Dosimetry of cones for radiosurgery system

Similar documents
Commissioning and Calibrating a Linear Accelerator State-of-the-Art in 2010

Development of the Use of Amorphous Silicon (ASi) Electronic Portal Imaging Devices as a Physics Tool for Routine Linear Accelerator QA

Comparison of measured Varian Clinac 21EX and TrueBeam accelerator electron field characteristics

Beam Commissioning and Annual QA Phantoms, Detectors & Accessories. iba-dosimetry.com

Comparison of peripheral dose measurements using Ionization chamber and MOSFET detector

Evaluation of dosimetry parameters of photons and electron beams using a linear ionization chamber array

Radiation transmission, leakage and beam penumbra measurements of a micro-multileaf collimator using GafChromic EBT film

Commissioning and validation of BrainLAB cones for 6X FFF and 10X FFF beams on a Varian TrueBeam STx

When small things matter. Small Field Dosimetry Application Guide

Accurate two-dimensional IMRT verification using a back-projection EPID dosimetry method

DETERMINATION OF INHERENT AND ADDITIONAL FILTRATION IN ORDER TO ESTABLISH RADIATION QUALITIES ACCORDING TO IEC 61267

SCINTILLATING FIBER DOSIMETER ARRAY

Characterization of an in vivo diode dosimetry system for clinical use

The evaluation of minimum detectable phantom thickness change using a scanning liquid filled ion chamber EPID dose response

Peace of Mind. Automated.

A Generalized Strategy for 3D Dose Verification of IMRT/VMAT Using EPID-measured Transit Images

CyberKnife Iris Beam QA using Fluence Divergence

Commissioning an Elekta Versa HD linear accelerator

8/3/2017. Use of EPIDs for Non-Routine Linac QA. Disclosure. Learning Objectives. Parts of this project received support from Varian Medical System.

A new approach to film dosimetry for high energy photon beams: Lateral scatter filtering

QA Considerations. QA for LGK Perfexion : : Follow NRC licensing guidelines (10( CFR ) Leksell Gamma Knife Perfexion

DOSE EVALUATION IN UROGRAPHY RADIOLOGICAL PROCEDURES

Isocenter and Field of View Accuracy Measurement Software for Linear Accelerator

SRS MapCHECK. SRS Patient QA, No Film. Your Most Valuable QA and Dosimetry Tools

Detectors for small field dosimetry

BRANDON RICE UNIVERSITY OF FLORIDA

The Ultimate 4D QA Solution A 4D isotropic cylindrical detector array for arc delivery QA and Dosimetry.

Characterization, Commissioning and Evaluation of Delta 4 IMRT QA System. Ram Sadagopan 1 UTMD Anderson Cancer Center Houston, TX.

Analysis of Post-exposure Density Growth in Radiochromic Film with Respect to the Radiation Dose

SRS MapCHECK. SRS Patient QA, No Film. Your Most Valuable QA and Dosimetry Tools

Accuracy of rapid radiographic film calibration for intensity-modulated radiation therapy verification

Preliminary studies of a new monitor ionization chamber

Monte Carlo study on a new concept of a scanning photon beam system for IMRT

Assessment of an Unshielded Electron Field Diode Dosimeter for Beam Scanning in Small- to Medium-Sized 6 MV Photon Fields

Joint ICTP/IAEA Advanced School on Dosimetry in Diagnostic Radiology and its Clinical Implementation May 2009

ArcCHECK. The Ultimate 4D QA Solution. Your Most Valuable QA and Dosimetry Tools

Calibration of KAP meters

7/23/2014. Acknowledgements. Implementing a new digital medical accelerator. New Generation of Medical Accelerators

CHAPTER 2 COMMISSIONING OF KILO-VOLTAGE CONE BEAM COMPUTED TOMOGRAPHY FOR IMAGE-GUIDED RADIOTHERAPY

IQM Detector Characteristics: Signal reproducibility

The Current State of EPID-Based Linear Accelerator Quality Assurance. Disclosures. Purpose of this First Talk 8/3/2017

Commissioning. Basic machine performance MLC Dose rate control Gantry speed control End-to-end tests

Electronic Brachytherapy Sources. Thomas W. Rusch

A 2-D diode array and analysis software for verification of intensity modulated radiation therapy delivery

Dose Reduction and Image Preservation After the Introduction of a 0.1 mm Cu Filter into the LODOX Statscan unit above 110 kvp

3D Diode Array Commissioning: Building Confidence in 3D QA Technology

Post-irradiation colouration of Gafchromic EBT radiochromic film

Cylindrical Ion Chambers

Development of new dosimeter for measuring dose distribution in CT

A comparison of two methods for the determination of freein-air geometric efficiency in MDCT

NOT FOR DISTRIBUTION JINST_083P_0914 v1

Physical and dosimetric aspects of a multileaf collimation system used in the dynamic mode for implementing intensity modulated radiotherapy

) was derived from 50% of the central axis dose created by nominal light field using geometry and mathematical methods. Leaf position (X mlc.

PHYSICS QUESTIONNAIRE FORM

SPECIFICATION. Kilovoltage X-ray calibration system for protection and diagnostic level dosimetry. Prepared by

Effect of slit scan imaging techniques on image quality on radiotherapy electronic portal imaging

Aim. Images for this section: Page 2 of 13

Determination of the tube voltage from clinic mammographic system using two types of detectors

RADIATION THERAPY DETECTORS. For Small Field Dosimetry

DETECTORS UNCOMPROMISING QUALITY. The standard in dosimetry measurements for over 40 years. EXRADIN DETECTORS

specifications TrueBeam STx System

Image processing techniques to evaluate mammography screening quality

Investigation of Kodak extended dose range (EDR) film for megavoltage photon beam

Physics of small megavoltage photon beam dosimetry

PIXE TECHNIQUE APPLIED TO ALMEIDA JÚNIOR MATERIALS

A diagnostic tool for basic daily quality assurance of a tomotherapy Hi Art machine

Cylindrical Ion Chambers Victoreen Model 550 Series

specifications Novalis Tx image-guided radiosurgery linear accelerator

Characterization of an in vivo diode dosimetry system for clinical use

Test Equipment for Radiology and CT Quality Control Contents

Rotational total skin electron irradiation with a linear accelerator

O USO DE REVISTAS ELETRÓNICAS POR PROFESSORES / INVESTIGADORES DE UMA FACULDADE DE ENGENHARIA

QUALITY CONTROL PHANTOMS FOR RADIOTHERAPY AND MEDICAL IMAGING

IMRT Delivery System QA. IMRT Dose Delivery. Acceptance testing. Why: specific tests for IMRT? Accuracy of leaf positioning (gaps) MLC Alignment

A Multichannel Dosimeter Based on Scintillating Fibers for Medical Applications

X-ray tube with needle-like anode

Exposure in Dental Radiology: A Comparison Between Intra-oral, Panoramic and Tomographic Examinations

specifications TrueBeam System

Evaluation of a diode array for QA measurements on a helical tomotherapy unit

Performance evaluation of the RITG148 + set of TomoTherapy quality assurance tools using RTQA 2 radiochromic film

Exposure Indices and Target Values in Radiography: What Are They and How Can You Use Them?

Using the frame averaging of as500 EPID for IMRT verification

A positioning QA procedure for 2D/2D (kv/mv) and 3D/3D (CT/CBCT) image matching for radiotherapy patient setup

GafChromic EBT2 and EBT3 Films for Ball Cube II Phantom

A new approach to measure dwell position inaccuracy in HDR ring applicators quantification and corrective QA

熊本大学学術リポジトリ. Kumamoto University Repositor

I. Introduction.

Comparative performance evaluation of a new a-si EPID that exceeds quad high-definition resolution

ABSORBED DOSE DISTRIBUTIONS USING THE ISODENSITOMETRIC METHOD FOR EXPOSURES WITH FILTER EMPLOYED FOR MAMMOGRAPHIES

Clinical helical tomotherapy commissioning dosimetry

Evaluation of a commercial flatbed document scanner and radiographic film scanner for radiochromic EBT film dosimetry

Effective energy measurement using radiochromic film: application of a mobile scanner

Response characteristics of a tandem ionization chamber in standard X-ray beams

X3D in Radiation Therapy Procedure Planning. Felix G. Hamza-Lup, Ph.D. Computer Science Armstrong Atlantic State University Savannah, Georgia USA

A Guide to Radiochromic Film Dosimetry with EBT2 and EBT3

Beam-Restricting Devices

ArcCHECKTM. The Ultimate 4D QA Solution. Your Most Valuable QA and Dosimetry Tools. VMAT RapidArc TomoTherapy Pinnacle 3 SmartArc Conventional IMRT

Quality control of Gamma Camera. By Dr/ Ibrahim Elsayed Saad 242 NMT

State of the Art Film Dosimetry

GAFCHROMIC. Therapy Dosimetry Media Models to

Transcription:

Artigo Original Revista Brasileira de Física Médica.211;5(1):25-3. Dosimetry of cones for radiosurgery system Dosimetria de cones para sistema radiocirúrgico Laura Furnari, Camila P. Sales, Gabriela R. Santos, Marco A. Silva and Gisela Menegussi Department of Radiology, Clinics Hospital, of São Paulo University School of Medicine (FMUSP), São Paulo (SP), Brazil. Abstract Dosimetry of small fields, such as cones for radiosurgery, requires a lot of care in its implementation. The acquisition of curves of Percentage depth dose () and profiles for nine circular cones with diameters from 4 to 2 mm for 6 MV photons was performed. Measurements with four types of dosimeters: diode, pinpoint ionization chamber, diamond detector and film were done. A comparison between the data obtained with the several detectors permitted to conclude that the diode is the detector more reliable. The effect of various methods to smooth the curves was studied and showed that there are methods that change very much the measured data. Interpolations were made in curves in order to eliminate the noise of small fields due to low signal in the detectors. The most important conclusion refers to the choice of suitable detector, in this case the diode, and to a careful handling of obtained data to not disturb or modify the results of the measurements. Keywords: cones, radiosurgery, commissioning, diode detector, diamond detector, pinpoint chamber, dosimetric film. Resumo A dosimetria de pequenos campos, como cones para radiocirurgia, requer muito cuidado em sua implementação. A aquisição das curvas de porcentagem de dose na profundidade (PDP) e perfis para nove cones circulares, com diâmetros de 4 a 2 mm para fótons de 6 MV, foi realizada. Foram realizadas as medidas com quatro tipos de dosímetros: diodo, câmara de ionização do tipo pinpoint, detector de diamantes e filme. Uma comparação entre os dados obtidos com diversos detectores permitiu concluir que o diodo é o detector mais confiável. O efeito dos diversos métodos para atenuar as curvas foi estudado e mostrou que existem métodos que realmente mudam os dados medidos. Interpolações foram feitas em curvas de PDP para eliminar os ruídos de pequenos campos, devido ao baixo ruído nos detectores. A conclusão mais importante refere-se à escolha do detector adequado, neste caso, o diodo, e ao manuseio cuidadoso dos dados obtidos para não transtornar ou modificar os resultados das medidas. Palavras-chave: cones, radiocirurgia, comissionamento, detector de diodo, detector de diamantes, câmara detalhada, filme dosimétrico. Introduction The cones are precise accessories used to apply radiosurgery. With them, it is possible to obtain circular fields with diameter down to 4 mm. At the moment of treatment, the jaws aperture needs to be smaller than the external circumference of the cone, because this is the shielded region of this accessory. The recent accidents related to the use of cones were originated just because the jaws had been set wrong. As a manufacturer recommendation, the adequated jaw s aperture must be the same for all cones size; for our set of cones, it must be 26 mm x 26 mm. So, the use of this accessory demands a lot of attention and care. This care must be taken in the entire process of a radiosurgery with cones, beginning by the data acquisition. The data needed to feed the planning system are numerous and must be obtained through direct measurements 1.This work presents the dosimetry data obtained with cones and the discussion about how to handle these data. Materials and methods The cones, showed in Figure 1, were commissioned in an accelerator Varian 6EX, 6 MV photon energy, equipped with micro-multileaf (mmlc) collimator m3 BrainLab. With an automatic scanning system, it was obtained the Percentage depth dose () and the profile curves at 7.5 mm depth in transverse directions for nine cones with 1.a 1.b Figure 1. Circular cone mounted at 6EX accelerator. 1.a shows mm and 1.b, 2 mm. Corresponding author: Laura Furnari Clinics Hospital (USP) Rua Dr. Eneas de Carvalho Aguiar, 255 Cerqueira Cesar São Paulo (SP), Brazil CEP 543-9 E-mail: laurafurnari@hotmail.com Associação Brasileira de Física Médica 25

Furnari L, Sales CP, Santos GR, Silva MA, Menegussi G diameter from 4 mm to 2 mm. The measurements were made with diode, pinpoint chamber, diamond detector 2 and dosimetric film. Table 1 presents the type and characteristics of the detectors used. The diode, pinpoint chamber and diamond detector were positioned vertically in the water phantom, because the fields were too small. For the measurements with the diode and pinpoint chamber, it was used the 3D scanning Blue Phantom of IBA, and for the diamond detector, it was used the PTW MP2 phantom, because it has a particular insulation system. The EDR2, an enveloped film, was put between slabs of solid water to obtain the buildup and lateral scatter. For the measurements of, the film was positioned vertically and this positioning had to be done very carefully, because, as the fields are very small, it is very difficult to put the film exactly in the direction of the central axis especially for the cone of 4 mm diameter. For the measurements of profiles, the film was positioned horizontally. The curves of and profile obtained were smoothed. Absolute measurements of output factor for a field 1 cm x 1 cm and for each cone were done, and the scatter factor was calculated. The treatment planning system iplan for radiosurgery with cones needs beam data used for the Clarkson dose calculation methods. The major beam data required for the commissioning of this unit include depth dose data measured at source-to-surface distance (SSD) of 98.5 cm, beam profiles measured at the depth of 7.5 cm with SSD of 92.5 cm and relative scatter factor data measured at the depth of 1.5 cm with SSD of 98.5 cm. 12 1 8 6 4 2 1 8 6 4 2 5 1 15 2 25 3 35 cone 2 cone 1 Figure 2. curves for 9 cones measured with diode and not smoothed. cone 1 cone 2 5 1 15 2 25 3 Figure 3. curves for 9 cones measured with diode and smoothed. Results and discussion Diode results curves The curves presented in Figure 2 are the for 9 cones obtained with diode, and Figure 3 is the same curves smoothed. The Figure 4 shows curves, the original and the smoothed manually for cone of 4 mm. The original curve presents a lot of noise that may be perceived because, as the electric signal is small, the size of noise is proportionally big. We tried to smooth the curves using several methods presented as tools in the software OmniPro-Accept, but finally it was chosen to do manual interpolations and corrections. The same procedure was applied to all other curves. 12 1 8 6 4 2 5 1 15 Edited Original 2 25 3 35 Figure 4. curves for cone with 4 mm diameter: original and edited. Table 1. Characteristics of the detectors Detector type Model Brand Effective measurement point Thick ness (mm) Diameter of active area Diode Stereotactic SFD IBA <.9 mm.6.6 mm Pinpoint chamber CC1 IBA 2.3 mm from tip 3.6 2 mm Diamond detector 63 PTW.5 mm.3 3 mm Film EDR2 Kodak 26 Revista Brasileira de Física Médica.211;5(1):25-3.

Dosimetry of cones for radiosurgery system The curves presented in Figure 5 are halves of the profiles for 9 cones obtained with diode. These curves furnished the off axis factor for each cone inserted in the planning system. Scatter factor The relative scatter data were measured with diode at the depth of 1.5 cm with SSD of 98.5 cm. The relative scatter factor was calculated by dividing the measured data for each collimator setting, by the measured data at the same depth and same SSD for a field size of 1 cm x1 cm. Figure 6 shows the relative scatter factors for different circular cones with collimator jaws of 2.6 cm x 2.6 cm. Pinpoint results curves The same results of were obtained with pinpoint chamber, but with much more noise because of the size of the chamber. (Figure 7) The profile curves obtained with pinpoint chamber, showed in Figure 8, are like that obtained with diode, but also present a lot of noise. Film results The Figure 9 is the image obtained irradiating the film perpendicularly to the central axis at the depth of 7.5 cm with SSD of 92.5 cm. This was scanned in an Epson Expression 1XL device to construct the profile curves presented in the Figure 1. 12 1 8 6 4 2 cone 1 cone 2 5 1 15 2 25 3 35 Figure 7. curves for 9 cones measured with pinpoint chamber. 1, 8, 6, 4, 2, mm mm mm mm cone 1 mm mm mm mm cone 2 mm 7 6 5 4 3 2 1 cone 2 cone17.5 cone15 cone12.5 cone1 cone5-4 -3-2 -1 1 2 3 4-1 Off axis distance (mm), 5 1 15 2 25 Distance off center (mm) Figure 5. for 9 cones measured with diode. Figure 8. for 9 cones measured with pinpoint chamber. 1,3 Scatter factor 1,1,9,7,5 5 1 15 2 25 cone diameter (mm) Figure 6. Scatter factor relative of 9 cones measured with diode. Figure 9. Image film irradiated to obtain profile curves for the cones 4, 5, 7.5, 8, 1, 12.5, 15, 17.5 and 2mm. Revista Brasileira de Física Médica.211;5(1):25-3. 27

Furnari L, Sales CP, Santos GR, Silva MA, Menegussi G Diamond results curves The Figure 11 represents the curves obtained with diamond detector. In the enlarged detail, it is possible to see the noise presented by this detector. In the Figure 12, we see the profile curves obtained with diamond detector. 35 3 25 2 15 1 5-4 -3-2 -1-5 1 2 3 4 off axis distance (mm) Figure 1. for 9 cones obtained from the film. 11 1 9 8 7 6 5 4 3 2 1 1 234567891 11 12 13 14 15 16 17 18 19 2 21 22 23 24 25 cone 2 cone 1 Figure 11. curves for 9 cones measured with diamond detector. cone 2 cone 1-3 -2-1 1 2 3 Off axis distance Figure 12. Profiles curves for 9 cones measured with diamond detector. Comparison of curves for different detectors The Figure 13 is the representation of the obtained with diode, pinpoint and film for cone of 4 mm. We may observe that they present different results: the measurements with pinpoint show more collection of charge and the film less than diode. This effect is important only in small cones. To establish a criterion to know what curve was the correct, we did measurements with diode, moving it manually and verifying after each movement if the diode was well positioned. The comparison of the data obtained with automatic and manual movement is in the Figure 14. It is possible to see that the curves are very similar, almost coincident, what indicates that the measurements done with automatic scanning of the diode were correct. Otherwise, that data were compared with gold standards BrainLab library and were approved. So, we concluded that the diode is the best detector for small cones and we used this data to feed the planning system. Effect of smooth curves The smooth of the curves offered by the software is an interesting procedure, because even if the curves have some noise they are usable. However, it is necessary to make it carefully as each smooth is going to change the parameters of the curve. As an example, the Figure 15 presents 1 8 6 4 2 pinpoint diode film 5 1 15 2 25 3 35 Figure 13. curves for mm obtained with pinpoint chamber, diode and film. 12 1 8 6 4 2 automatic manual 5 1 15 2 25 3 Figure 14. curves for mm obtained with automatic and manual movement of the diode. 28 Revista Brasileira de Física Médica.211;5(1):25-3.

Dosimetry of cones for radiosurgery system Quantity: R1: D5: D1: D2: Dmax: TPR2/1: Dose 9 mm 74.3 % 5.8 % 23.1 % 132.5 %.516 Dose 9 mm 74.4 % 5.3 % 22.9 % 132.7 %.516 Dose 9.8 mm 75.2 % 51.1 % 23.2 % 132.2 %.516 Figure 15. Results of maximum dose depth (R1), depth maximum dose (R5), percentage of dose in 1 mm depth (D1), percentage of dose in 2 mm depth (D2), percentage of maximum dose (Dmax), ratio of Tissue-phantom ratio depth 2 mm to 1 mm (TPR) for three types of smooth: colunm 1 - without smooth, column 2 - least square with mean value region of 5 mm, and column 3 - envelope with mean value region of 5 mm. 8 7 6 5 4 3 2 1-3 -2-1 1 (G) Off-axis distance (Inline) [mm] 2 (T) Figure 16. Profi le curve for cone 2 mm obtained with pinpoint chamber in the horizontal position. the effect of different type of smooth and it is possible to see a difference of almost 1% in the position of the maximum (R1) with the smooth envelope. Orientation of the detector The Figure 16 represents a profile curve obtained with pinpoint chamber in the horizontal position for the cone 2 mm. A comparison with the corresponding data of the Figure 9 shows that, if the chamber is put in the horizontal orientation, it cannot do the collection of charge properly. cones (4 to 7.5 mm). The best positioning of chambers for these measurements is vertical orientation, and the speed to acquisition is important to charge collection; it is convenient to use step by step instead of continuous measurements. It is also necessary a carefully handling of obtained data to not disturb or modify the results. The dosimeter size and its orientation can increase the signal s noise and disturb the results, and, as high speed, may result in wrong values. Another important concern is the evaluation of the data, because the smoothing tools can produce big distortions to the fi nal results. Conclusions The decision about the best dosimeter to data acquisition in small fi elds is a serious task. It involves each center s needs and availability of the equipments. For our purpose, the diode was the best detector for measures with small References 1. Fang-Fang Y. Dosimetric characteristics of Novalis shaped beam surgery unit. Med Phys. 22;29(8):1729-38. 2. Rustgi NS, Frye DM. Dosimetric characterization of radiosurgical beams with diamond detector. Med Phys. 1995;22(12):2117-21. Revista Brasileira de Física Médica.211;5(1):25-3. 29