Procedures Guide. Tektronix. HDMI Sink Instruments Differential Impedance Measurement

Similar documents
Test ID 5-15 Utility Line Impedance Test Procedures Guide

Tektronix Inc. DisplayPort Standard. Revision Tektronix MOI for Cable Tests (DSA8200 based sampling instrument with IConnect software)

80E01, 80E02, 80E03, 80E04 & 80E06 Electrical Sampling Modules

High Data Rate Characterization Report

High Speed Characterization Report

Reference. TDS6000 Series Digital Storage Oscilloscopes

Supplement. TDS5032 and TDS5034 Digital Phosphor Oscilloscopes

High Speed Characterization Report

High Data Rate Characterization Report

High Speed Characterization Report

Line Impedance Analyzer TDR 3000

High Data Rate Characterization Report

High Data Rate Characterization Report

Faculty of Engineering, Thammasat University

Agilent MOI for MIPI D-PHY Conformance Tests Revision 1.00 Dec-1, 2011

EENG-201 Experiment # 4: Function Generator, Oscilloscope

Keysight MOI for MIPI D-PHY Conformance Tests Revision Oct, 2014

Agilent MOI for MIPI M-PHY Conformance Tests Revision Mar 2014

Agilent 33522A Function Arbitrary Waveform Generator. Tektronix TDS 3012B Oscilloscope

IConnect and MeasureXtractor Signal Integrity, TDR, and S-parameter Analysis Software

DPO3PWR, MDO3PWR and DPO4PWR Power Analysis Modules ZZZ User Manual

Agilent Correlation between TDR oscilloscope and VNA generated time domain waveform

Introduction to Basic Laboratory Instruments

High Speed Characterization Report

Introduction to basic laboratory instruments

ECE65 Introduction to the Function Generator and the Oscilloscope Created by: Eldridge Alcantara (Spring 2007)

Oscilloscope Operation. Visualizing Signals and Making Measurements

Reference. TDS3000B Series Digital Phosphor Oscilloscopes Copyright Tektronix, Inc. All rights reserved.

CPE 310L EMBEDDED SYSTEM DESIGN LABORATORY

Introduction to Oscilloscopes Instructor s Guide

TDR Impedance Measurements: A Foundation for Signal Integrity

Keysight MOI for USB Type-C Connectors & Cable Assemblies Compliance Tests (Type-C to Legacy Cable Assemblies)

LAB #3: Virtual Instruments; Behavior of Second-Order Systems

Serial ATA International Organization

CPE 100L DIGITAL LOGIC DESIGN I DESIGN LABORATORY LABORATORY 1 LAB SAFETY QUIZ & LAB EQUIPMENT USE TUTORIAL UNIVERSITY OF NEVADA, LAS VEGAS GOALS:

Keysight MOI for USB Type-C Connectors & Cable Assemblies Compliance Tests (Type-C to Legacy Cable Assemblies)

Measuring Power Supply Switching Loss with an Oscilloscope

Combinational logic: Breadboard adders

MODEL745 FOUR CHANNEL DIGITAL DELAY/ PULSE GENERATOR

ArbStudio Triggers. Using Both Input & Output Trigger With ArbStudio APPLICATION BRIEF LAB912

LAB I. INTRODUCTION TO LAB EQUIPMENT

Measuring Hot TDR and Eye Diagrams with an Vector Network Analyzer?

Tektronix Courseware. Academic Labs. Sample Labs from Popular Electrical and Electronics Engineering Curriculum

UCE-DSO210 DIGITAL OSCILLOSCOPE USER MANUAL. FATIH GENÇ UCORE ELECTRONICS REV1

Documentation. PerfectPulse Fast Edge Signal Generator

Online Help. DPOPWR Power Measurement and Analysis Software Adapted from DPOPWR Power Measurement and Analysis Online Help

Introduction to basic laboratory instruments

LABORATORY 4. Palomar College ENGR210 Spring 2017 ASSIGNED: 3/21/17

3.003 Lab 3 Part A. Measurement of Speed of Light

TDS6604B & TDS6804B Digital Storage Oscilloscopes Specifications and Performance Verification

P7600 Series TriMode Probes

High Speed Competitive Comparison Report. Samtec MMCX-J-P-H-ST-TH1 Mated With MMCX-P-P-H-ST-TH1 Competitor A (Mated Set) Competitor B (Mated Set)

Limitations And Accuracies Of Time And Frequency Domain Analysis Of Physical Layer Devices

Ph 3455 The Franck-Hertz Experiment

Lab 0: Introduction to basic laboratory instruments. Revised by Dan Hoang & Tai-Chang Chen 03/30/2009

High Speed Characterization Report

PI3HDMI201 PI3HDMI201 HDMI-HDMI Demo Board Rev.A User Manual by Ada Yip

UCE-DSO212 DIGITAL OSCILLOSCOPE USER MANUAL. UCORE ELECTRONICS

Instruction Manual. P7350SMA 5 GHz Differential Probe

The oscilloscope and RC filters

The Oscilloscope. Vision is the art of seeing things invisible. J. Swift ( ) OBJECTIVE To learn to operate a digital oscilloscope.

Agilent 81133A/81134A

Introduction to Lab Instruments

UNIVERSITY OF CALIFORNIA, SANTA BARBARA Department of Electrical and Computer Engineering. ECE 2A & 2B Laboratory Equipment Information

TLA5000 Golden Demo for Hardware Engineers

Chapter 3 Simulation

Exponential Waveforms

UNCONTROLLED COPY OF RELEASED

Agilent N2740A Education Training Kit for 1000 Series Oscilloscopes

DEPARTMENT OF THE ARMY TECHNICAL BULLETIN

P a g e 1 ST985. TDR Cable Analyzer Instruction Manual. Analog Arts Inc.

How to Test OnFILTER Ground EMI Filters

High Speed Characterization Report

74ACQ241 Octal Buffer/Line Driver with 3-STATE Outputs

EE 210 Lab Exercise #3 Introduction to PSPICE

Introduction to ANSYS SIwave

B. Equipment. Advanced Lab

User Manual TDS 694C Digital Real-Time Oscilloscope

Appendix A: Specifications

User s Guide. DDS-3005 USB Operation Manual

Lab Reference Manual. ECEN 326 Electronic Circuits. Texas A&M University Department of Electrical and Computer Engineering

Get started with DSO-X 2014A

Instruction Manual. P7380SMA 8 GHz Differential Probe

Getting Started. MSO/DPO Series Oscilloscopes. Basic Concepts

Function Generator Guide Tektronix AFG3102

High Speed Characterization Report

DS 6000 Specifications

EE 1210 Op Amps, Gain, and Signal Integrity Laboratory Project 6

Specification. CTR 2 ESD calibration target

P7313SMA 13 GHz Differential Probe

1000BASE-T1 EMC Test Specification for Common Mode Chokes

Technical Reference. TDS 500D, TDS 600C, TDS 700D & TDS 714L Digitizing Oscilloscopes Performance Verification and Specifications

High Speed Characterization Report

Hantek2000 Series 3u1 Oscilloscope+DMM+AF generator

74ACTQ821 Quiet Series 10-Bit D-Type Flip-Flop with 3-STATE Outputs

PI3HDMI231-A/B PI3HDMI231-A/B Demo Board Rev.B User Manual by Ada Yip

PHY152 Experiment 4: Oscillations in the RC-Circuits (Measurements with an oscilloscope)

Exercise 2-1. PAM Signals EXERCISE OBJECTIVE DISCUSSION OUTLINE. Signal sampling DISCUSSION

MDO3000 Series Mixed Domain Oscilloscopes

Exercise 2. The Buck Chopper EXERCISE OBJECTIVE DISCUSSION OUTLINE. The buck chopper DISCUSSION

Transcription:

Procedures Guide Tektronix HDMI Sink Instruments Differential Impedance Measurement

Equipment Required Table 1: Equipment required Table 1 lists the equipment required to perform the differential impedance measurement. Item Qty. Recommended equipment Sampling oscilloscope 1 ea. Tektronix TDS8200 (or DSA8200/TDS8000B) TDR sampling module 1 ea. Tektronix 80E04 50 Ω SMA terminator 6 ea. Tektronix part number 015-1022-01 50 Ω SMA cable 2 ea. HUBER+SUHNER SUCOFLEX 104PEA TPA (test fixture) 1 ea. Efficere Technologies EFF-HDMI-TPA-P NOTE. To protect the sampling module from damage due to electro-overstress (EOS) and electrostatic discharge (ESD), power off a device under test (DUT) to discharge the static voltage completely from it before performing the procedures. While performing the following procedures, be sure to wear a grounded antistatic wrist strap to discharge the static voltage from your body. 1

Front-Panel Buttons Figure 1 shows the front-panel buttons on the TDS8200 used in the procedures. DEFAULT SETUP button CURSORS button VERTICAL MATH button ACQUISITION MENU button VERTICAL MENU button HORIZONTAL MENU button VIEW Mag1 button VIEW Mag2 button Figure 1: Front-panel buttons on the TDS8200 Equipment Connections 1. Use the 50 Ω SMA cable to connect the CH1 connector on the sampling module and the DATA0_P connector on the test fixture (see Figure 2). 2. Use the 50 Ω SMA cable to connect the CH2 connector on the sampling module and the DATA0_N connector on the test fixture (see Figure 2). 3. Use the 50 Ω SMA terminators to terminate the unused connectors on the test fixture (see Figure 2). 2

Sampling oscilloscope (TDS8200 and 80E04) 50 Ω SMA terminators Test fixture 50 Ω SMA cables Figure 2: Initial equipment connections for the differential impedance measurement Calibration 1. Output TDR step pulses from the sampling oscilloscope, and then set the oscilloscope so that both the TDR step pulses and waveforms at the open ends of the test fixture are displayed on the screen: a. Press the DEFAULT SETUP button, and then click Yes. b. Press the ACQUISITION MENU button to display the Setups window. c. Click the C1-C2 Diff button on the TDR tab, and then set the ACQ On Units to Ω. d. On the Vert tab, select C1 from the Waveform field, and then set the SetupScale to 10.00 Ω/div and Position to -5.000 div. e. On the Vert tab, select C2 from the Waveform field, and then set the SetupScale to 10.00 Ω/div and Position to -5.000 div. 2. Use the sampling oscilloscope Mag View feature to magnify the TDR step pulses and waveforms at the open ends of the test fixture in other windows: a. Press the VIEW Mag1 button, and then use the HORIZONTAL POSITION knob and HORIZONTAL SCALE knob to display the TDR step pulses properly (see Figure 3 on page 4). 3

b. Press the VIEW Mag2 button, and then use the HORIZONTAL POSITION knob and HORIZONTAL SCALE knob to display the waveforms at the open ends of the test fixture properly (see Figure 3). 3. Press the HORIZONTAL MENU button. Figure 3: Magnifying the TDR step pulses and the waveforms at the open ends of the test fixture 4. Set the sampling oscilloscope so that the delay between the CH1 and CH2 can be measured in the magnified window: a. Press the VIEW Mag1 button, and then click the Meas tab. b. Click the Select Meas button, and then select Pulse-Timing > Delay. c. Click Source1, and then select Mag1 C1 on the Source tab. d. Click the Region tab, and then set Edge Slope to + and Direction to -->. e. Click Source2, and then select Mag1 C2 on the Source tab. f. Click the Region tab, and then set Edge Slope to + and Direction to -->. g. Press the VIEW Mag2 button, and then click the Meas tab. h. Select Meas2, and then click the Select Meas button and select Pulse-Timing > Delay. 4

i. Click Source1, and then select Mag2 C1 on the Source tab. j. Click the Region tab, and then set Edge Slope to + and Direction to -->. k. Click the RefLevel tab, and then select Absolute. Set Mid to 75 Ω. l. Click Source2, and then select Mag2 C2 on the Source tab. m. Click the Region tab, and then set Edge Slope to + and Direction to -->. n. Click the RefLevel tab, and then select Absolute. Set Mid to 75 Ω. 5. Adjust step deskew so that the delay between the TDR step pulses is equal to the delay between waveforms at the open ends of the test fixture: a. Click the TDR tab, and then select C1_C2 in the Step Deskew drop-down list box. b. Adjust the readings of and in Figure 4 are about the same value by changing the Step Deskew value (see Figure 5 on page 6). Figure 4: Adjusting the step deskew 5

Figure 5: Changing the Step Deskew value 6. Adjust deskew so that the delay between the waveforms at the open ends of the test fixture has the minimum value: a. Click the Vert tab, and then select Deskew. b. Adjust the deskew so that the reading of in Figure 4 on page 5 has the minimum value by changing the Deskew value (see in Figure 6). 6

Figure 6: Changing the Deskew value 7. Turn off the Mag View feature. Setting the Rise Time 1. Click the TDR tab, and then set ACQ On Units to V. 2. Define a math waveform: a. Press the VERTICAL MATH button, and then enter Filter (CH1-CH2) in the Math Expression field in the Define Math window. b. Select the Math Waveform On check box, and then click OK. c. Press the VERTICAL MENU button (if the Setups window is not displayed), and then click the Vert tab and select M1. d. Set Scale to 100.0 mv/div and Position to -5.00 div. 3. Use the HORIZONTAL SCALE knob to set the horizontal scale to 200 ps/div. 7

4. Use the HORIZONTAL POSITION knob to display the math waveform at the center of the screen. 5. Set the sampling oscilloscope to measure the rise time of the waveform: a. Press the VERTICAL MENU button (if the Setups window is not displayed), and then click the Meas tab. b. Select Meas3, and then click the Select Meas button and select Pulse-Timing > Rise Time. c. Click Source1, and then select Main M1 on the Source tab. 6. Press the VERTICAL MATH button to display the Define Math window. 7. Adjust the rise time so that the reading of in Figure 7 is 200 ps by changing the Filter Risetime value (see in Figure 8). If you cannot adjust the value to 200 ps exactly, set it to the nearest value below 200 ps. Figure 7: Adjusting the rise time 8

Figure 8: Changing the Filter Risetime value Measuring the Impedance 1. Clear the C1 and C2 waveforms. 2. Press the ACQUISITION MENU button, and then click the TDR tab. 3. Set ACQ On Units to Ω. 4. Change the math waveform: a. Press the VERTICAL MATH button to display the Define Math window, and then click Clear. b. Enter Filter (CH1 +CH2) in the Math Expression field. 5. Press the VERTICAL MENU button. 6. Click the Vert tab, and then set M1 Scale to 10.00 Ω/div and M1 Position to -10.00 div. 7. Set the vertical scale to 100 ps/div. 8. Adjust the HORIZONTAL POSITION knob so that the rise-time waveform is displayed at near 1 major division from the left side of the screen. 9. Connect the test fixture to the HDMI connector on a device under test (DUT) (see Figure 9). 9

Sampling oscilloscope (TDS8200 and 80E04) 50 Ω SMA terminators Sync DUT Test fixture 50 Ω SMA cables Figure 9: Connecting a sync DUT to the test fixture 10. Display the horizontal cursors and align them to the 85 Ω and 115 Ω points on the impedance waveform: a. Press the CURSORS button twice to display the horizontal cursors. b. Use the multipurpose knob to adjust the reading value of v1 at the right side of the screen to 115 Ω. c. Press the SELECT button to toggle selection between the two cursors. d. Use the multipurpose knob to adjust the reading value of v1 at the right side of the screen to 85 Ω. 11. Use the automatic measurement feature to measure the Max and Min values of the impedance waveform. At this time, use the gating feature to specify the measurement area. a. Select Pulse and Amplitude from the pull-down menus in the measurement tool bar. b. Click Max and Min on the tool bar to specify the measurement items. c. Right-click Max in the measurement area at the left side of the screen to display the Setup dialog box for setting Max. d. On the Region tab, set the G1 (Gate1) to 0 and G2 (Gate2) to 50, and then select the On check box. e. Right-click Min in the measurement area at the left side of the screen to display the Setup dialog box for setting Min. 10

f. On the Region tab, set the G1 to 0 and G2 to 50, and then select the On check box. g. When you select Max or Min in the Measurement area at the left side of the screen, each gate position is indicated by the vertical cursor. h. Adjust the gates so that the measurement area is between the gates using a mouse. 12. Measure the impedance of the waveform: When the Sink_Diff_PowerOn field of the CDF is Y: Measure the impedance between the HDMI connector on the DUT and the receive IC pin. The measurement period (time width) is defined in the Sink_Term_Distance field of the CDF. If the measurement value is less than and equal to 75 Ω or 125 Ω, the differential lines being measured are Fail. If an area that has the value less than and equal to 85 Ω or more than and equal to 115 Ω lasts for more than and equal to 250 ps, the differential lines being measured are Fail (even if an area that has the any of the above values lasts for less than and equal to 250 ps, if the area appears twice, the differential lines being measured are Fail). If any of the above cases do not apply, the differential lines being measured are Pass. When the Sink_Diff_PowerOn field of the CDF is N: Measure the impedance between the HDMI connector on the DUT and the point that the impedance settles at the terminal impedance. If the measurement value is less than and equal to 75 Ω or 125 Ω, the differential lines being measured are Fail. If an area that has the value less than and equal to 85 Ω or more than and equal to 115 Ω lasts for more than and equal to 250 ps, the differential lines being measured are Fail (even if an area that has the any of the above values lasts for less than and equal to 250 ps, if the area appears twice, the differential lines being measured are Fail). If the terminal impedance is less than and equal to 75 Ω or 125 Ω, the differential lines being measured are Fail. If any of the above cases do not apply, the differential lines being measured are Pass. 13. Repeat Step 12 for all the TMDS lines. End of Procedures 11

12