter. ND: neutral density filter. QWP: quarter-wave plate at 45. M: reference mirrors. L: lens. OL: objective lens (f = 88mm). PTG: olume Phase olography Transmission Grating (1200lines/mm). CCD: line scan camera. The control signals are illustrated in Fig. 2. A digital pulse train (Ch.0) was generated from a DAQ card (PCI-6221, National Instruments, USA) to trigger the CCD line by line. This pulse train also served as the sample clock for generating saw tooth signals to control the scanner. Two voltage signals (Ch.1 and Ch.2) were used to drive the x- and y-axis galvanometer scanners to acquire the B- and C-scans, respectively. Fig. 2. Waveforms for system synchronization. Ch. 0: timing base camera trigger; Ch.1: B-scan signal; Ch.2: C-scan signal. 2.2 signal reconstruction and processing The acquired raw OCT signal represented the summation of the - and -polarized interference components and can be described as: ( ϕ ) ( ϕ ) I = I + A cos 2 k( z d) + + A ( k, x)cos 2 k( z+ d) +, (1) 0 k, x k, x k, x d k, x d (C) 2010 OSA 29 March 2010 / ol. 18, No. 7 / OPTICS EXPRESS 7283

where A k, x and A are the - and -polarized signal amplitudes at wave number k and k, x lateral scanning position x, respectively. Similar to that in [15], the optical paths of the and reference were symmetrically mapped to two opposite sides ( ± d) of the zero delay line z = 0 (the focus plane in our configuration) by adjusting the delays in the reference arm. This configuration was necessary to shift the and component image adjacent to each other in the final image. The φ d was the dynamic phase introduced by offset B-scan. φ d was a function of the offset h related to the scanner pivot axis [16,17]: 8π h θ ϕd =, (2) λ where θ is the scanning angle. In order to remove the autocorrelation, self-cross correlation and fixed pattern noise, the acquired signal was preprocessed first by subtracting the average of 1000 A-scans. Then the spatial spectrum of the raw interferogram I k,x, B(k, f x ), was calculated by applying Fourier transform to the preprocessed signal against lateral position x: 1 1 B( k, fx ) = ( B ( k, fx fd ) + B ( k, fx + fd )) + ( B ( k, fx fd ) + B ( k, fx + fd )), (3) 2 2 where B and B are the Fourier transform of the and components, respectively. The space frequency shift f d is: f d = 1 dϕd 4h 1, 2π dx = λ F (4) where F is the focal length of the imaging objective, and x = Fθ. In the experiment, the offset and the lateral scan step size x were adjusted so that the spatial modulation frequency f d was half of the maximal spatial sampling frequency 1/ x. After obtaining Fourier transforms [Eq. (3)] on every wave number k, the resulting - and -polarized spectra were mirror-symmetric with zero frequency. A band-pass filter was applied to select the single side spatial spectrum. An inverse Fourier transform was then used to obtain the complex spectrum. The longitudinal (to the wave number k) Fourier transform was carried out on every calculated complex spectrum. Full range complex image was obtained with corresponding complex conjugate mirror image removed, leaving two images adjacent to each other by a distance of 2d. These two images represent these formed in horizontal and vertical polarization channels: A z x = A z+ d x iψ z x + A z d x iψ z x (5) (, ) (, )exp( (. )) (, )exp( (. )). To calculate polarization parameters, one image was shifted by a distance of 2d to coincide with the other. The intensity I image can be calculated as I A z x A z x established method [3,9] as: 2 2 = (, ) + (, ). The retardance δ and fast-axis θ images were calculated using 1 A ( z, x) ψ ( z, x) ψ ( z, x) + π δ = tan ( ), θ =. (6) A ( z, x) 2 The ranges of retardation and fast axis were [0, 90 ] and [0, 180 ] respectively. 3. Results and discussion Figure 3 demonstrated the image processing procedure described in the last section. The acquired original B-scan OCT spectra is shown in Fig. 3(a). Figure 3(b) was obtained after removing the DC components in the raw data. After Fourier transform on x, the resulting complex spectra were shown in Fig. 3(c). Both real (R) and imagery (I) parts of the result are (C) 2010 OSA 29 March 2010 / ol. 18, No. 7 / OPTICS EXPRESS 7284

shown. The spectrum was modulated to be symmetric to the center zero frequency. The complex spectrum was band-pass filtered to select the single-sided spectrum. Such filtering also helped remove low frequency fixed pattern noise in the system. The Fig. 3(d) was computed by applying inverse Fourier transform to Fig. 3(c). Finally, applying Fourier transform against the wave number k produced the analytical OCT images where the - and -polarized components were adjacent to each other by a displacement of 2d. After shifting one image to coincide with the other orthogonally polarized component, the intensity and retardance images can be obtained from the amplitude image in Fig. 3(e) and the fast axis image can be calculated using the phase image [Fig. 3(f)] according to Eq. (6). Fig. 3. An illustration of the OCT image processing. (a) The raw B-scan spectra; (b) Preprocessed spectra after removing the DC component; (c) The imaginary (I) and real (R) parts of Fourier transformed (on lateral position x) spectra. The white boxes represent the band-pass filter. (d)the complex interferogram after inverse Fourier transform of (c). The final amplitude (e) and phase (f) of the depth-resolve image were calculated by Fourier transform of (d). During image processing, the depth-resolved signals were corrected for the sensitivity falling-off with depth whose profile approximately approached a Sinc function corresponding to Fourier transforms of the rectangular pixel sampling on the interferogram. The raw OCT signal was corrected by multiplying the inverse of the sensitivity curve. The present system had a maximum sensitivity of 104.5 db at the zero delay line (focal plane) and minimum sensitivity of 75.5 db at an imaging depth of 2.6 mm. The extinction ratio of complex conjugate suppression as defined in [16] was measured as 29.6 db. Fig. 4. The measured optical axis and retardance of a waveplate in front of a mirror. Three sets of data were measured when the mirror was placed at 0, 0.9 and 1.8mm from the focal plane. To validate the system, a waveplate (retardance of 67.5 at 844nm) placed in front of a mirror was used as the sample. The fast axis and retardance were measured while rotating the waveplate from 0 to 170. As shown in Fig. 4, three sets of data were measured when mirror was positioned at z 1 = 0 mm, z 2 = 0.9 mm and z 3 = 1.8 mm from the focal plane (zero delay (C) 2010 OSA 29 March 2010 / ol. 18, No. 7 / OPTICS EXPRESS 7285

line). The measured axis orientation had good agreement with preset values with errors of [- 2.5, 1.5 ], [-4.3, 1.2 ] and [-3.4, 3.1 ] at position z 1, z 2 and z 3, respectively. The corresponding measured retardance had errors between [-3.5, 3.1 ], [-4.9, 3.2 ] and [-5.9, 3.1 ]. Errors were slightly larger when the mirror was moved away from the zero delay line. Figure 5 shows the intensity, retardance and axis orientation images obtained in a piece of chicken cartilage sample. Similar to other polarization-sensitive OCT images, the internal intensity variations suggested structural changes. The periodic changes in retardance near the sample surface revealed birefringence information of the cartilage material. The fast-axis orientation patterns coincided with the retardance changes and indicated the directional information of the organized cartilage tissue. At larger depths, the periodic changes disappeared in both retardance and fast-axis images, indicating loss of birefringence materials. Fig. 5. the PSOCT images of a chicken cartilage sample in vitro. (a) Intensity; (b) retardance; and (c) fast axis orientation. The size bar indicates 500µm. To demonstrate real time imaging, 3D in vivo PSOCT images were acquired from a finger fold of a human volunteer (Fig. 6). The scan contained 1000 500 A-scans at 50 kz. The images were median-filtered. In the intensity images [Fig. 6(a)], the epidermal area could be discriminated from dermal area close to the nail fold. Cuticle, nail plate, and nail bed can be identified. The retardance image [Fig. 6(b)] revealed strong birefringence at the lower area of the nail fold. Strong birefringence was observed underneath the nail bed. ariations in dermis indicated the changes in birefringence although the same region appeared homogenous in the intensity image. Similar features were also shown in the axis image [Fig. 6(c)]. 4. Summary Fig. 6. 3D in vivo PSOCT images of a human finger fold. (a) Intensity (Media 1), (b) retardance (Media 2), and (c) fast-axis orientation (Media 3). The white size bar indicates 500 µm. We implemented a full range 3D PSOCT using a single camera for parallel detection of the polarization sensitive spectral interferograms generated from - and -polarized components. The two orthogonal polarization channels were spatially separated by a fixed distance in the full range image space by using a dual delay assembly in the reference arm and offset B-scan in the sample arm. The separated two polarization images were recombined to obtain the retardance and axis images. By utilizing the offset B-scan method [16 18], we eliminated the additional modulation devices in a previous implementation [15]. Because a common path (C) 2010 OSA 29 March 2010 / ol. 18, No. 7 / OPTICS EXPRESS 7286

was used in the detection arm for the two polarization channels, system alignment was much simplified than those required when splitting the two channels in the detection arm [13,14]. This simple system was demonstrated to be capable of acquiring in vivo 3D polarizationsensitive images at a speed of the 50k lines/sec, limited only by the camera speed. Acknowledgements This project was supported in part by a NSF grant CBET06431990. (C) 2010 OSA 29 March 2010 / ol. 18, No. 7 / OPTICS EXPRESS 7287