White Paper: Technology and Clinical Overview of the Genii TouchSoft Coagulator monopolar probe

Similar documents
Journal of Hepatology and Gastroenterology

Electrosurgery Units (ESU s) and their variants

Electrosurgery Units (ESU s) and their variants. D. J. McMahon rev cewood

What tests should be done?

Application Notes. Electrosurgical Analyzer Primer

RAISING THE LEVEL OF PROTECTION IN YOUR ORS

J-470 ELECTROSURGE. Radio Frequency Electrosurgery Unit. Operation Instructions

Crossfire Footswitch. User Guide

MINIMAL ENERGY. MAXIMUM PRECISION. 1,2, The newest addition to Valleylab energy made to enhance the performance of your electrosurgical handpieces1,2,

Leksell Neuro Generator

ARC 200 HF SURGICAL GENERATOR SIMPLY SAFE

perioperativecpd.com The basics of diathermy

EXCELL NHP/T. A new step AheAd to the future ELECTROSURGICAL UNITS FOR MONOPOLAR AND BIPOLAR USE ISO ALSA IT ISO

ALSATOM SU MPC SU 50 MPC SU 100 MPC SU 140 MPC SU 140/D MPC SU 140/BD MPC

MICROWAVE DIATHERMY AND SURGICAL DIATHERMY DIATHERMICS

QA-ES III Electrosurgical Analyzer

Electrosurgery. KLS Martin ME 102. The compact Electrosurgical Unit with high Capacity

Electrosurgery. KLS Martin ME MB 2 SAFETY BY EASY USE

Andy Falconer. Principles of surgical diathermy

ESU-2000 Series Product Overview

ELECTROSURGICAL GENERATOR USER S GUIDE

surtron.com SURTRON ELECTROSURGICAL units

QA-ES III Electrosurgical Analyzer

Surgitron Dual Frequency 4.0 MHz Radiowave Technology

USER S GUIDE. User s Guide Bovie IDS-400

Radio Frequency Electrosurgical System Specialist

RF leakage current in electrosurgical units: Influence of the layout in taking measurements

Designed to improve physician productivity and patient care.

Single balloon Enteroscope system SIF-Q180 / OBCU

Instruction Manual. (English-English)

User s Guide Aaron 950

Voltage Current and Resistance II

Sonicator Plus 994 Specifications

DIATERMO MB 122 AND MB 160 MONO-BIPOLAR W W

troubleshooting LEVEL 1 training

Lab E5: Filters and Complex Impedance

LEEP System 1000 Table of Contents

Designed to improve physician productivity and patient care.

WRIST ARTHROSCOPY SYSTEM

Introducing the LumiCoil Platinum Fiducial Marker. Features of LumiCoil Platinum Fiducial Markers. Expanding Innovation in EUS

SERVICE MANUAL. Dynatron 950plus & Dynatron 650plus

PULSAR Generator Operator s Manual

Low-Cost Power Sources Meet Advanced ADC and VCO Characterization Requirements

Miscellaneous Accessories and technical equipment.

Model BiConiLog Antenna. User Manual

EXPERIMENT 1 PRELIMINARY MATERIAL

Using X-Y Displays APPLICATION BRIEF LAB WM312. May 29, Introduction. Summary

Guide. Installation. Wilson Electronics, Inc. In-Building Wireless Amplifi er. Contents:

SA18N-20 DATA SHEET. 20 db Fixed Attenuator N Male To N Female Up To 18 GHz Rated To 2 Watts With Passivated Stainless Steel Body.

An Alternative Technique to Perform Surgeriesin Hospital by Surgical Diathermy

LAB I. INTRODUCTION TO LAB EQUIPMENT

Aquamantys 3 Pump Generator User Guide. Revision F

SA26B-10 DATA SHEET. 10 db Fixed Attenuator SMA Male To SMA Female Up To 26 GHz Rated To 2 Watts With Passivated Stainless Steel Body.

Lab #1 Lab Introduction

Operator s Manual E L E C T R O S U R G I C A L U N I T

OPERATING MANUAL ELECTROSURGICAL UNIT

DATASHEET HI-201HS. Features. Applications. Ordering Information. Pinout (Switches Shown For Logic 1 Input) High Speed, Quad SPST, CMOS Analog Switch

ERBE ERBOTOM ICC 300. Instruction manual 08.00

Low / Medium Power Electrosurgery Review 2002

Wideband Quasi-Omni Antenna

Panaray MB4 TECHNICAL DATA SHEET. modular bass loudspeaker. Key Features. Product Overview. Technical Specifications

OPERATING MANUAL MINIDOP ES-100VX POCKET DOPPLER

BOVIE SURGI-CENTER PRO. Electrosurgical Generator. User's Guide

Meg1 (with Ligasmart)

AHLxxx Low-Voltage Nanopower Digital Switches

DIATERMO MB 122 AND MB 160 MONO-BIPOLAR W W

Electrosurgery. maxium INGENUITY IS ALWAYS QUITE SIMPLE

Choose flexibility Preserve tissue and cut complete sections

VA L U E A N A LY S I S B R I E F

QUICK SETUP GUIDE SECULIFE ESPRO

FreeSpace DS 16F VA TECHNICAL DATA SHEET. loudspeaker. Key Features. Product Overview. Technical Specifications

focused on implantable pulse generators for restoration of function in individuals paralyzed by stroke or SCI.

3D FINITE ELEMENT ANALYSIS FOR NON- ASYMMETRY STRUCTURE ANTENNA FOR MICROWAVE ABLATION THERAPY

Perclose ProGlide. Deployment Steps. Suture-Mediated Closure System Abbott. All rights reserved. ANZ /11

LAB I. INTRODUCTION TO LAB EQUIPMENT

NGTB15N60EG. IGBT - Short-Circuit Rated. 15 A, 600 V V CEsat = 1.7 V

Superseded. Small Cell IDA Antenna IDA-30F-KE-H2 DATA SHEET. Overview. Applications

LABORATORY 3 v1 CIRCUIT ELEMENTS

Radiofrequency Ablation Needle Electrodes. Procedural Algorithms and Lesion Geometry

Crossfire 2 Integrated Resection and Energy System

T CaRe ENERGY MADE TO MEASURE

Performing Cyclic Voltammetry Measurements Using Model 2450-EC or 2460-EC Electrochemistry Lab System

Endoscopy Equipment. D. J. McMahon rev cewood

ERBE. ERBOTOM ICC 80 / 50 Instruction manual 08.00

M302RM OPERATING MANUAL

Antenna. Wilson. Wide-Band Panel Antenna Window, Ceiling, Wall & Outdoor Mounts. Appearance of device and accessories may vary.

34134A AC/DC DMM Current Probe. User s Guide. Publication number April 2009

Design and Analysis of Two Slot Antenna for Microwave Ablation

its power (1,250mW/cm 2 ) is greatly superior to that of most other lamps and nearly as efficient as plasma lamps, with no rise of temperature;

High Frequency. ECT-CPG com shop ECT-CPG com

SPECS. Impulse (4 and 8 ohm) Two-Way Weather-Resistant Injection-Molded Speaker System SPECIFICATIONS. Built under U.S.

Dynatel 2250E/2273E Advanced Cable and Fault Locator

Using the isppac-powr1208 MOSFET Driver Outputs

Excel Tool: Plots of Data Sets

EXpERIENCE THE power Of LIGHT

ME 365 EXPERIMENT 1 FAMILIARIZATION WITH COMMONLY USED INSTRUMENTATION

INSTRUCTIONS ESG-400

FIELD EFFECT TRANSISTOR (FET) 1. JUNCTION FIELD EFFECT TRANSISTOR (JFET)

Sonicator Plus 940. Instruction Manual

LABORATORY 3 v3 CIRCUIT ELEMENTS

Transcription:

White Paper: Technology and Clinical Overview of the Genii TouchSoft Coagulator monopolar probe Prepared by Genii, Inc. St. Paul, MN, January 2012. (Updated April, 2015) All rights reserved. Introduction US Endoscopy acquired Genii, Inc. on April 7 th, 2017 The Genii TouchSoft Coagulator monopolar probe (TSC) is indicated for use through flexible endoscopes to provide monopolar hemostasis or tissue ablation in the gastrointestinal tract. It was developed specifically to meet a need for a lower cost alternative to argon plasma coagulation or bipolar coagulation for small, discrete lesions and to take advantage of the increasing prevalence of low voltage, depth limited, electrosurgery generator outputs. These contact coagulation outputs are typically known by such names as TouchSoft coagulation, Gentle Coag, Soft Coag, and the like*. While the TSC is an ideal design for newer TouchSoft coagulation, and other low voltage, depth limited, contact coagulation outputs, it can also successfully be used with standard coagulation ( Coag ) outputs of the type commonly used for polypectomy. The TSC has a Teflon sheath, 2.3 mm outer diameter, and a working length of 300 cm. It is compatible with all flexible endoscopes that can accommodate these dimensions. The product is shipped sterile and is intended for single use. Certifications The TSC has been certified to meet all IEC 60601-2-2 electrical standards for a high frequency rating of 6 kilo volts peak to peak. The TSC has passed all applicable biocompatibility testing and was cleared by FDA August 2, 2010 (K093079). The product is manufactured and sterilized in an ISO 13485 Registered CE and FDA compliant facility in the United States. Current Use The TSC was released for sale on May 12, 2011. Since then, it has been used on more than 4,000 patients with no reported complications. Physicians report successful use of the TSC for hemostasis or residual polyp tissue cleanup after cold biopsy of small sessile polyps, hemostasis or residual polyp tissue cleanup after snare polypectomy, and hemostasis of discrete oozing or bleeding lesions, AVMs, ESD applications and others. 1 P a g e

TouchSoft Coagulator monopolar probe treatment of a Jejunal Angioectasias during Double- Balloon Endoscopy. Electrosurgery Generator Considerations In electrosurgery, the terms monopolar and bipolar refer to the way the circuit is completed. In a monopolar application all of the energy leaving the active electrode (snare, TSC, sphinctertome, etc) at the treatment site returns to the generator via a dispersive (grounding) pad. In a bipolar (or multipolar) application all of the output energy both leaves and returns to the generator through the accessory (BiCap, GoldProbe and others). The final tissue effect from the use of any type of electrosurgery is most determined by the tissue, the type of accessory (broad surface or thin wire for example), the type of waveform (coag, cut, etc), the power curve (how the microprocessor responds to changing impedance during the power application), the time that the operator delivers power, and the power setting (in watts). The final tissue effect is much less dependent on the way the circuit is completed (bi, multi, or mono polar). Bipolar circuits may sometimes be more advantageous for some patients including those with some types of implanted devices. Accessory choice, power selected and time of power delivery is quite straightforward; nearly entirely operator dependent, and will not be addressed further here, nor will differences in impedances of tissue types. The waveform type and power curve both affect an understanding of the technology associated with the TSC and will be reviewed. Electrosurgery (ES) technology depends on high frequency alternating current delivered in one of two wave form patterns: continuous or modulated. In a continuous waveform, power is flowing 100% of the time that the operator is pressing the activation foot pedal. Continuous high frequency sine waves are associated with electrosurgical cutting if they reach peak voltages over 200 which can create a high enough current density to heat cellular water so quickly that the cell membranes burst. These bursting cells along a cleavage plane are what we refer to as electrosurgical cutting. Continuous waveforms held below this 200 Vp threshold are too weak to initiate cutting or to drive coagulation deeply into tissue. Consequently they are an ideal waveform for gentle, limited depth, contact coagulation. To increase coagulation and hemostasis further, and to balance the blend of cutting and coagulation, waveforms are interrupted or modulated to allow slower heating with lower current density. Voltage peaks are increased to drive the coagulation deeper into tissue. Names given to particular waveforms are not standardized, but typical "Coag" waveforms are modulated, being interrupted with the "off" or 'resting' cycle taking up 94% of the time (a 6% duty cycle) that a foot pedal is depressed. 2 P a g e

Output Power A continuous waveform is shown on the left, a modulated type on the right: As tissue becomes coagulated, it becomes more resistant to electric flow. Electrosurgery follows the law of Ohms which dictates that as this impedance rises, the power must decrease unless either the voltage or the current rises to compensate. Modern ES units use microprocessor control to manage this impedance to output ratio. A graph of the power versus the rise in impedance over time is referred to as a power curve. It is used to demonstrate how the software has been designed to change the power delivered in changing impedance conditions. Shown below are examples of widely available power curves. Note that in each case the microprocessor is responding to increasing impedance with increasing coagulation (increasing from left to right along the horizontal axis). Three initial power settings are shown on the vertical axis. Note the blue arrow. It shows that at the initial power setting of twenty watts, this microprocessor is maintaining the power output pretty much constantly over time as the tissue becomes increasingly coagulated and impedance is increasing. A usual Broad power curve (often used for polypectomy) Note the red arrow below. This graph shows an ideal power curve for use with a bipolar endostasis probe (BiCap or Goldprobe) in GI. The selected power ramps up quickly in the low impedance beginning stage (as in frank blood) to produce coagulation quickly; but the power drops off rapidly as the tissue impedance increases over time with continued application of power. By the time that tissue is observed clinically to be quite white and coagulated, the power still being delivered is less than 3 or 4 3 P a g e

watts. This drop off from the starting power is what gives these therapies their desired selflimiting tissue effect. They tend to produce coagulation with a limited depth of injury. A Narrow Bipolar power curve Output Power Output Power Note the Green arrow below which points out how similar a TouchSoft coagulation power curve is to the Ideal outputs for GI bipolar. The initial power is set higher to be able to drive a monopolar circuit, but the power still ramps up quickly and drops off quickly to give the selflimited, superficial tissue effect which is so desirable. A TouchSoft coagulation Narrow power curve 140 120 100 80 60 40 TouchSoft coagulation Mode Power vs Load Resistance Soft COAG 120 Soft COAG 60 Soft COAG Default 50 20 0 100 200 300 400 500 600 700 800 900 1000 1100 1200 1300 1400 1500 1600 1700 1800 1900 2000 Most newer electrosurgery generators marketed to GI include correct waveforms for bipolar probes and TouchSoft coagulation or Soft Coag outputs similar to these graphs. They all control voltage peaks to stay under about 200Vp most of the time, which means that no electrosurgical cutting is promoted. Only gentle, limited depth coagulation is created. 4 P a g e

TouchSoft coagulation Pre-Clinical Performance Data The following data was generated from a third-party testing lab using an ex vivo tissue model protocol. Three generator outputs were tested: Bipolar, TouchSoft coagulation and a standard Coag output labeled Forced. Accessories tested were the TouchSoft Coagulator monopolar probe and a commercially available brand of bipolar endostasis probe. Bipolar Probe with Bipolar output. TouchSoft Coagulator monopolar probe with TouchSoft coagulation output. Bipolar 60 watts 40 watts 30 watts TouchSoft Coag 0.00 5.00 10.00 milimeters Depth TouchSoft Coagulator with Standard Coag (Forced) Power settings comparable per output mode. Forced Coag Comparable Watts Each Mode 25 watts Forced 15 20 watts Ave Depth Ave Width Soft 40 Ave Depth Ave Width 15 watts Bipolar 15 0.00 2.00 4.00 6.00 milimeters 0.00 1.00 2.00 3.00 milimeters 4.00 5.00 TouchSoft Coagulator monopolar probe with Bipolar and TouchSoft coagulation output modes 2.5 Ave Depth Bipolar and TouchSoft coagulation 1.5 Bipol 0.5 15 watts20 watts40 watts60 watts 5 P a g e

Conclusion If width measurement alone is used, a Bipolar application of 20 watts is most like a TouchSoft coagulation waveform with the TouchSoft Coagulator monopolar probe at 40 watts. For depth, TS with TouchSoft Coagulator monopolar probe at 40 is most like Bipolar at 15. TS with Forced Coag 25 watts is most like TS with TouchSoft coagulation at 60 watts. In all cases, the tissue effect produced and performance of the TouchSoft Coagulator monopolar probe falls within the expected ranges of similar devices. It is anticipated that in the clinical setting, the TS will produce safe and efficacious results comparable to current technology. Genii gi4000 settings recommendation With TouchSoft coagulation, Soft Coag or other low voltage, depth limited modes, it is recommended that the physician begin between a range of 40 to 60 watts. 50 is a good starting point. A range of 20 to 30 watts as a start point is best for a Conmed 6500 Beamer Gentle output mode. If a standard, wide power curve Coag ( Standard ) mode is used, the recommendation is that a power setting of 25 watts or less be used and the physician should carefully limit the application time. Please see the Instructions for Use insert shipped with each box of TouchSoft Coagulator monopolar probes for complete information. *TouchSoft Coagulator and TouchSoft are registered trademarks of US Endoscopy. Soft Coag is a registered trademark of ERBE Electromedizin. US Endoscopy is a subsidiary of STERIS Corporation. Listening and delivering solutions, the US Endoscopy listening and delivering solutions design, and all marks denoted with or are registered with the U.S. Patent and Trademark Office, or trademarks owned by US Endoscopy. 2017 US Endoscopy. All Rights Reserved. 761429 Rev. A 6 P a g e