PXIe Contents CALIBRATION PROCEDURE. 10 GHz or 20 GHz RF Analog Signal Generator

Similar documents
Calibrating the NI 5653 requires you to install one of the following packages on the calibration system. NI-RFSA 2.4 or later NI-RFSG 1.

Contents. CALIBRATION PROCEDURE NI PXIe-5698

PXIe Contents. Required Software CALIBRATION PROCEDURE

NI PXIe Contents. Required Software CALIBRATION PROCEDURE. Dual-Output Programmable DC Power Supply

NI PXIe Contents CALIBRATION PROCEDURE. Four Channel Precision Source-Measure Unit (SMU)

NI PXIe Contents CALIBRATION PROCEDURE. Single-Channel Precision Source-Measure Unit (SMU)

NI PXIe-5171R. Contents. Required Software CALIBRATION PROCEDURE

Required Software. Related Documentation. Password. Calibration Interval

Contents. CALIBRATION PROCEDURE PXIe-5673 Vector Signal Generator

PXIe, 7½-Digit, ±1,000 V, Onboard 1.8 MS/s Isolated Digitizer, PXI Digital Multimeter

Contents. Software Requirements. CALIBRATION PROCEDURE NI PXIe-5663E

Contents. Software Requirements CALIBRATION PROCEDURE NI PXI-5663

Contents. Software. Documentation. CALIBRATION PROCEDURE NI PXIe-6674T

CALIBRATION PROCEDURE PXIe-4302/4303 and TB-4302C 32 Ch, 24-bit, 5 ks/s or 51.2 ks/s Simultaneous Filtered Data Acquisition Module. ni.

PXIe Contents. Required Software CALIBRATION PROCEDURE

Contents. CALIBRATION PROCEDURE NI PXIe-4463 DSA Analog Output

Contents. Software Requirements

Contents. CALIBRATION PROCEDURE NI PXIe-5668R 14 GHz and 26.5 GHz Signal Analyzer

Contents. CALIBRATION PROCEDURE NI PXIe GHz and 14 GHz RF Vector Signal Analyzer

Reconfigurable 6 GHz RF Vector Signal Transceiver with 1 GHz Bandwidth

GETTING STARTED GUIDE NI khz to 100 MHz, 50 Ω, AC Coupled RF Receiver Module

PXIe Contents CALIBRATION PROCEDURE. Reconfigurable 6 GHz RF Vector Signal Transceiver with 200 MHz Bandwidth

CALIBRATION PROCEDURE NI PXIe-4330/4331. Contents

Power Amplifier. Input Characteristics. SPECIFICATIONS NI PXIe-4610 Power Amplifier. Signal Range

PXIe Contents CALIBRATION PROCEDURE

Contents. Software Requirements. Documentation Requirements. Calibration Interval CALIBRATION PROCEDURE NI TB-4300B

NI PXIe Contents CALIBRATION PROCEDURE

GETTING STARTED GUIDE NI AI, ±5 V, 24 Bit, 51.2 ks/s/ch Simultaneous, AC/DC Coupling, IEPE AC Coupling

NI PXIe-5667 (3.6 GHz)

GETTING STARTED GUIDE NI AI, ±60 V, 24 Bit, 1 ks/s/ch Simultaneous

GETTING STARTED GUIDE NI AI, ±10 V, 24 Bit, 50 ks/s/ch Simultaneous

DEVICE SPECIFICATIONS Phase Matrix PXI GHz PXI Vector Signal Analyzer

NI 6624 Calibration Procedure

NI PXIe-5601 Specifications

These specifications apply to the PXIe-5113 with 64 MB and 512 MB of memory.

Input Characteristics. Measurement Range. SPECIFICATIONS NI PXIe-4357

Contents CALIBRATION PROCEDURE NI 5412

NI PXIe-5630 Specifications

Contents CALIBRATION PROCEDURE NI PXI-5422

PXIe Contents SPECIFICATIONS. 250 khz to 20 GHz RF Analog Signal Generator

PXIe Contents SPECIFICATIONS. RF Preselector Module

Contents. CALIBRATION PROCEDURE NI 5421/ MS/s Arbitrary Waveform Generator

USRP-2940/2942/2943/2944/2945

PXIe, 24-Channel, ±24 V, 50 ma Precision PXI Source Measure Unit

DEVICE SPECIFICATIONS NI PXIe Ch, 24-bit, 25.6 ks/s Simultaneous AC LVDT Input Module

Note Using the PXIe-5785 in a manner not described in this document might impair the protection the PXIe-5785 provides.

PXIe Contents SPECIFICATIONS. RF Amplifier

Analog Characteristics

NI PXIe-5663 Specifications

if it is used in a manner not described in the user documentation.

PXIe Contents SPECIFICATIONS. 3.3 GHz RF Analog Signal Generator

NI PXI Contents DEVICE SPECIFICATIONS. RF Attenuator

PXI Contents SPECIFICATIONS. 6.6 GHz RF Analog Signal Generator

Current ON state current ma minimum, 6 ma typical, 10 ma maximum OFF state current ma maximum

PXIe Contents SPECIFICATIONS. 14 GHz and 26.5 GHz Vector Signal Analyzer

PXI Contents SPECIFICATIONS. 1.3 GHz RF Analog Signal Generator

100 MHz Bandwidth, 12-Bit PXI Waveform Generator

is used in a manner not described in the user documentation.

Reconfigurable 6 GHz Vector Signal Transceiver with I/Q Interface

NI PXI-2797 Specifications 40 GHz 6 1 Terminated 50 Ω Multiplexer

SCXI MHz Quad 8x1 50 Ω Multiplexer

NI PXI-2521 Specifications 40-Channel DPST Relay Module

PXIe Contents. Definitions SPECIFICATIONS. 8.5 GHz Vector Network Analyzer

DEVICE SPECIFICATIONS NI PXIe-4302/ Ch, 24-bit, 5 ks/s or 51.2 ks/s Simultaneous Filtered Data Acquisition Module

Contents. CALIBRATION PROCEDURE NI PXIe-6555/6556. ni.com/manuals

NI PXI-2520 Specifications 80-Channel SPST Relay Module

PXIe Contents. Definitions SPECIFICATIONS. 8.5 GHz Vector Network Analyzer

GETTING STARTED GUIDE NI AI Differential/32 AI Single-Ended, ±200 mv to ±10 V, 16 Bit, 250 ks/s Aggregate, Fuel-Cell Measurements

NI Contents CALIBRATION PROCEDURE

NI PXI/PCI-5412 Specifications 14-Bit 100 MS/s Arbitrary Waveform Generator

Specifications for the NI PXI/PCI-6552/6551

NI PXI/PCI-5411/5431 Specifications

NI Contents CALIBRATION PROCEDURE

Terminology. Output Characteristics. DEVICE SPECIFICATIONS NI PXIe-4463 DSA Analog Output. ni.com/manuals

NI PXI/PXIe-2540 Specifications

NI PXI/PXIe-2541 Specifications 300 MHz Ω Matrix

SPECIFICATIONS PXIe Ch (8 ADC), 2 MS/s, bit, Flexible Resolution PXI Analog Input Module

NI PXI/PCI-5402/5406 Specifications 14/16-Bit, 20/40 MHz Arbitrary Function Generator

NI PXI-5422 Specifications 16-Bit 200 MS/s Arbitrary Waveform Generator

SPECIFICATIONS PXIe channel, 16-bit, 400 ks/s/ch, Ch-Ch Isolated Analog Input Module. ni.com/manuals

USRP Contents. Definitions SPECIFICATIONS. 10 MHz to 6 GHz Tunable RF Transceiver

SCXI Terminal Block

DEVICE SPECIFICATIONS NI PXIe-4304/ Ch, 24-bit, ±42 V, 5 ks/s or 51.2 ks/s Simultaneous Filtered Data Acquisition Module. ni.

NI PXIe Contents. About These Specifications DEVICE SPECIFICATIONS. 3 GHz Ω Multiplexer (SP16T)

NI PXI/PXIe Contents. About These Specifications DEVICE SPECIFICATIONS. Quad 2 1 Terminated Multiplexer

PXIe Contents SPECIFICATIONS. 32-Channel Digital Pattern Instrument

CDA Contents. Definitions SPECIFICATIONS. Clock Distribution Device

NI PXI/PCI-5411/5431 Specifications

NI PXI Contents. About These Specifications DEVICE SPECIFICATIONS GHz Dual Transfer Switch 50 Ω Relay Module

NI PXI/PXIe-2529 Specifications

PXIe Contents SPECIFICATIONS. 100 MVector/s PXI Digital Pattern Instrument

USRP Contents. Definitions. Conditions SPECIFICATIONS. Software Defined Radio Device

Contents CALIBRATION PROCEDURE NI PXI-5404

USRP Contents. Definitions. Conditions SPECIFICATIONS. Software Defined Radio Device

PCI Contents SPECIFICATIONS. 8-Channel, 12-Bit, 60 MHz PCI Oscilloscope Device

NI sbrio-9632/9642 Verification Procedure

This document lists the specifications for the NI PXI The PXI-5900 is a differential amplifier.

PXI MICROWAVE LOCAL OSCILLATOR MODULE

USRP Contents. Definitions SPECIFICATIONS. Software Defined Radio Reconfigurable Device

USRP Contents. Definitions. Conditions SPECIFICATIONS. Software Defined Radio Reconfigurable Device

Transcription:

CALIBRATION PROCEDURE PXIe-5654 10 GHz or 20 GHz RF Analog Signal Generator This document contains the verification and adjustment procedures for the PXIe-5654 RF Analog Signal Generator. Refer to ni.com/calibration for more information about calibration solutions. Contents Required Software...2 Related Documentation...2 Test Equipment...2 Test Conditions...5 Initial Setup... 5 Test System Characterization...5 Zeroing and Calibrating the Power Sensor... 6 As-Found and As-Left Limits... 6 Verification...6 Verifying 10 MHz Reference Frequency Accuracy...6 Verifying 10 MHz Reference Amplitude Accuracy...8 Verifying 100 MHz Reference Frequency Accuracy...10 Verifying 100 MHz Reference Amplitude Accuracy...11 Verifying RF OUT Frequency Accuracy... 13 Verifying RF OUT Phase Noise...14 Verifying RF OUT Maximum Power...17 Verifying Frequency Settling... 19 Optional Verification...23 Verifying RF OUT Amplitude Accuracy... 23 Verifying RF OUT Minimum Power... 25 Verifying PULSE IN Operation... 27 Adjustment... 29 Adjusting Frequency Reference Accuracy...29 Adjusting RF OUT Power...31 Reverification...32 Updating Calibration Date and Time... 32 Worldwide Support and Services... 32

Required Software Calibrating the PXIe-5654 requires you to install one of the following packages on the calibration system: NI-RFSG 14.5 NI-RFSA 14.5 Note NI-RFSG automatically installs when you install NI-RFSA. You can download all required software from ni.com/downloads. The software supports programming the verification procedures in the LabVIEW, C, and LabWindows /CVI application development environments (ADE) and programming the adjustment procedures in LabVIEW. When you install the software, you need to install support only for the ADE that you intend to use. Related Documentation For additional information, refer to the following documents as you perform the calibration procedure: NI PXIe-5654 Getting Started Guide NI RF Signal Generators Help NI PXIe-5654 Specifications Visit ni.com/manuals for the latest versions of these documents. Test Equipment NI recommends that you use particular equipment for the performance verification and adjustment procedures. If the recommended equipment is not available, select a substitute using the minimum requirements listed in the following table. 2 ni.com PXIe-5654 Calibration Procedure

Table 1. Recommended Equipment for PXIe-5654 Calibration Equipment Recommended Where Used Minimum Model Requirements Signal source analyzer Rohde & Schwarz (R&S) FSUP26 Signal Source Analyzer Verifications: 10 MHz reference frequency accuracy 100 MHz reference frequency accuracy RF OUT frequency accuracy Phase noise Frequency settling PULSE IN operation Frequency range: 10 MHz to 20 GHz Frequency counter resolution: 0.1 Hz Phase noise measurement using cross-correlation Frequency counter marker feature Adjusting reference accuracy Spectrum analysis capabilities Frequency reference Symmetricon 8040C Rubidium Frequency Standard Verifications: 10 MHz reference frequency accuracy 100 MHz reference frequency accuracy Frequency: 10 MHz Frequency accuracy: 10-9 (typically ±5 10-11 ) RF OUT frequency accuracy Frequency settling Adjusting reference accuracy Arbitrary waveform generator (AWG) PXIe-5451 Waveform Generator Verifications: Frequency settling PULSE IN operation PXIe-5654 Calibration Procedure National Instruments 3

Table 1. Recommended Equipment for PXIe-5654 Calibration (Continued) Equipment Recommended Where Used Minimum Model Requirements Power meter Anritsu ML2438A with a SC7413A Power Sensor Verifications: 10 MHz reference amplitude accuracy 100 MHz reference amplitude accuracy RF OUT amplitude accuracy Range: -55 dbm to 20 dbm Frequency range: 100 khz to 40 GHz Accuracy: <4.0% RF OUT maximum power RF OUT minimum power Adjusting RF OUT output power K(m)-to-K(m) cable, 36 in. Florida RF Labs KMS-160-36.0- KMS Verifications: 10 MHz reference frequency accuracy 100 MHz reference frequency accuracy Length: 36 in. Loss: <0.7 db/ft (typical) at 20 GHz RF OUT frequency accuracy RF OUT phase noise Frequency settling PULSE IN operation Adjustments: Reference accuracy RF OUT output power 4 ni.com PXIe-5654 Calibration Procedure

Test Conditions The following setup and environmental conditions are required to ensure the PXIe-5654 meets published specifications. Keep cabling as short as possible. Long cables act as antennas, picking up extra noise that can affect measurements. Verify that all connections to the PXIe-5654, including front panel connections and screws, are secure. Maintain an ambient temperature of 23 C ±5 C. Keep relative humidity between 10% and 90%, noncondensing. Allow a warm-up time of at least 30 minutes after the chassis is powered on. The warmup time ensures that the PXIe-5654 is at a stable operating temperature. In each verification procedure, insert a delay between configuring all devices and acquiring the measurement. This delay may need to be adjusted depending on the instruments used but should always be at least 1,000 ms for the first iteration, 1,000 ms when the power level changes, and 100 ms for each other iteration. Plug the chassis and the calibrator into the same power strip to avoid ground loops. Use an appropriate torque wrench to tighten all module RF connectors (SMA, 3.5 mm, or K). NI recommends a 0.565 N m (5 lb in.) wrench for SMA connectors and a 0.90 N m (8 lb in.) wrench for 3.5 mm or K connectors. Connect the frequency reference source to the REF IN connector on the back of the signal source analyzer and from the signal source analyzer REF OUT to the REF IN connector on the PXI Express chassis with standard BNC(m)-to-BNC(m) cables. This connection replaces the connection from the PXIe-5654 REF OUT (10 MHz) connector to the PXI Express chassis REF IN connector, if present. Ensure that the PXI chassis fan speed is set to HIGH, that the fan filters (if present) are clean, and that the empty slots contain slot blockers and filler panels. For more information about cooling, refer to the Maintain Forced-Air Cooling Note to Users document available at ni.com/manuals. Frequencies less than and equal to 10 GHz apply to the 10 GHz and 20 GHz PXIe-5654; frequencies above 10 GHz apply only to the 20 GHz PXIe-5654. Initial Setup Refer to the NI PXIe-5654 Getting Started Guide for information about how to install the software and the hardware and how to configure the device in Measurement & Automation Explorer (MAX). Test System Characterization The following procedure characterizes the test equipment used during verification. PXIe-5654 Calibration Procedure National Instruments 5

Zeroing and Calibrating the Power Sensor 1. Connect channel A of the power meter to the power sensor. 2. Zero and calibrate the power sensor using the built-in functions in the power meter. As-Found and As-Left Limits The as-found limits are the published specifications for the PXIe-5654. NI uses these limits to determine whether the PXIe-5654 meets the specifications when it is received for calibration. Use the as-found limits during initial verification. The as-left calibration limits are equal to the published NI specifications for the PXIe-5654, less guard bands for measurement uncertainty, temperature drift, and drift over time. NI uses these limits to reduce the probability that the instrument will be outside the published specification limits at the end of the calibration cycle. Use the as-left limits when performing verification after adjustment. Verification The performance verification procedures assume that adequate traceable uncertainties are available for the calibration references. Verifying 10 MHz Reference Frequency Accuracy This procedure verifies that the internal oven-controlled crystal oscillator (OCXO) is adjusted for correct frequency accuracy. 1. Connect the PXIe-5654 REF OUT 10 MHz front panel connector to the signal source analyzer RF IN front panel connector using the K(m)-to-K(m) cable as shown in the following figure. 6 ni.com PXIe-5654 Calibration Procedure

ACCESS ACTIVE Figure 1. Reference Frequency Verification (10 MHz) Equipment Setup (Reference Connections Not Shown) 2 1 250 khz 20 GHz Signal Generator 3 REF IN 1 20 MHz ±10 dbm NOM 17 dbm MAX PULSE IN 0.5V MIN +5.5V MAX REF OUT 10 MHz FM IN REF OUT 2 100 MHz AM IN ALC IN ±15 VDC MAX TRIG 3.3V CMOS 0.5V MIN IN 5.5V MAX IN RF OUT 250 khz 20 GHz ESD SENSITIVE ALL PORTS 50 Ω 1. PXIe-5654 RF Analog Signal Generator 2. Signal Source Analyzer 3. K(m)-to-K(m) Cable 2. Create a new device session for the PXIe-5654. 3. Set the PXIe-5654 RF frequency to 4 GHz, 0 dbm. 4. Export the reference clock to the RefOut terminal. 5. Initiate signal generation. 6. Disable the RF output. 7. Check the signal generation status and verify that there are no reported errors or warnings. 8. Configure the signal source analyzer using the following settings: Center frequency: 110 MHz Reference level: -50 dbm Frequency span: 1 khz PXIe-5654 Calibration Procedure National Instruments 7

Reference Clock source: External Frequency counter resolution: 0.1 Hz 9. Use the signal source analyzer frequency counter to measure the frequency of the peak at approximately 110 MHz. 10. Calculate the deviation using the following formula: Δ = Measured MHz 110 MHz 110 MHz ppm 11. Ensure that the deviation found in the previous step is less than the result of the following equation: Initial Accuracy + Aging + Temperature Stability where Initial Accuracy = ±0.1 ppm Aging = ±(0.01 ppm Number of Days Since Last Adjustment) Temperature Stability (15 C to 35 C) = ±0.2 ppm 12. Close the device session. Verifying 10 MHz Reference Amplitude Accuracy This procedure verifies that the internal 10 MHz reference circuitry is adjusted for correct amplitude accuracy. 1. Connect the power sensor to the PXIe-5654 REF OUT 10 MHz front panel connector as shown in the following figure. 8 ni.com PXIe-5654 Calibration Procedure

ACCESS ACTIVE Figure 2. Reference Amplitude Accuracy Verification (10 MHz) Equipment Setup 2 1 250 khz 20 GHz Signal Generator 3 REF IN 1 20 MHz ±10 dbm NOM 17 dbm MAX PULSE IN 0.5V MIN +5.5V MAX REF OUT 10 MHz FM IN REF OUT 2 100 MHz AM IN ALC IN ±15 VDC MAX TRIG 3.3V CMOS 0.5V MIN IN 5.5V MAX IN RF OUT 250 khz 20 GHz ESD SENSITIVE ALL PORTS 50 Ω 1. PXIe-5654 RF Analog Signal Generator 2. Power Meter 3. Power Sensor 2. Create a new device session for the PXIe-5654. 3. Set the PXIe-5654 RF frequency to 4 GHz, 0 dbm. 4. Export the reference clock to the RefOut terminal. 5. Initiate signal generation. 6. Disable the RF output. 7. Check the signal generation status and verify that there are no reported errors or warnings. 8. Measure the PXIe-5654 REF OUT 10 MHz front panel connector power using the power meter. 9. Correct the power meter reading for the 10 MHz measurement frequency and ensure that the corrected power is between 3 dbm and 7 dbm. 10. Close the device session. PXIe-5654 Calibration Procedure National Instruments 9

ACCESS ACTIVE Verifying 100 MHz Reference Frequency Accuracy This procedure verifies that the internal 100 MHz reference circuitry is adjusted for correct frequency accuracy. 1. Connect the PXIe-5654 REF OUT 2 100 MHz front panel connector to the signal source analyzer RF IN front panel connector using the K(m)-to-K(m) cable as shown in the following figure. Figure 3. Reference Frequency Accuracy Verification (100 MHz) Equipment Setup (Reference Connections Not Shown) 2 1 250 khz 20 GHz Signal Generator REF IN 1 20 MHz ±10 dbm NOM 17 dbm MAX PULSE IN 0.5V MIN +5.5V MAX 3 REF OUT 10 MHz FM IN REF OUT 2 100 MHz AM IN ALC IN ±15 VDC MAX TRIG 3.3V CMOS 0.5V MIN IN 5.5V MAX IN RF OUT 250 khz 20 GHz ESD SENSITIVE ALL PORTS 50 Ω 1. PXIe-5654 RF Analog Signal Generator 2. Signal Source Analyzer 3. K(m)-to-K(m) Cable 2. Create a new device session for the PXIe-5654. 3. Set the PXIe-5654 RF frequency to 4 GHz, 0 dbm. 4. Export the reference clock to the RefOut terminal. 5. Initiate signal generation. 6. Disable the RF output. 10 ni.com PXIe-5654 Calibration Procedure

7. Check the signal generation status and verify that there are no reported errors or warnings. 8. Configure the signal source analyzer using the following settings: Center frequency: 100 MHz Reference level: 10 dbm Frequency span: 250 khz Reference Clock source: External Frequency counter resolution: 0.1 Hz 9. Use the signal source analyzer frequency counter to measure the frequency of the peak at approximately 100 MHz. 10. Calculate the deviation using the following formula: Δ = Measured MHz 100 MHz 100 MHz ppm 11. Ensure that the deviation found in the previous step is less than the result of the following equation: Initial Accuracy + Aging + Temperature Stability where Initial Accuracy = ±0.1 ppm Aging = ±(0.01 ppm Number of Days Since Last Adjustment) Temperature Stability (15 C to 35 C) = ±0.2 ppm 12. Close the device session. Verifying 100 MHz Reference Amplitude Accuracy This procedure verifies that the internal 100 MHz reference circuitry is adjusted for correct amplitude accuracy. 1. Connect the power sensor to the PXIe-5654 REF OUT 2 100 MHz front panel connector as shown in the following figure. PXIe-5654 Calibration Procedure National Instruments 11

ACCESS ACTIVE Figure 4. Reference Amplitude Accuracy Verification (100 MHz) Equipment Setup 2 1 250 khz 20 GHz Signal Generator REF IN 1 20 MHz ±10 dbm NOM 17 dbm MAX REF OUT 10 MHz PULSE IN 0.5V MIN +5.5V MAX FM IN 3 REF OUT 2 100 MHz AM IN ALC IN ±15 VDC MAX TRIG 3.3V CMOS 0.5V MIN IN 5.5V MAX IN RF OUT 250 khz 20 GHz ESD SENSITIVE ALL PORTS 50 Ω 1. PXIe-5654 RF Analog Signal Generator 2. Power Meter 3. Power Sensor 2. Create a new device session for the PXIe-5654. 3. Set the PXIe-5654 RF frequency to 4 GHz, 0 dbm. 4. Export the reference clock to the RefOut terminal. 5. Initiate signal generation. 6. Disable the RF output. 7. Check the signal generation status and verify that there are no reported errors or warnings. 8. Use the power meter to measure the PXIe-5654 REF OUT 2 100 MHz front panel connector power. 9. Correct the power meter reading for the 100 MHz measurement frequency. Ensure that the corrected power is between 3 dbm and 7 dbm. 10. Close the device session. 12 ni.com PXIe-5654 Calibration Procedure

ACCESS ACTIVE Verifying RF OUT Frequency Accuracy This procedure verifies that the internal signal generation circuitry is adjusted for correct frequency accuracy. 1. Connect the PXIe-5654 RF OUT front panel connector to the signal source analyzer RF IN front panel connector using the K(m)-to-K(m) cable as shown in the following figure. Figure 5. RF OUT Frequency Accuracy Verification Equipment Setup (Reference Connections Not Shown) 2 1 250 khz 20 GHz Signal Generator REF IN 1 20 MHz ±10 dbm NOM 17 dbm MAX REF OUT 10 MHz PULSE IN 0.5V MIN +5.5V MAX FM IN 3 REF OUT 2 100 MHz AM IN ALC IN ±15 VDC MAX TRIG 3.3V CMOS 0.5V MIN IN 5.5V MAX IN RF OUT 250 khz 20 GHz ESD SENSITIVE ALL PORTS 50 Ω 1. PXIe-5654 RF Analog Signal Generator 2. Signal Source Analyzer 3. K(m)-to-K(m) Cable 2. Create a new device session for the PXIe-5654. 3. Create a list of test frequencies, including endpoints, containing 100 MHz and 1 GHz to 20 GHz with 1 GHz increments. 4. Set the RF output to 0 dbm. 5. Initiate signal generation. 6. Enable the RF output. PXIe-5654 Calibration Procedure National Instruments 13

7. Commit the settings to hardware. 8. Export the reference clock to the RefOut terminal. 9. Initiate signal generation. 10. Disable the RF output. 11. Check the signal generation status and verify that there are no reported errors or warnings. 12. Configure the signal source analyzer using the following settings: Center frequency: Frequency from the list in step 3 Reference level: 20 dbm Frequency span: 1 khz Reference Clock source: External Frequency counter resolution: 0.1 Hz 13. Measure the frequency of the peak that is returned by the signal source analyzer at approximately the corresponding point in the frequency array you created in step 3. 14. Calculate the deviation using the following formula: Δ = Measured MHz Expected MHz Expected MHz ppm 15. Ensure that the deviation found in the previous step is less than the result of the following equation: Initial Accuracy + Aging + Temperature Stability where Initial Accuracy = ±0.1 ppm Aging = ±(0.01 ppm Number of Days Since Last Adjustment) Temperature Stability (15 C to 35 C) = ±0.2 ppm 16. Repeat steps 4 through 15 for each subsequent point in the frequency array created in step 3. 17. Close the device session. Verifying RF OUT Phase Noise This procedure verifies that the internal frequency generation circuitry is adjusted for correct phase noise accuracy. 1. Connect the PXIe-5654 RF OUT front panel connector to the signal source analyzer RF IN front panel connector using the K(m)-to-K(m) cable as shown in the following figure. 14 ni.com PXIe-5654 Calibration Procedure

ACCESS ACTIVE Figure 6. RF OUT Phase Noise Verification Equipment Setup (Reference Connections Not Shown) 2 1 250 khz 20 GHz Signal Generator REF IN 1 20 MHz ±10 dbm NOM 17 dbm MAX REF OUT 10 MHz PULSE IN 0.5V MIN +5.5V MAX FM IN 3 REF OUT 2 100 MHz AM IN ALC IN ±15 VDC MAX TRIG 3.3V CMOS 0.5V MIN IN 5.5V MAX IN RF OUT 250 khz 20 GHz ESD SENSITIVE ALL PORTS 50 Ω 1. PXIe-5654 RF Analog Signal Generator 2. Signal Source Analyzer 3. K(m)-to-K(m) Cable 2. Create a new device session for the PXIe-5654. 3. Configure the signal source analyzer using the following settings: Phase noise measurement using cross-correlation phase-locked loops (PLLs) and internal generators Automatic device under test (DUT) detection Spur suppression enabled Fast Fourier transform (FFT) mode with Blackman-Harris window Measurement span from 100 Hz to 10 MHz (offset frequencies) 4. Use the signal source analyzer to perform a manual frequency sweep at the settings listed in the following table. PXIe-5654 Calibration Procedure National Instruments 15

Table 2. RF OUT Manual Sweep Settings Carrier Frequency Offset Resolution Bandwidth Cross Correlation 100 Hz to 300 Hz 10 Hz 100 300 Hz to 1 khz 30 Hz 1,000 1 khz to 3 khz 100 Hz 10,000 3 khz to 10 khz 300 Hz 10,000 10 khz to 30 khz 1 khz 10,000 30 khz to 100 khz 1 khz 5,000 100 khz to 300 khz 3 khz 1,000 300 khz to 1 MHz 10 khz 1,000 1 MHz to 3 MHz 30 khz 1,000 3 MHz to 10 MHz 100 khz 1,000 5. Set the RF frequency to 500 MHz and +8 dbm output power. 6. Commit the settings to hardware. 7. Enable the RF output. 8. Set the signal source analyzer center frequency to 500 MHz if the signal source analyzer does not automatically center for phase noise measurements. 9. Check the signal generation status and verify that there are no reported errors or warnings. 10. Measure the phase noise using the signal source analyzer. Record the measurements. 11. Ensure that the recorded measurements are within the limits set in the following tables. Offset Table 3. RF OUT Phase Noise (dbc/hz) As-Found Limits Carrier Frequency 500 MHz 1 GHz 5 GHz 10 GHz 20 GHz 100 Hz -107-101 -87-81 -75 1 khz -127-121 -109-103 -97 10 khz -135-130 -120-114 -108 100 khz -137-131 -122-117 -111 1 MHz -138-132 -125-119 -113 16 ni.com PXIe-5654 Calibration Procedure

Offset Table 4. RF OUT Phase Noise (dbc/hz) As-Left Limits Carrier Frequency 500 MHz 1 GHz 5 GHz 10 GHz 20 GHz 100 Hz -108.5-102.5-88.5-82.5-76.5 1 khz -128.5-122.5-110.5-104.5-98.5 10 khz -135.5-130.5-120.5-114.5-108.5 100 khz -138-132 -123-118 -112 1 MHz -138.5-132.5-125.5-119.5-113.5 12. Repeat steps 5 through 11 for 1 GHz, 5 GHz, 10 GHz, and 20 GHz carrier frequencies. Verifying RF OUT Maximum Power This procedure verifies that the internal signal generation circuitry produces the correct maximum output power. 1. Connect the power sensor to the PXIe-5654 RF OUT front panel connector as shown in the following figure. PXIe-5654 Calibration Procedure National Instruments 17

ACCESS ACTIVE Figure 7. RF OUT Maximum Power Verification Equipment Setup 2 1 250 khz 20 GHz Signal Generator REF IN 1 20 MHz ±10 dbm NOM 17 dbm MAX PULSE IN 0.5V MIN +5.5V MAX REF OUT 10 MHz FM IN REF OUT 2 100 MHz AM IN 43 ALC IN ±15 VDC MAX TRIG 3.3V CMOS 0.5V MIN IN 5.5V MAX IN RF OUT 250 khz 20 GHz ESD SENSITIVE ALL PORTS 50 Ω 1. PXIe-5654 RF Analog Signal Generator 2. Power Meter 3. Power Sensor 2. Initialize a new external calibration session for the PXIe-5654. 3. Configure the PXIe-5654 using the following settings: Frequency: 10 GHz Coarse Amplitude DAC: 1 Automatic Thermal Correction: 0 (Disable) 4. Initiate signal generation. 5. Configure the PXIe-5654 using the following settings: Frequency: 1 MHz Coarse Amplitude DAC: 32767 6. Create a list of test frequencies using the following table. Table 5. RF OUT Maximum and Minimum Power Test Frequencies Start Frequency (MHz) Stop Frequency (MHz) Frequency Step (MHz) 0.25 0.25 250 20,000 250 18 ni.com PXIe-5654 Calibration Procedure

7. Configure the PXIe-5654 using the following settings: Frequency: Start Frequency from the list created in step 6 Coarse Amplitude DAC: 65535 Output Enabled: TRUE 8. Measure the PXIe-5654 RF OUT front panel connector power using the power meter. Correct the power meter reading for the RF frequency. 9. Verify that the output power measured in step 8 meets the limits in the following table. Table 6. RF OUT Maximum Power Limits Frequency (MHz) As-Found Limit (dbm) As-Left Limit (dbm) 0.25 to 250 >10 >12 >250 to 2,800 >13 >14.5 >2,800 to 4,370 >13 >15.2 >4,370 to 12,100 >13 >14 >12,100 to 18,000 >13 >15 >18,000 to 20,000 >12 >15.5 10. Repeat steps 7 through 9 for all test frequencies you created in step 6. 11. Stop signal generation. 12. Close the external calibration session. Verifying Frequency Settling This procedure verifies that the PXIe-5654 settles to a final frequency in the correct amount of time. 1. Connect the signal source analyzer RF input to the PXIe-5654 RF OUT front panel connector using the K(m)-to-K(m) cable. 2. Connect the PXIe-5451 CH 1+/Q+ output to two separate connectors using a Tee connector. a) Connect the PXIe-5451 CH 1+/Q+ output to the signal source analyzer rear panel EXT TRIG/GATE IN connector. (Not pictured.) b) Connect the PXIe-5451 CH 1+/Q+ output to the PXIe-5654 front panel TRIG connector as shown in the following figure. PXIe-5654 Calibration Procedure National Instruments 19

ACCESS ACTIVE 1 20 MHz ±10 dbm NOM 17 dbm MAX 10 MHz 100 MHz 3.3V CMOS 0.5V MIN IN 5.5V MAX IN ESD SENSITIVE ALL PORTS 50 Ω ±15 VDC MAX 0.5V MIN +5.5V MAX 250 khz 20 GHz Figure 8. Frequency Settling Verification Equipment Setup 3 1 2 NI PXIe-5451 16-bit 400 M/s Arbitrary Waveform Generator 250 khz 20 GHz Signal Generator REF IN PULSE IN REF OUT FM IN 4 REF OUT 2 AM IN 5 TRIG ALC IN RF OUT 1. PXIe-5451 Waveform Generator 2. PXIe-5654 RF Analog Signal Generator 3. Signal Source Analyzer 4. K(m)-to-K(m) Cable 5. K(m)-to-K(m) Cable 3. Initialize new device sessions for the PXIe-5654 and PXIe-5451. 4. Create a list of test frequencies within the frequency limits listed in the following tables. Start Frequency (MHz) Table 7. Frequency Settling Settings/Limits (10 GHz PXIe-5654) Stop Frequency (MHz) Fast Settling Time (µs) Standard Settling Time (µs) As-Found As-Left As-Found As-Left 160.957 164.960 150 140 1,150 1,115 164.960 160.957 150 140 1,150 1,115 160.957 10,000 100 90 1,000 970 10,000 160.957 1,150 1,115 2,000 1,900 10,000 240 1,150 1,115 2,000 1,900 10,000 500 100 90 1,000 970 20 ni.com PXIe-5654 Calibration Procedure

Table 7. Frequency Settling Settings/Limits (10 GHz PXIe-5654) (Continued) Start Frequency (MHz) Stop Frequency (MHz) Fast Settling Time (µs) Standard Settling Time (µs) As-Found As-Left As-Found As-Left 10,000 2,000 100 90 1,000 970 10,000 4,000 100 90 1,000 970 10,000 6,000 100 90 1,000 970 Start Frequency (MHz) Table 8. Frequency Settling Settings/Limits (20 GHz PXIe-5654) Stop Frequency (MHz) Fast Settling Time (µs) Standard Settling Time (µs) As-Found As-Left As-Found As-Left 160.957 164.960 150 140 1,150 1,115 164.960 160.957 150 140 1,150 1,115 160.957 10,557.49 100 90 1,000 970 10,557.49 160.957 1,150 1,115 2,000 1,900 10,000 240 1,150 1,115 2,000 1,900 10,000 500 100 90 1,000 970 10,000 2,000 100 90 1,000 970 10,000 4,000 100 90 1,000 970 10,000 6,000 100 90 1,000 970 10,000 15,000 100 90 1,000 970 20,602.51 164.96 1,150 1,115 2,000 1,900 164.96 20,602.51 100 90 1,000 970 20,602.51 10,557.49 100 90 1,000 970 10,557.49 20,602.51 100 90 1,000 970 5. Configure the PXIe-5654 for the frequency list created in step 4 using the following settings: Configuration List Step Trigger Type: Digital edge Configuration List Step Digital Edge Source: TrigIn Output Enabled: TRUE 6. Configure the signal source analyzer using the following setting: Reference level: 10 dbm 7. Abort the PXIe-5451 generation. PXIe-5654 Calibration Procedure National Instruments 21

8. Configure the PXIe-5451 using the following settings: Output Mode: 0 (Standard Function) Channel: 1 Waveform: Sine Frequency: 10 MHz Amplitude: 1 V DC Offset: 2.5 V Route Signal from: 0 (none) Route Signal to: 0 (RTSI 0) 9. Commit the PXIe-5451 settings to hardware. 10. Enable the PXIe-5451 output channel configured in step 8. 11. Repeat steps 7 through 10. 12. Wait 5 seconds. 13. Configure the signal source analyzer using the following settings: RF attenuation: 30 db Span: 5 MHz Center frequency: Stop frequency from the list created in step 4 Resolution bandwidth: Auto Video bandwidth: Auto 14. Configure the signal source analyzer for Single Sweep. 15. Move the marker to the center frequency from step 13. 16. Wait 100 milliseconds. 17. Read the marker amplitude. 18. Configure the signal source analyzer using the following settings: Reference level: 6 db above the amplitude in step 17 Sweep: Continuous Vertical scale: Linear Span: Zero span 19. Wait 1 second. 20. Read the marker amplitude. 21. Configure the signal source analyzer using the following settings: Trigger source: External External trigger level: 1 V Configure trigger polarity: Positive Sweep time: Settling time from the table in step 4 Sweep time: Coupled 22. Wait 1 second. 23. Repeat steps 7 through 10. 24. Wait 5 seconds. 25. Repeat steps 7 through 10. 22 ni.com PXIe-5654 Calibration Procedure

26. Wait the amount of settling time from the table in step 4. 27. Fetch the Y-axis from the signal source analyzer. 28. Convert the Y-axis from log to linear with a scale of 100, using 6 db above the level in step 20 as the reference level. 29. Calculate the switch time from the data in step 28. 30. Verify the switch time from step 29 meets the limits from the table in step 4. 31. Stop the PXIe-5451 signal generation. 32. Close the PXIe-5451 session. Optional Verification Use the following procedures to verify nonwarranted specifications for the PXIe-5654. Verifying RF OUT Amplitude Accuracy This procedure verifies that the internal signal generation circuitry is adjusted for correct amplitude accuracy. 1. Connect the power sensor to the PXIe-5654 RF OUT front panel connector as shown in the following figure. PXIe-5654 Calibration Procedure National Instruments 23

ACCESS ACTIVE Figure 9. RF OUT Amplitude Accuracy Verification Equipment Setup 2 1 250 khz 20 GHz Signal Generator REF IN 1 20 MHz ±10 dbm NOM 17 dbm MAX PULSE IN 0.5V MIN +5.5V MAX REF OUT 10 MHz FM IN REF OUT 2 100 MHz AM IN 43 ALC IN ±15 VDC MAX TRIG 3.3V CMOS 0.5V MIN IN 5.5V MAX IN RF OUT 250 khz 20 GHz ESD SENSITIVE ALL PORTS 50 Ω 1. PXIe-5654 RF Analog Signal Generator 2. Power Meter 3. Power Sensor 2. Create a new device session for the PXIe-5654. 3. Create a list not a configuration list of test frequencies using the settings in the following table. Power Level (dbm) Table 9. RF OUT Amplitude Accuracy Test Settings Start Frequency (MHz) Stop Frequency (MHz) Frequency Step (MHz) 10 0.25 0.25 250 20,000 250 0 0.25 0.25 250 20,000 250-7 0.25 0.25 250 20,000 250 24 ni.com PXIe-5654 Calibration Procedure

Power Level (dbm) Table 9. RF OUT Amplitude Accuracy Test Settings (Continued) Start Frequency (MHz) Stop Frequency (MHz) -10 0.25 0.25 250 20,000 250 Frequency Step (MHz) 4. Set the RF frequency to the first value of the frequency array you created in step 3 and set the RF output to 10 dbm, which is the Expected Power (dbm). 5. Commit the settings to hardware. 6. Enable the RF output. 7. Check the signal generation status and verify that there are no reported errors or warnings. 8. Measure the PXIe-5654 RF OUT front panel connector power using the power meter. Correct the power meter reading for the RF frequency. This measurement is the Measured Power (dbm). 9. Calculate the Power Deviation (db) using the following formula: Power Deviation (db) = Measured Power (dbm) - Expected Power (dbm) 10. Verify that the Power Deviation (db) results in the previous step are within the limits listed in the following table. Table 10. RF OUT Amplitude Accuracy Limits Frequency As-Found Limit As-Left Limit 250 khz to 250 MHz ±2.0 db ±2.0 db >250 MHz to 20 GHz ±1.5 db ±1.5 db 11. Repeat steps 7 through 10 for each test point in the frequency array created in step 3. 12. Repeat steps 7 through 11 for output power for each test point in the list created in step 3. 13. Close the device session. Verifying RF OUT Minimum Power This procedure verifies that the internal signal generation circuitry can produce the correct minimum output power. 1. Connect the power sensor to the PXIe-5654 RF OUT front panel connector as shown in the following figure. PXIe-5654 Calibration Procedure National Instruments 25

ACCESS ACTIVE Figure 10. RF OUT Minimum Power Verification Equipment Setup 2 1 250 khz 20 GHz Signal Generator REF IN 1 20 MHz ±10 dbm NOM 17 dbm MAX PULSE IN 0.5V MIN +5.5V MAX REF OUT 10 MHz FM IN REF OUT 2 100 MHz AM IN 43 ALC IN ±15 VDC MAX TRIG 3.3V CMOS 0.5V MIN IN 5.5V MAX IN RF OUT 250 khz 20 GHz ESD SENSITIVE ALL PORTS 50 Ω 1. PXIe-5654 RF Analog Signal Generator 2. Power Meter 3. Power Sensor 2. Initialize a new external calibration session for the PXIe-5654. 3. Configure the PXIe-5654 using the following settings: Frequency: 10 GHz Coarse Amplitude DAC: 1 Automatic Thermal Correction: Disabled 4. Initiate signal generation. 5. Configure the PXIe-5654 using the following settings: Frequency: 1 MHz Coarse Amplitude DAC: 32767 6. Create a list of test frequencies using the following table. Table 11. RF OUT Maximum and Minimum Power Test Frequencies Start Frequency (MHz) Stop Frequency (MHz) Frequency Step (MHz) 0.25 0.25 250 20,000 250 26 ni.com PXIe-5654 Calibration Procedure

7. Configure the PXIe-5654 using the following settings: Frequency: Start Frequency from the list created in step 6 Coarse Amplitude DAC: 0 Output Enabled: TRUE 8. Measure the PXIe-5654 RF OUT front panel connector power using the power meter. Correct the power meter reading for the RF frequency. 9. Verify that the power output measured in step 8 meets the requirements in the following table. Table 12. RF OUT Minimum Power Limits Frequency (MHz) As-Found Limit (dbm) As-Left Limit (dbm) 0.25 to 250 <-13 <-13 >250 to 2,800 <-10 <-10 >2,800 to 4,000 <-9 <-9 >4,000 to 10,400 <-10 <-10 >10,400 to 19,000 <-7 <-7 >19,000 to 20,000 <-10 <-10 10. Repeat steps 7 through 9 for each test frequency in the list you created in step 6. 11. Close the external calibration session. Verifying PULSE IN Operation This procedure verifies that the PXIe-5654 PULSE IN connection is functioning. 1. Connect the PXIe-5451 to the PXIe-5654 PULSE IN front panel connector using the K(m)-to-K(m) cable. 2. Connect the signal source analyzer to the PXIe-5654 RF OUT front panel connector. The completed hardware setup is shown in the following figure. PXIe-5654 Calibration Procedure National Instruments 27

ACCESS ACTIVE 1 20 MHz ±10 dbm NOM 17 dbm MAX 10 MHz 100 MHz 3.3V CMOS 0.5V MIN IN 5.5V MAX IN ESD SENSITIVE ALL PORTS 50 Ω ±15 VDC MAX 0.5V MIN +5.5V MAX 250 khz 20 GHz Figure 11. PULSE IN Operation Verification Equipment Setup 3 1 2 NI PXIe-5451 16-bit 400 M/s Arbitrary Waveform Generator 250 khz 20 GHz Signal Generator 4 REF IN REF OUT FM IN PULSE IN 5 REF OUT 2 AM IN ALC IN TRIG RF OUT 1. PXIe-5451 Waveform Generator 2. PXIe-5654 RF Analog Signal Generator 3. Signal Source Analyzer 3. Initialize a new device session for the PXIe-5654. 4. K(m)-to-K(m) Cable 5. K(m)-to-K(m) Cable 4. Configure the signal source analyzer using the following settings: Reference level: 5 dbm Start frequency: 0.25 MHz Stop frequency: 10,000 MHz Span (khz): 2 KHz Resolution bandwidth: 30 Hz Video bandwidth: 100 Hz RF attenuation: 30 db 5. Configure the PXIe-5654 using the following settings: Pulse Modulation Enabled: TRUE Frequency: 5 GHz Power Level: 10 dbm Output Enabled: TRUE 28 ni.com PXIe-5654 Calibration Procedure

6. Configure the PXIe-5451 using the following settings: Output Mode: 0 (Standard Function) Channel Name: 1 Waveform: 5 Frequency: 10 MHz Amplitude: 1 V DC Offset: 2.5 V Output Enable: TRUE 7. Sweep the signal source analyzer and read the marker. This value is the OnPower. 8. Configure the PXIe-5451 using the following settings: DC offset: 0 V Output Enable: TRUE 9. Sweep the signal source analyzer and read the marker. This value is the OffPower. 10. Verify that the following equation is true. OnPower - OffPower 80 db 11. Stop signal generation. 12. Close the instrument sessions. Adjustment This section describes the steps needed to adjust the PXIe-5654 to meet published specifications. Adjusting Frequency Reference Accuracy This procedure adjusts the PXIe-5654 frequency accuracy performance using a signal source analyzer. This adjustment yields a more accurately tuned voltage-controlled oscillator (VCO) frequency. 1. Connect the PXIe-5654 REF OUT 10 MHz front panel connector to the signal source analyzer RF IN front panel connector using the K(m)-to-K(m) cable as shown in the following figure. PXIe-5654 Calibration Procedure National Instruments 29

ACCESS ACTIVE Figure 12. Frequency Reference Accuracy Adjustment Equipment Setup (Reference Connections Not Shown) 2 1 250 khz 20 GHz Signal Generator 3 REF IN 1 20 MHz ±10 dbm NOM 17 dbm MAX PULSE IN 0.5V MIN +5.5V MAX REF OUT 10 MHz FM IN REF OUT 2 100 MHz AM IN ALC IN ±15 VDC MAX TRIG 3.3V CMOS 0.5V MIN IN 5.5V MAX IN RF OUT 250 khz 20 GHz ESD SENSITIVE ALL PORTS 50 Ω 1. PXIe-5654 RF Analog Signal Generator 2. Signal Source Analyzer 3. K(m)-to-K(m) Cable 2. Configure the signal source analyzer using the following settings: Center frequency: 10 GHz Reference level: 10 dbm Frequency counter resolution: 0.1 Hz Reference Clock source: External 3. Call the nirfsg Initialize External Calibration VI using the appropriate password. Note The default password for password-protected operations is NI. 4. Call the nirfsg 5654 OCXO Cal Initialize VI. 5. Call the nirfsg 5654 OCXO Cal Configure VI. 6. Use the signal source analyzer to measure the frequency at the peak of the signal. 30 ni.com PXIe-5654 Calibration Procedure

ACCESS ACTIVE 7. Use the frequency measured in step 6 as the value of the measured frequency input of the nirfsg 5654 OCXO Cal Adjust VI. 8. Repeat steps 5 through 7 until the OCXO calibration complete output of the nirfsg 5654 OCXO Cal Adjust VI returns a value of TRUE. 9. Call the nirfsg Close External Calibration VI to close the session. 10. Set the write calibration to hardware? parameter to TRUE to store the results to the EEPROM on the PXIe-5654. Adjusting RF OUT Power This procedure adjusts the PXIe-5654 RF OUT power using a power meter. This adjustment yields a more accurate output power for the PXIe-5654 RF OUT front panel connectors. 1. Connect the power sensor to the PXIe-5654 RF OUT front panel connector as shown in the following figure. Figure 13. RF OUT Power Adjustment Equipment Setup 2 1 250 khz 20 GHz Signal Generator REF IN 1 20 MHz ±10 dbm NOM 17 dbm MAX PULSE IN 0.5V MIN +5.5V MAX REF OUT 10 MHz FM IN REF OUT 2 100 MHz AM IN 43 ALC IN ±15 VDC MAX TRIG 3.3V CMOS 0.5V MIN IN 5.5V MAX IN RF OUT 250 khz 20 GHz ESD SENSITIVE ALL PORTS 50 Ω 1. PXIe-5654 RF Analog Signal Generator 2. Power Meter 3. Power Sensor 2. Call the nirfsg Initialize External Calibration VI using the appropriate password. Note The default password for password-protected operations is NI. PXIe-5654 Calibration Procedure National Instruments 31

3. Call the nirfsg 5654 Power Cal Initialize VI. 4. Call the nirfsg 5654 Power Cal Configure VI. 5. Measure the PXIe-5654 output power with the power meter. Correct the measurement for the frequency output in step 4. 6. Use the power measured in step 5 as input of the nirfsg Power Cal Adjust VI. 7. Repeat steps 4 through 6 until the power calibration complete output of the nirfsg 5654 Power Cal Adjust VI returns a value of TRUE. 8. Call the nirfsg Close External Calibration VI to close the session. 9. Set the write calibration to hardware? parameter to TRUE to store the results to the EEPROM on the PXIe-5654. Reverification Repeat the Verification section to determine the as-left status of the PXIe-5654. Note If any test fails reverification after performing an adjustment, verify that you have met the test conditions before returning your PXIe-5654 to NI. Refer to the Worldwide Support and Services section for information about support resources or service requests. Related Information Test Conditions on page 5 Updating Calibration Date and Time This procedure updates the date and time of the last calibration of the PXIe-5654. 1. Call the nirfsg Initialize External Calibration VI. 2. Call the nirfsg Update External Calibration Date and Time VI. 3. Call the nirfsg Close External Calibration VI to close the session. Set the write calibration to hardware? parameter to TRUE to store the results to the EEPROM on the PXIe-5654. Worldwide Support and Services The NI website is your complete resource for technical support. At ni.com/support, you have access to everything from troubleshooting and application development self-help resources to email and phone assistance from NI Application Engineers. Visit ni.com/services for NI Factory Installation Services, repairs, extended warranty, and other services. Visit ni.com/register to register your NI product. Product registration facilitates technical support and ensures that you receive important information updates from NI. A Declaration of Conformity (DoC) is our claim of compliance with the Council of the European Communities using the manufacturer s declaration of conformity. This system 32 ni.com PXIe-5654 Calibration Procedure

affords the user protection for electromagnetic compatibility (EMC) and product safety. You can obtain the DoC for your product by visiting ni.com/certification. If your product supports calibration, you can obtain the calibration certificate for your product at ni.com/calibration. NI corporate headquarters is located at 11500 North Mopac Expressway, Austin, Texas, 78759-3504. NI also has offices located around the world. For telephone support in the United States, create your service request at ni.com/support or dial 1 866 ASK MYNI (275 6964). For telephone support outside the United States, visit the Worldwide Offices section of ni.com/ niglobal to access the branch office websites, which provide up-to-date contact information, support phone numbers, email addresses, and current events. PXIe-5654 Calibration Procedure National Instruments 33

Refer to the NI Trademarks and Logo Guidelines at ni.com/trademarks for information on NI trademarks. Other product and company names mentioned herein are trademarks or trade names of their respective companies. For patents covering NI products/technology, refer to the appropriate location: Help»Patents in your software, the patents.txt file on your media, or the National Instruments Patent Notice at ni.com/patents. You can find information about end-user license agreements (EULAs) and third-party legal notices in the readme file for your NI product. Refer to the Export Compliance Information at ni.com/ legal/export-compliance for the NI global trade compliance policy and how to obtain relevant HTS codes, ECCNs, and other import/export data. NI MAKES NO EXPRESS OR IMPLIED WARRANTIES AS TO THE ACCURACY OF THE INFORMATION CONTAINED HEREIN AND SHALL NOT BE LIABLE FOR ANY ERRORS. U.S. Government Customers: The data contained in this manual was developed at private expense and is subject to the applicable limited rights and restricted data rights as set forth in FAR 52.227-14, DFAR 252.227-7014, and DFAR 252.227-7015. 2016 National Instruments. All rights reserved. 376182A-01 Oct16